Ohaus ISLDMPHDGL Manuel utilisateur

Taper
Manuel utilisateur

Ce manuel convient également à

Instruction Manual
Incubating Shaker, Lid, ISLDMPHDGL
Incubating Shaker, Light, ISLD04HDG
Incubating Shaker, Light, ISLDMPHDG
EN - English ......... 1
FR - Français ......... 11
ES - Español ......... 23
IT - Italiano ......... 35
DE - Deutsch ......... 47
PT - Português ......... 59
NL - Nederlands ......... 71
NO - Norsk ......... 75
DA - Dansk ......... 79
SV - Svenska ......... 83
FI - Suomi ......... 87
HU - Magyar ......... 91
PL - Polski ......... 95
CZ - Czech ......... 99
KR - Korean ......... 103
JP - Japanese ......... 114
Revision 6
2/14/2020
Service information
If the troubleshooting section does not resolve or describe your problem, contact your
authorized OHAUS service agent. For service assistance or technical support in the
United States call toll-free 1-800-672-7722 ext. 7852 between 8:00 AM and 5:00 PM EST.
An OHAUS product service specialist will be available to provide assistance. Outside the
USA, please visit our web site, www.ohaus.com to locate the Ohaus office nearest you.
Serial Number: _______________________________________________________
Date of Purchase: ____________________________________________________
Supplier: ___________________________________________________________
table of contentS
Package Contents . . . . . . . . . . . . 1
Service Information . . . . . . . . . . . . 1
Installation . . . . . . . . . . . . 2
Maintenance & Servicing . . . . . . . . . . . . 2
Environmental Conditions . . . . . . . . . . . . 2
Equipment Disposal . . . . . . . . . . . . 2
Safety Instructions . . . . . . . . . . . . 3
Standards & Regulations . . . . . . . . . . . . 3
Control Panel . . . . . . . . . . . . 4
Incubating Light Duty Shaker Specifications . . . . . . . . . . . . 5
Incubating Light Duty Shaker Set-Up . . . . . . . . . . . . 5
Incubating Microplate Shaker Specifications . . . . . . . . . 6
Incubating Microplate Shaker Set-Up . . . . . . . . . 7
Operating Instructions . . . . . . . . . . . . 8-9
Troubleshooting . . . . . . . . . . . . 10
Package contentS
Incubating Light Duty Shaker or
Incubating Microplate Shaker
Power Cord
Instruction manual
1
EN
Upon receiving the Ohaus Incubating Light Duty/Microplate Shaker, check to ensure
that no damage has occurred during shipment. It is important that any damage that
occurred in transport is detected at the time of unpacking. If you do find such damage
the carrier must be notified immediately.
After unpacking, place the Incubating Light Duty/Microplate Shaker on a level bench
or table, away from explosive vapors. Secure to an immovable work surface by
pressing down on the four (4) corners of the unit, creating a strong suction to the work
surface (DO NOT place on a bench mat). Ensure that the surface on which the unit is
placed will withstand typical heat produced by the unit. Always place the unit on a
sturdy work surface.
The Incubating Light Duty/Microplate Shaker is supplied with a power cord that is
inserted into the IEC connector on the back of the unit first, then it can be plugged into
a properly grounded outlet. The 120V unit plugs into a 120 volt, 50/60 Hz source. The
230V unit plugs into a 230 volt, 50/60 Hz source.
maintenance & Servicing
The Incubating Light Duty/Microplate Shaker is built for long, trouble-free, dependable
service. No lubrication or other technical user maintenance is required. However at
least every three (3) months you should:
Unplug the unit.
Remove any accumulated dirt from the base and tray.
Check all accessible items to make sure they are properly tightened.
The unit should be given the care normally required for any electrical appliance.
Avoid wetting or unnecessary exposure to fumes. Spills should be removed promptly.
DO NOT use a cleaning agent or solvent on the front panel or lid which is abrasive
or harmful to plastics, nor one which is flammable. Always ensure the power is
disconnected from the unit prior to any cleaning. If the unit ever requires service,
contact your Ohaus representative.
Operating Conditions: Indoor use only.
Temperature: 5 to 40°C (41 to 104°F)
Humidity: maximum 80% relative humidity, non-condensing
Altitude: 0 to 6,562 ft (2000 M) above sea level
Non-Operating Storage:
Temperature: -20 to 65°C (-4 to 149°F)
Humidity: maximum 80% relative humidity, non-condensing
Installation Category II and Pollution Degree 2 in accordance with IEC 664.
equiPment DiSPoSal
This equipment must not be disposed of with unsorted waste. It is your responsibility
to correctly dispose of the equipment at life-cycle-end by handing it over
to an authorized facility for separate collection and recycling. It is also
your responsibility to decontaminate the equipment in case of biological,
chemical and/or radiological contamination, so as to protect the persons
involved in the disposal and recycling of the equipment from health
hazards.
For more information about where you can drop off your waste of equipment, please
contact your local dealer from whom you originally purchased this equipment. By doing
so, you will help to conserve natural and environmental resources and you will ensure that
your equipment is recycled in a manner that protects human health.
inStallation environmental conDitionS
2
EN
Please read the entire instruction manual before operating the Ohaus Incubating Light
Duty/Microplate Shaker.
