Drive DeVilbiss 1025DS Manuel utilisateur

Taper
Manuel utilisateur

Ce manuel convient également à

en
Drive DeVilbiss® 10-Liter Oxygen Concentrator Instruction Guide
WARNING–Read instruction guide before operating this equipment.
ASSEMBLED IN USA
NO SMOKING
es
Guía de instrucciones del concentrador de oxígeno de 10-litros de Drive DeVilbiss®
ADVERTENCIA–Lea la guía de instrucciones antes de poner a funcionar este equipo.
ENSAMBLADO EN EE. UU.
NO FUMAR
fr
Guide d’instructions du concentrateur d’oxygène 10-litres Drive DeVilbiss®
AVERTISSEMENT–Lire le mode d’emploi avant d’utiliser ce dispositif.
ASSEMBLÉ AUX ÉTATS-UNIS
NE PAS FUMER
0044
1025 Series
Serie de 1025
Série 1025
EN - 2 A-1025
ENGLISH........................................................................................................................................................................................... en-2
ESPAÑOL .......................................................................................................................................................................................... es-14
FRANÇAIS ........................................................................................................................................................................................ fr-26
TABLE OF CONTENTS
Symbol Definitions .................................................................................................................... en - 3
Important Safeguards.................................................................................................................. en - 3
Introduction.......................................................................................................................... en - 5
Intended Use .................................................................................................................... en - 5
Indications For Use ............................................................................................................... en - 5
Contraindications ................................................................................................................ en - 5
Essential Performance ............................................................................................................ en - 5
Service Life ..................................................................................................................... en - 5
Why Your Physician Prescribed Supplemental Oxygen ................................................................................... en - 5
How Your Concentrator Works ...................................................................................................... en - 5
Important Parts of Your Concentrator...................................................................................................... en - 6
Setting Up Your Concentrator ........................................................................................................... en - 7
Operating Your Concentrator ............................................................................................................ en - 7
DeVilbiss OSD
®
Operation.......................................................................................................... en - 8
Reserve Oxygen System ............................................................................................................... en - 8
Caring for Your Concentrator ............................................................................................................ en - 8
Troubleshooting ...................................................................................................................... en - 9
Overview of Alarms ................................................................................................................... en - 9
Return and Disposal................................................................................................................... en - 10
Provider’s Notes...................................................................................................................... en - 10
Specifications ........................................................................................................................ en - 11
Electromagnetic Compatibility Information.................................................................................................. en - 12
Warranty............................................................................................................................ en - 13
WARNING
Under certain circumstances, oxygen therapy can be hazardous. Seek medical advice before using an oxygen concentrator.
Physician Information
Physician Name: __________________________________________________________________
Telephone: ______________________________________________________________________
Address: ________________________________________________________________________
Prescription Information
Name: _________________________________________________________________________
Oxygen liters per minute
at rest:____________________ during activity:____________________ other: _____________
Oxygen use per day
Hours:_____________________________________ Minutes: ___________________________
Comments: ______________________________________________________________________
Drive DeVilbiss 10-Liter Oxygen Concentrator w/OSD Serial Number: _______________________
Drive DeVilbiss Equipment Provider Information
Set-Up Person: ___________________________________________________________________
This instruction guide was reviewed with me, and I have been instructed on the safe use and care of
the Drive DeVilbiss Oxygen Concentrator.
Signature:____________________________________________________ Date: ________________
10 10
8 8
6 6
4 4
2 2
Increase
DeVilbiss 10-Liter Series
EN - 3A-1025
SYMBOL DEFINITIONS
It is mandatory to read and understand the
operating instructions prior to use.
This symbol has a blue background on the
product label.
O
I
Off
On
Catalog Number LOT Number
Electric Shock Hazard. Cabinet to be removed by
authorized personnel only.
This symbol has a yellow background on the
product label.
Reset Serial Number Manufacturer
Danger - No smoking near patient or device.
This symbol has a red circle and diagonal bar
on the product label.
Alternating Current Normal Oxygen
EC REP
European Representative
Use no Oil, Grease or Lubricants
This symbol has a red circle and diagonal bar
on the product label.
Type B applied part Low Oxygen European Rep CE mark
Do not use near heat or open ames
This symbol has a red circle and diagonal bar
on the product label.
Double Insulated Service Required
10
10 LPM
10
8 8
6 6
4 4
2 2
Maximum recommended
ow rate: 10 LPM
General Warning
This symbol is used throughout this manual to
indicate hazardous situations to avoid.
Hour Meter
C US
TUV Rheinland C-US
approval mark
Important Information
This symbol is used throughout this manual to
indicate important information you should know.
+5˚C
+35˚C
Operating Temperature
Range +5 to +35˚C (+41
to +95˚F)
TUV Rheinland Certied
approval mark
Note and Information Symbol
This symbol is used throughout this manual to
indicate notes, useful tips, recommendations and
information.
Atmospheric Pressure
Range 840 to 1010 hPa
(Approximate sea level to
5000 ft)
Inmetro approval mark
Federal (U.S.A.) law restricts this device to sale by or on the order of a physician.
IP21
Ingress Protection - Protected against nger access to hazardous parts; protected against vertically falling water drops.
This device contains electrical and/or electronic equipment that must be recycled per EU Directive 2012/19/EU- Waste Electrical and Electronic Equipment (WEEE)
IMPORTANT SAFEGUARDS
Read this entire guide before using your DeVilbiss concentrator. Important safeguards are indicated throughout this guide. Pay special attention to all safety information.
Imminently and potentially hazardous information is highlighted by these terms:
DANGER
Indicates an imminently hazardous situation which could result in death or serious injury to the user or operator if not avoided.
WARNING
Indicates a potentially hazardous situation which could result in death or serious injury to the user or operator if not avoided.
CAUTION
Indicates a potentially hazardous situation which could result in property damage, injury, or device damage if not avoided.
IMPORTANT
Indicates important information you should know.
NOTE
Indicates notes, useful tips, recommendations, and information.
READ ALL INSTRUCTIONS BEFORE USING.
EN - 4 A-1025
DANGER
Oxygen causes rapid burning. Do not smoke while your oxygen concentrator is operating, or when you are near a person utilizing oxygen therapy.
Smoking during oxygen therapy is dangerous and is likely to result in facial burns or death. Do not allow smoking within the same room where the oxygen
concentrator or any oxygen carrying accessories are located.
If you intend to smoke, you must always turn the oxygen concentrator off, remove the cannula and leave the room where either the cannula or mask or
the oxygen concentrator is located. If unable to leave the room, you must wait 10 minutes after you have turned off the oxygen concentrator before
smoking.
Oxygen makes it easier for a re to start and spread. Do not leave the nasal cannula or mask on bed coverings or chair cushions if the oxygen concentrator is
turned on but not in use. The oxygen will make the materials ammable. Turn the oxygen concentrator off when not in use to prevent oxygen enrichment.
Keep the oxygen concentrator and cannula at least 2 m (6.5 feet) from hot, sparking objects or naked sources of ame.
Open ames during oxygen therapy are dangerous and are likely to result in re or death. Do not allow open ames within 2 m (6.5 feet) of the oxygen
concentrator or any oxygen carrying accessories.
Drive DeVilbiss oxygen concentrators are equipped with a re mitigating outlet tting that prevents propagation of re into the unit.
