ResMed AirFit P10 for Her Mode d'emploi

Taper
Mode d'emploi

Ce manuel convient également à

User Guide
English • Fraais
Español • Português
Nasal Pillows System
618278 Frontcover.indd 1 8/10/2013 12:01:06 PM
ii iii
Fitting / Mise en place / Colocación / Colocação
1
2
3
Adjustment / Ajustement / Ajuste / Ajuste
618264r4 Book.indb 2 19/09/2013 8:33:42 AM
ii iii
Removal / Retrait / Remoción / Remoção
Disassembly / Démontage / Desmontaje /Desmontagem
1
2
618264r4 Book.indb 3 19/09/2013 8:33:56 AM
iv
Reassembly / Remontage / Montaje / Remontagem
1
2
3
618264r4 Book.indb 4 19/09/2013 8:34:08 AM
English
1
English
AirFit
P10
NASAL PILLOWS SYSTEM
Thank you for choosing the AirFit P10. This document provides the
user instructions for the AirFit P10 and AirFit P10 for Her masks
referred to collectively as AirFit P10 throughout this manual. AirFit
P10 has blue headgear while AirFit P10 for Her has pink headgear.
Using this guide
Please read the entire guide before use. When following instructions,
refer to the images at the front and back of the guide.
Intended Use
The AirFit P10 channels airflow noninvasively to a patient from a
positive airway pressure (PAP) device such as a continuous positive
airway pressure (CPAP) or bilevel device.
The AirFit P10 is:
to be used by patients (> 66 lb (30 kg)) for whom positive airway
pressure has been prescribed
intended for single-patient re-use in the home environment and
multipatient re-use in the hospital/institutional environment.
GENERAL WARNINGS
The mask must be used under qualified supervision for users
who are unable to remove the mask by themselves. The
mask may not be suitable for those predisposed to
aspiration.
The vent must be kept clear.
The mask should only be used with CPAP or bilevel devices
recommended by a physician or respiratory therapist.
The mask should not be used unless the device is turned on.
Once the mask is fitted, ensure the device is blowing air.
Explanation: CPAP and bilevel devices are intended to be
used with special masks (or connectors) which have vent
holes to allow continuous flow of air out of the mask. When
2
the device is turned on and functioning properly, new air
from the device flushes the exhaled air out through the mask
vent holes. However, when the device is not operating,
insufficient fresh air will be provided through the mask, and
the exhaled air may be rebreathed. Rebreathing of exhaled
air for longer than several minutes can, in some
circumstances, lead to suffocation. This applies to most
models of CPAP or bilevel devices.
Follow all precautions when using supplemental oxygen.
Oxygen flow must be turned off when the CPAP or bilevel
device is not operating, so that unused oxygen does not
accumulate within the device enclosure and create a risk of
fire.
Oxygen supports combustion. Oxygen must not be used
while smoking or in the presence of an open flame. Only use
oxygen in well ventilated rooms.
At a fixed rate of supplemental oxygen flow, the inhaled
oxygen concentration varies, depending on the pressure
settings, patient breathing pattern, mask, point of application
and leak rate. This warning applies to most types of CPAP or
bilevel devices.
The technical specifications of the mask are provided for
your clinician to check that they are compatible with the
CPAP or bilevel device. If used outside specification or if
used with incompatible devices, the seal and comfort of the
mask may not be effective, optimum therapy may not be
achieved, and leak, or variation in the rate of leak, may affect
the CPAP or bilevel device function.
Discontinue using this mask if you have ANY adverse
reaction to the use of the mask, and consult your physician
or sleep therapist.
Using a mask may cause tooth, gum or jaw soreness or
aggravate an existing dental condition. If symptoms occur,
consult your physician or dentist.
As with all masks, some rebreathing may occur at low CPAP
pressures.
English
3
Refer to your CPAP or bilevel device manual for details on
settings and operational information.
Remove all packaging before using the mask.
Using your mask
When using your mask with ResMed CPAP or bilevel devices that
have mask setting options, refer to the Technical specifications
section in this user guide for mask selection options.
