ResMed Mirage Quattro Manuel utilisateur

Taper
Manuel utilisateur

Ce manuel convient également à

Global leaders in sleep and respiratory medicine www.myresmed.com
Mirage Quattro™
FULL FACE MASK
User Guide
English Français • Español • Português
618321/1
2020-05
MIRAGE QUATTRO
USER
AMER ML
618321
618297 cover.indd 1618297 cover.indd 1 28/05/2020 4:21:28 PM28/05/2020 4:21:28 PM
Manufacturer:
ResMed Pty Ltd 1 Elizabeth Macarthur Drive Bella Vista NSW 2153 Australia
See www.resmed.com for other ResMed locations worldwide.
For patent information, see www.resmed.com/ip.
Mirage and Mirage Quattro are trademarks of ResMed and Mirage is registered in U.S. Patent and
Trademark Office. Velcro is a registered trademark of Velcro Industries B.V.
© 2020 ResMed.
618297 cover.indd 2618297 cover.indd 2 28/05/2020 4:21:28 PM28/05/2020 4:21:28 PM
A
Mask parts / Composants du masque / Piezas de la mascarilla /
Partes da máscara
Forehead support / Support frontal / Apoyo para la frente /
Suporte para a testa
Forehead support pad / Tampon frontal /
Almohadilla para la frente / Almofada do suporte
para a testa
Mask cushion / Bulle du
masque / Almohadilla de
la mascarilla / Almofada da
máscara
Mask frame / Entourage rigide / Armazón de
la mascarilla / Armação da máscara
Ports cap / Bouchon d’entrées / Tapón de
los puertos / Tampa das portas
Elbow assembly /
Coude / Unidad del
codo / Conjunto
do cotovelo
Vent / Orifices de
ventilation / Ventilación /
Respiradouro
Dial / Molette de
réglage / Disco / Botão
Valve clip / Clip de la valve / Traba de la válvula /
Presilha da válvula
Valve hinge / Charnière de la valve /
Bisagra de la válvula / Articulação da
válvula
Valve membrane / Membrane de
la valve / Membrana de la válvula /
Membrana da válvula
Valve / Valve / Válvula
/ Válvula
Elbow / Coude / Codo /
Cotovelo
Swivel / Pièce pivotante / Pieza
giratoria / Peça giratória
Upper straps / Sangles supérieures /
Correas superiores / Tiras superiores
Lower straps / Sangles inférieures
/ Correas inferiores / Tiras
inferiores
Headgear clip / Clip
d’attache du harnais / Broche
del arnés para la cabeza /
Presilha do arnês
B
Fitting the mask / Mise en place du masque / Colocación de la mascarilla / Ajuste da máscara
3
6
9
12
15
18
21
24
1
2
3
4
5
6
7
8
C
Disassembling for cleaning / Démontage pour le nettoyage / Desmontaje para la limpieza / Desmontagem para limpeza
1
2
1
7
2 3 4 5 6
8 9 10 11
© 2020 ResMed.
© 2020 ResMed.
© 2020 ResMed.
Elbow assembly / Ensemble coude / Unidad del codo /
Conjunto do cotovelo
Cushion clip / Clip de la bulle / Traba de
la almohadilla / Presilha da
almofada
618297 illo sheets.indd 3618297 illo sheets.indd 3 28/05/2020 4:16:46 PM28/05/2020 4:16:46 PM
D
Reassembling the mask / Remontage du masque / Para volver a montar la mascarilla / Remontando a máscara
Elbow Reassembly / Remontage du coude / Montaje del codo / Remontagem do cotovelo
Mask Reassembly / Remontage du masque / Montaje de mascarilla / Remontagem da máscara
1 2 3 4 5
1
2
1
2
6 7 8 9
3
6
9
12
15
18
21
24
10 11 12 13
Mirage Quattro
Full Face Mask / Masque facial / Mascarilla Facial / Máscara Facial
Illustrations Sheet / Fiche d’illustrations /
Hoja de figuras / Folheto de ilustrações
Note: For complete instructions, please read the text section in
conjunction with the illustrations on this sheet.
Remarque : pour les instructions complètes, veuillez lire le texte du
guide avec cette fiche d’illustrations.
Nota: para tener instrucciones completas, lea el texto en conjunto
con las figuras de esta hoja.
