Gilson PIPETMAX 268 Manuel utilisateur

Taper
Manuel utilisateur
PIPETMAX® 268
Users Guide
Gilson, Inc. | PO Box 620027 | Middleton, WI 53562-0027, USA
Tel: 608-836-1551 or 800-445-7661 | Fax: 608-831-4451
Gilson S.A.S. | 19, avenue des Entrepreneurs | BP 145 | F-95400 Villiers-le-bel, FRANCE
www.gilson.com | [email protected] | [email protected] | training@gilson.com
LT255519-07
©2018 Gilson, Inc. | All rights reserved.
Table of Contents
Gilson PIPETMAX® 268 User’s Guide Table of Contents-1
Safety
Voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Safety-2
Safety Interlocks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Safety-3
Pipette Heads. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Safety-3
Tip Waste Container. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Safety-3
Tray . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Safety-3
Liquids . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Safety-4
Replacement Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Safety-4
Sécurité
Tension. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Sécurité-2
Dispositifs de sécurité . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Sécurité-3
Têtes de pipetage. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Sécurité-3
Conteneur pour pointes usagées . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Sécurité-3
Plateau . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Sécurité-4
Liquides . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Sécurité-4
Pièces Détachées . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Sécurité-4
1 Introduction
Unpacking. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Standard Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Technical Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
Customer Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
Repair and Return Policies. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
Before Calling Us . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
Warranty Repair. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
Non-Warranty Repair. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
Return Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
Unit End-of-Life . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
Table of Contents-2 Gilson PIPETMAX® 268 User’s Guide
2 Setup/Operation
Placement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
On Lab Bench or Lab Cart . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
In Hood. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Removable Tray. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Options. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
Rear Panel Diagrams . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
Drain Connection (Optional). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6
Front Panel Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-7
Power On and Start Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-8
Log In . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-8
User Management . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-9
Change Password . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-9
Edit User Properties . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-9
Add Users (Administrators Only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-10
Delete Users (Administrators Only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-10
Touchscreen Tablet Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-11
Icons and Fields . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-11
On-Screen Keyboard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-11
Shutdown . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-11
Run Alignment Protocol. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-12
Prepare to Run Alignment Protocol . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-12
Run Alignment Protocol. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-13
Run Application Protocol. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-14
Import Application Protocol . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-14
Add Pipette Head(s). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-14
Set Date and Time. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-14
Run Application Protocol . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-15
Pause or Stop a Running Protocol . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-23
Pause. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-23
STOP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-23
View Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-24
Tray View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-24
Labware View. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-25
Volume View. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-26
Steps View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-26
Run Report . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-27
Export and View Run Report . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-27
Export Protocol . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-28
Log Out. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-28
Gilson PIPETMAX® 268 User’s Guide Table of Contents-3
3 Messages
Communication . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Simulation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Hood Open . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Insufficient Volume/Excessive Volume . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
Refill Tip Rack . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
4 Maintenance
Cleaning. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
Add Pipette Heads . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
Replace Pipette Heads . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
Convert Blank Head into Alignment Head. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
A Replacement Parts
PIPETMAX® 268 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-1
TRILUTION® micro Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-1
Pipette Heads. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-1
Removable Trays. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-1
Tip Reload Block . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-1
Tip Ejector Feet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-2
Tips. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-2
Tip Waste Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-2
Racks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-3
Passive Temperature (PT) Racks. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-3
Circulating Temperature (CT) Racks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-3
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-3
Miscellaneous. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-3
Gilson PIPETMAX® 268 User’s Guide Safety-1
Safety
Read this section before installing and operating the PIPETMAX® 268.
This instrument is intended to be used in a laboratory environment by trained technical personnel.
For safe and proper use of this instrument, it is required that both operating and service personnel follow the instructions
contained in this guide when installing, cleaning, and maintaining the instrument.
The following safety precautions must be observed during all phases of operation, service, and repair of the instrument. Failure
to comply with these precautions or with specific warnings elsewhere in this user’s guide violates safety standards of design,
manufacture, and intended use of the instrument. Gilson assumes no liability for the customers failure to comply with these
requirements.
The PIPETMAX 268 has been certified to safety standards required in Canada, Europe, and the United States. Refer to the
instrument rear panel label and the Declaration of Conformity document for the current standards to which the instrument has
been found compliant.
