Joerns Hoyer® Advance Le manuel du propriétaire

Taper
Le manuel du propriétaire

Ce manuel convient également à

User Instruction Manual
Hoyer® Advance
To avoid injury, read user manual prior to use.
Manuel de l’utilisateur
Hoyer® Advance
Afin d’éviter tout accident, veuillez lire
attentivement la notice avant utilisation.
Manual de Instrucciones
Hoyer® Advance
Para evitar posibles daños, lea
previamente el manual de usuario.
Benutzerhandbuch
Hoyer® Advance
Um Verletzungen zu vermeiden, lesen Sie bitte die
Gebrauchsanweisung vor der ersten Benutzung.
Hoyer®
Advance
2
English
Contents
1. Hoyer Advance Patient Lift ............................................................................................. 3
2. Introduction: The Hoyer Advance Patient Lift ................................................................. 4
3. Assembly and Commissioning Instructions .................................................................... 5
4. Disassembly .................................................................................................................... 7
5. Safety Precautions ........................................................................................................ 10
6. Control Box & Handset ................................................................................................. 12
7. Operating Instructions ................................................................................................... 13
8. Maintenance Schedule .................................................................................................. 16
9. Technical Specifications ................................................................................................ 18
10. Warranty ........................................................................................................................ 21
Hoyer®
Advance
3
English
1. The Hoyer Advance Patient Lift
Hoyer Advance lift (ready for use)
Boom
Cradle/Spreader Bar
Electric Actuator
Spreader Bar Retaining Cradle
Mast Tightening Knob
Braked Caster
Base Casting
Push Pad
Push Handle
Battery/
Control Pack
Leg Restraining Cord
Leg Extrusion
Hoyer Advance lift (storage position)
Hoyer®
Advance
4
English
2. Introduction: The Hoyer Advance Patient Lift
The Hoyer Advance is available in two versions, hydraulically operated and electrically operated.
This manual covers both Advance models. Each lift is fully assembled, load tested and certified
before being packed/shipped.
The packing consists of a strong, purpose built carton and is used for both export and domestic
markets to ensure the safe arrival of the lift. A number of documents are supplied in a wallet
packed with each lift and should be kept safely for future reference.
• TEST CERTIFICATE • USER MANUAL
• DEALER GUARANTEE CARD • CUSTOMER SATISFACTION CARD
The TEST CERTIFICATE is an important document and should be kept for reference purposes. To
properly maintain your lift please refer to the maintenance schedule included in this document. If
you are at all unsure what your country’s servicing requirements are, please check with your dealer
and/or a local government agency.
The CE mark:
Statement of Intended Use
The intended use of this lifting device is for the safe lifting and transfer of an individual from one
resting surface to another (such as a bed to a wheelchair). Joerns Healthcare recommends that the
transfer of a patient is fully risk assessed and conducted safely over a short distance only.
The Hoyer Advance is suitable for patients in the SITTING, SITTING/RECUMBENT and
RECUMBENT positions.
The Hoyer Advance is designed to support and promote safe patient handling and transfer for both
the patient and carer. It can be folded without the use of tools enabling it to be easily stored in envi-
ronments where space is limited. The Hoyer Advance can also be separated into two sections to
facilitate onward transportation in a vehicle with a suitably sized boot.
The Hoyer Advance patient lift incorporates a 6-point spreader bar and is designed to be used in
conjunction with the Hoyer range of slings. The examples of slings suitable for use with this device
are listed as follows:
• Hoyer Quickfit • Hoyer Long Seat
• Hoyer Full Back • Hoyer Access sling
• Hoyer Quickfit Deluxe • Hoyer Silkfit
Expected Service Life
Hoyer Professional lifts are designed and tested for a minimum service life of ten (10) years,
subject to the use and maintenance procedures stated in this manual. Use, other than in accordance
with these instructions, may compromise service life.
