Phcbi 170AICUVH-PE Mode d'emploi

Taper
Mode d'emploi

Ce manuel convient également à

Please read the operating instructions carefully before using this product, and keep the operating
instructions for future use.
See page 72 for model numbers.
Operating Instructions
CO2 Incubator
2
CONTENTS
INTRODUCTION .......................................................................................... 4
INTENDED USE AND PRECAUTIONS ........................................................ 4
PRECAUTIONS FOR SAFE OPERATION ................................................... 5
LABELS ON INCUBATOR .......................................................................... 10
SYMBOLS ON INCUBATOR ..................................................................... 11
ENVIRONMENTAL CONDITIONS .............................................................. 11
INCUBATOR COMPONENTS ................................................................... 12
Unit .......................................................................................................... 12
LCD touch panel ...................................................................................... 14
Remote alarm terminal ............................................................................ 16
INSTALLATION ........................................................................................... 17
Installation site ......................................................................................... 17
Installation ............................................................................................... 18
Connecting a CO2 gas cylinder ............................................................... 21
BEFORE COMMENCING OPERATION ..................................................... 22
Initial cleaning method............................................................................. 22
Removing inner attachments .................................................................. 23
Installing inner attachments ..................................................................... 25
Filling the humidifying pan ....................................................................... 26
FOR OPTIMAL CULTIVATION .................................................................... 27
Precautions for cultures........................................................................... 27
Preventing contamination ........................................................................ 28
CORRECT OPERATION ............................................................................ 29
BASIC OPERATION ON LCD TOUCH PANEL .......................................... 30
Operation from Menu key ........................................................................ 30
Entering numeric values into input window ............................................. 32
Operating slide keys ................................................................................ 33
BASIC PARAMETERS ................................................................................ 34
Setting temperature, CO
2
density, and high-limit temperature alarm
........... 34
Setting key lock ....................................................................................... 35
Turning off key lock ................................................................................. 35
ALARM PARAMETERS .............................................................................. 36
OPERATION/ALARM LOG ......................................................................... 38
Setting log interval ................................................................................... 38
Displaying operation log .......................................................................... 39
Exporting operation log ........................................................................... 41
Displaying alarm log ................................................................................ 43
Exporting alarm log ................................................................................. 44
OTHER PARAMETERS .............................................................................. 46
Setting date and time .............................................................................. 46
Setting brightness and sleep ................................................................... 47
3
CONTENTS
UV LAMP PARAMETERS ........................................................................... 48
Using the UV lamp .................................................................................. 48
Setting UV lamp ON period ..................................................................... 49
Lighting the UV lamp for 24 hours ........................................................... 50
H2O2 DECONTAMINATION ........................................................................ 51
H
2O2 decontamination ............................................................................. 51
Precautions when handling H2O2 reagent ............................................... 55
ELECTRIC LOCK (OPTIONAL) .................................................................. 56
Setting User-ID ........................................................................................ 56
Setting auto-lock ...................................................................................... 57
Using unlock key ..................................................................................... 59
Turning off auto-lock ................................................................................ 59
GAS AUTO-CHANGER (OPTIONAL) ......................................................... 60
Connecting CO2 gas cylinders ................................................................ 60
Automatic CO2 gas supply line changeover ............................................ 60
Manual CO2 gas supply line A/B changeover ......................................... 62
STD GAS AUTO-CALIBRATION KIT (OPTIONAL) .................................... 63
ALARMS, SAFETY, AND SELF-DIAGNOSIS ............................................. 65
TROUBLESHOOTING ................................................................................ 68
SPECIFICATIONS ...................................................................................... 70
PERFORMANCE ........................................................................................ 72
SAFETY CHECK SHEET ........................................................................... 73
English
4
INTRODUCTION
Ŷ Read the operating instructions carefully before using the product and follow the instructions for safe
operation.
Ŷ PHC Corporation takes no responsibility for safety if the product is not used as intended or is used with
any procedures other than those given in the operating instructions.
Ŷ Keep the operating instructions in a suitable place so that they can be referred to as necessary.
Ŷ The operating instructions are subject to change without notice for improvement of performance or
function.
Ŷ Contact our sales representative or agent if any page of the operating instructions is lost or the page
order is incorrect, or if the instructions are unclear or inaccurate.