Safety
Safety-2 Gilson PIPETMAX® 268 User’s Guide
Safety
The following electronic and hazard symbols may appear on the instrument:
The following safety notices may appear in this document:
Voltage
Access to the rear panel is necessary because the instrument must be detached from all voltage sources before service, repair,
or exchange of parts.
Operate the instrument using the approved power supply provided and only at the voltage specified on the rear panel label of
the instrument.
Symbol Explanation
Direct current Courant continu Gleichstrom
| Electrical power ON Sous tension Netzschalter ein
O Electrical power OFF Hors tension Netzschalter aus
Caution Attention Vorsicht
Mechanical hazard Danger méchanique Mechanische Gefährdung
WARNING indicates a potentially hazardous situation which, if not avoided, may result in serious injury
CAUTION indicates a potentially hazardous situation which, if not avoided, may result in minor or
moderate injury
NOTICE indicates a potentially hazardous situation which, if not avoided, may result in equipment damage
Gilson PIPETMAX® 268 User’s Guide Safety-3
Safety
Safety
Safety Interlocks
The instrument is configured with one of two safety interlocks. One interlock is used when the rotating cover is installed on the
instrument, and the other interlock is an external sensor and is used when the instrument is located in a hood.
The safety interlocks prevent the instrument from operating when the rotating cover or hood door is open.
When the instrument is installed in the hood, the external sensor must face the door of the hood and be located within
15 cm (6 in) of the door.
PIPETMAX 268 Configured with External Safety Interlock Sensor and Placed in Hood
Pipette Heads
The instrument is capable of generating significant forces that could cause potential injury to the user. For an instrument with
the rotating cover installed, the safety interlock in the cover will disable the instrument movement if the cover on the instrument
is open. For an instrument installed in a hood, the external safety interlock sensor mounted on the instrument will disable the
instrument movement if the hood door is open.
Tip Waste Container
The tip waste container may contain residues of flammable or infectious liquids in ejected tips.
If you use flammable liquids (e.g., ethanol 98%), treat the waste before disposing of it in accordance with your laboratory
guidelines.
Dispose of infectious material, waste, or tips in accordance with national and local safety regulations.
Tray
The moving tray is capable of causing injury by pinching. For an instrument with the rotating cover installed, the safety interlock
in the cover will disable the instrument movement if the cover on the instrument is open. For an instrument installed in a hood,
the external safety interlock sensor mounted on the instrument will disable the instrument movement if the hood door is open.
Safety
Safety-4 Gilson PIPETMAX® 268 User’s Guide
Safety
Liquids
Observe safe laboratory practices when handling liquids. If working with biological samples or chemical substances, ensure that
there is proper ventilation, and wear personal protective equipment (PPE), such as safety glasses, gloves, and protective clothing
at all times.
Refer to the Material Safety Data Sheets for solvents before use.
Replacement Parts
Be sure to use only replacement parts specified in this users guide. Do not repair or change parts which are not listed in this
users guide. If it is necessary to change parts not listed, please contact your local Gilson representative.
Gilson PIPETMAX® 268 User’s Guide Sécurité-1
Sécurité
Merci de lire ces instructions avant toute installation ou utilisation du PIPETMAX® 268.
Cet instrument est destiné à être utilisé dans un environnement de laboratoire, par un personnel technique qualifié.
Pour une utilisation correcte et en toute sécurité, il est nécessaire que le personnel qui utilise et réalise la maintenance de
l’instrument, suive les instructions contenues dans ce guide lors de l'installation, le nettoyage et la maintenance de l'instrument.
Les consignes de sécurité suivantes doivent être respectées durant toutes les phases de fonctionnement, d’entretien ou de
réparation de l'instrument. Le non-respect de ces précautions ou des avertissements spécifiques mentionnés dans ce guide
compromet les normes de sécurité de conception, de fabrication et d’utilisation prévue de l'instrument. Gilson décline toute
responsabilité en cas d’'incapacité du client à se conformer à ces exigences.
Le PIPETMAX 268 a été certifié conformément aux normes de sécurité en vigueur au Canada, en Europe et aux Etats-Unis. Merci
de vous reporter aux indications mentionnées sur le panneau arrière de l'instrument ainsi qu’au document de Déclaration de
Conformité aux normes pour lesquelles l'instrument a été déclaré conforme.
