ResMed AirFit P10 and AirFit P10 for Her Nasal Pillows Masks Manuel utilisateur

Taper
Manuel utilisateur
User guide
English | Français | Español | Português
User guide
ii
1
3
2
(a)
(b)
(c)
Fitting / Mise en place / Colocación / Colocação
Adjustment / Ajustement / Ajuste / Ajuste
iii
1
2 3
Removal / Retrait / Remoción / Remoção
Disassembly / Démontage / Desmontaje /
Desmontagem
iv
1
2
3
618264 Book.indb 6 21/08/2013 11:38:40 AM
Reassembly / Remontage / Montaje / Remontagem
English 1
ENGLISH
AirFit™ P10
Nasal Pillows System
Thank you for choosing the AirFit P10. This document provides the user
instructions for the AirFit P10 and AirFit P10 for Her masks referred to
collectively as AirFit P10 throughout this manual. AirFit P10 has blue
headgear while AirFit P10 for Her has pink headgear. Note your headgear
may come with or without a buckle.
Using this guide
Please read the entire guide before use. When following instructions,
refer to the images at the front and back of the guide.
Intended use
The AirFit P10 channels airflow noninvasively to a patient from a positive
airway pressure (PAP) device such as a continuous positive airway
pressure (CPAP) or bilevel device.
The AirFit P10 is:
to be used by patients (> 66 lb (30 kg)) for whom positive airway
pressure has been prescribed
intended for single-patient re-use in the home environment and multi-
patient re-use in the hospital/institutional environment.
GENERAL WARNINGS
The mask must be used under qualified supervision for users who
are unable to remove the mask by themselves. The mask may not
be suitable for those predisposed to aspiration.
The vent must be kept clear.
The mask should only be used with CPAP or bilevel devices
recommended by a physician or respiratory therapist.
The mask should not be used unless the device is turned on. Once
the mask is fitted, ensure the device is blowing air. Explanation:
CPAP and bilevel devices are intended to be used with special
masks (or connectors) which have vent holes to allow continuous
flow of air out of the mask. When the device is turned on and
functioning properly, new air from the device flushes the exhaled
2
air out through the mask vent holes. However, when the device is
not operating, insufficient fresh air will be provided through the
mask, and the exhaled air may be rebreathed. Rebreathing of
exhaled air for longer than several minutes can, in some
circumstances, lead to suffocation. This applies to most models of
CPAP or bilevel devices.
Follow all precautions when using supplemental oxygen.
Oxygen flow must be turned off when the CPAP or bilevel device
is not operating, so that unused oxygen does not accumulate
within the device enclosure and create a risk of fire.
Oxygen supports combustion. Oxygen must not be used while
smoking or in the presence of an open flame. Only use oxygen in
well ventilated rooms.
At a fixed rate of supplemental oxygen flow, the inhaled oxygen
concentration varies, depending on the pressure settings, patient
breathing pattern, mask, point of application and leak rate. This
warning applies to most types of CPAP or bilevel devices.The
technical specifications of the mask are provided for your clinician
to check that they are compatible with the CPAP or bilevel device.
If used outside specification or if used with incompatible devices,
the seal and comfort of the mask may not be effective, optimum
therapy may not be achieved, and leak, or variation in the rate of
leak, may affect the CPAP or bilevel device function.
Discontinue using this mask if you have ANY adverse reaction to
the use of the mask, and consult your physician or sleep therapist.
Using a mask may cause tooth, gum or jaw soreness or aggravate
an existing dental condition. If symptoms occur, consult your
physician or dentist.
As with all masks, some rebreathing may occur at low CPAP
pressures.
Refer to your CPAP or bilevel device manual for details on
settings and operational information.
The mask is not intended to be used simultaneously with
nebulizer medications that are in the air path of the mask/tube
Remove all packaging before using the mask.
Using your mask
When using your mask with ResMed CPAP or bilevel devices that have
mask setting options, refer to the Technical specifications section in this
user guide for mask selection options.
English 3
For a full list of compatible devices for this mask, see the Mask/Device
Compatibility List at ResMed.com/downloads/masks.