For a full list of compatible devices for this mask, see the
Mask/Device Compatibility List on www.resmed.com on the
Products page under Service & Support. If you do not have internet
access, please contact your ResMed representative.
Notes:
This mask is not compatible for use with ResMed
AutoSet CS
2 and VPAP
Adapt SV devices.
SmartStart may not operate effectively when using this mask
with some CPAP or bilevel devices that have this feature.
If you experience nasal dryness or irritation, using a humidifier is
recommended.
Fitting
1. Hold the pillows away from your nose ensuring that the pillow
size (eg, M) with the left (L) and right (R) indicators are correctly
facing towards you then guide the nasal pillows into your nostrils
(L into your left nostril, R into your right nostril). Pull the headgear
over your head by holding the bottom strap and stretching it
around the back of your head. The top strap should sit
comfortably on top of your head.
Tip! There are raised logos on top of the frame so you can feel
the correct way up when fitting in the dark.
2. Move the mask away from your face then place it back to get a
comfortable fit. Ensure the nasal pillows fit securely into your
nostrils with both pillows positioned upright.
4
3. Finally, connect the end of the mask tubing into your device air
tubing. Your mask is now fitted and ready to use. When the
device is running, a light breeze should be felt through the vent.
To inspect the vent place a finger close but not over the entire
vent area.
Adjustment
To achieve a comfortable fit, adjust the split straps by spreading
them apart to loosen or draw them closer together to tighten.
Removal
To take off your mask system, pull the mask and headgear by the
split-strap, up and over your head.
Disassembly
1. Holding the top of the mask tubing, squeeze and pull the pillows
away from the mask frame.
2. Remove the headgear from the stabilizers by holding the fabric
end of each strap and pulling out to reveal the buttonhole, then
sliding over the node and off the end of each stabilizer. Your
mask is now disassembled.
Reassembly
1. With the gray side facing out, insert each end of the headgear
strap over the stabilizers through the buttonholes. Secure in
place by wrapping each strap end up and around the nodes.
Holding the split straps, pull the headgear from the stabilizer to
ensure it is tightly secure around the nodes. Ensure the headgear
strap is not twisted.
2. Attach the pillows to the mask frame by inserting the top and
bottom clips into the correct slots (shorter clip into top slot,
longer clip into bottom slot) on the mask frame. Your mask is
now reassembled.
3. Ensure the inner walls of the nasal pillows are not compressed
or collapsed. If they are, squeeze each nasal pillow so they return
to the correct shape.
English
5
Cleaning your mask at home
Your mask and headgear should only be gently handwashed. The
mask vent may discolour over time due to dust. Follow the cleaning
instructions in this guide to maintain performance and safety of the
mask.
WARNING
Always follow cleaning instructions and use mild soap. Some
cleaning products may damage the mask, its parts and their
function, or leave harmful residual vapours that could be inhaled
if not rinsed thoroughly.
CAUTION
If any visible deterioration of a system component is
apparent (cracking, crazing, tears etc), the component should
be discarded and replaced.
Avoid connecting flexible PVC products (eg, PVC tubing)
directly to any part of the mask. Flexible PVC contains
elements that can be detrimental to the materials of the
mask, and may cause the components to crack or break.
Daily/After each use:
Disassemble the pillows from the frame according to the
disassembly instructions.
Thoroughly clean the separated mask components (excluding
headgear), by gently rubbing in warm (approximately 86°F/30°C)
water using mild soap.
To optimize the mask seal, facial oils should be removed from
the pillows after use.
Use a soft bristle brush to clean the vent.
Inspect each component and, if required, repeat washing until
visually clean.
Rinse all components well with drinking quality water and allow
to air dry out of direct sunlight. Dry the vent by gently wiping
with a cloth.
6
When all components including the vent are dry, reassemble
according to the reassembly instructions.
Weekly:
Handwash the headgear in warm (approximately 86°F/30°C)
water using mild soap. It may be washed without being
disassembled. Rinse well and allow it to air dry out of direct
sunlight.
Reprocessing the mask between patients
Only Sleep Lab Mask (SLM) variants of the mask are intended for
multi-patient re-use. When using between patients, these masks
must be reprocessed according to cleaning and disinfection
instructions available on ResMed.com/downloads/masks.
