Infusion set / Set de infusión /
Dispositif de perfusion
Made in Denmark
Fabricado en Dinamarca
Fabriqué au Danemark
In US, distributed by:
En EE.UU., distribuido por:
Aux États-Unis, distribué par :
Roche Diabetes Care, Inc.
Indianapolis, IN 46256, USA
Toll Free / Teléfono atención / Sans frais 1-800-280-7801
ACCU-CHEK and ACCU-CHEK TENDER are trademarks of Roche.
ACCU-CHEK y ACCU-CHEK TENDER son marcas registradas de Roche.
ACCU-CHEK et ACCU-CHEK TENDER sont des marques de Roche.
©2016 Roche Diabetes Care
Legal Manufacturer: / Fabricante legal: / Fabricant légal :
Unomedical a/s
Aaholmvej 1-3, Osted
4320 Lejre, Denmark
1
4
2
5
3
6
Manufacturer
Fabricante
Fabricant
Global Trade Item Number
Número mundial de artículo comercial
Code article international
2
Instructions for use
Instrucciones de uso
Instructions d’utilisation
en
es
fr
Infusion set
33
L K J I N
M
F
B
C
G D
A
E
H
transfer set
head set
4
en
Description
The Accu-Chek Tender infusion set is intended for the subcutaneous infusion
of insulin, administered with microdosage insulin pumps.
Contents of the package Accu-Chek Tender II
Inner packages 1 and 2.
Package 1: head set and transfer set
A protective cover
B soft cannula
C introducer needle
D side clips of the introducer needle
E self-adhesive
F front half of the adhesive backing
G back half of the adhesive backing
H cannula housing
I tubing
J connector with connector needle
K side clips of the connector
L circular protective cap
M disconnect cover
N luer-lock connector
5
Package 2: additional head set
Components A–H from package 1.
Contents of the package Accu-Chek Tender I
Package 1 only.
WARNING
Read the instructions carefully before use.
We recommend that you consult your insulin pump manual for information
regarding pump therapy.
When using the infusion set for the first time, your doctor or healthcare team
must be present.
This device remains sterile until the package is opened or becomes damaged.
Do not use if the sterile package has been opened or damaged.
Protect the product from excessive humidity, sunlight and heat. Store at
room temperature.
6
en
Dispose of your infusion set after use according to local regulations. Do not
clean or re-sterilize.
Reuse of the infusion set may cause infection, site irritation or damage to the
cannula/needle. A damaged cannula/needle may lead to inaccurate insulin
delivery.
The infusion set is for subcutaneous use only.
Always screw the luer-lock connector firmly into the pump adapter. If this is
not done, a loose connection between the infusion set and the pump could
cause leakage. Avoid using tools to screw the luer-lock connector into place,
otherwise damage could result to the luer-lock connector.
Do not leave air bubbles in the infusion set. Prime thoroughly.
A small percentage of the Accu-Chek Tender infusion sets using soft cannulas
can accidentally become crimped during insertion or displaced while in use.
If this should occur, use a new infusion set immediately.
Check the infusion site frequently to make sure the soft cannula remains
firmly in place. Because the cannula is so soft, you will not notice any pain if
7
it is pulled out. The soft cannula must always be completely inserted in order
to receive the full amount of insulin. Replace the infusion set as soon as the
self-adhesive loosens.
Do not reinsert the introducer needle into the soft cannula. Reinsertion could cause
tearing of the soft cannula which would result in unpredictable insulin delivery.
If you have to disconnect your infusion set, pay particular attention to hygiene.
Never use a soft cannula for more than 72 hours (3 days). Never use the
transfer set for more than 6 days. When changing the cartridge, make sure
that the transfer set is completely primed before connecting it to the head set.
lf the infusion site becomes inflamed, replace the infusion set and use a
new site.
Do not put disinfectants, perfumes, deodorants or insect repellent on your
infusion set as these may aect the integrity of the infusion set.