WARNING! DO NOT use the Incubating Light Duty/Microplate Shaker in a
hazardous atmosphere or with hazardous materials for which the unit was
not designed. Also, the user should be aware that the protection provided
by the equipment may be impaired if used with accessories not provided or
recommended by the manufacturer, or used in a manner not specified by the
manufacturer.
Always operate unit on a level surface for best performance and maximum
safety.
DO NOT lift unit by the tray or lid.
CAUTION! To avoid electrical shock, completely cut off power to the unit by
disconnecting the power cord from the unit or unplug from the wall outlet.
Disconnect unit from the power supply prior to maintenance and servicing.
Spills should be removed promptly. DO NOT immerse the unit for cleaning.
DO NOT operate the unit if it shows signs of electrical or mechanical damage.
CAUTION! The caution hot indicator light warns that the temperature of the
top plate is above 40°C. The light will illuminate and remain lit when the
temperature of the top plate reaches approximately 40°C. When the heat is
turned off, the caution hot indicator light will stay lit until the temperature of the
top plate is less than 40°C.
Earth Ground - Protective Conductor Terminal
Alternating Current
Compliance to the following standards and regulations is indicated by the corresponding mark
on the product.
Mark Standards and Regulations
OHAUS Corporation declares that the ISLD series shakers comply with directives
2011/65/EU, 2014/30/EU, 2014/35/EU and standards EN 50581, EN 61010-1, EN
61010-2-010, EN 61010-2-051, EN 61326-1.
The full text of the EU declaration of conformity is available at the following internet
address: www.ohaus.com/ce.
This product complies with directive 2012/19/EU. Please dispose of this product in
accordance with local regulations at the collecting point specified for electrical and
electronic equipment.
For disposal instructions in Europe, refer to www.ohaus.com/weee.
EN 61326-1
CAN/CSA C22.2 61010-1, CAN/CSA C22.2 61010-2-010, CAN/CSA C22.2 61010-
2-051
UL 61010-1, UL 61010-2-010, UL 61010-2-051
Global Notice
Warning: This is a Class A product. In a domestic environment this product may cause radio interfer-
ence in which case the user may be required to take adequate measures.
Canada Notice
This Class A digital apparatus complies with Canadian ICES-003.
FCC Notice
NOTE: This equipment has been tested and found to comply with the limits for a Class A digital
device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable
protection against harmful interference when the equipment is operated in a commercial envi-
ronment. This equipment generates, uses, and can radiate radio frequency energy and, if not
installed and used in accordance with the instruction manual, may cause harmful interference to
radio communications. Operation of this equipment in a residential area is likely to cause harmful
interference in which case the user will be required to correct the interference at his own expense.
Changes or modifications not expressly approved by Ohaus Corporation could void the user’s authority
to operate the equipment.
Safety inStructionS StanDarDS & regulationS
3
EN
H.
L.I.
D.
A.
K.
C.
J.
E.
F.
G. B.
The front panel of the Incubating Light Duty/Microplate Shaker contains all the controls
and displays needed to operate the unit.
A. Standby button/standby indicator light: The standby indicator light will illuminate
when the unit is plugged in. The unit will be in standby mode. Press the standby button
to activate the temperature, speed and time functions. The standby indicator light
will shut off and the temperature, speed and time displays will illuminate. Press the
standby button again and the unit will once again be in standby mode.
B. Temperature display: Displays the actual/set-point temperatures in conjunction
with the actual/set-point indicator lights. C. Up/down arrows for set-point control.
On/off button starts/stops the heating function. D. The heat indicator light will be
illuminated when the unit is heating.
E. Actual indicator light: Illuminates when the temperature displayed is the actual
temperature of the air in the chamber.
F. Set-point indicator light: Illuminates when the set-point temperature is displayed.
G. Caution hot indicator light: Illuminates when the air temperature of the chamber
is above 40°C (104°F).
H. Speed display: Displays the speed of the shaker. I. Up/down arrows for set-point
control. On/off button starts/stops shaking function. J. The speed indicator light will
be illuminated when the unit is shaking.
K. Time display: Displays accumulated time (continuous mode) or how much time is
remaining (timed mode). The display range is from 0 to 9,999 minutes in one (1)
second increments. The display will indicate minutes and seconds until the timer
reaches 99 minutes and 59 seconds (99:59), then the display will automatically
display minutes up to 9,999. L. Up/down arrows for set-point control. On/off button
starts/stops the time function.
control Panel
4
EN
Overall dimensions (L x W x H): 17.0 x 11.0 x 10.7” (43.2 x 27.9 x 27.0cm)
Interior dimensions (L x W x H): 11.3 x 8.3 x 5.7” (28.7 x 21.1 x 14.5cm)
Tray dimensions (L x W): 11 x 7.75” (27.9 x 19.7cm)
Electrical (50/60 Hz): 120 volts, 5 amps, 450 watts
230 volts, 5 amps, 450 watts
Fuses: 5mm x 20mm, 5 amp quick acting
Temperature range: ambient +5°C to 65°C
Temperature uniformity: ±0.5°C at 37°C
Speed range: 100 to 1200rpm
Speed accuracy: ±2% of set speed up to 999rpm
±5% 1000 to 1200rpm
Timer: 1 second to 9999 minutes
(increased in 1 second increments)
Orbit: 0.125” (3mm)
Maximum weight capacity: ~ 8lbs (3.6kg), up to 1000rpm
~ 5lbs (2.3kg), over 1000rpm
Controls: see page 4
Tray material: aluminum
Ship weight: 30lbs (13.6kg)
incubating light Duty Shaker SPecificationS
The Incubating Light Duty Shaker is supplied with a tray designed to hold a variety
of accessories.
