Mettler Toledo APS768x-120V Guide d'installation

Taper
Guide d'installation

Ce manuel convient également à

Guide for installers
METTLER TOLEDO MultiRange
Explosion-proof weighing system
with APS768x power supply unit
K...x-T4, PBA430x, PFA57.x,
PTA459x, PUA579x,
PBK-/PFK-series cat2/DIV1
K...x-T4, WMH Ex, PFA57.x,
PTA459x, PUA579x,
PBK-/PFK-series cat2/DIV1
RS422
CL20mA
RS232
Ex-i
Power
Supply
ACM200
Hazardous area
Safe area
IND226x
IND560x
ICS426x
ICS466x
www.mt.com/support
Congratulations on choosing the quality and precision of METTLER TOLEDO. Proper
use according to these instructions and regular calibration and maintenance by our
factory-trained service team ensure dependable and accurate operation, protecting
your investment. Contact us about a service agreement tailored to your needs and
budget.
We invite you to register your product at
www.mt.com/productregistration so we can
contact you about enhancements, updates and important notifications concerning
your product.
Contents
Guide for installers 22021223F 17/04 3
APS768x
Contents Page
1 Safety precautions ....................................................................... 4
2 System overview ......................................................................... 6
2.1 Configurations .............................................................................. 6
2.2 Description of the components ....................................................... 9
3 Installation.................................................................................. 10
3.1 Setting up system modules ............................................................ 10
3.2 Equipotential bonding.................................................................... 12
3.3 Connecting power supply............................................................... 12
4 Fabricating cables ....................................................................... 13
4.1 Fabricating power cable (APS768x-230V)....................................... 13
4.2 Fabricating connection cables ........................................................ 14
5 Connecting a barcode reader........................................................ 16
5.1 Barcode wiring option ................................................................... 16
5.2 Mounting ..................................................................................... 16
6 Installing a CL/CL interface .......................................................... 17
7 Technical data ............................................................................. 18
8 Control drawing........................................................................... 20
Safety precautions
4 Guide for installers 22021223F 17/04
APS768x
1 Safety precautions
The explosion-proof APS768x-230V power supply unit is approved for operation in
Zone 1 and Zone 21 hazardous areas.
The explosion-proof APS768x-120V power supply unit is approved for operation in
hazardous areas of Class I, II, III; Division 1; Groups A–G.
Special care must be taken when using weighing systems with the APS768x power
supply unit in hazardous areas. The code of practice is oriented to the "Safe
Distribution" concept drawn up by METTLER TOLEDO.
Competence The weighing system may only be installed, maintained and repaired by
authorised METTLER TOLEDO service personnel.
The mains supply may only be installed by a specialist authorised by the owner-
operator.
Ex approval No modifications may be made to the device and no repair work may be
performed on the modules. Any weighing platforms or system modules that are
used must comply with the specifications. Non-compliant equipment jeopardises
the intrinsic safety of the system, cancels the Ex approval and renders any
warranty or product liability claims null and void.
The safety of the weighing system is only guaranteed when the weighing system
is operated, installed and maintained in accordance with the respective instructions.
Also comply with the following:
the instructions for the system modules,
the relevant national regulations and standards,
the applicable statutory requirements for electrical equipment installed in
hazardous areas in the respective country,
all instructions related to safety issued by the owner.
The explosion-proof weighing system must be checked to ensure compliance
with the safety requirements before being put into service for the first time,
following any service work and at least every 3 years.
Operation Prevent the build-up of static electricity. Always wear suitable working clothes
when operating or performing service work in a hazardous area.
Do not use protective covers with the devices.
Prevent damage to the system components.
Safety precautions
Guide for installers 22021223F 17/04 5
APS768x
Installation Only install or perform maintenance work on the weighing system in the
hazardous area if the following conditions are fulfilled:
the intrinsically safe characteristic values and zone approval of the individual
components are in accord with one another,
the owner has issued a permit ("spark permit" or "fire permit"),
the area has been rendered safe and the owner's safety co-ordinator has
confirmed that there is no danger,
the necessary tools and any required protective clothing are provided (danger
of the build-up of static electricity).
The certification papers (certificates, manufacturer’s declarations) must be present.
Lay cabling securely so that it does not move and effectively protect it against damage.
Only route cables into the housing of the system modules via the suitable gland
and ensure proper seating of the seals.
