Baumer DAB10-AI Guide de démarrage rapide

Taper
Guide de démarrage rapide
Messverstärker
Measuring Amplifier
Amplificateur de mesure
DAB10-AI
Kurzanleitung
Quickstart
Guide rapide
IND. CONT. EQ.
2BD0
Baumer Electric AG
Hummelstrasse 17
CH − 8501 Frauenfeld
( +41 52 728 11 22
Weitere Baumer-Kontakte finden Sie unter:
For further Baumer contacts go to:
Autres contacts Baumer sous :
www.baumer.com
Änderungen vorbehalten
Right of modifications reserved
Modifications réservées
V3, 22.01.2021, 81362964
DE | EN | FR
Masszeichnung DAB10
Dimensional drawing DAB10
Plan coté DAB10
10 62 11,4
12
12
52
3
3
24
14
22
2 x Ø 9,5 x 90°
2 x Ø 4,5
M12 x 1
M8 x 1
Anschlussbild
Connection diagram
Schéma de raccordement
Transducer DAB10-AI Control unit
+V
s
-V
s
S-
S+
L+ / +V
s
DI/DQ
L- / 0V
C/Q
Supply voltage
Current output
I/O
Ground
IO-Link / I/O
I
OUT
max. 10 m max. 10 m
Steckerbelegung
Pin assignment
Affectation des connecteurs
Sensor side 1 +Vs
2 Sig +
3 -Vs
4 Sig -
Control side 1 +Vs/L+
2 Iout
3 0 V/L-
4 C/Q
5 DI/DQ
DE
Mitgeltende Dokumente
n
Als Download unter www.baumer.com:
Betriebsanleitung
Datenblatt
EU-Konformitätserklärung
n
Als Produktbeileger:
Beileger Allgemeine Hinweise (11042373)
Hinweise zur elektrischen Installation
Betriebsspannungsbereich: +Vs = 18 ... 30 VDC
Vor dem Anschliessen des Geräts die Anlage span-
nungsfrei schalten. Netzteil nach UL 1310, Class 2
oder externe Absicherung durch eine UL anerkannte
oder gelistete Sicherung mit max.100W/Vp oder max.
5A unter 20V.
Hinweis zur elektromagnetischen Verträglichkeit: Ge-
schirmtes Anschlusskabel empfohlen. Kabelschirm
beidseitig, grossflächig erden und Potentialausgleich
sicherstellen.
Montage
ð Um eine zuverlässige Dichtigkeit zu gewährleisten,
wird eine ebene Auflagefläche empfohlen (bearbei-
ten Sie bei Bedarf die Auflagefläche vor der Monta-
ge mechanisch).
Vorgehen:
Montieren Sie den Messverstärker mit einem An-
zugsdrehmoment von max. 1 Nm. Verwenden Sie
hierfür die 2 beigelegten Schrauben.
Werkseinstellungen
DAB10-AI
Wert
Eingang (sensorseitig):
Einheit Prozesswert: ppth (mV/V)
Nennkennwert Sensor: 1 mV/V
Offset: 0
Ausgang:
Ausgangssignal (an I
out
): 4 ... 20 mA
SIO1: Inactive
SIO2: Teach-In Offset
Allgemeine Funktionsweise
Der Messverstärker verarbeitet Signale von Sensoren
mit Dehnmessstreifen (DMS). Dazu zählen bspw.
Kraft-, Dehnungs-, Wägesensoren, welche typischer-
weise mit 4 DMS ausgerüstet und zu einer (Wheatsto-
ne-)Vollbrücke verschaltet sind.
Inbetriebnahme
Der Messverstärker-Eingang ist standardmässig auf
1 mV/V eingestellt und somit direkt an die Baumer
Kraftsensoren angepasst. Sie können den Messver-
stärker per Plug & Play in Betrieb nehmen.
Parametrierung
Die integrierte IO-Link Schnittstelle ermöglicht Ihnen,
den Messverstärker an die Applikation und an die Ei-
genschaften des angeschlossenen Sensors anzupas-
sen.
Das IODD-File zum Parametrieren des Messverstär-
kers finden Sie unter:
n
www.baumer.com (-> Produktdetailseite) und
n
ioddfinder.io-link.com
Parametrierung der Sensoreigenschaften
Mit der Parametrierung der Sensoreigenschaften defi-
nieren Sie die Umrechnung des Eingangssignals (mV/
V) in den Prozesswert (N, kN, µm/m, ...). Hierfür ste-
hen folgende Parameter zur Verfügung:
n
Process Value Unit: Einheit des Prozesswertes
n
Nominal Process Value: Nominaler Messbereich des
Prozesswertes
n
Nominal Sensitivity: Nennkennwert des Sensors in
mV/V
n
Offset: Eingangssignal in mV/V, bei dem ein Pro-
zesswert von 0 ausgegeben werden soll
Sie können die Parameter entweder manuell eingeben
oder teachen (Teach-in by Value, Teach-in Offset/ Ta-
re, Teach-in Sensitivity).
