NL Handboek van Methoden 01/20EN Handbook of Methods 01/20
KS4.3 T
0.1 - 4 mmol/l KS4.3
Acid / Indicator
20
S:4.3
Instrument specific information
The test can be performed on the following devices. In addition, the required cuvette
and the absorption range of the photometer are indicated.
Instrument Type Cuvette λMeasuring Range
MD 200, MD 600, MD 610,
MD 640, MultiDirect, PM 620,
PM 630
ø 24 mm 610 nm 0.1 - 4 mmol/l KS4.3
SpectroDirect, XD 7000,
XD 7500
ø 24 mm 615 nm 0.1 - 4 mmol/l KS4.3
Material
Required material (partly optional):
Reagents Packaging Unit Part Number
Alka-M-Photometer Tablet / 100 513210BT
Alka-M-Photometer Tablet / 250 513211BT
Application List
Waste Water Treatment
Drinking Water Treatment
Raw Water Treatment
Notes
1. The terms Alkalinity-m, m-Value, total alkalinity and Acid demand to KS4.3 are iden-
tical.
2. For accurate results, exactly 10 ml of water sample must be used for the test.
Display in the MD
100 / MD 110 /
MD 200
Bar code for the
detection of the
methods
Method name
Measuring range
Chemical Method
Method number
Language codes
ISO 639-1 Revision status
KS4.3 T / 20
How to use
ENMethod Reference Book4
EN
Implementation of the provision Acid capacity KS4.3 with Tablet
Select the method on the device
For this method, no ZERO measurements are to be carried out with the following de-
vices: XD 7000, XD 7500
10 ml
Fill 24 mm vial with 10 ml
sample.
Close vial(s).
Sample
Place sample vial in the
sample chamber. • Pay
attention to the positioning.
Dissolve tablet(s) by inver-
ting.
Sample
Place sample vial in the
sample chamber. • Pay
attention to the positioning.
Test
Press the TEST (XD:
START) button.
The result in Acid Capacity KS4.3 appears on the display.
Performing test
procedure
KS4.3 T / 20
NL Handboek van Methoden 01/20EN Handbook of Methods 01/20
How to use
ENMethod Reference Book 5
EN
Aluminium T / M40
ENMethod Reference Book1.06
EN
Aluminium T M40
0.01 - 0.3 mg/L Al AL
Eriochrom Cyanine R
Material
Required material (partly optional):
Reagents Packaging Unit Part Number
Aluminium No. 1 Tablet / 100 515460BT
Aluminium No. 1 Tablet / 250 515461BT
Aluminium No. 2 Tablet / 100 515470BT
Aluminium No. 2 Tablet / 250 515471BT
Set Aluminium No. 1/No. 2 100 Pc.#100 each 517601BT
Set Aluminium No. 1/No. 2 250 Pc.#250 each 517602BT
Preparation
1. To get accurate results the sample temperature must be between 20 °C and 25 °C.
2. To avoid errors caused by contamination, rinse the vial and the accessories
with Hydrochloric acid (approx. 20%) before the analysis. Then rinse them with
deionised water.
Aluminium T / M40
ENMethod Reference Book1.0 7
EN
Determination of Aluminium with Tablet
Select the method on the device.
10 mL
Sample
Fill 24 mm vial with 10 mL
sample.
Close vial(s). Place sample vial in the
sample chamber. Pay
attention to the positioning.
Zero
Press the ZERO button. Remove the vial from the
sample chamber.
24 mm
Add ALUMINIUM No.
1 tablet .
Crush tablet(s) by rotating
slightly and dissolve.
Add ALUMINIUM No.
2 tablet .
Aluminium T / M40
ENMethod Reference Book1.08
EN
24 mm
Crush tablet(s) by rotating
slightly.
Close vial(s). Dissolve tablet(s) by
inverting.
Sample
Test
9
4
5
6
7
8
10
11 12
1
2
3
5 min
Place sample vial in the
sample chamber. Pay
attention to the positioning.
