Thermo Fisher Scientific PrepSEQ Nucleic Acid Sample Preparation Kit Mode d'emploi

Taper
Mode d'emploi
For Research Use Only. Not for use in diagnostic procedures.
PrepSEQ Nucleic Acid Sample Preparation
Kit
USER GUIDE
Catalog NumbersA50485
Publication Number MAN0026641
Revision A.0
The customer is responsible for validation of assays and compliance with regulatory requirements that pertain to their procedures and
uses of the instrument.
The information in this guide is subject to change without notice.
DISCLAIMER: TO THE EXTENT ALLOWED BY LAW, THERMO FISHER SCIENTIFIC INC. AND/OR ITS AFFILIATE(S) WILL NOT BE
LIABLE FOR SPECIAL, INCIDENTAL, INDIRECT, PUNITIVE, MULTIPLE, OR CONSEQUENTIAL DAMAGES IN CONNECTION WITH OR
ARISING FROM THIS DOCUMENT, INCLUDING YOUR USE OF IT.
Revision history: Pub. No.MAN0026641
Revision Date Description
A.0 28 July 2022 New document for the PrepSEQ Nucleic Acid Sample Preparation Kit
(Cat.No.A50485). Content previously in Pub.No.4469838.
Important Licensing Information: These products may be covered by one or more Limited Use Label Licenses. By use of these
products, you accept the terms and conditions of all applicable Limited Use Label Licenses.
TRADEMARKS: All trademarks are the property of Thermo Fisher Scientific and its subsidiaries unless otherwise specified.
©2022 Thermo Fisher Scientific Inc. All rights reserved.
Contents
CHAPTER1Productinformation .................................................. 5
Productdescription ............................................................. 5
Contents and storage ............................................................ 5
Required materials not supplied for manual protocols ................................ 6
Required materials not supplied for automated protocols ............................. 7
Workflow ....................................................................... 8
CHAPTER2Prepare the reagents andsamples ................................. 9
Prepare the reagents: before first use of thekit ...................................... 9
Magneticbeads ............................................................ 9
BindingSolution ............................................................ 9
Wash Buer Concentrate .................................................... 9
Prepare reagents: before each use of thekit ........................................ 9
Proteinase K (PK)mix ....................................................... 9
Lysissolution ............................................................. 10
Guidelines for optimalyields ..................................................... 10
Sample preparationguidelines ................................................... 11
Sample dilution (if necessary) ............................................... 11
Triplicate extractions ....................................................... 11
Extraction controlguidelines ................................................ 12
CHAPTER3Manual protocol for DNA/RNA extraction ........................ 13
Digest the test samples and controls ............................................. 13
Bind theDNA/RNA ............................................................. 13
Wash theDNA/RNA ............................................................ 14
Elute theDNA/RNA ............................................................. 15
CHAPTER4Automated protocol for DNA/RNA extraction ................... 16
Before each use of thekit ....................................................... 16
Ensure that you have the correct plates ....................................... 16
Prepare the plates ......................................................... 16
Prepare the lysis plate .......................................................... 17
Process samples on theinstrument .............................................. 17
PrepSEQ Nucleic Acid Sample Preparation Kit User Guide 3
APPENDIXATroubleshooting .................................................... 19
APPENDIXBGood laboratory practices ....................................... 20
Good laboratory practices for PCR and RT-PCR ................................... 20
Avoiding false positives due to cross-contamination ........................... 20
Plate layout suggestions ....................................................... 20
APPENDIXCSafety ............................................................... 21
Chemicalsafety ................................................................ 22
Biological hazardsafety ......................................................... 23
Documentation and support ....................................................... 24
Relateddocumentation ......................................................... 24
Customer and technical support ................................................. 24
Limited product warranty ........................................................ 24
Contents
4PrepSEQ Nucleic Acid Sample Preparation Kit User Guide
Product information
IMPORTANT! Before using this product, read and understand the information in the “Safety” appendix
in this document.
