Mettler Toledo Transmitter M400 Mode d'emploi

Taper
Mode d'emploi
Quick Setup Guide Transmitter M400
en Quick Setup Guide 3
de Quick-Setup-Leitfaden 27
fr Guide de paramétrage rapide 51
it Guidaallacongurazionerapida 75
es Guíadeconguraciónrápida 99
pt GuiadeConguraçãoRápida 123
ru Руководствопобыстройнастройке 147
ja 171
Multilingual Quick Setup Guide
© 02/2023 METTLERTOLEDO.
Subjecttotechnicalchanges.PrintedinSwitzerland.52 121 395 I
3
Quick Setup Guide
Transmitter M400
Content
1 Operation 4
2 Menu Structure 5
3 TerminalBlock(TB)denitions 6
4 Wiring example for pH Transmitter 9
5 General Setup (applies for all parameters) 10
6 pHCalibrationincl.ISFET 11
7 O2 Calibration 13
8 CO2 Calibration (InPro 5000) 14
9 CO2 Hi (High) Calibration (InPro 5500 i) 17
10 M400 Type 1 Cond Ind only:
Conductivity Calibration for inductive Sensors 19
11 Conductivity Quick Setup 20
12 Conductivity Calibration
for 2-e and 4-e Sensors 21
13 TDLGPro™500 22
14 Environmentalprotection 26
English
en
4
1 Operation
Entry of data values, selection of data entry options
Use the m key to increase or the . key to decrease a digit. Use the same keys to
navigatewithinaselectionofvaluesoroptionsofadataentryeld.
h Note:Somescreensrequireconguringmultiplevaluesviathesamedataeld(ex:
conguringmultiplesetpoints).Besuretousethec or b key to return to the pri-
maryeldandthem or .keytotogglebetweenallcongurationoptionsbefore
entering to the next display screen.
Navigation with u in Display
If a u is displayed on the bottom right hand corner of the display, you can use the
c or bkey tonavigate toit.Ifyou click[ENTER] youwill navigatebackwards
through the menu (go back one screen). This can be a very useful option to move
back up the menu tree without having to exit into the measuring mode and re-enter
the menu.
Exit menu
h Note: Exit the menu at any time by pressing the b and c key simultaneously
(ESCAPE). The transmitter returns to the Measurement mode.
”Save changes” dialog
Three options are possible for the ”Save changes” dialog:
– ”Yes & Exit”: Save changes and exit to measuring mode.
– ”Yes & u”: Save changes and go back one screen.
–”No&Exit”: Don’tsavechangesandexittomeasuringmode.
The ”Yes & uoptionisveryusefulifyouwanttocontinueconguringwithouthaving
to re-enter the menu.
en
5
2 Menu Structure
Channel Select
Output
Set Points
Measurement
Mode M400
CalMenu Info
Quick Setup** Configure System PID Setup Service
Messages Calibration Data ISM Sensor Info* ISM Diagnostics*
Measurement
Analog Outputs
Set Points
Alarm/Clean
Display
Hold outputs
ISM Setup*
Set Language
USB
Passwords
Set/Clear Lockout
Reset
PID Display
PID/AM
Tune Parameters
Mode
Diagnostics
Calibrate
Tech Service
Model/Software
Revision
* Only available in combination with ISM sensors.
** Do not use after configuration
en
6
3 Terminal Block (TB) denitions
Power connections are labeled – N for Neutral and + LforLine,for100to240VACor20–30VDC.
TB2 for 2
1AO1+
2AO1–/AO2–
3AO2+
4AO3+
5AO3–/AO4–
6 AO4+
7DI1+
8DI1–/DI2–
9DI2+
TB2 for 2
1NO1 8 NC5
2COM1 9 COM6
3 NC1 10 NO6
4NO2 11 NO3
5COM2 12 COM3
6NC2 13 NO4
7COM5 14 COM4
TB2 TB3 TB4
TB1 1 14
1 9 1 9 1 9
Note:
This is a 4-wire-
product with an
active 4–20 mA
analog output.
Please do not
supply to
Pin1–Pin6of
TB2.
