ADA Undercounter Laboratory
Refrigerator
Installation and Operation
331315H01 Revision B 5/16/2023
IMPORTANT Read this instruction manual. Failure to follow the instructions in this manual can result in
damage to the unit, injury to operating personnel and poor equipment performance.
CAUTION All internal adjustments and maintenance must be performed by qualified service personnel.
Material in this manual is for informational purposes only. The contents and the product it describes are
subject to change without notice. Thermo Fisher Scientific makes no representations or warranties with
respect to this manual. In no event shall Thermo be held liable for any damages, direct or incidental, arising
from or related to the use of this manual.
© 2023 Thermo Fisher Scientific Inc. All rights reserved.
For your future reference and when contacting the factory, have the following information readily
available. It can be found on the dataplate attached to your unit.
Model Number:
Serial Number:
The following information, if available, is helpful for contacting the factory.
Date Purchased:
Purchase order number:
Source of Purchase: (manufacturer or specific agent/rep organization)
Contents
Models......................................................................... 1
Safety Information........................................................ 2
Alert Signals ............................................................ 2
Intended Use ............................................................... 3
ADA Undercounter Laboratory Refrigerator.................. 4
Product specification................................................ 4
Unpacking and Inspection............................................ 5
Unpacking................................................................ 5
Visible Loss or Damage............................................ 5
Concealed Loss or Damage ..................................... 5
Operating Conditions ................................................... 6
Installation.................................................................... 7
Selecting a Location................................................. 7
Leveling the Unit....................................................... 7
Electrical Connection................................................ 7
Reversing the Door Swing ........................................ 7
Be Advised............................................................... 8
Operation..................................................................... 9
Start-Up Procedure.................................................. 9
Restart Procedure.................................................... 9
How to Save Energy................................................. 9
Safety Tips ............................................................... 9
Maintenance ................................................................ 10
Cleaning the Unit...................................................... 10
Cleaning Interior/Exterior/Door Gaskets.................... 10
Cleaning the Condenser........................................... 10
Manual Defrost Procedure........................................ 10
Transportation and storage ...................................... 11
Troubleshooting........................................................... 12
Wiring Diagram ............................................................ 13
End of Life Care ........................................................... 14
Replacement Parts................................................... 14
Warranty ...................................................................... 15
Regulatory Compliance................................................ 16
Product Safety ......................................................... 16
Additional Regulations and Markings........................ 16
Contact Information .................................................... 17
ADA Undercounter Laboratory Refrigerator Models | 1
Models
This manual covers the installation and operational instructions for TSV05RPAA refrigerator.
2 | Safety Information ADA Undercounter Laboratory Refrigerator
Safety Information
It is your responsibility to understand the proper function and
operational characteristics of the unit. This instruction manual
should be thoroughly read and all operators must be trained
before attempting to place these units in service. Awareness
of the stated cautions and warnings, and compliance with
recommended operating parameters – together with
maintenance requirements – are important for safe and
satisfactory operation. The units should be used for its
intended application; alterations or modifications will void the
Warranty.
These products are not intended, nor can they be used, as a
sterile or patient connected device. In addition, these units are
not designed for use in Class I, II or III locations as defined by
the US National Electrical Code, unless otherwise noted.
Alert Signals
ON
OFF
Safety Alert: Important operating instructions.
To reduce the risk of injury or poor performance
of the unit. Read the user manual before putting
the equipment into operation.
WARNING: Indicates an immediately hazardous
situation, which if not avoided, will result in death
or serious injury.
CAUTION: Indicates an immediately hazardous
situation, which if not avoided, may result in
minor to moderate injury.
WARNING: Avoid contact with cold surfaces,
potential for cold burns or skin sticking to cold
surfaces.
WARNING: Use of this symbol involves risk of
electric shock.
Do not modify or use power supplies other than
OEM equipment. Connection of the power
supply may require a properly grounded
receptacle. Potential for electrical shock or
equipment damage exists if precautions are not
followed.
Note: Notes alert you to pertinent facts and conditions.