Press the TEST (XD:
START)button.
Wait for 5 minute(s)
reaction time.
Once the reaction period is finished, the measurement takes place automatically.
The result in mg/L Aluminium appears on the display.
Aluminium T / M40
ENMethod Reference Book1.0 9
EN
Analyses
The following table identifies the output values can be converted into other citation
forms.
Unit Cite form Scale Factor
mg/l Al 1
mg/l Al2O31.8894
Chemical Method
Eriochrom Cyanine R
Appendix
Interferences
Removeable Interferences
A low test result may be given in the presence of Fluorides and Polyphosphates. The
effect of this is generally insignificant unless the water has fluoride added artificially.
In this case, the following table should be used to determine the actual concentration
of aluminium.
A special tablet ingredient prevents the measurement being affected as a result of
iron and manganese.
Fluoride Displayed value: Aluminium [mg/L]
[mg/L
F]
0.05 0.10 0.15 0.20 0.25 0.30
0.2 0.05 0.11 0.16 0.21 0.27 0.32
0.4 0.06 0.11 0.17 0.23 0.28 0.34
0.6 0.06 0.12 0.18 0.24 0.30 0.37
0.8 0.06 0.13 0.20 0.26 0.32 0.40
1.0 0.07 0.13 0.21 0.28 0.36 0.45
1.5 0.09 0.20 0.29 0.37 0.48 ---
Aluminium T / M40
ENMethod Reference Book1.010
EN
Method Validation
Limit of Detection 0.02 mg/L
Limit of Quantification 0.044 mg/L
End of Measuring Range 0.3 mg/L
Sensitivity 0.17 mg/L / Abs
Confidence Intervall 0.014 mg/L
Standard Deviation 0.006 mg/L
Variation Coefficient 3.71 %
Bibliography
Richter, F. Fresenius, Zeitschrift f. anal. Chemie (1943) 126: 426
According to
APHA Method 3500-Al B
# including stirring rod, 10 cm
Aluminium PP / M50
ENMethod Reference Book1.012
EN
Aluminium PP M50
0.01 - 0.25 mg/L Al AL
Eriochrom Cyanine R
Material
Required material (partly optional):
Reagents Packaging Unit Part Number
VARIO Aluminium Reagent, Set F20 1 Pieces 535000
Preparation
1. To get accurate results the sample temperature must be between 20 °C and 25 °C.
2. To avoid errors caused by contamination, rinse the vial and the accessories
with Hydrochloric acid (approx. 20%) before the analysis. Then rinse them with
deionised water.
Aluminium PP / M50
ENMethod Reference Book1.0 13
EN
Determination of Aluminium with Vario Powder Pack
Select the method on the device.
Blank
20 mL
Prepare two clean 24 mm
vials. Mark one as a blank.
Put 20 mL sample in
100 mL measuring beaker
Add Vario ALUMINIUM ECR
F20 powder pack.
9
4
5
6
7
8
10
11 12
1
2
3
30 sec
Dissolve the powder by
mixing.
Press the ENTER button. Wait for 30 second(s)
reaction time.
Blank
1
Add Vario HEXAMINE
F20 powder pack.
Dissolve the powder by
mixing.
Place 1 drops Vario
ALUMINIUM ECR Masking
Reagent in the blank.
Aluminium PP / M50
ENMethod Reference Book1.014
EN
24 mm
10 mL
Place 10 mL pre-treated
sample in each vial.
Close vial(s). Press the ENTER button.
9
4
5
6
7
8
10
11 12
1
2
3
5 min
Blank
Zero
Wait for 5 minute(s)
reaction time.
Place blank in the sample
chamber. Pay attention to
the positioning.
Press the ZERO button.
Sample
Test
Remove the vial from the
sample chamber.
Place sample vial in the
sample chamber. Pay
attention to the positioning.
Press the TEST (XD:
START)button.
The result in mg/L Aluminium appears on the display.
