Roche CoaguChek XS Manuel utilisateur

Catégorie
Mesure, test
Taper
Manuel utilisateur
CoaguChek
®
XS
User’s Manual
Manuel d’utilisation
ccxs_user_EN.book Seite 1 Mittwoch, 3. März 2010 1:45 13
2
On the packaging, on the type plate of the meter and in the User’s
Manual you may encounter the following symbols and abbreviations
shown here with their meaning:
Use by
Batch code/ Lot number
In vitro diagnostic medical device
This product fulfills the requirements of the European
Directive 98/79/EC for in vitro diagnostic medical devices.
Catalogue number
Consult instructions for use
Caution (consult accompanying documents). Refer to
safety-related notes in the manual accompanying this
instrument.
Manufacturer
Temperature limitation (Store at)
LOT
IVD
0123
ccxs_user_EN.book Seite 2 Mittwoch, 3. März 2010 1:45 13
The CoaguChek XS System
3
English
The CoaguChek XS System
The CoaguChek XS System is used for quantitative monitoring of coagu-
lation values (common terms: thromboplastin time, PT, Quick value) with
CoaguChek XS PT test strips.
Self-monitoring and self-adjustment of oral coagulation therapy using
coagulation values determined with the aid of the CoaguChek XS System
may only be undertaken after consultation with your doctor and compre-
hensive instruction by a qualified healthcare professional. Your pharma-
cist / specialist supplier can provide you the training and/or put you in
touch with associations or institutions offering training in coagulation
self-monitoring. Please read the special notes on INR self-monitoring in
the package insert accompanying the test strips.
Last update user information: 2010-03
WARNING
The CoaguChek Softclix lancing device is intended only
for patient self-monitoring by a single person. It must
not be used to collect blood in a multi-patient
setting as it does not incorporate any features to
guard against cross-infection.
This manual contains all the information you need to operate and care
for the CoaguChek XS System. Please read this entire manual care-
fully before you use the meter.
Healthcare professionals: In addition, observe carefully the pre-
cautions and procedures applying to professional use provided in the
relevant sections of this manual.
ccxs_user_EN.book Seite 3 Mittwoch, 3. März 2010 1:45 13
The CoaguChek XS System
4
ccxs_user_EN.book Seite 4 Mittwoch, 3. März 2010 1:45 13
5
English
The CoaguChek XS System
3
Introduction
9
The CoaguChek XS System
......................................................................... 9
Test principle
..................................................................................................... 10
Contents of the pack
..................................................................................... 10
Operating conditions
.................................................................................... 11
Quality control
.................................................................................................. 12
Safety Information and Additional Information
13
The CoaguChek XS Meter
14
The CoaguChek XS meter
.......................................................................... 15
Batteries
............................................................................................................... 16
Starting Up
17
Inserting the batteries
.................................................................................. 18
Code Chip
21
Inserting the code chip
................................................................................ 22
Meter Setup
24
Settings summary
........................................................................................... 25
Setting up the meter (Set Mode)
............................................................ 26
Setting the date format
................................................................................ 29
Setting the date
............................................................................................... 30
Setting the time format
................................................................................ 32
Setting the time
............................................................................................... 33
Selecting the unit
............................................................................................ 34
Setting the beep tone
................................................................................... 35
Setting the therapeutic range (INR)
...................................................... 36
ccxs_user_EN.book Seite 5 Mittwoch, 3. März 2010 1:45 13
6
Testing a Capillary Blood Sample
41
Important notes
............................................................................................... 42
Getting a good capillary blood sample
............................................... 44
Preparing to test
............................................................................................. 45
Performing a test
............................................................................................ 46
Notes concerning the results display
.................................................. 57
Testing by Healthcare Professionals
61
Getting a good capillary blood sample in a healthcare
setting
.................................................................................................................. 62
Disposal by healthcare professionals
.................................................. 62
Memory
63
Viewing test results
....................................................................................... 63