Use only genuine METTLER TOLEDO accessories and cable assemblies with this
product. Use of anauthorized or counterfeit accessories or cable assemblies may
result in voided warranty, improper or erroneous operation or damage to property
(including the unit) and personal injury.
System overview
6 Guide for installers 22021223F 17/04
APS768x
2 System overview
2.1 Configurations
The following represents 7 example configurations of a weighing system with the
APS768x power supply unit.
Note
Some components are not available in all countries.
2.1.1 APS768x and VIPER EX compact scale
Ex-i
2.1.2 APS768x, IND226x, IND560x, ICS426x, ICS466x weighing terminal and
K...x-T4, PBK-/PFK-series cat2/DIV1, PBA430x, PFA57.x, PTA459x or PUA579x
weighing platform
Ex-i
Ex-i
max. 122 m for analog scales
max. 20 m for digital scales
-T4
PBK-/PFK-series cat2/DIV1**
IND226x
IND560x
ICS426x
ICS466x
* K...x-T4 only in combination with IND560x/IDNet or ICS466x
** PBK-/PFK-series cat2/DIV1 only in combination with IND560x, ICS426x or
ICS466x weighing terminals with SICSpro digital scale interface.
System overview
Guide for installers 22021223F 17/04 7
APS768x
2.1.3 APS768x, IND560x weighing terminal and Point Ex/Ex1 system solution
Point Ex**, Ex1
System solution
Weighing cell 1*
Weighing cell 2*
Weighing cell 3*
Weighing cell 4*
Ex-i
Ex-i
* Minimum system impedance of 87 
The weighing cells must be approved for the hazardous area. The intrinsically
safe characteristic values must be in agreement with those of the Point Ex system
solution.
** Only in combination with IND560x/IDNet
2.1.4 APS768x, weighing platform K...x-T4, PBK-/PFK-series cat2/DIV1, WMH...Ex,
PFA57.x, PTA459x, PUA579x
max. 100 m
max.
300 m
K...x-T4,
WMH...Ex
PFA57.x,
PTA459x,
PUA579x
ACM200
Interface
CL/RS
CL
IDNet
RS
or
IND...
PC
Interface
CL/CL
PBK*-/PFK*-series
cat2/DIV1
* M12 Ex-i plug / open wires cable 20 m
2.1.5 APS768x, compact scale VIPER EX and data interface ACM200
max. 50 m
max.
300 m
VIPER EX
ACM200
Interface
CL/RS
RS
PC
Interface
CL/CL
System overview
8 Guide for installers 22021223F 17/04
APS768x
2.1.6 APS768x, weighing terminal ICS426x, ICS466x and data interface ACM200
ACM200
Analog weighing
platform
APS768x
Ex-i
Hazardous area Safe area
PDC-
SG-Ex1
CL ACTIVE
RS422/
CL20mA
RS232*
alternative
CL20mA*
alternative
RS422/485*
max. 50 m
Ex-i
max. 122 m
Ex-i
max.
300 m
ICS426x
ICS466x
* Only one hardware interface available,
CL20mA and RS422/485 only with module ACM200-CL/RS422
2.1.7
Configuration with two digital weighing platforms in the hazardous area
Digital weighing
platform,
e.g. KA15sx-T4 or
PBK-/PFK-series cat2/DIV1**
Digital weighing
platform,
e.g. K...x-T4 or
PBK-/PFK-series cat2/DIV1**
APS768x ICS466x
Safe areaHazardous area
Ex-i
max.
300 m
Ex-i
max. 20 m
CL PASSIVE
CL ACTIVE
DIGITAL
SCALE
Ex-i
max. 50 m
CL-CL
APS768x
ACM200
RS422/
CL20mA
RS232*
alternative
CL20mA*
alternative
RS422/485*
* Only one hardware interface available,
CL20mA and RS422/485 only with module ACM200-CL/RS422
** The weighing terminal has to support a SICSpro interface.
System overview
Guide for installers 22021223F 17/04 9
APS768x
2.2 Description of the components
APS768x-230V
power supply unit
Type of protection II 2G Ex eb mb [ib] IIC T4 Gb
(EN, IECEx) II 2D Ex tb IIIC [ib] IP66 T70°C Db
BVS 10 ATEX E081 IECEx BVS 10.0054
APS768x-120V
power supply unit
Type of protection (
c
FM
US
) Class I, II, III; Division 1; Group A, B, C, D, E, F, G / T4
T
a
= 40 °C
VIPER EX compact scale See VIPER EX operating instructions.
