Tecno Control TS282EH Manuel utilisateur

Catégorie
Détecteurs de monoxyde de carbone (CO)
Taper
Manuel utilisateur
IST-2282.EC01.01 Istruzione / User’s Manual / Manuel d’utilisation Pag.1/17
(
IST
-
2282.EC01.01_TS282E (05.10.2017).docx
)
Rivelatore 4÷20mA di Gas tossici
4÷20mA Toxic Gas Detector
Détecteur 4÷20mA de gaz toxique
TECNOCONTROL S.r.l. - Via Miglioli, 47 20090 SEGRATE (MI) ITALY Tel. +39 02 26922890 - Fax +39 02 2133734
http: www.tecnocontrol.it e-mail: [email protected]
TS282E
Leggere Attentamente e Conservare
quest’Istruzione.
Please read and keep this manual
Lire avec soin et garder la notice d’instruction
Modello
/ Model / Modele
Calibrato per
/ Calibrated for / Tarée pour
Cartuccia/Cartridge/Cartouche
TS282EA (NH
3
) Ammoniaca / Ammonia ZSEA / IP
TS282EA-H (NH
3
) Ammoniaca / Ammonia ZSEA-H / IP
TS282EC-S (CO) Monossido di carbonio / Carbon monoxyde ZS EC / IP
TS282EC-H (CO) Monossido di carbonio / Carbon monoxyde ZS EC-H / IP
TS282ECL (CL
2
) Cloro / Chlorine ZS ECL / IP
TS282EH (H
2
S) Idrogeno Solforato / Hydrogen sulphide ZS EH / IP
TS282EHCL (HCL) Acido cloridrico / Hydrogen Chloride ZS EHCL / IP
TS282EHCN (HCN) Acido cianidrico / Hydrogen cyanide ZS EHCN / IP
TS282EN (NO) Ossido di azoto / Nitric oxide ZS EN / IP
TS282EN2 (NO
2
) Biossido di azoto / Dioxyde d’azote ZS EN2 / IP
TS282ES (SO
2
) Biossido di zolfo / Sulfur dioxide ZS ES / IP
Con Cartuccia Sensore Sostituibile
Inside Replaceable Cartridge Sensor / Avec Cartouche Capteur échangeable
Sommario – Contents – Sommaire
Caratteristiche tecniche / Technical specifications / Caractéristiques techniques ......................................... 2
IT
DESCRIZIONE
........................................................................................................................................... 4
NOTE SUL GAS RILEVATO DAI VARI MODELLI ................................................................................................. 4
FUNZIONAMENTO .................................................................................................................................................. 5
INSTALLAZIONE .................................................................................................................................................... 5
AVVERTENZE ......................................................................................................................................................... 6
VERIFICHE E CALIBRAZIONE .............................................................................................................................. 6
E
N
DESCRIPTION
........................................................................................................................................... 7
NOTES ON THE DETECTED GAS ......................................................................................................................... 7
OPERATIONAL DESCRIPTION ............................................................................................................................. 8
INSTALLATION ....................................................................................................................................................... 9
WARNING ................................................................................................................................................................ 9
TEST and CALIBRATION ..................................................................................................................................... 10
FR
DESCRIPTION
......................................................................................................................................... 11
MODÈLES ET CARACTÉRISTIQUES DU GAZ DÈTECTEÈ ............................................................................... 11
FONCTIONNEMENT ............................................................................................................................................. 12
INSTALLATION ..................................................................................................................................................... 13
INSTRUCTIONS .................................................................................................................................................... 13
VÉRIFICATIONS ET ETALONNAGE .................................................................................................................... 14
Tabella 1 / Table 1 / Tableau 1 ............................................................................................................................. 16
Tabella 2: Interferenze con altri Gas / Table 2: Typical cross Sensitivities / Tableau 2: Interférence ......... 16
Tabella 3: Interferenze con altri Gas / Table 3: Typical cross Sensitivities / Tableau 3: Interférence ......... 16
Tabella 4 / Table 4 / Tableau 4 ............................................................................................................................. 16
E
C
-
S
IST-2282.EC01.01 Istruzione / User’s Manual / Manuel d’utilisation Pag. 2/17
TECNOCONTROL S.r.l. - Via Miglioli, 47 20090 SEGRATE (MI) ITALY Tel. +39 02 26922890 - Fax +39 02 2133734
Caratteristiche tecniche / Technical specifications / Caractéristiques techniques
Alimentazione /
Power supply / Alimentation
12÷24Vcc (-10/+15%) 2W / 12÷24Vdc(-10/+15%) 2W / 124Vcc (-10/+15%) 2W
Sensore /
Sensor Type / Capteur
Elettrochimico / Electrochemical / électrochimique
Cartuccia Sensore
Cartridge Sensor / Cartouche capteur
Sostituibile / Replaceable / échangeable
Uscite /
Outputs / Sortie
4 ÷ 20 mA Lineare / Linear / linéaire
Resistenza di carico
load resistor / résistance de charge
50 ohm / 12Vdc (-10%) - 500 ohm / 24Vdc (-10%)
Pressione di funzionamento
Operation Pressure / Pression de fonctionnement
Atmosferica ±10% / Atmospheric±10% / Atmosphérique ±10%
Temperatura di immagazzinamento
Storage Temp / Température de stockage
0 ÷ + 20 °C
Umidità di immagazzinamento
Storage Temp / Hgrométrie de stockage
5 ÷ 95 % RH
non condensata / non condensed / non condensée
Grado di protezione -Dimensioni – Peso
IP Code - Size - Weight
Indice de protection - Dimensions du boîtier - poids
IP65 / 238 x 106 x 75 mm / 350g
Modelli / Models / Modele
TS282EA TS282EA-H TS282EC-S TS282EC-H TS282ES TS282EH
Campo di misura
Standard Range / Champ de mesure
0÷300 ppm
NH
3
0÷300 ppm
NH
3
0÷300 ppm
CO
0÷300 ppm
CO
0÷20 ppm
SO
2
0÷100 ppm
H
2
S
Limite Funzionamento Sensore
Sensor limits / Limite échelle
500 ppm 500 ppm 1000 ppm 1000 ppm 150 ppm 500 ppm
Vita media in aria pulita
Average Life in fresh air / Vie moyenne en air pur
2 anni
years / ans
2 anni
years / ans
3 anni
years / ans
2 anni
years / ans
2 anni
years / ans
2 anni
years / ans
Tempo massimo di immagazzinamento
Max Storage Time / Temps maximum de stockage
3 mesi
month / mois
1 mese
month / mois
3 mesi
month / mois
3
mesi
month / mois
3 mesi
month / mois
3 mesi
month / mois
Variazione dello Zero
/ Zero shift / Dérive du Zero
< 6 ppm
< 4 ppm
< 10 ppm
< 5 ppm
0.4 ppm
< 0.2 ppm
Deriva a lungo termine
Long time drift / Dérive à long terme
< 2% mese
month / mois
1% mese
month / mois
< 2% mese
month / mois
< 2% mese
month / mois
< 2% mese
month / mois
< 2% mese
month / mois
Risoluzione
/ Resolution / Résolution
1 ppm
1 ppm
1 ppm
1 ppm
0.1 ppm
1 ppm
Tempo di risposta T
90
Response Time T
90
/ Temps de réponse T
90
<120
secondi
Seconds / secondes
(*1)
<90 secondi
Seconds / seconds
)
<60 secondi
Seconds / secondes
40 secondi
Seconds / secondes
≤75
secondi
Seconds / secondes
50 secondi
Seconds / secondes
Ripetibilità / Repeatability / Répétitivité
±10
%
del segnale /
of
the signal / du signal
±10
%
del segnale /
of
the signal / du signal
±2
%
del segnale /
of
the signal / du signal
<2
%
del segnale /
of
the signal / du signal
<2
%
del segnale /
of
the signal / du signal
<2
%
del segnale /
of
the signal / du signal
Precisione /
Accuracy / Precision
10 %
10 %
10 %
5 %
2 %
5%
Temperatura di funzionamento
Operation Temp. / Température de fonctionnement
-20÷+40 °C -20÷+40 °C -20÷+50 °C -20÷+50 °C -10÷+50 °C -10÷+40°C
Umidità di funzionamento /
Operation Humidity
/
Hygrométrie de fonctionnement
15÷90 % RH
non condensata / non condensed / non condensée
(*1) NOTA / NOTE/ REMARQUE: calcolato da 5 min. di tempo d’esposizione / calculated from 5 min. exposure time / calculée à partir de 5 min. temps d'exposition.
IST-2282.EC01.01 Istruzione / User’s Manual / Manuel d’utilisation Pag.3/17
TECNOCONTROL S.r.l. - Via Miglioli, 47 20090 SEGRATE (MI) ITALY Tel. +39 02 26922890 - Fax +39 02 2133734
Modelli / Models / Modele
TS282EHCN TS282EN TS282EN2 TS282EHCL TS282ECL
Campo di misura
Standard Range / Champ de mesure
0÷10 ppm
HCN
0÷100 ppm
NO
0÷30 ppm
NO
2
0÷10 ppm
HCL
0÷10 ppm
CL
2
Limite Funzionamento Sensore
Sensor limits / Limite échelle
100 ppm 1000 ppm 150 ppm >30 ppm 50 ppm
Vita media in aria pulita
Average Life in fresh air / Vie moyenne en air pur
2 anni
years / ans
2 anni
years / ans
2
anni
years / ans
2 anni
years / ans
2 anni
years / ans
Tempo massimo di immagazzinamento
Max Storage Time / Temps maximum de stockage
3 mesi
month / mois
3 mesi
month / mois
3 mesi
month / mois
3 mesi
month / mois
3 mesi
month / mois
Variazione dello Zero
/ Zero shift / Dérive du Zero
< 1 ppm
< 4 ppm
0.4 ppm
< 1 ppm
< ± 0.2
ppm
Deriva a lungo termine
Long time drift / Dérive à long terme
< 2% mese
month / mois
< 2% mese
month / mois
< 2% mese
month / mois
< 3% mese
month / mois
< 2% mese
month / mois
Risoluzione
/ Resolution / Résolution
0.1 ppm
1 ppm
1 ppm
< 0,7 ppm
0.
1 ppm
Tempo di risposta T
90
Response Time T
90
/ Temps de réponse T
90
≤ 120
secondi
Seconds / secondes
60 secondi
Seconds / secondes
60 secondi
Seconds / secondes
(*2)
70 secondi
Seconds / secondes
50 secondi
Seconds / secondes
Ripetibilità / Repeatability / Répétitivité
<2
% del
segnale
/
of
the signal / du signal
<2
% del
segnale
/
of
the signal / du signal
2
% del
s
egnale
/
of
the signal / du signal
±10
% del
segnale
/
of
the signal / du signal
2
% del
Segnale
/
of
the signal / du signal
Precisione /
Accuracy / Precision
2
%
5
%
5
%
5 %
2 %
Temperatura di funzionamento
Operation Temp. / Température de fonctionnement
0÷+50°C -20÷+50°C -20÷+50°C -20÷+40 °C -20÷+50°C
Umidità di funzionamento /
Operation Humidity
/
Hygrométrie de fonctionnement
15÷90 % RH
non condensata / non condensed / non condensée
(*1)NOTA / NOTE/ REMARQUE: calcolato da 5 min. di tempo d’esposizione / calculated from 5 min. exposure time / calculée à partir de 5 min. temps d'exposition.
(*2) NOTA / NOTE/ REMARQUE: calcolato da 4 min. di tempo d’esposizione / calculated from 4 min. exposure time / calculée à partir de 4 min. temps d'exposition.
IST-2282.EC01.01 Istruzione / User’s Manual / Manuel d’utilisation Pag. 4/17
TECNOCONTROL S.r.l. - Via Miglioli, 47 20090 SEGRATE (MI) ITALY Tel. +39 02 26922890 - Fax +39 02 2133734
IT
DESCRIZIONE
I TS282E sono trasmettitori 4÷20mA con sensori a cella elettrochimica, utilizzati in sistemi centralizzati d’allarme
per parcheggi, laboratori, industrie e ambienti da proteggere da possibili fughe di gas tossici, indicati a pag. 1.
Il rilevatore è costituito da una custodia che contiene il circuito elettronico e i morsetti di collegamento. Nel porta-
sensore, posto nella parte inferiore della custodia, è posta la “Cartuccia Sensore Sostituibile” che contiene il sensore e
i dati identificativi e di taratura. I TS282E hanno un segnale d’uscita (S) 4÷20mA lineare con scala in ppm (parti per
milione) del gas misurato. Questa uscita va collegata a una centrale rilevazione gas. (Vedi tabella 1). Sul coperchio so-
no posti i tasti F1 e F2 per le operazioni di verifica e calibrazione utilizzabili solo tramite codice e i 3 Led.