Product description
The PrepSEQ Nucleic Acid Sample Preparation Kit uses chemical lysis and magnetic beads to
extract DNA from diverse sample types, including samples that contain high protein and low DNA
concentration.
The kit extracts nucleic acid (DNA or RNA) from common sample matrices in dierent steps of
the bioproduction workflow. For detection/quantification of viral RNA or DNA, use of the ViralSEQ
RealTime PCR Kits as described in the respective assay user guides.
To ensure accurate quantitative results, extract each sample in triplicate and perform a single PCR
reaction for each extraction.
Contents and storage
Table1PrepSEQ Nucleic Acid Sample Preparation Kit (Cat. No. A50485)
Contents Amount Storage
Box 1, PrepSEQ Nucleic Acid Extraction Kit
Lysis Buer 2×50mL
Room temperature
Binding Solution (Isopropanol), empty bottle 1
Wash Buer Concentrate 2×26mL
Elution Buer 25mL
Proteinase K (PK) Buer
Can be used for existing validated manual protocols.
50mL
ProteinaseK (PK) BuerII[1]
Recommended for new manual protocols.
Required for automated protocols.
11mL
Box 2, PrepSEQ Nucleic Acid Extraction Kit
Magnetic Particles 2×1.5mL Room temperature
1
PrepSEQ Nucleic Acid Sample Preparation Kit User Guide 5
Table 1 PrepSEQ Nucleic Acid Sample Preparation Kit (Cat. No. A50485) (continued)
Contents Amount Storage
Box 3, PrepSEQ Nucleic Acid Extraction Kit
Proteinase K, 20mg/mL 1.25mL –20°C or below
PrepSEQ Residual DNA Sample Preparation Kit
Proteinase K, 20mg/mL 1.25mL
–20°C or belowYeast tRNA, 10mg/mL 0.5mL
Glycogen, 5mg/mL 2×1.0mL
[1] Also sold separately (Cat.No.4415320).
Required materials not supplied for manual protocols
Unless otherwise indicated, all materials are available through thermofisher.com. "MLS" indicates that
the material is available from fisherscientific.com or another major laboratory supplier.
Item Source
Equipment
Magnetic stand, 16-position 12321D
Block heater for use with 2mL tubes.
Manual DNA/RNA extraction involves two incubations at
dierent settings, so two heaters may be convenient.
MLS
Benchtop microcentrifuge for 1.5mL and 2mL tubes MLS
Vortex MLS
Vortex Adapter-60, for use with the Vortex-GenieAM10014
Consumables
Disposable gloves MLS
Aerosol-resistant micropipette tips MLS
Pipettors, P1000, P200, P20 and P10:
Positive-displacement
Air-displacement
Multichannel
MLS
Nonstick, RNase-free Microfuge Tubes, 1.5mL (1 box;
250 tubes/box)
AM12450
Safe-Lock Tube, 2.0mL VWR 62111-754
Chapter1Product information
Required materials not supplied for manual protocols
1
6PrepSEQ Nucleic Acid Sample Preparation Kit User Guide
Item Source
Reagents
Ethanol, 95%
IMPORTANT! Do not use denatured ethanol. It
contains components that are not compatible with the
protocol.
MLS
Isopropanol, 100% MLS
1X PBS (free of Mg and Ca) MLS
Required materials not supplied for automated protocols
Unless otherwise indicated, all materials are available through thermofisher.com. "MLS" indicates that
the material is available from fisherscientific.com or another major laboratory supplier.
