Thermo Fisher Scientific Solinst 122 Interface Meter Manuel utilisateur

Taper
Manuel utilisateur
Interface Meter Operating Instructions
High Quality Groundwater and Surface Water Monitoring Instrumentation
(Page 1 of 2)
Model 122
Field Measurements
IMPORTANT
1. To meet safety requirements and to protect the
electronics from damage, always ground the meter by
attaching the grounding clip to the metal well casing or
to a suitable ground rod.
Operating Principle
The Solinst Model 122 Interface Meter has a narrow
5/8" (16 mm) diameter probe and uses laser-marked PVDF
flat tape. It is certified to CSA Standards, for use in
hazardous locations Class 1, Div. 1, Groups C & D T3C,
and is ATEX certified under directive 94/9/EC, as II 3 G
Ex ic IIB T4 Gc. It has an infra-red circuit which detects the
presence of a liquid. A conductivity circuit differentiates
between conductive liquid (water) and non-conductive liquid
(LNAPL or DNAPL product).
Equipment Check
Before beginning any measurements, check the electronics
and battery condition by pushing the ‘START/OFF’ button.
Five quick beeps with the green light flashing, indicates that the
meter is functional. A flashing green light every second indicates
the meter is on. It will automatically turn off after 5 minutes to
preserve battery life.
Using the Tape Guide
1. The tape guide has been designed to: improve accuracy
when reading interface measurements; to prevent the
laser-marked PVDF flat tape being cut by well casing; and
to allow the tape and probe to hang straight from the side
of the well.
2. If interface measurements are being taken in a 2" dia well,
then simply fit the small end of the tape guide onto the
edge of the well casing.
3. Insert the leg of the Interface Meter into the hole on the
Tape Guide (small reels only).
4. Once inserted, rest the Interface Meter on the side of the
well casing.
A
v
e
r
t
i
s
s
e
m
e
n
t
:
A
f
i
n
d
e
p
r
e
v
e
n
i
r
l
'
i
n
f
l
a
m
m
a
t
i
o
n
d
'
a
t
m
o
s
p
h
e
r
e
s
d
a
n
g
e
r
e
u
s
e
s
,
n
e
c
h
a
n
g
e
r
l
e
s
B
a
t
t
e
r
i
e
s
q
u
e
d
a
n
s
d
e
s
e
m
p
l
a
ce
m
e
n
t
s
d
e
s
i
g
n
e
s
n
o
n
d
a
n
g
e
r
e
u
x
.
A
v
e
r
t
i
s
s
e
m
e
n
t
:
N
e
p
a
s
u
t
i
l
i
s
e
r
d
a
n
s
d
e
s
e
n
v
i
r
o
n
n
e
m
e
n
t
s
h
u
m
i
d
e
s
o
u
p
o
u
s
s
i
e
r
e
u
x
I
nterface
M
eter
Model 122
Replace only with the following
batteries: Duracell MN1604
or Energizer 522
INTRINSICALLY SAFE or Exia
WARNING: To prevent ignition of
a hazardous atmosphere, batteries
must only be changed in an area
known to be nonhazardous
WARNING: Do not use in wet or
dusty environments
DETECTION SIGNALS
Product: Steady
Water: Intermittent
ON
START/OFF
Class I, Groups CD T3C.
Ta = -40˚C to 50˚C
Made In Canada
Solinst Canada Ltd.
www.solinst.com
Tel: +1(905) 873-2255
(800) 661-2023
Tape Guide
Fit onto well
casing
Measurement
Point Insert
Interface
Meter leg
through hole
Tape Guide: prevents tape
from scratching the top edge
of well casing
Interface Meter:
anchored to the well casing
with the tape guide for
stable operation. Small reels
only. Medium and large
reels should be placed on
the ground.
High quality PVDF at tape:
hangs straight in the well for
accurate interface
measurements
Notes:
When using the
tape guide, the
measuring point
is offset from the
top of casing
To adjust your
measurements,
simply subtract
the amount
indicated on the
front of the tape
guide (eg. 6 cm or
2/10 ft.)
Continued overleaf...
