Mettler Toledo EasyClean200e cleaning and calibration system Mode d'emploi

Taper
Mode d'emploi
EasyClean 200e
30 404 676 F
EasyClean 200e
Instruction manual
Bedienungsanleitung
Instructions d’utilisation
English: page 3
Deutsch: Seite 43
Français: page 83
EasyClean 200e 3
EasyClean 200e 02/2021 © METTLER TOLEDO
30 404 676 F Printed in Switzerland
EasyClean 200e
Instruction manual
4 EasyClean 200e
02/2021 © METTLER TOLEDO EasyClean 200e
Printed in Switzerland 30 404 676 F
How to use this instruction manual
This instruction manual is an integral part of the EasyClean 200e and contains notes
and instructions that are important for safety and operation.
All persons working on or with the EasyClean 200e must have first read and understood
the sections appropriate to the work in hand.
Please read this instruction manual carefully before using the EasyClean system. Keep this
document close to the unit, so operating personnel may be easily refer to it at any time.
First read section 2 “Safety”!
© It is forbidden to reprint this instruction manual in whole or part.
No part of this manual may be reproduced in any form, or modified, copied or distributed using electronic systems,
in particular in the form of photocopies, photographs, magnetic or other recordings, without written consent of
Mettler-Toledo GmbH, Process Analytics, CH-8902 Urdorf.
All rights reserved, in particular reproduction, translation and patenting/registration.
EasyClean is a trademark of the METTLER TOLEDO Group.
InDip is a registered trademark of the METTLER TOLEDO Group in Switzerland, Brazil, Belgium, Netherlands,
Luxembourg, Germany, France and Italy.
InTrac is a registered trademark of the METTLER TOLEDO Group in Switzerland, Brazil, UK, USA, Austria, Belgium,
Netherlands, Luxembourg, Germany, France, Italy and Russia.
EasyClean 200e 5
EasyClean 200e 02/2021 © METTLER TOLEDO
30 404 676 F Printed in Switzerland
Content
How to use this instruction manual 4
1 Product description 7
1.1 Introduction 7
1.2 Fittings and deliverables 7
1.3 Data and key characteristics EasyClean 200 e 8
2 Safety 10
2.1 Introduction 10
2.2 Intended use 10
2.3 Inappropriate use 10
2.4 Basic principles 11
2.5 Warning notices and symbols 11
2.6 Responsibility, organizational measures 12
2.7 Product-specific hazards 13
2.8 Residual hazards 13
2.9 Emergency measures 13
2.10 Safety measures 14
2.11 Modifications 14
3 Your device 15
3.1 Layout 15
3.2 Functions 17
3.3 Integration into a measurement system 18
4 Commissioning 19
4.1 Fitting 19
4.2 Connections 21
4.3 Commissioning procedure 28
4.4 Function test 29
4.5 Programming of control of EasyClean 200 e 30
4.5.1 User interface of the programmable control 30
4.5.2 SW parameter description 31
5 Operation 33
5.1 Checks by the operator 33
5.2 Refilling with cleaning fluid 33
5.3 Operating the EasyClean 200 e 33
5.4 Procedure in the event of a failure in the power or fuel supply 35
6 Maintenance and fault elimination 36
6.1 Maintenance 36
6.2 Fault elimination and repairs 36
7 Decommissioning, storage, disposal 39
7.1 Decommissioning 39
7.2 Storage 39
7.3 Correct disposal of the unit 39
8 Ordering information (product, spare parts, accessories) 40
8.1 Ordering information for EasyClean 200 e 40
6 EasyClean 200e
02/2021 © METTLER TOLEDO EasyClean 200e
Printed in Switzerland 30 404 676 F
EasyClean 200e 7
EasyClean 200e 02/2021 © METTLER TOLEDO
30 404 676 F Printed in Switzerland
1 Product description
1.1 Introduction
C
ertificates CE
I
dentification
1.2 Fittings and deliverables
Deliverables EasyClean 200e is shipped without connecting cables. At the installation site, cables are
connected to the appropriate terminals in the unit (please refer to section 4.2 “Connections”).
Packaging The unit is packed in plain and padded cardboard.
We recommend that you retain the packing materials for subsequent storage or transportation
of the unit. Should you choose to discard the packing materials, observe local regulations and
section 7.3 “Correct disposal of the unit”.
Good inspection Check the goods for damage while unpacking. Should you notice signs of damage,
inform the shipper and your supplier without delay.
