Lovibond 00386467 - MD1x0/200 - Chlorine HR (KI) T, Chlorine dioxide T, Chlorine L, Chlorine T Manuel utilisateur

Taper
Manuel utilisateur
Lovibond® Water Testing
Tintometer® Group
Manual of Methods
MD 100 • MD 110 • MD 200
Chlorine | Chlorine Dioxide
EN
Manual of Methods
Page 4
DE
Methodenhandbuch
Seite 32
ES
Manual de Métodos
Página 60
FR
Méthodes Manuel
Page 88
IT
Manuale dei Metodi
Pagina 116
PT
Métodos Manual
Página 144
NL
Handboek Methoden
Zijde 172
ZH
方法手册
Page 200
NL Handboek van Methoden 01/20EN Handbook of Methods 01/20
KS4.3 T
0.1 - 4 mmol/l KS4.3
Acid / Indicator
20
S:4.3
Instrument specific information
The test can be performed on the following devices. In addition, the required cuvette
and the absorption range of the photometer are indicated.
Instrument Type Cuvette λMeasuring Range
MD 200, MD 600, MD 610,
MD 640, MultiDirect, PM 620,
PM 630
ø 24 mm 610 nm 0.1 - 4 mmol/l KS4.3
SpectroDirect, XD 7000,
XD 7500
ø 24 mm 615 nm 0.1 - 4 mmol/l KS4.3
Material
Required material (partly optional):
Reagents Packaging Unit Part Number
Alka-M-Photometer Tablet / 100 513210BT
Alka-M-Photometer Tablet / 250 513211BT
Application List
Waste Water Treatment
Drinking Water Treatment
Raw Water Treatment
Notes
1. The terms Alkalinity-m, m-Value, total alkalinity and Acid demand to KS4.3 are iden-
tical.
2. For accurate results, exactly 10 ml of water sample must be used for the test.
Display in the MD
100 / MD 110 /
MD 200
Bar code for the
detection of the
methods
Method name
Measuring range
Chemical Method
Method number
Language codes
ISO 639-1 Revision status
KS4.3 T / 20
How to use
ENMethod Reference Book4
EN
Implementation of the provision Acid capacity KS4.3 with Tablet
Select the method on the device
For this method, no ZERO measurements are to be carried out with the following de-
vices: XD 7000, XD 7500
10 ml
Fill 24 mm vial with 10 ml
sample.
Close vial(s).
Sample
Place sample vial in the
sample chamber. • Pay
attention to the positioning.
Dissolve tablet(s) by inver-
ting.
Sample
Place sample vial in the
sample chamber. • Pay
attention to the positioning.
Test
Press the TEST (XD:
START) button.
The result in Acid Capacity KS4.3 appears on the display.
Performing test
procedure
KS4.3 T / 20
NL Handboek van Methoden 01/20EN Handbook of Methods 01/20
How to use
ENMethod Reference Book 5
EN
Chlorine T / M100
ENMethod Reference Book1.06
EN
Chlorine T M100
0.01 - 6.0 mg/L Cl2 a) CL6
DPD
Material
Required material (partly optional):
Reagents Packaging Unit Part Number
DPD No.1 Tablet / 100 511050BT
DPD No. 1 Tablet / 250 511051BT
DPD No. 1 Tablet / 500 511052BT
DPD No. 3 Tablet / 100 511080BT
DPD No. 3 Tablet / 250 511081BT
DPD No. 3 Tablet / 500 511082BT
DPD No. 1 High Calcium e) Tablet / 100 515740BT
DPD No. 1 High Calcium e) Tablet / 250 515741BT
DPD No. 1 High Calcium e) Tablet / 500 515742BT
DPD No. 3 High Calcium e) Tablet / 100 515730BT
DPD No. 3 High Calcium e) Tablet / 250 515731BT
DPD No. 3 High Calcium e) Tablet / 500 515732BT
DPD No. 4 Tablet / 100 511220BT
DPD No. 4 Tablet / 250 511221BT
DPD No. 4 Tablet / 500 511222BT
DPD No. 3 Evo Tablet / 100 511420BT
DPD No. 3 Evo Tablet / 250 511421BT
DPD No. 3 Evo Tablet / 500 511422BT
DPD No. 4 Evo Tablet / 100 511970BT
DPD No. 4 Evo Tablet / 250 511971BT
DPD No. 4 Evo Tablet / 500 511972BT
Available Standards
Title Packaging Unit Part Number
ValidCheck Chlorine 1,5 mg/l 1 Pieces 48105510
Chlorine T / M100
ENMethod Reference Book1.0 7
EN
Sampling
1. When preparing the sample, chlorine outgassing, e.g. through the pipette or
shaking, must be avoided.