Ŷ No part of the operating instructions may be reproduced in any form without the express written
permission of PHC Corporation.
IMPORTANT NOTICE
PHC Corporation guarantees this product under certain warranty conditions. However, please note that
PHC Corporation shall not be responsible for any loss or damage to the contents of the product.
INTENDED USE AND PRECAUTIONS
This equipment is designed for culture of cells, tissues, organs, and embryos.
Ŷ The adapted culture condition depends on the kind of sample. It is necessary to determine the culture
temperature, CO2 density, and culture period that is required.
Ŷ For culture of embryos, low O2 density is desirable. Using an O2/CO2 incubator is recommended.
Ŷ For IVF/ART use, special attention should be paid to traceability, since incidents may be disclosed
several months or years later, at the baby’s birth, or during its later life. It is therefore recommended that
the following data is maintained: product serial number, incubation term, and incubation parameters.
(Refer to details on MEDDEV 2.2/4.)
5
PRECAUTIONS FOR SAFE OPERATION
It is imperative that the user complies with the operating instructions
as they contain important safety advice.
Items and procedures are described so that this unit can be used correctly and safely.
Following these precautions will prevent possible injury to the user and any other person.
Precautions are illustrated in the following way:
WARNING
Warning indicates a potentially hazardous situation which, if not avoided,
could result in serious injury or death.
CAUTION
Failure to observe CAUTION signs could result in injury to personnel and
damage to the unit and associated property.
Symbols have the following meaning;
This symbol means caution.
T
his symbol means an action is prohibited.
T
his symbol means an instruction must be followed.
Be sure to keep the operating instructions in a place that is accessible to users of this
unit.
WARNING
As with any equipment that uses CO2 gas, there is a likelihood of oxygen depletion in the vicinity
of the equipment. It is important that you assess the work site to ensure there is suitable and
sufficient ventilation. If restricted ventilation is suspected, then other methods of ensuring a
safe environment must be considered. These may include atmosphere monitoring and warning
devices.
English
6
PRECAUTIONS FOR SAFE OPERATION
Do not use the unit outdoors. Exposure to rain may cause leakage and/or electric shock.
Only qualified engineers or service personnel should install the unit. The installation by
unqualified personnel may cause electric shock or fire.
Install the unit in a location capable of bearing the total combined weight (product + optional
accessories + stored items). After installing the unit, be absolutely sure to take precautions to
prevent the unit from falling over. If the unit is installed in a location which is not strong enough or if
the proper precautions are not taken, the unit may fall over and cause injuries.
Do not install the unit where there are high levels of moisture or where it may be splashed with
water. Installing the unit where there are high levels of moisture or where it may be splashed with
water may cause the insulation to deteriorate and give rise to leakage and/or electric shock.
Do not install the unit in a location where flammable or volatile substances are present.
Installing the unit in a location where flammable or volatile substances are present may cause
explosions and/or a fire.
Do not install the unit in a location where corrosive gases such as acids are present. Installing
the unit in a location where corrosive substances are present may cause electric components to
corrode, leading to leakage and/or electric shock due to the deterioration of insulation resulting from
corroded electrical components.
Do not place this unit in a location where it is difficult to disconnect the power supply plug.
Failure to disconnect the power supply plug may cause fire in the event of a problem or malfunction.
Be absolutely sure to earth (ground) the unit in order to prevent electric shock. Failure to
earth the product may give rise to electric shock. If necessary, ask a qualified contractor to do this
work.
Do not connect the earth wire to a gas pipe, water pipe, or lightning rod when earthing the
unit. Earthing the unit improperly may give rise to electric shock.
Connect the unit to a power source as indicated on the rating label attached to the unit. Use
of any other voltage or frequency other than that on the rating label may cause fire or electric shock.
Never store volatile or flammable substances in this unit except in a sealed container.
Such substances may cause explosion or fire if they leak.
Never insert metal objects such as pins and wires into any vent, gap or outlet on the unit.
This may cause electric shock or injury by accidental contact with moving parts.
When handling harmful samples (for example, those which consist of toxic, pathogenic,
or radioactive substances), install the unit inside a designated isolation facility. If the
unit is installed in a location which is not an isolation facility, there may be detrimental effects on
both people and the natural environment.