WARNING
To avoid electric shock, do not plug the concentrator into an AC outlet if the concentrator cabinet is broken. Do not remove the concentrator cabinet. The
cabinet should only be removed by a qualied Drive DeVilbiss technician. Do not apply liquid directly to the cabinet or utilize any petroleum-based solvents or
cleaning agents.
Improper use of the power cord and plugs can cause a burn, re or other electric shock hazards. Do not use the unit if the power cord is damaged.
Ensure the mains power cord is fully inserted into the concentrator connector (230 volt units) and the power cord plug is completely inserted into a fully
functioning AC wall outlet. Failure to do so may cause an electrical safety hazard.
The use of a re stop device as part of the oxygen accessories is optional, and is not contraindicated for use with this device. If used with the oxygen
accessories, the re stop device should be located as close to the patient as practicable.
Locate oxygen tubing and power supply cords to prevent tripping hazards and reduce the possibility of entanglement or strangulation.
Do not lubricate ttings, connections, tubing or other accessories of the oxygen concentrator to avoid the risk of re and burns.
Do NOT use lubricants, oils or grease.
Before attempting any cleaning procedures, turn the unit “Off.”
Use only water-based lotions or salves that are oxygen-compatible before and during oxygen therapy. Never use petroleum or oil-based lotions or salves to
avoid the risk of re and burns.
Use only spare parts recommended by the manufacturer to ensure proper function and to avoid the risk of re and burns.
When using the Transller Caddy with a Transll device, always keep the system on a at surface. Disassemble the system prior to moving.
If you feel discomfort or are experiencing a medical emergency while undergoing oxygen therapy, seek medical assistance immediately to avoid harm.
Geriatric, pediatric or any other patient unable to communicate discomfort can require additional monitoring and/or a distributed alarm system to convey the
information about the discomfort and/or the medical urgency to the responsible caregiver to avoid harm.
Use of this device at an altitude above 5000 feet (1524 meters) or above a temperature of 95°F (35°C) or greater than 93% relative humidity may affect the
ow rate and the percentage of oxygen and consequently the quality of the therapy. Refer to specications for details regarding parameters tested.
To ensure you receive the therapeutic amount of oxygen delivery according to your medical condition, the Oxygen Concentrator must:
be used only after one or more settings have been individually determined or prescribed for you at your specic activity levels.
be used with the specic combination of parts and accessories that are in line with the specication of the concentrator manufacturer and that were used
while your settings were determined.
Your delivery settings of the oxygen concentrator should be periodically reassessed for the effectiveness of therapy.
For your safety, the oxygen concentrator must be used according to the prescription determined by your physician.
Under certain circumstances, oxygen therapy can be hazardous. Seek medical advice before using an oxygen concentrator.
WARNING
MR Unsafe
Do not bring the device or accessories into a Magnetic Resonance (MR) environment as it may cause unacceptable risk to the patient or damage to the
oxygen concentrator or MR medical devices. The device and accessories have not been evaluated for safety in an MR environment.
Do not use the device or accessories in an environment with electromagnetic equipment such as CT scanners, Diathermy, RFID and electromagnetic security
systems (metal detectors) as it may cause unacceptable risk to the patient or damage to the oxygen concentrator. Some electromagnetic sources may not be
apparent, if you notice any unexplained changes in the performance of this device, if it is making unusual or harsh sounds, disconnect the power cord and
discontinue use. Contact your home care provider.
This device is suitable for use in home and healthcare environments except for near active HF SURGICAL EQUIPMENT and the RF shielded room of an ME
SYSTEM for magnetic resonance imaging, where the intensity of Electromagnetic DISTURBANCES is high.
Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operation. If such use is necessary,
this equipment and the other equipment should be observed to verify that they are operating normally.
Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12
inches) to any part of the oxygen concentrator, including cables specied by the manufacturer. Otherwise, degradation of the performance of this equipment
could result.
CAUTION
It is very important to follow your oxygen prescription. Do not increase or decrease the ow of oxygen consult your physician.
The surface temperature of the exhaust vents on the bottom of the unit may exceed 105.8˚F (41˚C) under certain conditions.
When device is used under extreme operating conditions, the temperature near the exhaust vents on the bottom of the unit may reach 145˚F (63˚C). Keep
body parts a minimum of 32” (81.2 cm) away from this area.
Use of harsh chemicals (including alcohol) is not recommended. If bactericidal cleaning is required, a non-alcohol based product should be used to avoid
inadvertent damage.
EN - 5
IMPORTANT
It is recommended that the homecare provider lock the ow control knob to prevent inadvertent adjustment. A ow setting other than prescribed may affect the
patient therapy.
Do not service or clean this device while in use with a Patient.
Do not use a low-output ow meter with this concentrator.
The Device is classied as IP21 which means it is protected against nger access to hazardous parts and protected against vertically falling water drops.
Equipment not suitable for use in the presence of a ammable anesthetic mixture with air or with oxygen or nitrous oxide.
This device contains electrical and/or electronic equipment. Follow local governing ordinances and recycling plans regarding disposal of device components.
SAVE THESE INSTRUCTIONS.
INTRODUCTION
This instruction guide will acquaint you with your Drive DeVilbiss oxygen concentrator. Make sure that you read and understand this guide before operating your unit. Important
safeguards are indicated throughout this guide. Pay special attention to all safety information. Contact your Drive DeVilbiss equipment provider should you have any questions.
Intended Use
The Drive DeVilbiss 10 Liter Oxygen Concentrator intended use is to provide supplemental low ow oxygen therapy for patients suffering from COPD, cardiovascular disease,
and lung disorders. The oxygen concentrator is used in home type environments, homes, nursing homes, patient care facilities, etc.
Indications For Use
The Drive DeVilbiss Oxygen Concentrator is intended for use as an oxygen concentrator to provide supplemental low ow oxygen therapy in the home, nursing homes, patient
care facilities, etc.
Contraindications
The device is not intended to be life supporting or life sustaining.
Essential Performance
Essential Performance of the Oxygen Concentrator is to deliver a continuous ow of oxygen enriched gas. Visual and audible alarms indicate if the device is not meeting
specication or a failure has been detected.
Service Life
The expected service life of the 1025 is 5 years of operation, when used in accordance with all manufacturer guidance for safe use, maintenance, storage, handling and
general operation. Expected service life of the unit, and in particular the sieve beds and compressor, may vary based on the operating environment, storage, handling and the
frequency and intensity of use.
Why Your Physician Prescribed Supplemental Oxygen
Today, many people suffer from heart, lung and other respiratory diseases. Many of these people can benet from supplemental oxygen therapy. Your body requires a steady
supply of oxygen to function properly. Your physician prescribed supplemental oxygen for you, because you are not getting enough oxygen from room air alone. Supplemental
oxygen will increase the amount of oxygen that your body receives.
Supplemental oxygen is not addictive. Your physician prescribed a specic oxygen ow to improve symptoms such as headaches, drowsiness, confusion, fatigue or increased
irritability. If these symptoms persist after you begin your supplemental oxygen program, consult your physician.
The oxygen delivery setting has to be determined for each patient individually with the conguration of the equipment to be used, including accessories.
The proper placement and positioning of the prongs of the nasal cannula in the nose is critical to the amount of oxygen delivered to the respiratory system of the patient.
Your Delivery settings of the oxygen concentrator should be periodically reassessed for the effectiveness of therapy.
How Your Drive DeVilbiss Oxygen Concentrator Works
Oxygen concentrators are the most reliable, efcient and convenient source of supplemental oxygen available today. The oxygen concentrator is electrically operated. The unit
separates oxygen from room air which allows high-purity supplemental oxygen to be delivered to you through the oxygen outlet. Although the concentrator lters the oxygen in
a room, it will not affect the normal amount of oxygen in your room.
