Hach CLF10sc Manuel utilisateur

Taper
Manuel utilisateur

Ce manuel convient également à

DOC023.97.80087
CLF10sc and CLT10sc Reagentless
Chlorine Analyzer
09/2012, Edition 3
User Manual
Manuel d'utilisation
Manual del usuario
Manual do Usuário
用户手册
取扱説明書
사용 설명서
English..............................................................................................................................3
Français.........................................................................................................................21
Español..........................................................................................................................40
Português......................................................................................................................59
中文.................................................................................................................................78
日本語.............................................................................................................................95
한글...............................................................................................................................113
2
Table of contents
Specifications on page 3 Maintenance on page 12
General information on page 3 Troubleshooting on page 16
Installation on page 8 Replacement parts and accessories on page 19
Operation on page 11
Specifications
Specifications are subject to change without notice.
Specification Detail
Component description Amperometric reagentless chlorine monitoring analyzer
panel, with chlorine, flow, optional pH combination or
pHD sensor, flow cells, sc controller and digital
gateway
Operating temperature 0 to 45 °C (0 to 113 °F)
Storage temperature (panel only) –20 to 60 °C (–4 to 149 °F)
Power requirements 12 VDC ± 10%, 100 mA maximum (supplied by sc
controller)
Panel dimensions (L x W x D) 48.3 x 49.5 x 15.1 cm (19 x 19.5 x 5.95 in.) with panel-
mounted components
Weight Approximately 5.4 kg (12 lb) (panel and empty panel-
mounted components only)
Flow rate Range: 30–50 L/hour (7.9–13.2 gal/hour)
Optimal: 40 L/hour (10.5 gal/hour)
Controller platform sc controller models
General information
In no event will the manufacturer be liable for direct, indirect, special, incidental or consequential
damages resulting from any defect or omission in this manual. The manufacturer reserves the right to
make changes in this manual and the products it describes at any time, without notice or obligation.
Revised editions are found on the manufacturer’s website.
Safety information
N O T I C E
The manufacturer is not responsible for any damages due to misapplication or misuse of this product including,
without limitation, direct, incidental and consequential damages, and disclaims such damages to the full extent
permitted under applicable law. The user is solely responsible to identify critical application risks and install
appropriate mechanisms to protect processes during a possible equipment malfunction.
Please read this entire manual before unpacking, setting up or operating this equipment. Pay
attention to all danger and caution statements. Failure to do so could result in serious injury to the
operator or damage to the equipment.
Make sure that the protection provided by this equipment is not impaired. Do not use or install this
equipment in any manner other than that specified in this manual.
English
3
Use of hazard information
D A N G E R
Indicates a potentially or imminently hazardous situation which, if not avoided, will result in death or serious injury.
W A R N I N G
Indicates a potentially or imminently hazardous situation which, if not avoided, could result in death or serious
injury.
C A U T I O N
Indicates a potentially hazardous situation that may result in minor or moderate injury.
N O T I C E
Indicates a situation which, if not avoided, may cause damage to the instrument. Information that requires special
emphasis.
Precautionary labels
Read all labels and tags attached to the instrument. Personal injury or damage to the instrument
could occur if not observed. A symbol on the instrument is referenced in the manual with a
precautionary statement.
This symbol, if noted on the instrument, references the instruction manual for operation and/or safety
information.
Electrical equipment marked with this symbol may not be disposed of in European public disposal
systems after 12 August of 2005. In conformity with European local and national regulations (EU
Directive 2002/98/EC), European electrical equipment users must now return old or end-of-life
equipment to the Producer for disposal at no charge to the user.
Note: For return for recycling, please contact the equipment producer or supplier for instructions on how to return end-
of-life equipment, producer-supplied electrical accessories, and all auxillary items for proper disposal.
Product overview
This product is primarily intended for use in municipal drinking-water applications. Wastewater and
industrial applications must be identified to be applicable by the manufacturer before installation.
The Reagentless Chlorine Analyzer monitors the concentration of free or total residual chlorine in
water, and is most reliable at residual chlorine concentrations higher than 0.1 ppm (mg/L). The
combination of a sc model controller with a three-electrode chlorine sensor, optional pH sensor and a
flow sensor give the best monitoring function. Both sensors read sample temperature. The pH sensor
transmits the temperature reading to the controller display. The chlorine sensor uses its temperature
measurement internally and does not transmit the measurement to the controller display. A pressure
regulator kit and an acidification and cleaning kit are available as accessories.