2. The analysis must take place immediately after taking the sample.
Preparation
1. Cleaning of vials:
As many household cleaners (e.g. dishwasher detergent) contain reducing
substances, this can lead to lower results with the determination of chlorine.
To avoid measurement errors, the glassware used should be free of chlorine
consumption. To achieve this, all glassware should be placed in a sodium
hypochlorite solution (0.1 g/L) for one hour and then rinsed thoroughly with
deionised water.
2. For individual testing of free and total chlorine, the use of different sets of
glassware is recommended (EN ISO 7393-2, 5.3)
3. The DPD colour development is carried out at a pH value of 6.2 to 6.5. The
reagents therefore contain a buffer for the pH adjustment. Strong alkaline or acidic
water samples must therefore be adjusted between pH 6 and pH 7 before the
analysis (use 0.5 mol/Lsulphuric acid or 1 mol/Lsodium hydroxide).
Remarks
1. Evotablets can be used as an alternative to the corresponding standard tablet (e.g.
DPD No.3 Evoinstead of DPD No.3).
Chlorine T / M100
ENMethod Reference Book1.08
EN
Determination of free chlorine with tablet
Select the method on the device.
10 mL
Sample
Fill 24 mm vial with 10 mL
sample.
Close vial(s). Place sample vial in the
sample chamber. Pay
attention to the positioning.
Zero
Press the ZERO button. Remove the vial from the
sample chamber.
Empty vial except for a few
drops.
24 mm
Add DPD No. 1 tablet . Crush tablet(s) by rotating
slightly.
Fill up vial with sample to the
10 mL mark.
Chlorine T / M100
ENMethod Reference Book1.0 9
EN
Sample
Close vial(s). Dissolve tablet(s) by
inverting.
Place sample vial in the
sample chamber. Pay
attention to the positioning.
Test
Press the TEST (XD:
START)button.
The result in mg/L free chlorine appears on the display.
Determination of total Chlorine with tablet
Select the method on the device.
10 mL
Sample
Fill 24 mm vial with 10 mL
sample.
Close vial(s). Place sample vial in the
sample chamber. Pay
attention to the positioning.
Chlorine T / M100
ENMethod Reference Book1.010
EN
Zero
Press the ZERO button. Remove the vial from the
sample chamber.
Empty vial except for a few
drops.
Add DPD No. 1 tablet . Add DPD No. 3 tablet .As an alternative to DPD
No. 1 and No. 3 tablets, a
DPD No. 4 tablet can be
added.
24 mm
Crush tablet(s) by rotating
slightly.
Fill up vial with sample to
the 10 mL mark.
Close vial(s).
Chlorine T / M100
ENMethod Reference Book1.0 11
EN
Sample
Test
Dissolve tablet(s) by
inverting.
Place sample vial in the
sample chamber. Pay
attention to the positioning.
Press the TEST (XD:
START)button.
9
4
5
6
7
8
10
11 12
1
2
3
2 min
Wait for 2 minute(s)
reaction time.
Once the reaction period is finished, the measurement takes place automatically.
The result in mg/L total Chlorine appears on the display.
Chlorine T / M100
ENMethod Reference Book1.012
EN
Chemical Method
DPD
Appendix
Interferences
Persistant Interferences
All oxidising agents in the samples react like chlorine, which leads to higher results.
Removeable Interferences
Interference from copper and iron (III) are eliminated by the addition of EDTA.
The use of reagent tablets in samples with high calcium content* and/or
high conductivity* can lead to turbidity of the sample and therefore incorrect
measurements. In this case, the alternative reagent tablet DPD No.1 High Calcium
and reagent tablet DPD No.3 High Calcium should be used.
*it is not possible to give exact values, because the development of turbidity depends
on the composition and nature of the sample.
Concentrations above 10 mg/L chlorine, in the event of using fluid reagents, can
lead to results within the measuring range of up to 0 mg/L. In the event of a high
concentration of chlorine, the sample must be diluted with chlorine-free water. 10
mLof the diluted sample should be mixed with the reagent and the measurement
taken again (plausibility test).