WARNING
7
Before proceeding with maintenance or checking of the unit, set the power switch to OFF and
disconnect the power supply plug. Performing the work while power is still flowing to the product or
while the power-supply plug is still connected may give rise to electric shock and/or injury.
Do not touch any electrical parts (such as power supply plug) or operate switches with a wet
hand. This may cause electric shock.
Wear protective gloves and mask during maintenance. Touching or inhaling chemicals or
aerosols from around the unit may be detrimental to health.
Never splash water directly onto the unit as this may cause electric shock or short circuit.
Never put containers with liquid on top of the unit as this may cause electric shock or short circuit if
the liquid is spilled.
Never damage the power supply cord or power supply plug (by breaking, adapting, placing near
a source of heat, bending with force, twisting, pulling, adding weight, or binding). A damaged
power supply cord or power supply plug may cause electric shock, short circuit, or fire.
Never disassemble, repair, or modify the unit yourself. A high-voltage area is located inside the
unit. Any work carried out by an unauthorized personnel may result in electric shock. Contact our
sales representative or agent for maintenance or repair.
Make sure the power supply plug is pushed fully in. Faulty insertion of the power supply plug may
cause electric shock or fire due to generation of heat. Never use a damaged power supply plug or
loose power outlet.
Disconnect the power supply plug if there is anything wrong with the unit. Continued abnormal
operation may cause electric shock or fire.
Grip the power supply plug when disconnecting the power supply cord from the outlet. Pulling
the power supply cord may cause electric shock or short circuit.
Remove dust from the power supply plug periodically. Dust on the power supply plug may cause
insulation failure due to moisture and thus cause a fire. Disconnect the power supply plug and wipe it
with a dry cloth.
Disconnect the power supply plug before moving the unit. Take care not to damage the power
supply cord. A damaged cord may cause electric shock or fire.
Disconnect the power supply cord when the unit is not in use for long periods. Keeping the unit
connected may cause electric shock, leakage, or fire due to the deterioration of insulation.
WARNING
English
8
PRECAUTIONS FOR SAFE OPERATION
If the unit is to be stored unused in an unsupervised area for a long period, ensure that children do
not have access and that doors cannot be closed completely.
Ask a qualified contractor to carry out disassembly and disposal of the unit. Leaving the unit in
a location that can be accessed by third parties may result in unexpected accidents (e.g. the unit may
be used for unintended purposes).
Do not leave the plastic bags used for packing in a place where they can be reached by small
children as this may result in unexpected accidents such as suffocation.
Use the reagent specified by our company for H2O2 decontamination. Using a different H2O2
solution may result in explosion or damage to the incubator.
When performing H2O2 decontamination, securely close the internal and external doors. Failure to
do so may be harmful to health due to leakage of H2O2 gas.
During H2O2 decontamination, plug the access hole with the silicon cap that is provided. Failure to
do so may be harmful to health due to leakage of H2O2 gas.
Always use the removal power supply cord that is provided. Other power supply cord may cause
electric shock or fire.
When using CO2 gas for control, make sure that there is adequate ventilation. Using CO2 gas in a
small room without adequate ventilation may cause gas poisoning or oxygen deprivation. In addition,
when opening the incubator doors, do not directly inhale the air in the chamber.
When connecting a gas cylinder to the incubator, confirm the gas type. Confirm that the
connections are secure and that no gas will leak. Be sure to use the specified pressure. Using
an incorrect gas or pressure may result in explosion or fire, or in gas poisoning or oxygen deprivation
due to gas leak.
Install the incubator in a location with adequate ventilation. If adequate ventilation cannot be
provided, then install an alarm system using CO2 and O2 densitometers.
Do not look directly at UV light. UV light is harmful to the eyes.
Never start H2O2 decontamination when “Warning: UV Bulb Life” is displayed in the message display
field. The UV resolve is not sufficient.
Do not use the unlock key to unlock the outer door during H2O2 decontamination or during H2O2
gas resolve by UV. Doing so may cause harm to health from H2O2 gas leakage.
The CO2 incubator must be dismantled and disposed of by qualified personnel only. If the CO2
incubator is left where outsiders enter, it may result unexpected accident (for example, children to
become locked inside).
WARNING
9
This unit must be plugged into a dedicated circuit protected by branch circuit breaker.
Use a dedicated power source as indicated on the rating label attached to the unit. A multiple-tap may
cause fire resulting from abnormal heating.