The chlorine and pH sensors rest in panel-mounted, series-connected flow cells. The flow cells
maintain the sensor-to-sample contact and help prevent the sensors from drying out when the
system is not in operation. Figure 1, Figure 2, Figure 3 show overviews of the system and the flow
cells.
Each sensor connects to the controller through the gateway mounted on the panel. The gateway
converts analog signals from the sensors to digital signals used by the controller. The controller
converts the digital signals to the chlorine, pH and temperature measurement values and shows the
readings in the display. The controller supplies power to the sensors and to the digital gateway.
The controller can be configured to cause an alarm condition or warning if measurement values rise
above or fall below specified values. For information on controller alarms, refer to the controller user
manual.
Note: Refer to the sensor user manuals for information about how to store the sensors when the system is not in
operation.
4
English
Two analyzer configurations are available: a Grab Sample option (without pH sensor) and a pH
option. The Grab Sample option includes the mounting panel, the chlorine sensor and flow cell, a pH
flow cell without sensor, a flow sensor, the digital gateway and the controller. The pH option includes
these items plus a pH sensor.
A flow control valve regulates the sample flow. A constant flow rate of 30-50 L/h is recommended for
correct sensor operation. The optimal flow rate is 40 L/hour. A single LED on the flow sensor shows if
the sample flow rate is sufficient. If the LED is not on, the flow rate must be adjusted until the LED is
on. The exact flow rate is not important as long as the LED is on, but the flow must stay constant
within the specifications.
Figure 1 System overview
1 Controller 7 Sample in
2 Gateway 8 Valve, shutoff (customer-supplied)
3 Chlorine sensor cable 9 Chlorine flow cell
4 pH sensor (optional) 10 Flow sensor cable
5 Sample drain 11 Chlorine sensor
6 pH flow cell
English 5
Figure 2 Chlorine flow cell overview
1 Flow sensor 8 Sample out fitting
English: Speed–fit fitting—3/8 in. OD tube
Metric: Speed–fit fitting—10 mm OD tube
2 Plug, 1/8 in. NPTF 9 Chlorine sensor alignment mark
3 Float alignment mark 10 Valve, cleaning port, 1/4 in. OD tube
4 Plug, 1/2 in. NPTF 11 Flow cell inlet
English: Speed–fit fitting—1/4 in. OD tube
Metric: Speed–fit fitting—6 mm OD tube
5 Lock ring 12 Flow cell
6 Sealing hub 13 Flow control valve
7 O-ring 14 Float
6 English
Figure 3 pH flow cell overview
1 Lock ring 4 Sample out fitting
English: Speed–fit elbow fitting—1/2 in. OD
Metric: Speed–fit elbow fitting—12 mm OD
2 Sealing blank 5 Grab sample port
3 O-ring 6 Flow cell inlet
English: Speed–fit fitting—3/8 in. OD
Metric: Speed–fit fitting—10 mm OD
Product components
Make sure that all components have been received. Refer to Figure 1 on page 5. If any items are
missing or damaged, contact the manufacturer or a sales representative immediately.
English
7
Figure 4 Reagentless chlorine analyzer panel
1 System panel 3 Tubing, 12 mm (½ in.)
2 Connector cable for connecting gateway to
controller, 1 m (3.3 ft)
4 Chlorine sensor
Installation
Installation considerations
Do the mechanical installation tasks first, the electrical installation tasks second, and the plumbing
installation tasks last.
All fittings are designed to be tightened by hand only.
The panel is designed to be wall or rack mounted. Mount or position the panel in an accessible
location.
Do not place the panel next to a heat source.
Do not mount the panel in an environment with frequent vibrations
Do not mount the panel in direct sunlight.
Install the panel indoors or in an environmental enclosure.
Prevent condensation from collecting on the panel and panel components, especially the flow cell,
sensor and where the sensor and cable connect.
Use Teflon
®
or PVDF linear low-density polyethylene tubing for the sample supply line and keep
the tubing as short as possible to minimize lag time.
Sample pressure must be sufficient to keep a continuous water supply to the sensors. An LED
indicator on the flow sensor will be lit as long as the flow is sufficient.
Keep the flow rate as constant as possible for best sensor operation.
Sample line guidelines
Select a good, representative sampling point for the best instrument performance. The sample must
be representative of the entire system.
To prevent erratic readings:
Collect samples from locations that are sufficiently distant from points of chemical additions to the
process stream.