The Hoyer Advance carries the CE mark and complies with the following
EC directives:
• Medical Device Directive (93/42/EEC)
• EMC Directive (89/336/EEC)
• Low Voltage Directive (73/23/EEC)
Hoyer®
Advance
5
English
3. Assembly and Commissioning Instructions
Carton Contents
Place the carton in a clear working area and open carefully. The carton contains:
HOYER ADVANCE
WALLET CONTAINING DOCUMENTS
HAND CONTROL (ELECTRIC ONLY)
BATTERY PACK (ELECTRIC ONLY)
BATTERY CHARGING LEAD (ELECTRIC ONLY)
HYDRAULIC UNIT (HYDRAULIC ONLY)
SAFETY NOTE: The Hoyer Advance is heavy and will need to be lifted with care. You may need
assistance to lift the Hoyer Advance from the carton.
Assembly Instructions
1. Remove all the parts from the carton and place on the floor, taking care to protect the finish
from damage.
Lift the Advance carefully from the carton. Stand the lift in the upright position (in a triangular stance).
2. Release the leg and mast assembly by unfastening the restraining strap from the wheels. This
strap will retract neatly into the top of the mast section.
Hoyer®
Advance
6
English
3. Lay the Advance flat on the floor (ensure rear casters are securely locked).
4. Lift the mast assembly upright (using the top of the push handle), and drop the mast into the slot (1),
tighten mast lock knobs (2), release braked casters (3).
SAFETY NOTE: Avoid trapping fingers. Keep fingers away from the end of the mast when inserting
into the mast slot. Ensure the mast is fully engaged before tightening the mast locking device.
5. Open legs fully. ENSURE leg-locking knob engages flush against the base casting and the legs
will no longer move freely. The legs will then only operate by the use of the foot pedal mechanism
located at the rear of the base.
WARNING
If operation of the foot pedal DOES NOT adjust the legs, DO NOT use the lift.
Check the leg locking knob is fully engaged.
If the leg locking knob will not fully engage and operating the foot pedal does not adjust the
legs, DO NOT use the lift and contact your authorised Joerns Healthcare service provider for
examination and repair.
Hoyer®
Advance
7
English
6. Release the spreader bar from the retaining cradle. The lift is now ready for operation.
4. Disassembly
1. Lower the mast to the lowest possible position, then insert spreader bar into retaining cradle.
2. Release leg-locking device (pulling it away from the cast base), close legs tightly together.
Hoyer®
Advance
8
English
3. Loosen the mast locking device sufficiently to allow the mast to be lifted upwards. Lift the
mast fully until the upper stop is reached. Carefully fold the mast towards the legs. Use the push
handle to assist this action.
CAUTION
When commencing folding of the mast, if resistance is felt, DO NOT force the mast; lift the
mast upwards again until the upper stop is reached and attempt the fold operation again.
NOTE: If you wish to separate the mast and boom completely from the base and legs, undo the
mast locking device fully (remove) and lift the mast and boom completely free from the base. Be
careful not to lose the mast lock device’s knob and location spindle.
4. Lift the folded Advance upright (using the legs and push handle as a guide) being careful that
the mast and boom stay together as you tilt upright.
Hoyer®
Advance
9
English
5. Pull the restraining strap sharply upwards and extend over the wheels. (This will keep the legs,
mast and boom neatly together during transit and storage).
NOTE: Care should be exercised when folding/unfolding an Advance as there is a possible danger
of pinching the fingers etc. Please follow the instructions carefully and ask for assistance if you are
unsure of the correct procedure.
ALWAYS CHECK THE FOLLOWING BEFORE OPERATION:
The legs are fully engaged from the stowed position (see assembly instructions, step 5)
The mast is fully locked into position (see assembly instructions, step 4).
The legs open and close satisfactorily when the foot pedal is operated.
Fit the power pack to the bracket on the mast and ensure the retaining latch is fully engaged
and ‘clicks’ into place.
The red emergency stop button, located on the control box, is in the OUT position.
Depress the RAISE and LOWER buttons on the hand control and confirm the boom operates
correctly.
HYDRAULIC MODELS ONLY:
Close the hydraulic unit release valve by turning the knob clockwise on the unit.
NOTE: The release valve requires only minimal tightening to operate and should only be closed finger
tight. DO NOT apply excessive force to the valve knob as this will result in damage to the valve.
Pump the handle of the hydraulic unit and confirm the ram raises the boom.