Never store corrosive substances such as acid or alkali in this unit if the container cannot be
sealed. These may cause corrosion of inner components or electric parts.
Check the setting when starting up of operation after power failure or turning off of power
switch. The stored items may be damaged due to the change of setting.
Be careful not to tip over the unit during movement to prevent damage or injury.
Prepare a safety check sheet (copy the last page) when you request any repair or maintenance for
the safety of service personnel.
Wear rubber gloves when handling the H2O2 reagent. Direct contact with the H2O2 reagent may
result in inflammation of the skin.
H2O2 decontamination can be performed only for the chamber and chamber attachments with standard
specifications, and not for any other objects.
Perform H2O2 decontamination with the chamber attachments arranged as specified by our
company. Arranging them in a different way may result in insufficient decontamination.
After H2O2 decontamination has been completed, wear rubber gloves and use a non-woven cloth to
wipe off the residual H2O2 fluid from the bottom of the chamber, any objects that were
decontaminated, and the bottoms of ducts.
Do not climb on top of the incubator or place any object on it. By doing so you may damage the
incubator or knock it over, which may result in injury.
Wear rubber gloves when performing maintenance on the chamber. Failure to wear gloves may
result in cuts or abrasions from sharp edges or corners.
CAUTION
English
10
LABELS ON INCUBATOR
Users are advised to read carefully the warnings and cautions contained on stickers at key locations on the
interior and exterior of the incubator.
Possible
Danger Warning/Caution Type
Location of Danger Warning/Caution Label Description of Danger
Burns Hot Surface
Cooling Unit &
Heat Cover
Avoid touching the cooling unit
and heat cover, which can reach
high temperatures and may cause
burns.
Personal
injury Hazardous UV Light
Interior The UV lamp is illuminated by
pressing the door switch. Do not
press the door switch: the UV light
is hazardous.
Personal
injury Hazardous UV Light
Interior
The UV light is hazardous. Never
turn on the UV lamp without the
cover.
Personal
injury Gas Poisoning or
Oxygen Deprivation
Environment
When using CO2 gas, ensure
there is adequate ventilation.
Using CO2 gas in a small room
without adequate ventilation may
cause gas poisoning or oxygen
deprivation. In addition, when
opening the incubator doors, do
not directly inhale the air in the
chamber.
Personal
injury Gas Poisoning or
Oxygen Deprivation
Interior
Excessive pressure may cause
gas supply lines inside the
incubator to come loose, which
may result in gas poisoning or
oxygen deprivation due to the
escape of gas.
Personal
injury Electric shock
Top cover Never remove this cover. Only a
service personnel remove the
cover to avoid the electric shock.
11
SYMBOLS ON INCUBATOR
The following symbols are attached to the incubator:
Attached to covers that access high-voltage electrical components to prevent electric
shock. Only a qualified engineer or service personnel should be allowed to open
these covers.
Indicates an ultraviolet light (UV) caution.
Indicates that caution is required. Refer to precautions for safe operation for details.
Indicates a hot surface.
Indicates an earthing.
Indicates “ON” for a power switch.
Indicates “OFF” for a power switch.
ENVIRONMENTAL CONDITIONS
This equipment is designed to be safe at least under the following conditions (based on the IEC 61010-1):
Ŷ Indoor use;
Ŷ Altitude up to 2000 m;
Ŷ Temperature 5 oC to 40 oC;
Ŷ Maximum relative humidity 80 % for temperature up to 31 oC decreasing linearly to 50 % relative
humidity at 40 oC;
Ŷ Mains supply voltage fluctuations up to ±10 % of the nominal voltage;
Ŷ Transient overvoltages up to the levels of OVERVOLTAGE CATEGORY II;
Ŷ Temporary OVERVOLTAGES occurring on the mains supply;
Ŷ Applicable pollution degree of the intended environment (POLLUTION DEGREE 2 in most cases);
English
12
INCUBATOR COMPONENTS
Unit
1 2 3 Tray 4 5
6
7*1
8
13
12*
2 Electric key*2 11 10 9
*1: MCO-170AICUVH/170AICUV or when an
optional UV system set MCO-170UVS is installed.
*2: MCO-170AICUVH or when an optional electric
lock MCO-170EL is installed.
*3: Type FG-7P or FG-7PL.