Make sure that the samples are sufficiently mixed.
8
English
Make sure that all chemical reactions are complete.
Mechanical installation
Mount the panel
Figure 5 shows the mounting dimensions. Refer to the figure and follow the steps to mount the panel.
1. Attach mounting screws to the mounting surface. Mount on wall studs or use wall anchors. Leave
enough room between the head of each screw and the mounting surface so the panel will fit in
between.
2. Slide the panel mounting holes over the heads of the mounting screws and allow the panel to
slide down until the top of each mounting hole rests on the screw body.
3. Tighten each screw to secure the panel to the mounting surface.
4. Install the chlorine and pH sensor (optional) in the flow cells. For information on how to install a
sensors, refer to the sensor user manual.
Figure 5 Reagentless chlorine analyzer panel dimensions
English 9
Electrical installation
Wire the panel
W A R N I N G
Potential Electrocution Hazard. Always disconnect power to the instrument when making electrical
connections.
The controller supplies power to the sensors and the digital gateway.
1. Remove power from the controller.
2. Connect the pH sensor (optional) wires to the gateway. For information on how to install sensors
in the flow cells, or how to connect the sensor to the gateway, refer to the sensor user manual.
3. Connect the chlorine sensor to the gateway. For information on how to connect the sensor to the
gateway, refer to the sensor user manual.
Note: When the panel is shipped, the chlorine sensor cable is taped to the front of the panel near the location
of the sensor connection.
4. Connect the controller to the gateway with the 1 m connector cable that is supplied with the
panel.
Plumbing
Plumb the panel
Make sure to follow the size specifications for all panel tubing and fittings. The flow path must
increase in diameter as water flows through the system to prevent build-up of backpressure. Figure 1
on page 5 shows an overview of the plumbing connections.
1. Attach tubing to the sample inlet and outlet fittings.
2. Make sure the cleaning port on the chlorine flow cell is closed.
3. Make sure the grab sample port on the pH flow cell is closed.
4. Open the flow valve on the sample line and allow water to move through the system.
5. Make sure there are no leaks from the tubing, valve fittings or the flow cells.
6. To minimize air bubbles, the pH flow cell may be tilted between 0º and 30º as shown in Figure 6.
10
English
Figure 6 pH flow cell tilt
Operation
User navigation
Refer to the controller documentation for keypad description and navigation information.
System configuration
Sensor configuration and calibration are done through the controller user interface. For more
information on how to configure or calibrate a sensor, refer to the sensor user manual.
An sc model controller must be connected to the gateway.
1. From the controller main menu, select Sensor Setup.
The Chlorine Gateway (CGW) Setup menu is displayed.
2. Select Configure and customize the entries. Confirm after each entry to save the changes and go
back to the Configure menu.
Option Description
Edit name Edit the sensor name
Select Parameter—Chlorine (Total or
Free), pH (optional), Temperature
(optional)
Selects the parameter to be configured. Configuring the
selected parameter requires additional steps. Refer to the
sensor user manual for more information.
Reset Defaults—Confirm to do the reset. Resets all settings to the factory default values.
English 11
System start up
Make sure all mechanical, electrical and plumbing tasks have been correctly done before starting
flow through the system.
1. Apply power to the controller.
When powered on for the first time, the controller displays the Language, Date Format and
Date/Time screens in order.
2. Set the language, date and time in the controller if this has not been done. Refer to the controller
user manual for more information on how to set these options.
The controller scans for connected devices.
3. Make sure the cleaning port and the grab sample port are closed, then open the ON/OFF valve
on the sample line to start the flow of water through the system.
The chlorine and pH flow cells will fill with water.
4. Make sure the flow sensor LED is on. If the LED is not on, adjust the flow control valve to
increase the flow.
The flow sensor LED will be on when the flow is sufficient for good sensor operation. If the flow is
too high, the LED may not be lit. Make sure the float is approximately aligned with the flow
alignment mark.
5. Check for leaks or blockages in the tubing and the flow cells. Shut off the flow to the panel if any
problems are discovered and make the necessary repairs.
6. When the system has sufficient flow, clear any warnings that appear on the controller display by
doing the suggested operations.
7. Push the HOME key.
The main measurement screen appears and shows the chlorine, temperature and pH (if a pH
sensor is connected) measurement values.
Maintenance
D A N G E R
Multiple hazards. Only qualified personnel must conduct the tasks described in this section of the
document.