1. Flat containers can be shaken by placing them on the tray.
2. The tray also has mounting holes ready for use with the flask clamps or test tube
racks.
incubating light Duty Shaker Set-uP
5
EN
Overall dimensions (L x W x H): 17.0 x 11.0 x 7.75” (43.2 x 27.9 x 19.7cm)
Tray dimensions (L x W): 11 x 7.75” (27.9 x 19.7cm)
Electrical (50/60 Hz): 120 volts, 5 amps, 450 watts
230 volts, 5 amps, 450 watts
Fuses: 5mm x 20mm, 5 amp quick acting
Temperature range: ambient +5°C to 65°C
Temperature uniformity: ±0.5°C at 37°C
Speed range: 100 to 1200rpm
Speed accuracy: ±2% of set speed up to 999rpm
±5% 1000 to 1200rpm
Timer: 1 second to 9999 minutes
(increased in 1 second increments)
Orbit: 0.125” (3mm)
Capacity: 4 microplates or 2 micro-tube racks
Controls: see page 4
Tray material: aluminum
Ship weight: 20.5lbs (9.3kg)
incubating microPlate Shaker SPecificationS
6
EN
NOTE: The tray on the Incubating Microplate Shaker is not designed to hold flask clamps.
(4) microplates(2) microplates (2) Micro-Tube Racks
The Incubating Microplate Shaker is designed to hold two (2) or four (4) microplates, or two (2) Micro-Tube Racks.
1. Place two (2) microplates or deep well blocks diagonally on the tray, or place four (4) microplates or deep well blocks on the tray. The plates/blocks do not have to be filled.
2. Place the corner of the plate/block under the spring located at each corner of the tray.
3. Slide plate/block into place. You are ready to use.
incubating microPlate Shaker tray configurationS
2. 3. 4.
incubating microPlate Shaker Set-uP
7
EN
The Incubating Light Duty/Microplate Shakers have been designed for the temperature,
speed and time functions to work independently of one another. The temperature and
speed can be reset without resetting the timer and the timer can be stopped and
started without interrupting the heating and shaking functions.
1. Getting ready:
a. Plug the power cord into a properly ground outlet. The standby indicator light
will illuminate, verifying power to the unit.
b. Press the standby button to move the unit from standby mode. The standby
indicator light will turn off and the temperature, speed and time displays will
illuminate, displaying the previously used settings.
2. Setting temperature:
a. Press the up/down arrows below the temperature display until you reach the
desired temperature. When you release the button, the display will blink off and
then on indicating the new set temperature has been accepted.
b. Press the on/off button to start the heating function. The indicator light below the
temperature display will illuminate to indicate the heating function is in use
and remain lit until heating has ceased.
c. Temperature adjustments can be made without interrupting heating by using
the up/down arrows below the temperature display. After the change has been
made and you release the button, the display will blink off and then on indicating
the new set temperature has been accepted.
d. To stop the heating function, press the on/off button below the temperature
display. The heat indicator light will turn off.
CAUTION HOT indicator:
The caution hot indicator light warns that the temperature of the air in the chamber
is above 40°C (104°F). The light will illuminate and remain lit when the temperature
of the air in the chamber reaches approximately 40°C (104°F).
When the heat is turned off, the caution hot indicator light will
stay lit until the temperature of the air in the chamber is less than
40°C (104°F).
3. Setting speed:
a. Press the up/down arrows below the speed display until you reach the desired
speed. When you release the button, the display will blink off and then on
indicating the new set speed has been accepted.
b. Press the on/off button to start the shaking function. The indicator light below
the speed display will illuminate to indicate the shaking function is in use and
remain lit until shaking has ceased. The microprocessor controlled ramping
feature slowly increases speed until the set-point is reached which helps to
avoid splashing, and provides excellent low end control.
c. Speed adjustments can be made without interrupting shaking by using the up/
down arrows below the speed display. After the change has been made and
you release the button, the display will blink off and then on indicating the new
set speed has been accepted.
d. To stop the shaking function, press the on/off button below the speed display. The
speed indicator light will turn off.