IND226x / IND560x /
ICS426x / ICS466x
weighing terminal
See IND226x / IND560x / ICS426x / ICS466x operating instructions/guide for
installers.
K...x-T4, PBK-/PFK-
series cat2/DIV1,
PBA430x, PFA57.x,
PTA459x, PUA579x,
WMH...Ex
weighing
platforms
See corresponding operating instructions/installation information.
Point Ex system solution See Point Ex guide for installers.
Ex1 system solution See Ex1 guide for installers.
Note
Some components are not available in all countries.
Installation
10 Guide for installers 22021223F 17/04
APS768x
3 Installation
EXPLOSION HAZARD
Installation of the explosion-proof weighing system must be carried out in
accordance with the control drawing at the end of this Guide for installers.
3.1 Setting up system modules
3.1.1 Setting up the VIPER EX compact scale
See VIPER EX operating instructions.
3.1.2 Setting up the IND226x / IND560x / ICS426x / ICS466x weighing terminal
See IND226x / IND560x / ICS426x / ICS466x Guide for installers.
3.1.3 Setting up the K...x-T4, PBK-/PFK-series cat2/DIV1, PBA430x, PFA57.x,
PTA459x, PUA579x or WMH...Ex weighing platform
See corresponding operating instructions and installation information.
3.1.4 Setting up the Point Ex system solution
See Point Ex Guide for installers.
3.1.5 Setting up the APS768x-230V
Fixed installation
1. Select a suitable location:
max. of 2 m from the power socket-outlet,
sufficient air circulation for cooling,
flat horizontal or vertical surface.
2. Use 4 screws M5 or M6 to screw the lower housing section via the attachment
lugs to the foundation. For hole pattern refer to the dimensional drawing.
Mobile installation
Depending on the operating specifications, use a power cable that is more
mechanically rugged than the cable mounted at the factory.
EXPLOSION HAZARD
Explosion hazard if not sufficiently dustproof!
If the APS768x power supply unit is used in dust explosion hazard areas ensure
that the degree of protection IP6x is ensured.
In this respect ensure the correct position of the seals and the maximum
tightening torque of the screwed connection.
During every maintenance interval check the respective tightening torque and
tighten if necessary.
Screw M4 x 0.7 M6 x 1 M16 x 1.5 M25 x 1.5
Max. tightening torque 1.2 Nm 2.0 Nm 2.0 Nm 3.0 Nm
Installation
Guide for installers 22021223F 17/04 11
APS768x
3.1.6 Setting up APS768x-120V
1. Select an appropriate location:
a flat horizontal or vertical surface,
sufficient air circulation for cooling.
2. Remove the lid of the enclosure and the screw on top of the explosion proof
cylinder.
3. Install the conduit to the 1/2" NPT pipe thread and pull the mains cable leads
(line and neutral) up to the connector block inside the cylinder. The cable ends
must be stripped bare to a length of 8 mm (3/8").
4. Use a screwdriver (inserted in the adjacent small hole) for opening the cage
clamp mechanism in the connector block, then insert the mains cable leads
(live: opposite wire black, neutral: opposite wire white).
5. If a ground wire is used in addition to Line and neutral, this can be inserted in the
third compartment of the connector block (opposite wire green/yellow). Proceed
as above.
6. Insert the top screw of the cylinder and fix the securing lever. Then mount the lid
to the enclosure.
EXPLOSION HAZARD
Explosion hazard if not sufficiently dustproof!
If the APS768x power supply unit is used in dust explosion hazard areas ensure
that the degree of protection IP6x is ensured.
In this respect ensure the correct position of the seals and the maximum
tightening torque of the screwed connection.
During every maintenance interval check the respective tightening torque and
tighten if necessary.
Screw M4 x 0.7 M6 x 1 M16 x 1.5 M25 x 1.5
Max. tightening torque
1.2 Nm
10.6 lbfin
2.0 Nm
17.7 lbfin
2.0 Nm
17.7 lbfin
3.0 Nm
26.5 lbfin
3.1.7 Connecting the power supply unit
The Ex-i connecting cable is packed unbound in the box. Depending on the system
or configuration the terminal assignment may differ.
