ENUSER MANUAL XTA
16
F11 rS232
ModE
The scale RS232 protocol will be determined by the Baud & Pr settings
oFF RS232 function disable
Cont Continues data transmission
StC Stable data transmission (sending continue frame)
toLE do Protocol type 1: Toledo
nC iE Cr Protocol type 2: NCI-ECR
nC iG En Protocol type 3: NCI-General
tEC Protocol type 4: TEC
EASY Protocol type 5: Easy Weigh
tpv0 trace
7PV-a
Manual
Protocol type 6 : TPV0
Command
Contin
7PV-b
Manual
Command
Contin
7PV 2A Protocol type 7 : TPV 2A
7PV 2b Protocol type 8 : TPV 2B
7PV 4 Protocol type 9 : TPV 4
7PV 7 Protocol type 10 : TPV 7
7EdC 1 Protocol type 11 : TED C1
tisa A Protocol type 12: TISA A
tisa B Protocol type 13 : TISA B
tisa v Protocol type 14 : BD TISA
bAUd 1200 - 2400 - 4800 - 9600 - 9200 - 38 400 - 115200 Transmission speed
Pr
8n 1 8 data bits, no verify, 1 stop bit
7E1 7 data bits, Even verify, 1 stop bit
7o1 7 data bits, Odd verify, 1 stop bit
F12
trAnSM
tYPE
Model
Comman Activate comman mode
Manual Activate manual mode
8. CALIBRATION
Turn on the scale. Keep pressed key during the self test.
Display will show “Pn” briefly PIN is activated. Press the PIN number to get into the parameter menu.
• The default PIN number is “0000”
• Press numeric key 0 four times, display will be shown Pn - - - -
• Press key to confirm, it will enter into the parameter settings and will to show “
F0 CAL
”.