Tubing replacement
Replace the tubing at least once a year. Replace brittle, cracked or leaky tubing immediately.
Replace old, discolored, fouled or damaged tubing with tubing of the same diameter and type
(Teflon
®
or PVDF tubing).
Clean a sensor flow cell
Clean a flow cell as necessary to remove deposits. Clean a flow cell if it becomes heavily
contaminated. Replace a flow cell if it becomes damaged, discolored or not transparent. It is not
necessary to remove a flow cell from the panel in order to clean it.
Note: Never use detergents or surfactants to clean a chlorine flow cell.
1. Stop the source flow to the panel.
2. Remove the chlorine sensor and put the sensor in a container filled with a sample of the
chlorinated water.
Note: Do not disconnect the sensor from the power. If power is removed from the chlorine sensor, the sensor
must be conditioned and calibrated before use.
Note: The removal of a sensor from a flow cell with the sensor cable attached will cause a change in the sensor
signal outputs. Refer to the controller manual for more information.
3. Rinse the flow cell with water and scrub the flow cell with a soft bristle brush to remove deposits.
4. Rinse the flow cell with water.
12
English
5. If the flow cell is not sufficiently clean, soak the flow cell for 10 to 15 minutes in a 1:3 mixture of
3-5% hydrogen peroxide solution and vinegar.
6. Rinse the flow cell with water.
7. Install the sensor in the flow cell and start the sample flow.
Replace the flow sensor
Under normal conditions, the flow sensor does not require regular maintenance or replacement.
However, if the LED flow indicator does not light when the float is at or above the float alignment
mark and power is applied, the sensor may need to be replaced. To replace the flow sensor, do the
illustrated steps that follow.
Note: It is not necessary to remove the chlorine sensor, drain the flow cell or power down the controller to replace
the flow sensor.
English 13
14 English
Replace the flow control valve assembly
English 15
Troubleshooting
Event log
The controller provides one event log for each sensor. The event log stores a variety of events that
occur on the devices such as calibrations done, calibration options changed, etc. A list of possible
events is shown below. The event log can be read out in a CSV format. For instructions on
downloading the logs, refer to the controller user manual.
Table 1 Event log
Event Description
Power On The power was turned on.
Flash Failure The external flash has failed or is corrupted.
1pointpHCalibration_Start Start of one-point sample calibration for pH
1pointpHCalibration_End End of one-point sample calibration for pH
2pointpHCalibration_Start Start of two-point sample calibration for pH
2pointpHCalibration_End End of two-point sample calibration for pH
1pointBufferpHCalibration_Start Start of one-point buffer calibration for pH
1pointBufferpHCalibration_End End of one-point buffer calibration for pH
2pointBufferpHCalibration_Start Start of two-point buffer calibration for pH
2pointBufferpHCalibration_End End of two-point buffer calibration for pH
TempCalibration_Start Start of temperature calibration
TempCalibration_End End of temperature calibration
1pointChemZeroCL2_Start Start of one-point chemical zero calibration for chlorine
1pointChemZeroCL2_End End of one-point chemical zero calibration for chlorine
1pointElecZeroCL2_Start Start of one-point electrical zero calibration for chlorine
1pointElecZeroCL2_End End of one-point electrical zero calibration for chlorine
1pointProcessConc_Start Start of one-point process concentration for chlorine
1pointProcessConc _End End of one-point process concentration for chlorine
2pointChemCL2_Start Start of two-point chemical calibration for chlorine
2pointChemCL2_End End of two-point chemical calibration for chlorine
2pointElecCL2_Start Start of two-point electrical calibration for chlorine
2pointElecCL2_End End of two-point electrical calibration for chlorine
CL2CalSetDefault The chlorine calibration was reset to the default.
pHCalSetDefault The pH calibration data was reset to the default.
TempCalSetDefault The temperature calibration data was reset to the default.
AllCalSetDefault All sensor calibration data was reset to the default.
CL2CalOptionChanged The chlorine calibration option was changed.
pHCalOptionChanged The pH calibration option was changed.
TempCalOptionChanged The temperature calibration option was changed.
SensorConfChanged The sensor configuration was changed.
16 English
Table 1 Event log (continued)
Event Description
ResetCL2CalHist The CL2 calibration history was reset.
ResetpH CalHist The pH calibration history was reset.
ResetTemp CalHist The temperature calibration history was reset.
ResetAllSensorsCalHist All sensor calibration history was reset.
ResetCL2Sensor The CL2 calibration data (sensor days, calibration history and calibration data)
was reset to the default.