4. Setting time to zero (0:00) and continuous mode: Accumulated time.
a. Press and hold the on/off button below the time display. After three (3) seconds
the display will indicate the previous set time.
b. Simultaneously press both the up and down arrows, the display will indicate
zero (0:00). The unit time is now set to zero (0:00) minutes. Alternately, you can
use the up/down arrows to get to zero (0:00).
c. Press the on/off button below the time display. The display will indicate the
accumulated time. The up/down arrows will become inactive. To stop timer,
press the on/off button again. IMPORTANT: This will NOT interrupt the
shaking function. Press the on/off button below the speed display to interrupt
the shaking function.
d. To reset, press and hold the on/off button below the time display. After three (3)
seconds the display will indicate the previous set time, which was zero (0:00).
incubating light Duty/microPlate Shaker oPerating inStructionS
8
EN
5. Setting timed mode: Programmed time.
a. Press the up/down arrows below the time display until you reach the desired
time.
b. Start this function by pressing the on/off button below the time display. The unit
will run for the selected time, the up/down arrows will become inactive while the
timer is running. The unit will stop shaking when the time display reaches zero
(0:00). Four (4) audible beeps will indicate the count down function is complete.
The time display will default back to the set time. To repeat for the same time,
simply press the on/off button again.
c. To interrupt an automatic timing cycle before it is completed, press the on/off
button below the time display. The display will flash off and on to indicate the
time function is on “hold”. IMPORTANT: This will NOT interrupt the shaking
function. Press the on/off button below the speed display to interrupt the shaking
function. Restart the timer by pressing the on/off button below the time display.
Unit will continue counting down to zero (0:00). When the display reaches zero
(0:00), you will hear the four (4) audible beeps that indicate the count down function
is complete and the shaking function will cease.
6. Turning unit off:
a. To turn the unit off, press the standby button. The temperature, speed and time
displays will be blank, the standby indicator light will illuminate. The Incubating
Light Duty/Microplate Shaker should be kept in standby mode when not in use.
To completely cut off power to the unit, disconnect the power cord from the unit
or unplug from the wall outlet.
OPERATING TIPS
As a safety feature, a built-in program will shut power off to the motor if the tray is
prevented from rotating, or the unit is overloaded beyond its recommended weight
capacity.
Built-in memory maintains the last used temperature, speed and time settings during
a power interruption.
SINGLE POINT CALIBRATION PROCEDURE
This procedure is used to fine tune and calibrate the Incubating Light Duty-Microplate
Shaker at a specific temperature setting. This process may be repeated for up to three
(3) separate set-points. If a fourth calibration set-point is entered, the first set-point
entered will be overwritten.
1. Turn unit on.
2. Set desired temperature.
3. Stabilize one (1) hour or more, measuring the temperature with a temperature
probe/thermometer.
4. Press and hold standby button, then press the temperature up arrow once. The
unit will beep two (2) times, confirming calibration mode. The display will now be
flashing.
5. Press the temperature up/down arrows until the display matches the temperature
probe/thermometer.
6. Press standby button to exit calibration mode and return to normal heating.
This process may be repeated at the same set-point, multiple times for fine tuning if
desired.
The unit will now use the biased offset for that specific temperature setting and
increase or decrease temperature accordingly to bring the temperature to set
temperature. The decimal point of the display will flash to indicate a biased offset is
being used. All other temperature settings will use the standard internal calibration.
This offset will be stored in memory and retained until reset.
To restore unit to factory setting:
Press and hold the standy button while pressing the temperature down arrow once.
The reset will be confirmed with two (2) audible beeps. Press the standby button to
exit calibration mode and return to normal heating.
During operation, any rattling or ticking sounds may indicate a loose screw on the tray,
a tray attachment or an accessory. All accessories should be sufficiently tightened in
place before starting the unit.
incubating light Duty/microPlate Shaker oPerating inStructionS
9
EN
troubleShooting
Problem Cause Solution
Unit fails to power on Missing or blown fuse Add or replace fuse as necessary. If problem persists, please contact your Ohaus Representative for repair.
Unit is excessively noisy
Rattling or ticking sounds may
indicate a loose screw on the tray
Ensure that tray is secured tightly. If problem persists, please contact your Ohaus Representative for repair.
E1
RTD open or over temperature
above 100ºC
This error cannot be fixed by end user. Please contact your Ohaus Representative for repair.
E2
RTD shorted or temperature
below 0ºC
This error cannot be fixed by end user. Please contact your Ohaus Representative for repair.
E3
Drive system failure
Loose suction cup foot
Ceased bearing
Drive belt broken
Remove the mechanical obstruction. In the event a foot (suction cup) has come loose from the bench top, firmly
press down on the four (4) corners of the unit creating a strong suction to the work surface (DO NOT place on
bench mat). If problem persists, please contact your Ohaus Representative for repair.
E4
Unit overload or
Loose suction cup foot
Ensure the load is evenly distributed and does not exceed the maximum load capacity for the unit. In the event
a foot (suction cup) has come loose from the bench top, firmly press down on the four (4) corners of the unit
creating a strong suction to the work surface (DO NOT place on bench mat). If problem persists, please contact
your Ohaus Representative for repair.
E5 Unit temperature is above 80˚C
Press the standby button to turn the unit off and allow it to cool. If the E5 error persists, switch the unit off and
contact your Ohaus representative for repair.
