Covidien Nellcor SCNEO Manuel utilisateur

Taper
Manuel utilisateur

Ce manuel convient également à

Indications/Contraindications
The Nellcor™ Neonatal SpO
2
Sensor Non-Adhesive, model
SCNEO, is indicated for single patient use when continuous
noninvasive arterial oxygen saturation and pulse rate moni-
toring are required for neonates weighing 1.5 kg to 5 kg.
Use this sensor only with Nellcor instruments and
instruments containing Nellcor OxiMax™ technology,
or with instruments licensed to use Nellcor OxiMax sen-
sors (Nellcor-compatible instruments). When connected
to an OxiMax - compatible instrument, this sensor uses
OxiMax technology to provide additional advanced
sensor performance features. Consult individual manu-
facturers for features and compatibility of particular
instruments and sensor models.
Each manufacturer of Nellcor-compatible instruments
is responsible for determining whether and under
what conditions its instruments are compatible for safe
and effective use with each Nellcor sensor model. This
may include different specifications and/or warnings,
cautions, or contraindications. Refer to the instru-
ment operator’s manual or consult manufacturer for
complete instructions for use of this sensor with their
Nellcor-compatible instrument.
Note: Non-adhesive sensor consists of sensor, cable,
wrap and loop fastener.
Instructions for Use
1) Locate transparent windows on the sensor.
Windows cover optical components. This is
the side that contacts the patient. Note corre-
sponding alignment marks (a) on the outside
(b) between the marks .
2) Orient the SCNEO so the graphics are on the
lateral edge of the site:
The preferred site is a foot. Alternatively, use a
hand. The window next to the cable goes on
the sole of the foot as shown .
Note: When selecting a sensor site, priority
should be given to an extremity free of an
arterial catheter, blood pressure cuff, or intra-
vascular infusion line.
3) Wrap the SCNEO firmly, but not too tightly
around the foot. Windows must oppose each
other. Secure the hook and loop fastener.
4) Use the cable wrap—attached to the sensor
cable—to secure the sensor cable to the patient’s
ankle. Secure with the hook and loop fastener.
5) Plug the SCNEO into the oximeter and verify
proper operation as described in the oximeter
operator’s manual.
Note: If the sensor does not track the pulse
reliably, it may be incorrectly positioned—or
the sensor site may be too thick, thin, or
deeply pigmented, or otherwise deeply
colored (for example, as a result of externally
applied coloring such as nail polish, dye, or
pigmented cream) to permit appropriate light
transmission. If any of these situations occurs,
reposition the sensor or choose an alternate
Nellcor sensor for use on a different site.
Reapplication
The SCNEO can be reused on the same patient as long as
the sensor attaches without slippage.
Indications/Contre-indications
Le capteur SpO
2
non adhésif pour nouveau-nés
Nellcor™, modèle SCNEO, est indiqué chez les nouveau-
nés pesant 1,5 à 5 kg qui nécessitent un monitorage
en continu non effractif de la saturation en oxygène
artériel et de la fréquence du pouls, dans le cadre d’une
utilisation chez un seul patient.
Ce capteur doit être utilisé seulement avec les
instruments Nellcor et les instruments équipés
de la technologie Nellcor OxiMax™ ou ayant une
licence d’utilisation des capteurs Nellcor OxiMax
(instruments compatibles Nellcor). Lorsqu’il est
connecté à un instrument compatible OxiMax, ce
capteur utilise la technologie Nellcor OxiMax qui
lui confère des caractéristiques de performances
avancées supplémentaires. Consulter chaque
fabricant pour connaître la compatibilité des
différents instruments et modèles de capteurs.
Il est de la responsabilité de chaque fabricant
d’instruments compatibles Nellcor, de déterminer
dans quelles conditions ses instruments sont
compatibles pour une utilisation efficace et en toute
sécurité avec chaque modèle de capteur Nellcor.
Différentes spécifications et/ou avertissements,
précautions à prendre ou contreindications peuvent
par exemple, être requis. Pour obtenir des instructions
complètes sur l’utilisation de ce capteur avec les
instruments compatibles Nellcor, se référer au manuel
de l’utilisateur de l’instrument ou consulter le fabricant.
Remarque : Le capteur non adhésif comporte un
capteur, un câble et un système de fixation à ergot.
Mode d’emploi
1) Repérer les fenêtres transparentes sur le cap-
teur. Elles recouvrent les composants
optiques. Le côté où elles se trouvent est
le côté en contact avec le patient. Noter les
repères d’alignement correspondants (a) sur le
côté extérieur (b) entre ces repères .
2) Orienter le SCNEO de sorte que les graphiques
se trouvent sur le bord du site :
Le pied constitue le site préférentiel, mais la
main peut également être utilisée. La fenêtre
située à proximité du câble doit se trouver sur
