Ohaus ST400D Manuel utilisateur

Catégorie
Mesure, test
Taper
Manuel utilisateur

Ce manuel convient également à

STARTER 400D
Portable DO Meter
Instruction Manual
Medidor de DO Portátil
STARTER 400D
Manual de instrucciones
STARTER 400D
Oxymètre portable
Manuel d'instructions
TABLE OF CONTENTS
1 INTRODUCTION ................................................................... 1
1.1 Definition of Signal Warnings and Symbols ...................................... 1
1.2 Safety Precautions ............................................................................ 2
1.3 Display and controls .......................................................................... 3
2 INSTALLATION ..................................................................... 5
2.1 Package contents .............................................................................. 5
2.2 Installing the batteries ....................................................................... 5
2.3 Installing the electrode clip ................................................................ 6
2.4 Installing the IP54 cover caps ........................................................... 6
2.5 Integrated stand for table top use ..................................................... 6
3 SETUP ................................................................................... 7
3.1 Setting Auto/Manual Endpoint ........................................................... 7
3.2 Setting Salinity Value ........................................................................ 7
3.3 Setting Temperature Unit .................................................................. 8
3.4 Setting the Date and Time ................................................................ 8
3.5 Setting Power State........................................................................... 8
3.6 Setting Factory Reset ........................................................................ 8
3.7 Setting New Probe Cap Parameter ................................................... 9
4 OPERATION ........................................................................ 10
4.1 Calibration ....................................................................................... 10
4.1.1 Zero-oxygen solution ................................................................ 10
4.1.2 Performing 1-point calibration .................................................. 10
4.1.3 Performing 2-point calibration ................................................... 11
4.2 Sample measurement ...................................................................... 11
4.3 Temperature measurement .............................................................. 11
4.4 Barometer pressure measurement .................................................. 11
4.5 Using the memory ........................................................................... 12
4.5.1 Storing a reading ...................................................................... 12
4.5.2 Recalling from memory ............................................................ 12
4.5.3 Clearing the memory ................................................................ 12
5 MAINTENANCE .................................................................. 13
5.1 Error message ................................................................................. 13
5.2 Meter maintenance.......................................................................... 13
5.3 Probe maintenance ......................................................................... 14
6 TECHNICAL DATA ............................................................. 15
6.1 Specifications ................................................................................ 15
STARTER 400D Portable DO Meter EN-1
1 INTRODUCTION
Thank you for choosing OHAUS. Please read the manual completely before using the
STARTER 400D portable Optical DO (Dissolved Oxygen) meter to avoid incorrect operation.
The STARTER 400D has an excellent performance/price ratio and is designed with many useful
features. Optical DO probe Other accessories are also available from OHAUS.
Starter 400D offers many practical features such as:
Ergonomic, multifunctional design for ambidextrous measurements or bench top use
Meter can measure barometric pressure automatically to compensate the DO value; Salinity
value can be set to compensate the DO value.
IP54 protection against dust and water
Ohaus’ renowned user friendly software
1.1 Definition of Signal Warnings and Symbols
Safety notes are marked with signal words and warning symbols. These show safety issues and
warnings. Ignoring the safety notes may lead to personal injury, damage to the instrument,
malfunctions and false results.
Signal Words
WARNING For a hazardous situation with medium risk, possibly resulting in
injuries or death if not avoided.
CAUTION For a hazardous situation with low risk, resulting in damage to the
device or the property or in loss of data, or injuries if not avoided.
Attention For important information about the product
Note For useful information about the product
Warning Symbols
General Hazard
Explosion Hazard Corrosive Hazard
EN-2 STARTER 400D Portable DO Meter
1.2 Safety Precautions
CAUTION Read all safety warnings before installing, making connections, or
servicing this equipment. Failure to comply with these warnings could result in
personal injury and/or property damage. Retain all instructions for future reference.
Use the equipment only in dry locations.
Dry off any liquid spills immediately. The instrument is not watertight.
Use only approved accessories and peripherals.
Operate the equipment only under ambient conditions specified in these instructions.
Do not operate the equipment in hazardous or unstable environments.
Service should only be performed by authorized personnel.
If the equipment is used in a manner not specified by the manufacturer, the protection
provided by the equipment may be impaired.
WARNING Do not use this instrument in an environment subject to explosion
hazards. The housing of the instrument is not gas tight (explosion hazard due to
spark formation, corrosion caused by the ingress of gases).
CAUTION When using chemicals and solvents, comply with the instructions of the
producer and the general lab safety rules.