Connect the power supply unit to the weighing terminal or compact scale in
accordance with the control drawing.
Installation
12 Guide for installers 22021223F 17/04
APS768x
3.2 Equipotential bonding
Equipotential bonding must be installed by an electrician authorised by the owner.
METTLER TOLEDO Service only has a monitoring and consulting function here.
Connect equipotential bonding (PA) of all devices (APS768x, weighing terminal,
weighing platform, compact scale) in accordance with the control drawing and
the country-specific regulations and standards. In the process it must be ensured
that
all device housings are connected to the same potential via the PA connecti-
ons,
no circulating current flows via the cable shielding for intrinsically safe circuits,
the neutral point for equipotential bonding is as close to the weighing system
as possible.
3.3 Connecting power supply
EXPLOSION HAZARD
The mains connection must be made by an electrician authorised by the owner in
accordance with the control drawing and country-specific regulations.
Notes
The APS768x power supply unit has no device of its own for mains disconnection.
This must be provided by the owner.
The power supply may only be conneced to a mains circuit in which the short-
circuit current does not exceed 35 A.
Fabricating cables
Guide for installers 22021223F 17/04 13
APS768x
4 Fabricating cables
If necessary, the power and Ex-i cables can be fabricated in accordance with the
requirements of the customer.
EXPLOSION HAZARD
Before opening the APS768x, ensure that non-intrinsically safe circuits are not
energised.
If the APS768x power supply unit is used in dust explosion hazard areas ensure
that the degree of protection IP6x is ensured.
4.1 Fabricating power cable (APS768x-230V)
When fabricating the power cable, the following requirements must be adhered to:
Maximum cable length in accordance with the specification of the owner
Cable quality: 3 x 1 mm
2
, additional qualities in accordance with the specification
of the owner
Maximum outside cable diameter: 5.0 – 8.0 mm
Procedure
1. Strip 40 mm of cable sheathing from the APS768x-230V and the mains
connection, as required. The free end of the grounding cable must be long
enough so that the grounding connection would be the last one to rip away in the
event of tensile strain.
2. Remove the old cable from the APS768x-230V and pull the new cable through
the cable gland. Ensure the IP protection type through proper installation.
3. Crimp a 6.3 x 0.8 mm receptacle onto the grounding cable on the APS768x-
230V; provide the other cable leads with end ferrules.
4. Connect the cable, tighten cable gland according to the manufacturer’s
specifications and screw on strain relief.
Fabricating cables
14 Guide for installers 22021223F 17/04
APS768x
4.2 Fabricating connection cables
EXPLOSION HAZARD
Before opening the APS768x, ensure that non-intrinsically safe circuits are not
energised.
If the APS768x power supply unit is used in dust explosion hazard areas ensure
that the degree of protection IP6x is ensured.
Customer-specific connection cables for intrinsically safe circuits must be fabricated
as follows:
Cable Dim. A
Dim. B
(APS768x)
Max.
length
VIPER EX – APS768x
4x2x0.5 mm
2
+ 1x0.5 mm
2
240 mm
1
80 mm
1
50 m
1
IND226x / IND560x /
ICS4_6x – APS768x
4x2x0.5 mm
2
+ 1x0.5 mm
2
240 mm
1
80 mm
1
50 m
1
IND560x / ICS466x –
APS768x – Barcode
2x2x0.5 mm
2
240 mm
1
50 mm
1
10 m
1
ACM200 – APS768x
2x2x0.5 mm
2
50 mm
1
80 mm
1
300 m
1
Cable Dim. A
Dim. B (IND...x/
APS768x))
Max.
length
Weighing platform /
Point Ex System
solution – IND226x /
IND560x
3x2x0.75 mm
2
80 mm
1
215 mm
1
20 m
1
Weighing platform /
Point Ex System
solution – APS768x
3x2x0.75 mm
2
80 mm
1
80 mm
1
100 m
1
For customer-specific connection cables M12 Ex-i plug / open wires used with
PBK-/PFK-series cat2/DIV1, see the corresponding installation information.
A B
. . .
. . .
Earthing cable gland
Cable shield
Push sleeve over
wires and shield
Cable as per country-specific
regulations for intrinsically safe circuits
Wire end ferrules
with plastic collar,
crimp connection
Fabricating cables
Guide for installers 22021223F 17/04 15
APS768x
1. Cut cable to length and strip cable ends according to dimension A/B, see table
above.