A-1025
EN - 6 A-1025
Front View (Figure A)
1. Operating instructions
2. Power Switch
| = ON
O = OFF
3. Flow meter knob
4. Flow meter
5. Circuit breaker resets the unit after electrical overload shutdown
6. Oxygen outlet oxygen is dispersed through this port
7. Normal Oxygen (green) light (see page 8)
8. Low Oxygen (yellow) light (see page 8)
9. Red Service Required (red) light – when illuminated contact your
DeVilbiss provider
10. Hour meter
Back View (Figure B)
11. Handgrip
12. Exhaust
CAUTION
When device is used under extreme operating conditions, the
temperature near the exhaust vents on the bottom of the unit
may reach 138.2˚F (59˚C). Keep body parts a minimum of 30”
(76.2 cm) away from this area.
13. Power cord and/or IEC power connector.
14. Line cord strap
15. Filter Door with venting and compartment for optional gross particle lter
16. Auxiliary Oxygen Port: Your concentrator is equipped with an auxiliary
oxygen port that can be used to ll oxygen cylinders with an FDA-cleared
cylinder lling device that is designed to use oxygen from a concentrator
to ll a cylinder. The port is only for use with FDA-cleared lling devices
with compatible oxygen input specications. Refer to the cylinder lling
device instruction guide for the oxygen input/output specications,
connection and operating instructions.
IMPORTANT PARTS OF YOUR CONCENTRATOR
Please take time to familiarize yourself with your Drive DeVilbiss oxygen concentrator before operating.
Accessories
Transfiller Caddy ........................................................................................................................................................................................................................DeVilbiss 525DD-650
High Flow (6-15 LPM) Bubble Humidifier .................................................................................................................................................................... Salter Labs 7900 or equivalent
Low Flow (up to 6 LPM) Bubble Humidifier ................................................................................................................................................................Salter Labs 7600 or equivalent
High Flow Nasal Cannula ....................................................................................................................................................................................... Salter Labs 1600HF or equivalent
There are many types of humidiers, oxygen tubing and cannulas/masks that can be used with this device. Certain humidiers and accessories may impair the device’s
performance. A mask or any nasal cannula can be used with continuous ow delivery and may be sized according to your prescription as recommended by your homecare
provider who should also give you advice on the proper usage, maintenance and cleaning.
WARNING
When using the Transller Caddy with a Transll device, always keep the system on a at surface. Disassemble the system prior to moving.
NOTE– The device should only be used with bubble humidiers that are designed for use with ows up to 10 liters per minute and 20 psi pressure.
NOTEA maximum of 50 feet (15 meters) of crush-proof oxygen tubing, plus 7 feet (2.1 meters) of cannula, plus a bubble humidier is allowed between the concentrator
and the patient.
NOTEYour healthcare provider should verify the compatibility of the oxygen concentrator and all of the parts used to connect to the patient before use.
1
2
3
4
5
6
A
11
12
13
12
14
11
15
B
16
10
7
9
8
EN - 7
SETTING UP YOUR OXYGEN CONCENTRATOR
1. Position your unit near an electrical outlet in the room where you spend most of your time.
NOTEDo not connect to an electrical outlet controlled by a wall switch.
DANGER
Oxygen causes rapid burning. Do not smoke while your oxygen concentrator is operating, or when you are near a person
utilizing oxygen therapy. Keep the oxygen concentrator and cannula at least 2 m (6.5 feet) from hot, sparking objects or naked
sources of ame.
2. Position your unit on a at surface at least 6 inches (16 cm) from walls, draperies or any other objects that might prevent the proper
ow of air in and out of your oxygen concentrator. The oxygen concentrator should be located in a well-ventilated area to avoid
pollutants or fumes.
NOTETo move the unit, rmly grasp the handle located on the top of the unit, rolling and/or lifting the unit over pathway obstacles.
3. Before operating your unit, always check to be sure the lter door vents (located on the back of your unit) are clean. Proper cleaning is
discussed in the Caring For Your Concentrator section on page 8.
4. Attach the appropriate oxygen accessories to the oxygen outlet.
Oxygen Tubing Connection:
a. Thread the supplied oxygen outlet connector onto the oxygen outlet.
b. Attach the oxygen tubing directly to the connector (Figure 1).
OxygenTubingConnectionWithHumidication:
If your physician has prescribed an oxygen humidier as part of your therapy, follow these steps (If using a prell, go to step b.):
a. Fill the humidier bottle as per manufacturer’s instructions.
b. Thread the wing nut located on the top of the humidier bottle to the oxygen outlet so that it is suspended (Figure 2). Make sure
it is securely tightened.
c. Attach the oxygen tubing directly to the humidier bottle outlet tting (Figure 3).
NOTE– Your physician has prescribed either a nasal cannula or face mask. In most cases, they are already attached to the oxygen
tubing. If not, follow the manufacturer’s instructions for attachment.
NOTE– Your healthcare provider should verify the compatibility of the oxygen concentrator and all of the parts used to connect to the
patient before use.
5. Remove the power cord completely from the line cord strap. Make sure the power switch is in the “Off” position and insert the plug
into the wall outlet. The unit is double insulated to guard against electric shock.
WARNING
Ensure the mains power cord is fully inserted into the concentrator connector (230 volt units) and the power cord plug is
completely inserted into a fully functioning AC wall outlet. Failure to do so may cause an electrical safety hazard.
NOTE– (only 120 volt units) The plug on the Drive DeVilbiss oxygen concentrator has one blade wider than the other. To reduce the
risk of electric shock, this plug is intended to t in a wall outlet only one way. Do not attempt to defeat this safety feature.
NOTE– To check your oxygen concentrator and accessories for proper operation; 1. Check the output ow by placing the end of the nasal cannula under the surface of a
half-full cup of water and look for the bubbles. 2. Check the system for leaks by bending the nasal prongs over and squeeze tight to stop the ow of oxygen. Look at the ow
meter to see that the indicator ball on the ow meter drops to zero. If the indicator ball does not drop to zero, check all connections for possible leaks. Parts to check for leaks
are: tubing connections, humidier bottle and other accessories like rebreaks. Repeat these steps until the ow meter ball drops to zero. Contact your provider or service
supplier immediately if you encounter any problems.
WARNING
Improper use of the power cord and plugs can cause a burn, re or other electric shock hazards. Do not use the unit if the power cord is damaged.
OPERATING YOUR DEVILBISS OXYGEN CONCENTRATOR
DANGER
Oxygen causes rapid burning. Do not smoke while your oxygen concentrator is operating, or when you are near a person
utilizing oxygen therapy. Keep the oxygen concentrator and cannula at least 2 m (6.5 feet) from hot, sparking objects or
naked sources of ame.
Drive DeVilbiss oxygen concentrators are equipped with a re mitigating outlet tting that prevents propagation of re into
the unit.
1. Press the power switch to the “On” position. When the unit is turned “On,” all three lights (Service Required, Low Oxygen and
Normal Oxygen) on the front panel will illuminate briey and an audible signal will briey alarm conrming that the LEDs and audible
signal are functioning properly. The unit will then operate in “start up” mode with the Low Oxygen light lit until a normal oxygen level
is achieved, at which time the Normal Oxygen light will remain lit. The “start up” may take up to 15 minutes.