Led rosso "ALARM": Segnalazione ottica d’allarme.
Led verde "ON": Funzionamento normale.
Led giallo "FAULT": Sensore guasto o scollegato o a fondo scala o scaduto.
NOTE SUL GAS RILEVATO DAI VARI MODELLI
TS282EA ha un sensore in grado di rilevare Ammoniaca (NH
3
), che è un composto dell’Azoto, è un gas tossico e
infiammabile, incolore e con un odore molto pungente, più leggero dell’aria (ha densità relativa all'aria di 0,6). Il suo LIE
(Limite Inferiore di Esplosività) è 15%volume. Il TWA è 25 ppm (Time-Weighted Average) e lo STEL è 35 ppm (Short Term
Exposure Limit). L’NH
3
a 50÷100 ppm causa irritazione alle vie respiratorie e oltre i 100ppm irrita agli occhi, a 5000
ppm è letale. L’NH
3
è usata come: base per fertilizzanti agricoli, componente per vernici, refrigerante nell'industria
del freddo, solvente, sbiancante nell'industria cartaria e per la produzione di esplosivi, materie plastiche e polimeri.
TS282EA-H è identico al precedente, ma utilizza un sensore di livello superiore adatto a industrie e ambienti con
presenza d’inquinati e altri gas tossici.
TS282EC-S ha un sensore in grado di rilevare Monossido di carbonio (CO) ed è utilizzato in sistemi d’allarme
per parcheggi, industrie e ambienti da proteggere dalla possibile presenza di CO. Il CO è un gas tossico, inodore,
incolore, leggero come l'aria (ha densità, relativa all'aria di 0,97) è esplosivo ma a concentrazioni elevatissime, infatti, il
suo LIE (Limite Inferiore d’Esplosività) è 10,9%vol. Il TWA è 25 ppm (Time-Weighted Average) e lo STEL è 35 ppm (Short
Term Exposure Limit). Il CO è prodotto dalla combustione incompleta dei combustibili a base di carbonio. Questi com-
bustibili, gas, oli, legno e carbone, quando non bruciano correttamente, si forma un eccesso di CO che, diffonden-
dosi nell’ambiente è respirato dalle persone, dai polmoni raggiunge il sangue, dove impedisce il trasporto
dell’ossigeno ai tessuti e agli organi.
TS282EC-H è identico al precedente ma utilizza un sensore di livello superiore adatto a industrie e ambienti con
presenza d’inquinati e altri gas tossici.
TS282ES ha un sensore in grado di rilevare Biossido di zolfo (SO
2
) (sinonimi Anidride solforosa, Ossido di zolfo, Ossido
solforoso) è un gas tossico, incolore con un odore molto pungente, è più pesante dell’aria (ha densità, relativa all'aria di
2,25). Il suo TWA è 2 ppm (Time-Weighted Average) e lo STEL è 5 ppm (Short Term Exposure Limit).
L’SO
2
è un forte irritante per gli occhi e le vie respiratorie. Per inalazione può causare edema polmonare e una
prolungata esposizione può portare alla morte. Nel caso si combini con l’acqua forma Idrogeno Solforato (H
2
S).
TS282EH ha un sensore in grado di rilevare Idrogeno Solforato (H
2
S) (sinonimo Acido Solfidrico) che è un gas tossi-
co e infiammabile, incolore con un odore caratteristico di uova marce, più pesante dell’aria (ha densità, relativa all'aria
di 1,19). Il suo TWA è 5ppm (Time-Weighted Average) e lo STEL è 10ppm (Short Term Exposure Limit). Il suo LIE (Limite Infe-
riore di Esplosività) è 4%volume). L’odore dell’H
2
S nell’aria si percepisce a concentrazioni superiori a 0,2ppm, a bassi
livelli 50÷100ppm produce già irritazione agli occhi e alla gola, tosse, accelerazione del respiro e formazione di flui-
do nelle vie respiratorie. Oltre i 150ppm, il gas agisce sul nervo olfattivo rendendo impossibile la percezione del
suo odore, a concentrazioni più alte causa incoscienza in pochi minuti e può portare alla morte.
TS282EHCN ha un sensore in grado di rilevare Acido cianidrico (HCN) (sinonimi acido prussico, cianuro d’idrogeno) che
è un gas tossico, incolore con un odore di mandorle amare, più leggero dell’aria (ha densità, relativa all'aria di 0,94). Il
suo TWA è 4,7 ppm (Time-Weighted Average) e lo STEL è 10 ppm (Short Term Exposure Limit). L’HCN è estremamente
tossico, è irritante per gli occhi e il tratto respiratorio, se inalato provoca l'anossia, con perdita di coscienza e le
convulsioni. A 300 ppm, in soli 2 minuti è mortale per gli esseri umani. L'HCN è prodotto industrialmente impiegato
come reagente nella produzione di numerosi composti chimici, quali materie plastiche, coloranti, esplosivi, farmaci.
TS282EN ha un sensore in grado di rilevare Ossido d’Azoto (NO) che è un gas tossico, incolore, pesante circa
come l’aria (ha densità, relativa all'aria di 1,04). Il suo TWA (Time-Weighted Average) e lo STEL (Short Term Exposure Limit)
sono 25 ppm. L’NO è un forte ossidante e reagisce con materiali combustibili e riducenti. Al contatto con l'aria si
trasforma in biossido di azoto (NO
2
). L'NO è irritante per gli occhi e il tratto respiratorio. L'inalazione può causare
edema polmonare e formazione di metaemoglobina nel sangue. Nei casi più gravi può portare alla morte. L’NO è
un inquinante che si genera, nei processi di combustione; dalle emissioni naturali (eruzioni vulcaniche,) o da attività
civili e industriali che comportano processi di combustione (incendi, motori diesel, benzina, GPL, Caldaie ecc.).
TS282EN2 ha un sensore in grado di rilevare Biossido d’azoto (NO
2
) che è un gas tossico, d’odore pungente, di
colore rosso bruno, più pesante dell'aria (ha densità, relativa all'aria di 1,58). Il suo TWA è 3 ppm (Time-Weighted Average)
e lo STEL è 5 ppm (Short Term Exposure Limit). Combinandosi con l’acqua si trasforma in acido nitrico e ossido
d’azoto. L’NO
2
si forma per ossidazione del monossido d’azoto (NO), che si forma per combinazione dell'azoto e
dell'ossigeno alle alte temperature durante i processi di combustione (centrali termoelettriche, impianti di riscaldamento e
veicoli, soprattutto i motori diesel). L’NO
2
già a basse concentrazioni, provoca tosse acuta, dolori al torace, convulsioni e
insufficienza circolatoria. Può inoltre provocare danni irreversibili ai polmoni, edemi polmonari e decesso in quanto
formando metaemoglobina nel sangue impedisce il trasporto d’ossigeno ai tessuti.
IST-2282.EC01.01 Istruzione / User’s Manual / Manuel d’utilisation Pag. 5/17
TECNOCONTROL S.r.l. - Via Miglioli, 47 20090 SEGRATE (MI) ITALY Tel. +39 02 26922890 - Fax +39 02 2133734
TS282EHCL ha un sensore in grado di rilevare Acido cloridrico (HCL) (sinonimi Cloruro di idrogeno, Acido muriatico)
che è normalmente un liquido, che rilascia un gas tossico, incolore con un odore pungente, più pesante dell’aria
(ha densità, relativa all'aria di 1,3). Il suo TWA è 5ppm (Time-Weighted Average) e lo STEL è 10ppm (Short Term Exposure Li-
mit). In soluzione acquosa diventa uno dei liquidi più corrosivi esistenti. I vapori sono corrosivi per gli occhi, la cute
e il tratto respiratorio. L’inalazione di elevate concentrazioni del gas può causare polmoniti e edema polmonare.
TS282ECL ha un sensore in grado di rilevare Cloro (CL
2
) che è un gas verde giallastro, ha un odore soffocante
estremamente sgradevole, è molto velenoso e molto più pesante dell’aria (ha densità, relativa all'aria di 2,5). Il suo
TWA è 0,5ppm (Time-Weighted Average) e lo STEL è 1ppm (Short Term Exposure Limit). In soluzione acquosa diventa un
acido corrosivo. I vapori sono corrosivi per gli occhi, la cute e il tratto respiratorio. L’inalazione del gas può causare
polmoniti e edema polmonare. L'esposizione prolungata può portare alla morte.
FUNZIONAMENTO
I sensori elettrochimici utilizzati sono compensati in temperatura, ma sono sensibili alle variazioni d’umidità estre-
me. La taratura è eseguita con lo specifico gas per i differenti modelli come indicato a pag. 1, ma è in grado di rile-
vare anche altri gas, se presenti nello stesso locale come indicato in Tabella 2 e 3.
Preriscaldo: quando il trasmettitore è alimentato, inizia la fase di stabilizzazione del sensore, segnalata dal lam-
peggio del Led giallo "FAULT". Dopo circa 60 secondi, il Led giallo si spegne e si accende il Led verde ON”, che in-
dica il normale funzionamento. Dopo questo tempo il sensore è in grado di rilevare il gas, ma raggiunge le condi-
zioni di stabilità ottimali dopo circa 4 ore di funzionamento continuo.
Funzionamento Normale: deve essere acceso il solo Led Verde (ON).
Allarmi: Il Led rosso (ALARM) si accende, (solo se è stato abilitato con il Dip-Switch n.1 posizionato su ON) se la concentra-
zione di gas supera il valore indicato sotto la tabella, in funzione del modello.
TS282
EA
TS282EA-H
TS282
EC
-
TS282EC-H
TS282ES TS282EH TS282EHCN TS282EN TS282EN2 TS282EHCL TS282ECL
50 ppm NH
3
200 ppm CO
10 ppm SO
2
20 ppm H
2
S
5 ppm HCN
25 ppm NO
30 ppm NO
2
5 ppm HCL
1 ppm CL
I Guasti: possibili, sotto elencati, sono indicati dall’accensione del Led giallo (FAULT) e portando l’uscita “S” a 0mA.
Il Led giallo si accende ogni 4 secondi (con il Led Verde acceso): per avvisare che la Cartuccia Sensoreha supe-
rato il suo limite teorico di vita e non è più garantito il corretto funzionamento. Il rilevatore continua a funzionare
normalmente, ma è necessario, al ppresto, sostituire la Cartuccia Sensorecon una nuova, il tipo da richiedere è
indicato a pag. 1. La procedura di sostituzione è descritta nella documentazione ad essa allegata.
Se il Led giallo è acceso e il verde è spento (uscita 0mA): indica ppossibilità di guasto. 1) la configurazione
dei Dip-Switch non è corretta, verificarne la posizione (Vedi Tabella 4). 2) la Cartuccia Sensoreè guasta, sostituirla
con una nuova. 3) se è installata una Cartuccia nuova, o non è collegata correttamente o non è stata montata
quella compatibile. Controllare le connessioni con la cartuccia e la compatibilità (vedi a pag. 1). Eseguite le verifiche
spegnere e riaccendere l’apparecchio. Se la condizione persiste, sarà necessario sostituire e/o inviare il rilevatore
al fornitore per la riparazione.
Se il Led giallo e il verde sono accesi (uscita 0mA): indica probabilmente il guasto della Cartuccia Sensore”. Pri-
ma, provare a eseguire la Regolazione dello ZERO come descritto nella sezione Verifiche e Calibrazione, poi spe-
gnere e riaccendere l’apparecchio e infine provare a sostituire la “Cartuccia Sensore”. Se la condizione persiste, sarà
necessario sostituire e/o inviare il rilevatore al fornitore per la riparazione.
Se è stato abilitato il Led rosso e quindi tutti i Led sono accesi (uscita >20mA): indica, o guasto della “Cartuccia
Sensore”, o una concentrazione di gas superiore al Fondo Scala. Se non è presente alcun gas e la condizione per-
siste, anche dopo la sostituzione della “Cartuccia” sarà necessario inviare il rilevatore al fornitore per la riparazione.
INSTALLAZIONE
I rilevatori vanno installati, posizionati ed eseguite le manutenzioni seguendo tutte le norme nazionali vigenti per
gli impianti elettrici e le norme di sicurezza degli impianti.
Montaggio: in Fig.1 sono indicate le dimensioni. Installare il TS282E verticale con il sensore rivolto verso il basso.
Posizione dei TS282EA - TS282EA-H: vanno fissati a circa 30÷50 cm dal soffitto (il gas NH
3
è più leggero dell’aria).
Posizione dei TS282EC-S-TS282EC-H: vanno fissati a circa 1,60 cm dal pavimento (il gas CO è circa come l’aria).
Posizione del TS282ES: va fissato a circa 30÷50 cm dal pavimento (il gas SO
2
è più pesante dell’aria).
