GW Instek The firmware upgrade guide for GDS-3000 Mode d'emploi

Taper
Mode d'emploi
Digital Storage Oscilloscope
GDS-1000B / GDS-2000E/ GDS-2000A/ GDS-3000 Series
FIRMWARE AND KERNEL UPGRADE GUIDE
GW INSTEK PART NO. N/A
ISO-9001 CERTIFIED MANUFACTURER
This guide contains proprietary information, which is protected by
copyright. All rights are reserved. No part of this guide may be
photocopied, reproduced or translated to another language without
prior written consent of Good Will company.
The information in this guide was correct at the time of printing.
However, Good Will continues to improve products and reserves the
right to change specification, equipment, and maintenance
procedures at any time without notice.
Good Will Instrument Co., Ltd.
No. 7-1, Jhongsing Rd., Tucheng District, New Taipei City 236, Taiwan.
UPGRADE PROCEDURE
3
UPGRADE PROCEDURE
This firmware upgrade guide describes how to upgrade both the
DSO firmware and the operating system kernel.
Upgrade firmware
1. Insert a USB flash disk containing the firmware
file, xxx.upg into the front panel USB slot.
2. Remove all probes and cables from the BNC
terminals.
Utility
VARIABLE
Up Down
Select
GDS-1000B/2000E/2000A/3000 Series Upgrade Guide
4
5. Rotate the Variable knob to select
the xxx.upg file and then press
the Select key. Press the Select key
again to begin the upgrade
procedure.
VARIABLE
Up Down
Select
6. When the procedure has completed restart the
scope.
Note
For the GDS-3000 models:
Steps 7 & 8 only apply to 5GSa/s models (GDS-
3154, GDS-3254, GDS-3352/4).
2.5GSa/s models (GDS-3152 & GDS-3252) will
bypass these two steps.
Upgrade Kernel
7. When the screen goes “blank”
during the booting process, press
the lit CH1 key three times. (For
GDS-3154, GDS-3254, GDS-
3352/4 , 5GSa/s models only)
CH1 CH1
à
X3
UPGRADE PROCEDURE
5
8. The scope will return to the main screen after
the CH1 key has been pressed. (For GDS-3154,
GDS-3254, GDS-3352/4 , 5GSa/s models only)
9. Repeat steps 3~6 to upgrade the kernel. The
same file is used again to upgrade the kernel.
10. When the calibration message appears for the
second time after the oscilloscope has restarted,
press TEST in order to perform the phase
calibration officially. (For GDS-3154, GDS-3254,
GDS-3352/4 , 5GSa/s models only)
Please note that the entire phase calibration may
take about 10 minutes.
11. The upgrade procedure is complete after the
second upgrade.
GDS-1000B/2000E/2000A/3000 Series Upgrade Guide
6
Note
The following note is for the GDS-3000
only.
If the existing firmware version is earlier
than v1.07, please repeat this firmware
upgrade procedure twice.
For firmware versions earlier than v1.07
(v1.00~v1.06), please upgrade both the DSO
firmware and the OS kernel (follow the operation
steps 1 to 10). This means that you need to do the
same upgrade procedure twice .
If upgrading the firmware from V1.07 to v1.08 or
later, only update the firmware. The kernel
upgrade is not required (follow operation steps 1
to 6).
If the upgrade procedure crashes and freezes
during the upgrade, please use the safe mode
upgrade method instead. Please see the next
chapter on details on how to use the “safe mode
upgrade method.
SAFE MODE UPGRADE PROCEDURE
7
SAFE MODE UPGRADE
PROCEDURE
Note
If the upgrade procedure crashes and freezes
during the upgrade, you use the safe mode
upgrade method instead.
The upgrade file name for this method are either
xxx.opg(for GDS-3000 150-350 ) or xxx.upg
(for GDS-1000B, 2000A, 2000E, GDS-3000 500 )
Upgrade firmware
by safe mode
1. Remove all the files on the USB flash disk.
2. Copy the firmware file xxx.opg or xxx.upg
into the USB flash disk (root directory only).
Insert the USB flash disk into the front panel
USB slot.
M Hz
M Hz
M Hz
GDS-1000B/2000E/2000A/3000 Series Upgrade Guide
8
3. Power up the oscilloscope and at the same time,
keep rotating the VARIABLEknob a number of
times until the oscilloscope boots into the
firmware upgrade safe mode, as shown in the
snapshot below.
4. When the xxx.opg or xxx.upg file on USB
flash disk has been recognized by oscilloscope, the
message usb device ready will appear on the
lower left corner.
5. Press the F1 key. The oscilloscope will
automatically start upgrading the firmware.
When the status indicator at the center of the
display has completed, the upgrade grade
procedure (using the safe mode) is complete.
ongoing
complete
SAFE MODE UPGRADE PROCEDURE
9
6. The oscilloscope will restart automatically.
After the scope has restarted, check the
firmware version by pressing Utilityà
Systemà System Info. The system
information screen should now show the new
firmware version that was updated.
For more information, contact your local dealer or GWInstek at
www.gwinstek.com / marketing@goodwill.com.tw.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9

GW Instek The firmware upgrade guide for GDS-3000 Mode d'emploi

Taper
Mode d'emploi