Aluminium PP / M50
ENMethod Reference Book1.0 15
EN
Analyses
The following table identifies the output values can be converted into other citation
forms.
Unit Cite form Scale Factor
mg/l Al 1
mg/l Al2O31.8894
Chemical Method
Eriochrom Cyanine R
Appendix
Interferences
Removeable Interferences
A low test result may be given in the presence of Fluorides and Polyphosphates. The
effect of this is generally insignificant unless the water has fluoride added artificially.
In this case, the following table should be used to determine the actual concentration
of aluminium.
Fluoride Displayed value: Aluminium [mg/L]
[mg/L
F]
0.05 0.10 0.15 0.20 0.25 0.30
0.2 0.05 0.11 0.16 0.21 0.27 0.32
0.4 0.06 0.11 0.17 0.23 0.28 0.34
0.6 0.06 0.12 0.18 0.24 0.30 0.37
0.8 0.06 0.13 0.20 0.26 0.32 0.40
1.0 0.07 0.13 0.21 0.28 0.36 0.45
1.5 0.09 0.20 0.29 0.37 0.48 ---
Bibliography
Richter, F. Fresenius, Zeitschrift f. anal. Chemie (1943) 126: 426
According to
APHA Method 3500-Al B
DE Methodenhandbuch 01/20
KS4.3 T
0,1 - 4 mmol/l KS4.3
Säure / Indikator
20
S:4.3
Instrumentspezifische Informationen
Der Test kann auf den folgenden Geräten durchgeführt werden. Zusätzlich sind die
benötigte Küvette und der Absorptionsbereich der Photometer angegeben.
Geräte Küvette λMessbereich
MD 200, MD 600, MD 610,
MD 640, MultiDirect, PM 620,
PM 630
ø 24 mm 610 nm 0,1 - 4 mmol/l KS4.3
SpectroDirect, XD 7000,
XD 7500
ø 24 mm 615 nm 0,1 - 4 mmol/l KS4.3
Material
Benötigtes Material (zum Teil optional):
Reagenzien Form/Menge Bestell-Nr.
Alka-M-Photometer Tablette / 100 513210BT
Alka-M-Photometer Tablette / 250 513211BT
Anwendungsbereich
Abwasserbehandlung
Trinkwasseraufbereitung
Rohwasserbehandlung
Anmerkungen
1. Die Begrie Alkalität-m, m-Wert, Gesamtalkalität und Säurekapazität KS4.3 sind
identisch.
2. Die exakte Einhaltung des Probevolumens von 10 ml ist für die Genauigkeit des
Analysenergebnisses entscheidend.
Displayanzeige
im MD 100 MD 110
/ MD 200
Barcode zur
Methodenerkennung
Methoden Name
Messbereich
Chemische
Methode
Methodennummer
Sprachkürzel nach
ISO 639-1 Revisionsstand
KS4.3 T / 20
Leitfaden für den Anwender
ENMethod Reference Book16
DE
Durchführung der Bestimmung Säurekapazität KS4.3 mit Tablette
Die Methode im Gerät auswählen.
Für diese Methode muss bei folgenden Geräten keine ZERO-Messung durchgeführt
werden: XD 7000, XD 7500
10 ml
24-mm-Küvette mit 10 ml
Probe füllen.
Küvette(n) verschließen.
Sample
Die Probenküvette in
den Messschacht stellen.
Positionierung beachten.
Tablette(n) durch Um-
schwenken lösen.
Sample
Die Probenküvette in
den Messschacht stellen.
Positionierung beachten.
Test
Taste TEST (XD: START)
drücken.
In der Anzeige erscheint das Ergebnis als Säurekapazität KS4.3.
Durchführung der
Messung
KS4.3 T / 20
DE Methodenhandbuch 01/20
Leitfaden für den Anwender
ENMethod Reference Book 17
DE
Aluminium T / M40
DEMethod Reference Book1.018
DE
Aluminium T M40
0,01 - 0,3 mg/L Al AL
Eriochromcyanin R
Material
Benötigtes Material (zum Teil optional):
Reagenzien Form/Menge Bestell-Nr.