Posizione del TS282EH: va fissato a circa 30÷50 cm dal pavimento (il gas H
2
S è più pesante dell’aria).
Posizione del TS282EHCN: va fissato a circa 1,60 m dal pavimento per la protezione delle persone o sopra il
probabile punto di emissione (il gas HCN è poco più leggero dell'aria).
Posizione del TS282EN: va fissato a circa 1,60 m dal pavimento (il gas NO è leggero circa come dell’aria).
Posizione del TS282EN2: va fissato a circa 30÷50 centimetri dal pavimento (il gas NO
2
è più pesante dell’aria).
Posizione del TS282EHCL: va fissato a circa 30÷50 centimetri dal pavimento (il gas HCL è più pesante dell’aria).
Posizione del TS282ECL: va fissato a circa 30÷50 centimetri dal pavimento (il gas CL
2
è molto ppesante dell’aria).
Collegamenti elettrici (Fig.2): Va utilizzato un cavo schermato a 3 conduttori. La calza va collegata a massa solo
dal lato Centrale. La distanza massima, cui può essere installato ogni singolo rivelatore dall'alimentatore, è indicata
nella Tabella 1 in funzione della sezione del cavo utilizzato e della Centrale Gas utilizzata.
Il morsetto, (+ - S) è a innesto e polarizzato, è necessario sfilarlo per effettuare i collegamenti.
NOTA: Il Dip-Switch va posizionato prima d’alimentare l’apparecchio. Se si utilizza l’apparecchio con il
Dip-Switch 1 su ON, sarà attivato il funzionamento del Led rosso d’allarme (vedi Tabella 1).
IST-2282.EC01.01 Istruzione / User’s Manual / Manuel d’utilisation Pag. 6/17
TECNOCONTROL S.r.l. - Via Miglioli, 47 20090 SEGRATE (MI) ITALY Tel. +39 02 26922890 - Fax +39 02 2133734
Importante: terminata l’installazione, alimentare l’apparecchio, attendere circa 30÷60 minuti e poi per adattare il
sensore alle condizioni ambientali, eseguire, se necessario, la “Regolazione dello Zero (vedi Verifiche e Calibrazione).
COMPATIBILITÀ CON ALTRE CENTRALI: Se non si usa una Centrale Gas Tecnocontrol, calcolare la resistenza
di carico massima utilizzando il grafico in Fig.4. La resistenza di carico massima con alimentazione 12Vcc è 50
ohm, mentre con 24Vcc è 400 ohm. Si consiglia di utilizzare un’alimentazione a 24Vcc.
AVVERTENZE
La vita utile del sensore in aria pulita è mediamente di 2 anni. (Vedi a pag.1). Al termine di questo periodo, indi-
cato dallo strumento con un lampeggio del Led Giallo ogni 4 secondi, è necessario sostituire la “Cartuccia Sensore”.
Verifiche Periodiche: si consiglia eseguire almeno ogni 6/12 mesi la verifica di funzionamento, Test Elettrico, Rego-
lazione dello Zero e Verifica e Calibrazione con la specifica miscela di gas titolato, vedi Verifiche e Calibrazione.
Nota: Il rilevatore non è in grado di rivelare perdite che avvengono fuori del locale in cui è installato o all'interno
dei muri o sotto il pavimento.
Limiti di funzionamento: Temperature inferiori a -40°C e superiori a +60°C provocano la rottura del Sensore.
Può danneggiarsi, anche se è esposto a concentrazioni oltre i limiti indicati sotto in tabella, in funzione del modello. In
questi casi il sensore deve essere sostituito.
TS282
EA
TS282EA-H
TS282
EC
-
TS282EC-H
TS282ES TS282EH TS282EHCN TS282EN TS282EN2 TS282EHCL TS282ECL
500ppm NH
3
1000ppm CO
150ppm SO
2
500ppm
H
2
S
100ppm EHN
1000ppm NO
150ppm NO
2
>30ppm HCL
>50ppm CL
2
Attenzione: i sensori elettrochimici contengono una piccola quantità d’acido. Nel caso avvenga una perdita dal
sensore, (causata da un'azione meccanica violenta o dall’uso oltre i limiti di funzionamento) evitare di toccare il li-
quido fuoriuscito. Nel caso di contatto con la pelle o gli occhi lavare immediatamente con abbondante acqua.
Interferenze con altri gas: l’apparecchio rileva gas, in ambiente, a una concentrazione molto bassa. Le celle elet-
trochimiche hanno una buona resistenza ai prodotti d'uso comune: spray, detersivi, colle o vernici. Questi possono
contenere sostanze che, in quantità elevata, potrebbero interferire con il sensore; si consiglia di ventilare il locale,
quando si utilizzano questi prodotti. Considerare che in ambienti particolarmente inquinati o con vapori d’altre so-
stanze in quantità molto elevate (in particolare solventi organici, idrogeno, alcoli), possono provocare falsi allarmi e
il sensore deve poi rimanere molto tempo in aria pulita per tornare a funzionamento normale (1-2 ore).
VERIFICHE E CALIBRAZIONE
NOTA IMPORTANTE: le seguenti operazioni vanno eseguite da personale esperto e addestrato, perché
l’uscita in mA cambiando valore, attiva i dispositivi d’allarme della Centrale cui è collegata.
Test Elettrico, Regolazione dello Zero e Calibrazione: per accedere a queste tre funzioni è necessario inserire
il relativo Codicetramite i pulsanti F1 e F2. Per far che la pressione su un pulsante sia riconosciuta, tenerlo
premuto per circa un secondo (finche non si spegne per un attimo il Led Verde). Dopodiché si può passare al pul-
sante successivo. In caso d’errore basta aspettare circa 10 secondi e la sequenza è automaticamente cancellata.
Per la Verifica non è necessario alcun codice.
Kit di Taratura TC014 e Bombole con Miscela Gas titolato (per Calibrazione e Verifica):
La miscela da utilizzare in funzione del modello installato è:
MODELLO
Bombola di Gas Titolato Bombola/Valvola Codice KIT
TS282EA
-
TS282EA
-
H
100 ppm
Ammoniaca (NH
3
)
in azoto
BO501
/ BO303
TC014
TS282EC
-
S
-
TS282EC
-
H
300 ppm
Monossido di carbonio (CO)
in azoto
BO210
/ compresa
TC011/TC014
TS282ES
10 ppm B
iossido di zolfo (SO
2
)
in azoto
BO518
/ BO303
TC014
TS282EH
50 ppm Idrogeno solforato (H
2
S) in azoto
BO520
/ BO303
TC014
TS282EHCN
10 ppm Acido cianidrico (HCN) in azoto
BO529
/ BO303
TC014
TS282EN
100 ppm Ossido d’azoto (NO) in azoto
BO522
/ BO303
TC014
TS282EN2
30ppm Biossido d’azoto (NO
2
) in aria Fornibile a richiesta TC014
TS282EHCL
10ppm Acido cloridrico (HCL) in azoto
BO527
/ BO305
TC014
TS282ECL
10ppm Biossido d’azoto (NO
2
) in aria Fornibile a richiesta TC014
È possibile usare le bombole monouso, con valvola d’erogazione, sia le ricaricabili ad alta pressione con ridutto-
re. Alcune bombole hanno riduttori a flusso fisso, verificare il valore necessario in Tabella, colonna 3 a pag. 7. Inoltre è
necessario usare il KIT di Calibrazione Tecnocontrol mod. TC014 (Inox). Per garantire che il gas raggiunga il sen-
sore, usare riduttori d’acciaio, tubi in Teflon o altri materiali compatibili che non lo disperdano.
“TEST ELETTRICO” (Codice Test: F2, F2, F1, F1): questa funzione permette di eseguire un test funzionale
dell’apparecchiatura. Dopo aver messo l’impianto in sicurezza e aver inserito il Codice Test”, si spengono tutti Led.
Quindi si accenderanno in sequenza i Led, dal giallo fino al rosso. L’uscita 4÷20mA rimane invariata. Alla fine, tutti i
Led rimarranno accesi per circa 5 secondi, poi la centralina tornerà nelle condizioni di funzionamento normale. È
consigliabile eseguire quest’operazione ogni 6-12 mesi in base all’utilizzo.
Nota: Non è possibile eseguire l’operazione se è acceso il Led Rosso.
“REGOLAZIONE DELLO ZERO” (Codice di Zero: F2, F1, F1, F2): permette di regolare manualmente il sensore a
Zero e va eseguita esclusivamente in aria pulita (ambiente senza la presenza di gas tossici o altri inquinanti).
Considerare che i rilevatori, hanno anche un sofisticato inseguitore di zero, che ogni ora azzera il sensore, se na-
turalmente non c’è presenza di gas. Se necessario, eseguire quest’operazione dopo l’installazione o dopo il cam-
bio della cartuccia o ogni 6-12 mesi in base alle condizioni ambientali. Dopo aver inserito il Codice di Zero”, come
conferma dell’avvenuta operazione, ci sarà 1 lampeggio del LED rosso e l’uscita diventerà 4,0 mA.
IST-2282.EC01.01 Istruzione / User’s Manual / Manuel d’utilisation Pag. 7/17
TECNOCONTROL S.r.l. - Via Miglioli, 47 20090 SEGRATE (MI) ITALY Tel. +39 02 26922890 - Fax +39 02 2133734
AVVERTENZA: Non è possibile eseguire l’operazione se l’uscita in mA è superiore al valore indicato sotto in ta-
bella, in funzione del modello, oppure se è acceso il Led rosso. In questo caso è necessario procedere alla Cali-
brazione” oppure sostituire la “Cartuccia”.
TS282
EA
TS282EA-H
TS282
EC
-
TS282EC-H
TS282ES TS282EH TS282EHCN TS282EN TS282EN2 TS282EHCL TS282ECL
4,8 mA
15ppm NH
3
4,8 mA
15ppm CO
6,4 mA
3ppm SO
2
4,8 mA
5ppm H
2
S
8 mA
2,5ppm HCN
5,6 mA
10ppm NO
6,6 mA
5ppm NO
2
8 mA
2,5ppm HCL
4,8 mA
0,5ppm HCL
“CALIBRAZIONE” (Codice: F2, F2, F2, F1, F2, F1): serve per ritarare il sensore con la miscela di gas sopra indicata.
AVVISO: per garantire la sicurezza che non avvengano errori d’elaborazione, esiste la rara possibilità che durante
la Calibrazione, il Led Giallo si spenga ogni 8 secondi, in questo caso interrompere la procedura, spegnere e riac-
cendere l’apparecchio. Ripetere la Calibrazione, se la condizione persiste, sarà necessario inviare il rilevatore al
fornitore per la riparazione. Attenzione: Durante la Calibrazione, l’uscita in mA diventerà 0mA.
La Calibrazioneva eseguita solo in aria pulita (ambiente senza la presenza di gas tossici o altri inquinanti). Con i tasti ese-
guire il Codice Calibrazione”. Attendere finché si accendano fissi il Led giallo e verde e il Led rosso cominci a lam-
peggiare. Infilare il TC014 sul portasensore, regolare l’afflusso del gas, in modo che il flussometro indichi circa il
valore indicato in colonna 3 (vedi sotto tabella e Fig.3). Attendere (circa 3÷5 minuti) fino quando il Led rosso si accende
fisso (e mentre è ACCESO), premere il tasto F2 e tenerlo premuto finché il Led rosso non rimane spento per almeno 2
secondi, se invece il Led rosso continua a lampeggiare oltre 5-6 minuti, significa che il flusso di gas non è costante
o insufficiente o la concentrazione non è quella richiesta. Interrompere l’operazione, spegnere lo strumento e ripe-
tere la calibrazione. Chiudere la bombola e rimuovere il TC014. A questo punto si possono verificare due casi:
LED giallo e verde accesi: la calibrazione è corretta, dopo 8 secondi il rilevatore si riavvia automaticamente in
funzionamento normale (vedi capitolo FUNZIONAMENTO “ Preriscaldo).
LED giallo acceso: la calibrazione è fallita dopo 8 secondi il rilevatore si riavvia automaticamente e dopo il preri-
scaldo, ripetere la Calibrazione senza reinserire il Codice”. Se la condizione persiste anche dopo la sostituzione
della “Cartuccia Sensore”, inviare il rilevatore al fornitore per la riparazione.
“VERIFICA” (non serve codice): serve per controllare, utilizzando la miscela di gas sopra indicata, la corretta rispo-
sta al gas e può essere eseguita dopo la Calibrazione o dopo l’installazione, ma va eseguita soprattutto durante le
manutenzioni periodiche, perché è l’unico metodo per controllare l’effettivo funzionamento del rilevatore.