Erasing memory
.............................................................................................. 66
Cleaning and Disinfecting the Meter
69
Recommended cleaning/disinfecting solutions
............................. 70
Cleaning/disinfecting the exterior (meter housing)
..................... 71
Cleaning/disinfecting the test strip guide
......................................... 72
Error Messages
75
Error messages overview
........................................................................... 76
Error messages after turning the meter on
...................................... 78
Error messages when preparing to test
............................................ 80
Error messages after the code number was confirmed
............. 84
Error messages during or after blood application
........................ 86
Error message during infrared downloads
....................................... 94
ccxs_user_EN.book Seite 6 Mittwoch, 3. März 2010 1:45 13
7
English
Display Information and Symbols
96
Further Information
99
Ordering
.............................................................................................................. 99
Product limitations
......................................................................................... 99
Product Specifications
100
Operating conditions and technical data
........................................ 100
Sample material
............................................................................................ 101
Storage and transport conditions
........................................................ 101
Disposal of the CoaguChek XS meter
............................................... 102
Disposal of used batteries
....................................................................... 102
Information service
..................................................................................... 103
Repairs
.............................................................................................................. 103
Guarantee
104
Alphabetical Index
105
ccxs_user_EN.book Seite 7 Mittwoch, 3. März 2010 1:45 13
8
ccxs_user_EN.book Seite 8 Mittwoch, 3. März 2010 1:45 13
Introduction
9
English
Introduction
The CoaguChek XS System
The CoaguChek XS System (CoaguChek XS meter and CoaguChek XS PT
test strips) quantitatively determines prothrombin time (PT/Quick value/
INR) using capillary blood from a fingertip or untreated venous whole
blood. The CoaguChek XS System makes coagulation testing easy.
The CoaguChek XS meter guides you through the test step by step
using the symbols in the display.
The code chip accompanying the test
strips contains lot-specific information about these test strips, including
calibration data that allow the correct results to be calculated, and the
expiry date for the test strips. You only need to insert the code chip, turn
the meter on, insert the test strip, and apply a blood sample. When the
CoaguChek XS meter is ready for a measurement, the result will be
displayed about 1 minute after application of the sample. After the
measurement, the meter automatically stores the result in memory.
If you have any questions about the CoaguChek XS meter, please contact
your local customer support and service centre. You will find contact
details on page 103.
Please note: Before you use the meter for the first time (i.e. after you
have first inserted the batteries), you must set the date and time correctly
to allow you to carry out measurements properly. Each time you replace
the batteries you need to check (and, if necessary adjust) the date and
time.
ccxs_user_EN.book Seite 9 Mittwoch, 3. März 2010 1:45 13
Introduction
10
Test principle
The CoaguChek XS PT test strip contains a lyophilized reagent (reagent
in dried form). The reactive components of this reagent consist of throm-
boplastin and a peptide substrate. When a sample is applied, thrombo-
plastin activates coagulation, which leads to the formation of thrombin.
At the same time the meter starts to measure the time. The enzyme
thrombin cleaves the peptide substrate, generating an electrochemical
signal. Depending on the time elapsed when it first appears, this signal
is then converted by means of an algorithm into customary coagulation
units (INR, %Quick, seconds) and the result is displayed.
Contents of the pack
CoaguChek XS meter
4 alkali-manganese batteries, 1.5 V, type AAA
CoaguChek Softclix lancing device with instructions for use
CoaguChek Softclix Lancet
System carry-case
User's Manual
ccxs_user_EN.book Seite 10 Mittwoch, 3. März 2010 1:45 13
Introduction
11
English
Operating conditions
To ensure that your CoaguChek XS System functions properly, please
observe the following guidelines:
Only use the meter at a room temperature between 15°C and 32°C.
Only use the meter at a relative humidity between 10% and 85%.
When testing, place the meter on a level, vibration-free surface or
hold it so it is roughly horizontal.
If the meter is to remain unused for a longer period of time, keep it in
the carry-case supplied.
Use the meter only up to a maximum altitude of 4300 metres.
Electromagnetic interference
Strong electromagnetic fields may interfere with the
proper operation of the meter. Do not use the meter near
strong electromagnetic fields.
ccxs_user_EN.book Seite 11 Mittwoch, 3. März 2010 1:45 13
Introduction
12
Quality control
The CoaguChek XS System has a number of inbuilt quality-control
functions such as:
A check of the electronic components and functions every time the
meter is turned on.