Sécurité
Sécurité-2 Gilson PIPETMAX® 268 User’s Guide
Sécuri
Les symboles suivants sont susceptibles d’apparaître sur l'instrument:
Les consignes de sécurité suivantes peuvent apparaître dans ce document:
Tension
L'accès au panneau arrière doit être libre car l'instrument doit pouvoir être déconnecté de sa source d’alimentation avant toute
opération d’entretien, de réparation ou de remplacement de pièces.
Ne faîtes fonctionner l'appareil qu’en utilisant le bloc d'alimentation fourni et uniquement à la tension indiquée sur l'étiquette
située à l'arrière de l'instrument.
Symbole Signification
Direct current Courant continu Gleichstrom
| Electrical power ON Sous tension Netzschalter ein
O Electrical power OFF Hors tension Netzschalter aus
Caution Attention Vorsicht
Mechanical hazard Danger mécanique Mechanische Gefährdung
AVERTISSEMENT signale une situation potentiellement dangereuse qui, si elle n'est pas évitée, peut
entraîner des blessures graves
ATTENTION signale une situation potentiellement dangereuse qui, si elle n'est pas évitée, peut entraîner
des blessures mineures ou légères
AVIS signale une situation potentiellement dangereuse qui, si elle n'est pas évitée, peut entraîner des
dommages matériels
Gilson PIPETMAX® 268 User’s Guide Sécurité-3
Sécurité
Sécurité
Dispositifs de sécurité
L'instrument présente deux dispositifs de sécurité. Le premier est activé lorsque le capot rotatif est installé sur l'appareil ; le
second est un capteur externe, activé lorsque l'instrument est placé sous une hotte.
Les dispositifs de sécurité empêchent tout fonctionnement de l'instrument lorsque le capot ou la porte de la hotte sont ouverts.
Lorsque l'appareil est installé sous une hotte, le capteur externe doit faire face à la porte de la hotte et être positionné à 15 cm
de cette porte.
PIPETMAX 268 Configuration sous hotte où le dispositif de sécurité est un capteur externe
Têtes de pipetage
L'instrument est capable de générer des forces importantes, susceptibles d'entraîner des blessures à l'utilisateur. Lorsque
l’appareil est équipé de son capot rotatif, le dispositif de sécurité désactive tout mouvement de l'appareil lors de louverture du
capot. Lorsque l’instrument est installé sous une hotte, le capteur externe permet de désactiver tout mouvement de l'instrument
lors de l’ouverture de la porte de la hotte.
Conteneur pour pointes usagées
Les pointes usagées peuvent contenir des résidus liquides dangereux, inflammables ou infectieux.
Si vous utilisez des liquides inflammables (ex. éthanol 98%), veillez au traitement approprié des déchets avant leur élimination,
en vous conformant aux directives de votre laboratoire.
Eliminez le matériel infectieux, les déchets ou les pointes usagées en vous conformant aux réglementations de sécurité
nationale et locale.
Sécurité
Sécurité-4 Gilson PIPETMAX® 268 User’s Guide
Sécuri
Plateau
Le plateau mobile est susceptible d'occasionner des blessures par pincement. Lorsque l’appareil est équipé de son capot rotatif,
le dispositif de sécurité désactive tout mouvement de l'appareil lors de l’ouverture du capot. Lorsque l’instrument est installé
sous une hotte, le capteur externe permet de désactiver tout mouvement de l'appareil lors de l’ouverture de la porte de la hotte.
Liquides
Respectez les pratiques de sécurité du laboratoire lors de la manipulation de liquides. Si vous travaillez avec des échantillons
biologiques ou des substances chimiques, assurez-vous que la ventilation est adéquate et portez en permanence un
équipement de protection individuelle (EPI), tel que : lunettes, gants et vêtements de protection.
Reportez-vous aux Fiches de Données de Sécurité pour les solvants avant toute utilisation.
Pièces Détachées
Assurez-vous de n’utiliser exclusivement que les pièces détachées préconisées dans ce guide. Ne tentez pas de réparer ou
remplacer des pièces ne figurant pas dans ce guide.
Si le remplacement de pièces ne figurant pas dans ce guide s’avérait nécessaire, merci de contacter votre représentant Gilson
local.
Gilson PIPETMAX® 268 User’s Guide 1-1
Introduction 1
The PIPETMAX® 268 is a small footprint personal pipetting station used to perform repetitive, precise, pipetting actions.