Infilare il TC014 sul portasensore, in funzione del modello, regolare il riduttore della bombola in modo che il flus-
sometro indichi il valore indicato in colonna 3 (vedi Fig.3). Controllare che il multimetro, con i puntali sui Test-Point
TEST mA (vedi Fig.2) raggiunga in 3÷5 minuti circa il valore in mV, indicato sotto in colonna 3 della tabella. [ovvero che
l’uscita in mA aumenti fino al valore indicato in colonna 4 e la centrale, cui è collegato il rilevatore, indichi circa il valore indicato in colonna
5]. Se il valore fosse diverso, eseguire la Calibrazione”. Terminata la Verifica”, chiudere la bombola e togliere il
TC014. Attendere 3-5 minuti per tornare alle condizioni di funzionamento normale.
1-Modello 2-Gas da utilizzare 3-Flussomentro
4-TEST mA 5-Uscita in mA
6-Indicazione in ppm
TS282
EA
100ppm Ammoniaca (NH
3
) in azoto
0,3
l/min
90÷96 mV 9,0÷9,6 mA 100 ppm 5)
TS282
EA
-
H
1 l/min
TS282
EC
-
S
300ppm Monossido di carbonio(CO) in azoto
0,3 l/min 190÷210 mV
19÷21 mA 300 ppm (±20)
TS282
EC
-
H
TS282
ES
10ppm Biossido di zolfo (SO
2
) in azoto
0,3 l/min
118÷122 mV
11,7÷12,3 mA
10.0
ppm (±0.5
)
TS282
EH
50ppm Idrogeno solforato (H
2
S) in azoto
0,3 l/min
117÷123 mV
11,7÷12,3 mA
50
ppm (±2)
TS282
EHCN
10ppm Acido cianidrico (HCN) in azoto
0,3÷0,5
l/min
192÷208 mV
19,2÷20,8 mA
10
.0
ppm (±1
)
TS282
EN
100ppm Ossido d’azoto (NO) in azoto
0,3 l/min
192÷208 mV
19,2÷20,8 mA
100
ppm (±5)
TS282
EN2
30ppm Biossido d’azoto (NO
2
) in aria
1 l/min
190÷210 mV
19÷21 mA
30
ppm (±2)
TS282
EHCL
10ppm Acido cloridrico (HCL) in azoto
0,3÷0,5 l/min
192÷208 mV
19,2÷20,8 mA
10
ppm (±1)
TS282
E
C
L
10ppm Biossido d’azoto (NO
2
) in aria
1 l/min
19
2
÷20
8
mV
19
,2÷20,8
mA
1
0
.0
ppm (±
0,5
)
E
N
DESCRIPTION
The TS282E series are gas detectors able to detect toxic gases by employing an electrochemical sensor calibrated
to different gases as listed on page 1 and find their best application in centralized alarm systems for car parks, la-
boratories, manufacturing industries and environments to be protected from possible leakage of gas.
The instruments comprise of a thermoplastic enclosure in which the electronic circuit and the terminals are mount-
ed and a downward facing cylindrical sensor housing with inside a “Replaceable Cartridge Sensor”.
The TS282E has a 4÷20mA linear output (S) with F.S. of the detected gas, expressed in ppm (part per million). This
output is connectable to a remote Gas Central Unit as listed in Table 1. On the front panel, F1 and F2 key using for
Test and Calibration routine protected by a code and 3 Led shows the working conditions:
Red Led "ALARM": alarm indication
Green Led "ON": normal working condition
Yellow Led "FAULT": the sensor should be faulty, disconnected, out of scale or expired.
NOTES ON THE DETECTED GAS
TS282EA has a sensor that can detect Ammonia (NH
3
) gas which is a toxic, colourless and with a characteristic
pungent smell, lighter than air, (its density as to air is 0.6). His TWA is 25ppm (Time-Weighted Average) and STEL is
35ppm (Short Term Exposure Limit). His LEL (Lower Explosivity Limit) is 15% v/v (volume). This gas is irritating to the eyes
IST-2282.EC01.01 Istruzione / User’s Manual / Manuel d’utilisation Pag. 8/17
TECNOCONTROL S.r.l. - Via Miglioli, 47 20090 SEGRATE (MI) ITALY Tel. +39 02 26922890 - Fax +39 02 2133734
and mucous membranes (respiratory and digestive tracts). Ammonia is used in industrial refrigeration applications.
It is also used as fertilizers (either as its salts), as a component in paints, for the production of plastics and poly-
mers, such as solvent or as household cleaners.
TS282EA-H model is as previous, but it uses a sensor suitable for industry and polluted environments and with
presence of other toxic gases.
TS282EC-S has a sensor that can detect Carbon monoxide (CO) gas which is a poisonous, colourless, odour-
less, and tasteless. CO is a little lighter than air (its density as to air is 0.97) and his LEL (Lower Explosive Limit) is
10.9%v/v (%volume). His TWA is 25 ppm (Time-Weighted Average) and STEL is 35 ppm (Short Term Exposure Limit).
CO is produced by incomplete burning of carbon-based fuels, including gas, oil, wood and coal. Carbon-based
fuels are safe to use. It is only when the fuel does not burn properly that excess CO is produced, which is poison-
ous. When CO enters the body, it prevents the blood from bringing oxygen to cells, tissues, and organs.
TS282EC-H model is as previous, but it uses a sensor suitable for industry and polluted environments and with
presence of other toxic gases.
TS282ES has a sensor that can detect Sulphur dioxide (SO
2
) (also Sulphurous anhydride, Sulphurous oxide, Sulphur oxide)
gas which is a toxic, colourless, with pungent odour, heavier than air (its density as to air is 2.25). His TWA (Time-
Weighted Average) is 2 ppm and STEL (Short Term Exposure Limit) is 5 ppm. The SO
2
is a strong irritant to eyes and res-
piratory system. Inhalation may cause pulmonary edema and prolonged exposure can lead to death and when it
combines with water forms H
2
S (Hydrogen sulphide).
TS282EH has a sensor that can detect Hydrogen sulphide (H
2
S) (synonymous with hydrogen sulphide) gas which is a
toxic, flammable, colourless, with a characteristic odour of rotten eggs, heavier than air (its density as to air is 1.19).
His TWA is 5ppm (Time-Weighted Average) and STEL is 10 ppm (Short Term Exposure Limit). His LEL (Lower Explosive Limit)
is 4% volume. The H
2
S smell is perceived at concentrations above 0.2 ppm, to low levels (50÷100 ppm) already
produces irritation to the eyes and throat, cough, fast breathing and training of fluid in the respiratory tract. This
gas, over 150 ppm, acts on the olfactory nerve, making it impossible for the perception of its odour, and higher
concentrations can cause unconsciousness within a few minutes and can lead to death.
TS282EHCN has a sensor that can detect Hydrogen cyanide (HCN) (synonymous with Hydrocyanic acid, Prussic acid)
gas which is a toxic, extremely poisonous gas, colourless, with has an almond-like odour, a little lighter than air
(its density as to air is 0.94). His TWA (Time-Weighted Average) is 4.7 ppm and STEL (Short Term Exposure Limit) is 10 ppm.
The HCN is a strong irritant to eyes and respiratory system. Inhalation may cause effects on the cellular respira-
tion, resulting in convulsions and unconsciousness. Exposure to 300ppm for 2 minutes may result in death.
TS282EN has a sensor that can detect Nitric oxide (NO) gas which is a toxic and colourless, heavy as air (its den-
sity as to air is 1.04). His TWA (Time-Weighted Average) and STEL (Short Term Exposure Limit) are 25 ppm. NO is a strong
oxidant and reacts with combustible and reducing materials. On contact with the air turns it into nitrogen dioxide
(NO
2
). Nitric oxide is irritating to eyes and respiratory tract. Inhalation can cause pulmonary edema, may also have
effects on the blood, causing the formation of methaemoglobin. In more severe cases it can lead to death. Nitric ox-
ide is a pollutant that is generated in part in combustion processes, in part by emissions from natural volcanic erup-
tions, fires or civil activities and industrial processes involving combustion, such as transport (vehicles with diesel en-
gine, gasoline, LPG, etc.) and the production of heat and electricity.
TS282EN2 has a sensor that can detect Nitrogen dioxide (NO
2
) gas which is a reddish-brown toxic gas with a
characteristic sharp, biting odour, heavier than air (its density has to air is 1.58). His TWA (Time-Weighted Average) is 3
ppm and STEL (Short Term Exposure Limit) is 5 ppm. NO
2
combined with water turns into nitric acid and nitric oxide.
NO
2
is formed in most combustion processes using air as the oxidant. At elevated temperatures nitrogen combines
with oxygen to form nitrogen dioxide. The most important sources of NO
2
are internal combustion engines (power
plants, heating and vehicles, especially diesel engines). NO
2
is toxic by inhalation, it can cause irreversible damage to the
lungs, pulmonary edema and death as it is capable of combining with the haemoglobin by changing the chemical
and physiological, forming methaemoglobin which is no longer able to carry oxygen to the tissues. Also, low con-
centrations (4 ppm) will anesthetize the nose.
TS282EHCL has a sensor that can detect Hydrogen chloride (HCL) (synonymous with Hydrogen chloride, muriatic acid)
which is normally a liquid, which releases a toxic gas, colourless with a pungent odour, heavier than air (its density
has to air is 1.3). His TWA (Time-Weighted Average) is 5 ppm and STEL (Short Term Exposure Limit) is 10 ppm.
This gas, in aqueous solution becomes one of the most corrosive liquids. The vapours are corrosive to eyes, the
skin and the respiratory tract. Inhalation of high gas concentrations may cause pneumonitis and pulmonary edema.
TS282ECL has a sensor that can detect Chlorine (CL
2
) which is a yellow-green gas that has a distinctive strong
odour, is extremely poisonous and heavier than air (its density has to air is 2.5). His TWA (Time-Weighted Average) is 0.5
ppm and STEL (Short Term Exposure Limit) is 1 ppm. This gas, in aqueous solution becomes a corrosive acid. The
vapours are corrosive to eyes, skin and the respiratory tract. Inhalation of the gas may cause pneumonitis and
pulmonary edema. Prolonged exposure can cause death.
OPERATIONAL DESCRIPTION
The electrochemical sensor is temperature compensated, but is sensitive to extreme humidity variations. The cali-
bration is carried out with specific gas to be detected. Anyway, it can contemporaneously detect other gases that
should be present in the same environment as listed in tables 2 and 3.
Preheating: when powered, the sensor needs a time of preliminary heating of about 60 seconds. During this peri-
od the yellow Led FAULTflashes. After this period, the yellow Led light off, the green Led ONilluminates to indi-
cate normal functioning. After this period the unit is able to detect gas even if it attains the optimum stability condi-
tions after about 4 hours continual functioning.
Normal operation: the green Led “ON” should be light on.
IST-2282.EC01.01 Istruzione / User’s Manual / Manuel d’utilisation Pag. 9/17
TECNOCONTROL S.r.l. - Via Miglioli, 47 20090 SEGRATE (MI) ITALY Tel. +39 02 26922890 - Fax +39 02 2133734
Alarm: The red Led (ALARM) illuminates (only if it be activate by Dip-Switch no.1 set to ON) if the gas concentration ex-
ceeds the value indicated in the table below, depending on the model.
TS282
EA
TS282EA-H
TS282
EC
-
TS282EC-H
TS282ES TS282EH TS282EHCN TS282EN TS282EN2 TS282EHCL TS282ECL
50
ppm NH
3
200 ppm CO
10 ppm SO
2
20 ppm H
2
S
5 ppm HCN
25 ppm NO
30 ppm NO
2
5 ppm HCL
1 ppm CL
Faults: the yellow Led illuminates and the "S" output falls down to 0mA. (The different faults are listed below).
Yellow Led illuminates each 4 seconds (with green Led activate): this happens when the Cartridge Sensor has
overcome its theoretical period of life and its correct operation is no longer guaranteed. The detector keeps on op-
erating normally but it is necessary to replace, as soon as possible, the Cartridgewith a new one. The type to be
required is described on Page 1. The replacement procedure is described in the attached manual.
Yellow Led activate, green Led off (0 mA output signal): this signal different kind of faults. 1) The Dip-Switch set
up is wrong, please verify (see Table 4). 2) The Cartridgeis not working, please replace with new one. 3) If a new
Cartridgeis installed or it is not correctly connected or a not compatible one is mounted. Please check the car-
tridge connections and compatibility (see on page 1) these checks are made switching off and on the device. If the
condition does not change, it will be necessary to replace the unit and/or send it back to the supplier to repair.
Yellow and green Led activates (0 mA output signal): this happens when the Cartridgeis not working. First try to
perform the procedure of ZEROas described in the section Test and Calibration > Zero adjust then disconnect and
connect the unit, finally try to replace a new Cartridge”. If the condition is not change, please replace the unit and/or
send it back to the supplier to repair.