STARTER 400D Portable DO Meter EN-3
1.3 Display and controls
Display
1
Time: HH:MM:SS
2
Date: YYYY-MM-DD
3
calibration data: |K-1|0.3 and |B|<0.1
Probe condition is very good
calibration data: 0.3|K-1|0.4 and |B| <0.1
Probe condition is acceptable.
4
Measurement icon - measurement (calibration) is running
5
Calibration icon - calibration is running
6
Power state icon, include power always on icon , means meter will not automatically turn
off; and auto-off icon which means after 10 minutes no operation meter will turn off(see
4.4); power state icon position is at the top-left of the battery status icon.
7
Battery status icon – the charged status of the battery
8
DO reading: mg/L and %
9
Temperature during measurement or in calibration process
10
Barometric pressure during measurement or in calibration process
11
Data log: memory number
12
ppt for salinity entry
11
12
1
6
5
4
3
2
7
8
9
10
EN-4 STARTER 400D Portable DO Meter
Controls
Button
Press & release
Press & hold
for 2 seconds
- Start or endpoint measurement
(manual endpoint mode)
- Confirm setting, confirm entered
value for setting
- switch between auto endpoint
and manual endpoint mode
- Start calibration
- Review the latest calibration
data
- Meter turn on
- Back to measurement screen
- Meter turn off
- Store current reading to memory
- Increase value during setting
- Scroll up through the memory
- Recall stored data
- Enter setup screen
- Scroll down through the memory
- Reset factory setting
STARTER 400D Portable DO Meter EN-5
2 INSTALLATION
Carefully unpack the box.
2.1 Package contents
The ST400D kit has the following items:
We also offer ST400D-G kit which includes ST400D kit and a portable bag.
Accessories:
Model
Description
P/N
STDO21
Optical DO probe
30219117
3020
DO Zero Oxygen Chemical
30059257
Portable Bag for portable meters
30031635
2.2 Installing the batteries
Remove the battery cover screw using a coin or screwdriver and remove the battery cover
Insert the batteries in the battery compartment as shown
Replace the battery cover and tighten the battery cover screw
ST400D
Units
STARTER 400D meter
1
STDO21 (with 1.0m cable)
1
AAA battery
4
Electrode Clip
1
IP54 cover caps
1 set
Wrist Strap
1
EN-6 STARTER 400D Portable DO Meter
2.3 Installing the electrode clip
The electrode clip is an electrode holder that can be placed next to the display on either side of
the housing.
Attach the clip by inserting the tabs into the recess.
Slide the clip forward to lock in position. Slide the shaft of the probe into the clip from above.
2.4 Installing the IP54 cover caps
Use the small plastic tool as shown:
2.5 Integrated stand for table top use
Starter 400D has an integrated table stand as shown, which can be pushed out to use the meter on
a lab bench or desk. To close the stand, push the middle edge of the stand toward the meter.
STARTER 400D Portable DO Meter EN-7
3 SETUP
Press to enter the setup interface. We have 7 setting items, they are:
1) Auto/Manual Endpoint
2) Salinity Value
3) Temperature Unit
4) Date and Time
5) Power State
6) Reset Factory Setting
7) New Probe Cap Parameter
3.1 Setting Auto/Manual Endpoint
When in calibration, ST400D is only work at Auto End Point Mode.
The default endpoint mode of ST400D for measurement is Auto End Point Mode. When users
want to judge the endpoint manually, you can shift to Manual Endpoint Mode.
End point Mode:
o When in Manual Mode, to manually reach and lock a DO measurement, you need
to press button- when reading is stable then the measurement value
freezes; And meter displays .
o When in Auto End Point Mode, the meter determines when the reading is stable
then locks the measurement or calibration value automatically, the reading freezes;
the meter displays .
Users can also change the endpoint mode by shortcut -- Press and hold button- to shift the
End Point Mode.
3.2 Setting Salinity Value
Since salinity has an influence on the dissolved oxygen value, it is important to input the correct
salinity value. The meter can compensate salinity for measurements in solutions with salinity from
0 to 50 ppt. The default value for salinity is 0.0 ppt.
Press button- or button- to adjust the salinity value.
Press button- to confirm the value or press button- to return to the Setting screen.
Normally seawater salinity is 35ppt, which means 35g of salt per 1000g seawater.
EN-8 STARTER 400D Portable DO Meter
3.3 Setting Temperature Unit
Ohaus optical DO probe STDO21 include a built-in temperature probe, the meter will be in
Automatic Temperature Compensation (ATC) mode automatically. You can select the temperature
unit according to your preferred: °C or °F .
Note: °C = 5/9 (°F -32)
The default setting is 25 °C (77°F).
3.4 Setting the Date and Time
The Date format is yy-mm-dd. Press button- or button- to adjust the value of each
numerical digit. Press button- to confirm current value then goes to next numerical digit.