2. Shorten shield on both sides to 10 mm.
3. Strip wire ends.
4. Crimp wire end ferrules onto wire ends with a crimping tool.
5. Push second rear section of earthing cable gland onto cable.
6. On both sides: Push sleeve over wires and shield. Fold over cable shield.
7. Push on front section of cable gland and screw onto rear section.
Connecting a barcode reader
16 Guide for installers 22021223F 17/04
APS768x
5 Connecting a barcode reader
5.1 Barcode wiring option
A barcode reader, for example ELB SK200 (SC20-MTA1), can be connected to the
IND560x weighing terminal via an intrinsically safe RS232 data interface.
The Barcode wiring option of the APS768x power supply unit is used to supply the
barcode reader and consists of the following components:
APS768x
Barcode wiring option: 10 m Ex-i cable (2 x 2 x 0.5 mm
2
), pre-assembled
A suitable approved barcode reader does not form part of the scope of delivery from
METTLER TOLEDO and has to be ordered separately.
5.2 Mounting
EXPLOSION HAZARD
Before opening the APS768x, ensure that non-intrinsically safe circuits are not
energised.
RxD of the barcode reader may not be connected in the hazardous area. At the
barcode reader ELB SK200 this conductor has a shrink-on sleeve.
1. Connect the 4-conductor Ex-i cable of the barcode wiring option to the COM1 of
the IND560x in accordance with the control drawing 22006397 as of Version C.
2. If applicable, apply a shrink-on sleeve to the RxD of the barcode reader, if it is not
supplied in this state from the factory.
3. Pull the barcode reader cable into the APS768x and mount it with the cable
gland.
4. Connect the TxD of the barcode reader via a terminal part with the RxD of the
IND560x (prewired).
5. Connect the remaining conductors of the barcode reader cable at the connector
K3 to U5 and GND in accordance with the control drawing 22006397 as of
Version C.
Colour yellow brown white green
Signal RxD +5 V DC GND TxD
Version Shrink-on sleeve 3 cm stripped, with wire end ferrules
Barcode reader cable ELB SK200
Installing a CL/CL interface
Guide for installers 22021223F 17/04 17
APS768x
6 Installing a CL/CL interface
Digital (IDNet or SICSpro) weighing platforms or compact scales VIPER EX can be
connected via the optional CL/CL interface module of the APS768x to the
communication module ACM200 in the safe area.
Communication with operating terminals or computer systems is effected in the
ACM200 either via an RS232, CL20mA or RS422/485 hardware data interface.
EXPLOSION HAZARD
Before opening the APS768x ensure that non-intrinsically safe circuits are not
energised.
1. Loosen the screws of the APS768x housing cover and remove the cover.
2. Realise the connection cable to the weighing platform and communication
module via the two M16 holes. Ensure the correct position of the flat packings
when doing so.
3. Screw the APS768x-CL/CL interface mode to the two securing bolts.
4. Carry out the wiring of the two connecting cables and of the 2-wire
supplementary cable of the CL/CL interface in accordance with the connection
diagram 22006397 as of Version E.
When connecting a VIPER EX check the wiring of the compact scale and if
necessary correct at the terminal block.
In the configuration with a K...x-T4 or PBK-/PFK-series cat2/DIV1 weighing
platform shorten the wires of the connecting cable on the side of the APS768x
to approx. 80 mm. The individual wires may only be connected to the CL/CL
PCB with correctly crimped wire end ferrules.
5. Close the housing cover. Ensure proper seating of the seal when doing so.
6. Tighten the cover screws. Take the max. permissible tightening torque into
account when doing so, see page 10 resp. page 11.
Technical data
18 Guide for installers 22021223F 17/04
APS768x
7 Technical data
Mains connection
APS768x-230V: 230 V +10 % / –15 %, 50/60 Hz, 0.160 A, U
m
253 V
APS768x-120V: 120 V +10 % / –15 %, 50/60 Hz, 0.250 A, U
m
253 V
Cable for mains
connection
APS768x-230V: 1.9 m; H 05 RN
APS768x-120V: Explosion proof conduit system required
Ex-i connection cable From the factory 5 m long cable enclosed, intrinsically safe connection, on both
sides cable gland M16x1.5 pre-assembled. If required longer cable possible, see
Section 4.2.