Troubleshooting
Problem/possible cause Solution
Pillows won’t seal properly, mask is uncomfortable or causes red
marks
Pillows may have been fitted
incorrectly, adjusted incorrectly
or the headgear is fitted
incorrectly.
Check that the nasal pillows with the
left (L) and right (R) indicators are
inserted correctly (L into your left
nostril, R into your right nostril).
Carefully follow the fitting instructions.
Readjust the headgear to ensure it is
fitted comfortably.
Inner wall position of one or both
pillows is incorrect.
Squeeze pillow sides to release inner
wall. The correct position is shown in
the reassembly instructions.
Pillows are the wrong size. Talk to your clinician.
Pillows may be dirty. Clean pillows according to the
instructions.
Headgear is too loose. Wash headgear to help it return to its
original form. Follow the washing
instructions.
Adjust the headgear split straps.
English
7
Problem/possible cause Solution
Vent is blocked or
dirty/discolored.
If the vent requires cleaning, use a
soft bristle brush. The vent may
discolor over time and this is not a
safety concern. Follow the cleaning
instructions in this guide to maintain
performance and safety of the mask.
Vent is wet. Ensure the vent is dry after cleaning
and in the case of excess
condensation.
Mask too noisy
System is assembled incorrectly. Disassemble the mask, then
reassemble according to the
instructions.
System is leaking air. Refit or reposition your mask. Adjust
your headgear. Check your mask is
assembled correctly.
Technical specifications
Pressure-flow
curve
The mask contains passive venting to protect against
rebreathing. As a result of manufacturing variations, the
vent flow rate may vary.
Pressure
(cm H
2
O)
Flow
(L/min)
4 20
8 29
12 37
16 43
20 49
Dead space
information
Physical dead space is the empty volume of the
mask to the end of the swivel. Using the large
pillows it is 123 mL.
Therapy pressure
4 to 20 cm H
2
O
8
Resistance
Drop in pressure measured (nominal)
at 50 L/min: 0.4 cm H
2
O
at 100 L/min: 1.4 cm H
2
O
Environmental
conditions
Operating temperature: +41°F to 104°F (+5°C to
+40°C)
Operating humidity: 15% to 95% non-condensing
Storage and transport temperature: -4°F to 140°F
(-20°C to +60°C)
Storage and transport humidity: up to 95% relative
humidity non-condensing
Gross dimensions Large:
23.23" (H) x 5.87" (W) x 1.18" (D)
(590 mm (H) x 149 mm (W) x 30 mm (D))
Mask frame system (no headgear)
Mask setting options For S9: Select 'Pillows'
For other devices: Select 'SWIFT' (if available),
otherwise select 'MIRAGE' as the mask option.
Notes:
The mask system does not contain PVC, DEHP or phthalates.
This product is not made with natural rubber latex.
The manufacturer reserves the right to change these
specifications without notice.
Storage
Ensure that the mask is thoroughly clean and dry before storing it for
any length of time. Store the mask in a dry place out of direct
sunlight.
Disposal
This mask does not contain any hazardous substances and may be
disposed of with your normal household refuse.
English
9
Symbols
The following symbols may appear on your product or packaging:
Catalog number
Batch code
Humidity limitation
Temperature limitation
Fragile, handle with care
Not made with natural
rubber latex
Manufacturer
European Authorized
Representative
Keep away from rain
This way up
Indicates a Warning or Caution and alerts you to a possible
injury or explains special measures for the safe and effective
use of the device
Caution, consult accompanying documents
Prescription only (In the US, Federal law restricts these devices
to sale by or on the order of a physician)
Polypropylene recyclable
Limited warranty
ResMed Pty Ltd (hereafter ’ResMed’) warrants that y
our ResMed
mask system (including mask frame, cushion, headgear and tubing)
shall be free from defects in material and workmanship from the date
of purchase for the period of 90 days or in the case of disposable
masks the maximum use period. This warranty is only available to
the initial consumer. It is not transferable. If the product fails under
conditions of normal use, ResMed will repair or replace, at its option,
the defective product or any of its components. This limited warranty
does not cover: a) any damage caused as a result of improper use,
abuse, modification or alteration of the product; b) repairs carried out
by any service organization that has not been expressly authorized by
ResMed to perform such repairs; and c) any damage or
contamination due to cigarette, pipe, cigar or other smoke. Warranty
is void on product sold, or resold, outside the region of original
purchase.
