Drive Handy-Ox Manuel utilisateur

Taper
Manuel utilisateur

Ce manuel convient également à

Handy-Ox
DIGITAL COLOR
PULSE OXIMETER
OXÍMETRO DE PULSO DIGITAL
OXYMÈTRE DE POULS DIGITAL
AVEC AFFICHAGE EN COULEURS
User Manual
Instrucciones para su uso
Guide d’utilisation
18720
www.drivemedical.com
REF.
2English
INSTRUCTIONS FOR USE
This manual is written in compliance with the council directive MDD93/42/EEC for medical
devices and harmonized standards. In case of modifications and software upgrades, the infor-
mation contained in this document is subject to change without notice.
The manual describes, in accordance with the Digital Handy-Ox Pulse Oximeter’s features and
requirements, main structure, functions, specifications, correct methods for transportation, in-
stallation, usage, operation, repair, maintenance and storage, as well as the safety procedures
to protect both the user and equipment. Refer to the respective chapters for details.
Please read the User Manual carefully before using this product. The User Manual describes
the operating procedures which should be followed strictly. Failure to follow the User Man-
ual may cause measurement inaccuracies, equipment damage and/or human injury. The
manufacturer is NOT responsible for the safety, reliability and performance issues and any
monitoring abnormality, human injury and equipment damage due to users’ negligence of the
operation instructions. The manufacturer’s warranty service does not cover such faults.
3
English
WARNING:
An uncomfortable or painful feeling may occur if using the device continuously, especially
for the microcirculation barrier patients. It is recommended that the sensor should not be
applied to the same finger for over 2 hours.
For special patients, there should be prudent inspection of the skin when placing the nger
probe onto the patient. This device should not be clipped onto an edema or onto tender
tissue.
The light (the infrared is invisible) emitted from the device is harmful to the eyes, so the user
and the technician should not stare at the light.
Patient should not use enamel or other makeup.
Patient’s ngernail should not be too long.
Please refer to the user instruction manual for clinical use restrictions and cautions.
This device is not intended for treatment.
Caution: Federal (US) Law restricts this device to sale by or on the order of a physician.
4English
CONTENT
1 Safety ................................................................................................................................. 6
1.1 Instructions for Safe Operations ............................................................................ 6
1.2 Warning .................................................................................................................. 6
1.3 Attention ................................................................................................................. 7
2 Overview ............................................................................................................................ 9
2.1 Features ................................................................................................................. 9
2.2 Environment Requirements ................................................................................. 10
3 Principle ........................................................................................................................... 10
4 Technical Specifications ................................................................................................. 11
4.1 Main Performance ................................................................................................ 11
4.2 Main Parameters .................................................................................................. 12
5 Installation ....................................................................................................................... 14
5.1 View of the Front Panel ........................................................................................ 14
5.2 Underside View and Left View ............................................................................. 14
5.3 Rear View ............................................................................................................. 15
5.4 Accessories .......................................................................................................... 15
6 Operating Guide .............................................................................................................. 16
6.1 Application Method .............................................................................................. 16
6.2 Attention for Operation ........................................................................................ 26
6.3 Clinical Restrictions.............................................................................................. 27
5
English
7 Maintain, Transportation and Storage ........................................................................... 28
7.1 Cleaning and Disinfecting ................................................................................... 28
7.2 Maintain ................................................................................................................ 28
7.3 Transportation and Storage ................................................................................. 28
8 Troubleshooting .............................................................................................................. 29
9 Key of Symbols ............................................................................................................... 30
10 Function Specification .................................................................................................... 31
Appendix 1 ............................................................................................................................ 33