DANGER RISK OF CHILD ENTRAPMENT
Before you throw away your old refrigerator:
Take off doors.
Leave the shelves in the place so that
children may not easily climb inside.
WARNING: If the equipment is used in a manner
not specified by the manufacturer, the protection
provided by the equipment may be impaired.
WARNING: Units are charged with hydrocarbon
refrigerant (R600a). Only qualified service
personnel should service this unit.
WARNING: Risk of fire or explosion. Flammable
refrigerant used. To be repaired only by trained
service personnel. Do not puncture refrigerant
tubing.
WARNING: No equipment that uses an open
flame should be placed inside the refrigerator.
This will harm the unit, hamper functionality and
compromise your safety.
WARNING: Risk of fire or explosion. Flammable
refrigerant used. Do not use mechanical devices
to defrost the unit.
WARNING: This product can expose you to
chemicals including Nickel (Metallic) which is
known to the State of California to cause
cancer.For more information go to
www.P65Warnings.ca.gov
Note: Nickel is a component in all stainless steel
and some other metal components
ADA Undercounter Laboratory Refrigerator Intended Use | 3
Intended Use
This unit is intended for storage of non-flammable general
laboratory products.
The Refrigerator described in this manual is for professional
use only. The intended use of this product is in research, for
the storage of samples or inventory in the following
temperature ranges.
Refrigerators 33.8°F to 53.6°F (+1°C to +12°C)
This unit is not considered as medical device and have
therefore not been registered with a medical device regulatory
body (e.g. FDA): that is, this is not been evaluated for the
storage of samples for diagnostic use or for samples to be
re-introduced to the body.
Note: Do not store corrosive materials in these units. Any
damage which occurs due to storage of corrosives will not be
covered under warranty claims.
4 | ADA Undercounter Laboratory Refrigerator ADA Undercounter Laboratory Refrigerator
ADA Undercounter Laboratory
Refrigerator
This manual provides installation, operation and maintenance instructions for ADA under-counter laboratory Refrigerator
model..
Product specification
Note: Amps listed are at normal run mode, starting amps may be higher.
MODEL TSV05RPAA - 5.5 Cu. ft.
23.63" Wide, Under-counter Refrigerator
Features:
Adjustable wire shelves
White color
Key lock
Quiet, hermetically sealed
refrigeration compressors.
TSV05RPAA
Refrigerator Chamber Dimensions
H x W x D inches (cm)
27.75" x 20" x 17.5"
(70.5 x 50.8 x 44.5 cm)
Freezer Chamber Dimensions
H x W x D inches (cm) N/A
Exterior Dimensions
H x W x D inches (cm)
32.25" x 23.63" x 24"
(81.91 x 60.02 x 60.96 cm)
Electrical Characteristics
Volts/Hz, Amps 120/60, 0.8
Refrigerator Temperature Range
°F (°C)
33.8° to 53.6°
(1° to 12°)
Freezer Temperature Range
°F (°C) N/A
Net Weight lbs. (kg) 86 (39)
Shipping Weight lbs. (kg) 95 (43)
ADA Undercounter Laboratory Refrigerator Unpacking and Inspection | 5
Unpacking and Inspection
Unpacking
This should be inspected upon delivery. When received,
carefully examine for any shipping damage before unpacking.
If damage is discovered, the delivering carrier should both
specify and sign for the damage on your copy of the delivery
receipt.
Visible Loss or Damage
Note any external evidence of loss or damage on the freight
bill, or express receipt, and have it signed by the carrier’s
agent. Failure to adequately describe such external evidence
of loss or damage may result in the carrier’s refusing to honor
your damage claim. The form required to file such a claim will
be supplied by the carrier.
IMPORTANT: Failure to request an inspection of damage
within a few days after receipt of shipment absolves the
carrier from any liability for damage.You must call for a
damage inspection promptly.
Concealed Loss or
Damage
Concealed loss or damage refers to loss or damage, which
does not become apparent until the merchandise has been
unpacked and inspected. If either occur, make a written
request for the carrier’s agent within 15 days of the delivery
date; then file a claim with the carrier since the damage is the
carrier’s responsibility.