Check the delivery against your order and the delivery note.
w
ww.mt.com/pro
M
ettler-Toledo GmbH
I
m Hackacker 15
8
902 Urdorf
S
witzerland
S
/N: 1234567890
P/N: 52403776
M
ODEL: EasyClean 200e
PWR: 100 - 230VAC 50/60Hz,
0.18 - 0.3A
M
ETTLER TOLEDO
w
0.18 - 0.3A
15
8 EasyClean 200e
02/2021 © METTLER TOLEDO EasyClean 200e
Printed in Switzerland 30 404 676 F
1.3 Data and key characteristics EasyClean 200e
Enclosure Stainless steel enclosure, powder-coated with polyester resin
Attention: housing may not be cleaned with solvents
containing acetone
Weight Control unit approx. 3.0 kg (6.6 lb)
Valve unit approx. 2.5 kg (5.5 lb)
Protection rating IP65
Ambient conditions Temp. in operation: 0…+55°C (+32…+131°F)
Temp. storage/transportation: –10…+70°C (+14…+158°F)
Power supply 100…230V AC, 50/60 Hz, 0.18…0.3 A (± 15%)
System control Process initiation:
via a wash contact in the associated transmitter
Sequence:
controlled by the integrated programable PLC
Control input 1 control input for wash contact in transmitter 24V
Transmitter METTLER TOLEDO transmitters:
M300 Process
M400 4-wire
Alternatively via any standard transmitter with wash
contact and “HOLD” function
Electrical connections Terminal strip 0.08…2.5 mm
2
Connection to valve unit 10-wire control cable, 1.5 mm
2
Length: 1.5 m
Connection transmitter 4-wire control cable, 43 1.5 mm
2
Length: 5 m
Cable ducts 3.5…10 mm Æ M16
2.5…6.5 mm Æ M12
Compressed air supply 4…8 bar (400…800 kPa, 58…116 psi)
Air quality according to ISO 8573-1
Moisture class 4 (water dew point +3°C)
Particles class 5 (filter 40 µm)
Max. oil content class 2 (0.1 mg/m
3
)
Connection: thread G
1
/8"
EasyClean 200e 9
EasyClean 200e 02/2021 © METTLER TOLEDO
30 404 676 F Printed in Switzerland
Connection to Pneumatics: LDPE hose 6/4 mm (
1
/4"/
1
/8")
retractable housing (control air)
Medium: PTFE hose 6/4 mm (
1
/4"/
1
/8")
Hose length typ. 5 m (16.4 ft)
Hose length max. 10 m (32.8 ft)
Flushing water supply Pressure: 2…8 bar (200…800 kPa, 29…116 psi)
Particle filter: 50 µm
Volume: $4 l/min
Connection: thread G
1
/4" female
Adapter:
1
/4" NPT male (PP)
Pump Pump height: max. 5 m (16.4 ft)
Suction height: max. 3 m (9.8 ft)
Feed rate: 3…5 ml/s (depending on hose length
and medium viscosity)
Materials: PPS (pump body)
EPDM (membrane and sealings)
Connection: thread G
1
/8" female, polymer (PP)
Viscosity All liquids 25 mPa s/25°C (77°F)
Wetted materials Materials in contact with:
– Rinsing water: PP, EPDM, PFTE, PPS, Ni
– Compressed air: PP, EPDM, PFTE, PPS, LDPU, Ni
– Cleaning agents: PP, FFKM, PTFE, EPDM, PEEK
Liquid temperatures Water: 0…80°C, 95°C max. 10 min/h
(32…176°F, 203°F max. 10 min/h)
Cleaning agents: 0…50°C (32…122°F)
Subject to technical changes.
10 EasyClean 200e
02/2021 © METTLER TOLEDO EasyClean 200e
Printed in Switzerland 30 404 676 F
2 Safety
2.1 Introduction
This instruction manual lists the main points of how to use the EasyClean 200e efficiently,
in the intended manner. For safe, trouble-free operation, operators must be aware of these safety
instructions and comply with further warning notices in the instruction manual.
This instruction manual, and the safety requirements in particular, must be observed by the
operator and all other persons working with the EasyClean 200e. Compliance with local
regulations and statutory requirements is also essential.
The instruction manual must always be stored close at hand, in a place accessible to all
people working with the EasyClean 200e.
2.2 Intended use
EasyClean 200e is intended solely for automatically controlled or manual flushing and cleaning
of sensors in retractable housings, as described in this instruction manual.
Other prerequisites for appropriate use include:
observing the instructions, notes and requirements set out in this instruction manual
observing the prescribed inspection and maintenance intervals
correct care and maintenance of the unit, according to the instruction manual
observing the prescribed environmental and operational conditions, and permitted
installation positions
compliance with local statutes.
2.3 Inappropriate use
Any other use, or use not mentioned here, that is incompatible with the technical specifications
is deemed inappropriate. The operator is solely responsible for any damage arising from such
use.