Table2Pharma KingFisher Flex Purification System with 96 Deep-Well Head (Cat.No.A31508) and
Pharma MagMAX Express-96 instrument[1] accessories
Item Source
Pharma MagMAX Express-96 DW plate A31540
Pharma MagMAX Express-96 Deep-Well Tip Combs A31537
Pharma KingFisher Flex Magnetic Head for 96 Deep-Well Plate A31542
Pharma MagMAX 96 PCR Well Magnetic Head 4472991
Pharma MagMAX Express-96 Standard Plates A31541
(Optional) Pharma Magnetic Stand-96 A31543
[1] This instrument is no longer available for purchase
Table3Additional materials
Item Source
Consumables
Disposable gloves MLS
Aerosol-resistant micropipette tips MLS
Chapter1Product information
Required materials not supplied for automated protocols 1
PrepSEQ Nucleic Acid Sample Preparation Kit User Guide 7
Item Source
Pipettors, P1000, P200, P20 and P10:
Positive-displacement
Air-displacement
Multichannel
MLS
Nonstick, RNase-free Microfuge Tubes, 1.5mL (1 box; 250
tubes/box)
AM12450
Reagents
Ethanol, 95%
IMPORTANT! Do not use denatured ethanol. It contains
components that are not compatible with the protocol.
MLS
Isopropanol, 100% MLS
1X PBS (free of Mg and Ca) MLS
Workflow
Manual DNA/RNA extraction Automated DNA/RNA extraction
Digest the test samples and controls
(page13)
Prepare the plates (page16)
▼ ▼
Bind the DNA/RNA (page13) Prepare the lysis plate (page17)
▼ ▼
Wash the DNA/RNA (page14) Process samples on the instrument
(page17)
Elute the DNA/RNA (page15)
Chapter1Product information
Workflow
1
8PrepSEQ Nucleic Acid Sample Preparation Kit User Guide
Prepare the reagents and samples
Prepare the reagents: before first use of the kit
Magnetic beads
1. Set a block heater to 37°C.
2. Incubate the Magnetic Particle suspension at 37°C for a minimum of 10 minutes with intermittent
vortexing at 900 rpm, or until the particles are completely suspended.
Binding Solution
1. Add 45 mL of 100% isopropanol to the Binding Solution bottle.
2. Label the bottle to indicate that it contains isopropanol, then store the bottle at room temperature.
Wash Buer Concentrate
1. Add 74 mL of 95% ethanol to one bottle of PrepSEQ Wash Buer Concentrate, then mix
completely.
2. Label the bottle to indicate that it contains ethanol, then store the bottle at room temperature.
Prepare reagents: before each use of the kit
Proteinase K (PK) mix
Use ProteinaseK (PK) BuerII for all protocols.
Note: Proteinase K (PK) Buer is provided in the kit for use by laboratories that had previously
validated with this buer. If required, use exactly as described for ProteinaseK (PK) BuerII.
Prepare a fresh mix before each use of the kit.
Include a 10% overage to account for pipetting losses.
Component
Number of extractions
1 7 10 13 25
Proteinase K, 20mg/mL 10μL 70μL 100μL 130μL 250μL
ProteinaseK (PK) BuerII 60μL 420μL 600μL 780μL 1,500μL
2
PrepSEQ Nucleic Acid Sample Preparation Kit User Guide 9
Lysis solution
Prepare a fresh mixture immediately before use or during Proteinase K incubation.
Prepare 360 µL (amount required) of lysis solution mix per sample.
Reagent Volume for 1 extraction Volume for ~20 extractions
Glycogen, 5mg/mL 8.32µL 180µL
Yeast tRNA, 10mg/mL 0.18µL 4µL
Lysis Buer 351.5µL 7,600µL
Total 360µL 7,784µL
Guidelines for optimal yields
Maintain a homogenous suspension of the magnetic beads to maximize the surface area to which
the DNA/RNA can bind. The appearance of the mixture should be homogenous after mixing.
After drying, the DNA/RNA remains bound to the magnetic beads. Do not allow the magnetic
beads to over-dry because this reduces the elution eciency; over-dried beads are not easily
resuspended.
During manual elution, vortex every 2 minutes to assist elution. This will result in better yield during
recovery.
Note: Some test samples cause the beads to adhere very firmly to the tube wall, while others form
loose pellets that detach during the vortex steps. All pellets should dissolve with vortexing during
heated elution. If vortexing does not result in full resuspension, then wash the beads o the tube by
pipetting.