NO = normally
open (contact is
open if un-
actuated).
NC = normally
closed (contact
is closed if un-
actuated).
TB4 – ISM (digital) Sensors
TDL GPro™ 500 Optical Oxygen,
InPro™ 5500 i
(CO2)
InPro 6860 i
(VP8 cable) pH, amp. Oxygen,
Cond 4-e,
InPro 5000 i
Term. Function Color Color Color Color
124VDC brown grey
2GND(24VDC) black blue
3 1-Wire transparent
(cable core)
4GND(5VDC) green / yellow green / yellow red (shield)
5 – – – – –
6 GND(5VDC) brown –
7RS485-B yellow blue brown
8RS485-A green white pink
95VDC – – – –
en
7
TB3 – Conventional (analog) Sensors
Cond 2-E or 4-E amp. Oxygen
Term. Function Color* Function Color
1 Cnd inner1 white
2 Cnd outer1 white / blue Anode red
3 Cnd inner2 blue
4Cnd outer2 / Shield black Shield/GND green / yellow
5 – Cathode transparent
6 RTDret/GND bare shield GND/NTC green
7RTDsense red –
8RTD green NTC white
9+5V +5V
Terminal4and6are
internally connected,
either terminal can be
used to connect a wire.
* Transparent
not connected.
TB3 – Conventional (analog) Sensors
pH Redox (ORP)
Term. Function Color* Function Color
1 Glass tansparent Platinum transparent***
2 –
3Reference** red Reference Shield
4SolutionGND/Shield** green / yellow and blue – –
5 –
6 RTDret/GND white – –
7RTDsense – –
8RTD green – –
9+5V – –
Terminal4and6areinter-
nally connected, either ter-
minal can be used to con-
nect a wire.
* Grey wire not used.
** ForpHwithoutsolution
ground, install jumper
3 to 4.
***Removeblackcoating.
Terminal4and6areinter-
nally connected, either ter-
minal can be used to con-
nect a wire.
TB3 – Conventional (analog) Sensors
Cond Ind
Term. Color. InPro 7250 ST / PFA Color. InPro 7250 HT Function
1Coax inner / transparent Coax inner / transparent receive hi
2 red yellow receive lo
3green / yellow green / yellow shield/GND
4 brown, white violet send lo
5 blue black send hi
6white white RTDret/GND
7 grey grey RTDsense
8 green green RTD
9 not used
en
8
TB3 – Conventional (analog) Sensors
ISFET
Term. Color Function
1 Coax inner / pink FET
2 not used
3* yellow Reference
4* green / yellow GND/Shield
5 not used
6 white RTDret/GND
7 not used
8 grey RTD
9 brown +5V
Terminal4and6areinter-
nally connected, either ter-
minal can be used to con-
nect a wire.
* Jumper 3 to 4 has to be
installed.
TB3 – Conventional (analog) Sensors
Dissolved carbon dioxide
Term. Color* Function
1 Coax inner / transparent Glass
2 not used
3** Coax shield / red Reference
4** green / yellow GND/Shield
5 not used
6 white RTDret/GND
7 not used
8 green RTD
9 +5V
Terminal4and6areinter-
nally connected, either ter-
minal can be used to con-
nect a wire.
* Grey wire not used.
** Jumper 3 to 4 has to
be installed.
en
9
4 Wiring example for pH Transmitter (using TB3)
pH measurement with monitoring of glass electrode.
h Note: Wire colors only valid for connec tion
with VP cable, grey not connected.
1
2
3
4
5
6
7
8
9
yellowgreen
green
red
transparent
white
Combination
pH electrode
with RTD
and SG
Temperature
probe
blue
Cable
SG (Solution
Ground)
Jumper if
pH electrode
without SG
9 +5V
8 RTD
7 RTD sense
6 RTD ret/GND
5 not used
4 Solution GND/Shield
3 Reference
2 not used
1 Glass
en
10
5 General Setup (applies for all parameters) (PATH: Menu / Quick Setup)
 WhileinMeasurementmodepressthe[MENU]keytobringuptheMenuselection.