Nota: para obter instruções completas, leia a seção de texto
enquanto consulta as ilustrações neste folheto.
Component of: 618321/1
© 2020 ResMed.
618297 illo sheets.indd 4618297 illo sheets.indd 4 28/05/2020 4:16:48 PM28/05/2020 4:16:48 PM
English
1
Mirage Quattro™
Full Face Mask
Thank you for choosing ResMed’s Mirage Quattro Full Face Mask.
The Mirage Quattro is a full face mask system that covers your nose and
mouth.
This means that you can receive effective therapy even if you breathe
through your mouth.
The mask incorporates vent holes and a built-in valve so that you can continue
to breathe fresh air if the airflow to your mask is impeded for any reason.
Intended Use
The Mirage Quattro channels airflow noninvasively to a patient from a positive
airway pressure device such as a continuous positive airway pressure (CPAP)
or bilevel system.
The Mirage Quattro is to be used by adult patients (>66 lb / 30 kg) for whom
positive airway pressure has been prescribed.
The Mirage Quattro is intended for single-patient re-use in the home
en
vironment and multipatient re-use in the hospital/institutional environment.
CAUTION
In the USA, federal law restricts this device to sale by or on the order of a
physician.
Before Using the Mask
WARNINGS
The Mirage Quattro must be used under qualified supervision with
patients who are unable to remove the mask by themselves. The mask
may not be suitable for those predisposed to aspiration.
The elbow and valve assembly are intended to be used with the
Mirage Quattro and have specific safety functions. The mask should
not be worn if the valve is damaged as it will not be able to perform its
618297 Eng.fm Page 1 Thursday, April 12, 2018 4:11 PM
2
safety function. The valve should be replaced if it is damaged,
distorted or torn.
The vent holes must be kept clear.
Follow all precautions when using supplemental oxygen.
Any oxygen flow must be turned off when the flow generator is not
operating, so that unused oxygen does not accumulate within the flow
generator enclosure and create a risk of fire.
At a fixed flow rate of supplemental oxygen flow, the inhaled oxygen
concentration will vary, depending on the pressure settings, patient
breathing pattern, mask, point of application and leak rate.
The technical specifications of the mask and flow generator are
provided for your clinician to check that they are compatible. If used
outside specification or if used with incompatible devices, the seal and
comfort of the mask may not be effective, optimum therapy may not
be achieved, and leak, or variation in the rate of leak, may affect the
flow generator function.
Discontinue using the Mirage Quattro if you have ANY adverse
reaction to any part of the mask system, and consult your physician or
sleep therapist.
Refer to your flow generator manual for details on settings and
operational information.
Using a mask may cause tooth, gum or jaw soreness or aggravate an
existing dental condition. If symptoms occur, consult your physician or
dentist.
CAUTION
Avoid connecting flexible PVC products (eg, PVC tubing) directly to any
part of this mask. Flexible PVC contains elements that can be detrimental
to the materials of the mask, and may cause the components to crack or
break.
Using the Mask
Select ‘Mir Full’ or ‘Full Face’ as the mask option when using the Mirage
Quattro with ResMed flow generators that have mask setting options.
Fitting the Mask
The diagrams in section A on the illustrations sheet show how the parts of the
mask fit together.
The diagrams in section B sho
w the sequence for obtaining the best fit and
comfort in using your mask. Please follow these instructions carefully.
618297 Eng.fm Page 2 Thursday, April 12, 2018 4:11 PM
3
English
1 Twist the forehead support dial to Position 24, where the forehead support
is fully extended (Fig. B-1).
2 Hold the mask steadily on your face (Fig. B-2) and pull the headgear over
your head.
3 Bringing the lower straps below your ears, clip the headgear to the lower
part of the mask frame (Fig. B-3).
4 Unfasten the Velcro
®
, pull the upper headgear straps evenly until they are
comfortable and firm, and reattach the Velcro (Fig. B-4).
5 Repeat step 4 with the lower headgear straps (Fig. B-5).
6 Test the functioning of the valve by sealing the swivel with your hand and
ensuring that you can still breathe freely.
7 Connect the main air tubing to the swivel (Fig. B-6). Then connect the free
end of the air tubing to your flow generator and turn the flow generator on.