Note: White or brown precipitate may form in the Magnetic Particles tube if it is stored at 2–8°C.
The precipitate will dissolve when it is heated to 37°C for a minimum of 10 minutes with intermittent
vortexing. Make sure the precipitate is completely dissolved before using the beads.
Chapter2Prepare the reagents and samples
Guidelines for optimal yields
2
10 PrepSEQ Nucleic Acid Sample Preparation Kit User Guide
Sample preparation guidelines
Sample dilution (if necessary)
Test samples from the early purification process often contain levels of DNA/RNA that are above the
highest point of the assay standard curve. You must dilute these samples (from 1:100 up to 1:10,000)
before PrepSEQ sample preparation.
Dilute test samples before DNA/RNA extraction with a solution of 1X PBS (free of Mg and Ca) or
50 mM Tris, pH 8.0, 0.5M NaCl.
Note: Diluting samples in water or TE reduces extraction eciency.
Use the sample dilution buer as the negative extraction control instead of water.
Alternatively, dilute extracted DNA/RNA with elution buer before running the PCR reaction.
Triplicate extractions
Triplicate extractions are required for post-PCR analysis calculation of mean quantity, standard
deviation, and coecient of variation.
In addition to test samples, we recommend triplicate extractions for the negative control and the
extraction/recovery control (ERC).
Perform a single PCR reaction for each extraction.
The table below illustrates the total number of extractions required based on the 1, 2, and 3 samples
extracted in a batch.
Table4Total number of extractions per batch of test samples
Number of test samples Total number of extractions for the batch
1 3 extractions required for each:
Test sample
Test sample extraction/recovery
control (ERC)
Negative extraction control[1]
9
2 15
3 21
[1] Optional during routine testing.
Chapter2Prepare the reagents and samples
Sample preparation guidelines 2
PrepSEQ Nucleic Acid Sample Preparation Kit User Guide 11
Extraction control guidelines
We recommend that you use the following extraction controls:
Type of control Contains Number to run Used to
Negative (NEG)[1] 1X PBS (free of Mg and Ca) 1 per batch of
extractions
Test for cross-contamination of
DNA/RNA extraction reagents.
Extraction/recovery (ERC) Positive control from the kit 3 per sample Evaluate the eciency of DNA
or RNA extraction, recovery,
and quantification from test
samples.
Verify reagent and system
performance.
[1] Optional during routine testing.
For the Extraction/recovery (ERC) :
Prepare the control standard dilutions as described in the relevant ViralSEQ user guide for your
assay.
Add a volume of positive control standard dilution to each test sample so that the total DNA/RNA
amount is 2–10 times the amount of DNA/RNA measured in the test sample without the addition of
the DNA/RNA control.
For example (amount):
The DNA amount measured in a test sample is ≤1 pg.
To prepare a 10 pg ERC for a PCR elution volume of 50 µL, spike samples with 16.7µL of the 3
pg/µL positive control standard dilution (SD3) = 50 pg spike to yield 10 pg per PCR reaction.
For example (copies):
The number of copies in a test sample is ≤2,000 copies.
Spike control sample at 1x 105 copies using appropriate standard control concentration.
If 1:20 (10 µL out of 200 µL eluate) of the KingFisher Flex extraction is used for qPCR, there
are 5,000 expected copies of spike control per PCR reaction.
If 1:5 (10 µL out of 50 µL eluate) of the manual extraction is used for qPCR, there are 20,000
expected copies of spike control per PCR reaction.
Prepare three separate extractions for each test sample, then add the ERC to each reaction. Do not
prepare a large volume of ERC, then aliquot it into three reactions.
Note: To calculate the eciency of DNA/RNA recovery and quantification from the test samples,
subtract the amount of DNA/RNA measured in the sample without the addition of DNA/RNA control
from the amount of DNA/RNA measured in the ERC sample.