This chapter provides information on the following topics:
Unpacking
Technical Specifications
Customer Service
Repair and Return Policies
PIPETMAX® 268
with optional riser for off-bed tip disposal
Introduction
1
1-2 Gilson PIPETMAX® 268 User’s Guide
Unpacking
Unpacking
PIPETMAX is delivered with most major components already assembled. Keep the original packaging in case PIPETMAX must be
returned to the factory.
It is recommended that two people lift PIPETMAX out of the box, as it
weighs approximately 24.9 kg (55 lbs.).
1 Remove the open-ended cardboard box.
2 Remove the box of accessories and open it.
3 Lift the printed, outer cardboard box up to remove.
4 Grip PIPETMAX at the recesses near the base. There is one recess in the front, one in the
back, and one on each side. Use these recesses when lifting PIPETMAX out of the
foam-lined, cardboard tray.
Do not attempt to lift PIPETMAX from the cover or from the X-arm (the horizontal arm). Always lift the
instrument from its base.
5 Place PIPETMAX on a lab bench or cart or in hood.
6 Remove the bag covering PIPETMAX. You may need to lift the front and the back
slightly.
7 Remove the film protecting the rotating cover (if installed).
8 Remove shipping screws and bracket using the tool provided.
9 Store shipping screws and bracket in rear panel for future use if PIPETMAX needs to be
shipped or transported.
Do not store screws in rear panel without bracket as that will damage
PIPETMAX during operation.
10 Remove the shipping brace from the rear panel by loosening the thumb screws
securing the shipping brace to PIPETMAX. After removing the shipping brace, tighten
the thumb screws.
11 Remove the foam block preventing the tray from moving and securing the
alignment cable cord.
Standard Equipment
After PIPETMAX and its accessories are unpacked, you should have the following:
PIPETMAX® 268 with alignment head, dual eight-channel tip ejector foot, and rotating cover or external safety interlock
sensor installed
•USB Cable
•Tip Reload Block
•Tip Disposal Bin
•USB Drive
Plug for Outlet on the Rear Panel
Power Supply
Power Cords
Gilson PIPETMAX® 268 User’s Guide 1-3
Introduction
1
Unpacking
Documentation
The following documents are included with the PIPETMAX® 268:
Installation Qualification/Operational Qualification (IQ/OQ) Procedures
Declaration of Conformity
Quality Control (QC) Checklist
Items Included Checklist
•Setup Guide
Gilson Quality Control Report Pipette Head (one for each pipette head)
Protocols
Alignment Protocol
Standard Liquid Handling Protocols with Documentation
Accessories
Required
Some accessories are required, but are ordered separately.
Tablet with TRILUTION® micro pre-loaded or TRILUTION® micro on a USB drive for installation on a PC
•Pipette Head(s)
•Removable Tray
Blister Packs of Tips
•Racks
For part numbers, refer to Appendix A,
Replacement Parts.
Optional
The following optional accessories are also available:
Tablet Stand
Riser Kit for Off-bed Tip Disposal
•Orbital Shaker
Microscan 2D Barcode Reader
For part numbers, refer to Appendix A,
Replacement Parts.
Introduction
1
1-4 Gilson PIPETMAX® 268 User’s Guide
Technical Specifications
Technical Specifications
Please be aware of the following before operating the PIPETMAX® 268.
Changes or modifications to this device not expressly approved by Gilson could void the warranty.
The instrument complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This instrument
may not cause harmful interference, and (2) this instrument must accept any interference received, including interference that
may cause undesired operation.
Shielded cables must be used with the instrument to ensure compliance with the FCC Class A limits.
PIPETMAX® 268
Technical
Specification
Definition
Communication USB
Connections Three USB host ports and one USB device port
Two inputs (contact closure, TTL), two relay outputs, and one switched +12V DC 1A output
Switching voltages higher than 30V or greater than 1A of current
may damage the instrument.