If you follow the above instructions carefully, we will guarantee
our support of your claim to be compensated for loss from
concealed damage.
DO NOT, FOR ANY REASON RETURN THIS UNIT WITHOUT
FIRST OBTAINING AUTHORIZATION.
CAUTION: DO NOT REMOVE, under any
circumstance, the grounding prongs from the
3-prong power cord supplied with all units.
CAUTION: DO NOT USE electrical extension
cords that may result in voltage loss and possible
hazardous operation.
6 | Operating Conditions ADA Undercounter Laboratory Refrigerator
Operating Conditions
The unit described in this manual is classified for use as
stationary equipment in a Pollution Degree 2 and Over voltage
Category II environment.
These units are designed to operate under the following
environmental conditions:
Indoor use
Altitude up to 2,000 m (6512 feet)
Maximum relative humidity 65%, non-condensing
Temperatures: 59°F to 90°F (15°C to 32°C)
Main supply voltage fluctuations should not exceed by
±10% of the nominal voltage.
ADA Undercounter Laboratory Refrigerator Installation | 7
Installation
Selecting a Location
Choose a location for the TSV05RPAA (Under-counter
Refrigerator) that provides a clearance of 4 inches. on each
side.
Appropriate electrical power must be available. Locate the
refrigerator within eight feet of the power outlet so that no
extension cord is required. Attach the refrigerator to the
facility’s electrical supply as directed by the National Electrical
Code.
Leveling the Unit
This refrigerator must be in level, in order to provide adequate
condensate drainage as well as proper door alignment and
operation. The refrigerator should be in its final operating
location and set so that it is firmly positioned on the floor.
Electrical Connection
Determine the total amount of current presently being used by
other apparatus connected to the circuit that will be used by
this refrigerator. It is critical that this added current demand
and other equipment on this circuit not exceed the rating of
the fuse or circuit breaker in use.
The frequency and nominal voltage requirements for the unit
are specified on the data plate. Supply this unit with an
electrical source that meets these requirements only. Low line
voltage is often the cause of service complaints. With the unit
running, check that the line voltage is within ±10% of that
specified on the data plate.
Unit must be connected to a grounded outlet matching the
nameplate and / or the information furnished in this manual. If
you are not sure about the outlet, you should contact a
qualified electrician for assistance.
The Unit should always be connected to a dedicated power
source.
Reversing the Door Swing
1. Unscrew the lower hinge and take off the door.
2. Push the kick plate to the right in the new position.
3. Undo the upper hinge fastener and fix it to the opposite
side of the housing, where you have already removed
and replaced the fastener with the hole to the opposite
side.
4. Put the door on the upper hinge fastener.
5. Fix the lower hinge in the new position.
6. Unscrew the handle and place it on the opposite side.
7. Some screws are covered with caps. Take off these caps
before undoing the screws.
8. Check the door gasket. It should not be creased and
should seal well.
CAUTION: Be sure the voltage supplied to the
refrigerator is equal to that specified on the data
plate.
WARNING: For personal safety, this unit must
be properly grounded before use.
WARNING: DO NOT REMOVE, under any
circumstance, the grounding prongs from the
3-prong power cord supplied with all units.
WARNING: DO NOT USE electrical extension
cords, it may result in voltage loss and possible
hazardous operation.
8 | Installation ADA Undercounter Laboratory Refrigerator
Be Advised
WARNING: Storage by user of any materials in
the product that may cause a deterioration of the
product shall be deemed to constitute abnormal
and improper usage of the product for purposes
of this warranty.
WARNING: Before connecting the final power
supply, check the electrical characteristics of the
unit nameplate to see that it is in agreement with
the power supplied. In addition, power should be
wired to the unit according to the electrical
schematic and all applicable codes. Only
qualified electricians should work on the
electrical portion of any unit installation.