Select QUICK SETUPandpressthe[ENTER]key.
Display Convention:
1st line on display a 3rd line on display c
2nd line on display b 4th line on display d
Selecttheunitsofmeasurementforaandb.Onlylinesaandbcanbeconguredin
QUICK SETUP.GototheCongurationMenutocongurelinescandd.
Channel Selection
Please select the type of Sensor:
Analog: Forconventionalanalogsensors(willbedisplayedonchannel”A”).
ISM: ForISMsensors(willbedisplayedonchannel”B”).
Please select the parameter requirement:
The choice of parameter depends on the level of transmitter. If an ISM sensor is se-
lected,thesetting”Auto”means,allpossibleISMsensorswillberecognizedand
accepted.Ifaspecialparameterischosen,onlythisparameterwillberecognized
and accepted on the transmitter.
By using analogue 2-e or 4-e conductivity sensors,
see ”Conductivity Quick Setup” below for intermediate steps.
Analog out puts
By selecting YES the linear 4–20 mA analog output Aout1 will be set up for measure-
mentawhen[ENTER]ispressed.SelectingNO means that no analog output is set
up.
Aout1 min, Aout1 max are the minimum and maximum measurement values for the
4 and 20 mA values respectively.
h Note for multi channel devices:TheusercancongureoutputAout3tomesure-
ment c by going back to the previous menu and selecting c.
21.7 %sat
25.0 °C
MENU
Quick Setup
A 6.0 pH
A 25.0 °C
Channel Select = ISM
Parameter = Auto
7.0 pH
25.0 °C
Aout1 min= 0.000 pH
Aout1 max= 14.00 pH u
en
11
Setpoints
AfterconguringtheAnalogOutputasetpointcanbeconguredforthatoutput.IfNO
isselectedand[ENTER]ispressedthenthequicksetupisdoneandthemenusare
exited without setting up any setpoint.
Selecting YESandpressing[ENTER]meansasetpointcanbeconguredforchannela.
FollowingsetpointTypescanbeselected:
– High (High value has to be set).
– Low (Low value has to be set).
– Between (High and Low value has to be set).
–Outside(HighandLowvaluehastobeset).
Aftersettingthesetpointvalue(s)aRelay(none,1,2,3,4)canbeconguredforthat
setpoint.TheRelaydelayissetto10secondsandtheHysteresisissetto5%.Ifa
relayisconguredasCLEAN relay, it will not be selectable in this menu.
For multi-channel devices:Alsosetpoints5and6canbeconguredandtheuser
cancongureasetpointtomeasurementcbygoingbackthourghpreviousmenus
and selecting c.
6 pH Calibration incl. ISFET (PATH: Cal)
For Display setup, analog outputs and setpoints see the section ”General Setup”.
pH (resp. ISFET) Two point Calibration
Remark:WhenusingISFETsensors,youhavetoadjustthezeropointrst.
While in Measurement mode press the c key. If the display prompts you to enter the
calibration security code, press the m or . key to set the calibration security code,
thenpressthe[ENTER]keytoconrmthecalibrationsecuritycode.
Press the m or .keytoselectthepH(resp.ISFET)calibrationsubfunction.
Aashing”H”showstheongoingcalibrationprocess.
7.0 pH
25.0 °C
a Set Point Yes
SP1 Type= Off u
7.0 pH
25.0 °C
SP1 BetweenH= 0.000
SP1 BetweenL= 0.000 u
7.0 pH
25.0 °C
SP1 use Relay #1 u
A 7.0 MΩ-cm
A 25.0 °C
Calibrate Sensor
Channel A pH m
en
12
Select 2POINTCALIBRATIONbypressingthe[ENTER]key.
Placetheelectrodeintherstbuffersolutionandthenpressthe[ENTER]key.
Assoonasthestabilisationcriteriahavebeenfullled(or[ENTER]waspressedin
manual mode) the display changes and prompts you to place the electrode in the
second buffer solution.