Open the release valve fully anti-clockwise and check the boom descends. An unloaded boom
will not come down under its own weight. It may be necessary to apply some pressure to the
boom before it will descend.
NOTE: The release valve is fully open and encounters a positive end stop in less than two full
turns of the knob.
WARNING
When lowering a patient, open the valve very slowly. Rapidly opening the valve may cause
the boom to lower quickly which may cause injury to the patient.
Hoyer®
Advance
10
English
5. Safety Precautions
Please read and follow the safety precautions listed below. The operation and use of Hoyer
patient lifts is simple and straightforward. Following these few basic safety precautions will make
lifting operations easy and trouble free.
READ AND UNDERSTAND THE USER INSTRUCTION MANUAL BEFORE USING YOUR
“ADVANCE LIFT
WARNING: Important safety information for hazards that might cause serious injury.
CAUTION: Information for preventing damage to the product.
NOTE: Information to which you should pay special attention.
WARNING
ALWAYS plan your lifting operations before commencing.
ALWAYS carry out the DAILY CHECK LIST (page 16) before using the lift.
ALWAYS familiarise yourself with the operating control and safety features of a lift
before lifting a patient.
DO NOT use a sling unless it is recommended for use with the lift.
ALWAYS check the sling is suitable for the particular patient and is of the correct size
and capacity.
NEVER use a sling, which is frayed or damaged.
ALWAYS fit the sling according to the instructions provided (user instructions).
ALWAYS check the safe working load of the lift is suitable for the weight of the patient.
ALWAYS carry out lifting operations according to the instructions in the user manual.
NEVER disconnect or bypass a control or safety feature because it seems easier to
operate the lift.
DO NOT lift a patient with the caster brakes on. Always let the lift find the correct center
of gravity.
DO NOT attempt to maneuver the lift by pushing on the mast, boom or patient.
ALWAYS maneuver the lift with the handle provided.
ALWAYS lower the patient to the lowest comfortable position before transfers.
DO NOT push a loaded lift at speeds, which exceed a slow walking pace
(3 Kilometers/hour 0.8 meters/second).
DO NOT push the lift over uneven or rough ground, particularly when loaded.
DO NOT attempt to push/pull a loaded lift over a floor obstruction, which the casters
are unable to ride over.
NEVER force an operating/safety control. All controls are easy to use and do not require
excessive force.
The MINIMUM load required to manually lower the hoist using the mechanical emergency
down function is 30 kg.
DO NOT park a loaded lift on ANY sloping surface.
DO NOT use electric lifts in a shower.
Hoyer®
Advance
11
English
DO NOT charge an electric lift in a bathroom or shower room.
DO NOT lift a patient unless you are trained and competent to do so.
YOUR lift is for patient lifting. DO NOT use it, or allow it to be used, for any other purpose.
DO NOT bump the lift down steps, loaded or unloaded.
DO NOT attempt to negotiate a loaded lift on a slope, which exceeds 1:12 (approximately
5 degrees).
DO NOT attempt to negotiate a slope without a second helper being present.
ALWAYS check the hoist is not charging before moving as the electrical connection
may be damaged.
Hoyer®
Advance
12
English
6. Control Box & Handset
Emergency Stop Button
Emergency Raise & Lower
Redundant Controls
LCD
Information
Screen
Handset
Connection
Socket
Actuator
Connection
Socket
Handset
Connection
Plug
Raise & Lower
Buttons
Handset
Actuator
Detachable Battery Pack
Hoyer®
Advance
13
English
7. Operating Instructions
1. Leg adjustment
The legs on the Hoyer Advance are adjustable for width. The legs can be opened to enable
access around armchairs or wheelchairs. For transferring and negotiating doorways and
passages, the lift legs should be securely locked in the closed position.
To operate adjustment, the leg adjustment pedal, located at the rear of the base, is pressed
down on the right hand side to open the legs outwards and is pressed down on the left hand side
to close the legs. The adjustment can be carried out whilst the patient is in the lift.
WARNING
Never perform a lift/transfer with the legs in the closed/transport position (front casters
touching). The closed/transport position is for storage and transport only.