To clear error press the standby button
10
EN
Manuel d’instructions
Shaker de couve, Bouchon, ISLDMPHDGL
Shaker de couve, Léger, ISLD04HDG
Shaker de couve, Léger, ISLDMPHDG
EN - English ......... 1
FR - Français ......... 11
ES - Español ......... 23
IT - Italiano ......... 35
DE - Deutsch ......... 47
PT - Português ......... 59
NL - Nederlands ......... 71
NO - Norsk ......... 75
DA - Dansk ......... 79
SV - Svenska ......... 83
FI - Suomi ......... 87
HU - Magyar ......... 91
PL - Polski ......... 95
CZ - Czech ......... 99
KR - Korean ......... 103
JP - Japanese ......... 114
contenu De l’emballage
Agitateur léger d’incubation ou
Agitateur de micro plaque d’incubation
Cordon d’alimentation détachable
Manuel d’instruction
table DeS matièreS
Contenu de l’emballage . . . . . . . . . . . . 12
Des informations de service . . . . . . . . . . . . 12
Installation . . . . . . . . . . . . 13
Maintenance et réparation . . . . . . . . . . . . 13
Conditions ambiantes . . . . . . . . . . . . 13
Élimination du matériel . . . . . . . . . . . . 13
Consignes de sécurité . . . . . . . . . . . . 14
Normes et réglementations . . . . . . . . . . . . 14
Panneau de commande . . . . . . . . . . . . 15
Spécifications de l’agitateur léger d’incubation . . . . . . . . . . . . 16
Réglage de l’agitateur léger d’incubation . . . . . . . . . . . . 16
Spécifications de l’agitateur de micro plaque d’incubation . . . . . . . . . 17
Spécifications de l’agitateur de micro plaque d’incubation . . . . . . . . . . 18
Mode d’emploi . . . . . . . . . . . . 19-21
Dépannage . . . . . . . . . . . . 22
FR
12
DeS informationS De Service
Si la section de dépannage ne permet pas de résoudre ou ne décrit pas le problème,
contacter l’agent d’entretien agréé OHAUS. Pour un service d’assistance ou une prise en
charge technique aux États-Unis, composer le numéro gratuit 1-800-672-7722, poste 7852
entre 8h00 et 17h00 (GMT - 5). Un spécialiste du service produit OHAUS est disponible
pour apporter son aide. En dehors des États-Unis, bien vouloir se rendre sur notre site web,
www.ohaus.com pour rechercher la filiale OHAUS la plus proche.
Numéro de série: ____________________________________________________
Date d’achat: _______________________________________________________
Fournisseur: ________________________________________________________
inStallation
Dès réception de l’agitateur incubateur droit léger/microplaques Ohaus, vérifiez
l’absence de dommages survenus pendant le transport. Il est important que tout dom-
mage résultant du transport soit détecté lors du déballage. Informez immédiatement
le transporteur en cas de dommage.
Après le déballage, placez votre agitateur incubateur droit léger/microplaques sur
un plan de travail ou une table de niveau, à l’écart de toute vapeur explosive. Fixez
l’appareil à un plan de travail stable en appuyant sur ses quatre (4) coins afin de créer
une forte adhérence au plan de travail (NE l’installez PAS sur une table antistatique).
Assurez-vous que la surface sur laquelle le système est placé résiste à la chaleur
généralement produite par le système. Placez toujours l’appareil sur un plan de travail
solide.
L’agitateur incubateur droit léger/microplaques est fourni avec un cordon d’alimentation
qu’il vous faut tout d’abord insérer dans le connecteur IEC situé au dos de l’appareil
avant de le brancher dans une prise correctement mise à la terre. L’appareil de 120 V
se branche dans une source d’alimentation de 120 volts, 50/60 Hz. L’appareil de 230
V se branche dans une source d’alimentation de 230 volts, 50/60 Hz.
maintenance et réParation
Votre agitateur incubateur droit léger/microplaques est construit pour fonctionner long-
temps, de façon efficace et sans problème. Aucun graissage ou tout autre entretien n’est
requis de la part de l’utilisateur. Toutefois, vous devez, au moins tous les trois (3) mois:
Débrancher l’appareil.
Nettoyer la saleté accumulée sur la base et le plateau.
Vérifier tous les éléments possibles afin de vous assurer qu’ils sont
bien serrés.
Votre appareil doit être traité avec le même soin que n’importe quel autre appareil
électrique. Évitez de le mouiller ou de l’exposer inutilement à des émanations. Essuyez
promptement tout liquide renversé. Pour nettoyer le panneau avant, N’utilisez PAS de
nettoyant ou de solvant abrasif ou pouvant endommager le plastique, ni de produit
inflammable. Assurez-vous toujours que l’alimentation est débranchée avant d’effectuer
tout nettoyage. Si l’appareil doit être réparé, contactez votre représentant Ohaus.
conDitionS ambianteS
Conditions de fonctionnement: Pour une utilisation à l’intérieur uniquement.
Températures: 5 à 40°C (41 à 104°F)
Humidité: 80 % HR maximum, sans condensation
Altitude: jusqu’à 6,562 ft (2000 M) au-dessus du niveau de la mer
Stockage:
Températures: -20 à 65 °C (-4 à 149°F)
Humidité: 80 % HR maximum, sans condensation
Installation de Catégorie II et palier de pollution 2 suivant la norme CEI 664.