Assoonasthestabilisationcriteriahavebeenfullled(or[ENTER]waspressedin
manual mode) the display changes to show the slope calibration factor S and the
offset calibration factor Z. Select ADJUST to save the calibration values and the suc-
cessfulCalibrationisconrmedonthedisplay.IfyouselectCALIBRATE, the values
will not be taken. If an ISM Sensor is used, the Calibration data will be stored in the
calibration history.
Process calibration
Select PROCESSCALIBRATION by pressing the mkeyoncefollowedbythe[ENTER]
key. To show the ongoing Calibration Process an ”H” is displayed in the top left hand
corner.
The ”H” changes to ”A” (or ”B”, depends on type of sensor) if Process Calibration is
selected to show the user the ongoing calibration on Channel ”A or ”B”.
Takeasampleandpressthe[ENTER]keyagaintostorethecurrentmeasuringValue.
After determining the pH Value of the Sample press the c key again to proceed with
the calibration. If the display prompts you to enter the calibration security code, press
the m or .keytosetthecalibrationsecuritycode,thenpressthe[ENTER]keyto
conrmthecalibrationsecuritycode.
H 7.0 pH
25.0 °C
pH Calibration
Type = 2 point u
H 7.0 pH
25.0 °C
Press ENTER when
Sensor is in Buffer 1 u
H 7.0 pH
25.0 °C
Press ENTER when
Sensor is in Buffer 2 u
H 7.0 pH
25.0 °C
Point2 = 6.86 pH
pH = 7.00 pH u
H 7.0 pH
25.0 °C
pH Calibration
Type = Process u
A 7.0 pH
25.0 °C
Point1 = 7.00 pH
pH = 6.87 pH u
en
13
EnterthepHvalueofthesamplethenpressthe[ENTER]keytostartcalibration.
After the calibration the slope calibration factor S and the offset calibration factor Z
are displayed. Select ADJUST to save the new calibration values and the successful
Calibrationisconrmedonthedisplay.IfyouselectCALIBRATE, the values will not
be taken. If an ISM Sensor is used, the Calibration data will be stored in the calibration
history. The ”A” in the top left hand corner disappears.
7 O2 Calibration (PATH: Cal)
For Display setup, analog outputs and setpoints see the section ”General Setup”.
O2 Calibration
For multi-channel devices: Using the m and .keyonthe”ChannelAeldletsthe
user change the channel to be calibrated.
Enter Calibration mode by pressing the c key.
Aashing”H”(alternatingwith”A”or”B”toshowthechannelbeingcalibratedfor
multi- channel devices) in the top left hand corner shows the ongoing calibration
process.
ADOsensorcalibrationisalwaysaonepointcalibrationeitherinAir(Slope)ora
zero(Offset)calibration.
It is possible to select a SLOPE or ZEROCALIBRATION. A one point slope calibration
isdoneinair andaonepoint offsetcalibrationisdoneat 0ppbDO.Pressthe
[ENTER]keyafterselectingSLOPE or OFFSET.
EnterthevalueforPoint1includingadecimalpoint.DOisthevaluebeingmeasured
bythetransmitterandsensorintheunitssetbytheuser.Press[ENTER]whenthis
value is stable to perform the calibration.
After the calibration the slope calibration factor S and the offset calibration factor Z
are displayed.
A 7.0 pH
25.0 °C
pH S=100.0 % Z=7.124
Save Calibration Yes u
A 21.7 pH
A 25.0 °C
Calibrate Sensor
Channel B Oxygen m
H 21.7 %sat
25.0 °C
Point1 = 100.0 %sat
DO = 0.033 %sat u
H 21.7 %sat
25.0 °C
O2 S=0.019nA Z=0.000nA
Save Calibration Yes u
en
14
SelectADJUSTtosavethecalibrationvaluesandthesuccessfulcalibrationiscon-
rmedonthedisplay.IfyouselectCALIBRATE, the values will not be taken. If an ISM
Sensor is used, the Calibration data will be stored in the calibration history.
8 CO2 Calibration (InPro 5000) (PATH: Cal)
For Display setup, analog outputs and setpoints see the section ”General Setup”.