8 Lie down in your sleeping position. Twist the forehead support dial
(Fig. B-7) until you feel that the mask is evenly positioned on your face. It
should be comfortable, with a satisfactory seal. Gently tighten the lower
headgear straps if necessary.
9 The fitted mask is shown in Fig. B-8.
Optimizing Mask Seal and Comfort
To receive effective therapy it is important to have the best seal possible
between your mask and your face. If you are experiencing leaks, please ensure
that you have followed the fitting instructions carefully. Carry out a quick check
after fitting your mask:
Is the mask cushion creased? Creases can allow air to leak out of the mask.
Have you cleaned your mask cushion recently? Oils on the cushion can
affect the seal.
Are your headgear straps overtightened? Try not to alter the upper headgear
straps once they have been tightened by your clinician.
You may need to slightly adjust the dial. Note the dial setting when
comfortable.
Does the mask seem too close to your eyes? You may need a smaller mask.
If you cannot resolve any problems, look at the Troubleshooting section, or talk
to your clinician. You may require another mask size.
Removing the Mask
To remove the Mirage Quattro, disconnect one of the lower headgear clips and
pull the mask and headgear up over your head.
618297 Eng.fm Page 3 Thursday, April 12, 2018 4:11 PM
4
Disassembling the Mask
See section C on the illustrations sheet.
1 Det
ach the air tubing.
2 Unclip the lo
wer headgear clips (Fig. C-1) and unhook the upper headgear
straps (Fig. C-2). It is not necessary to remove the headgear clips from the
headgear.
3 Remove the elbow assembly from the mask frame (Fig. C-3). It’s easier to
pivot the bottom of the elbow away from the mask frame than to pull the
elbow straight off.
4 Remove the valve clip from the elbow (Fig. C-4).
5 P
eel the valve away from the valve clip (Fig. C-5). Do not pull on the
membrane to remove the valve.
6 Det
ach the swivel from the elbow (Fig. C-6).
7 Note the position on the dial f
or the next time you use the mask.
8 F
ully unscrew and remove the dial then pull the forehead support out of the
mask frame (Fig. C-7).
9 R
emove the forehead support pads from the forehead support (Fig. C-8).
10 Squeez
e and push the upper side tabs on the cushion clip out of the mask
frame. Pivot away from the mask, and gently pull out (Fig. C-9).
11 Separate the cushion from the cushion clip (Fig. C-1
0).
12 R
emove the ports cap from the mask frame (Fig. C-11).
Cleaning the Mask in the Home
Notes:
The mask and headgear can only be cleaned by handwashing.
The headgear may be washed without removing the headgear clips.
WARNING
Do not use aromatic-based solutions or scented oils (eg, eucalyptus or
essential oils), bleach, alcohol or products that smell strongly (eg,
citrus) to clean any of the mask components. Residual vapours from
these solutions can be inhaled if not rinsed thoroughly. They may also
damage the mask, causing cracks.
Do not iron the headgear as the material is heat sensitive and will be
damaged.
Daily/After Each Use
To optimize the mask seal, facial oils should be removed from the cushion
before use.
618297 Eng.fm Page 4 Thursday, April 12, 2018 4:11 PM
5
English
Handwash the mask by gently rubbing in warm (86°F / 30°C) water using mild
soap or diluted dishwashing detergent.
Do not twist the valve during cleaning as this may damage or tear the valve
membrane or hinge. Take care when handling the valve.
If the vent requires cleaning use a soft bristle brush.
Rinse all components well with drinking quality water and allow them to air dry
out of direct sunlight. Do not soak the mask.
Weekly
Handwash the headgear in warm (approx. 86°F/30°C) water using mild soap or
diluted dishwashing detergent. Rinse well and allow it to air dry out of direct
sunlight.
Reprocessing the Mask between Patients
Only Sleep Lab Mask (SLM) variants of the mask are intended for multi-patient
re-use. When using between patients, these masks must be reprocessed
according to cleaning and disinfection instructions available on ResMed.com/
downloads/masks.
Reassembling the Mask
See section D on the illustrations sheet.
Before reassembling the mask, check that all components are clean and dry.
Check that the valve is not damaged, distorted or torn, and that none of the
other components are broken or cracked. Replace any parts that are damaged.
Elbow Reassembly
1 Insert the valve into the valve clip (Fig. D-1).
2 Push the valve clip onto the elbow (Fig. D-2). Make sure both sides click into
place (Fig. D-3).