The following are deemed particularly inappropriate:
use of hydraulic fluid in place of compressed air
use of suspensions as cleaning media
use of salt acid of > 10%.
EasyClean 200e 11
EasyClean 200e 02/2021 © METTLER TOLEDO
30 404 676 F Printed in Switzerland
2.4 Basic principles
The EasyClean system is a state-of-the-art construction and complies with engineering safety
g
uidelines.
All the same, there can remain certain risks and hazards:
if the units are operated by inadequately trained personnel.
if the units are not used in the intended manner.
if the units are not inspected periodically or the periodical inspection
is not carried out properly.
EasyClean units must only be used while in perfect working order, for their intended purpose,
with due awareness of associated safety and hazard issues and in compliance with the
instruction manual.
Defects or damage that impact the unit’s safety and functionality must be immediately repaired
by the operator or a suitably qualified person, and reported in writing to the manufacturer.
2.5 Warning notices and symbols
This instruction manual identifies safety instructions by means of the following symbols:
DANGER
Warns of a hazard that could lead to extensive material damage, death or grievous bodily
injuries.
CAUTION
Warns of a potentially hazardous situation that could lead to minor bodily injury and/
or material damage.
NOTICE
Information relating to technical requirements. Non-observance could lead to defects,
inefficient operation and possible loss of production.
12 EasyClean 200e
02/2021 © METTLER TOLEDO EasyClean 200e
Printed in Switzerland 30 404 676 F
2.6 Responsibility, organizational measures
Operator’s
responsibilities
The operator must only permit personnel to work with the EasyClean if they are acquainted
with basic regulations concerning workplace safety and accident prevention, and familiar
with operating the unit. This instruction manual serves as a basis.
In addition to the instruction manual, the operator must provide instruction in general statu-
tory and other binding regulations concerning workplace safety and accident prevention,
as well as environmental protection.
Personnel safety and hazard awareness must be monitored regularly.
Steps must be taken to ensure that the EasyClean is operated only while in a safe, properly
functional state.
Personnel
responsibilities
All persons working with the EasyClean must have read the section on safety and respon -
sibilities the warning notices contained in this instruction manual.
In addition to the instruction manual, generally applicable statutory and other binding
regulations concerning workplace safety and accident protection must also be observed.
Avoid any hazardous working practices that exceed the unit’s intended purpose.
Personnel selection and
qualification
– basic responsibilities
Only dependable, trained or supervised personnel may work with the EasyClean.
Operators must read this instruction manual beforehand.
Responsibilities for operating, maintaining, installing, etc. the unit must be clearly defined.
Appropriate steps must be taken to ensure that the unit is used by authorized personnel only.
Personnel under instruction, supervision or general training may only work with the unit
under constant supervision by an experienced person.
Work on electrical parts of the unit may only be performed by a qualified electrician in
compliance with electrical engineering regulations.
Constructional
modifications on the unit
The unit may not be modified, extended or rebuilt in a way that impacts its safety without
written permission from Mettler-Toledo GmbH, Process Analytics.
Any parts that are not in perfect working order must be replaced without delay.
EasyClean 200e 13
EasyClean 200e 02/2021 © METTLER TOLEDO
30 404 676 F Printed in Switzerland
2.7 Product-specific hazards
D
ANGER
Electrical hazards!
O
pening the unit for installation or repair work exposes live electrical parts.
C
lose the unit immediately after work is completed!
DANGER
Compressed air/water hazards!
Pressurized pipes (compressed air, water) must be isolated and depressurized before
commencing repair work!
2.8 Residual hazards
DANGER
Despite all precautionary measures, there remain residual hazards such as:
a pipe bursting
hazards arising from a defect in a supervisory controller.
2.9 Emergency measures
DANGER
Should fluid come into contact with the eyes, rinse thoroughly under running water
for at least 15 minutes. A doctor must be consulted for subsequent diagnosis.
Fluid coming into contact with the skin must be immediately washed away with water.
Medical assistance is mandatory for symptoms that clearly arise from skin or eye
contact with a medium. Inform the doctor about the type and specific identity of all
materials used.
14 EasyClean 200e
02/2021 © METTLER TOLEDO EasyClean 200e
Printed in Switzerland 30 404 676 F
2.10 Safety measures
Always observe local statutes and regulations. They are not incorporated in this instruction
manual.
Personal safety equipment like protective goggles and clothing must be worn.
T
he unit must only be operated while closed.
The operator is responsible for personnel instruction. By way of assistance, the instruction
manual can also be ordered in other languages. Being an integral part of the unit, this instruc-
tion manual must always be available at the unit’s location.
The operator should immediately inform the manufacturer about all safety-related events
occurring during use of the unit.