All Led activates only if reed Led be activate by Dip-Switch no.1 set to ON (>20 mA output signal): this happens when the
Cartridgeis not working or gas concentration is higher than F.S. If there are not any gas leaks and the condition is
not change, it will be necessary to replace the unit and/or send it back to the supplier to repair.
INSTALLATION
The detector must be accurately installed and testing according to the national dispositions in force on the safety of
the plants and installation of electric devices.
Mounting: The Fig. 2 shows the instrument size. The unit must be positioned vertically with the sensor downwards.
TS282EA and TS282EA-H positioning: should be fixed at 30÷50 cm from the ceiling (NH
3
gas is lighter than air).
TS282EC-S and TS282EC-H positioning: should be fixed at 1.60 m from the floor (CO gas is a little lighter than air).
TS282ES positioning: should be fixed at 30÷50 cm from the floor (SO
2
gas is heavier than air)
TS282EH positioning: should be fixed at 30-50 cm from the floor (H
2
S gas is heavier than air).
TS282EHCN positioning:
should be fixed at 1.60 m from the floor for the protection of persons or over the likely
point of emission. (
HCN
gas is a little lighter than air).
TS282EN positioning: should be fixed at 1.60 m from the floor (NO gas is heavy as air).
TS282EN2 positioning: should be fixed at 30÷50 cm from the floor (NO
2
gas is heavier than air).
TS282EHCL positioning: should be fixed at 30÷50 cm from the floor (HCL gas is heavier than air).
TS282ECL positioning: should be fixed at 30÷50 cm from the floor (CL
2
gas is much heavier than air).
Electrical Connection (see Fig.2): the maximum distance to install each detector from the Gas Central Unit is
show in the Table 1. Normally use a tree wire shielded cable.
The terminals (+ - S), are polarized plug-in type, it is necessary to extract them to make the connection.
Note: Dip-Switch should be set with instrument powered off. Dip-Switch 1 settled ON activates ALARM red
Led indication (see Table 2).
Important: Once installation is completed, power up the unit, wait about 30 to 60 minutes and then to adjust the
sensor to the environment, only if it is necessary, carry out the Zero Adjust (see Test and Calibration).
COMPATIBILITY WITH OTHER CENTRAL UNITS: In case of a central unit other than Tecnocontrol, please
verify the max load resistor as shown in Fig. 4. The max load resistor is 50 ohm with 12Vdc power supply, while is
400 ohm with 24Vdc power supply. We suggest using 24Vdc power supply.
WARNING
Average life: In clean air and in normal working condition the sensor's life is about 2 years. After this period the
yellow Led “FAULT” flashes every 4 seconds, is necessary replacing the “Cartridge Sensor”.
Periodical testing: we advise to carry out working tests every 6 or 12 months. Tests, Zero Adjust and Calibration
with sample gas bottle, as explained in the next chapter “Tests and Calibration”.
Note: the detector is not able to detect gas leaks occurring outside the room where it is installed, either inside
walls or under the floor.
Sensor operating limits: the electrochemical cell, in any case, should not be exposed to temperatures lower than
-40°C or higher than 60°C. The electrochemical cell can be damaged, if it is exposed to concentrations over the
limit specified in the table, depending on the model. If this condition occurs, please replace the “Cartridge Sensor.
TS282
EA
TS282EA-H
TS282
EC
-
TS282EC-H
TS282ES TS282EH TS282EHCN TS282EN TS282EN2 TS282EHCL TS282ECL
500ppm NH
3
1000ppm CO
150ppm SO
2
500ppm
H
2
S
100ppm EHN
1000ppm NO
150ppm NO
2
>30ppm HCL
>50ppm CL
2
Safety: electrochemical Cell contains a small quantity of acid. It is possible that either a very strong mechanical
action or an incorrect instrument use beyond the limits of its operational functioning could provoke a liquid leak out
of the sensor. Should this event occurs, please avoid touching the come out liquid. Should a contact with either the
skin or eyes occur, wash immediately and abundantly with water.
IST-2282.EC01.01 Istruzione / User’s Manual / Manuel d’utilisation Pag. 10/17
TECNOCONTROL S.r.l. - Via Miglioli, 47 20090 SEGRATE (MI) ITALY Tel. +39 02 26922890 - Fax +39 02 2133734
Cross sensitivity to other gases: this detector is designed to detect gas leaks at a very low concentration. The
electrochemical cells employed have a good resistance towards products such as sprays, detergents, glues and
paints. However, in case of products containing substances in great quantity, these could interfere with the Sensor
and cause false alarms. We recommend ventilating the room when products like these are used. Alcohols, Hydro-
gen and volatile organic solvents in great quantity cause false alarms, and then the cell needs one or more hours in
clean air to recover the normal sensitivity conditions.
TEST and CALIBRATION
PAY ATTENTION: This procedure has to be made with extreme attention and by authorized and trained
people; because starting this procedure it will increase mA output causing the activation of connected
alarm devices.
Operation Check, Zero Adjust, and Calibration: are different code protected functions. To access these func-
tions is necessary to insert the relevant Code" through the keys F1s and F2. To have the key pressure recognized,
hold pressing it for around a second (until the green Led doesn't switch off for a moment). Then the next key can be
pressed. In case of error all it takes is waiting around 10 seconds and the sequence is automatically erased.
The Calibration Check is a function code free.
Calibration Kit, Sample Gas Bottles (for Calibration Check and Calibration) the mixture to be used depending on the
model installed is shown in the table below:
MODEL
Sample gas bottle
Our Code / Valve KIT Code
TS282EA
-
TS282EA
-
H
100 ppm Ammonia (NH
3
) in Nitrogen
BO501
/ BO303
TC014
TS282EC
-
S
-
TS282EC
-
H
300 ppm Carbon monoxide (CO) in Nitrogen
BO210
/ included
TC011/TC014
TS282ES
10 ppm Sulphur dioxide (SO
2
) in Nitrogen
BO518
/ BO303
TC014
TS282EH
50 ppm Hydrogen sulphide (H
2
S) in Nitrogen
BO520
/ BO303
TC014
TS282EHCN
10 ppm Hydrogen cyanide (HCN) in Nitrogen
BO529
/ BO303
TC014
TS282EN
100 ppm Nitric oxide (NO) in Nitrogen
BO522
/ BO303
TC014
TS282EN2
30ppm Nitrogen dioxide (NO
2
) in air Available on request TC014
TS282EHCL
10ppm Hydrogen cloride (HCL) in Nitrogen
BO527
/ BO305
TC014
TS282ECL
10ppm Nitrogen dioxide (NO
2
) in air Available on request TC014
It is possible to use either the disposable cylinders with adjust valve or the high pressure ones with reduction
gear. Some cylinders have fixed flow reducers, make the necessary value in Table, column 3, page 11. Is also neces-
sary the Tecnocontrol calibration kit model TC014 (Stainless steel). To ensure that the gas concentration reach-
es sensor, use steel reduction gear and Teflon tubes or other compatible materials which do not disperse gas.
“INSTRUMENT OPERATION CHECK” (Check Code: F2, F2, F1, F1): this function allows to effect a functional test
of the detector. After having inserted the "Check Code", all Led are switched off. Then Led will switch on in se-
quence, from the yellow up to the red. At the end all the Led will remain lighted for around 5 seconds, then the de-
tector returns at the conditions of normal operation. It is advisable to perform this operation every 12 months ac-
cording to the use. Note: this function is not working if the red LED is already switched on.
“ZERO ADJUST” (Zero Code: F2, F1, F1, F2): this function is to adjust the Zero sensor and can be done in clean
air only (environment without the presence of toxic gas or other pollutants). Consider that the TS282E series have a sophisti-
cated follower of zero, that every hour will reset the sensor, if of course there is the presence of gas. If necessary,
perform this operation after installation or after changing the Cartridgeor every 6-12 months depending on envi-
ronmental conditions. After entering the "Zero Code" as confirmation of the operation, there will be a flashing red Led
and output becomes 4.0 mA.
Note: This function is not working if mA output is higher than the value shown in the table below, depending on the
model, or if the red Led is already ON. In this case, please recalibrate the unit and/or replacing the “Cartridge”.
TS282
EA
TS282EA-H
TS282
EC
-
TS282EC-H
TS282ES TS282EH TS282EHCN TS282EN TS282EN2 TS282EHCL TS282ECL
4,8 mA
15ppm NH
3
4,8 mA
15ppm CO
6,4 mA
3ppm SO
2
4,8 mA
5ppm H
2
S
8 mA
2,5ppm HCN
5,6 mA
10ppm NO
6,6 mA
5ppm NO
2
8 mA
2,5ppm HCL
4,8 mA
0,5ppm HCL
“CALIBRATION” (Calibration Code: F2, F2, F2, F1, F2, F1): this function allows completely recalibrate the sensor
using the gas mixture indicated above.
Warning: to guarantee that no errors of elaboration happen, the possibility exists that during the Calibration the yel-
low Led switch off every 8 seconds, in this case interrupt the procedure, switch off and switch on the instrument and
repeat the Calibration. If condition persists it will be necessary to send the detector to the supplier for the reparation.
Important note: during Calibration routine the mA output indicates 0mA.
The “Calibration” can be done in clean air only (environment without the presence of toxic or other polluting gas). With the keys
perform the Calibration Code". Wait until the yellow and green Led switch on fix and the red Led starts to flash. In-
sert the TC014 on the sensor holder and adjust the influx of the gas the way that the flow meter reaches the value
indicates in table below in column 3, depending on the model (see also Fig.3). Wait around 3÷5 minutes, until the
red Led lights (and while it is ON), press the key F2 and hold it pressed until the red LED is switched off for at least 2
seconds (but if the red LED continues to blink more than 5-6 minutes means that the gas flow is not constant or
unsatisfactory, or the concentration is not as required, stop the operation, turn off the instrument and recalibrate).
Then, close the gas cylinder and remove TC014. At this point we can have two possibilities:
Yellow and green Led illuminates: the calibration routine is correct. Wait 8 seconds, until the instrument auto-
matically restores the normal working conditions. (See “Operational Description > Preheating).
IST-2282.EC01.01 Istruzione / User’s Manual / Manuel d’utilisation Pag. 11/17
TECNOCONTROL S.r.l. - Via Miglioli, 47 20090 SEGRATE (MI) ITALY Tel. +39 02 26922890 - Fax +39 02 2133734
Yellow Led illuminates: the routine has failed. In this case, wait 8 seconds, until the instrument automatically re-
peat Preheating, then repeat the Calibration routine without inserting again the code. If condition still persists after
the replacement of the “Cartridge”, please send the detector back to the manufacturer for reparation.
“CALIBRATION CHECKS” (no Code required): using the gas mixture indicated above, this function is used to con-
trol, the correct response to the gas and can be made after the "Calibration" or the installation, but should be done
during the periodic maintenances, as it is the only method to verify the effective functioning of the detector.
Insert the TC014 on the sensor holder and adjust the sample gas bottle’s valve, so that the flow meter reaches the
value indicates in table below in column 3, depending on the model (see also Fig.3). Check that the voltmeter, with
probes connected on tests points TEST mV”, (see Fig.2) reaches (in 2÷3 minutes) the value shown in column 4, de-
pending on the model, [corresponding to mA output indicated in column 5 and the Central Unit which is connected to the detector, dis-
plays the value indicated in column 6]. If the result is different, is necessary to recalibrate the sensor (see Calibration).
Then, close the gas bottle, remove TC014. The mA output will slowly decrease up to 4 mA (normal working conditions).
1-Model 2-Sample gas bottle to be use 3-Flowmeter ADJ
4-TEST mA 5- mA Output 6- Central Unit Display
TS282
EA
100ppm Ammonia (NH
3
) in Nitrogen
0,3 l/min
90÷96 mV 9,0÷9,6 mA 100 ppm 5)
TS282
EA
-
H
1 l/min
TS282
EC
-
S
300ppm Carbon monoxide (CO) in Nitrogen 0,3 l/min 190÷210 mV
19÷21 mA 300 ppm (±20)
TS282
EC
-
H
TS282
ES
10ppm Sulphur dioxide (SO
2
) in Nitrogen
0,3
l/min
118÷122 mV
11,7÷12,3 mA
10.0
ppm (±0.5)
TS282
EH
50ppm Hydrogen sulphide (H
2
S) in Nitrogen
0,3 l/min
117÷123 mV
11,7÷12,3 mA
50
ppm (±2)
TS282
EHCN
10ppm Hydrogen cyanide (HCN) in Nitrogen
0,3÷0,5 l/min
192÷208 mV
19,2÷20,8 mA
10.0
ppm (±1)
TS282
EN
100ppm Nitric oxide (NO) in Nitrogen
0,3 l/min
192÷208 mV
19,2÷20,8 mA
100
ppm (±5)
TS282
EN2
30ppm Nitrogen dioxide (NO
2
) in air
1 l/min
190÷210 mV
19÷21 mA
30
ppm (±2)
TS282
EHCL
10ppm Hydrogen cloride (HCL) in Nitrogen
0,3÷0,5 l/min
192÷208 mV
19,2÷20,8 mA
10
ppm (±1)
TS282
ECL
10ppm Nitrogen dioxide (NO
2
) in air
1 l/min
192÷208 mV
19,2÷20,8 mA
10.0
ppm (±0,5)
FR
DESCRIPTION
Les TS282E, sont sondes à transmetteur (4÷20mA) pour gaz toxiques, équipée d’un capteur a cellule électrochimique
utilisé en systèmes centralisés d'alarme pour parkings, laboratoires, industries et environnements qui doit être protégé par la
présence de gaz. Le détecteur est constitué par un boîtier contenant le circuit électronique et les borniers de raccor-
dement; dans le porte capteur, placé dans la partie inférieure du boîtier, est logée une "cartouche capteur échan-
geable" contenant l'élément sensible et les données identificatrices et de réglage.