CAUTION: Solutions used to clean coils or
neutralize bacteria growth must not be corrosive
to metals (enamel-coated steel) and materials
used in the maintenance of this equipment—
damage can result: use a soft cloth and warm
water to clean
ADA Undercounter Laboratory Refrigerator Operation | 9
Operation
Start-Up Procedure
Turn the knob clockwise from the STOP (0) position toward
the higher numbers for colder temperatures. Higher settings
on the knob (higher numbers) mean lower temperatures
(colder) in the refrigerator.
On the coldest settings, temperatures in some spots inside
the cabinet may drop below 32°F. Use higher thermostat
settings only when required or recommended.
To check chamber temperature, place a dial thermometer on
a shelf in the center of the chamber. Initially, rotate the
temperature control knob to an arbitrary setting. Allow
approximately 2 hours for the temperature to initially stabilize.
Check the temperature and compare with the dial setting.
Adjust dial further to reach the desired operating temperature.
After chamber initially stabilizes, allow 1/2 to 1 hour for the
chamber temperature to stabilize after subsequent
temperature adjustments.
If the room thermostat is turned below 60ºF at night, consider
setting the temperature control one step colder. It should be
left at this setting for the nighttime period; return temperature
control to original setting when the room thermostat is
returned to its normal setting.
Restart Procedure
If unit is unplugged or turned off, allow 3 minutes before
restarting or plugging it back in.
How to Save Energy
Be sure to follow location suggestions as mentioned in
the Installation.
Wipe moisture from glassware or other materials before
placing them in a unit.
Don't overcrowd the unit. Too many items can increase
electrical energy demand in order to keep everything
cool.
Close the door as soon as possible in hot, humid
weather.
Make certain that the door is closed tightly.
As soon as frost has accumulated to 1/4", defrost.
Keep containers covered, when possible, to reduce
moisture buildup.
Set operating temperature no colder than necessary for
the items being refrigerated.
Safety Tips
CAUTION: Do not use in the presence of
flammable or combustible materials or explosive
gases. Fire or explosion could result, causing
death or severe injury.
After a unit is in operation, do not touch the cold
surfaces, particularly when hands are damp. Skin
may adhere to the cold surfaces.
We recommend handling samples by wearing
gloves to avoid frost bite.
Never disconnect your unit by pulling the power
cord. Always grip the plug securely and pull
straight out from the outlet.
Do not use a power cord that shows cracks or
abrasions. Have a qualified electrician repair or
replace damaged cords immediately.
10 | Maintenance ADA Undercounter Laboratory Refrigerator
Maintenance
Cleaning the Unit
Before beginning cleaning unit follow these instruction listed
below:
Disconnect power cord from its outlet.
Set the temperature control to the OFF position.
The unit designs permit easy and rapid cleaning and
should not take more than a few minutes. Remember to
wear protective gloves to prevent frost bite, especially
when removing items from the unit.
Do not use abrasive scouring powders, waxes, solvents,
furniture polish, undiluted detergents or cleansers
containing petroleum products on the surfaces of units.
Cleaning Interior/Exterior/
Door Gaskets
A solution of mild soap/detergent and lukewarm water can be
used for cleaning the interior, exterior and door gaskets with a
soft, clean cloth. Rinse with clean water and dry thoroughly
before reconnecting and turning on the unit.
The cabinet interior should be cleaned frequently. Any spilled
liquid should be wiped off immediately since stains resulting
from some spills could be permanent if not quickly removed.
The most convenient time to clean the interior is after
defrosting.
Cleaning the Condenser
With forced-fan vented units, remove the screws that mount
the grill to the unit. Pull temperature control knobs straight out
this will expose condenser for cleaning.
These surfaces may be warm to the touch. The condenser
should be cleaned before becoming clogged with dirt/dust.
Construction or other dirt causing environments may
significantly increase the required frequency. The condenser
should be cleaned at least once a year.
Manual Defrost Procedure
Use the following procedure to defrost manually:
Rotate the temperature control knob to the OFF setting
or zero position according to the model.
Disconnect the power cord from its outlet.
Remove contents of the unit. If practical, wrap contents
in paper and then in a heavy blanket to maintain
temperature of items. Wear protective gloves to prevent
from frost bite when handling cold items.
Open door and allow free circulation of ambient air.