NOTE– DeVilbiss recommends for optimal service life that the Drive DeVilbiss Oxygen Concentrator be operated for at least 30 minutes after it is powered on. Shorter
periods of operation, operating in extreme temperature/humidity conditions or in the presence of contaminates, and/or handling and storage conditions outside those
specied, may affect the long term reliable operation of the product.
DANGER
Oxygen makes it easier for a re to start and spread. Do not leave the nasal cannula or mask on bed coverings or chair cushions if the oxygen concentrator is
turned on but not in use, the oxygen will make the materials ammable. Turn the oxygen concentrator off when not in use to prevent oxygen enrichment.
NOTE– If the audible signal alarms but the unit is not operating, there is no power to the unit. Refer to the Minor Troubleshooting chart on page 9 and contact your
Drive DeVilbiss provider if necessary.
A-1025
NO
SMOKING
3
2
1
EN - 8
NOTE– If an audible low-frequency vibration sound is detected, the unit is not operating properly. Refer to the Minor Troubleshooting chart on page 9, and contact
your Drive DeVilbiss provider if necessary.
2. Check the ow meter to make sure that the ow meter ball is centered on the line next to the prescribed number of your ow rate.
CAUTION
It is very important to follow your oxygen prescription. Do not increase or decrease the ow of oxygen consult your physician.
NOTEYour Drive DeVilbiss provider may have preset the ow meter so that it cannot be adjusted.
NOTE– If the ow meter knob is turned clockwise, the ow decreases (and eventually will shut off the oxygen ow). If the knob is turned counterclockwise, the ow
increases.
NOTE– For prescriptions of 10 LPM, be sure the ball is centered on the 10 liter line. The ball should not touch the red line. Setting the ow higher than 10 may cause
the oxygen purity level to drop.
NOTE– The low-ow alarm may activate if the ow meter ball is set below 2 lpm. The unit will continue to run; however, the Service Required light will come on
accompanied by an audible alarm. Adjust the ow meter to your prescribed ow.
IMPORTANT
Do not use a low-output ow meter with this concentrator.
3. Your Drive DeVilbiss concentrator is now ready for use, properly position the cannula with the nasal prongs facing upward, insert the
prongs into nose. Wrap the cannula tubing over the ears and position in front of body (Figure 4). Allow 15 minutes for the oxygen
concentrator to reach stated performance.
DeVilbiss OSD
®
Operation
The OSD (Oxygen Sensing Device) is a device within your concentrator that monitors the oxygen produced by your unit.
The OSD lights on the top panel are dened as follows:
Green Normal Oxygen light–acceptable oxygen level.
Yellow Low Oxygen light– below an acceptable oxygen level.
If the oxygen purity falls below the acceptable level: The green Normal Oxygen light will shut off, the yellow Low Oxygen light will illuminate, and an intermittent audible signal
will sound.
Refer to the Minor Troubleshooting section in this guide on page 9, and switch to your reserve oxygen system. Do not attempt any other maintenance. Contact your Drive
DeVilbiss provider immediately.
RESERVE OXYGEN SYSTEM
As a precaution, your Drive DeVilbiss provider may supply you with a reserve oxygen system. If your unit loses electrical power or fails to operate correctly, the Patient Alert
System will sound to signal you to switch to your reserve oxygen system (if provided) and contact your Drive DeVilbiss provider.
CARING FOR YOUR DEVILBISS OXYGEN CONCENTRATOR
Drive DeVilbiss recommends using only original DeVilbiss parts and lters in order to guarantee reliable operation of the product.
WARNING
Do NOT use lubricants, oils or grease.
Before attempting any cleaning procedures, turn the unit “Off.”
Cannula/Mask, Tubing and Humidifier Bottle
Clean and replace the cannula/mask, tubing and humidier bottle according to the manufacturer’s instructions.
Filter Door with Vents
Inspect the vents periodically, and wipe with a dry cloth as needed to remove dust.
Exterior Cabinet
Clean the concentrator exterior cabinet weekly by using a damp cloth and wiping dry; the vents can also be wiped with a damp cloth.
Cleaning
Recommended cleaning
interval
Number of cleaning
cycles * Compatible cleaning method
Outer Cabinet 7 days 260 Water, use only a damp cloth
Filter Door Vents 7 days 260 Wipe with dry cloth, or a cloth dampened with
water to remove dust.
Oxygen Outlet Connector 7 days 104 Mild dish soap (2 tbsp) and warm water (2 cups)
* number of cleaning cycles determined by recommended cleaning interval and expected service life
WARNING
To avoid electric shock, do not plug the concentrator into an AC outlet if the concentrator cabinet is broken. Do not remove the concentrator cabinet. The cabinet
should only be removed by a qualied DeVilbiss technician. Do not apply liquid directly to the cabinet or utilize any petroleum-based solvents or cleaning agents.
CAUTION
Use of harsh chemicals (including alcohol) is not recommended. If bactericidal cleaning is required, a non-alcohol based product should be used to avoid
inadvertent damage.
4
A-1025
EN - 9
TROUBLESHOOTING
The following troubleshooting chart will help you analyze and correct minor oxygen concentrator malfunctions. If the suggested procedures do not help, switch to your reserve
oxygen system and call your DeVilbiss homecare provider. Do not attempt any other maintenance.
WARNING
To avoid electric shock, do not plug the concentrator into an AC outlet if the concentrator cabinet is broken. Do not remove the concentrator cabinet. The cabinet
should only be removed by a qualied DeVilbiss technician.
Minor Troubleshooting Chart
SYMPTOM POSSIBLE CAUSE REMEDY
A. Unit does not operate. All lights are off when the
power switch is “On.Audible alert is pulsing.
1. Power cord not properly inserted
into wall outlet.
1. Check power cord connection at the wall outlet. On 230 volt units, also
check the mains connection on the back of the unit.
2. No power at wall outlet. 2. Check your home circuit breaker and reset if necessary. Use a different
wall outlet if the situation occurs again.
3. Oxygen concentrator circuit
breaker activated.
3. Press the concentrator circuit breaker reset button located below the
power switch. Use a different wall outlet if the situation occurs again.
If the above remedies do not work, contact your DeVilbiss provider.
B. Unit operates. Red Service Required light is
illuminated. Audible alert may be sounding.
1. Filter door vents are blocked. 1. Check lter door vents and ensure that the openings are not blocked.
2. Exhaust is blocked. 2. Check the exhaust area and make sure there is nothing restricting the
unit exhaust.
3. Blocked or defective cannula, face
mask, or oxygen tubing.
3. Detach cannula or face mask. If proper ow is restored, clean or replace
if necessary. Disconnect the oxygen tubing at the oxygen outlet. If
proper ow is restored, check oxygen tubing for obstructions or kinks.
Replace if necessary.
4. Blocked or defective humidier
bottle.
4. Detach the humidier from the oxygen outlet. If proper ow is obtained,
clean or replace humidier.
5. Flow meter set too low. 5. Set ow meter to prescribed ow rate.
If the above remedies do not work, contact your DeVilbiss provider.
C. Unit operates. Audible low-frequency vibration
sound is detected.
1. Electronic Assembly Malfunction. 1. Turn your unit “Off.” Switch to your reserve oxygen system and contact
your DeVilbiss provider immediately.
D. Yellow Low Oxygen light is on. 1. Unit in “start up” mode. 1. Allow unit up to 15 minutes to complete start up period.
E. The yellow Low Oxygen light is on and the
intermittent audible signal is sounding.