Les TS282E, sont sondes à transmetteur sur 3 fils avec un signal de sortie (S) 20mA linéaire avec fin d’échelle du
gaz mesuré (comme indiqué sur la page 1). Il s’utilise en se raccordant sur les centrales d’alarmes mono et multivoies
Tecnocontrol. Sur le couvercle, se trouvent les touches de codage F1 et F2 pour les opérations de vérification et
calibration et 3 LED indiquant les conditions de fonctionnement:
Signalisations et commandes:
Led rouge "ALARM": signal d'alarme ALARM.
Led vert "ON": fonctionnement normal.
Led jaune "FAULT": capteur en panne ou déconnecté ou saturé ou échu.
Touches F1 et F2 pour les opérations de vérification et calibrage
MODÈLES ET CARACTÉRISTIQUES DU GAZ DÈTECTEÈ
Le TS282EH permet de détecter l'Ammoniac (NH
3
) qui est un gaz toxique et inflammable, incolore avec une odeur
très âcre, est plus léger que l'air (sa densité relative à l'air est 0,59). Sa Limites d’Exposition Professionnelle (LEP) est TWA
(Time-Weighted Average) 25 ppm et STEL (Short Term Exposure Limit) 35 ppm. Sa LIE (Limite Inférieure d'Explosivité) est 15%
volume. L'Ammoniac à 50÷100 ppm provoquer une irritation des voies respiratoires et au-delà de 100 ppm pro-
voque une irritation des yeux. A 5000 ppm est mortelle. L’ammoniac est utilisé comme base pour les engrais, en
tant que composant dans les peintures, comme réfrigérant dans le l'industrie du froid, pour la production d'explo-
sifs, pour la production de plastiques et polymères, et pour les solvants.
Le TS282EA-H est comme le modèle précédente, mais il utilise un capteur adappour les environnements pol-
lués et de l'industrie et à la présence d'autres gaz toxiques
Le TS282EC-S permet de détecter le Monoxyde de Carbone (CO) qui est un gaz toxique, inodore, sans couleur,
est lourd comme l'air, (Sa densité relative à l’air est 0,97). Il est explosif aussi mais à des concentrations élevées, en
effet sa LIE (Limite Inférieure d'Explosivité) est 10,9% volume. Sa Limites d’Exposition Professionnelle (LEP) est TWA (Time-
Weighted Average) 25ppm et STEL (Short Term Exposure Limit) 50ppm. Le CO est produite par la combustion incomplète
des combustibles à base de carbone. Quand ces combustibles, gaz huile, bois et charbon ne brûlent pas correc-
tement, il se forme un excès de CO. Quand le CO, en les pandant dans l’ambiance, il est respiré par les personnes,
en transitant par les des poumons il atteint le sang, il empêche le transport de l'oxygène aux tissus et aux organes.
Le TS282EC-H est comme le modèle précédente, mais il utilise un capteur adappour les environnements pol-
lués et de l'industrie et à la présence d'autres gaz toxiques
Le TS282ES permet de détecter le Dioxyde de soufre (SO
2
) (ou Anhydride sulfureux, Oxyde sulfureux, Oxyde de soufre) qui
est un gaz toxique, incolore, d'odeur âcre, son inhalation est fortement irritante. Est plus lourd que l'air (sa densité
relative à l'air est 2,25). Sa Limites d’Exposition Professionnelle (LEP) est TWA (Time-Weighted Average) 2ppm et STEL (Short
IST-2282.EC01.01 Istruzione / User’s Manual / Manuel d’utilisation Pag. 12/17
TECNOCONTROL S.r.l. - Via Miglioli, 47 20090 SEGRATE (MI) ITALY Tel. +39 02 26922890 - Fax +39 02 2133734
Term Exposure Limit) 5ppm. Le SO
2
est fortement irritante pour les yeux et les voies respiratoires. L'inhalation du gaz
peut causer un œdème pulmonaire. L'exposition peut entraîner la mort.
Le TS282EH permet de détecter le Sulfure d'hydrogène (H
2
S) qui est un gaz toxique et inflammable, incolore
avec une odeur caractéristique d'œufs pourris, est plus lourd que l'air (Sa densirelative à l'air est 1,19). Sa Limites
d’Exposition Professionnelle (LEP) est TWA (Time-Weighted Average) 5ppm et STEL (Short Term Exposure Limit) 10ppm.
Sa LIE (Limite Inférieure d'Explosivité) est 4% volume. Son odeur est perçue à des concentrations supérieures à 0,2
ppm, aux niveaux autour de 50-100 ppm il produit déjà une irritation des yeux et de la gorge, toux, augmentation
de la respiration et de la formation de liquide dans les voies respiratoires. Au-delà de 150 ppm, ce gaz agit sur le
nerf olfactif, rendant impossible la perception de l'odeur, et des concentrations élevées peuvent provoquer l'incons-
cience en quelques minutes et peuvent entraîner la mort.
Le TS282EHCN permet de détecter le Cyanure d'hydrogène (HCN) (ou Acide cyanhydrique, Acide prussique) qui est un
gaz toxique, incolore exhalant une odeur caractéristique d'amandes amères, son inhalation est fortement irritante.
Est plus léger que l'air (sa densité relative à l'air est 0,94). Sa Limites d’Exposition Professionnelle (LEP) est TWA (Time-
Weighted Average) 4,7 ppm et STEL (Short Term Exposure Limit) 10 ppm. Le HCN est fortement irritante pour les yeux et
les voies respiratoires. L'inhalation du gaz est extrêmement toxique et mortel par anoxie. Une concentration de 300
ppm dans l'air tue un homme en quelques minutes.
Le TS282EHN permet de détecter l’Oxyde nitrique (NO) qui est un gaz toxique, incolore, est lourd comme l'air,
(sa densité relative à l'air est 1,04). Sa Limites d’Exposition Professionnelle (LEP) est TWA (Time-Weighted Average) et STEL (Short
Term Exposure Limit) 25 ppm. Le NO est un oxydant fort et il réagit avec le combustible et les matériaux réduisant.
Avec le contact de l'air se transforme en dioxyde d'azote (NO
2
). L’Oxyde nitrique est irritant pour les yeux et les
voies respiratoires. L'inhalation peut causer un œdème pulmonaire, mais aussi avoir des effets sur le sang, provo-
quant la formation de méthémoglobine. Dans les cas plus graves il peuve entraîner la mort. L’Oxyde nitrique est un
polluant qui est générée en partie dans les procédés de combustion, en partie par les émissions naturelles des
éruptions volcaniques, des incendies ou des activités civiles et industriels de combustion, tels que les transports
(véhicules à moteur diesel, essence, GPL, etc.) et la production du chaleur et d'électricité.
Le TS282EN2 permet de détecter le Dioxyde d'azote (NO
2
) est un gaz toxique, d'odeur âcre, brun rougeâtre, plus
lourd que l'air (Sa densité relative à l’air est 1,58). Sa Limites d’Exposition Professionnelle (LEP) est TWA (Time-Weighted Ave-
rage) 3 ppm et STEL (Short Term Exposure Limit) 5 ppm. Lorsqu'il est combiné à l’humidité, il se transforme en acide ni-
trique et oxyde nitrique. Le NO
2
se forme par oxydation du monoxyde d’azote (NO), lui-même formé par combinai-
son directe de l'azote et de l'oxygène à haute température durant certains processus de combustion (centrales
thermiques, installations de chauffage et véhicules surtout à moteurs diesel). Le NO
2
même à concentration modé-
rée, provoque: toux aigues, douleurs thoraciques, convulsions et insuffisances circulatoires. Il peut, en outre, pro-
voquer des dommages irréversibles aux poumons, des œdèmes et le décès car en se combinant avec
l’hémoglobine, il forme une molécule empêchant le transport de l’oxygène aux tissus et organes.
Le TS282EHCl permet de détecter le Acide chlorhydrique (HCl) (ou Chlorure d'Hydrogène, le Chlorhydrique acide) qui est
normalement un liquide, ce qui libère un gaz toxique, incolore avec une odeur piquante, plus lourd que l'air (sa
densité relative à l'air est 1,3). Sa Limites d’Exposition Professionnelle (LEP) est TWA (Time-Weighted Average) 5 ppm et STEL
(Short Term Exposure Limit) 10 ppm. En solution aqueuse devient l'un des liquides les plus corrosifs existent. Les va-
peurs sont corrosives pour les yeux, la peau et les voies respiratoires. L'inhalation de concentrations élevées de
gaz provoquer un œdème pulmonaire et une pneumonie.
Le TS282ECL permet de détecter le Chlore (CL
2
), qui est un gaz vert jaunâtre, avec une odeur suffocante très dé-
sagréable, il est très toxique, plus lourd que l'air (sa densité relative à l'air est 2,5). Sa Limites d’Exposition Professionnelle
(LEP) est TWA (Time-Weighted Average) 0,5 ppm et STEL (Short Term Exposure Limit) 1 ppm. En solution aqueuse devient
un acide corrosif. Les vapeurs sont corrosives pour les yeux, la peau et les voies respiratoires. L'inhalation du gaz
peut causer un œdème pulmonaire et une pneumonie. Une exposition prolongée peut conduire à la mort.
FONCTIONNEMENT
Le capteur électrochimique, est compensé en température mais il est sensible aux variations extrêmes d'humidité.
L'étalonnage est effectué pour le gaz à détecter, mais en même temps l’on peut détecter d’autres gaz présents
dans la même ambiance, comme indiqués au Tableaux 2 et 3.
Préchauffage: à partir de la mise sous tension le détecteur commence la phase de préchauffage du capteur, si-
gnalée par le clignotement de la LED jaune "FAULT". Après environ 60 secondes, la LED jaune s'éteint et la LED
verte "ON" s'allume, indiquant le fonctionnement normal. Après ce temps le capteur est apte à détecter le gaz, mais
il n’atteint les conditions de stabilité optimale qu’après quatre heures environ de fonctionnement continu.
Fonctionnement Normal: seule la LED vert (ON) doit être allumée.
Alarme: Le LED rouge, ALARM s'allume si la concentration de gaz dépasse la valeur indiquée dans le tableau ci-
dessous, selon le modèle (Habilitation avec Dip-Switch n.1 ON).
TS282
EA
TS282EA-H
TS282
EC
-
TS282EC-H
TS282ES TS282EH TS282EHCN TS282EN TS282EN2 TS282EHCL TS282ECL
50 ppm NH
3
200 ppm CO
10 ppm SO
2
20 ppm H
2
S
5 ppm HCN
25 ppm NO
30 ppm NO
2
5 ppm HCL
1 ppm CL
Dérangement: La centrale signale les anomalies, décrites ci-dessous, en allumant la LED jaune (FAULT), en por-
tant la sortie "S" à 0mA.
La LED jaune clignote toute les 4 secondes, (avec la LED vert allumée): pour avertir que la "Cartouche Capteur" a dépas-
sa limite de vie, et que le fonctionnement correct n'est plus garanti. Le détecteur continue à fonctionner normalement,
IST-2282.EC01.01 Istruzione / User’s Manual / Manuel d’utilisation Pag. 13/17
TECNOCONTROL S.r.l. - Via Miglioli, 47 20090 SEGRATE (MI) ITALY Tel. +39 02 26922890 - Fax +39 02 2133734
mais il estcessaire, au plus tôt de remplacer la "Cartouche Capteur" par une nouvelle. Le type à commander est indiqué
en
Page 1. La procédure de substitution est décrite dans la documentation jointe à la cartouche.
