Page 1-2
10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110 • sales@tegam.com
Accuracy may become degraded in the presence of significant
levels of RF field. At a RF field strength of 3 V/m, accuracy is
1% of reading ± 2 counts for the four lowest ranges (1.999
mΩ full scale through 1.999 Ω full scale) and 2% of reading ±
2 counts on the 19.99 Ω full scale range.
Operating Temperature: -20 °C to +50 °C.
Relative humidity: 0 to 90%, non-condensing.
1.3 Description of Equipment
A rugged heavy-duty static-dissipative ABS/Polycarbonate
plastic case is supplied to contain and protect the R1L-E2A.
When closed, a gasket seals the lid to keep out water and dirt,
while the R1L-E2A is transported through rain or other
environmentally hostile conditions. When the lid is open, a
second gasket provides additional, although reduced
protection, for the front panel. The removable battery access
cover also has a gasket. The front panel is aluminum with
durable polycarbonate graphical overlay. The lid of the R1L-
E2A case further serves as a storage area for the supplied pair
of test probes and instruction manual.
Dimensions of the case are 17” W x 11.5” D x 7.1” H.
Weight is 18 pounds (including test probes, power cord,
battery cells, and manual).
With the lid open for operation, the front panel and its controls
are readily accessible. Controls and connectors may be
operated while wearing safety gloves.
Electrical: Power is from three alkaline D-cells. Current drain
depends on the measurement range selected, and battery
capacity. Based on an average drain of 150 mA, and brand
new D cells and a 25 °C ambient, the battery cells will typically
give up to 80 hours of operation (with the backlight OFF).
Increasing battery drain and decreasing temperature both
result in decreased battery capacity. Extended use on the 2
mΩ range or with backlight on will reduce operating time
dramatically. At low ambient in the 2 mΩ range, battery life