1. Flow meter is not properly set. 1. Ensure the ow meter is properly set to the prescribed number. (The
maximum ow meter setting is 3 LPM when an oxygen bottle is being
lled with oxygen from the auxiliary port.)
2. Filter door vents are blocked. 2. Check lter door vents and ensure that the openings are not blocked.
3. Exhaust is blocked. 3. Check the exhaust area and make sure there is nothing restricting the
unit exhaust.
If the above remedies do not work, contact your DeVilbiss provider.
F Red Service Required light is on and an
intermittent audible signal is sounding.
1. Flow meter is not properly set. 1. Ensure the ow meter is properly set to the prescribed number. (The
maximum ow meter setting is 3 LPM when an oxygen bottle is being
lled with oxygen from the auxiliary port.)
2. Filter door vents are blocked. 2. Check lter door vents and ensure that the openings are not blocked.
3. Exhaust is blocked. 3. Check the exhaust area and make sure there is nothing restricting the
unit exhaust.
If the above remedies do not work, contact your DeVilbiss provider.
4. Electronic Assembly Malfunction. 4. Turn your unit “Off.” Switch to your reserve oxygen system and contact
your DeVilbiss provider immediately.
G. If any other problems occur with your oxygen
concentrator.
1. Turn your unit “Off.” Switch to your reserve oxygen system and contact
your DeVilbiss provider immediately.
OVERVIEW OF ALARMS
This device contains an alarm system which monitors the state of the device and alerts of abnormal operation, loss of essential performance or failures. Alarm conditions are
shown on the LED display. The alarm system functions are tested at power up by lighting all visual alarm indicators and sounding the audible alarm (beep).
All alarms are Low Priority Technical Alarms.
Alarm Condition
LED
Indicator
Visual Alarm Signal
Meaning
Audible Alarm
Signal
Visual Alarm Signal
Cleared by Action to take
Start-up Period
YELLOW
Low O
2
LED ON
No
After startup period,
O
2
≥82%
Wait for unit to nish startup
period, up to 15 minutes
Low Oxygen
Concentration at <82%
YELLOW
Low O
2
LED ON
Yes Turn unit off See Troubleshooting table
Malfunction
RED
Service Required LED ON
Yes Turn unit off
Return unit to provider for
service
A-1025
EN - 10
RETURN AND DISPOSAL
This device may not be disposed of with household waste. After use of the device, please return the device to the provider for disposal. This device contains electrical and/or
electronic components that must be recycled per EU Directive 2012/19/EU-Waste Electrical and Electronic Equipment (WEEE). Non-infectious used accessories (e.g. nasal
cannula) can be disposed of as residential waste. The disposal of infectious accessories (e.g. nasal cannula from an infected user) must be made via an approved waste
disposal company. Names and addresses can be obtained from the local municipality.
PROVIDER’S NOTES - Cleaning and Disinfection When There is a Patient Change
Drive DeVilbiss Healthcare recommends that at least the following procedures be carried out by the manufacturer or a qualied third party between uses by different patients.
NOTE If the following described complete processing of the concentrator by an appropriately trained individual is not possible, the device should not be used by another
patient.
NOTE If preventive maintenance is due at this time, these procedures should be carried out in addition to the servicing procedures.
1. Use disinfectants safely. Always read the label and product information before use.
2. Always wear personal protective equipment when performing this procedure. Use suitable gloves and safety glasses. Cover exposed skin on arms to prevent
accidental contact with bleach solution that has been applied to the concentrator.
3. Dispose of all accessories that are not suitable for reuse. This includes but may not be limited to the oxygen tubing, tubing connectors, nasal cannula and/or mask,
oxygen outlet connector, and humidier bottle.
4. Clean the exterior of the concentrator with a clean lint-free cloth. Heavy soil should be removed with a clean lint-free cloth dampened with water. A soft bristled
brush dampened with water can be used to remove stubborn soil. Dry the concentrator using a clean lint-free cloth if water was used to remove soil.
5. Use 5.25% chlorine bleach (Clorox Regular Liquid Bleach or equivalent). Mix one (1) part bleach with four (4) parts water in an appropriate clean container. This
ratio produces a one (1) part bleach to ve (5) total parts solution (1:5). The total volume (amount) of solution required is determined by the number of concentrators
in need of disinfection.
NOTE An alternate suitable disinfecting agent (e.g. Mikrobac® forte or Terralin® Protect) may also be used. Follow disinfectant
manufacturer’s instructions.
6. Apply the bleach solution in an even manner to the cabinet and power cord using a clean lint-free cloth. The cloth should be dampened only and not dripping of
solution. Do not use a spray bottle to apply the solution. Do not saturate the device with the solution. Take care that no solution enters the vent areas on the
concentrator base or the Auxiliary O2 tting area on the back of the unit. Avoid over-saturating the cabinet seams so that no solution residue builds up in these
areas. Avoid the caster wells located on the bottom of the unit.
7. Exposure time of the disinfectant solution should be 10 minutes minimum to 15 minutes maximum.
8. After the recommended exposure time, all surfaces of the concentrator should be wiped with a clean lint-free cloth dampened with drinking quality water no warmer
than room temperature. Dry the unit with a dry, clean lint-free cloth. This is to remove residue that may stain or leave a lm on the unit, especially after repeated
disinfections.
9. Check the cord, the plug on the back of the device, the power switch, the fuse holder, and the indicator lights for possible damage. Replace all damaged or worn
components.
10. Check the oxygen concentration. If the device is within specication, the extended life intake bacteria lter does not need to be replaced between patients. If the oxygen
concentration is not within specication, the provider should refer to the service manual section on Troubleshooting.
NOTE – There is no portion of the gas pathways through the concentrator that could be contaminated with body uids under normal conditions.
The device patient connection may unintentionally become contaminated with expired gases for a single fault condition i.e., a hose internal to the device becomes
disconnected. This condition will cause no ow out of the device and/or an alarm condition. Should this occur, refer to the service manual for additional instructions.
Disinfection
NOTE – The disinfection process can only be completed by the manufacturer or by an appropriately trained individual.
Recommended
disinfection interval
Number of
disinfection cycles
Compatible disinfection
method
Cabinet, power cord Between patients 20 1:5 chlorine bleach (5.25%) and water
solution, Mikrobac forte, Terralin Protect
Oxygen tubing, tubing connectors, nasal cannula/mask,
oxygen outlet connector, humidier bottle
Do not clean, replace between
patients
N/A N/A
A-1025
EN - 11
SPECIFICATIONS
DEVILBISS 10-LITER SERIES
Catalog Number 1025DS 1025KS
1025UK
Delivery Rate 2 to 10 LPM 2 to 10 LPM 2 to 10 LPM
Maximum Recommended Flow (@
nominal outlet pressures of zero and
7 kPa)**
10 LPM 10 LPM 10 LPM
Outlet Pressure 20.0 ± 1.0 psi (138 kPa +/- 7 kPa) 20.0 ± 1.0 psi (138 kPa +/- 7 kPa) 20.0 ± 1.0 psi (138 kPa +/- 7 kPa)
Auxiliary Oxygen Port **
Outlet Pressure: <15 psi
Outlet Flow: 2 LPM
Outlet Pressure: <15 psi
Outlet Flow: 2 LPM
Outlet Pressure: <15 psi
Outlet Flow: 2 LPM
Electrical Rating 120 V, 60 Hz, 6.1 Amp 230 V~, 50 Hz, 3.2 Amp 240 V~, 50 Hz, 3.2 Amp
Operating Voltage Range 102-132 V~, 60 Hz 195-253 V~, 50 Hz 204-264 V~, 50 Hz
Oxygen Percentage 2-10 LPM=87%-96% 2-10 LPM=93% +3%/-6% 2-10 LPM=93% +3%/-6%
Operating Altitude
(tested at 70˚F {21˚C} only)
0-1500 M (0-5000 ft)
Across the voltage range:
No degradation of performance
Across the voltage range:
No degradation of performance
Across the voltage range:
No degradation of performance
Operating Environment Range*
41˚F (5˚C) to 95˚F (35˚C), humidity
range of 15% to 93% non-
condensing
No degradation in performance across the
operating voltage range.