Notes:
This mask is not compatible for use with ResMed AutoSet CS™ 2 and
VPAP™ Adapt SV devices.
SmartStart may not operate effectively when using this mask with
some CPAP or bilevel devices that have this feature.
If you experience nasal dryness or irritation, using a humidifier is
recommended.
Fitting
1. Hold the pillows away from your nose ensuring that the pillow size
(eg, M) with the left (L) and right (R) indicators are correctly facing
towards you then guide the nasal pillows into your nostrils (L into your
left nostril, R into your right nostril). Pull the headgear over your head
by holding the bottom band and stretching it around the back of your
head. The top band should sit comfortably on top of your head.
Tip! There are raised logos on top of the frame so you can feel the
correct way up when fitting in the dark.
2. Move the mask away from your face then place it back to get a
comfortable fit. Ensure the nasal pillows fit securely into your nostrils
with both pillows positioned upright.
3. Finally, connect the end of the mask tubing into your device air tubing.
Your mask is now fitted and ready to use. When the device is running,
a light breeze should be felt through the vent. To inspect the vent
place a finger close but not over the entire vent area.
Adjustment
To achieve a comfortable fit, adjust the split bands by spreading them
apart to loosen or draw them closer together to tighten.
If your headgear comes with a buckle, adjust the headgear by pulling
the headgear through the buckle to fix mask leaks. Adjust only enough
for a comfortable seal and do not overtighten. You may need to adjust
the headgear as it stretches over time.
4
Removal
To take off your mask system, pull the mask and headgear by the split-
band, up and over your head.
Disassembly
1. Holding the top of the mask tubing, squeeze and pull the pillows away
from the mask frame.
2. Unhook the headgear from the stabilizer by pulling on the end cap to
reveal the buttonhole.
3. Slide the headgear off the end of the stabilizer.
Reassembly
1. With the gray side facing out, insert each end of the headgear band
over the stabilizers through the buttonholes. Secure in place by
wrapping each band end up and around the nodes. Holding the split
bands, pull the headgear from the stabilizer to ensure it is tightly
secure around the nodes. Ensure the headgear band is not twisted.
2. Attach the pillows to the mask frame by inserting the top and bottom
clips into the correct slots (shorter clip into top slot, longer clip into
bottom slot) on the mask frame. Your mask is now reassembled.
3. Ensure the inner walls of the nasal pillows are not compressed or
collapsed. If they are, squeeze each nasal pillow so they return to the
correct shape.
Cleaning your mask at home
Your mask and headgear should only be gently handwashed. The mask
vent may discolour over time due to dust. Follow the cleaning instructions
in this guide to maintain performance and safety of the mask.
WARNING
Always follow cleaning instructions and use mild soap. Some
cleaning products may damage the mask, its parts and their function,
or leave harmful residual vapours that could be inhaled if not rinsed
thoroughly.
English 5
CAUTION
If any visible deterioration of a system component is apparent
(cracking, crazing, tears etc), the component should be discarded
and replaced.
Avoid connecting flexible PVC products (eg, PVC tubing) directly
to any part of the mask. Flexible PVC contains elements that can
be detrimental to the materials of the mask, and may cause the
components to crack or break.
If your headgear comes with a buckle, do not overtighten the
headgear when fitting the mask as this may lead to skin redness
or sores around the mask cushion.
Do not iron the headgear as the material is heat sensitive and will
be damaged.
Daily/After each use:
Disassemble the pillows from the frame according to the disassembly
instructions.
Thoroughly clean the separated mask components (excluding
headgear), by gently rubbing in warm (approximately 86°F/30°C) water
using mild soap.
To optimize the mask seal, facial oils should be removed from the
pillows after use.
Use a soft bristle brush to clean the vent.
Inspect each component and, if required, repeat washing until visually
clean.
Rinse all components well with drinking quality water and allow to air
dry out of direct sunlight.
When all components including the vent are dry, reassemble
according to the reassembly instructions.
Weekly:
Soak the headgear in warm water with a mild liquid detergent.
Hand wash the headgear with a soft bristle brush.