CO2 One point Calibration
Whileinmeasurementmodepressthekey[CAL].Pressthem or . key to select the
CO2CALIBRATIONandpressthe[ENTER]keytocallthesubfunction.
Select 1POINTandpress[ENTER].
Placethesensorinthebuffersolutionandpressthe[ENTER]keytostartthecalibra-
tion.
Thedisplayshowsthebufferthetransmitterhasrecognized(Point1)andthemea-
sured value.
Assoonasthedriftconditionshavestabilizedthedisplaychangestoshowtheslope
calibration factor S and the offset calibration factor Z.
After a successful calibration, the calibration values are taken over (Adjust) or were
aborted (Calibrate or Abort).
H 21.7 %sat
25.0 °C
Calibration Successful
A 180.4 hPa
A 26.1 °C
Calibrate Sensor
Channel A CO2 u
H 180.4 hPa
A 26.1 °C
CO2 Calibration
Type = 1 Point u
H 137.5 hPa
A 26.1 °C
Press ENTER when
Sensor is in Buffer 1 u
154.5 hPa
A 26.1 °C
A Point1 = 7.00 pH
A CO2 = 7.07 pH u
H 154.5 hPa
A 26.1 °C
pH S=100.0% Z=7.048pH
Save Adjust u
en
15
CO2 Two point Calibration
StartasinchapterCO2 ONEPOINTCALIBRATION described.
Select 2POINTCALIBRATION.
Placethesensorintherstbuffersolutionandpressthe[ENTER]keytostartthe
calibration.
Thedisplayshowsthebufferthetransmitterhasrecognized(Point1)andthemea-
sured value.
Assoonasthedriftconditionshavestabilized,thedisplaychangesandpromptsyou
to place the electrode in the second buffer.
Placethesensorinthesecondbuffersolutionandpressthe[ENTER]keytogoon
with the calibration.
Thedisplayshowsthesecondbufferthetransmitterhasrecognized(Point2)and
the measured value.
Assoonasthedriftconditionshavestabilizedthedisplaychangestoshowtheslope
calibration factor S and the offset calibration factor Z.
After a successful calibration, the calibration values are taken over (Adjust) or were
aborted (Calibrate or Abort).
A 154.5 hPa
A 26.1 °C
CO2 Calibration
Type = 2 Point u
H 137.5 hPa
A 26.1 °C
Press ENTER when
Sensor is in Buffer 1 u
154.5 hPa
A 26.1 °C
A Point1 = 7.00 pH
A CO2 = 7.07 pH u
122.4 hPa
A 26.1 °C
Press ENTER when
Sensor is in Buffer 2 u
2.8 hPa
A 26.1 °C
A Point2 = 9.21 pH ...
A CO2 = 8.80 pH u
2.8 hPa
A 26.1 °C
pH S=74.21% Z=6.948pH
Save Adjust u
en
 16
CO2 Process Calibration
StartasinchapterCO2 ONEPOINTCALIBRATION described.
Select PROCESSCALIBRATION.
Takea sampleandpress the[ENTER]key againto storethecurrent measuring
value. To show the ongoing calibration process, A or B (depending on the channel)
isblinkinginthedisplay.AfterdeterminingtheCO2 value of the sample, press the c
key again to proceed with the calibration.
EntertheCO2valueofthesamplethenpressthe[ENTER]keytostartcalibration.
The display shows the slope calibration factor S and the offset calibration factor Z.
After a successful calibration, the calibration values are taken over (Adjust) or were
aborted (Calibrate or Abort).
A 17.3 hPa
A 27.3 °C
CO2 Calibration
Type = Process u
A 17.3 hPa
A 27.3 °C
A Point1 = 00000 hPa
A CO2 = 17.3 hPa u
A 17.3 hPa
A 27.3 °C
A Point1 = 16.90 hPa
A CO2 = 17.3 hPa u
A 17.3 hPa
A 27.3 °C
pH S=100.0% Z=7.009pH
Save Adjust u
en
17
9 CO2 Hi (High) Calibration (InPro 5500 i) (PATH: Cal)
For Display setup, analog outputs and setpoints see the section ”General Setup”.