la plante du pied, comme indiqué .
Remarque : Le site du capteur doit en priorité
être un membre sans cathéter artériel, brassard
à tension ou voie veineuse de perfusion.
3) Envelopper le SCNEO fermement mais sans
serrer autour du pied. Les fenêtres doivent
être à l’opposé l’une de l’autre. Attacher le
crochet à la boucle du fermoir.
4) Utiliser la bande de fixation attachée au câble
du capteur pour fixer ce câble à la cheville du
patient. Attacher le crochet à la boucle du
fermoir.
5) Introduire le raccord du SCNEO dans l’oxy-
mètre et s’assurer du fonctionnement correct
conformément au manuel d’utilisation de
l’oxymètre.
Remarque : Si le capteur ne peut détecter le
pouls de manière fiable, cela signifie qu’il est
probablement mal positionné, ou que le site
du capteur est trop épais, trop fin ou encore
trop profondément pigmenté ou coloré (par
une coloration externe telle qu’un vernis à
ongles, un colorant ou une crème pigmentée,
par ex.) pour permettre une transmission
correcte de la lumière. Dans ces cas, reposi-
tionner le capteur ou choisir un autre capteur
Nellcor applicable sur un autre site.
Réutilisation
Le SCNEO peut être réutilisé sur le même patient aussi
longtemps que le capteur reste enplace sans glisser.
Warnings
1) Failure to apply the SCNEO properly may
cause incorrect measurements.
2) High oxygen levels may predispose a prema-
ture infant to develop retinopathy. Therefore,
the upper alarm limit for oxygen saturation
must be carefully selected in accord with ac-
cepted clinical standards and considering the
accuracy range of the oximeter being used.
3) While the SCNEO is designed to reduce the
effects of ambient light, excessive light may
cause inaccurate measurements. In such
cases, cover the sensor with opaque material.
4) Circulation distal to the sensor site should be
checked routinely. The site must be inspected
every 8 hours to ensure skin integrity, and
correct optical alignment. If skin integrity
changes, move the sensor to another site.
5) Intravascular dyes or externally applied color-
ing such as nail polish, dye, or pigmented
cream, may lead to inaccurate measurements.
6) Excessive motion may compromise performance.
In such cases, try to keep the patient still, or
change the sensor site to one with less motion.
7) Do not immerse in water or cleaning solu-
tions. Do not resterilize.
8) If the sensor is wrapped too tightly or supple-
mental tape is applied, venous pulsations may
lead to inaccurate saturation measurements.
9) Do not use the SCNEO or other oximetry sen-
sors during MRI scanning. Conducted current
may cause burns. Also, the SCNEO may affect
the MRI image, and the MRI unit may affect
the accuracy of oximetry measurements.
10) Do not alter or modify the SCNEO. Alterations
or modifications may affect performance or
accuracy.
11) For additional warnings, cautions or contrain-
dications when using this sensor
with Nellcor-compatible instruments, refer to
the instrument operator’s manual or contact
the manufacturer of the instrument.
12) In the event of damage to the sterile packag-
ing, do NOT resterilize. Follow local governing
ordinances and recycling instructions regard-
ing disposal or recycling of sensors.
13) If the sensor is misapplied with excessive
pressure for prolonged periods, a pressure
injury can occur.
Accuracy Specifications
For the accuracy specification range when used with
Nellcor monitors, refer to information provided with the
monitor, or (in the U.S.), contact Covidien’s Technical
Services Department. Outside the U.S., contact your
local Covidien representative.
For the accuracy specification range of this sensor when
used with Nellcor-compatible instruments, refer to the
instrument operator’s manual or contact the instru-
ment manufacturer.
Additional Copies of Instructions
Additional copies of these Instructions are available at
no charge by calling Covidien or its authorized distribu-
tors. Also, permission is hereby granted under Covidien
copyrights to purchasers of products obtained from Co-
vidien or its authorized distributors to make additional
copies of these instructions for use by such purchasers.
Any other use of this sensor is not authorized by Covi-
dien under any patent.
Mises en garde
1) Une mauvaise application du SCNEO peut
entraîner une inexactitude dans les mesures.
2) Une augmentation significative du taux
d’oxygène dans le sang peut provoquer, chez
le prématuré, une rétinopathie. C’est la raison
pour laquelle les limites d’alarme de la SpO
2
doivent être conformes aux normes cliniques
reconnues en tenant compte de la précision
de l’oxymètre utilisé.
3) Bien que le SCNEO soit conçu pour réduire les
effets de la lumière ambiante, une lumière
excessive peut affecter l’exactitude des mesures.
Dans ce cas, recouvrir le capteur d’un tissu
opaque.
4) La circulation en aval su site du capteur doit
être contrôlée régulièrement. Le site doit être