Control Touchscreen tablet or PC control via USB and TRILUTION® micro software
Dimensions
(W x D x H)
PIPETMAX® 268 with rotating cover
54.4 x 65.5 x 53.1 cm (21.4 x 25.8 x 20.9 in)
PIPETMAX® 268 with rotating cover installed on optional riser assembly for off-bed tip disposal
54.4 x 65.5 x 69.6 cm (21.4 x 25.8 x 27.4 in)
PIPETMAX® 268 without cover
50.8 x 64.3 x 49.5 cm (20 x 25.3 x 19.5 in)
PIPETMAX® 268 without cover installed on optional riser assembly for off-bed tip disposal
52.3 X 65 X 65.8 cm (20.6 x 25.6 x 25.9 in)
Environmental
Conditions
Indoor use
Altitude: up to 2000 m
Temperature range: 5°C–40°C
Humidity: Maximum relative humidity 80% for temperatures up to 31°C, decreasing linearly to 50%
relative humidity at 40°C
Front Panel Two USB host ports and STOP button
Liquid Contact
Materials
Description Material
Tips 100% Virgin Polypropylene
Tip Disposal Bin Pro-fax 6523 (Polypropylene with colorant)
Tip Disposal Bin (Off-bed) Polypropylene
Tip Chute
Pro-fax 6523 (Polypropylene with colorant)
Aluminum 5052 with Hentzen, URA-ZEN, White, Matte Finish,
Fine Texture Paint
Pipette Heads and
DIAMOND® Tips
Pipette Head
Flow Rate
Range (mL/min)
DIAMOND®
Tips
Volume
Range (μL)
DIAMOND®
Filter Tips
Volume
Range (μL)
MAX8x20 0.0225–3.75
D200 1–20 DF30 1–20
DL10 1–20 DFL10 1–10
MAX8x200 0.225–37.5 D200 20–200 DF200 20–200
MAX1000 1.125–187.5 D1000 100–1000 DF1000 100–1000
Sterilized option available for some tips.
PIPETMAX® 268 Technical Specifications
(Page 1 of 2)
Gilson PIPETMAX® 268 User’s Guide 1-5
Introduction
1
Technical Specifications
Positioning
Performance
Machine Accuracy (without tips) in XYZ = +/-0.2 mm (0.008")
System Accuracy (with current tips) in XYZ = +/-0.9 mm (0.035")
Power
Requirements
External Power Supply
Voltage Input
Frequency: 50 to 60 Hz
Voltage: 100–240V AC
Voltage Output
Voltage: 24V DC
Current Rating: 6.25A, 150W
Removable Tray
Capacity
9-position removable tray (microplate footprints, but not for 384-well microplates)
9-position removable tray for 384-well microplates
Safety and
Compliance
The PIPETMAX® 268 has been certified to safety standards specified for Canada, Europe, and the
United States. Refer to the instrument rear panel label and the Declaration of Conformity document for
the current standards to which the instrument has been found compliant.
Volumetric
Specification for
Transfers
Maximum Permissible Errors
Pipette Head
Volume of
Distilled Water
Systematic
Error (μL)
Random
Error (μL)
Systematic
Error (%)
Random
Error (%)
MAX8x20
1 μL ±0.08 0.05 8 5
10 μL ±0.15 0.10 1.5 1
20 μL ±0.25 0.12 1.25 0.6
MAX8x200
20 μL ±0.50 0.16 2.5 0.8
100 μL ±1.00 0.30 1 0.3
200 μL ±2.00 0.50 1 0.25
Weight 24.9 kg (55 lbs.)
XYZ Motion
Parameters
X Min Max Default
Power 1% 100% 50%
Speed 1 mm/sec 550 mm/sec 350 mm/sec
Acceleration Speed Value
4000 mm/sec
2
2200 mm/sec
2
Y Min Max Default
Power 1% 100% 50%
Speed 1 mm/sec 550 mm/sec 350 mm/sec
Acceleration Speed Value
4000 mm/sec
2
900 mm/sec
2
Z Min Max Default
Power 1% 100% 60%
Speed 1 mm/sec 140 mm/sec 120 mm/sec
Acceleration Speed Value
4000 mm/sec
2
650 mm/sec
2
PIPETMAX® 268 (Continued)
Technical
Specification
Definition
PIPETMAX® 268 Technical Specifications
(Page 2 of 2)
Introduction
1
1-6 Gilson PIPETMAX® 268 User’s Guide
Customer Service
Customer Service
Gilson, Inc. and its worldwide network of authorized representatives provide customers with the following types of assistance:
sales, technical support, applications, and instrument repair.
If you need assistance, please contact your local Gilson representative. Specific contact information can be found at
www.gilson.com
. To help us serve you quickly and efficiently, please refer to Before Calling Us on page 1-7.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59

Gilson PIPETMAX 268 Manuel utilisateur

Taper
Manuel utilisateur

dans d''autres langues