(When some inner
attachments are removed)
20
21
Access port
cover Silicon cap
Rear side
19
Leveling
feet Leveling
Foot
Handles
(When 13 is removed)
Sample air outlet cap
18 17 16*3 15 14
Lower right side
Removable power
supply cord port
13
1. Outer door: The outer door is held to the frame with the magnetic seal. The door heater is installed in
the door panel. The door opening is reversible. Contact our sales representative or agent to change the
door hinge from left to right or vice versa.
2. Inner door: The inner door is made of tempered glass. Do not subject the glass to excessive impact.
Contact our sales representative or agent to change the inner door to optional small doors (MCO-170ID).
3. Tray catches: Insert tray to fit the concave portion on the chamber.
4. Fan cover (above the duct): Serves as the inlet for circulating air. It is removable.
5. Duct: The duct for the path for circulating air. It is removable.
6. Fan (inside the duct): The fan is made from polypropylene resin. It can be disinfected in an autoclave.
7. UV lamp*1:This UV lamp does not generate ozone. For replacement, contact our sales representative
or agent.
8. Humidity control bar: Automatically reduces condensation caused by the external environment when
the door is frequently opened and closed. The humidity control bar has a bactericidal effect due to its
plated surface. However, it is recommended that it is replaced every 5 to 6 years to maintain this effect
(The duration of bactericidal effect differs depending on the environment).
9. Humidifying pan cover: Prevents UV light entering the chamber.
10. Humidifying pan: Fill with sterile distilled water.
11. Door switch: Detects the door opening/closing and stops the fan and electromagnetic valve for CO2
when the door is open. The UV lamp*1 is also deactivated by the door opening.
12. Key hole*2: This is unlocked with the unlock key when the outer door is locked by the electric lock.
13. Switch cover: Prevents shutting down of the unit in case of accidental pressing of the main power
switch.
14. Power supply cord cover plate: Prevents the power supply cord becoming detached.
15. Connecting port A for CO2 gas pipe: Refer to page 21 for gas cylinder connection.
Note: When the optional MCO-21GC gas auto-changer is installed, both ports A and B are available. Refer
to page 60.
16. Glow starter*3:Starts the glow for the UV lamp.
17. Power switch: This is the main switch for the incubator (ON-“I”, OFF-“O”). It also functions as an
overcurrent breaker.
18. Sample air outlet: This also functions as an internal gas outlet. In normal use, cover this outlet with
the sample air outlet cap.
19. USB port: Insert USB memory device to export operations and alarms log. Refer to pages 41–45.
20. Access port: Place the silicon caps on both the outside and inside of the port when the port is not
being used.
21. Remote alarm terminal: Sends the alarm to a remote location by connecting to an external alarm unit.
Refer to page 16.
English
14
INCUBATOR COMPONENTS
LCD touch panel
The following display (called the Top screen) will appear when the power switch is turned ON.
Note: It takes approximately 20 seconds until the Top screen is displayed. During warming-up, “Status:
Gas sensor initializing” is displayed in the Message display field (13), and “--.-” is displayed in the present
CO2 density display field (4).
2 6
㻌 㻌
7 8
3
1
1. Present temperature display field
The current chamber temperature is displayed.
2. Set temperature value display field
The set value of chamber temperature is
displayed. Default setting: 37 oC
3. Heating indicator
This lights up when the heater is on.
4. Present CO2 density display field
The current chamber CO2 density is displayed.
Nothing is displayed when CO2 density is set to
0 %.
5. Set CO2 density value display field
The set value of the chamber CO2 density is
displayed. Default setting: 0 %.
6. CO2 gas injection indicator
This lights up when CO2 gas is being injected.
7. CO2 gas supply line indicator A and B*
Current supplying CO2 gas supply line
(connecting port for CO2 gas pipe) is displayed.
When the CO2 gas automatic changer function
changes the empty CO2 gas cylinder to the
other, the empty indicator is displayed in reverse
video and blinks.
8. CO2 gas supply line select key*
This key selects CO2 gas supply line A or B
(connecting port A or B for CO2 gas pipe). When
an optional gas auto-changer MCO-21GC is
installed, CO2 gas supply line A/B changes over
automatically when CO2 gas cylinder is empty.
* Only displayed when optional component MCO-21GC (gas auto-changer) is installed.