Parametrierung des Analogausgangs
Über die Parametrierung des Analogausgangs können
Sie folgende Einstellungen vornehmen:
n
Signal Alarm: Signal bei Alarmzustand (kein gültiges
Signal). Das ist z. B. der Fall, wenn der Wertebe-
reich des Prozesswerts/Eingangssignals "überlauft"
oder der Prozesswert sich ausserhalb des Ausgabe-
bereichs des Analogsignals befindet
n
Signal Adjustment: 2-Punkt-Abgleich von Prozess-
wert und Analogsignal
n
Signal Limits: Anpassung des Ausgabebereichs
Parametrierung des Eingangsfilters
Für das Eingangssignal stehen folgende Filter zur Ver-
fügung:
n
Filter Moving Average für Teach-In: Dieser Filter er-
höht die Abgleichgenauigkeit während des Teach-In.
n
Tiefpassfilter: Dieser Filter ist ein optionaler Filter,
der sich in den folgenden Fällen empfiehlt:
Erhöhung der effektiven Auflösung, indem das
Rauschen reduziert wird
Bandbegrenzung, wenn das Ausgangssignal für
weitere Signalverarbeitung verwendet wird, die
langsamer abtastet
IO-Link Daten (Übersicht)
n
Prozessdaten (zyklisch):
Prozesswert
Scale exponent
SSC 1
SSC 2
Quality
Alarm
n
Einstellbare Daten (azyklisch):
Sensor Adjustment
Teach-in Offset/ Tare
Teach-in Sensitivity
Input-Filter
Analog Output
Source Process Value
SIO1/2 Function Selection
SIO1/2 Input Delay
n
Diverse Diagnosedaten (azyklisch)
EN
Applicable documents
n
Download from www.baumer.com:
Operating manual
Data sheet
EU conformity declaration
n
As a product insert:
General information insert (11042373)
Notes for electric installation
Operating voltage range: +Vs = 18 ... 30 VDC
Disconnect the system from power before connecting
the device. Power supply according to UL 1310,
Class 2 or external protection via a UL-approved or
listed fuse with max. 100 W/Vp or max. 5 A below
20V.
Note on electromagnetic compatibility: Shielded con-
nection cable recommended. Ground the cable shield
on both sides over a large area and ensure potential
equalization.
Installation
ð To ensure reliable leak tightness, we recommend
using a level contact surface (if required treat the
contact surface mechanically prior to the installati-
on).
Instruction:
Install the measurement amplifier with a torque of
max. 1 Nm. Use the 2 included screws for this pur-
pose.
Factory settings
DAB10-AI
Value
Input (sensor side):
Process value unit: ppth (mV/V)
Nominal specific sensor value: 1 mV/V
Offset: 0
Output:
Output signal (on I
out
): 4 ... 20 mA
SIO1: Inactive
SIO2: Teach-In Offset
General functionality
The measurement amplifier processes signals from
sensors with strain gauges. These include, for exam-
ple, force, strain, and weighing sensors that are typi-
cally equipped with 4 strain gauges and connected into
a (Wheatstone) full bridge.
Commissioning
The measurement amplifier input is set as standard to
1mV/V and thus directly matched to the Baumer force
sensors. You can commission the measurement ampli-
fier via plug & play.
Parameterization
The integrated IO-Link interface allows you to adjust
the measurement amplifier to the application and the
properties of the connected sensor.
You can find the IODD file for parameterizing the mea-
surement amplifier at:
n
www.baumer.com (-> product detail page) and
n
ioddfinder.io-link.com
Parameterization of the sensor characteristics
With the parameterization of the sensor characteristics
you can define the conversion of the input signal (mV/
V) to the process value (N, kN, µm/m, ...). The follow-
ing parameters are available for this purpose:
n
Process Value Unit: unit of the process value
n
Nominal Process Value: nominal measurement
range of the process value
n
Nominal Sensitivity: nominal specific value of the
sensor in mV/V
n
Offset: input signal in mV/V at which a process value
of 0 should be output
You can either enter the values manually or teach
them in (Teach-in by Value, Teach-in Offset/ Tare,
Teach-in Sensitivity).
Parameterization of the analog output
With the parameterization of the analog output you can
carry out the following settings:
n
Signal Alarm: signal for the alarm status (no valid
signal). This is the case, e.g., when the value range
of the process value / input signal “flows over” or the
process value is outside the output range of the ana-
log signal
n
Signal Adjustment: 2-point adjustment of process
value and analog signal
n
Signal Limits: adjustment of the output range
Parameterization of the input filter
The following filters are available for the input signal:
n
Moving Average filter for teach-in: This filter in-
creases the calibration accuracy during the teach-in.
n
Deep pass filter: This is an optional filter that is rec-
ommended in the following cases:
Increasing the effective resolution by reducing the
noise
Bandwidth limitation when the output signal is used
for other signal processing that scans more slowly
IO-Link data (overview)
n
Process data (cyclical):
Process value
Scale exponent
SSC 1
SSC 2
Quality
Alarm
n
Adjustable data (cyclical):
Sensor Adjustment
Teach-in Offset/ Tare
Teach-in Sensitivity
Input-Filter
Analog output
Source Process Value
SIO1/2 Function Selection
SIO1/2 Input Delay
n
Various diagnosis data (non-cyclical)
FR
Documents valables
n
Téléchargement sous www.baumer.com:
Mode d’emploi
Fiche technique
Déclaration de conformitéUE
n
En tant qu’annexe du produit:
Remarques générales supplémen-
taires(11042373)
Remarques sur l’installation électrique
Tension de service: +Vs = 18 ... 30VCC
Mettre l’installation hors tension avant de raccorder
l’appareil. Bloc d’alimentation selon UL1310, classe2
ou protection externe par un fusible homologué ou listé
UL avec max.100W/Vp ou max. 5A à 20V.