élimination Du matériel
Ce dispositif ne doit pas être éliminé parmi les déchets non triés. Il est votre respon-
sabilité d’éliminer de façon appropriée le dispositif à la fin de son
cycle de vie en le confiant à une installation autorisée spécialisée
dans la collecte de déchets tries et le recyclage. Il est aussi votre
responsabilité de décontaminer tout dispositif en cas de contamina-
tion biologique, chimique ou radiologique, de façon à protéger les
personnes impliquées dans l’élimination et le recyclage des risques
sanitaires.
Pour obtenir plus d’informations concernant les endroits où vos dis-
positifs peuvent être déposés pour l’élimination, contactez le concessionnaire auprès
duquel vous avez acheté ce dispositif. En faisant cela, vous contribuerez à conserver
des ressources naturelles et environnementales et vous garantirez que votre dispositif
est recyclé d’une manière qui préserve la santé humaine.
FR
13
conSigneS De Sécurité
Veuillez lire le manuel d’instruction dans sa totalité avant d’utiliser l’agitateur incuba-
teur droit léger/microplaques.
AVERTISSEMENT! NE vous servez PAS de l’agitateur incubateur Droit
léger/ microplaques dans un environnement dangereux ou avec des maté-
riaux dangereux pour lesquels cet appareil n’a pas été conçu. D’autre part,
sachez que la protection offerte par l’appareil devient obsolète si celui-ci est
utilisé avec des accessoires non fournis ou recommandés par le fabricant ou
s’il est utilisé de façon non appropriée.
Utilisez toujours l’appareil sur une surface de niveau pour assurer une perfor-
mance optimale et une sécurité maximale.
NE soulevez PAS l’appareil par le plateau ou le couvercle.
AVERTISSEMENT Pour éviter tout risque d’électrocution, coupez complète-
ment l’alimentation de l’appareil, débranchez le cordon d’alimentation de
l’appareil ou de la prise murale. Débranchez l’alimentation avant toute procé-
dure d’entretien et de dépannage.
Essuyez promptement tout liquide renversé. N’immergez PAS l’appareil pour
le nettoyer.
NE faites PAS fonctionner l’appareil s’il semble avoir subi des dommages
électriques ou mécaniques.
AVERTISSEMENT Le témoin Caution Hot signale que la température ambi-
ante de la chambre est supérieure à 40°C. Le témoin s’allume et reste allumé
lorsque la température ambiante de la chambre atteint environ 40°C. Une
fois le chauffage éteint, le témoin Caution Hot reste allumé jusqu’à ce que la
température ambiante de la chambre soit inférieure à 40°C.
Terre - Borne du conducteur de protection
Courant alternatif
FR
La conformité aux normes et réglementations suivantes est indiquée par la marque correspon-
dante sur le produit.
Marque Normes et réglementations
OHAUS Corporation déclare que le Série ISLD agitateurs est conforme aux directives
2011/65/UE, 2014/30/UE, 2014/35/UE et aux normes EN 50581, EN 61010-1, EN
61010-2-010, EN 61010-2-051, EN 61326-1.
Le texte complet de la déclaration UE de conformité est disponible à l’adresse internet
suivante: www.ohaus.com/ce.
Ce produit est conforme à la directive 2012/19/UE. Veuillez jeter ce produit conformé-
ment à la réglementation locale au point de collecte spécifié pour les équipements
électriques et électroniques.
Pour obtenir des instructions d’élimination en Europe, consultez www.ohaus.com/
weee.
EN 61326-1
CAN/CSA C22.2 61010-1, CAN/CSA C22.2 61010-2-010, CAN/CSA C22.2 61010-
2-051
UL 61010-1, UL 61010-2-010, UL 61010-2-051
Avis Global
Avertissement: Il s’agit d’un produit de classe A. Dans un environnement domestique, ce produit
peut causer des interférences radio, auquel cas l’utilisateur peut être amené à prendre des mesures
adéquates.
Canada Avis
Cet appareil numérique de classe A est conforme à la norme ICES-003 du Canada.
Avis de la FCC
REMARQUE: Cet équipement a été testé et déclaré conforme aux limites d’un appareil numérique de
classe A, conformément à la partie 15 des règles de la FCC. Ces limites sont conçues pour fournir
une protection raisonnable contre les interférences nuisibles lorsque l’équipement est utilisé dans un
environnement commercial. Cet équipement génère, utilise et peut émettre de l’énergie radiofréquence
et, s’il n’est pas installé et utilisé conformément au manuel d’instructions, peut causer des interférences
nuisibles aux communications radio. Le fonctionnement de cet équipement dans une zone résiden-
tielle est susceptible de causer des interférences nuisibles, auquel cas l’utilisateur devra corriger
l’interférence à ses propres frais.
Les modifications ou modifications non expressément approuvées par Ohaus Corporation peuvent
annuler l’autorisation de l’utilisateur d’utiliser l’équipement.
normeS et réglementationS
14
Le panneau avant de l’agitateur incubateur droit léger/microplaques réunit toutes les
commandes et tous les affichages nécessaires au fonctionnement de l’appareil.