Appendix 2 ............................................................................................................................ 37
6English
1 SAFETY
1.1 Instructions for Safe Operations
Check the main unit and all accessories periodically to make sure that there is no visible
damage that may affect patient’s safety and monitoring performance. It is recommended
that the device is inspected at least once a week. When there is obvious damage, stop
using the device.
Necessary maintenance must be performed by qualied service technicians ONLY. Users
are not permitted to maintain the pulse oximeter themselves.
The oximeter cannot be used together with accessories not specied in User Manual. Only
the accessory that is recommended by Drive Medical can be used with this device.
This product is calibrated before leaving factory.
1.2 Warning
Explosive hazard—DO NOT use the oximeter in the environment with ammable material
such as anesthetic.
DO NOT use the oximeter while the patient is being scanned by MRI or CT.
The person who is allergic to rubber cannot use this device.
The disposal the pulse oximeter and its accessories and packing (including battery, plastic
bags, foams and paper boxes) should follow the local laws and regulations.
7
English
Please check the packing before use to make sure the device and accessories are in accor-
dance with the packing list.
Please only use accessories and probes approved by Drive Medical, or else it may damage
the device.
The device can only be matched with the compatible probe.
Please don’t measure this device with functional tester for the device’s related information.
1.3 Attention
Keep the oximeter away from dust, vibration, corrosive substances, explosive materials,
high temperature and moisture.
If the oximeter gets wet, please stop operating it.
When it is carried from cold environment to warm or humid environment, please do not use
it immediately.
DO NOT operate keys on front panel with sharp materials.
High temperature or high pressure steam disinfection of the oximeter is not permitted. Refer
to User Manual in the relative chapter (7.1) for instructions of cleaning and disinfection.
Do submerge the oximeter in liquid. When it needs cleaning, please wipe its surface with
medical alcohol by soft material. Do not spray any liquid on the device directly.
When cleaning the device with water, the temperature should be lower than 140° F (60° C).
8English
Fingers which are too thin or too cold may affect the measurement accuracy. Please clip
the thicker nger, such as thumb and middle nger deeply into the probe.
The pulse oximeter can be used for adult and pediatric patients, based on the manufacturer
supplied finger probe.
The update period of data is less than 5 seconds, and is subject to change based on varia-
tions in individual pulse rates.
Please read the measurements on the waveform screen once the readings have stabilized.
This measure value is most accurate.
If some abnormal conditions appear on the screen during test process, pull out the patient
finger and reinsert to restore normal use. If continued abnormal conditions appear, contact
your equipment provided.
The device has a normal life expectancy of three years since the rst electried use.
This device has alarm functions. Please reference chapter 6.1 for details.
The device may not work for all patients. If you are unable to achieve stable readings, dis-
continue use.
9
English
2 OVERVIEW
The pulse oxygen saturation is the percentage of HbO2 in the total Hb in the blood, so-called
the O2 concentration in the blood. It is an important bio-parameter for the respiration. A number
of diseases relating to respiratory system may cause the decrease of SpO2 in the blood, fur-
thermore, some other causes such as the malfunction of human body’s self-adjustment, dam-
ages during surgery, and the injuries caused by some medical checkup would also lead to the
difficulty of oxygen supply in human body, and the corresponding symptoms would appear as
a consequence, such as vertigo, impotence, vomit etc. Serious symptoms might bring danger
to human’s life. Therefore, prompt information of patients’ SpO2 is of great help for the doctor to
discover the potential danger, and is of great importance in the clinical medical field.
The Handy-Ox Pulse Oximeter is a compact, efficient and convenient portable pulse oximeter. It is
only necessary for a patient to put one finger into probe for diagnosis, and the display screen will
directly show the measured value of SpO2 and pulse rate with a high level of accuracy.
2.1 Features
A. Operation of the product is simple and convenient.
B. The product is compact, light weight and portable.
C. Low power consumption.
The problem of overrating would emerge when the patient is suffering from toxicosis
which caused by carbon monoxide, the device is not recommended to be used
under this circumstance.
10 English
2.2 Environment Requirements
Storage Environment
a) Temperature -40° ~ 140° F (-40 ° ~60 ° C)
b) Relative humidity: 95%
c) Atmospheric pressure: 500hPa~1060hPa
Operating Environment
a) Temperature: 50 ° ~ 104 ° F (10 ° ~40 ° C)
b) Relative Humidity: 75%
c) Atmospheric pressure:700hPa~1060hPa
3. PRINCIPLE
Principle of the Oximeter is as follows: An experience formula of data process is established
taking use of Lambert Beer Law according to Spectrum Absorption Characteristics of Reduc-
tive Hemoglobin (Hb) and Oxyhemoglobin (HbO2) in glow & near-infrared zones. Operation
principle of the device is: Photoelectric Oxyhemoglobin Inspection Technology is adopted in
accordance with Capacity Pulse Scanning & Recording Technology, so that two beams of
different wavelength of lights can be focused onto human nail tip through perspective clamp
finger-type sensor. Then measured signal can be obtained by a photosensitive element, infor-
mation acquired through which will be shown on screen through treatment in electronic circuits
and microprocessor.
11
English
Figure 1
4. TECHNICAL SPECIFICATIONS
4.1 Main Performance
A. SpO2 value display