Warranty claims on defective product must be made by the initial
consumer at the point of purchase.
This warranty replaces all other expressed or implied warranties,
including any implied warranty of merchantability or fitness for a
particular purpose. Some regions or states do not allow limitations on
how long an implied warranty lasts, so the above limitation may not
apply to you.
ResMed shall not be responsible for any incidental or consequential
damages claimed to have resulted from the sale, installation or use
of any ResMed product. Some regions or states do not allow the
exclusion or limitation of incidental or consequential damages, so the
above limitation may not apply to you. This warranty gives you
specific legal rights, and you may also have other rights which vary
from region to region. For further information on your warranty
rights, contact your local ResMed dealer or ResMed office.
10
Français
1
Français
AirFit
P10
MASQUE NARINAIRE
Merci d'avoir choisi l'AirFit P10. Ce document indique comment
utiliser les masques AirFit P10 et AirFit P10 for Her, désignés par
AirFit P10 dans le présent guide. Le harnais de l'AirFit P10 est bleu,
alors que celui de l'AirFit P10 for Her est rose.
Comment utiliser ce guide
Veuillez lire le guide dans son intégralité avant d'utiliser ce masque.
Pour vous aider à suivre les instructions, référez-vous aux
illustrations en première et dernière page de ce guide.
Usage prévu
L'AirFit P10 achemine au patient de façon non-invasive le débit d'air
produit par un appareil à pression positive tel qu'un appareil de PPC
ou d'aide inspiratoire avec PEP.
L'AirFit P10 est prévu pour :
une utilisation par des patients (poids > 66 lb (30 kg)) à qui un
traitement par pression positive a été prescrit;
pour un usage multiple par un seul patient à domicile ou pour un
usage multiple par plusieurs patients en milieu médical.
AVERTISSEMENTS D'ORDRE GÉNÉRAL
Le masque doit être utilisé sous le contrôle d'une personne
qualifiée si le patient n'est pas en mesure de l'enlever de lui-
même. Le masque peut ne pas convenir aux patients sujets
aux vomissements.
Les orifices de ventilation ne doivent jamais être obstrués.
Le masque doit être utilisé uniquement avec les appareils de
PPC ou d'aide inspiratoire avec PEP recommandés par un
médecin ou un kinésithérapeute respiratoire.
Le masque ne doit être porté que si l'appareil est sous
tension. Une fois le masque en place, assurez-vous que
l'appareil produit un débit d'air. Explication : Les appareils de
2
PPC et d'aide inspiratoire avec PEP sont conçus pour être
utilisés avec des masques (ou raccords) spéciaux possédant
des orifices de ventilation qui permettent l'évacuation
continue de l'air hors du masque. Lorsque l'appareil est sous
tension et qu'il fonctionne correctement, l'air neuf provenant
de l'appareil expulse l'air vicié par les orifices de ventilation
du masque. Cependant, lorsque l'appareil ne fonctionne pas,
le débit d'air propre délivré dans le masque est insuffisant et
une réinhalation de l'air expiré peut se produire. Si la
réinhalation d'air expiré dure plus de quelques minutes, elle
peut, dans certains cas, entraîner une suffocation. Cela vaut
pour la plupart des appareils de PPC et d'aide inspiratoire
avec PEP.
Prendre toutes les précautions applicables lors de
l'adjonction d'oxygène.
L'arrivée d'oxygène doit être fermée lorsque l'appareil de
PPC ou d'aide inspiratoire avec PEP n'est pas en marche afin
d'empêcher l'oxygène inutilisé de s'accumuler dans le boîtier
de l'appareil et de créer un risque d'incendie.
L'oxygène entretient la combustion. Veiller à ne pas fumer ni
approcher de flamme nue près de l'appareil lors de
l'utilisation d'oxygène. L'oxygène ne doit être utilisé que
dans un endroit bien aéré.