3 Insert the swivel into the end of the elbow (Fig. D-4).
4 Insert the assembled elbow (Fig. D-5) into the mask frame.
Mask Reassembly
1 Fit the cushion clip into the grooved edge of the mask cushion. Fit the top
section first, then work your way around, pushing the clip into the cushion
(Fig. D-6). Ensure that the cushion is not twisted or distorted around the
clip.
2 Align the cushion clip with the mask frame. Push the cushion clip into the
mask frame, ensuring that all three clips click into place (Fig. D-7).
3 Push the forehead pads firmly into the forehead support (Fig. D-8).
618297 Eng.fm Page 5 Thursday, April 12, 2018 4:11 PM
6
4 Insert the forehead support into the mask frame and screw the dial
clockwise onto the pole (Fig. D-9). Turn the dial until it clicks (Fig. D-10).
Ensure the dial sits flush against the frame.
5 Insert the ports cap (Fig. D-11).
6 R
eattach the top headgear by threading it through the loop (Fig. D-12).
7 T
urn the dial until you reach the setting that gives you good comfort and seal
(Fig. D-13).
Storage
Ensure that the mask is thoroughly clean and dry before storing it for any
length of time. Store the mask in a cool dry place out of direct sunlight.
Disposal
The Mirage Quattro does not contain any hazardous substances and may be
disposed of with your normal household refuse.
Troubleshooting
Problem Possible reason Possible solution
Mask won’t seal properly
or is uncomfortable
Mask cushion may be dirty. Clean cushion according to
instructions in “Cleaning the
Mask in the Home”.
Mask may have been fitted
incor
rectly or is is incorrectly
adjusted.
Carefully follow instructions in
“Fitting the Mask”. Check
headgear and dial are not over-
tightened. Extend dial to apply
more force to chin. Contract dial
to increase force at nose bridge.
Mask cushion is skewed in
cushion clip.
Check insertion of cushion and
reinsert cor
rectly.
Mask cushion is misaligned or
creased on f
ace.
Ensure cushion is properly
aligned and that there are no
creases in it: undo headgear clip,
check cushion and reposition the
mask on your face.
Mask is wrong size. Talk to your clinician.
Mask is too noisy Vents are blocked or partially
bloc
ked.
Clean vents according to
instructions in “Cleaning the
Mask in the Home”.
Valve is misassembled. Reassemble valve.
Cushion is not fully clipped in. Check insertion of cushion.
Mask is assembled incorrectly. Reassemble mask.
Valve is torn or distorted and is
not sealing
.
Replace valve.
Dial inserted into mask
frame before forehead support
Misassembly. Push dial base onto a hard
surf
ace until the dial comes out
of the frame.
618297 Eng.fm Page 6 Thursday, April 12, 2018 4:11 PM
7
English
Technical Specifications
Note: The mask system does not contain latex, PVC or DEHP materials.
Pressure–Flow Curve
Vent Flow Rate (L/min)
Mask Pressure (cm H
2
O)
Pressure
(cm H
2
O)
Flow
(L/min)
422
832
12 41
16 48
20 54
24 60
28 66
32 72
36 77
40 82
Flow (L/min)
Note: The mask contains passive venting that meets the requirements for
protection against rebreathing as specified in ISO 17510-2. As a result of
manufacturing variations, the vent flow rate may vary.
Dead Space Information
Dead space: the empty volume of the mask up to the swivel.
The dead space of the mask varies according to cushion size. It is 242 mL for the large
size.
Standards
CE designation in accordance with EC directive 93/42/EEC, class IIa.
ISO 14971, ISO 13485, ISO 10993-1 and ISO 17510-2.
Therapy Pressure
4 to 40 cm H
2
O
Inspiratory and expiratory resistance
with Anti Asphyxia Valve (AAV) open to atmosphere.