2. Push the ‘START/OFF’ button. Five brief tones and
green light indicate that the meter is functional and the
flashing green light indicates that the meter is on. The
meter automatically turns off after 5 minutes. Press the
‘START/OFF’ button as necessary during operation to
turn the meter back on.
3. Place the slotted part of the tape guide onto the edge of the
well casing. Lay the Interface Meter laser-marked PVDF
flat tape into the groove on the top of the tape guide.
Measurements are read at the V-notch on the tape guide.
Note: When using the tape guide, remember to
subtract the compensation factor
stamped onto the side of the guide from
each measurement.
4. A steady tone and red light indicates a non-conductive
liquid (e.g. product). An intermittent tone and red light
indicates a conductive liquid (e.g. water).
5. For floating product (LNAPL), take the air/product
interface measurement on the way into the liquid, and
the water/product interface on the way up. When passing
through product into water, some product may adhere
to the probe sensors due to surface tension. Therefore,
when water is detected below product, the probe should
be raised and lowered rapidly in a short vertical motion to
remove any product that may have been carried down with
the probe.
®Solinst is a registered trademark of Solinst Canada Ltd.
START/OFF Button
Product/Water
Indicator Light
Battery
Drawer
Handle
Sonalert
ON Light
Made In Canada
Replace only with the following
batteries: Duracell MN1604
or Energizer 522
Solinst Canada Ltd.
www.solinst.com
Tel: +1(905) 873-2255
(800) 661-2023
INTRINSICALLY SAFE or Exia
WARNING: To prevent ignition of
a hazardous atmosphere, batteries
must only be changed in an area
known to be nonhazardous
WARNING: Do not use in wet or
dusty environments
I
nterface
M
eter
Model 122
ON
START/OFF
DETECTION SIGNALS
Product: Steady
Water: Intermittent
A
v
e
r
t
i
s
s
e
m
e
n
t
:
A
f
i
n
d
e
p
r
e
v
e
n
i
r
l
'
i
n
f
l
a
m
m
a
t
i
o
n
d
'
a
t
m
o
s
p
h
e
r
e
s
d
a
n
g
e
r
e
u
s
e
s
,
n
e
c
h
a
n
g
e
r
l
e
s
B
a
t
t
e
r
i
e
s
q
u
e
d
a
n
s
d
e
s
e
m
p
la
c
e
m
e
n
t
s
d
e
s
i
g
n
e
s
n
o
n
d
a
n
g
e
r
e
u
x
.
A
v
e
r
t
i
s
s
e
m
e
n
t
:
N
e
p
a
s
u
t
i
l
i
s
e
r
d
a
n
s
d
e
s
e
n
v
i
r
o
n
n
e
m
e
n
t
s
h
u
m
i
d
e
s
o
u
p
o
u
s
s
i
e
r
e
u
x
Class I, Div 1, Groups CD T3C.
Ta = -40˚C to 50˚C
ATEX Directive 94/9/EC
II 3 G Ex ic IIB T4 Gc
Printed in Canada
May 31, 2018
(#110931 Rev.f)
(Page 2 of 2)
Other suitable cleaning method:
Steam clean the PVDF flat tape only.
Battery Replacement
Push the battery drawer in and up and then release. The battery
drawer should eject slightly, allowing it to be pulled out. Replace
the 9V alkaline battery.
Other General Tips:
1. The probe should be cleaned after each use.
2. Always use the grounding cable.
3. Do not drop probe: damage to probe tip may result.
4. If battery is weak, the start tone will not sound, and flashing
“green” light will be off. Replace the 9V alkaline battery.
5. Where possible, use a Solinst tape guide to protect the tape
from scraping on well casing.
6. Before storage, make sure the meter is turned off.
If the Interface Meter is going to be stored for longer than
two months, the 9V alkaline battery should be removed to
avoid potential leakage.
7. The meter can be checked by placing the probe in distilled
(non-conductive) water or pure phase product, for example
lamp oil (avoid bright sunlight during testing and resting
the probe on the bottom of the container). A steady
tone and light should be observed.
8. To maintain Intrinsic Safety Certifications, do not splice the
tape.