Si la LED jaune est allumée et la vert est éteinte (sortie 0mA): indication de plusieurs possibilités de dérange-
ment, c'est-à-dire: 1°) la configuration des Dip-Switch n'est pas correcte, vérifier la position, (Tableau 4). 2°) la "Car-
touche Capteur" est en panne, la remplacer par une nouvelle. 3°) si une nouvelle "Cartouche" est installée: ou bien
elle n'est pas correctement connectée, ou bien elle n’est pas compatible. Contrôler les connexions avec la "Car-
touche" et la compatibilité, (Voire en page 1). Exécutez le contrôle en coupant et en rétablissant l’alimentation du dé-
tecteur. Si la condition perdure il sera nécessaire de remplacer et/ou de renvoyer le détecteur au fournisseur pour
réparation.
Si la LED jaune et la vert sont allumées, (sortie 0mA): indication de la panne probable de la "Cartouche ". Essayer
tout d’abord du "glage du ZÉRO" comme crit dans la rubrique "Vérification et Etalonnage", puis exécutez le contrôle
en coupant et en rétablissant l’alimentation du détecteur. En cas de persistance remplacer la "Cartouche ". Si malgré
cela, la condition persiste, il sera nécessaire de remplacer et/ou renvoyer le détecteur au fournisseur pour réparation.
Si toutes les LED sont allumées (tous les relais activés et sortie >20mA): indication de: ou panne de la "Cartouche " ou
une concentration de gaz supérieure au fond d’échelle. S'il n'est pas constade fuite de gaz et que la condition per-
dure après la substitution de la "Cartouche" il sera nécessaire d'envoyer le détecteur au fournisseur pour réparation.
INSTALLATION
Les détecteurs doivent être installés, positionnés et vérifiés en suivant toutes les règles nationales en vigueur pour
les installations électriques dans les zones avec dangers d'explosions et les normes de sécurité des installations.
Montage: en Fig.1 sont indiquées les dimensions. Le détecteur doit être monté en position verticale avec le cap-
teur tourné vers le bas.
Positionnement du TS282EA et TS282EA-H: Il doit être fixé à environ 30÷50 cm du plafond, le gaz NH
3
étant plus
léger que l'air.
Positionnement du TS282EC-S et TS282EC-H: Il doit être fixé à environ 160 cm du plancher, le gaz CO étant d'envi-
ron léger que l'air.
Positionnement du TS282ES: Il doit être fixé à environ 30-50cm du plancher, le gaz SO
2
étant plus lourdes que l'air.
Positionnement du TS282EH: Il doit être fixé à environ 30-40cm du plancher, le gaz H
2
S étant plus lourdes que l'air.
Positionnement du TS282EHCN: Il doit être fixé à environ 160 cm du plancher pour la protection des personnes
ou sur le point susceptible d'émission, le gaz HCN est légèrement plus léger que l'air.
Positionnement du TS282EN: Il doit être fixé à environ 160 cm du plancher, le gaz NO étant d'environ léger que l'air.
Positionnement du TS282EN2: Il doit être fixé à environ 30-40cm du plancher, le gaz NO
2
étant plus lourdes que l'air.
Positionnement du TS282EHCL: Il doit être fixé à environ 30-40cm du plancher, le gaz HCL étant plus lourdes que l'air.
Positionnement du TS282ECL: Il doit être fixé à environ 30-40cm du plancher, le gaz CL
2
étant plus lourdes que l'air.
Raccordements électriques (Fig.2): la distance maximale à laquelle chaque détecteur peut être raccordées à la
centrale est indiquée dans le Tableau 1 en fonction de la section du câble utilisé. Normalement on utilise un câble à
écran à 3 conducteurs. Raccorder l’écran du câble a la masse de la centrale.
Les borniers (+ - S), de type brochable, sont situés sur la carte principale, et il est nécessaire de le débrocher pour
effectuer les connexions. Prêter attention en les réinsérant étant donné qu'ils sont polarisés.
Remarque: Les Dip-Switch doivent être paramétrés avant d'alimenter le détecteur. Si on utilise le détecteur
avec le Dip-Switch 1 en position ON l’indication d’alarme (LED rouge) sera activée (Tableau 2).
Important: Une fois terminée l'installation, alimenter le détecteur, attendre 30÷60 minutes environ et puis pour
adapter le capteur aux conditions ambiantes, exécuter la "Régulation du Zéro" (voir Vérifications et Calibrage).
COMPATIBILITE AVEC D’AUTRES CENTRALES: En cas d’utilisation d’une centrale autre que de marque Tecno-
control, calculer la résistance de charge maxi en utilisant l’abaque selon la Fig. 4. L’alimentation sous 24Vcc est
conseillée. Si l’on utilise une alimentation 24Vcc la résistance maximale de charge (RL) sera de 500 Ω.
INSTRUCTIONS
La vie utile du capteur en air propre est de 2 ans en moyenne. Au terme de cette période, indiqué par l'instrument
par un clignotement de la LED jaune toutes les 4 secondes, il est nécessaire de remplacer la "Cartouche Capteur".
Vérifications Périodiques: nous conseillons d'exécuter tous les 6 à 12 mois la vérification de fonctionnement du
détecteur: Test Électrique, Réglage du Zéro et Vérification et Etalonnage avec les bouteilles avec mélange de gaz titré (voir
section Vérifications et Etalonnage).
Remarque: Le détecteur n'est pas apte à révéler des fuites survenant hors de l’ambiance dans laquelle il est ins-
tallé ou à l'intérieur des murs ou sous un plancher.
Limites de fonctionnement: des températures inférieures à -40°C et supérieures à +60°C provoquent la détério-
ration du capteur électrolytique. Il peut également être endommagé s’il est exposé à des concentrations de gaz
très élevées indiquée ci-dessous, selon le modèle. Dans ce cas la "Cartouche Capteur" doit être remplacé.
TS282
EA
TS282EA-H
TS282
EC
-
TS282EC-H
TS282ES TS282EH TS282EHCN TS282EN TS282EN2 TS282EHCL TS282ECL
500ppm NH
3
1000ppm CO
150ppm
SO
2
500ppm
H
2
S
100ppm EHN
1000ppm NO
150ppm NO
2
>30ppm HCL
>50ppm CL
2
Attention: Les capteurs électrochimiques contiennent une petite quantité d’acide sulfurique qui est corrosif. Nor-
malement il ne doit y avoir aucune perte au capteur, dans le cas contraire ceci est à une action mécanique vio-
lente ou à un usage hors-limites de fonctionnement. Dans le cas d’un contact accidentel entre du liquide provenant
du capteur et la peau ou les yeux, il est conseillé de laver immédiatement avec de l’eau en abondance.
IST-2282.EC01.01 Istruzione / User’s Manual / Manuel d’utilisation Pag. 14/17
TECNOCONTROL S.r.l. - Via Miglioli, 47 20090 SEGRATE (MI) ITALY Tel. +39 02 26922890 - Fax +39 02 2133734
Interférence avec d’autres gaz: La cellule électrochimique utilisée possède une bonne résistance aux produits
d’usage courant comme les aérosols, tergents, alcools, colles ou vernis; toutefois si ceux-ci sont présents en
quantités élevées, ils peuvent interférer avec le capteur; il est conseilde ventiler l’ambiance après emploi de ces
substances. Considérer qu’en ambiances particulièrement polluées par des vapeurs de solvants organiques, hydro-
gène, alcool, en quantités élevées, des fausses alarmes peuvent apparaître et que le capteur devra, pour retrouver
ses conditions de fonctionnement normales rester longtemps en air propre.
VÉRIFICATIONS ET ETALONNAGE
Remarque Importante: Toutes les opérations suivantes doivent être exécutées seulement par un person-
nel compétent et formé, étant donné que durant ces opérations, les sorties en courant (mA) en provoquant
l'activation des dispositifs d'alarmes connectés à la centrale
Le TS282E possède trois différentes fonctions protégées par un "Code": Test Électrique, Réglage du Zéro et Étalonnage.
Pour accéder à ces fonctions il est nécessaire d'insérer le "Code" en utilisant les touches F1 et F2. Pour que la
pression sur un touche soit reconnue, la tenir pressée durant environ une seconde (jusqu’à ce que s’éteigne un
bref instant la LED verte). Ensuite, l’on peut passer au bouton suivant. En cas d’erreur, il suffit d'attendre environ
10 secondes et la séquence est effacée automatiquement.
Kit d’étalonnage Tecnocontrol TC014 et Bouteille avec mélange de gaz titré (pour Vérification et Etalonnage):
Modèle
Bouteille avec mélange de gaz titré
notre modèle
/robinet
KIT
modèle
TS282EA
-
TS282EA
-
H
100 ppm
Ammoniac (NH
3
)
danse l’azote
BO501
/ BO303
TC014
TS282EC
-
S
-
TS282EC
-
H
300 ppm
Monoxyde de Carbone (CO)
danse l’azote
BO210
/ inclus
TC011/TC014
TS282ES
10 ppm
Dioxyde de soufre (SO
2
)
danse l’azote
BO518
/ BO303
TC014
TS282EH
50 ppm
Sulfure d'hydrogène
(H
2
S) danse l’azote
BO520
/ BO303
TC014
TS282EHCN
10 ppm
Cyanure d'hydrogène
(HCN) danse l’azote
BO529
/ BO303
TC014
TS282EN
100 ppm
Oxyde nitrique
(NO) danse l’azote
BO522
/ BO303
TC014
TS282EN2
30ppm
Dioxyde d'azote
(NO
2
) danse l’air
Disponible sur demande
TC014
TS282EHCL
10ppm
Acide
chlorhydrique
(HCL) danse l’azote
BO527
/ BO305
TC014
TS282ECL
10ppm
Dioxyde d'azote
(NO
2
) danse l’air
Disponible sur demande
TC014
Il est possible d'utiliser soit les bouteilles jetables avec robinet sortie de gaz, soit celles à la haute pression avec
détendeur. Quelques bouteilles disposent d'un réducteurs de débit fixes, rendre la valeur nécessaire dans le ta-
bleau, colonne 3, page 7. Utiliser le Kit d’étalonnage Tecnocontrol TC014 (Inox). Afin de garantir que le gaz re-
joigne la "Cartouche-Capteur", il est absolument nécessaire d’utiliser des réducteurs en acier et des tubes en Téflon
ou d’autres matériaux compatibles qui ne risquent pas d’entraîner sa dispersion.
TEST ELECTRIQUE (Code Test: F2, F2, F1, F1): cette fonction permet d'effectuer un test fonctionnel de l'appareil-
lage. Après avoir mis l'installation en sécurité et avoir inséré le "Code Test", toutes les LED s'éteignent. Ils s'allume-
ront ensuite en séquence, de la LED jaune jusqu'à la rouge. À la fin toutes les LED resteront allumées pendant en-
viron 5 secondes, puis le détecteur reviendra dans les conditions de fonctionnement normal. Il est souhaitable
d'exécuter cette opération tous les 12 mois en fonction de l’utilisation.
Remarque: Il n'est pas possible d'exécuter l'opération si la LED rouge est déjà allumés.
GLAGE DU ZÉRO (Code du Zéro: F2, F1, F1, F2) cette fonction sert pour régler le zéro du capteur et doit être
effectuée en air propre exclusivement (milieu sans présence de gaz polluants inflammables ou autres). Après avoir
inséré le "Code du Zéro", comme confirmation de l'opération 1 clignotement de la LED rouge et la sortie devient 4,0
mA. Nous conseillons d'exécuter cette opération après l'installation ou après le changement de la "Cartouche Capteur"
et chaque semestre en fonction des conditions ambiantes. Considérer que les TS282E sont dotée d’un suiveur de zé-
ro sophistiqué qui toutes les heures refait le zéro du capteur si, naturellement il n’y a pas de présence de gaz.
Remarque: Il n'est pas possible d'exécuter l'opération si la LED rouge est allumée ou si la sortie en mA est supérieure
à la valeur indiquée dans le tableau ci-dessous, selon le modèle. Dans le cas il cessaire de procéder à la calibra-
tion ou bien de substituer la "Cartouche Capteur".
TS282
EA
TS282EA-H
TS282
EC
-
TS282EC-H
TS282ES TS282EH TS282EHCN TS282EN TS282EN2 TS282EHCL TS282ECL
4,8 mA
15ppm NH
3
4,8 mA
15ppm CO
6,4 mA
3ppm SO
2
4,8 mA
5ppm H
2
S
8 mA
2,5ppm HCN
5,6 mA
10ppm NO
6,6 mA
5ppm NO
2
8 mA
2,5ppm HCL
4,8
mA
0,5ppm HCL
"ETALONNAGE" (Code de Etalonnage: F2, F2, F2, F1, F2, F1): cette fonction permet l’étalonnage du capteur
avec le mélange de gaz prévu.
Durant la procédure, une rare possibilité existe que la LED jaune clignote chaque 8 secondes, dans ce cas inter-
rompre la procédure puis exécutez le contrôle en coupant et en rétablissant l’alimentation du détecteur. En cas de
persistance il sera nécessaire de remplacer et/ou renvoyer le détecteur au fournisseur pour réparation.