2. Casters and braking
The lift has two braked casters, which can be applied for parking. When lifting, the casters should
be left free and un-braked, so that the lift will then be able to move to the center of gravity of the
lift. DO NOT apply the brakes. If the brakes are applied, the patient may swing to the center of
gravity and this may prove distressing and uncomfortable.
3. Raising and lowering the boom
The movement of the boom is achieved by a powerful electric actuator, which is controlled by
a simple hand control unit. The hand control has two buttons with directional arrows RAISE
and LOWER. The actuator stops automatically at the limit of travel in both directions. The hand
control plugs into a socket at the base of the control box. There is a hook on the rear of the hand
control, which allows it to be “parked” on the mast or boom when not in use.
4. Emergency stop
The red Emergency Stop Button is located on the rear of the control box and is activated by pressing
in. This will cut all power to the lift and only be reset by twisting the button counterclockwise and
releasing.
5. Redundant controls
The standard control box is fitted with emergency LOWER/RAISE buttons located underneath the
emergency stop button. They can be operated by inserting the tip of a ball point pen into the
buttons highlighted “lower/raise”. This will raise or lower the boom should the hand control fail.
6. Mechanical Emergency Down
In the case of a complete electrical failure, the electric actuator is fitted with a mechanical lowering
device (RED BOSS). This will only operate when the lift is under load. The device must be pulled
upwards to activate and a slow decent will commence.
NOTE: The MINIMUM load required to manually lower the hoist using the mechanical emergency
down function is 30 kg.
7. Batteries
The batteries are protected from deep discharge by a LOW VOLTAGE ALARM. This will sound
when the batteries need recharging and the hand control is being operated. It will not sound
independently of the hand control being operated. CAUTION: DO NOT IGNORE THIS WARNING
ALARM. Complete the lifting operation and place the lift on charge (see charging instructions).
Hoyer®
Advance
14
English
8. Raising and lowering the boom (hydraulic models)
WARNING
Never loosen or unscrew the retaining bolt of the hydraulic jack unless you are an authorized compe-
tent engineer. Never load whilst user needs to rotate hydraulic jack. If the user intends to rotate
the jack strongly, it may impact the patient or cause unforeseen danger to patient and/or caregiver.
The raising and lowering of the boom is achieved by a powerful hydraulic ram, which is operated
by two simple controls. The release valve, which is identified by a silver star shaped knob, and
the pump handle which is a long lever on the side of the hydraulic unit to raise the boom. The
valve is closed by gently turning the knob fully clockwise. When closed, pump the long handle
with smooth even strokes for maximum effect. The handle strokes from an upright position
through an arc of 90 degrees. Leave the handle in the upright position when not in use.
DO NOT force the handle beyond the upper or lower stops. The hydraulic unit can be rotated to
allow the handle to be used from either side of the lift. To lower the boom, turn the release valve
anti-clockwise. The release valve is progressive, the more it is opened the faster the descent.
The valve is restricted so even when fully open the descent is controlled. This facility allows for a
“hands free” descent. If the release valve is opened a fraction (a quarter turn) a very slow speed
of descent will allow the caregiver to work “hands free” while assisting or comforting the patient.
REMEMBER to close the release valve before commencing lifting operations. The release valve
only requires gentle pressure to open or close. DO NOT apply excessive force to the release
valve, either to close or to open. It is not necessary and will only damage the valve.
WARNING
When lowering a patient, open the valve very slowly. Rapidly opening the valve may cause
the boom to lower quickly which may cause injury to the patient.
9. Slings
The selected sling is attached to the spreader bar hooks. Each sling is supplied with instructions.
Please study the instruction guide for the lift and sling before using. Once the correct sling has been
selected for the patient, attach it to the spreader bar hooks.
NOTE: For detailed fitting instructions, please refer to the user guide supplied with each sling.
WARNING
Joerns Healthcare recommends slings are checked regularly and before use for fraying or
damage. DO NOT use slings that are worn or damaged.
WARNING
HOYER RECOMMENDS THE USE OF GENUINE HOYER PARTS. Hoyer slings and lifters
are not designed to be interchangeable with other manufacturer’s products. Using other
manufacturer’s products on Hoyer products is potentially unsafe and could result in
serious injury to patient and/or caregiver.