A. Touche/témoin veille: Le témoin veille s’allume lorsque l’appareil est branché.
L’appareil passe en mode veille. Appuyez sur la touche veille pour activer les fonc-
tions de température, de vitesse et de minuterie. Le témoin veille s’éteint et les
affichages de la durée, de la température et de la vitesse s’allument. Appuyez sur
la touche veille à nouveau et l’appareil repasse en mode veille.
B. Affichage Temperature (température): Indique la température réelle/de consigne
en association avec les témoins actual/setpoint (réelle/consigne). C. Flèches haut/
bas pour le contrôle des valeurs de consigne. La touche marche/arrêt active et dés-
active la fonction de chauffage. D. Le témoin de vitesse s’allume lorsque l’appareil
est en train de chauffer.
E. Témoin actual (valeur réelle): S’allume lorsque la température affichée est la
température ambiante réelle de la chambre.
F. Témoin point de réglage: S’allume lorsque la température de réglage s’affiche.
G. Témoin Caution Hot (Attention plaque chaude): S’allume lorsque la température
ambiante de la chambre est supérieure à 40°C.
H. Affichage Speed (vitesse): Indique la vitesse de l’agitateur. I. Flèches haut/bas
pour le contrôle des valeurs de consigne. La touche marche/arrêt active et désac-
tive la fonction d’agitation. J. Le témoin de vitesse s’allume lorsque l’appareil est
en train d’agiter.
K. Affichage Time (durée): Indique la durée accumulée (mode continu) ou le temps
restant (mode temporisé). La plage d’affichage est de 0 à 9999 minutes par incré-
ments d’une (1) seconde. L’affichage indique les minutes et les secondes jusqu’à
ce que la minuterie atteigne 99 minutes et 59 secondes (99:59), puis il affiche
automatiquement les minutes, jusqu’à 9999. L. Flèches haut/bas pour le contrôle
des valeurs de consigne. La touche marche/arrêt active et désactive la fonction
de minuterie.
Panneau De commanDe
FR
H.
L.I.
D.
A.
K.
C.
J.
E.
F.
G. B.
15
Dimensions hors-tout (l x L x h):
17,0 x 11,0 x 10,7”(43,2 x 27,9 x 27,0cm)
Dimensions internes (l x L x h): 11,3 x 8,3 x 5,7” (28,7 x 21,1 x 14,5cm)
Dimensions du plateau (l x L): 11 x 7,75” (27,9 x 19,7cm)
Spécifications électriques (50/60 Hz): 120 Volts, 5 Ampères, 450 Watts
230 Volts, 5 Ampères, 450 Watts
Fusibles : 5mm x 20mm, 5 A à action rapide
Plage de température: Ambiante + 5° C à 65° C
Uniformité des températures: ±0,5°C à 37°C
Plage de vitesses: 100 à 1200rpm
Précision de la vitesse: ±2% de la vitesse réglée jusqu’à 999rpm
± 5% 1000 à 1200rpm
Minuterie : d’1 seconde à 9999 minutes
(augmente par incréments d’1 seconde)
Orbite de mouvement: 0,125” (3mm)
Capacité massique maximum: ~ 8lbs (3,6kg), jusqu’à 1000rpm
~ 5lbs (2,3kg), plus de 1000rpm
Commandes: consultez la page 15
Matériau du plateau: aluminium
Poids avec emballage: 30lbs (13,6kg)
agitateur incubateur miniPlaqueS réglage
L’agitateur incubateur miniplaques est fourni avec un plateau destiné à accueillir une
gamme d’accessoires.
1. Les récipients plats peuvent être agités en les plaçant sur le plateau.
2. Le plateau est aussi muni de trous de montage prêts à l’emploi avec des pinces
à fiole ou des supports de tubes à essai.
SPécificationS - agitateur incubateur miniPlaqueS
FR
16
agitateur incubateur microPlaqueS - SPécificationS
Dimensions hors-tout (l x L x h):
17,0 x 11,0 x 7,75” (43,2 x 27,9 x 19,7cm)
Dimensions du plateau (l x L): 11 x 7,75” (27,9 x 19,7cm)
Spécifications électriques (50/60 Hz): 120 Volts, 5 Ampères, 450 Watts
230 Volts, 5 Ampères, 450 Watts
Fusibles : 5mm x 20mm, 5 A à action rapide
Plage de température: Ambiante + 5° C à 65° C
Uniformité des températures: ±0,5°C à 37°C
Plage de vitesses: 100 à 1200rpm
Précision de la vitesse: ±2% de la vitesse réglée jusqu’à 999rpm
± 5% 1000 à 1200rpm
Minuterie : d’1 seconde à 9999 minutes
(augmente par incréments d’1 seconde)
Orbite de mouvement: 0,125” (3mm)
Capacité: 4 microplaques ou
2 supports de microtubes
Commandes: consultez la page 15
Matériau du plateau: aluminium
Poids avec emballage: 20,5lbs (9,3kg)
FR
17
(4) microplaques(2) microplaques (2) supports de microtubes
agitateur incubateur microPlaqueS réglage
L’agitateur incubateur microplaques est destiné à accueillir deux (2) ou quatre (4) microplaques ou deux (2) supports de microtubes.