Never prime the transfer set or attempt to free a clogged line while the transfer
set is connected to the head set. You may accidentally infuse an uncontrolled
quantity of insulin.
8
en
Avoid mechanical stress on the infusion site and on the infusion set. Pay
attention when carrying heavy weights.
Before inserting the infusion set, prepare the infusion site in accordance with
the instructions of your healthcare professional.
Check your blood glucose level 1 to 3 hours after inserting your infusion set
and inspect the infusion site on a regular basis. Do not insert a new soft
cannula just prior to bedtime. Check blood glucose frequently (check with
your healthcare professional).
lf there is an unexplained increase in your blood glucose level, causing it to
be too high, or an “occlusion” alarm occurs, check for clogs and leaks. lf in
doubt, change your infusion set.
This product contains small parts. There is a risk of suocation if small parts
(e.g. covers, caps or similar objects) are swallowed.
Pay attention to the tubing of the infusion set connected to your body. There
is a risk of strangulation if the tubing becomes wrapped around your neck.
9
In case of doubt, use the shortest tubing length to minimise the risk of
strangulation.
Insert the infusion set
Picture 1
Choose a site away from the waistline, bones, scar tissue, belly button and
recent infusion sites. Wash your hands and clean the infusion site according
to the instructions from your healthcare professional. Allow the infusion site to
dry before the soft cannula is inserted.
Remove the front half of the adhesive backing (F) from the self-adhesive (E) and
pull the protective cover (A) o the introducer needle (C). Grasp the infusion set
between the thumb and middle finger of one hand behind the side clips and
pull back the edge of the self-adhesive with the index finger.
10
en
WARNING
Ensure clean application. Do not touch the soft cannula or the introducer needle.
Picture 2
With your free hand, pinch together the subcutaneous tissue at the disinfected
site. Insert the introducer needle (C) under the skin at an angle of 20–45°,
proceeding quickly and in one motion. Watch the soft cannula (B) while
inserting to assure that it goes in smoothly without kinking or crimping.
WARNING
People with very small or large amounts of subcutaneous fatty tissue should take
care when choosing the angle of insertion or the cannula length. The cannula could
become lodged in the muscular tissue or epidermis, where absorption of the insu-
lin is limited or impossible. Ask your healthcare professional for more information.
Make sure that the soft cannula is fully inserted into the subcutaneous tissue
and not kinked at the skin surface.
11
Insert the introducer needle in a single, smooth movement, otherwise the soft can-
nula may kink or crimp upon insertion, causing reduced or blocked insulin delivery.
Picture 3
Press the self-adhesive (E) firmly onto the skin in order to ensure secure
positioning of the infusion set. Place one finger in front of the inspection window
on the infusion set to fix it. Press the side clips of the introducer needle (D) together
gently and at the same time pull the introducer needle (C) out of the skin.
WARNING
Dispose of the introducer needle safely, so that nobody can prick or injure
themselves on it.
The safest way to dispose of a used introducer needle is to immediately place it
in a sharps disposal container to reduce the risk of accidental pricks or injuries
from loose sharps. If you cannot find a sharps disposal container right away,
12
en
you may need to recap the needle or use a needle clipper until you have an
opportunity to dispose of sharps in an appropriate sharps disposal container.
Never throw away loose introducer needles and other sharps in trash cans or
recycling bins, and never flush them down the toilet.
Picture 4
Remove the back half of the adhesive backing (G) from the self-adhesive (E)
and press the self-adhesive firmly onto the skin.
Picture 5
Attach the luer-lock connector (N) to the pump by screwing it into the pump
adapter. Make sure there is a cartridge containing sucient insulin in the
cartridge compartment. Remove the circular protective cap (L) by pressing the
side clips gently inwards. Follow the pump manufacturer’s User’s Manual to
prime the transfer set until insulin drips from the connector needle tip. Do not
leave air bubbles in either the cartridge or the transfer set.
