10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110
3-6
WARNING
Disconnect probes from external circuits and turn the
instrument off before removing the bottom cover. Reinstall
the cover before resuming use of the instrument.
AVERTISSEMENT
Débrancher les sondes de circuits externes et éteindre
l'appareil avant de retirer le couvercle inférieur. Remettre le
couvercle avant de reprendre l'utilisation de l'instrument.
Open the voltmeter case (see Battery Installation/Replacement
section), and cut the (2) strain relief cable ties that secure the test-
leads to the circuit board. Next, remove the circuit board from the
top case by unscrewing the hexagonal spacer in the center of the
board. Use a small soldering iron (< 75 W) to remove the test-leads
from the circuit board eyelets. Pull the old probe/test-lead set free of
the voltmeter case.
To install the new lead set, push the leads through the top case
holes, and solder into the appropriate eyelets (red-probe connects
to +eyelet, black-probe connects to –eyelet). To secure the test
leads, use the cable ties supplied with the replacement lead set.
Allow a lead length of 23-28 mm (0.9-1.1 inches) between the
circuit-board eyelets and the respective cable ties. Use a cable-tie
tool with a tension setting of 35-55 N (8-12 lb.) to install the cables
ties.
Reinstall the circuit board into the top case. Check that the fish
paper barrier at the switch opening is securely attached to the top
case. Before the bottom case is reattached, be sure that the switch
actuator cover is in place.
Confirm meter operation (refer to Performance Verification, page 3-
7).