Inspiration at 50 L/min 0.8 cm H
2
O
Expiration at 50 L/min 0.8 cm H
2
O
Resistance
Drop in pressure measured (nominal)
at 50 L/min: 0.1 cm H
2
O
at 100 L/min: 0.4 cm H
2
O
Environmental Conditions
Operating temperature: +5°C to +40°C
Operating humidity: 15%–95% relative humidity non-condensing
Storage and transport: -20°C to +60°C
Storage and transport humidity: 10%–95% relative humidity non-condensing
618297 Eng.fm Page 7 Thursday, April 12, 2018 4:11 PM
8
Mask sizes
(Mask fully assembled with elbow assembly – no headgear)
size height (mm) width (mm) depth (mm)
extra small 176 109 108
small 186 109 108
medium 196 109 108
large 206 109 108
Note: The manufacturer reserves the right to change these specifications
without notice.
Symbols
System and Packaging
Caution, consult accompanying documents.
LOT
Batch Code
REF
Catalogue Number
European Authorised Representative
Does not contain natural rubber latex
Humidity limitation
Temperature limitation
Manufacturer
User Guide
Indicates a warning or caution and alerts you to a possible injury or explains
special measures for the safe and effective use of the device.
618297 Eng.fm Page 8 Thursday, April 12, 2018 4:11 PM
9
English
Limited Warranty
ResMed warrants that your ResMed mask system, including its components:
mask frame, cushion, headgear, tubing and other accessories (“Product”) shall
be free from defects in material and workmanship for a period of ninety (90)
days from the date of purchase by the initial consumer.
For further details refer to the Warranty.
618297 Eng.fm Page 9 Thursday, April 12, 2018 4:11 PM
Français
11
Mirage Quattro™
Masque facial
Merci d'avoir choisi le masque facial Mirage Quattro de ResMed.
Le Mirage Quattro est un masque facial qui couvre à la fois votre nez et votre
bouc
he. Cela signifie que même si vous respirez par la bouche, vous pouvez
recevoir un traitement efficace.
Le masque comprend des orifices de ventilation et une valve qui vous
permettent de continuer à respirer de l'air frais en cas de blocage de l'arrivée
d'air à votre masque.
Usage prévu
Le Mirage Quattro achemine au patient de façon non-invasive le débit d'air
produit par un appareil à pression positive tel qu'un appareil de PPC (CPAP) ou
d'aide inspiratoire avec PEP.
Le Mirage Quattro est prévu pour une utilisation par des patients adultes
(> 66 lb / 30 kg) à qui une pression positive a été prescrite.
Le Mirage Quattro est prévu pour un usage multiple par un seul patient à
domicile ou pour un usage multiple par plusieurs patients en milieu hospit
alier.
ATTENTION
Selon la loi fédérale américaine, ce dispositif ne peut être vendu aux
U.S.A. que par un médecin ou sur prescription médicale.
Avant d'utiliser le masque
AVERTISSEMENTS
Le Mirage Quattro doit être utilisé sous le contrôle d’une personne
qualifiée lorsque le patient n’est pas en mesure d’enlever le masque
de lui-même. Le masque peut ne pas convenir aux patients sujets aux
vomissements.
618297 Fre.fm Page 11 Thursday, April 12, 2018 4:13 PM
12
L'ensemble coude et valve est prévu pour une utilisation avec le
Mirage Quattro et possède des caractéristiques de sécurité
spécifiques. Le masque ne doit pas être porté si la valve est
endommagée, auquel cas le masque n'est pas en mesure de remplir
sa fonction de sécurité. La valve doit être remplacée si elle est
endommagée, déformée ou déchirée.
Les orifices de ventilation ne doivent jamais être obstrués.
Prendre toutes les précautions applicables lors de l'utilisation
d'oxygène additionnel.
L'arrivée d'oxygène doit être fermée lorsque l'appareil n'est pas en
marche afin d'empêcher l'oxygène inutilisé de s'accumuler dans le
boîtier de l'appareil et de créer un risque d'incendie.
Lorsque de l’oxygène additionnel est délivré à un débit fixe, la
concentration de l’oxygène inhalé varie en fonction des réglages de
pression, de la respiration du patient, du masque, du point d'arrivée de
l'oxygène ainsi que du niveau de fuites.
Les caractéristiques de votre masque et de votre appareil sont
fournies afin que votre clinicien puisse vérifier qu'ils sont compatibles.
L'utilisation du masque en dehors de ses caractéristiques ou avec des
appareils incompatibles peut compromettre son étanchéité et son
confort ainsi que l'efficacité du traitement, ou la variation du niveau de
fuites peut affecter le fonctionnement de l'appareil.