13
Picture 6
Connect the transfer set to the newly inserted head set with the bulging side
upwards. An audible “click” ensures the connector has locked into place. Fill
the empty space of the cannula housing (H) with a bolus of 0.7 units of U100
insulin (about 7µl). Start insulin delivery with the pump.
WARNING
Change the head set every 2 to 3 days (max. 72 hours), otherwise you risk
insulin resistance or inflammation of the infusion site.
Never use the transfer set for more than 6 days.
You must fill the empty space of the newly inserted head set after connecting
the transfer set. Failure to do so results in missed insulin.
Always check your blood glucose level 1 to 3 hours after inserting your infusion set.
If the coupling does not click noticeably and audibly together, use a new head set
and a new transfer set.
Do not use any tools to screw the luer-lock connector into place.
14
en
Priming volume:
60 cm/24" approx 10 units U100 insulin approx 100 µl
80 cm/31" approx 13 units U100 insulin approx 130 µl
110 cm/43" approx 18 units U100 insulin approx 180 µl
Disconnect the infusion set
The infusion set allows you to temporarily disconnect from your pump for
showering, swimming, etc., without changing your infusion set.
WARNING
Consult with your healthcare professional on how to compensate for any missed
insulin while you are disconnected. Carefully monitor your blood glucose level
while disconnected from the pump and after you reconnect.
15
Step 1
Stop the pump. Hold the self-adhesive (E) in place, gently press the side clips of
the connector (K) together and pull the connector (J) from the cannula housing (H).
Step 2
Put the disconnect cover (M) on the cannula housing.
Reconnect the infusion set
WARNING
Before reconnecting, ensure that a drop of insulin appears at the tip of the
connector needle.
Step 1
Prime the infusion set until insulin comes out of the connector needle.
16
en
Step 2
Hold the self-adhesive (E) in place and remove the disconnect cover (M) from
the cannula housing (H).
Reattach the connector (J) to the cannula housing. An audible “click” ensures
the connector is locked.
WARNING
It is not necessary to give an additional bolus.
In US, distributed by:
Roche Diabetes Care, Inc.
Indianapolis, IN 46256, USA
Toll Free: 1-800-280-7801
ACCU-CHEK Customer Care
Service Center: 1-800-688-4578
www.accu-chek.com
Canada
Accu-Chek Pump Support:
1-800-688-4578 (toll free)
www.accu-chek.ca
Set de infusión
17
L K J I N
M
F
B
C
G D
A
E
H
catéter
cabezal
18
es
Descripción
El set de infusión Accu-Chek Tender ha sido diseñado para la administración
subcutánea de insulina realizada con micro-infusoras de insulina.
Contenido del envase Accu-Chek Tender II
Envases interiores 1 y 2.
Envase 1: Cabezal y tubo
A capuchón protector
B cánula flexible
C aguja guía
D clips laterales de la aguja guía
E adhesivo
F mitad anterior del protector del
adhesivo
G mitad posterior del protector
del adhesivo
H cuerpo de la cánula
I catéter
J conector con aguja del conector
K clips laterales del conector
L cubierta protectora circular
M tapón de desconexión
N conector de cierre luer
19
Envase 2: Cabezal adicional
Componentes A–H del envase 1.
Contenido del envase de Accu-Chek Tender I
Sólo envase 1.
ADVERTENCIA
Lea detenidamente las instrucciones antes de usar el producto.
Es recomendable que consulte las instrucciones de uso de la micro-infusora
de insulina para obtener información acerca de esta terapia.
Si es la primera vez que lo utiliza, deberá colocarlo bajo la supervisión de un
profesional sanitario.
Si es la primera vez que utiliza el set de infusión, deberá hacerlo en presencia
de su médico o equipo de personal sanitario.
Este dispositivo se conserva estéril hasta que se abre o se daña el envase.
No utilice el producto si el envase está abierto o ha sufrido algún daño.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50

Accu-Chek Tender infusion set Mode d'emploi

Taper
Mode d'emploi
Ce manuel convient également à