Important: Durant l’opération d’étalonnage l’appareil porte la sortie "S" à 0mA.
L’étalonnage doit être exécuté en air propre exclusivement (milieu sans la présence de gaz toxiques ou autres). A l’aide
des touches exécuter le "Code d’Etalonnage". Attendre que les LED jaune et verte s’allument en feu fixe et que la LED
rouge clignote. Coiffer la tête de tection avec le TC014, gler le débit du gaz, de manière que le débitmètre in-
dique la valeur indique dans le tableau ci-dessous dans la colonne 3 (
voir aussi Fig.3). Attendre 3-5 minutes environ,
puis, quand la LED rouge s'allume (et pendant qu'elle est allumée), appuyer sur la touche F2 et maintenir cette pres-
sion tant que la LED rouge ne reste pas éteinte durant au moins 2 secondes (si, en revanche la LED rouge continue à cli-
gnoter plus de 5-6 minutes, cela signifie que le flux de gaz n’est pas constant, ou insuffisant, ou la concentration
IST-2282.EC01.01 Istruzione / User’s Manual / Manuel d’utilisation Pag. 15/17
TECNOCONTROL S.r.l. - Via Miglioli, 47 20090 SEGRATE (MI) ITALY Tel. +39 02 26922890 - Fax +39 02 2133734
n’est pas celle requise. Interrompre l’opération, éteindre l’instrument et répéter la calibration) Fermer la bouteille et
ôter le TC014. À ce point, deux cas peuvent se vérifier:
LED jaune et verte allumées: l’étalonnage est correct. Après 8 secondes, le détecteur reviendra automatique-
ment dans les conditions de fonctionnement normal. (Chapitre Fonctionnement > Préchauffage).
Led jaune allumée: l’étalonnage a échoué. Dans ce cas, après 8 secondes, le détecteur reviendra automatiquement
dans les conditions d’étalonnage, répéter la procédure sans réinsérer la séquence. Si la condition persiste même
après la substitution de la "Cartouche Capteur", il sera nécessaire d'envoyer le détecteur au fournisseur pour réparation.
"VÉRIFICATION D’ETALONNAGE" (code pas nécessaire): cette fonction permet la vérification du fonctionnement cor-
rect du détecteur avec le mélange de gaz prévu. La vérification doit être exécutée après l’"Etalonnage", ou après l'instal-
lation ou pendant les entretiens périodiques, étant donné qu’il s’agit de la seule méthode permettant de contrôler la
fonction effective du détecteur.
Coiffer la tête de détection avec le TC014, selon le modèle, régler le débit du gaz, de manière que le débitmètre in-
dique la valeur indique dans le tableau ci-dessous dans la colonne 3 (voir aussi Fig.3). Vérifiez avec le voltmètre sur les
bornes de test "TEST mA" (Fig.2) que la valeur en mV, indiquée dans la colonne 4 est atteinte. [Ou bien que la sortie en mA
augmente jusqu’à la valeur indiquée dans la colonne 5 et que la centrale à laquelle le détecteur est connecté indique une valeur en % de la
LIE comme indiquée dans la colonne 6 selon le modèle]. Un fois terminé la “Vérification”, ôter le TC014. Apres cela, la sortie
reviens progressivement à 4 mA. Si le résultat est différent, il convient de refaire l’étalonnage.
1-Modèle 2- Bouteille de gaz à utiliser 3- Débitmètre
4-TEST mA
5-Sortie mA 6-Indication de la Centrale
TS282EA
100ppm Ammoniac (NH
3
) dans azote
0,3 l/min
90÷96 mV 9,0÷9,6 mA 100 ppm (±5)
TS282EA-H
1 l/min
TS282EC-S
300ppm Monoxyde de Carbone (CO) dans azote
0,3 l/min 190÷210 mV
19÷21 mA 300 ppm (±20)
TS282EC-H
TS282ES
10ppm Dioxyde de soufre (SO
2
) dans azote
0,3 l/min 118÷122 mV
11,7÷12,3 mA
10.0 ppm (±0.5)
TS282EH
50 ppm Sulfure d'hydrogène (H
2
S) dans azote
0,3 l/min 117÷123 mV
11,7÷12,3 mA
50 ppm (±2)
TS282EHCN
10 ppm Cyanure d'hydrogène (HCN) dans azote
0,3÷0,5 l/min
192÷208 mV
19,2÷20,8 mA
10 ppm (±1)
TS282EN
100 ppm l’Oxyde nitrique (NO) dans azote
0,3 l/min 192÷208 mV
19,2÷20,8 mA
100 ppm (±5)
TS282EN2
30ppm Dioxyde d'azote (NO
2
) dans air
1 l/min 190÷210 mV
19÷21 mA 30 ppm (±2)
TS282EHCL
10ppm Acide chlorhydrique (HCL) dans azote
0,3÷0,5 l/min
192÷208 mV
19,2÷20,8 mA
10 ppm (±1)
TS282ECL
10ppm Dioxyde d'azote (NO
2
) dans air
1 l/min 192÷208 mV
19,2÷20,8 mA
10 ppm (±0,5)
IST-2282.EC01.01 Istruzione / User’s Manual / Manuel d’utilisation Pag. 16/17
TECNOCONTROL S.r.l. - Via Miglioli, 47 20090 SEGRATE (MI) ITALY Tel. +39 02 26922890 - Fax +39 02 2133734
Tabella 1 / Table 1 / Tableau 1
Sezione Cavo
schermato
Shielded Cable
Size
Section du câble
à écran
Resistenza Cavo
[Singolo Conduttore]
Cable Resistance
[Single wire]
Résistance câble
[par conducteur]
La max distanza cui può essere insta
l-
lato ogni trasmettitore dalla centralina
Tecnocontrol Modello ID250 e SE148 è:
the maximum distance to install each de-
tector from the Gas Central Unit Model
ID250 and SE148 is:
La maxi distance à laquelle peuvent être
raccordées les sondes à les centrales
ID250 et SE148 est:
La max distanza cui può essere installato ogni
trasmettitore dalla centrali Modello
CE100, CE400, CE600, CE700 (CE380UR) è:
The maximum distance to install each detector from
the Gas Central Unit Model
CE100, CE400, CE600, CE700 (CE380UR) is:
La maxi distance à laquelle peuvent être raccordées
les sondes à les centrales CE100, CE400, CE600,
CE700(CE380UR) est:
0,75
mm
2
26
Ω/
km
100 m
300 m
1
mm
2
20
Ω/
km
150 m
400 m
1,5
mm
2
14
Ω/
km
200 m
500 m
2,5
mm
2
8
Ω/
km
400 m
800 m
Tabella 2: Interferenze con altri Gas / Table 2: Typical cross Sensitivities / Tableau 2: Interférence
GAS / GAZ
Gas Utilizzato
Test Gas Used
Gaz utilisé
ppm letti come gas
Gas concentration equivalent (ppm)
ppm lu comme gaz
TS282
EA
(1)
TS282
EA
-
H
(2)
TS282
EC
-
(3)
TS282
EC
-
H
(4)
TS282
ES
(5)
TS282
EH
(
(6)
(1) (2)
Ammoniaca /
Ammonia / Ammoniac
100 ppm
100
100
- - - 0
Anidride carbonica
Carbon dioxide / Dioxyde de carbone
5000 ppm 0 - 0.4 - - - -
(5)
Biossido di zolfo
Sulphur dioxide / Dioxyde de soufre
10 ppm
- 2 < 15 0 0
10
≈ 2
Biossido di azoto / Nitrogen dioxide / dioxyde d’azote
20 ppm < 2 < 2 - ≤ ±1
≈ - 16
Cloro /
Clorine / Chlore
1 ppm 0 0 - 0 - -
Etanolo /
Ethanol / Éthanol
200 ppm 0 0.2 < 5 0 - -
Etilene /
Ethylene
/ Éthylène
100 ppm 0 0 < 80 < 50 - ≤ 0.1
Idrogeno /
Hydrogen / Hydrogène
100 ppm 0 < 0.2 < 46 < 40 - < 1
(6) Idrogeno Solforato
Hydrogen sulphide / Sulfure d'hydrogène
15 ppm
- 3÷0 ~3 0 < 0.75 0
15
(3) (4)
Monossido di carbonio
Carbon Monoxide / Monoxyde de carbone
100 ppm
0 <1
100 100
< 1 ≤ 1
Ossido di azoto /
Nitric oxide / Oxyde nitrique
35
ppm
0 0 ~ 35 < 3 0 < 0.7
Tabella 3: Interferenze con altri Gas / Table 3: Typical cross Sensitivities / Tableau 3: Interférence
GAS / GAZ
Gas Utilizzato
Test Gas Used
Gaz utilisé
ppm letti come gas
Gas concentration equivalent (ppm)
ppm lu comme gaz
TS282
EHCN
(7)
TS282
EN
(8)
TS282
EN2
(9)
TS282
EHCL
(10)
TS282
ECL
(11)
(7)
Acido cianidrico
Hydrogen cyanide / Cyanure d'hydrogène
10 ppm
10
- - 3.5 -
(10)
Acido cloridrico
Hydrogen cloride / Acide chlorhydrique
10 ppm -
- -
10
-
Ammoniaca /
Ammonia / Ammoniac
100 ppm - - 0 0 0
Acetato di etile /
Ethyl acetate / Acétate d'éthyle
100 ppm - - < 0.5 - -
Biossido di zolfo /
Sulphur di
oxide / Dioxyde de
soufre
20 ppm 40 ÷ 75 0 - 0.4 8 < 2
(9)
Biossido di azoto
Nitrogen dioxide / dioxyde d’azote
5 ppm
- 20 ÷ -10
< 1.5
5
< ± 0.2
5
(11) Cloro / Clorine / Chlore
5 ppm
- - ≈ 10 < ± 0.2
Etanolo /
Ethanol / Éthanol
100 ppm - - 0 - < 1
Etilene /
Ethylene / Éthylène
100 ppm < 25 - 0 - < 1
Idrogeno /
Hydrogen / Hydrogène
1000 ppm - - 0 0 < 1
(6) Idrogeno Solforato
Hydrogen sulphide / Sulfure d'hydrogène
15 ppm
~ 90 ≈ 1.5 - 19 45 < -15
Monossido di carbonio
Carbon Monoxide / Monoxyde de carbone
300 ppm < 15 0 0 0 < 2
(8) Ossido di azoto /
Nitric oxide / Oxyde nitrique
35 ppm
-28 ÷ 0
35
0 0 < 1
Toluene /
Toluene / Toluène
50 ppm - - < 1.5 - -
Tabella 4 / Table 4 / Tableau 4
“S1
-
SET”(Dip
-
Switch)
Led Allarme
Alarm Led / Led d'alarme
4
ON
ON
ON
ON
OFF
ON
OF
F
OFF
OFF
ON
Il Dip-Switch va posizionato prima d’alimentare l’apparecchio.
Dip-Switch should be set with instrument powered off.
Les Dip-Switch doivent être paramétrés avant d'alimenter le détecteur
IST-2282.EC01.01 Istruzione / User’s Manual / Manuel d’utilisation Pag. 17/17
TECNOCONTROL S.r.l. - Via Miglioli, 47 20090 SEGRATE (MI) ITALY Tel. +39 02 26922890 - Fax +39 02 2133734
Fig. 1 –
Dimensioni /
Size /
Dimensions
1
2
3
4
ON
OFF
Dip Switch
SET
Fig. 2 -
Schema di collegamento
/ Wiring diagram / Schéma
TESTm
TP2
TP1
-
S
12
-
24V
mA
Centrale rivelazione gas
Gas Central Unit
Centrale de détection
RL
-
S
+
Test
Uscit
a
Output Test
Sortie Test
40÷200mV
+/
-
(12÷24Vdc
)
Alimentazione / Power Supply / Alimentation
S
” =
Uscita 4÷20 mA
4÷20 mA Output / Sortie en 4÷20 mA
m
V
Fig.
3
-
Tester di calibrazione
/ Calibration Tester / Kit de Calibration
F1
F2
Tasti /
Key /
Touches
TC014
0,3 L/min
1 L/min
Alimentazione
/ Power Supply / Alimentation
100
300
400
500
50
0
5
200
Vdc
RL=Resistenza di carico
/ Load resistor / Resistance de charge
RL
Ohm
Fig.4 - Alimentazione / Resistenza di Carico 4÷20mA
Power supply / Load resistance diagram 4÷20mA
Alimentation / Résistance de charge 4÷20mA
600
0
10
15
20
25
28
650
10,8
24
Area di funzionamento
Allowable oparating region
350
700
800
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17

Tecno Control TS282EH Manuel utilisateur

Catégorie
Détecteurs de monoxyde de carbone (CO)
Taper
Manuel utilisateur