Thoroughly rinse the headgear under running water.
Leave the headgear to air dry out of direct sunlight.
6
Reprocessing the mask between patients
Only Sleep Lab Mask (SLM) variants of the mask system are intended for
multi-patient re-use. When using between patients, these masks must be
reprocessed according to cleaning and disinfection instructions available
on ResMed.com/downloads/masks.
Troubleshooting
Problem/possible cause
Solution
Pillows won’t seal properly, mask is uncomfortable or causes red marks
Pillows may have been fitted
incorrectly, adjusted incorrectly
or the headgear is fitted
incorrectly.
Check that the nasal pillows with the left (L) and right
(R) indicators are inserted correctly (L into your left
nostril, R into your right nostril). Carefully follow the
fitting instructions. Readjust the headgear to ensure it
is fitted comfortably.
Inner wall position of one or
both pillows is incorrect.
Squeeze pillow sides to release inner wall. The correct
position is shown in the reassembly instructions.
Pillows are the wrong size. Talk to your clinician.
Pillows may be dirty. Clean pillows according to the instructions.
Headgear is too loose. Adjust the headgear split bands.
If your headgear does not come with a buckle, wash
the headgear to help it return to its original form.
Follow the washing instructions.
If your headgear comes with a buckle, adjust the
headgear by pulling the headgear through the buckle.
Adjust only enough for a comfortable seal and do not
overtighten. You may need to adjust the headgear as it
stretches over time.
Vent is blocked or
dirty/discolored.
If the vent requires cleaning, use a soft bristle brush.
The vent may discolor over time and this is not a safety
concern. Follow the cleaning instructions in this guide
to maintain performance and safety of the mask.
Vent is wet.
Ensure the vent is dry after cleaning and in the case of
excess condensation.
Mask too noisy
System is assembled incorrectly.
Disassemble the mask, then reassemble according to
the instructions.
System is leaking air.
Refit or reposition your mask. Adjust your headgear.
Check your mask is assembled correctly.
English 7
Technical specifications
Pressure-flow curve
The mask contains passive venting to protect against
rebreathing. As a result of manufacturing variations, the vent
flow rate may vary.
Flow
(L/min)
20
29
37
43
49
Dead space information
Physical dead space is the empty volume of the mask to the
end of the swivel. Using the large cushions it is 123 mL.
Therapy pressure
4 to 20 cm H
2
O
Resistance
Drop in pressure measured (nominal)
at 50 L/min: 0.4 cm H
2
O
at 100 L/min: 1.4 cm H
2
O
Sound
DECLARED DUAL-NUMBER NOISE EMISSION VALUES in
accordance with ISO 4871. The A-weighted sound power
level of the mask is 21 dBA, with uncertainty of 3 dBA. The
A-weighted sound pressure level of the mask at a distance
of 1 m is 13 dBA, with uncertainty of 3 dBA.
Environmental
conditions
Operating temperature: +41°F to 104°F (+5°C to +40°C)
Operating humidity: 15% to 95% non-condensing
Storage and transport temperature: -4°F to 140°F (-20°C to
+60°C)
Storage and transport humidity: up to 95% relative humidity
non-condensing
Gross dimensions
Large:
23.23" (H) x 5.87" (W) x 1.18" (D)
(590 mm (H) x 149 mm (W) x 30 mm (D))
Mask fully assembled with short tube assembly (no
headgear)
8
Mask setting options
For S9:
Select 'Pillows'
For other devices: Select 'SWIFT' (if available), otherwise
select 'MIRAGE' as the mask option.
Notes:
The mask system does not contain PVC, DEHP or phthalates.
This product is not made with natural rubber latex.
The manufacturer reserves the right to change these specifications without notice.
Storage
Ensure that the mask is thoroughly clean and dry before storing it for any length of time.
Store the mask in a dry place out of direct sunlight.
Disposal
This mask does not contain any hazardous substances and may be disposed of with your
normal household refuse.