Interference from / [mg/L]
CrO4 2- 0.01
MnO20.01
Method Validation
Limit of Detection 0.02 mg/L
Limit of Quantification 0.06 mg/L
End of Measuring Range 6 mg/L
Sensitivity 2.05 mg/L / Abs
Confidence Intervall 0.04 mg/L
Standard Deviation 0.019 mg/L
Variation Coefficient 0.87 %
Conformity
EN ISO 7393-2
Chlorine T / M100
ENMethod Reference Book1.0 13
EN
a) determination of free, combined and total | e) alternative reagent, used instead of DPD No.1/No.3 in case of turbidity
in the water sample caused by high concentration of calcium and/or high conductivity
Chlorine L / M101
ENMethod Reference Book1.014
EN
Chlorine L M101
0.02 - 4.0 mg/L Cl2 a) CL6
DPD
Material
Required material (partly optional):
Reagents Packaging Unit Part Number
DPD 1 Buffer Solution, Blue Bottle 15 mL 471010
DPD 1 Buffer Solution 100 mL 471011
DPD 1 Buffer Solution 1 Pieces 471016
DPD 1 Reagent Solution, Green Bottle 15 mL 471020
DPD 1 Reagent Solution 100 mL 471021
DPD 1 Reagent Solution 1 Pieces 471026
DPD 3 Solution, Red Bottle 15 mL 471030
DPD 3 Solution 100 mL 471031
DPD 3 Solution 1 Pieces 471036
DPD Reagent Set 1 Pieces 471056
Available Standards
Title Packaging Unit Part Number
ValidCheck Chlorine 1,5 mg/l 1 Pieces 48105510
Sampling
1. When preparing the sample, Chlorine outgassing, e.g. through the pipette or
shaking, must be avoided.
2. The analysis must take place immediately after taking the sample.
Chlorine L / M101
ENMethod Reference Book1.0 15
EN
Preparation
1. Cleaning of vials:
As many household cleaners (e.g. dishwasher detergent) contain reducing
substances, this can lead to lower results with the determination of Chlorine.
To avoid measurement errors, the glassware used should be free of chlorine
consumption. To achieve this, all glassware should be placed in a sodium
hypochlorite solution (0.1 g/L) for one hour and then rinsed thoroughly with
deionised water.
2. For individual testing of free and total Chlorine, the use of different sets of
glassware is recommended (EN ISO 7393-2, 5.3)
3. The DPD colour development is carried out at a pH value of 6.2 to 6.5. The
reagents therefore contain a buffer for the pH adjustment. Strong alkaline or acidic
water samples must therefore be adjusted between pH 6 and pH 7 before the
analysis (use 0.5 mol/l Sulphuric acid or 1 mol/l Sodium hydroxide).
Remarks
1. After use, ensure the cuvettes are once again closed with the respective same-
coloured screw caps.
2. Reagent sets are to be stored in the cool at +6 °C to +10 °C.
Chlorine L / M101
ENMethod Reference Book1.016
EN
Determination of free chlorine with liquid reagent
Select the method on the device.
10 mL
Sample
Fill 24 mm vial with 10 mL
sample.
Close vial(s). Place sample vial in the
sample chamber. Pay
attention to the positioning.
Zero
Press the ZERO button. Remove the vial from the
sample chamber.
Empty vial.
Sample
6
Sample
2
Hold cuvettes vertically
and add equal drops by
pressing slowly.
Add 6 drops DPD 1 Buffer
Solution to the sample
vial.
Add 2 drops DPD
1 Reagent Solution to the
sample vial.
Chlorine L / M101
ENMethod Reference Book1.0 17
EN
Fill up vial with sample to
the 10 mL mark.
Close vial(s). Invert several times to mix
the contents.
Sample
Test
Place sample vial in the
sample chamber. Pay
attention to the positioning.
Press the TEST (XD:
START)button.
The result in mg/L free chlorine appears on the display.
Determination of totale Chlorine with liquid reagent
Select the method on the device.
10 mL
Sample
Fill 24 mm vial with 10 mL
sample.
Close vial(s). Place sample vial in the
sample chamber. Pay
attention to the positioning.
Chlorine L / M101
ENMethod Reference Book1.018
EN
Zero
Press the ZERO button. Remove the vial from the
sample chamber.
Empty vial.
Sample
6
Sample
2
Hold cuvettes vertically
and add equal drops by
pressing slowly.