No degradation in performance across
the operating voltage range.
No degradation in performance across
the operating voltage range.
Power Consumption 120 vac, 60HZ: 639 watts average 230 vac, 50Hz: 664 watts average 240 vac, 50Hz: 670 watts average
Weight 42 lbs. (19 Kilograms) 42 lbs. (19 Kilograms) 42 lbs. (19 Kilograms)
Safe Working Load
53 lbs. (24 Kilograms) 53 lbs. (24 Kilograms) 53 lbs. (24 Kilograms)
Sound Pressure Level at 3 and
10LPM
<59 dBA <57 dBA <57 dBA
Sound Power Level at 3 and 10 LPM <69 dBA <67 dBA <67 dBA
Dimensions
24.5”H x 13.5”W x 12”D
(62.2 x 34.2 x 30.4 cm)
24.5”H x 13.5”W x 12”D
(62.2 x 34.2 x 30.4 cm)
24.5”H x 13.5”W x 12”D
(62.2 x 34.2 x 30.4 cm)
Maximum Limited Pressure 35 PSIG (241 kPa) 35 PSIG (241 kPa) 35 PSIG (241 kPa)
Operating System Time Cycle / Pressure Swing Time Cycle / Pressure Swing Time Cycle / Pressure Swing
Low Oxygen Indicator
<82% low oxygen
<60% very low oxygen
<82% low oxygen
<60% very low oxygen
<82% low oxygen
<60% very low oxygen
Storage Conditions -13˚F (-25˚C) to 158˚F (70˚C), humidity
range of 15% to 93% non-condensing
-13˚F (-25˚C) to 158˚F (70˚C), humidity
range of 15% to 93% non-condensing
-13˚F (-25˚C) to 158˚F (70˚C), humidity
range of 15% to 93% non-condensing
Equipment Class and Type
Class II Equipment Double Insulated;
Type B Applied Part
Ordinary Equipment, IP21
Class II Equipment Double Insulated;
Type B Applied Part
IP21
Class II Equipment Double Insulated;
Type B Applied Part
IP21
Approval Body and Safety Standard
TUV
ANSI/AAMI ES60601-1:2005+A2 (R2012)
+A1
IEC 60601-1-6:2010
IEC 60601-1-11:2015
ISO 80601-2-69:2014
CAN/CSA-C22.2 No. 60601-1:14
CAN/CSA-C22.2 No. 60601-1-6:11
CAN/CSA-C22.2 No. 60601-1-11:15
CAN/CSA-C22.2 No. 80601-2-69:16
TUV
IEC 60601-1:2012
IEC 60601-1-6:2010+A1
IEC 60601-1-11:2015
EN ISO 80601-2-69:2014
TUV
IEC 60601-1:2012
IEC 60601-1-6:2010+A1
IEC 60601-1-11:2015
EN ISO 80601-2-69:2014
CE mark No Yes Yes
EMC Compliance To EN60601-1-2 EN60601-1-2 EN60601-1-2
*
NOTEThe OSD performance at 41˚F (5˚C) to 95˚F (35˚C), 93% R.H. through voltage range on the 1025 veried at 670m.
**The maximum recommended ow is 6 LPM when an oxygen bottle is being lled with oxygen from the auxiliary oxygen port.
Specications subject to change without notice.
A-1025
EN - 12
Oxygen Concentration vs Flow Rate
Flow L/m
%O
2
10 87% - 92%
9 87% - 93%
8 87% - 95%
7 87% - 96%
6 87% - 96%
5 87% - 96%
4 87% - 95%
3 87% - 95%
2 87% - 94%
ELECTROMAGNETIC COMPATIBILITY INFORMATION
WARNING
MR Unsafe
Do not bring the device or accessories into a Magnetic Resonance (MR) environment as it may cause unacceptable risk to the patient or damage to the oxygen
concentrator or MR medical devices. The device and accessories have not been evaluated for safety in an MR environment.
Do not use the device or accessories in an environment with electromagnetic equipment such as CT scanners, Diathermy, RFID and electromagnetic security
systems (metal detectors) as it may cause unacceptable risk to the patient or damage to the oxygen concentrator. Some electromagnetic sources may not be
apparent, if you notice any unexplained changes in the performance of this device, if it is making unusual or harsh sounds, disconnect the power cord and
discontinue use. Contact your home care provider.
This device is suitable for use in home and healthcare environments except for near active HF SURGICAL EQUIPMENT and the RF shielded room of an ME
SYSTEM for magnetic resonance imaging, where the intensity of Electromagnetic DISTURBANCES is high.
WARNING
Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operation. If such use is necessary, this
equipment and the other equipment should be observed to verify that they are operating normally.
WARNING
Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to
any part of the oxygen concentrator, including cables specied by the manufacturer. Otherwise, degradation of the performance of this equipment could result.
A-1025
EN - 13
WARRANTY
Drive DeVilbiss Healthcare warrants the DeVilbiss 10 Liter Oxygen Concentrator under the conditions and limitations stated below. Drive DeVilbiss warrants this equipment to
be free from defects in workmanship and materials for three (3) years from date of factory shipment to the original purchaser, (typically the healthcare provider) unless
contractually specied otherwise. This warranty is limited to the Buyer of new equipment purchased directly from Drive DeVilbiss, or one of its Providers, Distributors, or
Agents. Drive DeVilbiss’ obligation under this warranty is limited to product repair (parts and labor) at its factory or at an Authorized Service Center. Routine maintenance
items, such as lters, are not covered under this warranty, nor does it cover normal wear and tear.
Warranty Claims Submissions
The original purchaser must submit any warranty claim to Drive DeVilbiss or to an Authorized Service Center. Upon verication of the warranty status, instructions will be
issued. For all returns, the original purchaser must (1) properly package the unit in a DeVilbiss approved shipping container, (2) properly identify the claim with the Return
Authorization Number, and (3) send the shipment freight prepaid. Service under this warranty must be performed by Drive DeVilbiss and/or an Authorized Service Center.
NOTE– This warranty does not obligate Drive DeVilbiss to provide a loaner unit during the time that an oxygen concentrator is undergoing repair.
NOTE– Replacement components are warranted for the unexpired portion of the original Limited Warranty.
This warranty shall be voided, and Drive DeVilbiss shall be relieved of any obligation or liability if:
The device has been misused, abused, tampered with, or used improperly during this period.
Malfunction results from inadequate cleaning or failure to follow the instructions.
The equipment is operated or maintained outside the parameters indicated in the Drive DeVilbiss operating and service instructions.
Unqualied service personnel conduct routine maintenance or servicing.
Unauthorized parts or components (i.e., regenerated sieve material) are used to repair or alter the equipment.
Unapproved lters are used with the unit.