WARNING
Refer to maximum weight capacity of lift. Sling capacity is limited by the maximum
capacity of the lift.
Hoyer®
Advance
15
English
Charging Instructions
The batteries are located in the power pack and are charged through a socket at the base of the
control box.
1. Insert the power supply plug into the charge point at the base of the control box. The plug is
inserted with a straight push. DO NOT twist the plug in the socket.
2. Plug the charger mains lead/cord into a suitable mains/power outlet.
3. Charging is fully automatic. The LCD indicator indicates the status of the charging.
NOTE: Even if the charger is left plugged in for extended periods it will not allow the batteries
to overcharge.
a) Green Light - Indicates main power is on.
b) Yellow Light - Indicates battery is charging.
c) Battery will be fully charged when yellow light goes off.
NOTE: Charging will take up to 4 hours.
It is recommended that the battery be charged immediately up on receipt.
4. To return the lift to service, pull the charger socket out and disengage from the mains supply.
WARNING
The battery, charger, hand control & control box are NOT to be opened by unauthorized
personnel. (Contact your distributor for warranty and repairs).
DO NOT touch battery/charger terminals.
KEEP the batteries fully charged. Place the battery on charge whenever it is not in use.
If it is more convenient to do so, place on charge every night. The charger will not allow
the batteries to overcharge.
NEVER run the batteries completely flat. As soon as the audible warning sounds,
complete the lifting operation in hand and place on charge.
To avoid possible permanent damage to the battery, the battery should be placed on
charge as soon as the display indicates the half empty battery symbol.
NEVER store the battery for long periods without regular charging throughout the storage
period.
ALWAYS make sure the mains power to the charger is switched off before connecting or
disconnecting the battery pack.
NEVER leave the battery pack connected with the charger mains power switched off.
DO NOT leave the charger switched on with the battery disconnected.
DO NOT use the lift while charging is taking place.
CHECK the lift is not charging before moving as the electrical connection may be
damaged.
NEVER disconnect the charger plug by pulling on the cable.
BE CAREFUL not to trip over the charge lead.
DO NOT charge an electric lift in a bathroom or shower room.
Hoyer®
Advance
16
English
8. Maintenance Schedule
All Hoyer products are designed for minimum maintenance, however some safety checks and
procedures are required. A schedule of DAILY tasks is detailed below. A list of replacement parts
is available upon request.
DAILY CHECK LIST
Joerns Healthcare strongly recommends the following checks be carried out on a daily basis and
before using lift.
- MAKE sure the lift moves freely on its casters.
- MAKE sure the spreader bar is free to rotate and swing. Check the spreader bar is firmly
attached to the boom.
- ENSURE the mast is fully engaged into the mast slot and the mast locking device is securely
tightened.
- EXAMINE the sling hooks on the spreader bar and side suspenders for excessive wear. If in
doubt - do not use.
- EXAMINE slings for fraying or other damage. DO NOT use any sling if damaged or if the sling
shows signs of wear.
- MAKE sure the legs open and close correctly.
- OPERATE the hand control or the hydraulic unit to confirm the boom raises and lowers
satisfactorily.
- CONFIRM the lift is not giving a low battery alarm when the hand control is operated. If the
alarm sounds DO NOT use and place on charge immediately.
- ON electric powered lifts check the operation of the emergency stop button.
- ON hydraulically operated lifts check for hydraulic fluid leakage. Any leakage should
be reported to a service engineer immediately and the lift should not be used until it
has been repaired.
MAINTENANCE, INSPECTION AND TEST
Joerns Healthcare recommends a thorough inspection and test of the Hoyer Advance lift and lifting
accessories; slings, etc. are carried out on a regular basis. Inspection frequency varies depending on
locality, so you must check with your dealer or local government agency as appropriate regarding
how often an inspection is required. The examination and test should be conducted according
to the recommendations and procedures provided in this manual. Whenever possible Joerns
Healthcare recommends maintenance, inspection and certified testing is carried out by authorized
service agent / dealers only.
NOTE: These recommendations are in compliance with the requirements of 1998 No2307 Health
and Safety: The Lifting Operations and Lifting Equipment Regulations 1998. This is a UK regulation.