Chapter2Prepare the reagents and samples
Sample preparation guidelines
2
12 PrepSEQ Nucleic Acid Sample Preparation Kit User Guide
Manual protocol for DNA/RNA
extraction
Digest the test samples and controls
1. Set a block heater to 56°C. If available, set a second block heater to 70°C.
2. Label the Safe-Lock Tubes, 2.0mL:
3 for each sample
3 for each sample + ERC
3 for NEG
3. Add 100 µL of sample, sample + ERC, or 1X PBS (free of Mg and Ca) into each tube.
4. Add 10µL of 5M NaCl and 70μL Proteinase K/Proteinase K Buer II mix.
5. Briefly vortex and centrifuge.
6. Incubate at 56°C for 30 minutes.
If only one block heater is available, after this incubation step is complete, reset the block heater to
70°C for the elution step.
Note: For samples with high protein concentration, extending the incubation time to 60 minutes
can increase recovery.
7. Cool samples to room temperature.
8. Add 360 µL freshly made Lysis solution mix to each tube.
Bind the DNA/RNA
1. Vortex the Magnetic Particles to resuspend the particles.
Note: The appearance of the mixture should be homogeneous.
2. Add 30µL of the Magnetic Particles to each tube.
3. Add 400µL Binding Solution to the first sample, then close the cap and invert twice to mix.
3
PrepSEQ Nucleic Acid Sample Preparation Kit User Guide 13
4. Repeat step3 for each additional sample.
Note: Low DNA recovery can be observed if the sample and the Binding Solution are not
immediately mixed.
5. Vortex all the tubes in the vortex adaptor for 5 minutes at 900rpm.
6. Briefly centrifuge the tubes for 15 seconds at top speed (>15,000 ×g) to collect the
Magnetic Particles at the bottom of the tubes.
7. Place the tubes in the magnetic stand with the pellet against the magnet, then let the tubes stand
for 5 minutes or until the solution is clear.
8. Without disturbing the magnetic beads, remove the supernatant using a pipette or by aspiration.
Wash the DNA/RNA
1. Remove the tube rack (with tubes) from the magnetic stand, then add 300 µL of Wash Solution to
the tubes. Vortex the tubes for 5 seconds at room temperature at 900rpm.
2. Centrifuge the tubes in a microcentrifuge at top speed (>15,000 ×g) for a maximum of 20seconds.
Do not centrifuge for >20 seconds.
3. Place the tubes in the magnetic stand, then let the tubes stand for 1 minute.
Note: The Magnetic Particles with the bound DNA/RNA are magnetically captured after
approximately 1 minute.
4. Without disturbing the Magnetic Particles, remove the supernatant by pipette or by aspiration.
5. Remove the tube rack (with tubes) from the magnetic stand, then add 300 µL of Wash Solution to
each tube for a second wash. Vortex the tubes for 5 seconds at room temperature at 900rpm.
6. Centrifuge the tubes in a microcentrifuge at top speed (>15,000 ×g) for a maximum of 20seconds.
Do not centrifuge for >20 seconds.
7. Place the tubes in the magnetic stand, then let the tubes stand for 1 minute.
Note: The Magnetic Particles with the bound DNA/RNA are magnetically captured after
approximately 1 minute.
8. Open all tubes, then start the 5minute timer.
9. Without disturbing the Magnetic Particles, remove the supernatant by pipette or aspiration.
Use a P200 to remove the remaining solution from the bottom of the tube.
Chapter3Manual protocol for DNA/RNA extraction
Wash the DNA/RNA
3
14 PrepSEQ Nucleic Acid Sample Preparation Kit User Guide
10. With the tube lid open, air-dry the Magnetic Particles pellet in the magnetic stand for no more than
5 minutes at room temperature.
IMPORTANT! Air-dry to remove ethanol from the Wash Solution. Once dry, the DNA/RNA stays
bound to the magnetic beads. Do not over-dry; over-dried beads are not easily resuspended.
Elute the DNA/RNA
1. Add 50µL of Elution Buer to each tube.
Note: Add 200µL of Elution Buer to each tube when intended for use with the ViralSEQ
Lentivirus Physical Titer Kit.