Cesser d'utiliser le Mirage Quattro en cas d'une quelconque réaction
indésirable à l'un des composants du masque et consulter votre
médecin ou votre spécialiste du sommeil.
Voir le manuel de l'appareil pour de plus amples informations sur les
réglages et le fonctionnement.
L’utilisation d’un masque peut provoquer des douleurs dans les dents,
les gencives ou la mâchoire ou aggraver un problème dentaire
existant. En cas de symptômes, consulter un médecin ou un dentiste.
ATTENTION
Éviter de raccorder des produits en PVC souple (par ex. un tube en PVC)
directement aux composants de ce masque. Le PVC souple contient des
éléments qui peuvent altérer les matériaux du masque et entraîner leur
affaiblissement ou leur rupture.
618297 Fre.fm Page 12 Thursday, April 12, 2018 4:13 PM
13
Français
Utilisation du masque
Lorsque vous utilisez le masque Mirage Quattro sur des appareils ResMed
disposant d'une fonction de sélection du masque, sélectionnez l'option de
masque « facial ».
Mise en place du masque
Les diagrammes de la section A de la fiche d'illustrations vous indiquent
comment les composants de votre masque s'emboîtent les uns dans les
autres.
Les diagrammes de la section B vous indiquent comment mettre votre masque
en place pour obtenir le meilleur ajustement et le meilleur confort possibles.
Veuillez suivre ces instructions à la lettre.
1 Tournez la molette de réglage du support frontal jusqu'à la position 24, dans
laquelle le support est étiré au maximum (Fig.
B-1).
2 Maintenez le masque sur votre visage (Fig. B-2) et faites passer le harnais
par-dessus votre tête.
3 Ramenez les sangles inférieures sous vos oreilles et fixez le harnais à la
partie inférieure du masque (Fig.
B-3).
4 Desserrez les bandes Velcro
®
, tirez sur les deux sangles supérieures du
harnais en même temps jusqu'à ce qu'elles soient bien en place tout en
étant confortables et rattachez les bandes Velcro (Fig.
B-4).
5 Répétez l'étape 4 avec les sangles inférieures du harnais (Fig. B-5).
6 Testez le fonctionnement de la valve en bouchant la pièce pivotante avec
votre main et en vous assurant que vous pouvez toujours respirer librement.
7 Raccordez le circuit respiratoire principal à la pièce pivotante (Fig. B-6).
Raccordez ensuite l’extrémité libre du circuit respiratoire à votre appareil et
mettez ce dernier sous tension.
8 Allongez-vous dans la position dans laquelle vous dormez normalement.
Tournez la molette de réglage du support frontal (Fig.
B-7) jusqu'à ce que
vous ayez l'impression que le masque repose uniformément sur votre
visage. Il devrait être confortable, avec une étanchéité satisfaisante. Serrez
légèrement les sangles inférieures du harnais si nécessaire.
9 Le masque mis en place est représenté à la Fig. B-8.
Optimisation de l'étanchéité et du confort du masque
Pour que votre traitement soit efficace, il est important d'avoir une étanchéité
aussi bonne que possible entre votre masque et votre visage. Si vous êtes
confronté à des fuites, vérifiez que vous avez bien suivi les instructions de mise
en place. Vérifiez les points suivants rapidement après avoir mis votre masque
en place
:
618297 Fre.fm Page 13 Thursday, April 12, 2018 4:13 PM
14
Y a-t-il des plis au niveau de la bulle du masque ? Les plis peuvent laisser
l'air s'échapper du masque.
Avez-vous nettoyé la bulle de votre masque récemment ? Le sébum sur la
bulle peut affecter l'étanchéité.
Les sangles de votre harnais sont-elles trop serrées ? Essayez de ne pas
changer le serrage des sangles supérieures du harnais une fois que votre
technicien les a ajustées.
Vous devez peut-être ajuster la molette de réglage du support frontal. Notez
le réglage une fois que le masque est bien en place.
Avez-vous l'impression que le masque est trop près de vos yeux ? Vous avez
peut-être besoin d'un masque plus petit.
Si vous ne réussissez pas à résoudre un problème, reportez-vous à la section
Dépannage ou contactez votre prestataire de soins. Vous avez peut-être
besoin d'une autre taille de masque.