Outside the UK please check your local country requirements.
Hoyer®
Advance
17
English
*THESE CHECKS SHOULD INCLUDE:
1. SPREADER BAR: Check the spreader bar for freedom of rotation and
ü ü
ü
swing. Check for wear on the central pivot. Check for firm attachment to
the boom.
2. BOOM: Check the attachment of the boom to the mast. Make sure there
ü ü ü
is only minimal side movement of the boom and the boom is free to rotate
on the boom bearing.
3. MAST: Check the operation of the mast-locking device. Make sure the
ü ü ü
mast fully engages into the socket. Check the bottom actuator or hydraulic
unit mounting.
4. ACTUATOR: (electric only): The actuator should require no maintenance X ü
other than checking for correct operation and listening for unusual noise.
5. CONTROL BOX (electric only): Check the function of the emergency stop
button. Inspect the hand control socket for correct fitting. Check functioning
of the hand control. Check the redundant controls and confirm they operate ü ü ü
as intended. Note: The MINIMUM load required to manually lower the hoist
using the emergency lower function is 30 kg.
6. BATTERIES: (electric only): The batteries are housed in the power pack
ü
and should not require maintenance other than the regular charging as
detailed in the charging instructions. Check that the connections remain clean.
7. HYDRAULIC UNIT: The hydraulic unit should require no maintenance other ü ü
than checking for correct operation and leakage of hydraulic fluid.
8. LEG ADJUSTMENT: Check the legs operate in both full extensions ü ü ü
(inward/outward).
9. CASTERS: Check all casters for firm attachment to the legs. Check for ü ü ü
free rotation of the caster and the wheels.
10. CLEANING: Clean with ordinary soap and water and/or any hard surface
X ü
disinfectant. Harsh chemical cleaners or abrasives should be avoided as
these may damage the surface finish of the lift. Avoid wetting any of the
electrical parts.
11. LOAD TEST: The load test should be carried out in accordance with the X
manufacturer’s test procedures.
12. BASE AND WHEELS: Ensure base is even and level ü ü ü
(all four wheels are on the floor).
13. SLINGS: Check for wear and fraying. ü ü ü
14. LUBRICATION: Oil pivot joints, including mast and boom connections, ü
pedal assembly, spreader bar joint (only if required).
15. FIXINGS: Check all nuts, bolts, screws and fasteners for excessive wear 
ü
and for tightness. Replace as required.
Before Use
Initially
Service
Intervals *
ü = Required X = Recommended
* 6 months in North America
Hoyer®
Advance
18
English
9. Technical Specifications
Specification Electric Hydraulic
Safe Working Load .....................................341 lbs. ........155Kgs ....................... 341 lbs ....... 155Kgs
Maximum Overall Length ........................51 inches ......1300mm ................ 49.2 inches ..... 1250mm
Minimum Overall Length .........................49 inches ......1250mm ................ 47.2 inches ..... 1200mm
Maximum Overall Height .........................73 inches ......1860mm ................ 72.6 inches ..... 1845mm
Minimum Overall Height .......................53.5 inches ......1360mm ................ 53.1 inches ..... 1350mm
Folded Dimensions
Height ...................................................17.1 inches ........ 450mm ................ 16.9 inches ....... 470mm
Depth ....................................................46.4 inches ...... 1180mm ................ 46.5 inches ..... 1230mm
Width ....................................................21.6 inches ........ 550mm ................ 23.0 inches ....... 550mm
Spreader Bar Max. Height ....................66.5 inches ......1690mm ................ 64.6 inches ..... 1640mm
Spreader Bar Min. Height (usable).......15.3 inches ........390mm ................ 15.4 inches ....... 390mm
Height at Maximum Reach ...................... 46 inches ......1170mm ................ 33.5 inches ....... 850mm
Reach at Maximum Height ................... 25.6 inches ........650mm ................ 28.0 inches ....... 710mm
Reach at Minimum Height .................... 13.7 inches ........350mm ................ 15.4 inches ....... 390mm
Maximum Reach* .................................32.8 inches ........835mm ................ 33.5 inches ....... 850mm
Turning Radius .....................................55.9 inches ...... 1420mm ................ 41.0 inches ..... 1040mm