Symbols
The following symbols may appear on your product or packaging:
Nasal Pillows mask
Polypropylene recyclable
Size - Extra small
Size - Small
Size - Medium
Size - Large
Caution, consult accompanying documents
Prescription only (In the US, Federal law restricts these devices to sale by or
on the order of a physician)
See symbols glossary at ResMed.com/symbols.
English 9
Limited warranty
ResMed Pty Ltd (hereafter ’ResMed’) warrants that your ResMed mask system (including
mask frame, cushion, headgear and tubing) will be free from defects in material and
workmanship from the date of purchase for 90 days, or in the case of disposable masks
and disposable mask components for 7 days. This warranty is only available to the initial
consumer. It is not transferable. During the warranty period, if the product fails under
conditions of normal use, ResMed will repair or replace, at its option, the defective
product or any of its components. This limited warranty does not cover: a) any damage
caused as a result of improper use, abuse, modification or alteration of the product; b)
repairs carried out by any service organization that has not been expressly authorized by
ResMed to perform such repairs; c) any damage or contamination due to cigarette, pipe,
cigar or other smoke; and d) any damage caused by exposure to ozone, activated oxygen or
other gases. Warranty is void on product sold, or resold, outside the region of original
purchase.
Warranty claims on defective product must be made by the initial consumer at the point of
purchase.
This warranty replaces all other expressed or implied warranties, including any implied
warranty of merchantability or fitness for a particular purpose. Some regions or states do
not allow limitations on how long an implied warranty lasts, so the above limitation may
not apply to you.
ResMed shall not be responsible for any incidental or consequential damages claimed to
have resulted from the sale, installation or use of any ResMed product. Some regions or
states do not allow the exclusion or limitation of incidental or consequential damages, so
the above limitation may not apply to you. This warranty gives you specific legal rights,
and you may also have other rights which vary from region to region. For further
information on your warranty rights, contact your local ResMed dealer or ResMed office.
Visit ResMed.com for the latest information on ResMed's Limited Warranty.
Français 1
FRANÇAIS
AirFit™ P10
Système De Coussins Narinaires
Merci d’avoir choisi l’AirFit P10. Ce document indique comment utiliser
les masques AirFit P10 et AirFit P10 for Her, collectivement désignés
AirFit P10 dans le présent manuel. Le harnais d’AirFit P10 est bleu, tandis
que celui d’AirFit P10 for Her est rose. Notez que votre harnais peut être
livré avec ou sans boucle.
Utilisation de ce guide
Veuillez lire le guide dans son intégralité avant d’utiliser ce masque. Les
instructions doivent être lues en consultant les illustrations au début et à
la fin de ce guide.
Usage prévu
L'AirFit P10 achemine au patient de façon non-invasive le débit d'air
produit par un appareil à pression positive tel qu'un appareil de PPC ou
d'aide inspiratoire avec PEP.
Le masque AirFit P10 est :
utilisé par des patients adultes de plus de 66 lb (30 kg) à qui une
pression positive a été prescrite;
prévu pour un usage multiple par un seul patient à domicile ou pour un
usage multiple par plusieurs patients en milieu hospitalier.
AVERTISSEMENTS D’ORDRE GÉNÉRAL
Le masque doit être utilisé sous le contrôle d’une personne
qualifiée si le patient n’est pas en mesure de l’enlever de lui-
même. Le masque peut ne pas convenir aux patients sujets aux
aspirations trachéo-bronchiques.
L'orifice de ventilation ne doit jamais être obstrué.
Le masque doit être utilisé uniquement avec les appareils de PPC
ou d’aide inspiratoire avec PEP recommandés par un médecin ou
un kinésithérapeute respiratoire.
Le masque ne doit être porté que si l’appareil est sous tension.