Retrait du masque
Pour retirer le Mirage Quattro, détachez l’un des clips inférieurs du harnais et
tirez sur le masque et le harnais pour les faire passer par-dessus votre tête.
Démontage du masque
Voir la section C de la fiche d'illustrations.
1 Détachez le circuit respiratoire.
2 Détachez les clips inférieurs du harnais (Fig. C-1) et décrochez les sangles
supérieures du harnais (Fig.
C-2). Il n’est pas nécessaire de démonter les
clips du harnais.
3 Retirez le coude de l'entourage rigide (Fig. C-3). Il est plus facile de tourner
l’ensemble coude pour le retirer de l'entourage rigide que de tirer dessus
sans tourner.
4 Retirez du coude le clip de la valve (Fig. C-4).
5 Enlevez la valve de son clip (Fig. C-5). Ne tirez pas sur la membrane pour
enlever la valve.
6 Détachez la pièce pivotante du coude (Fig. C-6).
7 Notez la position de la molette de réglage pour la prochaine fois que vous
utiliserez votre masque.
8 Dévissez complètement la molette de réglage pour la retirer et pour ensuite
retirer le support frontal de l'entourage rigide (Fig.
C-7).
9 Enlevez le tampon frontal du support frontal (Fig. C-8).
10 Pressez les pattes d'attache supérieures du clip de la bulle et poussez
dessus pour détacher la bulle de l'entourage rigide. Faites pivoter la bulle
vers l'arrière et tirez légèrement pour la désolidariser du masque (Fig.
C-9).
618297 Fre.fm Page 14 Thursday, April 12, 2018 4:13 PM
15
Français
11 Démontez le clip de la bulle (Fig. C-10).
12 R
etirez le bouchon d'entrées de l’entourage rigide (Fig. C-11).
Nettoyage du masque à domicile
Remarques:
Le masque et le harnais ne peuvent être nettoyés qu'à la main.
Il est possible de nettoyer le harnais sans avoir à détacher ses clips.
AVERTISSEMENT
Ne pas utiliser de solutions à base d'aromates ou d'huiles parfumées
(par ex. de l'essence d'eucalyptus ou des huiles essentielles), d’eau de
Javel, d’alcool ou de produits fortement parfumés (par ex. au citron)
pour le lavage des composants du masque. Les vapeurs résiduelles de
ces produits peuvent être inhalées si les composants ne sont pas
correctement rincés. Ces produits peuvent également endommager le
masque et provoquer des fissures.
Ne pas repasser le harnais au risque d'endommager le matériau dans
lequel il est fabriqué, celui-ci étant sensible à la chaleur.
Quotidien/Après chaque utilisation
Pour optimiser l'étanchéité du masque, toute trace de sébum doit être retirée
de la bulle avant utilisation.
Nettoyez le masque à la main en les frottant légèrement dans de l'eau tiède
(86°F / 30°C) à l'aide d'un sa
von doux ou d'un liquide vaisselle dilué.
Ne tordez pas la valve pendant le nettoyage au risque d'endommager ou de
déc
hirer la membrane ou la charnière de la valve. Manipulez la valve avec
précaution.
Si les orifices sont sales, utilisez une brosse à poils doux pour les nettoyer.
Rincez tous les composants soigneusement à l'eau potable et laissez-les
séc
her à l'abri de la lumière directe du soleil. Ne laissez pas le masque tremper.
Hebdomadaire
Nettoyez le harnais à la main en le frottant légèrement dans de l'eau tiède
(86°F / 30°C) à l'aide d'un savon doux ou d'un liquide vaisselle dilué. Rincez
soigneusement et laissez-le séc
her à l'abri de la lumière directe du soleil.
618297 Fre.fm Page 15 Thursday, April 12, 2018 4:13 PM
16
Retraitement du masque entre les patients
Seules les versions SLM (Sleep Lab Mask) du masque sont destinées à un
usage multiple par plusieurs patients. Avant d’être utilisés par un autre patient,
ces masques doivent être retraités conformément aux instructions de
nettoyage et de désinfection disponibles sur ResMed.com/downloads/masks.
Remontage du masque
Voir la section D de la fiche d'illustrations.
Avant de remonter le masque, vérifiez que tous les composants sont bien
propres et secs. Vérifiez que la valve n'est pas endommagée et qu'aucun des
composants restants n'est cassé ou fissuré. Remplacez tout composant
endommagé.