2. Vortex the tubes for 20 seconds at high speed, then incubate the tubes at 70°C for 7 minutes.
Vortex the tubes two to three times during the incubation to help resuspend the beads.
3. Centrifuge the tubes in a microcentrifuge at top speed (>15,000 ×g) for a maximum of 20seconds.
Do not centrifuge for >20 seconds.
4. Place the tubes in the magnetic stand, then let the tubes stand for 1 minute.
5. Without disturbing the Magnetic Particles, transfer the liquid phase containing the eluted DNA/RNA
to a new nonstick 1.5mL microcentrifuge tube.
6. Centrifuge the tube at top speed (>15,000 ×g) for 3 minutes to collect the Magnetic Particles at
the bottom of the tube, then place the tubes in the magnetic stand for 1 minute.
7. Without disturbing the Magnetic Particles, transfer the liquid phase containing the eluted DNA/RNA
to a new nonstick 1.5mL microcentrifuge tube.
Note: Magnetic Particles can inhibit PCR.
Store eluted DNA/RNA for up to 6 hrs on ice, or up to 24hrs at −20°C.
Chapter3Manual protocol for DNA/RNA extraction
Elute the DNA/RNA 3
PrepSEQ Nucleic Acid Sample Preparation Kit User Guide 15
Automated protocol for DNA/RNA
extraction
You can use the KingFisher Flex or MagMAX Express 96-deep well automation platforms to
automate the extraction of DNA/RNA. For all chemicals, read the Safety Data Sheet (SDS) and follow
the handling instructions. Wear appropriate protective eyewear, clothing, and gloves.
Before each use of the kit
Ensure that you have the correct plates
The KingFisher Flex or the MagMAX Express require 5 plates.
Plate name Plate type
Lysis 96 deep-well plate
Wash 1 96 deep-well plate
Wash 2 96 deep-well plate
Elution 96 deep-well plate
Comb loading plate 96 deep-well tip comb combined with 96 standard plate
Prepare the plates
Prepare the Wash 1, Wash 2, and Elution plates:
Plate name Plate type Volume of buer to add
Wash 1 96 deep-well plate 300 µL of Wash buer
Wash 2 96 deep-well plate 300 µL of Wash buer
Elution 96 deep-well plate 200µL of Elution buer
4
16 PrepSEQ Nucleic Acid Sample Preparation Kit User Guide
Prepare the lysis plate
In all steps that require pipetting, dispense liquid at bottom center of the wells.
1. Add 100μL to the appropriate wells of the 96 deep-well Lysis plate:
3 wells for each sample
3 wells for each sample + ERC
3 wells for NEG
2. Add 10μL of 5 M NaCl to each sample well.
3. Add 70 µL Proteinase K/ProteinaseK (PK) BuerII mix to each sample well.
Process samples on the instrument
1. Select the script for the instrument and kit that you are using:
Kit Instrument Script
PrepSEQ Nucleic Acid
Sample Preparation Kit
KingFisher Flex PrepSEQ_NucleicAcid_v1
2. Load the plates into the instrument in the order listed below. After loading each plate, press START
to move the turntable.
a. Comb loading plate
b. Elution plate with 200μL of Elution Buer
c. Wash 2 plate with 300μL of wash buer
d. Wash 1 plate with 300μL of wash buer
e. Lysis plate
3. Press START to begin the PK digestion process.
The instrument mixes the samples for 10 seconds at fast speed, then incubates the samples at
56°C for 30 minutes, mixing at slow speed. When digestion is complete, the instrument pauses
and returns the Lysis plate to the loading position.
4. After the digestion step is complete, add additional components to the Lysis plate:
a. Remove the Lysis plate from the instrument.
b. Add 360 µL of Lysis Solution to each sample well.
c. Add 30 µL of Magnetic Particle suspension to each sample well.