IP protection type IP66
Temperature range –10 °C to +40 °C
Relative humidity 20 % – 80 %, non-condensing
Degree of soiling 2
Max. application height Up to 4000 m above sea level
Application location Inside locations only
Installation category II
Dimensions (L x W x H) APS768x-230V: 303 x 172 x 80 mm (without connection parts)
320 x 172 x 80 mm (with connection parts)
APS768x-120V: 320 x 172 x 80 mm
Weight (incl. cables) APS768x-230V: 4.1 kg (net)
APS768x-120V: 4.7 kg (net)
Technical data
Guide for installers 22021223F 17/04 19
APS768x
Dimension drawing APS768x-230V
80 mm
3.2"
30 mm
1.2"
255 mm
10.0"
24 mm / 0.9"
41 mm / 1.6"
10 mm
0.4"
284 mm / 11.2"
303 mm / 11.9"
320 mm / 12.6"
172 mm / 6.8"
100 mm / 3.9"
68 mm / 2.7"
Dimension drawing APS768x-120V
80 mm
3.2"
30 mm
1.2"
255 mm
10.0"
24 mm / 0.9"
10 mm
0.4"
284 mm / 11.2"
303 mm / 11.9"
172 mm / 6.8"
100 mm / 3.9"
68 mm / 2.7"
11 mm / 0.4"
Control drawing
20 Guide for installers 22021223F 17/04
APS768x
8 Control drawing
Scale 1
ID7sx
S1/3
S1/1
T1/4
T1/1
T1/3
T1/2
T1/5
S2/1
max. 20 m
7
6
8
9
5
10
TxD
RxD
GND
GND
U1
U2
U3
NC
U6
U4
Connection
in terminal box
of measuring celll
TBrick 15-Ex/32-Ex
A
K...x-T4 with TBrick-Ex
A
W1/5
W1/6
W1/4
W1/3
W1/2
W1/1
green
white
brown
yellow
pink
grey
B
B
passive
U6
S1/2
+
1
...
+
8
I, passive
+
1
...
+
8
O, passive
Mainboard ID7sx
ID7sx
S1/3
S1/1
T1/4
T1/1
T1/3
T1/2
T1/5
S2/1
A B
passive
U6
S1/2
Mainboard ID7sx
TxD
RxD
GND
GND
U1
U2
W1/5
W1/6
W1/4
W1/3
W1/2
W1/1
Scale 1
max. 20 m
7
6
8
9
5
10
B
green
white
brown
yellow
pink
gray
SEN
EXC
SIG
SIG +
SEN +
EXC +
Scale 1
B
Analog scale
Connecting terminals
Point Ex
7
6
8
9
5
10
+
1
...
+
8
I, passive
+
1
...
+
8
O, passive
Analog interface
EXC
SIG
SIG +
SEN +
EXC +
active
B
Alternative to
active
via internal
cable in ID7sx
SEN
NC
NC
passive
CENELEC / FM approved
equipment for category 2 /
Div. 1 in agreement with the
following intrinsically safe
parameters
I
i
L
U
i
U
o
I
o
C
i
+C
cable
C
o
i
+L
cable
L
o
PBA430x, PFA57.x,
PUA579x, PTA459x,
System solution Point Ex
with Point Ex
A/D converter
gray
white
yellow
green
brown
red
blue
black
pink
gray
white
yellow
green
brown
red
blue
black
pink
gray
white
yellow
green
brown
red
blue
black
pink
EB
3)
EB
3)
EB
3)
EB
3)
EB
3)
EB
3)
EB
3)
EB
3)
EB
3)
P: brown
N: blue
power
supply
4), 5)
P: brown
N: blue
power
supply
4), 5)
P: brown
N: blue
GND: gre/yel
power
supply
4), 5)
1), 5)
1), 5)
1), 5)
2), 5)
2), 5)
active
intrinsically safe
active
intrinsically safe
active
intrinsically safe
Non-hazardous
(safe) area
Hazardous area
Class I, II, III Div.1 Groups A-G, T4
C
C
C
PSUx
APS768x
PSUx
APS768x
PSUx
APS768x
B
B
GND: gre/yel
GND: gre/yel
passive
passive
passive
passiveactive
7)
max. 20 m
2), 5)
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40

Mettler Toledo APS768x-120V Guide d'installation

Taper
Guide d'installation
Ce manuel convient également à

dans d''autres langues