Lorsque l'adjonction d'oxygène est délivrée à un débit fixe, la
concentration de l'oxygène inhalé varie en fonction des
réglages de pression, de la respiration du patient, du
masque, du point d'arrivée de l'oxygène et du niveau de
fuite. Cet avertissement s'applique à la plupart des types
d'appareils de PPC ou d'aide inspiratoire avec PEP.
Les caractéristiques techniques de votre masque sont
fournies afin que votre prestataire de soins puisse vérifier
que celui-ci est compatible avec votre appareil de PPC ou
d'aide inspiratoire avec PEP. L'utilisation du masque en
dehors de ses caractéristiques ou avec des appareils
incompatibles peut compromettre son étanchéité et son
confort ainsi que l'efficacité du traitement, et les fuites ou
des variations du niveau de fuite peuvent affecter le
Français
3
fonctionnement de l'appareil de PPC ou d'aide inspiratoire
avec PEP.
En cas de réaction indésirable QUELCONQUE au masque,
cesser de l'utiliser et consulter un médecin ou un spécialiste
du sommeil.
L'utilisation d'un masque peut provoquer des douleurs dans
les dents, les gencives ou la mâchoire ou aggraver un
problème dentaire existant. Si ces symptômes apparaissent,
consulter un médecin ou un dentiste.
Comme c'est le cas avec tous les masques, une réinhalation
peut se produire à une pression PPC basse.
Consulter le guide d'utilisation de l'appareil de PPC ou d'aide
inspiratoire avec PEP pour davantage d'informations sur les
réglages et le fonctionnement.
Retirer l'intégralité de l'emballage avant d'utiliser le masque.
Utilisation du masque
Si vous utilisez le masque avec un appareil de PPC ou d'aide
inspiratoire avec PEP ResMed disposant d'une fonction de sélection
du masque, veuillez consulter la partie Caractéristiques techniques
de ce guide utilisateur pour davantage d'informations sur cette
fonction.
Pour une liste complète des appareils compatibles avec ce masque,
veuillez consulter la liste de compatibilité appareil/masque sur le site
www.resmed.com dans la page Produits sous la rubrique SAV et
assistance. Si vous n'avez pas accès à Internet, veuillez contacter
votre représentant ResMed.
Remarques :
Ce masque n'est pas compatible avec les appareils ResMed
AutoSet CS
2 et VPAP
Adapt SV.
La fonction SmartStart peut ne pas fonctionner correctement
lorsque ce masque est utilisé avec certains appareils de PPC ou
d'aide inspiratoire avec PEP dotés de cette fonctionnalité.
En cas de sécheresse ou d'irritation nasale, il est recommandé
d'utiliser un humidificateur.
4
Mise en place
1. Maintenez les coussins à une certaine distance de votre nez pour
vérifier que la taille des coussins (p. ex. M) et les marqueurs
gauche (L) et droite (R) sont bien tournés vers vous, puis insérez
les coussins narinaires dans vos narines (L dans votre narine
gauche, R dans votre narine droite). Faites passer le harnais par-
dessus votre tête en tenant la lanière inférieure et en la tirant sur
l'arrière de votre tête. La lanière supérieure doit être installée de
manière confortable sur le dessus de votre tête.
Astuce : Des logos en relief sont présents sur la partie
supérieure de l'entourage rigide pour vous permettre de mettre
en place le masque dans le bon sens même dans le noir.
2. Éloignez le masque de votre visage, puis remettez-le en place
pour l'ajuster confortablement. Assurez-vous que les coussins
narinaires sont bien en place dans vos narines, avec les deux
coussins en position verticale.
3. Pour finir, branchez l'extrémité du circuit du masque dans le
circuit respiratoire de votre appareil. Votre masque est
maintenant prêt à être utilisé. Pendant le fonctionnement de
l'appareil, un mince filet d'air devrait s'échapper des orifices de
ventilation. Pour contrôler les orifices de ventilation, placez un
doigt au-dessus, sans les recouvrir entièrement.
Ajustement
Pour un positionnement confortable, ajustez les lanières de la sangle
en les séparant pour un ajustement plus lâche ou en les rapprochant
pour un ajustement au contraire plus serré.