5
4
15
10 11 9 12
13
18 17 16 15
9. Present date/time display field
The date and time is set when the incubator is
shipped from the factory. Refer to page 46 for
details.
10. Over heat display
High-limit temperature alarm is activated:
“Over Heat” is displayed alternately in normal
characters and reverse video.
11. UV lamp condition display
UV lamp ON: “UV : On” is displayed.
UV lamp OFF: “UV : Off” is displayed.
Note: Displayed in models MCO-170AICUVH/
170AICUV or when the optional UV system set
MCO-170UVS is installed on the model
MCO-170AIC.
12. Outer door (opening/closing) display
Open: “Door : Open” is displayed alternately in
normal characters and reverse video.
Closed: “Door : Closed” is displayed.
Locked: “Door : Locked” is displayed.
13. Message display field
Alarms, errors, or messages are displayed
when a fault occurs. Refer to pages 65–67.
Note: When there are several errors, the display
shows the message number. For example, if
there are two errors, the display shows “1/2”.
14. Message select key
When there are several alarms/errors, the
message on the screen can be changed.
15. Menu key
Press this key to access the Menu screen. Refer
to page 30.
16. H2O2 key
This key initiates H2O2 decontamination. Refer
to pages 51–55. If the incubator is not equipped
for H2O2 decontamination, this key is not
displayed on the LCD touch panel.
17. Unlock key
Press this key to unlock the outer door when it is
auto-locked by the electric lock. Refer to page
58. If the incubator is equipped for auto-lock and
the auto-lock function is ON, this key is
displayed.
18. Buzzer key
Press this key to silence the buzzer. Refer to
pages 37 and 65–67.
14
English
16
INCUBATOR COMPONENTS
Remote alarm terminal
The alarm of this product can be transferred to a remote location by connecting the external alarm unit to
the remote alarm terminal. For the behaviour of remote alarm output, refer to pages 65–67.
The terminal of the remote alarm is installed at the
rear upper right of the unit (see the figure opposite).
The alarm outputs from this terminal. Contact
capacity is DC 30 V, 2 A.
Table 1 shows the behaviour of the remote alarm
when the Buzzer key is pressed.
Notes:
• The door alarm does not work remotely. Refer to
pages 65–66.
• Wiring of the remote alarm should be performed by
qualified service personnel.
Table 1. Behaviour of remote alarm when pressing Buzzer key
Remote Alarm setting
(Refer to pages 36–37) Connecting
terminal Normal
condition
Abnormal condition
(Including in the cases of power outage and
of where the power supply plug is pulled out)
When pressing the buzzer key
ON: Remote alarm setting not
connected with Buzzer key
(Remote alarm is cancelled)
COM.-N.C. Close Open Open (Maintain in abnormality)*
COM.-N.O. Open Close Close (Maintain in abnormality)*
OFF: Remote alarm setting
connected with Buzzer key
(Remote alarm is not
cancelled)
COM.-N.C. Close Open Close (Return to normal)
COM.-N.O. Open Close Open (Return to normal)
* In the case of Err01 (CO2 gas cylinder empty), Err11 and Err12 (CO2 sensor error), the condition returns
to normal.
Use a twisted shield wire for the connection.
Type: UL 2343, UL 2448, UL 2464, UL 2552, UL2623.
Length: 30 m max.
Remote alarm
terminal
17
INSTALLATION
Installation site
For correct operation, install the incubator in a location with the following conditions (referring also to the
precautions for safe operation on pages 5–9):
z
Install in an environment with normal air.
z
Do not expose to direct sunlight.
z
Do not install near significant heat sources, such as heaters, boilers, ovens, or autoclaves.
z
Ensure ambient temperature is at least 5
q
C lower than set temperature.
z
Do not allow the ambient temperature to become too high.
z
Install in a site with a relative humidity of 80 %R.H. or lower.
z
Do not expose to flammable or corrosive gas.
z
Do not install where objects may fall on the incubator from above.
English
18
INSTALLATION
Installation
1. Remove the packing tape and clean up.
Remove all the tapes that are securing the doors and the inner attachments. Open the doors for ventilation.
If the outer panels are dirty, use a cloth to wipe them with a diluted neutral detergent (Undiluted detergent
can damage the plastic components. For the dilution, refer to the instructions on the detergent).
Wipe off the residual detergent with a wet cloth and then wipe off any moisture.
