Brief Instruction FIB-SEM 002
Safely Powering Off/On your Crossbeam System
Systems concerned Crossbeam Series
Prerequisite documents Instruction Manual Crossbeam
Launch stage
Required time
Issued by Carl Zeiss Microscopy GmbH
Date 19 Jan 2022
P/N
Table of Contents
1 Shuting down the Microscope............................................................... 2
1.1 Shuting down the FIB Gun .......................................................................................... 2
1.2 Shutting down the SEM EHT ....................................................................................... 3
1.3 Shutting down the SEM Gun....................................................................................... 4
1.4 Shuting down the SmartSEM User Interface................................................................ 4
1.5 Shuting down the EM Server....................................................................................... 4
1.6 Shuting down the PC .................................................................................................. 5
1.7 Turning off the Microscope at the Front Panel ............................................................ 5
1.8 Turning off the Microscope at the Back Panel ............................................................. 6
2 Restarting the Microscope..................................................................... 6
2.1 Switching on the Microscope at the Back Panel .......................................................... 6
2.2 Switching on the Microscope at the Front Panel ......................................................... 7
2.3 Starting the PC............................................................................................................ 7
2.4 Starting SmartSEM...................................................................................................... 7
2.5 Waiting for SEM Vacuum............................................................................................ 9
2.6 Waiting for FIB Vacuum .............................................................................................. 9
Brief Instruction FIB-SEM 002 | en-US | Rev. 1 1
1 Shuting down the Microscope ZEISS
1 Shuting down the Microscope
This manual will safely guide you to power your Crossbeam Off and On.
Info
Please note the following
4This method works with the pumping system off for one day maximum. For a longer period of
time, contact Zeiss service to help you with an eventual bakeout of the system or a more com-
plex vacuum recovering procedure.
4Not every user profile is allowed to switch off the SEM gun. For this procedure, select at least
the SYSTEM user profile.
1.1 Shuting down the FIB Gun
Procedure 1. Select the FIB Control panel.
2. Click Off.
2 Brief Instruction FIB-SEM 002 | en-US | Rev. 1
ZEISS Shuting down the Microscope
3. Wait for FIB EHT = 0
4. Close the FIB Control panel.
1.2 Shutting down the SEM EHT
Procedure 1. Go to the lower right corner of the user interface.
2. Select All > EHT Off.
3. Wait for the Running down.. bar to fill and disap-
pear.
9 is displayed.
Brief Instruction FIB-SEM 002 | en-US | Rev. 1 3
1 Shuting down the Microscope ZEISS
1.3 Shutting down the SEM Gun
Procedure 1. Go to the lower right corner of the user interface.
2. Select Gun > Shutdown Gun.
3. Confirm with Yes.
4. Wait for the Running down.. bar to fill and disap-
pear.
9 is displayed.
1.4 Shuting down the SmartSEM User Interface
Procedure 1. Go to the upper right corner of the user interface
and click X .
2. Confirm to close the software.
9The SmartSEM user interface closes.
1.5 Shuting down the EM Server
Procedure 1. Locate the EM Server window. It may be minimized in the Windows toolbar.
2. To close the server, click X in the upper right corner.
4 Brief Instruction FIB-SEM 002 | en-US | Rev. 1
ZEISS Shuting down the Microscope
3. Confirm to shut down the EM Server.
4. Wait until the server window disappears.
1.6 Shuting down the PC
Procedure 1. Close all running applications.
2. Shut down the PC.
1.7 Turning off the Microscope at the Front Panel
Procedure 1. Locate the three buttons at the front of the micro-
scope.
Brief Instruction FIB-SEM 002 | en-US | Rev. 1 5
2 Restarting the Microscope ZEISS
2. Press the red OFF button and wait until it is the
only one that remains illuminated.
1.8 Turning off the Microscope at the Back Panel
Procedure 1. Locate the black rotating knob at the back panel
of the microscope.
2. Turn it to the OFF position.
9The microscope is now off.
2 Restarting the Microscope
2.1 Switching on the Microscope at the Back Panel
Procedure 1. Locate the black rotating knob at the back panel
of the microscope.
2. Turn it to the ON position.
9The microscope is now ON.
6 Brief Instruction FIB-SEM 002 | en-US | Rev. 1
ZEISS Restarting the Microscope
2.2 Switching on the Microscope at the Front Panel
Prerequisite üOnly the red OFF button is illuminated
Procedure 1. Locate the three buttons at the front of the micro-
scope.
2. Press the ON button and wait until it is the only
one that stays illuminated.
2.3 Starting the PC
Procedure 1. Switch on the PC.
2. Log on to your Windows user account (default is User: SEM; Password: sem).
3. Make sure the remote icon is displayed in the bottom right corner before launching SmartSEM.
2.4 Starting SmartSEM
Procedure
1. On the desktop, double click the SmartSEM user interface icon .
Brief Instruction FIB-SEM 002 | en-US | Rev. 1 7
2 Restarting the Microscope ZEISS
àThe EM Server starts.
2. Log on to the "system" account by entering user name and password.
Alternatively, press <Ctrl+s>. There is no default password for the system user.
àAn error message should appear in the bottom
right corner of the screen.
3. Do not click anything, but wait for SmartSEM to finish booting.
8 Brief Instruction FIB-SEM 002 | en-US | Rev. 1
ZEISS Restarting the Microscope
2.5 Waiting for SEM Vacuum
Procedure 1. Navigate to the Crossbeam SEM Control panel and
wait until Vac Status is “Ready”.
2.6 Waiting for FIB Vacuum
Procedure
1. On the top right corner of the UI, click the FIB Control icon .
Brief Instruction FIB-SEM 002 | en-US | Rev. 1 9
2 Restarting the Microscope ZEISS
àThe FIB Control panel is displayed.
2. In the bottom right corner, confirm the error mes-
sage with Yes.
àSmartSEM launches the power fail recovery process.
àAfter pumping time the last power fail recovery
message should be displayed.
3. Click OK.
9The microscope is now ready for normal operation.
Info
If you have any further questions, contact:
4Microscopy Headquarters
10 Brief Instruction FIB-SEM 002 | en-US | Rev. 1
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10

Zeiss FIB-SEM 002 Manuel utilisateur

Taper
Manuel utilisateur

dans d''autres langues