B. Pulse rate value display, bar graph display
C. Pulse waveform display
D. Low-voltage indication: low-voltage indicator will appear before operation and
accuracy are affected.
E. The display mode can be changed
F. Screen brightness can be changed
G. Pulse sound indication
12 English
H. Alarm function
I. With SpO2 value and pulse rate value record function, the stored data can be uploaded to
computer
J. It can be connected with an external oximeter probe
K. Real-time data can be transmitted to computer
4.2 Main Parameters
A. Measure of SpO2
Measure range: 0%~100%
Accuracy:
When the SpO2 measure range is 70%~100%,the permission of absolute error is ±2%;
below 70% unspecified
B. Measure of pulse rate
Measure range:30bpm~250bpm
Accuracy: ±2 bpm or ±2% (select larger)
13
English
C. Resolution
SpO2: 1%, Pulse rate: 1bpm.
D. Measure Performance in Weak Filling Condition
SpO2 and pulse rate can be shown correctly when pulse-filling ratio is 0.4%. SpO2 error is ±4%,
pulse rate error is ±2 bpm or ±2% (select larger).
E. Resistance to surrounding light
The deviation between the value measured in the condition of man-made light or indoor natural
light and that of darkroom is less than ±1%.
F. Power supply requirement: 3.6 V DC ~ 4.2V DC.
G. Optical Sensor
Red light (wavelength is 660nm, 6.65mW)
Infrared (wavelength is 880nm, 6.75mW)
H. Adjustable alarm range
SpO2: 0%~100%
Pulse Rate:0bpm~254bpm
14 English
5 INSTALLATION
5.1 View of the Front Panel
Figure 2
Front View
5.2 Underside View and Left View
Figure 3
Underside
and Left View
1. Probe jack: It is used to connect a ngertip probe to measure the oxygen saturation and
pulse rate.
2. USB port: It is used to connect a personal computer to export the trend data(or real-time
data) or charge the lithium battery.
15
English
5.4 Rear View
Figure 4
Rear View
(Refer to Figure 4 and insert the lithium battery properly in the right direction.)
5.4 Accessories
A. A lithium battery
B. A User Manual
C. A power adapter
D. USB Cord
E. A disk (PC software)
F. An adult-oximeter probe 18720-ADPROBE
An infant-oximeter probe (optional) 18720-PDPROBE
16 English
6 OPERATING GUIDE
6.1 Application Method
A. Measuring the data
a) Open the lid of battery box and put the battery inside, then close.
b) Put the ngertip probe into the jack on the underside of the oximeter. Only use Drive Medical
supplied accessories.
c) Put the nger into the probe.
d) Press and hold the “power on/off button” for a few seconds. The device will power on within a
few seconds and will start to self-check. The device displays the user interface after self-check.
e) Do not shake the nger and keep the patient in a calm state during measurement.
f) The data can be read directly from the screen in the user interface.
Fingernails and the luminescent tube should be at the same side.
If the alarm function is on, the device will provide a medium priority alarm signal when the
probe plug or finger is out. An intermittent alarm will occur and the user interface presents
“FINGER OUT”.
Medium priority indicates that prompt operator response is required.
The corner position of the plug should
be inserted into the right place of the
recessed portion of the socket, just as
shown in the picture.
17
English
Figure 5
Figure 6
18 English
B. Change display direction
On the user interface, press the”screen change button” to change the display direction.
C. Pause alarm
a) The Digital Handy-Ox will alarm if the PR or SpO2 exceed preset low and high parameters, if
the battery has a low charge, and if the patient’s finger is removed.
b) When alarm is on, press the ”alarm pause button” to pause the alarm. In order to reset the
alarm, press the “alarm pause button” again within 60s.
c) If you want to turn off the alarm for good, enter the menu.
D. Menu operation
In the user interface, press the ”menu button” to enter the menu options screen ((Figure 7).
Users can adjust the setting through the main menu, such as alarm, pulse sound indication, PR
and SpO2 low and high alarm settings, backlight, data storage, and ID function.
Figure 7
Main Menu Interface
19
English
a) Alarm setting
In the main menu interface, press the ”up button” or ”down button”, and move the menu choice
bar to “Alarm”, then press the “menu button” to enter the alarm setting menu of (Figure 8).
Figure 8
Alarm Setting Menu
a. The high/low alarm limit setting
Press the ”up button” or “down button” to choose the parameter to be adjusted, then press
the “menu button” again to enter the similar dialog box shown as gure 9, then press the “up
button” or “down button” to change data. Each press of the “up button” or “down button”, the
data will raise or descend in single increments. Continue until the desired setting is reached,
then press the “menu button” to save the setting.
If the alarm function is on, the device will provide medium priority alarm signal when
the data of SpO2 or pulse rate is beyond the limit. Intermittent alarm will occur and
the measure value becomes yellow.
20 English
Medium priority indicating that prompt operator response is required.
When setting up operation, the operator inserts the patient’s finger, and sets the SpO2
ALM HI lower than the current measure value, then the machine will start alarm. The
pulse rate alarm test is the same with the above.
Figure 9
b. The alarm state setting
Press the “up button” or “ down button”, and move the menu choice bar to “Alarm”, then
choose the alarm state “on/off” by pressing the “menu button”. Choose “on” to start the alarm
function, and choose “off” to disable it.
b) Pulse sound indication setting
In the main menu interface, press the “up button” or “down button”, and move the menu
choice bar to “Pulse Sound”, then choose the pulse sound indication state “on/offby pressing
the “menu button”. Choose “on” to start the function, and choose “offto disable it.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114

Drive Handy-Ox Manuel utilisateur

Taper
Manuel utilisateur
Ce manuel convient également à

dans d''autres langues