3.5 Setting Power State
We have 2 power states: "Auto OFF" or "Always ON".
When "Auto OFF" is selected, which means the meter will turn off the powder without any press
operation after 10 minutes and auto-off icon will display at the top-left of the battery status icon.
"Always ON" means power will not turn off automatically, and the power always-on icon
displays.
Ohaus suggest you to select "Auto OFF” so to save power of batteries.
3.6 Setting Factory Reset
Press button- and button- together to enter the Reset Factory Setting screen.
STARTER 400D Portable DO Meter EN-9
3.7 Setting New Probe Cap Parameter
When replace a new probe cap, user must input the parameter of this new cap into the meter. Or
the measurement value will be incorrect.
Press button- , button- and button- together to enter the new probe cap parameter
setting screen. Input the parameter of the new cap. Then do calibration before any measurement.
Probe Body
Probe
Cap
Protection
Cover
O-ring
EN-10 STARTER 400D Portable DO Meter
4 OPERATION
WARNING Do not operate the equipment in hazardous environments. The
equipment is not explosion protected.
WARNING When using chemicals and solvents, comply with the instructions of the
chemical producer and the general lab safety rules.
4.1 Calibration
STARTER 400D allows you to perform 1-point or 2-point calibrations; Ohaus suggest 1-point
calibration(100% DO calibration) for samples which DO value is not less than 2 mg/L.
Only when measure samples which DO value is lower than 2 mg/L, Ohaus suggest 2-point
calibration.
4.1.1 Zero-oxygen solution
Preparing a zero-oxygen solution: put approximately 20ml of pure or deionized water into the
bottle of DO Zero Oxygen Chemical (30059257, there are about 5g powder in it), stir for at least 5
minutes.
This zero-oxygen solution is used for 2-point calibration.
4.1.2 Performing 1-point calibration
When performing calibration, the meter only uses Auto End Point Mode.
1 point calibration can be performed in the air or oxygen-saturated water. There will be a little
difference between calibration in the air and in oxygen-saturated water.
Before calibration you need to rinse the DO probe with pure water and carefully wipe the water off
with a tissue.
o Place the probe in the air (or in oxygen-saturated water).
o Press button- , the calibration icon appear on the display.
o The meter reaches endpoints automatically, 100% will display. Measurement icon
disappears.
STARTER 400D Portable DO Meter EN-11
Once you have completed your 1-point calibration, you have 3 options new:
o Press button- to finish 1-point calibration then return to measurement screen.
o Press button- to reject calibration and return to measurement screen.
o Press button- to complete a 2-point calibration (after preparing the zero
oxygen solution and putting STDO21 into it).
4.1.3 Performing 2-point calibration
Perform the 1-point calibration.
Place the probe in the zero oxygen solution and press button- .
The meter reaches the endpoint automatically. 0% will display then return to measurement
screen.
Pressing and holding button- will display the calibration data.
4.2 Sample measurement
Dissolved oxygen readings can display unit of mg/L or %.
Place the DO probe in the sample. Press to start the measurement.
When the meter reaches endpoint, the measurement reading freezes on the screen.
4.3 Temperature measurement
For better accuracy, we recommend the use of the temperature probe. If a temperature probe is
used, ATC and the sample temperature are displayed.
STDO21 has a built-in temperature probe.
4.4 Barometer pressure measurement
Barometric pressure influences the DO value.
STARTER 400D has a built-in barometric pressure sensor which can measure the barometric
pressure automatically. This meter uses barometric pressure unit KPa (kilopascal).
EN-12 STARTER 400D Portable DO Meter
4.5 Using the memory
4.5.1 Storing a reading
The STARTER 400D can store up to 99 measurement readings. Press button- when the
measurement reaches end point. indicates that is the 2
nd
measurement reading had
been stored. The starter 400D can store up to 99 measurement readings.
4.5.2 Recalling from memory
Press and hold button- to recall the stored values from memory.
Press button- or button- to scroll through the stored results. Press button- to go
back to the measurement screen.
4.5.3 Clearing the memory
Continue pressing button- to scroll through the stored results until "Clear All" appears.
Press button- to confirm the deletion or press button- to return to the measurement
mode without deleting all the data in the memory.
STARTER 400D Portable DO Meter EN-13
5 MAINTENANCE
5.1 Error message
Low battery
If low battery, the battery icon will blink to show need
new battery. Then turn off after 10 seconds.
Communication fail
When do calibration or new cap parameter input, meter
will detect if the probe is connected properly. If not,
meter displays " Fail, please confirm the STDO21 probe
is connected”
Measured temperature out
of range (< -5°C or >55 °C)
Meter will displays "Temperature Out of Range!"