examiné toutes les 8 heures afin de vérifier
l’intégrité de la peau et le bon alignement
optique. Si l’intégrité de la peau est modifiée,
déplacer le capteur sur un autre site.
5) ILa présence de colorants intravasculaires ou
externes tels qu’un vernis à ongles ou une
crème pigmentée, peut affecter l’exactitude
des mesures.
6) Des mouvements excessifs peuvent
compromettre les résultats. Dans ce cas,
essayer de calmer le patient ou choisir un
autre site où les mouvements sont limités.
7) Ne pas immerger le capteur dans l’eau ni dans
des produits de nettoyage. Ne pas restériliser.
8) Si le capteur est trop serré ou si une bande
supplémentaire est utilisée, les pulsations
veineuses peuvent entraîner des mesures de
saturation inexactes.
9) Ne pas utiliser le SCNEO ou d’autres capteurs
d’oxymétrie pendant l’exploration par IRM. Le
courant de conduction peut provoquer des
brûlures. Le SCNEO peut altérer la qualité de
l’image et l’unité IRM peut affecter la précision
des mesures de l’oxymètre.
10) Ne pas adapter ou modifier le SCNEO; cela
pourrait affecter les résultats ou la précision.
11) Pour les avertissements, précautions à prendre
ou contre-indications additionnels lorsque
ce capteur est utilisé avec des instruments
compatibles Nellcor, se référer au manuel de
l’utilisateur de l’instrument ou contacter le
fabricant de l’instrument.
12)
ISi l’emballage stérile est endommagé, NE PAS
restériliser. Se conformer à la réglementation
et aux instructions locales en vigueur
concernant la destruction et le recyclage des
capteurs.
13) Une blessure par pression peut se produire
si le capteur est mal utilisé avec une pression
excessive sur des périodes prolongées.
Spécifications relatives à la précision
Pour la gamme de spécification de précision lorsqu’il
est utilisé avec les moniteurs Nellcor, reportez-vous aux
informations fournies avec le moniteur, ou (aux États-
Unis), veuillez communiquer avec Covidien Département
des Services Techniques. En dehors des Etats-Unis,
contactez notre Covidien service technique.
Pour la plage de spécifications de précision de ce
capteur lorsqu’il est utilisé avec des instruments com-
patibles Nellcor, se référer au manuel de l’utilisateur de
l’instrument ou contacter le fabricant de l’instrument.
Copies additionnelles des Instructions
Des copies supplémentaires de ces modes d’emploi
sont disponibles gratuitement auprès de Covidien ou
de ses représentants agréés. De plus, les acheteurs
des produits obtenus auprès de Covidien ou de ses
distributeurs agréés ont l’autorisation de réaliser des
copies supplémentaires de ces modes d’emploi pour
leur usage personnel dans le respect des droits d’auteur
de Covidien.
Jede andere Verwendung dieses Sensors ist nicht von
Covidien im Rahmen von keinen Patenten autorisiert.
Neonatal SpO
2
Sensor Non-Adhesive
Directions for Use
Capteur SpO
2
non adhésif pour nouveau-nés
Mode d’emploi
Nellcor
TM
Neonatal SpO
2
Sensor
Non-Adhesive
SCNEO
SCNEO-I
This product cannot be adequately cleaned and/or sterilized by the user in order to facilitate safe reuse, and is
therefore intended for single use. Attempts to clean or sterilize these devices may result in a bio-incompatibility,
infection or product failure risks to the patient.
This device is not made with natural rubber latex or DEHP.
Ce produit ne peut pas être nettoyé et/ou stérilisé correctement par l’utilisateur pour permettre sa réutilisation sans
risque ; c’est donc un produit à usage unique. Toute tentative de nettoyer ou de stériliser ces instruments peut se
traduire par une bio-incompatibilité, une infection ou des risques de défaillance du produit au détriment du patient.
Ce dispositif nest pas fabriqué à partir de latex de caoutchouc naturel ni de DEHP.
Identification of a substance that
is contained or present within the
product or packaging.
Do not use if
package is opened
or damaged
Identification of a substance that is
not contained or present within the
product or packaging.
en
Instructions For Use
fr
Mode d’emploi
de
Gebrauchsanweisung
nl
Gebruiksaanwijzing
es
Instrucciones de uso
it
Istruzioni per l'uso
sv
Bruksanvisning
da
Brugsanvisning
pt
Instruções de uso
ru
Инструкция по применению
zh
使用说明
Identification d’une substance
contenue ou présente dans le
produit ou l’emballage.
Ne pas utiliser si
l’emballage est ouvert
ou endommagé
Identification d’une substance qui
n’est pas contenue ou présente
dans le produit ou l’emballage.
IP22
Protection contre l’accès à des parties
dangereuses avec un doigt et protection
contre la chute verticale de gouttes d’eau
lorsque le capteur est incliné à 15°.
IP22
Protected against access to hazardous parts with
a finger and protected against vertically falling
water drops when sensor tilted up to 15°.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6

Covidien Nellcor SCNEO Manuel utilisateur

Taper
Manuel utilisateur
Ce manuel convient également à