ResetpHSensor The pH calibration data (sensor days, calibration history and calibration data)
was reset to the default.
ResetTempSensor The temperature calibration data (sensor days, calibration history and
calibration data) was reset to the default.
ResetAllSensors All sensor calibration data (sensor days, calibration history and calibration data)
was reset to the default.
Flow Detection The state of the proximity sensor has changed (sufficient or not sufficient flow).
Error list
Errors may occur for various reasons. The sensor reading on the measurement screen flashes. All
outputs are held when specified in the controller menu. To show the sensor errors, push the MENU
key and select Sensor Diag, Error List. A list of possible errors is shown below.
Table 2 Error list for the sensor
Error Description Resolution
CL CAL REQD A chlorine calibration and/or pH calibration
is required.
The chlorine and/or pH measurement has
changed enough to cause a Cal Watch
alarm to occur. Refer to the chlorine sensor
manual for more information.
Calibrate the chlorine sensor and/or pH
sensor.
CL MAINT REQD Chlorine sensor maintenance is required.
The slope is less than 30% or more than
300% of the default.
The default slope is 100 mV/ppm (100%).
Do sensor maintenance and then repeat the
calibration, or replace the sensor. Refer to the
chlorine sensor manual for more information.
CL MAINT REQD Chlorine sensor maintenance is required.
The offset is too high (more than ±50 mV).
Do sensor maintenance and then repeat the
calibration, or replace the sensor. Refer to the
chlorine sensor manual for more information.
PH TOO LOW The pH value is less than 0 pH. Calibrate or replace the pH sensor.
PH TOO HIGH The pH value is more than 14 pH.
PH SLOPE FAIL The slope is outside of the -45 to -65 mV/pH
range.
Clean the pH sensor, then repeat the
calibration with a fresh buffer or sample, or
replace the sensor.
PH OFFSET FAIL The offset is outside of ±60 mV. Clean the pH sensor and then repeat the
calibration with a fresh buffer or sample, or
replace the sensor.
PH OFFSET FAIL The offset is outside of ±60 mV. Clean the pH sensor and replace the salt
bridge and standard cell solution. Repeat the
calibration with a fresh buffer or sample, or
replace the sensor.
English 17
Table 2 Error list for the sensor (continued)
Error Description Resolution
TEMP TOO LOW The temperature is less than 0 °C. Calibrate the temperature or replace the pH
sensor.
TEMP TOO HIGH The temperature is more than 100 °C.
TEMP FAIL The offset is higher than 5.0 °C or lower
than -5.0 °C.
Calibrate the temperature or replace the pH
sensor.
NO FLOW The flow level is too low. Increase the flow.
Warning list
A warning does not affect the operation of menus, relays and outputs. A warning icon flashes and a
message is shown on the bottom of the measurement screen. To show the sensor warnings, push
the MENU key and select Sensor Diag, Warning List. A list of possible warnings is shown below.
Table 3 Warning list for the sensor
Warning Description Resolution
CL CAL RECD A chlorine and/or pH calibration is recommended.
The chlorine and/or pH measurement has changed
enough to cause a Cal Watch warning alarm to occur.
Refer to the sensor manuals for more information.
Calibrate the chlorine sensor
and/or pH sensor.
CL CAL RECD A chlorine calibration is recommended.
Chlorine calibration data is not available (sensor with
default calibration data).
Calibrate the chlorine sensor.
PH CAL RECD A pH calibration is recommended.
pH calibration data is not available (sensor with default
calibration data).
Calibrate the pH sensor.
TEMP CAL RECD A temperature calibration is recommended.
Temperature calibration data is not available (sensor
with default calibration data).
Calibrate the temperature.
CL CAL TO DO The Sensor Days value for the chlorine sensor is
greater than the Cal Reminder value.
Calibrate the chlorine sensor.
PH CAL TO DO The Sensor Days value for the pH sensor is greater
than the Cal Reminder value.
Calibrate the pH sensor.
TEMP CAL TO DO The Sensor Days value for the temperature sensor is
greater than the Cal Reminder value.
Calibrate the temperature.
CL MAINT RECD Chlorine sensor maintenance is recommended.
The slope is 30 to 45% of default or the slope is 250 to
300% of default.
The default slope is 100 mV/ppm (100%).
Do sensor maintenance and then
repeat the calibration, or replace
the sensor.
CL MAINT RECD Chlorine sensor maintenance is recommended.