Please keep the sample temperature within the range for
measurement.
Calibration out of range
Meter displays "Calibration Fail".
|K-1| >0.4 or a |B| >0.1
You may need to do a proper calibration again, if
calibration data is still not good, you may need to replace
a new probe cap or probe.
Recall memory fail
Meter displays "No Data"
When recall memory while no data memorized yet.
The current data set has
already been stored once
Meter displays" Already saved!"
A measurement can only be stored once. Perform a new
measurement to store a new data set.
Any more problems, you can contact Ohaus.
5.2 Meter maintenance
The STARTER 400D does not require any maintenance other than replacement of depleted batteries.
To clean, use a damp cloth.
Attention: Do not disassemble the instrument. There are no user serviceable parts inside.
Attention: The housing is made of acrylonitrile butadiene styrene (ABS). This material is
susceptible to damage by some organic solvents, such as toluene, xylene and methyl ethyl ketone
(MEK). Any spillage should be immediately wiped off.
EN-14 STARTER 400D Portable DO Meter
5.3 Probe maintenance
The DO probe measures the oxygen based on luminescence principle. The optical DO probe
measures dissolved oxygen by emitting a blue light of the proper wavelength that causes the
sensing element to luminesce. Oxygen dissolved in the sample continually passes through the
diffusion layer to the sensing layer, affecting the luminescence lifetime. STDO21 can measure the
lifetime of the luminescence as it is affected by the presence of oxygen and compared with the
reference. This technology has been approved by EPA.
When measurement, optical probe did not consume the DO in the sample, that means the sample
do not need stir
Optical probe STDO21 do not need to replace the membrane or any maintenance of anode or
cathode. The optical probe only requires that its sensor cap be replaced every 12-18 months.
Please input the new cap parameter when new probe cap in installed, then do calibration before
any measurement.
When storing the probe for a longer period, make sure it is placed in the wetting protection cover.
Calibration frequency: Ohaus suggest verifying the calibration regularly according to your
measurement requirement. STDO21 only has very little calibration drift when compared to
electrochemical probes. This advantage results in less frequent calibrations. In fact, optical probes
are so stable they are capable of holding their calibration for many months.
Please also read the STDO21 probe manuals before using it.
STARTER 400D Portable DO Meter EN-15
6 TECHNICAL DATA
6.1 Specifications
Ambient conditions
Altitude: Up to 2000 m
Specified Temperature range: 5°C to 40°C
Humidity: maximum relative humidity 80 % for temperatures up to 31°C decreasing linearly to
50% relative humidity at 40°C
Installation category: N/A
Pollution degree: N/A
Operability is assured at ambient temperatures between 5°C to 40°C
Model: ST400D
Measurement Range
%
0.0 to 200.0%
mg/L
0.00 to 20.00mg/L (ppm)
Temp.
0 to 50 °C
Measurement Resolution
%
0.1%
mg/L
0.01 mg/L
Temp.
0.1 °C
Accuracy
± 0.2 mg/L (<8 mg/L)
± 0.3 mg/L (8 to 20 mg/L);
± 0.3 °C
Barometric
Range
50.0 to 115.0 kPa
Resolution
0.1 kPa
Accuracy
1.5 kPa
Temp compensation
ATC
Salinity compensation
0.0 to 40.0 ppt
Calibration
1 point or 2 points
Memory
99 sets, last calibration data
Power
4 x AAA(LR03) batteries
Size/Weight
90 W x 150 D x 35 H mm
Weight
0.16kg
Display
liquid crystal
IP rating
IP54
T input
4-pin screw
Materials
ABS
EN-16 STARTER 400D Portable DO Meter
Compliance
Compliance to the following standards is indicated by the corresponding mark on the product.
Mark
Standard
This product complies with the applicable harmonized standards of EU Directives
2011/65/EU (RoHS) and 2014/30/EU (EMC). The EU Declaration of Conformity is
available online at www.ohaus.com/ce.
This product complies with the EU Directive 2012/19/EU (WEEE). Please dispose of
this product in accordance with local regulations at the collecting point specified for
electrical and electronic equipment.
For disposal instructions in Europe, refer to www.ohaus.com/weee.
EN 61326-1
ISED Canada Compliance Statement
This Class A digital apparatus complies with Canadian ICES-003.
ISO 9001 Registration
The management system governing the production of this product is ISO 9001 certified.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64

Ohaus ST400D Manuel utilisateur

Catégorie
Mesure, test
Taper
Manuel utilisateur
Ce manuel convient également à

dans d''autres langues