Retrait
Pour retirer votre masque, utilisez la sangle pour tirer le masque et le
harnais par-dessus votre tête.
Démontage
1. En tenant la partie supérieure du circuit du masque, pincez et
retirez les coussins de l'entourage rigide du masque.
2. Détachez le harnais des stabilisateurs en tenant l'extrémité en
tissu de chaque sangle et en tirant pour faire apparaître la
Français
5
boutonnière. Passez-les ensuite par-dessus le nœud à chaque
bout des stabilisateurs pour les retirer. Le démontage de votre
masque est à présent terminé.
Remontage
1. Avec la partie grise dirigée vers l'extérieur, insérez chacune des
extrémités des stabilisateurs dans les boutonnières de la sangle
du harnais, de l'extérieur vers l'intérieur. Pour sécuriser le
montage, faites passer chaque extrémité de la sangle autour des
nœuds. En tenant les lanières de la sangle, tirez le harnais au
niveau des stabilisateurs pour vérifier qu'il est suffisamment
serré autour des nœuds. Assurez-vous que la sangle du harnais
n'est pas vrillée.
2. Attachez les coussins à l'entourage rigide du masque en insérant
les clips en haut et en bas dans l'emplacement correspondant (le
clip le plus court en haut, le plus long en bas) sur l'entourage
rigide. Le remontage de votre masque est désormais terminé.
3. Vérifiez que les jupes internes du masque narinaire ne sont ni
comprimées, ni affaissées. Si c'est le cas, pincez chacun des
coussins narinaires pour les aider à reprendre leur forme.
Nettoyage de votre masque à domicile
Nettoyez votre masque et votre harnais exclusivement à la main et
en douceur. La poussière peut entraîner une décoloration des orifices
de ventilation avec le temps. Respectez les consignes de nettoyage
de ce guide pour ne pas altérer la performance et la sécurité du
masque.
AVERTISSEMENT
Systématiquement respecter les instructions de nettoyage et
utiliser un savon doux. Certains produits de nettoyage peuvent
endommager le masque, ses composants et leur
fonctionnement, ou laisser des vapeurs résiduelles nocives qui
peuvent être inhalées si les composants ne sont pas
correctement rincés.
6
ATTENTION
En cas de détérioration visible de l'un des composants du
système (p. ex. fissures, fendillements, déchirures, etc.), le
composant en question doit être jeté et remplacé.
Éviter de raccorder des produits en PVC souple (par ex. un
tube en PVC) directement aux composants de ce masque. Le
PVC souple contient des éléments qui peuvent altérer les
matériaux du masque et entraîner leur fissuration ou leur
rupture.
Quotidien/Après chaque utilisation :
Retirez les coussins de l'entourage rigide en suivant les
instructions de démontage.
Nettoyez soigneusement chaque composant du masque (sauf le
harnais) en frottant légèrement avec de l'eau tiède (environ
30 °C/86 °F) et un savon doux.
Pour optimiser l'étanchéité du masque, toute trace de sébum
doit être retirée des coussins après utilisation.
Utilisez une brosse à poils doux pour nettoyer les orifices de
ventilation.
Procédez à un contrôle visuel de chaque composant du masque
et, si nécessaire, répétez l'opération de nettoyage jusqu'à ce que
les composants vous semblent propres.
Rincez soigneusement tous les composants à l'eau potable et
laissez-les sécher à l'abri de la lumière directe du soleil. Pour
sécher l'orifice de ventilation, essuyez-le délicatement avec un
chiffon.
Lorsque tous les composants, y compris l'orifice, sont secs,
remontez-les en suivant les instructions de remontage.
Hebdomadaire :
Nettoyez le harnais à la main dans de l'eau tiède (à environ 86 °F,
soit 30 °C) à l'aide d'un savon doux. Il n'est pas nécessaire de le
démonter pour le nettoyer. Rincez soigneusement le harnais et
laissez-le sécher à l'abri de la lumière directe du soleil.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50

ResMed AirFit P10 for Her Mode d'emploi

Taper
Mode d'emploi
Ce manuel convient également à