Note: Remove the cable tie around the power supply cord to prevent corrosion of the cord coating.
2. Set up the humidifying pan and humidifying pan cover (Fig. 1).
3. Install 4 trays (Fig. 2).
(When optional small doors (MCO-170ID) are installed)
To use the gastight split doors effectively, set 3 trays at the same level as shown in Fig 3 below.
At the same level
as bottom side of
each gastight
split door
Gastight split door Inner door
Fig. 3
Humidifying pan Fig. 1
Humidifying pan cover
4 Trays
Fig. 2
19
4. Adjust the levelling feet.
Adjust the levelling feet by turning them
anticlockwise until the incubator is level (Fig. 4).
Note: Incubating on a sloping tray may adversely
affect the cultivation.
5. Earth the incubator.
Earth the incubator during installation to prevent electric shock. If there is no earth wire at the location,
consult with qualified service personnel.
z
When an earth must be installed
If an earthed 3-pole outlet is not available, then an earth must be installed. Consult with qualified service
personnel.
z
Installing an earth fault circuit breaker
If you must use the incubator in a moist or humid location, then it is recommended that an earth fault circuit
breaker be installed in the power supply circuit (i.e., the power supply at the incubator). Have the circuit
breaker installed by qualified service personnel.
Fig. 4
Leveling feet
Shrink
Stretch
English
20
INSTALLATION
z
Double-stacking
For stacking the incubators securely, refer to the
procedure included with the optional
double-stacking bracket MCO-170PS or the
stacking plate MCO-170SB.
Note: Two hooks are attached at the upper rear of
the incubator. When stacking incubators, fix the
upper incubator to the wall with these hooks and
wire or chain (Fig. 5).
Note: When stacking the incubators with our CO2
incubator or another O2/CO2 incubator, use the
stacking plate MCO-170SB. Refer to Table 11 on
page 71.
z
When the incubator is not in use
Empty the water from the humidifying pan and remove moisture from the chamber. Make sure that the
chamber is completely dry before closing the doors. Failure to do so may result in damage.
z
Before moving the incubator
Before moving the incubator, empty the water from the humidifying pan, disconnect the power supply plug
from the outlet, and make sure that the power supply cord will not be damaged. Failure to do so may result
in electric shock or fire.
z
If the key is fitted in the door lock
Remove the key from the door lock and keep it in a secure place. Make a note of the key symbol and
number in case the key is lost. To remove the key from the door lock, open the outer door and turn the key
in the direction of the lock. Remove the key and close the door.
Note: The outer door can only be locked with the unlock key while the outer door is open. Attempting to
turn the unlock key while the outer door is closed may damage the electric lock system.
Fig. 5
Hooks
Wire
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114
  • Page 115 115
  • Page 116 116
  • Page 117 117
  • Page 118 118
  • Page 119 119
  • Page 120 120
  • Page 121 121
  • Page 122 122
  • Page 123 123
  • Page 124 124
  • Page 125 125
  • Page 126 126
  • Page 127 127
  • Page 128 128
  • Page 129 129
  • Page 130 130
  • Page 131 131
  • Page 132 132
  • Page 133 133
  • Page 134 134
  • Page 135 135
  • Page 136 136
  • Page 137 137
  • Page 138 138
  • Page 139 139
  • Page 140 140
  • Page 141 141
  • Page 142 142
  • Page 143 143
  • Page 144 144
  • Page 145 145
  • Page 146 146
  • Page 147 147
  • Page 148 148
  • Page 149 149
  • Page 150 150
  • Page 151 151
  • Page 152 152
  • Page 153 153
  • Page 154 154
  • Page 155 155
  • Page 156 156
  • Page 157 157