Une fois le masque en place, s’assurer que l’appareil produit un
débit d’air. Explication : les appareils de PPC et d’aide inspiratoire
2
avec PEP sont conçus pour une utilisation avec des masques (ou
raccords) spéciaux possédant des orifices de ventilation qui
permettent l’écoulement continu de l’air hors du masque. Lorsque
l’appareil est sous tension et qu’il fonctionne correctement, l’air
neuf provenant de l’appareil expulse l’air vicié par les orifices de
ventilation du masque. Cependant, lorsque l'appareil ne
fonctionne pas, l'air frais fourni par le masque est insuffisant et
l'air expiré peut être réinspiré. Le fait de respirer l'air expiré
pendant plus de plusieurs minutes peut, dans certaines
circonstances, entraîner une suffocation. Cela s’applique à la
plupart des modèles d’appareils de PPC ou d’aide inspiratoire
avec PEP.
Prendre toutes les précautions applicables lors de l’adjonction
d’oxygène.
L’arrivée d’oxygène doit être fermée lorsque l’appareil de PPC ou
d’aide inspiratoire avec PEP n’est pas en marche afin d’empêcher
l’oxygène inutilisé de s’accumuler dans le boîtier de l’appareil et
de créer un risque d’incendie.
L’oxygène est combustible. Veiller à ne pas fumer près de
l’appareil ni l'approcher de flamme nue lors de l’utilisation
d’oxygène. L’adjonction d’oxygène ne doit avoir lieu que dans des
salles bien aérées.
Lorsque l’adjonction d’oxygène est délivrée à un débit fixe, la
concentration de l’oxygène inhalé varie en fonction des réglages
de pression, de la respiration du patient, du masque, du point
d’arrivée de l’oxygène et du niveau de fuite. Cet avertissement
s’applique à la plupart des types d’appareils de PPC ou d’aide
inspiratoire avec PEP.Les caractéristiques techniques de votre
masque sont fournies afin que votre prestataire de soins puisse
vérifier que celui-ci est compatible avec votre appareil de PPC ou
d’aide inspiratoire avec PEP. L’utilisation du masque en dehors de
ses caractéristiques ou avec des appareils incompatibles peut
compromettre son étanchéité et son confort ainsi que l’efficacité
du traitement, et les fuites ou la variation du niveau de fuites
peuvent affecter le fonctionnement de l’appareil de PPC ou d’aide
inspiratoire avec PEP.
En cas de réaction indésirable QUELCONQUE au masque, cesser
d’utiliser le masque et consulter votre médecin traitant ou un
spécialiste du sommeil.
Français 3
L’utilisation d’un masque peut provoquer des douleurs dans les
dents, les gencives ou la mâchoire ou aggraver un problème
dentaire existant. Si de tels symptômes apparaissent, consulter
un médecin ou un dentiste.
Comme c’est le cas avec tous les masques, une réinhalation peut
se produire à une pression PPC basse.
Consulter le guide d’utilisation de l’appareil de PPC ou d’aide
inspiratoire avec PEP pour davantage d’informations sur les
réglages et le fonctionnement.
Ce masque n’est pas destiné à être utilisé conjointement à des
médicaments normalement administrés par nébuliseur qui se
trouvent dans tout composant du passage de l’air du masque ou
du tuyau
Retirer l’intégralité de l’emballage avant d’utiliser le masque.
Utilisation de votre masque
En cas d'utilisation du masque avec un appareil de PPC ou d'aide
inspiratoire avec PEP ResMed disposant d'une fonction de sélection du
masque, veuillez consulter la partie Caractéristiques de ce guide
utilisateur pour davantage d'informations relative à cette fonction.
Pour une liste complète des appareils compatibles avec ce masque,
veuillez consulter la liste de compatibilité appareil/masque sur le site
ResMed.com/downloads/masks.
Remarques :
Ce masque n'est pas compatible avec les appareils ResMed
AutoSet CS™ 2 et VPAP™ Adapt SV.
La fonction SmartStart peut ne pas fonctionner correctement lorsque
ce masque est utilisé avec certains appareils de PPC ou d'aide
inspiratoire avec PEP dotés de cette fonctionnalité.
En cas de sécheresse ou d'irritation nasale, il est recommandé
d'utiliser un humidificateur.