Chapter4Automated protocol for DNA/RNA extraction
Prepare the lysis plate 4
PrepSEQ Nucleic Acid Sample Preparation Kit User Guide 17
d. Add 400 µL of Binding Solution to the first sample well, then immediately pipet up-and-down
three times to mix. Repeat for each additional sample well.
Note: Low DNA recovery can be observed if the sample and the Binding Solution are not
immediately mixed.
e. Place the plate back into the instrument loading position, then press START to begin binding.
5. When DNA/RNA extraction is finished, the instrument returns the Elution plate to the loading
position.
Store eluted DNA/RNA for up to 6 hrs on ice, or up to 24hrs at −20°C.
Upon completion of the sample extraction procedure, see the relevant assay user guide for specific
assay instructions.
Chapter4Automated protocol for DNA/RNA extraction
Process samples on the instrument
4
18 PrepSEQ Nucleic Acid Sample Preparation Kit User Guide
Troubleshooting
Observation Possible cause Action
Poor extraction eciency (low yields) Overdrying the sample. Start the 5minute timer before
removing ~300 µL from the first 6–
8 samples. Then continue removing
wash buer from the remaining
samples.
Magnetic Particles are dicult to
resuspend during the elution.
Incubate the pellets at 70°C for
>7 minutes. Vigorously vortex the
tubes three times during incubation
to help resuspension.
Do not overdry.
If necessary, repeat the incubation
and vortexing steps.
Formation of precipitate in Magnetic
Particles.
Incubate the Magnetic Particle
suspension at 37°C with intermittent
vortexing at 900rpm until the
particles are completely suspended.
PK Buer was used instead of PKII
Buer.
Use PKII Buer.
Particles no longer produce
consistent results (fine brown sandy
particles and brown color are
observed in the supernatant)
Samples have low pH. Measure the pH of the sample and
adjust the pH to between 6 and 8.
Magnetic Particles were stored
at −20°C.
Order new materials and store them
at room temperature.
Clumps or white precipitate in the
Magnetic Particles tube
The Magnetic Particles were shipped
or stored at temperatures below
room temperature.
Incubate the Magnetic Particle tube
at 37°C for 30 min with intermittent
vortexing at 900 rpm. Incubate and
vortex until all clumps or white
precipitate is dissolved.
A
PrepSEQ Nucleic Acid Sample Preparation Kit User Guide 19
Good laboratory practices
Good laboratory practices for PCR and RT-PCR
Wear clean gloves and a clean lab coat.
Do not wear the same gloves and lab coat that you have previously used when handling
amplified products or preparing samples.
Change gloves if you suspect that they are contaminated.
Maintain separate areas and dedicated equipment and supplies for:
Sample preparation and reaction setup.
Amplification and analysis of products.
Do not bring amplified products into the reaction setup area.
Open and close all sample tubes carefully. Avoid splashing or spraying samples.
Keep reactions and components capped as much as possible.
Use a positive-displacement pipettor or aerosolresistant barrier pipette tips.
Clean lab benches and equipment periodically with a 10% bleach solution or decontamination
solutions (for example, DNAZap PCR DNA Degradation Solutions (Cat.No.AM9890) and
RNaseZap RNase Decontamination Solution (Cat.No.AM9780).
Avoiding false positives due to cross-contamination
To avoid false positives due to cross-contamination:
Prepare and close all negative control and unknown sample tubes before pipetting the positive
control.
Do not open tubes after amplification.
Use dierent sets of pipettors when pipetting negative control, unknown, and positive control
samples.
Plate layout suggestions
For each plate row, dispense in sequence from left to right the: negative controls, unknown
samples and ERCs, and positive controls (at the end of the row or column).
Place positive controls in one of the outer columns (10–12).
If possible, separate all samples from each other by at least one well; if space is limiting, place at
least one well between unknown samples and controls.
B
20 PrepSEQ Nucleic Acid Sample Preparation Kit User Guide
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25

Thermo Fisher Scientific PrepSEQ Nucleic Acid Sample Preparation Kit Mode d'emploi

Taper
Mode d'emploi