One point calibration
Whileinmeasurementmodepressthekey[CAL].Pressthe. or m key to select the
CO2Hiandpressthe[ENTER]keytocallthesubfunction.
Select“1point”andpress[ENTER].
Placethesensorinthecalibrationgas(e.g.100%CO2) resp. solution.
Press[ENTER].
Enter the value for Point 1 including a decimal point and units. The value in the
second text line is the value being measured by the transmitter and sensor in the
units selected by the user.
DependingontheDriftcontrol(seeUserManualchapter8.2.3.10“CO2 Hi Para-
meters”), one of the two following modes is active.
If Drift is auto, as soon as the signal of stabilization criteria have been fullled,
transmitter will display the calibration result automatic.
IfDriftisManual,press[ENTER]toproceedassoonasthevalueisstable,transmitter
willdisplaythecalibrationresultaftermanualpress[ENTER]
If ”Adjust” or ”Calibrate” are chosen, the message ”Calibration successful” is displayed.
Inanycaseyouwillgetthemessage”Re-installsensor”and”PressENTER”onthe
display.Afterpressing”ENTER”theM400returnstothemeasuringmode.
B 189.0 hPa
B 25.0 °C
Calibrate Sensor
Channel B CO2 Hi u
B 189.0 hPa
H 25.0 °C
CO2 Hi Calibration
Type = 1 Point u
B 189.0 hPa
H 25.0 °C
Press ENTER when
Sensor is in Gas (CO2) u
B 189.0 hPa
H 25.0 °C
B Point1=1013. hPa .
B CO2=189.0 hPa u
B 189.0 hPa
H 25.0 °C
CO2 S= 9.28 mV BL= 253 mV
Save Adjust u
en
18
Process calibration
StartasinchapterOnepointCalibrationdescribed.
Select Process Calibration
Takea sampleandpress the[ENTER]key againto storethecurrent measuring
value. To show the ongoing calibration process, A or B (depending on the channel)
is blinking in the display.
AfterdeterminingtheCO2valueofthesamplepressthe[CAL]keyagaintoproceed
with the calibration.
EntertheCO2valueofthesamplethenpressthe[ENTER]keytostartthecalculation
of the calibration results.
After the calibration the slope calibration factor S and the offset calibration factor Z
are displayed.
In case of a successful calibration, the calibration values are stored in the cal his-
tory* and taken over (Adjust), stored in the cal history* and not taken over (Calibrate)
or discarded (Abort).
If ”Adjust” or ”Calibrate” are chosen, the message ”Calibration successful” is dis-
played. The M400 returns to the measuring mode.
B 189.0 hPa
H 25.0 °C
CO2 Hi Calibration
Type Process u
B 189.0 hPa
B 25.0 °C
Press ENTER to Capture
B CO2=189.0 hPa u
B 189.0 hPa
H 25.0 °C
B Point1=1013. hPa
B CO2=189.0 hPa u
B 189.0 hPa
H 25.0 °C
CO2 S= 9.28 mV BL= 253 mV
Save Adjust u
en
19
10 M400 Type 1 Cond Ind only:
Conductivity Calibration for inductive Sensors (PATH: Cal)
For Display setup, analog outputs and setpoints see the section ”General Setup”.
Cond Ind Zero point Calibration
Whileinmeasurementmodepressthekey[CAL].Pressthem or . key to select the
CONDUCTIVITYCALIBRATIONandpressthe[ENTER]keytocallthesubfunction.
Select the kind of calibration (standard, linear or n. water).
Select ZEROPOINTandpress[ENTER].Removethesensoranddryit.
Pressthe[ENTER]keyagaintogoonwiththecalibration.
Assoonastheconditionshavestabilized(or[ENTER]pressedinthemanualmode)
the display changes to show the multiplier and adder.
After a successful calibration, the calibration values are taken over and stored in the
cal history (Adjust), only stored in the cal history (Calibrate) or the calibration can
be aborted (Abort).