A check of the test strip temperature while a test is in progress.
A check of the expiry date and lot information on the test strip.
A quality control function is incorporated into the test strip. Quality
control and system checks using control test solutions that you may
be familiar with from other systems are no longer required.
ccxs_user_EN.book Seite 12 Mittwoch, 3. März 2010 1:45 13
Safety Information and Additional Information
13
English
Safety Information and Additional
Information
This section explains how safety-related messages and information
related to the proper handling of the system are presented in the
CoaguChek XS User’s Manual. Read these passages carefully.
The safety alert symbol by itself (without a signal word) is
used to promote awareness to hazards which are generic
or to direct the reader to related safety information
WARNING
Indicates a hazardous situation which, if not avoided,
could result in death or serious injury.
CAUTION
Indicates a hazardous situation which, if not avoided,
could result in minor or moderate injury.
NOTICE
Indicates a hazardous situation which, if not avoided,
may result in damage to the system.
Important information that is not safety relevant is presented against
a colored background (without a symbol). Here you will find addi-
tional information on correct use of the meter or useful tips.
Instructions that require you to actively do something also appear
on a blue background.
ccxs_user_EN.book Seite 13 Mittwoch, 3. März 2010 1:45 13
The CoaguChek XS Meter
14
The CoaguChek XS Meter
G
D
A
B
E
C
F
I
H
H
ccxs_user_EN.book Seite 14 Mittwoch, 3. März 2010 1:45 13
The CoaguChek XS Meter
15
English
The CoaguChek XS meter
A Display
Shows results, information,
symbols and results recalled
from memory.
B M (memory) button
Press this button to recall
results from memory and to
change the meter settings.
You also press this button
to confirm the code number
displayed before each test.
C On/Off button
Press this button to turn the
meter on or off.
D Test strip guide cover
Remove this cover to clean
the test strip guide.
E Test strip guide
Insert the test strip here.
F Battery compartment
cover
Covers the battery compart-
ment (4 1.5 V alkali-manga-
nese batteries type AAA).
G Code chip slot
Insert the code chip here.
H Set button
Press this button to enter or
change the meter settings.
I Infrared window
You can transfer data stored
in memory via this infrared
interface.
ccxs_user_EN.book Seite 15 Mittwoch, 3. März 2010 1:45 13
The CoaguChek XS Meter
16
Batteries
To save power, the CoaguChek XS meter automatically turns itself off
after 3 minutes unless a button has been pressed or a new test strip has
been inserted. When the meter turns itself off, all results obtained up to
that point remain in memory. The meter briefly indicates the battery
power level when it is turned on. The battery symbol is divided into four
segments which correspond to the battery power level.
Results are retained in memory together with the date and time even
when no batteries are inserted. All other settings are retained, too.
Think of the environment. Dispose of used batteries sensibly.
When replacing the batteries you must insert the new batteries within
one minute of removing the old ones, to keep the date and time settings.
If you take longer than this, you must re-enter the date and time. Use only
alkali-manganese batteries type AAA.
The CoaguChek XS meter is shipped with high quality batteries. Since
quality varies among different battery brands, Roche recommends
replacing your used batteries with new batteries of a similar high quality
and not mixing different brands. Do not mix new and already used
batteries.
Please be aware that battery life may be affected by many factors, such
as battery quality, operating conditions (e.g. ambient temperature),
frequency of use, and test duration.
WARNING
Do not throw batteries onto a fire. They may explode!
ccxs_user_EN.book Seite 16 Mittwoch, 3. März 2010 1:45 13
Starting Up
17
English
Starting Up
Before using the meter for the first time, carry out the following steps:
1 Insert the batteries
2 Insert the code chip (you can also carry out this step immediately
before testing)
3 Set the current date and time
4 Select the coagulation unit in which the results are to be displayed
ccxs_user_EN.book Seite 17 Mittwoch, 3. März 2010 1:45 13
Starting Up
18
Inserting the batteries
1 With the meter turned off,
turn it over.
2 Gently press the tab on the
battery compartment cover
towards the centre of the
meter and lift off the cover.
3 Insert the four batteries in
the battery compartment as
indicated. Pay attention to
the positions of “+” (top of
battery) and “–” (flat end).