The offset is -50 mV to 45 mV or the offset is 45 mV to
50 mV.
Do sensor maintenance and then
repeat the calibration, or replace
the sensor.
PH MAINT RECD pH sensor maintenance is recommended.
The slope is outside of the -50 to -61 mV/pH range.
Clean the pH sensor, then repeat
the calibration with a fresh buffer
or sample, or replace the sensor.
PH MAINT RECD pH sensor maintenance is recommended.
The offset is outside of ±45 mV but within ±60 mV.
Clean the sensor and then repeat
the calibration, or replace the
sensor.
18 English
Table 3 Warning list for the sensor (continued)
Warning Description Resolution
PH MAINT RECD pH sensor maintenance is recommended.
The offset is outside of ±45 mV but within ±60 mV.
Clean the sensor and replace the
salt bridge and standard cell
solution and then repeat the
calibration, or replace the sensor.
T MAINT RECD The temperature offset is outside of ±3 °C but within
±5 °C.
Calibrate the temperature.
Replacement parts and accessories
W A R N I N G
Personal injury hazard. Use of non-approved parts may cause personal injury, damage to the
instrument or equipment malfunction. The replacement parts in this section are approved by the
manufacturer.
Note: Product and Article numbers may vary for some selling regions. Contact the appropriate distributor or refer to
the company website for contact information.
Parts
Description Item no.
Cable, gateway to controller, 1 m 6122400
Cable assembly, 4 pin, shielded, flow sensor, 2 m 9160900
Cable assembly, 5 pin, shielded, M12, chlorine sensor, 1 m 9160300
Fitting, inlet port for chlorine flow cell
English:
9159200 (1/4 in.)
Metric:
9196400 (6 mm)
Fitting, speed fit, outlet port for chlorine flow cell or inlet port for pH flow cell
English:
9159300 (3/8 in.)
Metric:
9196500 (10 mm)
Fitting, elbow, outlet port for pH flow cell
English:
9196900 (1/2 in.)
Metric:
9197000 (12 mm)
Float, flow sensor 9159100
Flow cell assembly, chlorine
English: 9198400
Metric: 9198500
Flow cell assembly, pH combination
English: 9180200
Metric: 9180400
Flow cell assembly, pHD
includes sealing hub
English: 9180100
Metric: 9180300
Flow cell body only (pH) 9159700
Flow cell block, chlorine 9158900
Flow control valve 9159000
Flow control valve kit 9198600
Flow sensor 9160800
English 19
Replacement parts and accessories (continued)
Description Item no.
Gateway, digital 9180500
Lock ring, chlorine flow cell 9196200
Lock ring, pH flow cell 60F2032-001
O-ring, flow control valve (2x) (#2-108, Viton) 9159600
O-ring, pH flow cell (#2-228, Viton) 5H1223
O-ring, sealing hub, chlorine flow cell ( #2-123, Viton) 9196300
O-ring, Split Ring (M25 x 2.5) 9181800
Plug, chlorine flow cell, 1/2 in. 9159500
Screw, chlorine flow cell (2x) (10-32 x 1.75 Lg Phillps pan head) 4422200
Screw, pH flow cell (2x) (10-32 x .5 Lg Phillps pan head) 561211
Sealing blank, pH flow cell 9198000
Sealing hub, chlorine flow cell 9196100
Sealing hub, pH combination sensor, 3/4 in. 9159800
Sealing hub, pHD flow cell, 1 in. 60F2021-001
Split ring, chlorine flow cell 9181700
Tubing, sample stream, 1/4 in. x 2 m 4743800
Tubing, sample stream, 3/8 in. x 18 in. 9198100
Tubing, waste stream, 1/2 in. x 2 m 9198200
Valve, grab sample port 9005300
Valve, speed fit, cleaning port 6166300
Accessories
Description Item no.
Acidification Kit LZY051
Pressure Regulator Sample Conditioning Kit 9159900
20 English
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114
  • Page 115 115
  • Page 116 116
  • Page 117 117
  • Page 118 118
  • Page 119 119
  • Page 120 120
  • Page 121 121
  • Page 122 122
  • Page 123 123
  • Page 124 124
  • Page 125 125
  • Page 126 126
  • Page 127 127
  • Page 128 128
  • Page 129 129
  • Page 130 130
  • Page 131 131
  • Page 132 132

Hach CLF10sc Manuel utilisateur

Taper
Manuel utilisateur
Ce manuel convient également à