Active switch
Please refer to the user manual of
the LCC-S to make the conguration.
Output setting – active switch
To set the type of analogue output, please put the on-off switch of
the output as shown beside.
4-20 mA 0-10 V
1
2
3
4
1
2
3
4
Units setting – active switch
To set a measurement unit, put the on-off switches 2, 3 and 4
of the units as shown in the table below.
mbar mmHG hPa
1
2
3
4
1
2
3
4
1
2
3
4
Conguration via LCC-S software (option)
An easy and friendly conguration with the software!
To access the conguration via software:
• Set the DIP switches as shown beside. Nota: the on-off switch 1
of the active DIP switch can be in any position (selection of the
analogue output 0-10 V or 4-20 mA).
• Connect the cable of the LCC-S to the connection of the
transmitter.
Caution: the conguration of the parameters can be done
either with the DIP switch or via software (you can not
combine both solutions).
To mount the transmitter, mount the ABS plate on the
wall (drilling: Ø 6 mm, screws and pins are supplied).
Insert the transmitter on the xing plate (see A on
the drawing beside). Rotate the housing in clockwise
direction until you hear a “click” which conrms that
the transmitter is correctly installed.
7.5 mm
8 mm
4.5 mm
40 mm
50 mm
68 mm
75 mm
37.5 mm
23.75 mm
14 mm
A
A
Mounting
Accessories
Maintenance: please avoid any aggressive
solvent. Please protect the transmitter
and its probes from any cleaning product
containing formalin, that may be used for
cleaning rooms or ducts.
Precautions for use: please always use the device
in accordance with its intended use and within
parameters described in the technical features in order
not to compromise the protection ensured by the
device.
Please refer to the data sheet to get more information about available accessories.