1. Placez deux (2) microplaques ou blocs de cupules profondes en diagonale sur le plateau ou placez quatre (4) microplaques ou blocs de cupules profondes sur le plateau.
Les blocs et plaques ne doivent pas être remplis.
2. Placez le coin de la plaque/bu bloc sous le ressort situé à chaque coin du plateau.
3. Faites glisser la plaque/le bloc en place. Vous pouvez utiliser l’appareil.
agitateur incubateur microPlaqueS configuration Du Plateau
2. 3. 4.
REMARQUE: Le plateau de l’agitateur incubateur microplaques n’est pas destiné à accueillir des pinces à fiole.
FR
18
Les agitateurs incubateurs droit léger/microplaques ont été conçus de façon à ce que
les fonctions de température, de vitesse et de minuterie fonctionnent indépendam-
ment les unes des autres. La température et la vitesse peuvent être réinitialisées
sans réinitialiser la minuterie et la minuterie peut être arrêtée et démarrée sans
interrompre les fonctions de chauffage et d’agitation.
1. Préparation:
a. Branchez le cordon d’alimentation dans une prise correctement mise à la terre.
Le témoin veille s’allume pour indiquer que l’appareil est bien alimenté.
b. Appuyez sur la touche veille pour basculer l’appareil du mode veille. Le témoin
veille s’éteint et les affichages de la durée, de la température et de la vitesse
s’allument pour indiquer les réglages précédents.
2. Réglage de la température:
a. Appuyez sur les flèches haut/bas situées sous l’affichage de la température
jusqu’à ce que la température souhaitée soit obtenue. Lorsque vous relâchez
la touche, l’affichage clignote pour indiquer que la nouvelle température a été
acceptée.
b. Appuyez sur la touche marche/arrêt pour activer la fonction de chauffage. Le
témoin situé sous l’affichage de la température s’allume pour indiquer que la
fonction de chauffage est en cours d’utilisation et reste allumé jusqu’à ce que
le chauffage cesse.
c. La température peut être modifiée sans interrompre le chauffage à l’aide
des flèches haut/bas situées sous l’affichage de la température. Une fois la
modification effectuée, et après avoir relâché la touche, l’appareil clignote pour
indiquer que la nouvelle température a été acceptée.
d. Appuyez sur la touche marche/arrêt située sous l’affichage de la température
pour arrêter le chauffage. Le témoin de chauffage s’éteint.
Témoin CAUTION HOT (Attention plaque chaude):
Le témoin Caution Hot signale que la température ambiante de la chambre
est supérieure à 40°C. Le témoin s’allume et reste
allumé lorsque la température ambiante de la cham-
bre atteint environ 40°C. Une fois le chauffage éteint,
le témoin Caution Hot reste allumé jusqu’à ce que la
température ambiante de la chambre soit inférieure à 40°C.
3. Réglage de la vitesse:
a. Appuyez sur les flèches haut/bas situées sous l’affichage de la vitesse jusqu’à
ce que la vitesse souhaitée soit obtenue. Lorsque vous relâchez la touche,
l’affichage clignote pour indiquer que la nouvelle vitesse a été acceptée.
b. Appuyez sur la touche marche/arrêt pour activer la fonction d’agitation. Le
témoin situé sous l’affichage de la vitesse s’allume pour indiquer que la fonction
d’agitation est en cours d’utilisation et reste allumé jusqu’à ce que l’agitation
cesse. La fonctionnalité de dosage contrôlée par microprocesseur augmente
lentement la vitesse jusqu’à atteindre la valeur de réglage afin d’éviter les
éclaboussements et fournit un excellent contrôle occasionnel.
c. La vitesse peut être modifiée sans interrompre l’agitation à l’aide des flèches
haut/bas situées sous l’affichage de la vitesse. Une fois la modification effec-
tuée, et après avoir relâché la touche, l’affichage clignote, indiquant que la
nouvelle vitesse a été acceptée.
d. Pour arrêter la fonction d’agitation, appuyez sur la touche marche/arrêt située
sous l’affichage de la vitesse. Le témoin de vitesse s’éteint.
4. Réglage de la durée à zéro (0 :00) et du mode continu: Durée cumulée.
a. Appuyez sans relâcher la touche marche/arrêt en dessous de l’affichage de
la minuterie. Au bout de trois (3) secondes, l’affichage indique la durée de
consigne précédente.
b. Appuyez simultanément sur les flèches haut et bas ; l’affichage indique zéro
conSigneS D’utiliSation - agitateurS incubateurS Droit léger/microPlaqueS
FR
19
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114
  • Page 115 115
  • Page 116 116
  • Page 117 117
  • Page 118 118
  • Page 119 119
  • Page 120 120
  • Page 121 121
  • Page 122 122
  • Page 123 123
  • Page 124 124
  • Page 125 125
  • Page 126 126

Ohaus ISLDMPHDGL Manuel utilisateur

Taper
Manuel utilisateur
Ce manuel convient également à