Once you have inserted the
batteries, the meter turns
itself on after about
5 seconds.
Only use alkaline manganese
batteries (1.5 V, AAA). Think
of the environment. Dispose
of used batteries sensibly.
ccxs_user_EN.book Seite 18 Mittwoch, 3. März 2010 1:45 13
Starting Up
19
English
4 Close the battery compart-
ment.
5 Check that all of the display
symbols are displayed prop-
erly. Results may be misread
if a segment is missing.
Tip
If the display check when you
turn on the meter is too quick,
you can put it “on hold” by
pressing and holding down
the On/Off button the
next time you turn the meter
on.
For as long as you keep the
button pressed, the display
remains frozen.
code
pm
am
mem
error
set
%Q
Sec
INR
28:88 88
--
8888
888
QC C
18:88
ccxs_user_EN.book Seite 19 Mittwoch, 3. März 2010 1:45 13
Starting Up
20
6 On the next screen that
appears check that the date
and time are correctly set.
If these have not yet been set
or if they have been lost
(because the batteries were
removed from the meter for
more than one minute), your
CoaguChek XS meter now
automatically enters Set
Mode. After you have set the
date and time, the meter
enters Test Mode. If you wish
to enter any further settings,
turn to the chapter Setting up
the meter (Set Mode) starting
on page 26.
7 Turn the meter off again
when the settings are
correctly entered.
11:23 10
--
0312
set
00
--
1231
ccxs_user_EN.book Seite 20 Mittwoch, 3. März 2010 1:45 13
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114
  • Page 115 115
  • Page 116 116
  • Page 117 117
  • Page 118 118
  • Page 119 119
  • Page 120 120
  • Page 121 121
  • Page 122 122
  • Page 123 123
  • Page 124 124
  • Page 125 125
  • Page 126 126
  • Page 127 127
  • Page 128 128
  • Page 129 129
  • Page 130 130
  • Page 131 131
  • Page 132 132
  • Page 133 133
  • Page 134 134
  • Page 135 135
  • Page 136 136
  • Page 137 137
  • Page 138 138
  • Page 139 139
  • Page 140 140
  • Page 141 141
  • Page 142 142
  • Page 143 143
  • Page 144 144
  • Page 145 145
  • Page 146 146
  • Page 147 147
  • Page 148 148
  • Page 149 149
  • Page 150 150
  • Page 151 151
  • Page 152 152
  • Page 153 153
  • Page 154 154
  • Page 155 155
  • Page 156 156
  • Page 157 157
  • Page 158 158
  • Page 159 159
  • Page 160 160
  • Page 161 161
  • Page 162 162
  • Page 163 163
  • Page 164 164
  • Page 165 165
  • Page 166 166
  • Page 167 167
  • Page 168 168
  • Page 169 169
  • Page 170 170
  • Page 171 171
  • Page 172 172
  • Page 173 173
  • Page 174 174
  • Page 175 175
  • Page 176 176
  • Page 177 177
  • Page 178 178
  • Page 179 179
  • Page 180 180
  • Page 181 181
  • Page 182 182
  • Page 183 183
  • Page 184 184
  • Page 185 185
  • Page 186 186
  • Page 187 187
  • Page 188 188
  • Page 189 189
  • Page 190 190
  • Page 191 191
  • Page 192 192
  • Page 193 193
  • Page 194 194
  • Page 195 195
  • Page 196 196
  • Page 197 197
  • Page 198 198
  • Page 199 199
  • Page 200 200
  • Page 201 201
  • Page 202 202
  • Page 203 203
  • Page 204 204
  • Page 205 205
  • Page 206 206
  • Page 207 207
  • Page 208 208
  • Page 209 209
  • Page 210 210
  • Page 211 211
  • Page 212 212
  • Page 213 213
  • Page 214 214
  • Page 215 215
  • Page 216 216
  • Page 217 217
  • Page 218 218
  • Page 219 219
  • Page 220 220
  • Page 221 221
  • Page 222 222

Roche CoaguChek XS Manuel utilisateur

Catégorie
Mesure, test
Taper
Manuel utilisateur

dans d''autres langues