THERE IS NO OTHER EXPRESS WARRANTY. IMPLIED WARRANTIES, INCLUDING THOSE OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE,
ARE LIMITED TO THE DURATION OF THE EXPRESS LIMITED WARRANTY AND TO THE EXTENT PERMITTED BY LAW ANY AND ALL IMPLIED WARRANTIES ARE
EXCLUDED. THIS IS THE EXCLUSIVE REMEDY AND LIABILITY FOR CONSEQUENTIAL AND INCIDENTAL DAMAGES UNDER ANY AND ALL WARRANTIES ARE
EXCLUDED TO THE EXTENT EXCLUSION IS PERMITTED BY LAW. SOME STATES DO NOT ALLOW LIMITATIONS ON HOW LONG AN IMPLIED WARRANTY LASTS,
OR THE LIMITATION OR EXCLUSION OF CONSEQUENTIAL OR INCIDENTAL DAMAGES, SO THE ABOVE LIMITATION OR EXCLUSION MAY NOT APPLY TO YOU.
This warranty gives you specic legal rights, and you may also have other rights that vary from state to state.
NOTE International warranties may vary.
ORDERING AND RETURNING PARTS
Drive DeVilbiss Customer Service Contact Information
Customer Service (USA): 800-338-1988
International Department: 814-443-4881 / [email protected]
Ordering Non-Warranty Replacement Parts
Order non-warranty parts and literature from your Drive DeVilbiss provider.
A-1025
ES - 14 A-1025
INDICE
Definiciones de los símbolos ............................................................................................................ es - 15
Normas de Seguridad Importantes ....................................................................................................... es - 15
Introducción ......................................................................................................................... es - 17
Uso previsto..................................................................................................................... es - 17
Indicaciones de Uso .............................................................................................................. es - 17
Contraindicaciones ............................................................................................................... es - 17
Rendimiento esencial ............................................................................................................. es - 17
Vida útil ........................................................................................................................ es - 17
Por Qué su Médico le Prescribió un Suplemento de Oxígeno ............................................................................. es - 17
Cómo Funciona su Concentrador ................................................................................................... es - 17
Partes Importantes de su Concentrador ................................................................................................... es - 18
Preparación de su Concentrador ......................................................................................................... es - 19
Uso de su Concentrador ............................................................................................................... es - 19
Funcionamiento de DeVilbiss OSD
®
.................................................................................................. es - 20
Sistema de Oxígeno de Reserva ........................................................................................................ es - 20
Cuidado de su Concentrador ............................................................................................................ es - 20
Detección y Solución de Problemas ...................................................................................................... es - 21
Información General de las Alarmas ...................................................................................................... es - 22
Devolución y Eliminación ............................................................................................................... es - 22
Recomendaciones del Proveedor ........................................................................................................ es - 23
Especificaciones...................................................................................................................... es - 24
Información de compatibilidad electromagnética............................................................................................. es - 25
Garantia ............................................................................................................................ es - 25
ADVERTENCIA
Bajo ciertas circunstancias, la terapia de oxígeno puede ser peligrosa. Se aconseja pedir asesoría médica antes de usar el concentrador de oxígeno.
Información del médico
Nombre del médico: ________________________________________________________________
Teléfono: _________________________________________________________________________
Dirección: ________________________________________________________________________
Información de la prescripción
Nombre: _________________________________________________________________________
Litros de oxígeno por minuto
en reposo: ________________ durante actividad: ______________ otra: _________________
Uso de oxígeno por día:
horas: ______________________ minutos: ______________________
Comentarios: ________________________________________________________________________
Número de Serie del Concentrador de 10 Litros Drive DeVilbiss ____________________________
Información del proveedor de equipo médico en el hogar Drive DeVilbiss
Persona que prepara el aparato: ________________________________________________________
Esta guía de instrucciones se revisó conmigo y se me ha instruido en el uso y cuidado seguro del
concentrador de oxígeno Drive DeVilbiss.
Firma: ______________________________________________________ Fecha: ________________
10 10
8 8
6 6
4 4
2 2
Aumento
Serie de 10 Litros de DeVilbiss
ES - 15A-1025
DEFINICIONES DE LOS SÍMBOLOS
Es obligatorio leer y comprender las
instrucciones de funcionamiento antes de usar.
Este símbolo tiene un fondo azul en la
etiqueta del producto.
O
I
Apagado
Encindido
Número de catálogo/
modelo
Número de LOTE
Peligro de Elecroshock. La cubierta debe ser
retirada solo por personal autorizado.
Este símbolo tiene un fondo amarillo en la
etiqueta del producto.
Reinicio Número de serie Fabricante
Peligro: no fume cerca del paciente ni del
aparato.
Este símbolo tiene un círculo rojo y una
barra diagonal en la etiqueta del producto.
Corriente alterna Oxígeno normal
EC REP
Representante europeo
No usar aceite, grasa ni lubricantes
Este símbolo tiene un círculo rojo y una
barra diagonal en la etiqueta del producto.
Parte aplicada Tipo B Oxígeno bajo
Marcado CE como
Representante europeo
No usar cerca del calor ni del fuego directo
Este símbolo tiene un círculo rojo y una
barra diagonal en la etiqueta del producto.
Doble aislamiento
Mantenimiento
requerido
10
≤ 10 LPM
10
8 8
6 6
4 4
2 2
Nivel de ujo máximo
recomendado: 10 l/min
Advertencia general
Este símbolo se usa a lo largo de este
manual para indicar situaciones peligrosas que
debe evitar.
Medidor de horas
C US
Marcado de
aprobación de TUV
Rheinland C-US
Información importante
Este símbolo se usa a lo largo de este
manual para indicar información importante
que usted debe conocer.
+5˚C
+35˚C
Rango de temperatura de
operación + 5 a + 35 ˚C
(+ 41 a + 95 ˚F)
Marcado de
aprobación de la
Certicación TUV
Rheinland
Nota y símbolo de información
Este símbolo se usa a lo largo de este
manual para indicar notas, consejos útiles,
recomendaciones e información.
Rango de presión
atmosférica 840 a 1010
hPa (nivel del mar
aproximado a 1524 metros
[5000 pies]).
Marca de aprobación
Inmetro
La ley federal (de los EE. UU.) restringe la venta de este aparato, el cual solo puede ser vendido por un médico o con la prescripción de este.
IP21
Protección contra ingresos: protegido contra el acceso de los dedos a las piezas peligrosas; protegido contra la caída vertical de gotas de agua.
Este dispositivo contiene componentes eléctricos y/o electrónicos que se deben reciclar conforme a la directiva de la EU 2012/19/EU relativa a residuos de
aparatos eléctricos y electrónicos (RAEE).
NORMAS DE SEGURIDAD IMPORTANTES
Lea toda la guía antes de usar su concentrador DeVilbiss. Se indican las normas de seguridad importantes a través de esta guía; preste atención especial a toda la
información de seguridad. Las información inminente y potencialmente peligrosa se destaca con estos términos:
PELIGRO
Indica una situación de peligro inminente que puede causar la muerte o lesiones severas al usuario u operador si no se evita.
ADVERTENCIA
Indica una situación potencialmente peligrosa que, de no evitarse, podría provocar la muerte o lesiones graves al usuario u operador.
PRECAUCIÓN
Indica una situación potencialmente peligrosa que, de no evitarse, podría provocar la muerte o lesiones graves al usuario u operador.
IMPORTANTE
Indica información importante que usted debe conocer.
NOTA
Indica notas, consejos útiles, recomendaciones e información.