  • Page 158 158
  • Page 159 159
  • Page 160 160
  • Page 161 161
  • Page 162 162
  • Page 163 163
  • Page 164 164
  • Page 165 165
  • Page 166 166
  • Page 167 167
  • Page 168 168
  • Page 169 169
  • Page 170 170
  • Page 171 171
  • Page 172 172
  • Page 173 173
  • Page 174 174
  • Page 175 175
  • Page 176 176
  • Page 177 177
  • Page 178 178
  • Page 179 179
  • Page 180 180
  • Page 181 181
  • Page 182 182
  • Page 183 183
  • Page 184 184
  • Page 185 185
  • Page 186 186
  • Page 187 187
  • Page 188 188
  • Page 189 189
  • Page 190 190
  • Page 191 191
  • Page 192 192
  • Page 193 193
  • Page 194 194
  • Page 195 195
  • Page 196 196
  • Page 197 197
  • Page 198 198
  • Page 199 199
  • Page 200 200
  • Page 201 201
  • Page 202 202
  • Page 203 203
  • Page 204 204
  • Page 205 205
  • Page 206 206
  • Page 207 207
  • Page 208 208
  • Page 209 209
  • Page 210 210
  • Page 211 211
  • Page 212 212
  • Page 213 213
  • Page 214 214
  • Page 215 215
  • Page 216 216
  • Page 217 217
  • Page 218 218
  • Page 219 219
  • Page 220 220
  • Page 221 221
  • Page 222 222
  • Page 223 223
  • Page 224 224
  • Page 225 225
  • Page 226 226
  • Page 227 227
  • Page 228 228
  • Page 229 229
  • Page 230 230
  • Page 231 231
  • Page 232 232
  • Page 233 233
  • Page 234 234
  • Page 235 235
  • Page 236 236
  • Page 237 237
  • Page 238 238
  • Page 239 239
  • Page 240 240
  • Page 241 241
  • Page 242 242
  • Page 243 243
  • Page 244 244
  • Page 245 245
  • Page 246 246
  • Page 247 247
  • Page 248 248
  • Page 249 249
  • Page 250 250
  • Page 251 251
  • Page 252 252
  • Page 253 253
  • Page 254 254
  • Page 255 255
  • Page 256 256
  • Page 257 257
  • Page 258 258
  • Page 259 259
  • Page 260 260
  • Page 261 261
  • Page 262 262
  • Page 263 263
  • Page 264 264
  • Page 265 265
  • Page 266 266
  • Page 267 267
  • Page 268 268
  • Page 269 269
  • Page 270 270
  • Page 271 271
  • Page 272 272
  • Page 273 273
  • Page 274 274
  • Page 275 275
  • Page 276 276
  • Page 277 277
  • Page 278 278
  • Page 279 279
  • Page 280 280
  • Page 281 281
  • Page 282 282
  • Page 283 283
  • Page 284 284
  • Page 285 285
  • Page 286 286
  • Page 287 287
  • Page 288 288
  • Page 289 289
  • Page 290 290
  • Page 291 291
  • Page 292 292
  • Page 293 293
  • Page 294 294
  • Page 295 295
  • Page 296 296
  • Page 297 297
  • Page 298 298
  • Page 299 299
  • Page 300 300
  • Page 301 301
  • Page 302 302
  • Page 303 303
  • Page 304 304
  • Page 305 305
  • Page 306 306
  • Page 307 307
  • Page 308 308
  • Page 309 309
  • Page 310 310
  • Page 311 311
  • Page 312 312
  • Page 313 313
  • Page 314 314
  • Page 315 315
  • Page 316 316
  • Page 317 317
  • Page 318 318
  • Page 319 319
  • Page 320 320
  • Page 321 321
  • Page 322 322
  • Page 323 323
  • Page 324 324
  • Page 325 325
  • Page 326 326
  • Page 327 327
  • Page 328 328
  • Page 329 329
  • Page 330 330
  • Page 331 331
  • Page 332 332
  • Page 333 333
  • Page 334 334
  • Page 335 335
  • Page 336 336
  • Page 337 337
  • Page 338 338
  • Page 339 339
  • Page 340 340
  • Page 341 341
  • Page 342 342
  • Page 343 343
  • Page 344 344
  • Page 345 345
  • Page 346 346
  • Page 347 347
  • Page 348 348
  • Page 349 349
  • Page 350 350
  • Page 351 351
  • Page 352 352
  • Page 353 353
  • Page 354 354
  • Page 355 355
  • Page 356 356
  • Page 357 357
  • Page 358 358
  • Page 359 359
  • Page 360 360
  • Page 361 361
  • Page 362 362
  • Page 363 363
  • Page 364 364
  • Page 365 365
  • Page 366 366
  • Page 367 367
  • Page 368 368
  • Page 369 369
  • Page 370 370

Phcbi 170AICUVH-PE Mode d'emploi

Taper
Mode d'emploi
Ce manuel convient également à