6. The water/product interface should then be measured
as the probe is raised very slowly back up. Once the
interface is detected the probe can be raised and
lowered in small increments to precisely determine
the interface.
7. Repeat measurements to confirm water/product interface.
8. To determine the thickness of product, subtract the water/
product interface from the product/air interface.
9. To determine if there is any sinking product (DNAPL) in the
well, continue lowering the probe slowly. If steady signals
activate, determine the top of the sinking layer by reading
directly from the PVDF flat tape. Continue lowering the
probe slowly until the tape slackens when the well bottom
is reached. Read the level directly from the PVDF flat tape
and subtract one from the other to determine thickness.
10. Upon completion of readings clean the tape and probe; as
described in the Cleaning and Maintenance section.
Cleaning and Maintenance
After each use, the laser-marked PVDF tape should be wiped
clean and carefully rewound onto the reel.
The probe should be cleaned as follows:
Wash probe thoroughly with a non-abrasive mild detergent.
DO NOT USE ANY SOLVENTS. Use a soft cloth around
the pins on the end of the probe to remove all product. Use
the brush provided to remove all product from inner part
of the probe.
USE LUKE-WARM, NOT HOT WATER. DAMAGE TO
THE PROBE MAY RESULT.
Rinse probe thoroughly with distilled water, wipe dry.
Return the probe to the holder.
Note: In rare circumstances it is possible that the 122 might
sound when directed toward sunlight, and not in a
liquid. This is normal and does not affect proper
operation in a monitoring well.
Troubleshooting
When instrument is turned ‘ON’ there is a solid
red light (no tone)
1. Indicates a connection issue. Contact Solinst for further
troubleshooting options.
Instrument will not turn ‘ON’ (no starting tone)
1. Replace the battery
2. Check the polarity of the battery in the drawer: make sure
the + and - on the battery and the drawer match. The
probe may be harmed by a reversed battery.
3. ON/OFF button could be faulty. Contact Solinst.
When instrument is turned ‘ON’, it immediately
sounds product tone or intermittent water tone
1. Probe sensor may be dirty. Clean according to Cleaning
and Maintenance instructions.
2. Water may have leaked into the probe. Carefully, remove
the probe, keeping the wires connected. Dry out the
probe, wipe and inspect the o-ring, replace if necessary
and/or lubricate with silicone. To avoid any nicks, make
sure the wires are tucked back into the probe body when
replacing the probe. See Probe Replacement Instructions.
3. Tape may be damaged. Clean the tape and look for any
cuts or nicks. If necessary, replace the damaged tape. To
maintain the 122 Intrinsic Safety rating, do not splice or
repair a damaged tape. Contact Solinst for assistance.
4. The reel or probe circuitry could be damaged. Contact
Solinst.
Instrument does not detect liquid
1. Check battery. Replace if necessary.
2. Clean probe tip following the Cleaning and Maintenance
instructions.
3. Probe may be damaged. Please contact Solinst.
Instrument detects “Product” as “Water”
1. Note that this can happen if the probe is pulled into
product too quickly and therefore pulls water in with it.
Thoroughly dry the probe tip or shake the probe and try
again at a slower speed.
2. Product may have degraded or is now disturbed enough
to become an emulsion. If it has a detectable level of
conductivity, it will read water. Wait for it to settle and try
again.
Instrument does not detect water
1. Clean the probe tip. Follow the Cleaning and Maintenance
instructions.
2. The water could be pure and non-conductive or product
may be coating the probe, in which case, shake the probe
for a while in the water column to clean product from the
probe.
3. The probe circuitry could be damaged due to high
voltage (static) in the well. Always use a ground cable.
Please contact Solinst.
Authorised distributor
In Australia:
For customer service, call 1300-735-292
To email an order, ordersau@thermofisher.com
To order online: thermofisher.com
In New Zealand:
For customer service, call 0800-933-966
To email an order, ordersnz@thermofisher.com
To order online: thermofisher.com
  • Page 1 1
  • Page 2 2

Thermo Fisher Scientific Solinst 122 Interface Meter Manuel utilisateur

Taper
Manuel utilisateur