A 1.25 µS/cm
A 25.00 °C
Calibrate Sensor
Channel A Conductivity m
A 1.25 µS/cm
A 25.00 °C
Cal Compensation
Standard m
H 40.5 mS/cm
A 23.9 °C
Conductivity Calibration
Type = Zero Point u
A 1.035 mS/cm
A 21.9 °C
A Point 1 = 0.000 mS/cm
A C = 1.035 mS/cm u
en
20
Cond Ind One point Slope Calibration
Start as in chapter CONDINDZEROPOINTCALIBRATION described. Select the kind of calib ration.
Select 1POINTSLOPEandpress[ENTER].
Enterthevalueofcalibrationpoint1andthenpressthe[ENTER]keytostartcalibra-
tion.
The cell multiplier and the adder are displayed. After a successful calibration, the
calibration values are taken over and can be stored in the cal history (Adjust), only
stored in the cal history (Calibrate) or the calibration can be aborted (Abort).
11 Conductivity Quick Setup (PATH: Menu / Quick Setup)
For Display setup, analog outputs and setpoints see the section ”General Setup”.
Sensor Type Selection
Select the type of sensor to be used with the M400 transmitter. Choices are ”Cond(2)”,
used for all 2-Electrode type sensors and ”Cond (4)” for all 4-electrode sensors. Press
the[ENTER]key.
Cell Constant
Enterthe appropriate cell constant(s):from thesensor label or certicate(M) for
2-electrode sensors, leaving (A) at 0.000; or (M) and (A) values for 4-electrode
sensors.Pressthe[ENTER]key.
Measurement units
Select the measurement (conductivity or temperature) and units for measurement.
If using analog output, select YES.Referbacktosection”GeneralSetup”tocontinue
setup.
217.4 µS/cm
A 25.0 °C
Conductivity Calibration
Type = 1 Point Slope u
217.4 µS/cm
A 25.0 °C
A Point 1 = 215.0 µS/cm
A C = 217.4 µS/cm u
1.25 µS/cm
25.0 °C
Sensor Type = Cond(2) m
1.25 µS/cm
25.0 °C
p M=0.1003 A=0.0000
s M=1.0000 A=0.0000
1.25 µS/cm
25.0 °C
a S/cm
Analog Output? Yes m
en
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114
  • Page 115 115
  • Page 116 116
  • Page 117 117
  • Page 118 118
  • Page 119 119
  • Page 120 120
  • Page 121 121
  • Page 122 122
  • Page 123 123
  • Page 124 124
  • Page 125 125
  • Page 126 126
  • Page 127 127
  • Page 128 128
  • Page 129 129
  • Page 130 130
  • Page 131 131
  • Page 132 132
  • Page 133 133
  • Page 134 134
  • Page 135 135
  • Page 136 136
  • Page 137 137
  • Page 138 138
  • Page 139 139
  • Page 140 140
  • Page 141 141
  • Page 142 142
  • Page 143 143
  • Page 144 144
  • Page 145 145
  • Page 146 146
  • Page 147 147
  • Page 148 148
  • Page 149 149
  • Page 150 150
  • Page 151 151
  • Page 152 152
  • Page 153 153
  • Page 154 154
  • Page 155 155
  • Page 156 156
  • Page 157 157
  • Page 158 158
  • Page 159 159
  • Page 160 160
  • Page 161 161
  • Page 162 162
  • Page 163 163
  • Page 164 164
  • Page 165 165
  • Page 166 166
  • Page 167 167
  • Page 168 168
  • Page 169 169
  • Page 170 170
  • Page 171 171
  • Page 172 172
  • Page 173 173
  • Page 174 174
  • Page 175 175
  • Page 176 176
  • Page 177 177
  • Page 178 178
  • Page 179 179
  • Page 180 180
  • Page 181 181
  • Page 182 182
  • Page 183 183
  • Page 184 184
  • Page 185 185
  • Page 186 186
  • Page 187 187
  • Page 188 188
  • Page 189 189
  • Page 190 190
  • Page 191 191
  • Page 192 192
  • Page 193 193
  • Page 194 194
  • Page 195 195
  • Page 196 196

Mettler Toledo Transmitter M400 Mode d'emploi

Taper
Mode d'emploi