Remarque concernant la compatibilité électromagné-
tique: câble de connexion blindé recommandé. Effec-
tuer une mise à la terre sur une grande surface aux
deux extrémités du blindage du câble et assurer la liai-
son équipotentielle.
Montage
ð Afin d’assurer une étanchéité fiable, il est recom-
mandé d’utiliser une surface d’appui plane (au be-
soin, modifier la surface d’appui mécaniquement
avant le montage).
Procédure :
Montez l’amplificateur de mesure avec un couple de
serrage de 1 Nm maximum. Utilisez pour ce faire les
2 vis fournies.
Paramètres par défaut
DAB10-AI
Valeur
Entrée (côté détecteur) :
Unité valeur de process : ppth (mV/V)
Valeur nominale caractéristique du capteur : 1 mV/V
Offset: 0
Sortie :
Signal de sortie (an I
out
): 4 ... 20 mA
SIO1: Inactive
SIO2: Teach-In Offset
Fonctionnement général
L’amplificateur de mesure traite les signaux des cap-
teurs grâce aux jauges de contraintes (DMS). Parmi
ces détecteurs, on compte par ex. les capteurs de
force, de contrainte, de pesage, typiquement équipés
de 4jauges de contrainte et interconnectés à un pont
plein de Wheatstone.
Mise en service
L’entrée de l’amplificateur de mesure est pré-réglée
par défaut sur 1mV/V et est ainsi directement adaptée
aux capteurs de force (DLM). Vous pouvez mettre en
service votre amplificateur de mesure via Plug & Play.
Paramétrage
L’interface intégrée IO-Link vous permet d’adapter
l’amplificateur de mesure à l’application et aux caracté-
ristiques du capteur raccordé.
Vous trouverez ci-dessous le fichier IODD permettant
de paramétrer l’amplificateur de mesure:
n
www.baumer.com (-> page détaillant le produit) et
n
ioddfinder.io-link.com
Paramétrage des caractéristiques du détecteur
Le paramétrage des caractéristiques du détecteur
vous permet de définir la conversion du signal d’entrée
(mV/V) en valeur de process (N, kN, µm/m, ...). Pour
ce faire, différents paramètres sont à votre disposition:
n
Process Value Unit: unité de la valeur de process
n
Nominal Process Value: plage de mesure nominale
de la valeur de process
n
Nominal Sensitivity: valeur nominale caractéristique
du capteur en mV/V
n
Offset: signal d’entrée en mV/V pour lequel une va-
leur de process de 0 doit être émise
Vous pouvez soit saisir manuellement les paramètres,
soit effectuer un apprentissage (Teach-in by Value,
Teach-in Offset/ Tare, Teach-in Sensitivity).
Paramétrage de la sortie analogique
Via le paramétrage de la sortie analogique, vous pou-
vez procéder aux réglages suivants:
n
Signal Alarm : signal pour l’état de l’alarme (aucun
signal valable). Se produit par ex. lorsque la plage
de valeurs de la valeur de process/du signal d’en-
trée «déborde» ou lorsque la valeur de process se
trouve en-dehors de la plage d’émission du signal
analogique
n
Signal Adjustment: comparaison à 2points de la va-
leur de process et du signal analogique
n
Signal Limits: adaptation de la plage d’émission
Paramétrage du filtre d’entrée
Pour le signal d’entrée, le filtre suivant est disponible:
n
Filtre Moving Average pour Teach-In: ce filtre aug-
mente la précision d’ajustement pendant le proces-
sus Teach-In.
n
Filtre passe-bas: il s’agit d’un filtre optionnel recom-
mandé dans les cas suivants:
Augmentation de la résolution effective en rédui-
sant le grésillement
Limitation de la bande lorsque le signal de sortie
est utilisé pour un nouveau traitement du signal qui
balaie plus lentement
Données IO-Link (vision d’ensemble)
n
Données de processus (cycliques) :
Valeur de process
Scale exponent
SSC 1
SSC 2
Quality
Alarm
n
Données réglables (acycliques) :
Sensor Adjustment
Teach-in Offset/ Tare
Teach-in Sensitivity
Input-Filter
Analog Output
Source Process Value
SIO1/2 Function Selection
SIO1/2 Input Delay
n
Données de diagnostic diverses (acycliques)
  • Page 1 1
  • Page 2 2

Baumer DAB10-AI Guide de démarrage rapide

Taper
Guide de démarrage rapide