Add 6 drops DPD
1 Buffer Solution to the
sample vial.
Add 2 drops DPD
1 Reagent Solution to the
sample vial.
Sample
3
Add 3 drops DPD
3 Solution to the sample
vial.
Fill up vial with sample to
the 10 mL mark.
Close vial(s).
Chlorine L / M101
ENMethod Reference Book1.0 19
EN
Sample
Test
Invert several times to mix
the contents.
Place sample vial in the
sample chamber. Pay
attention to the positioning.
Press the TEST (XD:
START)button.
9
4
5
6
7
8
10
11 12
1
2
3
2 min
Wait for 2 minute(s)
reaction time.
Once the reaction period is finished, the measurement takes place automatically.
The result in mg/L total Chlorine appears on the display.
Chlorine L / M101
ENMethod Reference Book1.020
EN
Chemical Method
DPD
Appendix
Interferences
Persistant Interferences
All oxidising agents in the samples react like chlorine, which leads to higher results.
Removeable Interferences
Interference from Copper and Iron (III) are eliminated by the addition of EDTA.
Concentrations above 4 mg/L Chlorine, in the event of using fluid reagents, can lead
to results within the measuring range of up to 0 mg/L. In this case, the sample must be
diluted with chlorine-free water. 10 ml of the diluted sample should be mixed with the
reagent and the measurement taken again (plausibility test).
Interference from / [mg/L]
CrO4 2- 0,01
MnO20,01
Conformity
EN ISO 7393-2
a) determination of free, combined and total
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114
  • Page 115 115
  • Page 116 116
  • Page 117 117
  • Page 118 118
  • Page 119 119
  • Page 120 120
  • Page 121 121
  • Page 122 122
  • Page 123 123
  • Page 124 124
  • Page 125 125
  • Page 126 126
  • Page 127 127
  • Page 128 128
  • Page 129 129
  • Page 130 130
  • Page 131 131
  • Page 132 132
  • Page 133 133
  • Page 134 134
  • Page 135 135
  • Page 136 136
  • Page 137 137
  • Page 138 138
  • Page 139 139
  • Page 140 140
  • Page 141 141
  • Page 142 142
  • Page 143 143
  • Page 144 144
  • Page 145 145
  • Page 146 146
  • Page 147 147
  • Page 148 148
  • Page 149 149
  • Page 150 150
  • Page 151 151
  • Page 152 152
  • Page 153 153
  • Page 154 154
  • Page 155 155
  • Page 156 156
  • Page 157 157
  • Page 158 158
  • Page 159 159
  • Page 160 160
  • Page 161 161
  • Page 162 162
  • Page 163 163
  • Page 164 164
  • Page 165 165
  • Page 166 166
  • Page 167 167
  • Page 168 168
  • Page 169 169
  • Page 170 170
  • Page 171 171
  • Page 172 172
  • Page 173 173
  • Page 174 174
  • Page 175 175
  • Page 176 176
  • Page 177 177
  • Page 178 178
  • Page 179 179
  • Page 180 180
  • Page 181 181
  • Page 182 182
  • Page 183 183
  • Page 184 184
  • Page 185 185
  • Page 186 186
  • Page 187 187
  • Page 188 188
  • Page 189 189
  • Page 190 190
  • Page 191 191
  • Page 192 192
  • Page 193 193
  • Page 194 194
  • Page 195 195
  • Page 196 196
  • Page 197 197
  • Page 198 198
  • Page 199 199
  • Page 200 200
  • Page 201 201
  • Page 202 202
  • Page 203 203
  • Page 204 204
  • Page 205 205
  • Page 206 206
  • Page 207 207
  • Page 208 208
  • Page 209 209
  • Page 210 210
  • Page 211 211
  • Page 212 212
  • Page 213 213
  • Page 214 214
  • Page 215 215
  • Page 216 216
  • Page 217 217
  • Page 218 218
  • Page 219 219
  • Page 220 220
  • Page 221 221
  • Page 222 222
  • Page 223 223
  • Page 224 224
  • Page 225 225
  • Page 226 226
  • Page 227 227
  • Page 228 228
  • Page 229 229
  • Page 230 230
  • Page 231 231

Lovibond 00386467 - MD1x0/200 - Chlorine HR (KI) T, Chlorine dioxide T, Chlorine L, Chlorine T Manuel utilisateur

Taper
Manuel utilisateur