Remontage du coude
1 Introduisez la valve dans son clip (Fig. D-1).
2 Mettez le clip de la valve en place sur le coude en poussant (Fig. D-2).
Veillez à ce que les deux côtés se verrouillent en position (Fig.
D-3).
3 Introduisez la pièce pivotante dans l'extrémité du coude (Fig. D-4).
4 Introduisez le coude assemblé (Fig. D-5) dans l'entourage rigide.
Remontage du masque
1 Mettez le clip de la bulle dans le bord à rainure de la bulle. Commencez par
monter la partie supérieure et poursuivez tout autour de la bulle en poussant
le clip pour le fixer à la bulle (Fig.
D-6). Veillez à ce que la bulle ne soit ni
entortillée ni tordue autour du clip.
2 Alignez le clip de la bulle avec l'entourage rigide. Poussez le clip sur
l'entourage rigide pour le fixer, en vous assurant que les trois clips se
verrouillent en position (Fig.
D-7).
3 Poussez fermement sur le tampon frontal pour le fixer au support frontal
(Fig.
D-8).
4 Introduisez le support frontal dans l'entourage rigide et vissez la molette de
réglage sur le montant (Fig.
D-9). Tournez la molette de réglage jusqu'à ce
qu'elle se verrouille en position (Fig.
D-10). Vérifiez que la molette de
réglage est bien à niveau avec l'entourage.
5 Mettez le bouchon d'entrées en place (Fig. D-11).
6 Rattachez le haut du harnais en le faisant passer dans la boucle (Fig. D-12).
7 Tournez la molette de réglage jusqu'à ce que vous atteigniez la position qui
vous procure un bon niveau de confort et d'étanchéité (Fig.
D-13).
618297 Fre.fm Page 16 Thursday, April 12, 2018 4:13 PM
17
Français
Rangement
Veillez à ce que le masque soit bien propre et sec avant de le ranger. Rangez
le masque dans un endroit frais et sec à l'abri de la lumière directe du soleil.
Élimination
Le Mirage Quattro ne contient aucune substance dangereuse et peut être jeté
avec les ordures ménagères.
Dépannage
Problème Raison possible Solution possible
Le masque n'est pas étanche ou
est inconfortable.
La bulle du masque est peut-
être sale.
Nettoyez la bulle conformément
aux instructions de la section
« Nettoyage du masque à
domicile ».
Le masque n'a peut-être pas été
cor
rectement mis en place ou
n'est pas correctement ajusté.
Suivez les instructions de la
section « Mise en place du
masque » à la lettre. Vérifiez que
le harnais et la molette de
réglage ne sont pas trop serrés.
Dévissez la molette pour exercer
plus de pression sur le menton.
Serrez la molette pour
augmenter la pression exercée
sur l'arête du nez.
La bulle du masque a été mise
en biais dans le clip.
Vérifiez la position de la bulle et
introduisez-la à nou
veau
correctement.
La bulle du masque est décalée
par rapport au visage ou
présente des plis au niv
eau de la
surface de contact avec le
visage.
Veillez à ce que la bulle soit
correctement alignée et
exempte de plis : détachez le
clip du harnais, vérifiez la bulle et
repositionnez le masque sur
votre visage.
La taille de votre masque n'est
pas la bonne.
Contactez votre clinicien.
Le masque fait trop de bruit. Les orifices de ventilation sont
tot
alement ou partiellement
obstrués.
Nettoyez les orifices de
ventilation conformément aux
instructions de la section
« Nettoyage du masque à
domicile ».
La valve est mal montée. Remontez la valve.
La bulle n'est pas complètement
fix
ée au clip.
Vérifiez l'insertion de la bulle.
Le masque n'est pas
cor
rectement monté.
Remontez le masque.
La valve est déchirée ou
déf
ormée et ne forme pas de
joint étanche.
Remplacez la valve.
La molette de réglage a été fixée
à l'entourage rigide a
vant le
support frontal.
Mauvais montage. Poussez la base de la molette
sur une surface dure jusqu'à ce
qu'elle se détache de
l'entourage.
618297 Fre.fm Page 17 Thursday, April 12, 2018 4:13 PM
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43

ResMed Mirage Quattro Manuel utilisateur

Taper
Manuel utilisateur
Ce manuel convient également à