Aluminium No. 1 Tablette / 100 515460BT
Aluminium No. 1 Tablette / 250 515461BT
Aluminium No. 2 Tablette / 100 515470BT
Aluminium No. 2 Tablette / 250 515471BT
Set Aluminium No. 1/No. 2#je 100 517601BT
Set Aluminium No. 1/No. 2#je 250 517602BT
Vorbereitung
1. Zur Erzielung genauer Analyseergebnisse muss eine Probentemperatur von 20 °C
bis 25 °C eingehalten werden.
2. Zur Vermeidung von Fehlern durch Verunreinigungen, die Küvette und das Zubehör
vor der Analyse mit Salzsäurelösung (ca. 20%ig) und anschließend mit VE-Wasser
spülen.
Aluminium T / M40
DEMethod Reference Book1.0 19
DE
Durchführung der Bestimmung Aluminium mit Tablette
Die Methode im Gerät auswählen.
10 mL
Sample
24-mm-Küvette mit 10 mL
Probe füllen.
Küvette(n) verschließen. Die Probenküvette in
den Messschacht stellen.
Positionierung beachten.
Zero
Taste ZERO drücken. Küvette aus dem
Messschacht nehmen.
24 mm
Eine ALUMINIUM No.
1 Tablette zugeben.
Die Tablette(n) unter
leichter Drehung
zerdrücken und lösen.
Eine ALUMINIUM No.
2 Tablette zugeben.
Aluminium T / M40
DEMethod Reference Book1.020
DE
24 mm
Tablette(n) unter leichter
Drehung zerdrücken.
Küvette(n) verschließen. Tablette(n) durch
Umschwenken lösen.
Sample
Test
9
4
5
6
7
8
10
11 12
1
2
3
5 min
Die Probenküvette in
den Messschacht stellen.
Positionierung beachten.
Taste TEST (XD: START)
drücken.
5 Minute(n) Reaktionszeit
abwarten.
Nach Ablauf der Reaktionszeit erfolgt automatisch die Messung.
In der Anzeige erscheint das Ergebnis in mg/L Aluminium.
Aluminium T / M40
DEMethod Reference Book1.0 21
DE
Auswertung
Die folgende Tabelle gibt an wie die ausgegebenen Werte in andere Zitierformen
umgewandelt werden können.
Einheit Zitierform Umrechnungsfaktor
mg/l Al 1
mg/l Al2O31.8894
Chemische Methode
Eriochromcyanin R
Appendix
Störungen
Ausschließbare Störungen
Durch die Anwesenheit von Fluoriden und Polyphosphaten können die
Analyseergebnisse zu niedrig ausfallen. Dieser Einfluss hat im allgemeinen keine
signifikante Bedeutung, es sei denn, das Wasser wird künstlich fluoriert. In diesem
Fall kann die unten angegebene Tabelle angewandt werden, um die tatsächliche
Aluminiumkonzentration zu bestimmen.
Störungen durch Eisen und Mangan werden durch einen speziellen
Tabletteninhaltsstoff verhindert.
Fluorid Wert im Display: Aluminium [mg/L]
[mg/L
F]
0,05 0,10 0,15 0,20 0,25 0,30
0,2 0,05 0,11 0,16 0,21 0,27 0,32
0,4 0,06 0,11 0,17 0,23 0,28 0,34
0,6 0,06 0,12 0,18 0,24 0,30 0,37
0,8 0,06 0,13 0,20 0,26 0,32 0,40
1,0 0,07 0,13 0,21 0,28 0,36 0,45
1,5 0,09 0,20 0,29 0,37 0,48 ---
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103

Lovibond 00386440 - MD1x0/200 - Aluminium PP, Aluminium T Manuel utilisateur

Taper
Manuel utilisateur