To speed the process, place pan of warm water inside
the chamber.
Wipe out the interior.
Replace the contents.
Reconnect the power cord to outlet and set temperature
control to desired operating temperature
CAUTION: When servicing the unit, disconnect
from the electrical power source.
CAUTION: Refer servicing to qualified
personnel.
CAUTION: Never clean around the condensers
with your fingers. Some surfaces are sharp.
CAUTION: Do not use any sharp instruments,
blade or scraper to remove ice and frost on
refrigerator surfaces because of the very real
danger of puncturing the cooling coil.
CAUTION: Do not use any electrical devices to
defrost the unit.
ADA Undercounter Laboratory Refrigerator Maintenance | 11
When frost accumulates to 1/4" or more, the operating
efficiency of the unit will be affected.
All laboratory refrigerators feature automatic defrosting and in
normal conditions no manual defrosting is required. However,
in certain conditions (e.g. when the refrigerator is very full or
when any new items are stored in the refrigerator) frost may
form on the rear wall of the equipment as the compressor
must run for long periods of time. As automatic defrosting
takes place while the compressor is not running, the
refrigerator may not defrost completely before the
compressor restarts. In these instances, some manual
defrosting may be required.
Before manually defrosting the cabinet, make sure the power
lead is disconnected from the mains. With cold wall
refrigerator, condensation may form on the interior of the
cabinet and may require periodic wipe- down with an
absorbent cloth.
In high usage and/or high ambient scenarios, excessive ice
may form on the inner back wall of the refrigerator and in the
worst case or sudden power cut-off event, drain pan might
overflow. Hence it is recommended to defrost frequently and
collect any build up ice from inside in a container and safely
dispose in the sink. If the drain pan has over flowed, empty
the pan and dry up the wet regions around before next
restart.
Transportation and
storage
Use all required local facility procedures and PPE while
transporting and storing the product.
12 | Troubleshooting ADA Undercounter Laboratory Refrigerator
Troubleshooting
This section is a guide to troubleshoot general operational problems.
Symptom Probable Cause Action
Unit does Not Run Unit Unplugged
Blown fuse or tripped circuit breaker.
Temperature control knob is turned to OFF.
Plug in Unit
Check fuse or circuit breaker at breaker box.
Turn temperature control knob to ON.
Runs Continuously Frost build up on refrigeration coils.
Dust or lint buildup on condenser.
Door gasket worn out.
Temperature too low
Ambient temperature over 109ºF (43ºC)
Door opening frequency is high.
Defrost unit
Clean the condenser with a dry brush or
vacuum
A leak around the door gasket will allow cold
air to escape. This causes unit to work harder
than necessary to maintain cold temperatures.
Re-seat or replace the gasket if worn
This may cause unit to run continuously.
Check optimum running temperature
Is the ambient air over 109ºF (43ºC), or the
units located close to heat sources? If
possible, move to a different location.
An unusually high frequency of door openings
and closings can increase operating load. Unit
will stabilize as these are decreased.
Noise problems Contents of unit set too close and rattling
against each other.
Hissing or gurgling noise is caused by
refrigerating fluid circulating.
Unit is not level on floor
Fan noise
Rearrange contents as needed.
This is normal
Check the level of unit
Airflow can cause this. This is normal
ADA Undercounter Laboratory Refrigerator Wiring Diagram | 13
Wiring Diagram
Figure 1. Wiring diagram for TSV05RPAA
%.
:+
*1
32:(5&25'
//
1
,/
%8%.
%15'
*1
&5
5
0
%8
5'
%.
4
*1
:+
7+(50267$7
%8
:+
/$03
/$036:,7&+
429(5/2$'3527(&725
567$575(/$<
*125*1:,7+<(//2:75$&(5
6
6+(//*5281'
:,5,1*',$*5$0
14 | End of Life Care ADA Undercounter Laboratory Refrigerator
End of Life Care
Be sure to follow local regulations when disposing of an old
unit. Some suggestions are listed below:
1. Remove items and defrost unit. Be sure to clean up any
biological safety hazards.
