Working conditions
1. Environment Temp.: 5~32℃ (if setting temp.≦10℃,environment≦28℃).
2. Humidity:≦80%RH 3. Air Pressure: 80-106Kpa
4. There must be no shaking of the incubator and no caustic air around the chamber.
5. Should avoid light and cooling environment.
6. Keep distance of dust, install horizontally and make sure there is distance between the equipment and wall.
Safety information
1. After you receive the equipment, to avoid the damage of the compressor during a long transportation period,
please keep the machine without moving during one day before using it. To make sure equipment safety, please
install outer protection and also supply power as the equipment requires.
2. Do not test the flammable, combustible, poisonous or corrosive material.
3. The equipment should be installed horizontally.
4. Only trained personnel can disassembly and repair the equipment.
5. If you add one working equipment into the chamber, if the power is bigger than 2A, plug the power outside,
do not use this equipment power.
6. When the chamber temp.≧50 ℃, please do not set the low temp. settings to make sure the compressor can
be used for a long time.
7. Anyone who operates the equipment must first read this manual
Operation cautions
1. There is a test hole at the top of the incubator. If you add another equipment inside the chamber, please do it
through this hole.
2. If there is no need for U.V. light, turn off the U.V. lamp to avoid influence on temperature.
3. Before operating the cooling function, please dry the chamber for 1 hour under 50 ºC.
4. The chamber has a vertical air system. Please do not fully fill the shelfs with materials, the test material must
occupy less than 1/3 of the shelf.
5. Once the test environment > 35℃ or Use Temp. > 50℃, do not set the low temp. setting
6. Please clean the chamber and surface of the machine with pure water. When the machine is not used, please
plug off the electricity, and make sure the inside and outside is clean and dry.