2.11 Modifications
Modifications are permitted only after prior consultation and written approval from the
manufacturer.
Only those parts of the unit essential to fulfilling its function may be removed and refitted.
Except for authorized service personnel, it is forbidden to remove protective guards and
covers using a regular tool. Touching live electrical parts represents a lethal hazard.
EasyClean 200e 15
EasyClean 200e 02/2021 © METTLER TOLEDO
30 404 676 F Printed in Switzerland
3 Your device
3.1 Layout
Control unit
1 “In/Out” button (housing control)
2 “Water” button
3 “Clean” button
4 Information display
5 Power supply unit
E
asyClean 200e
6 Programmable logic controller
7 Operating section PLC
8 Connection clamps
9 Remote PCB
1 2 3
4
5 6 7
9
8
16 EasyClean 200e
02/2021 © METTLER TOLEDO EasyClean 200e
Printed in Switzerland 30 404 676 F
V
alve unit (figure shows valve unit EasyClean 200e)
1 Compressed air valve
2 Water valve
3 Pressure reducing valve
4 Detergent valve
5 Housing control valve
6 Diaphramgm pump
6
1 2 3 4 5
EasyClean 200e 17
EasyClean 200e 02/2021 © METTLER TOLEDO
30 404 676 F Printed in Switzerland
3.2 Functions
General The EasyClean 200e is designed for the automatic rinsing and cleaning of electrodes in
r
etract able housings.
The EasyClean is supplied with a predefined rinsing and cleaning process. Both the rinsing
and cleaning time can be adjusted by the operator to suit the individual local conditions.
The individual processes are instigated by switching on the washing contact on the connected
transmitter.
An integrated programmable logic controller controls the whole rinsing and cleaning procedure.
Types of operation The EasyClean has no main switch, and is always in active status.
Basic setting When the EasyClean 200e is switched off, the electrode is removed from the process medium.
If the EasyClean 200e is connected to the mains, the housing and sensor move into
the measuring position.
The message “RUN” appears on the display. The EasyClean is ready for operation.
“HOLD” function The “HOLD” function is a function of the transmitter. This function suppresses the measured
values of the sensor during a work cycle of the EasyClean.
NOTICE
If the transmitter is usually used with exit points, a “HOLD” function is absolutely necessary.
Signals that can be measured through cleaning with acids or alkalis might otherwise activate
the transmitter control.
Messages
The integrated control is equipped with a display.
This displays the current function status of the control.
18 EasyClean 200e
02/2021 © METTLER TOLEDO EasyClean 200e
Printed in Switzerland 30 404 676 F
3.3 Integration into a measurement system
T
he following diagram show typical examples of the integration of the EasyClean 200e
respectively into a measurement system.
EasyClean 200e
1 Control unit
2 Valve unit
3 Connection cable control unit > valve unit
4 Water connection
5 Drainwater hose
6 Connecting hose valve unit > cleaning medium
7 Connecting hose valve unit > retractable housing
8 Wall mounting kit
9 Compressed air hose inlet
10 Compressed air hoses
11 Control cable transmitter > control unit
12 Pneumatic set retractable housing
13 Panic button (optional)
14 Canister for cleaning medium
15 Remote control
16 Alarm light
11
16
1
3
3
8
1
1
5
9
4
2
10
7
5
12
6
14
Panic button
«Panik»-Knopf
Bouton «anti-panique»
10 0 – 230 VA C
20 – 2 53 VAC
EasyClean 200e 19
EasyClean 200e 02/2021 © METTLER TOLEDO
30 404 676 F Printed in Switzerland
DANGER
Please refer to section 2 “Safety”.
All work on the system’s electrical components may only be performed by authorized,
s
pecialist staff.
4.1 Fitting
The EasyClean 200e is designed to be fitted directly (c-rails), on to walls or posts/pipes.
The EasyClean 200e is built into the same type of housing and therefore have the
same dimensions.
The fitting location must have sufficient stability and must as far as possible be free of vibration.
Please make sure that the lengths of the leads are as short as possible.
4 Commissioning
Wall mounting
20 EasyClean 200e
02/2021 © METTLER TOLEDO EasyClean 200e
Printed in Switzerland 30 404 676 F
Post/pipe mounting
NOTICE
For the detailed procedure when fitting the suspended item, please refer to the installation
manual supplied separately.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114
  • Page 115 115
  • Page 116 116
  • Page 117 117
  • Page 118 118
  • Page 119 119
  • Page 120 120
  • Page 121 121
  • Page 122 122
  • Page 123 123
  • Page 124 124

Mettler Toledo EasyClean200e cleaning and calibration system Mode d'emploi

Taper
Mode d'emploi