2. Remove the cabinet door to help prevent entrapment
inside of a unit.
3. Have a certified technician remove the refrigerant and
compressor, then drain the compressor and oil from the
system. Disposal of the unit should be done by following
local regulations.
Replacement Parts
To obtain replacement parts information and pricing, please
call the Customer Service Department at 800-438-4851 and
have the unit’s model, serial and code numbers available.
This information is located on data plates on the rear of the
unit or on the upper left side of the interiors.
ADA Undercounter Laboratory Refrigerator Warranty | 15
Warranty
13 Month Limited Warranty
This product is warranted to be free of defects in materials and workmanship for one (13) months from the first to occur of (i)
the date the product is sold by the manufacturer or (ii) the date the product is purchased by the original retail customer (the
“Commencement Date”). Except as expressly stated above, the MANUFACTURER MAKES NO OTHER WARRANTY,
EXPRESSED OR IMPLIED, WITH RESPECT TO THE PRODUCTS AND EXPRESSLY DISCLAIMS ANY AND ALL
WARRANTIES, INCLUDING BUT NOT LIMITED TO, WARRANTIES OF DESIGN, MERCHANT ABILITY AND FITNESS FOR A
PARTICULAR PURPOSE.
An authorized representative of the manufacturer must perform all warranty inspections. In the event of a defect covered by the
warranty, we shall, as our sole obligation and exclusive remedy, provide free replacement parts to remedy the defective
product. In addition, for products sold within the continental United States or Canada, the manufacturer shall provide free labor
to repair the products with the replacement parts, but only for a period of ninety (90) days from the Commencement Date.
The warranty provided hereunder shall be null and void and without further force or effect if there is any (i) repair made to the
product by a party other than the manufacturer or its duly authorized service representative, (ii) misuse (including use
inconsistent with written operating instructions for the product), mishandling, contamination, overheating, modification or
alteration of the product by any customer or third party or (iii) use of replacement parts that are obtained from a party who is not
an authorized dealer of products.
Heating elements, because of their susceptibility to overheating and contamination, must be returned to the factory and if, upon
inspection, it is concluded that failure is due to factors other than excessive high temperature or contamination, the
manufacturer will provide warranty replacement. As a condition to the return of any product, or any constituent part thereof, to
the factory, it shall be sent prepaid and a prior written authorization from the manufacturer assigning a Return Materials Number
to the product or part shall be obtained.
IN NO EVENT SHALL THE MANUFACTURER BE LIABLE TO ANY PARTY FOR ANY DIRECT, INDIRECT, SPECIAL,
INCIDENTAL, OR CONSEQUENTIAL DAMAGES, OR FOR ANY DAMAGES RESULTING FROM LOSS OF USE OR PROFITS,
ANTICIPATED OR OTHERWISE, ARISING OUT OF OR IN CONNECTION WITH THE SALE, USE OR PERFORMANCE OF ANY
PRODUCTS, WHETHER SUCH CLAIM IS BASED ON CONTRACT, TORT (INCLUDING NEGLIGENCE), ANY THEORY OF
STRICT LIABILITY OR REGULATORY ACTION.
16 | Regulatory Compliance ADA Undercounter Laboratory Refrigerator
Regulatory Compliance
Product Safety
Product Testing
This product family has been tested to applicable product safety standards by Underwriters Laboratories
(UL), which is a Nationally Recognized Test Laboratory (NRTL). The product bears the following mark
Hydrocarbon Refrigerants
According to U.S. Code of Federal Regulation 40 Part 82, this refrigerator employs the natural hydrocarbon refrigerant R600a.
Because of the nature of hydrocarbon refrigerants, for mechanical repair, such as recharge or compressor replacement, should
only be carried out by a certified refrigeration technician. The safety of this equipment is listed by Underwriter Laboratory (UL)
under Standard UL471, section SB – “natural refrigerant”.
Additional Regulations and Markings
This product is not marked with a CE marking, as it does not operate in the voltage range to be sold to the EU Member States
or European Economic area (EEA). Please reach out to the manufacturer for questions regarding any additional regulatory
conformity.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53