Legs Open - External Width ................. 42.5 inches ......1080mm ................ 43.7 inches ..... 1110mm
Legs Open - Internal Width ..................39.3 inches ......1000mm ................ 45.7 inches ..... 1160mm
Legs Closed - External Width...............26.3 inches ........ 670mm ................ 25.2 inches ....... 640mm
Legs Closed - Internal Width ................ 22.4 inches ........ 570mm ................ 21.7 inches ....... 550mm
Overall Height of Legs ............................4.5 inches ........115mm .................. 4.7 inches ....... 120mm
Ground Clearance ..................................1.1 inches ..........30mm .................. 1.2 inches ......... 30mm
Front Twin Casters .................................3.0 inches .......... 75mm ..................... 3 inches ......... 75mm
Rear Braked Casters .............................. 3.9 inches ........ 100mm ..................... 4 inches ....... 100mm
*Reach = center of spreader bar to the front of the mast
Weights
Mast, Base & Boom Assembly ...................63.1 lbs ....... 28.9kgs ...................... 63.7 lbs .......26.8kgs
Power Pack ..................................................6.2 lbs ......... 2.8kgs .............................N/A ............. N/A
Total............................................................69.9 lbs ....... 31.7kgs ......................... 59 lbs .......26.8kgs
Base Assembly (not incl. battery)...............32.8 lbs ....... 14.9kgs ...................... 32.9 lbs .......14.9kgs
Mast & Boom (not incl. battery) ..................30.9 lbs ....... 14.0kgs ...................... 26.3 lbs .......11.9kgs
All measurements are within a +5/-5 degree of tolerance.
Hoyer®
Advance
19
English
ELECTRICAL SPECIFICATIONS:
BATTERIES............................................................24 volt Rechargeable sealed lead acid type
BATTERY CAPACITY ............................................ 3.2 Ampere hours
CHARGER RATED INPUT .................................... 100 - 240 V AC/ 50/60 Hz
CHARGER RATED OUTPUT ................................ 29.5 VDC, Max. 19 W
ELECTRIC SHOCK PROTECTION
CHARGER ..............................................................CLASS II (EN 60601-1)
LIFT ........................................................................ INTERNAL POWER SOURCE
DEGREE OF SHOCK PROTECTION
CHARGER ..............................................................TYPE B (EN 60601-1)
LIFT ........................................................................ TYPE B (EN 60601-1)
ENVIRONMENTAL CONDITIONS:
Outside this environment functionality and safety may be compromised.
OPERATING:
TEMPERATURE ................................................... 5°C to 40°C
RELATIVE HUMIDITY .......................................... 20% to 90% @ 30°C - not condensing
ATMOSPHERIC PRESSURE .............................. 700 to 1060 hPa
STORAGE:
TEMPERATURE........................................... ........ -10°C to +50°C
RELATIVE HUMIDITY .......................................... 20% to 90% @ 30°C - not condensing
ATMOSPHERIC PRESSURE .............................. 700 to 1060 hPa
IP RATINGS
CONTROL BOX .................................................... IPX4
ACTUATOR ...........................................................IPX4
OFF BOARD CHARGER .....................................IPX5
BATTERY .............................................................. IPX5
HANDSET .............................................................IPX5
DUTY CYCLES
ACTUATOR ...........................................................10% (2 min./18 min.)
CHARGER .............................................................Approx. 4 hrs.
ACOUSTICS
A-Weighted sound power level .............................. 60 dB (A)
Hoyer®
Advance
20
English
KEY SYMBOLS:
The following symbols are used on the charger, control unit and battery:
Type B equipment, as per EN 60601-1
Class 2 equipment
The disposal of the charging and control unit should not be mixed with general
household waste.
The disposal of batteries should not be mixed with general household waste.
The disposal of electronics should not be mixed with general household waste.
For indoor use
ATTENTION, consult accompanying documents.
WEEE Producers registration number WEE/GG0464RZ
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84

Joerns Hoyer® Advance Le manuel du propriétaire

Taper
Le manuel du propriétaire
Ce manuel convient également à