Mise en place
1. Maintenez les coussins à une certaine distance de votre nez pour
rifier que le marqueur de taille des coussins (p. ex. M) et les
marqueurs gauche (L) et droite (R) sont bien tournés vers vous, puis
insérez les coussins narinaires dans vos narines (L dans votre narine
gauche, R dans votre narine droite). Faites passer le harnais par-
4
dessus votre tête en tenant la bande inférieure et en la tirant sur
l'arrière de votre tête. La bande supérieure doit être installée de
manière confortable sur le haut du crâne.
Astuce! Des logos en relief sont présents sur la partie supérieure de
l'entourage rigide pour vous permettre de mettre en place le masque
dans le bon sens même dans le noir.
2. Éloignez le masque de votre visage, puis remettez-le en place pour
l'ajuster confortablement. Assurez-vous que les coussins narinaires
sont bien en place dans vos narines, avec les deux coussins en
position verticale.
3. Pour finir, raccordez l'extrémité du circuit du masque dans le circuit
respiratoire de votre appareil. Votre masque est maintenant prêt à être
utilisé. Pendant le fonctionnement de l'appareil, un mince filet d'air
devrait s'échapper des orifices de ventilation. Pour contrôler les
orifices de ventilation, placez un doigt au-dessus, sans les recouvrir
entièrement.
Ajustement
Pour un positionnement confortable, ajustez les lanières de la bande
en les séparant pour un ajustement plus lâche ou en les rapprochant
pour un ajustement plus serré.
Si votre harnais est muni d'une boucle, ajustez-le en le tirant vers la
boucle pour réparer les fuites du masque. Ne les serrez pas trop :
ajustez-les uniquement pour obtenir une étanchéité confortable. Vous
pourriez être amené à ajuster le harnais au fur et à mesure qu'il
s'allonge.
Retrait
Pour retirer votre masque, utilisez les lanières de la bande pour tirer le
masque et le harnais par-dessus votre tête.
Démontage
1. En tenant la partie supérieure du circuit du masque, pincez et retirez
les coussins de l'entourage rigide du masque.
2. Décrochez le harnais du stabilisateur en tirant sur l'embout pour faire
apparaître la boutonnière.
3. Faites glisser le harnais de l'extrémité du stabilisateur.
Français 5
Remontage
1. Avec la partie grise dirigée vers l'extérieur, insérez chacune des
extrémités des stabilisateurs dans les boutonnières de la bande du
harnais, de l'extérieur vers l'intérieur. Pour sécuriser le montage,
faites passer chaque extrémité de la bande autour des nœuds. En
tenant les lanières de la bande, tirez le harnais au niveau des
stabilisateurs pour vérifier qu'il est suffisamment serré autour des
nœuds. Assurez-vous que la bande du harnais n'est pas vrillée.
2. Attachez les coussins à l'entourage rigide du masque en insérant les
clips en haut et en bas dans l'emplacement correspondant (le clip le
plus court en haut, le plus long en bas) sur l'entourage rigide. Le
remontage de votre masque est désormais terminé.
3. Vérifiez que les jupes internes des coussins narinaires ne sont ni
comprimées ni affaissées. Si c'est le cas, pincez chacun des coussins
narinaires pour les aider à reprendre leur forme.
Nettoyage de votre masque à domicile
Nettoyez votre masque et votre harnais exclusivement à la main et en
douceur. La poussière peut entraîner une décoloration des orifices de
ventilation avec le temps. Respectez les consignes de nettoyage de ce
guide pour ne pas altérer la performance et la sécurité du masque.
AVERTISSEMENT
Veiller à toujours respecter les instructions de nettoyage et à utiliser
un savon doux. Certains produits de nettoyage peuvent
endommager le masque, ses composants et leur fonctionnement, ou
laisser des vapeurs résiduelles nocives qui peuvent être inhalées si
les composants ne sont pas correctement rincés.
PRÉCAUTIONS D’USAGE
En cas de détérioration visible de l’un des composants du
système (p. ex. fissures, fendillements, déchirures, etc.), le
composant en question doit être jeté et remplacé.
Éviter de raccorder des produits en PVC souple (par ex. un circuit
en PVC) directement aux composants de ce masque. Le PVC
souple contient des éléments qui peuvent altérer les matériaux du
masque et entraîner leur fissuration ou leur rupture.
6
Si votre harnais est muni d'une boucle, ne le serrez pas trop
pendant le port du masque, car cela peut mener à l’apparition de
rougeurs et d’irritations cutanées autour de la bulle du masque.
Ne pas repasser le harnais au risque d’endommager le matériau
avec lequel il est fabriqué, celui-ci étant sensible à la chaleur.
Tous les jours/Après chaque utilisation :
Retirez les coussins de l'entourage rigide en suivant les instructions
de démontage.
Nettoyez soigneusement chaque composant du masque (sauf le
harnais) en frottant légèrement avec de l'eau tiède (environ
30 °C/86 °F) et un savon doux.
Pour optimiser l'étanchéité du masque, toute trace de sébum doit être
retirée des coussins après utilisation.
Utilisez une brosse à poils doux pour nettoyer les orifices de
ventilation.
Procédez à une inspection visuelle de chaque composant du masque
et, si nécessaire, répétez l'opération de lavage jusqu'à ce qu'il soit
propre.
Rincez soigneusement tous les composants à l'eau potable et laissez-
les sécher à l'abri de la lumière directe du soleil.
Lorsque tous les composants, y compris l'orifice, sont secs, remontez
le masque en suivant les instructions de remontage.
Chaque semaine :
Faites tremper le harnais dans une eau tiède avec un détergent liquide
doux.
Nettoyez le harnais à la main à l’aide d’une brosse à soies souples.
Rincez soigneusement le harnais dans l’eau courante.
Laissez le harnais sécher à l'abri de la lumière directe du soleil.
Retraitement du masque entre les patients
Seules les versions SLM (Sleep Lab Mask) du masque sont destinées à
un usage multiple par plusieurs patients. Avant d’être utilisés par un autre
patient, ces masques doivent être retraités conformément aux
instructions de nettoyage et de désinfection disponibles sur
ResMed.com/downloads/masks.
Français 7
Dépannage
Problème/Cause possible
Solution
Les coussins ne sont pas étanches, le masque est inconfortable ou provoque des
marques rouges
Les coussins ont peut-être été mal mis
en place ou mal ajustés, ou le harnais
n'est pas correctement mis en place.
Vérifiez que les coussins narinaires, qui
présentent des marquages gauche (L) et droite
(R), sont insérés correctement (L dans votre
narine gauche, R dans votre narine droite).
Suivez les instructions de port du masque à la
lettre. Réajustez le harnais pour qu'il soit
correctement mis en place et confortable sans
être trop serré.
La jupe interne de l'un ou des deux
coussins est mal positionnée.
Pincez le coussin pour dégager la jupe interne.
La position adéquate de la jupe interne est
illustrée dans les instructions de remontage.
Les coussins ne sont pas de la bonne
taille.
Contactez votre clinicien.
Les coussins sont peut-être sales. Nettoyez les coussins en suivant les
instructions.
Le harnais est trop lâche. Ajustez les lanières de la bande du harnais.
Si votre harnais n'est pas muni d'une boucle,
lavez-le pour qu’il reprenne sa forme initiale.
Respectez les instructions de lavage.
Si votre harnais est muni d'une boucle, ajustez-
le en le tirant vers la boucle. Ne les serrez pas
trop : ajustez-les uniquement pour obtenir une
étanchéité confortable. Vous pourriez être
amené à ajuster le harnais au fur et à mesure
qu'il s'allonge.
L'orifice de ventilation est obstrué ou
sale/décoloré.
Si les orifices de ventilation sont sales, utilisez
une brosse à poils doux pour les nettoyer.
L'orifice de ventilation peut se décolorer avec
le temps ; cette décoloration est normale et
sans danger. Respectez les consignes de
nettoyage de ce guide pour ne pas altérer la
performance et la sécurité du masque.
L'orifice est mouillé. Vérifiez que l'orifice est correctement séc
après un nettoyage ou en cas de condensation
excessive.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46

ResMed AirFit P10 and AirFit P10 for Her Nasal Pillows Masks Manuel utilisateur

Taper
Manuel utilisateur