LEA TODAS LAS INSTRUCCIONES ANTES DE UTILIZAR ESTE EQUIPO.
ES - 16
PELIGRO
El oxígeno se puede quemar rápidamente. No fume cuando el concentrador de oxígeno está funcionando o cuando está cerca de una persona que está bajo
terapia de oxígeno.
Fumar durante la oxigenoterapia es peligroso y posiblemente produzca quemaduras en el rostro e incluso causar la muerte. No permita que se fume dentro
de una habitación en la que se encuentre el concentrador de oxígeno o cualquier accesorio que contenga oxígeno.
Si desea fumar, siempre debe apagar el concentrador de oxígeno, quitar la cánula y abandonar la habitación en la que se encuentren la cánula, la
mascarilla o el concentrador de oxígeno. Si no puede abandonar la habitación, debe esperar 10 minutos después de que se haya apagado el concentrador
de oxígeno antes de comenzar a fumar.
El oxígeno facilita que se produzca y se extienda un incendio. No deje la cánula nasal ni la mascarilla sobre acolchados ni cojines de silla si el concentrador
de oxígeno está encendido pero no está en uso, ya que el oxígeno podría provocar el incendio de los materiales. Apague el concentrador de oxígeno cuando
no lo esté utilizando para evitar la sobreoxigenación.
Mantenga el concentrador de oxígeno y la cánula a al menos 2 m (6,5 pies) alejados del calor, de los objetos con chispas o de fuentes de llama descubierta.
El fuego directo durante la oxigenoterapia es peligroso y posiblemente produzca un incendio e incluso provoque la muerte. Mantenga una distancia de 2 m
(6,5 pies) entre el fuego directo y el concentrador de oxígeno o cualquier accesorio que contenga oxígeno.
Los concentradores de oxígeno de Drive DeVilbiss están equipados con una salida de mitigación de fuego que evita la propagación del fuego hacia el interior
de la unidad.
ADVERTENCIA
Para evitar una descarga eléctrica, no conecte el concentrador a un tomacorrientes de CA si el gabinete del concentrador está roto. No quite el gabinete del
concentrador. Solamente un técnico calicado de Drive DeVilbiss debe quitar la cubierta. No aplique el líquido directamente en la cubierta ni utilice ningún
solvente derivado del petróleo o agentes limpiadores.
El uso inapropiado del cordón de suministro y de las clavijas puede causar quemaduras, incendios y riesgos de choque eléctrico. No use la unidad si el
cordón de suministro está deteriorado.
Asegúrese de que el cable de alimentación principal esté bien insertado en el conector del concentrador (230 voltios) y que el tomacorriente del cable de
alimentación esté completamente insertado en un tomacorriente de pared de CA que esté funcionando correctamente. En caso de no cumplirse, puede
ocurrir un riesgo relacionado con la seguridad eléctrica.
El uso de un dispositivo antiincendios como parte de los accesorios de oxígeno es opcional, y no está contraindicado para usarse con este dispositivo. Si se
usa con los accesorios de oxígeno, el dispositivo antiincendios debería ubicarse lo más cerca posible del paciente.
Ubique las tuberías de oxígeno y los cables de suministro eléctrico para evitar riesgos de tropiezos y reducir la posibildiad de enredos o estrangulamiento.
No aplique lubricante en las piezas, conexiones, tubos u otros accesorios del concentrador de oxígeno para evitar el riesgo de incendios y quemaduras.
No use lubricantes, aceites ni grasas.
Antes de realizar los procedimientos de limpieza apague la unidad “Off.”
Únicamente utilice lociones o ungüentos a base de agua que sean compatibles con el oxígeno antes y durante la oxigenoterapia. Nunca utilice lociones o
ungüentos a base de aceites o petróleo para evitar el riesgo de incendios y quemaduras.
Únicamente utilice piezas de repuesto que estén recomendadas por el fabricante para garantizar el funcionamiento adecuado y evitar el riesgo de incendios
y quemaduras.
Al usar el carro de trasvase con un dispositivo de trasvase, mantenga el sistema sobre una supercie plana en todo momento. Desmonte el sistema antes de
trasladarlo.
Si siente molestias o experimenta una emergencia médica mientras realiza la oxigenoterapia, obtenga asistencia médica de inmediato para evitar lesiones.
Los pacientes geriátricos, pediátricos o que no puedan comunicar molestias pueden necesitar monitoreo adicional o un sistema de alarma distribuido para
transmitir la información acerca de la molestia y/o la urgencia médica al cuidador responsable y evitar lesiones.
El uso de este dispositivo en una altitud superior a los 1524 m (5000 pies) metros o por encima de una temperatura de 35 ˚C (95 ˚F) o superior a los 93% de
humedad relativa puede afectar el nivel de ujo y el porcentaje de oxígeno y, por consiguiente, la calidad de la terapia. Consulte las especicaciones para
obtener detalles sobre los parámetros evaluados.
Para garantizar que recibe el suministro de oxígeno terapéutico adecuado según su enfermedad, el concentrador de oxígeno debe:
utilizarse únicamente después de que una o dos conguraciones se hayan determinado o recetado individualmente para usted según sus niveles de
actividad especícos.
utilizarse con la combinación especíca de piezas y accesorios de acuerdo con la especicación del fabricante del concentrador, y que se utilizaron
mientras se establecían las conguraciones.
Sus conguraciones de suministro del concentrador de oxígeno deben ser revaluadas periódicamente para garantizar la ecacia de la terapia.
Por su seguridad el concentrador de oxígeno debe usarse de acuerdo con la prescripción determinada por su médico.
Bajo ciertas circunstancias, la terapia de oxígeno puede ser peligrosa. Se aconseja pedir asesoría médica antes de usar el concentrador de oxígeno.
ADVERTENCIA
No es seguro para RM
No llevar el dispositivo ni los accesorios a un entorno de resonancia magnética (RM), ya que puede representar un riesgo inaceptable para el paciente o se
puede dañar el concentrador de oxígeno o los dispositivos médicos de resonancia magnética. El dispositivo y los accesorios no fueron evaluados en cuanto a
seguridad dentro de un entorno de resonancia magnética.
No usar el dispositivo ni los accesorios en un entorno con equipos electromagnéticos, como equipos de tomografía computarizada, diatermia, RFID y
sistemas de seguridad electromagnéticos (detectores de metal) ya que puede causar riesgos inaceptables para el paciente o se puede dañar el concentrador
de oxígeno. Algunas fuentes electromagnéticas pueden no ser aparentes. Si se notan cambios inexplicables en el funcionamiento de este dispositivo, si está
emitiendo sonidos inusuales o fuertes, desconectar el cable de suministro eléctrico y discontinuar el uso. Comuníquese con su proveedor de atención médica
domiciliaria. Este dispositivo es solo para uso exclusivo de un paciente.
Este dispositivo es adecuado para su uso en el hogar y en entornos de atención médica, excepto cerca del EQUIPO QUIRÚRGICO HF activo y la sala
blindada de RF de un SISTEMA ME para imágenes de resonancia magnética, donde la intensidad de las PERTURBACIONES electromagnéticas es alta.
Se debe evitar el uso de este equipo de forma adyacente o apilado con otros equipos, ya que podría provocar un funcionamiento incorrecto. Si tal uso es
necesario, se debe observar este equipo y el otro equipo para vericar que estén funcionando normalmente.
Los equipos de comunicación por radiofrecuencia portátiles (incluyendo periféricos como cables de antena y antenas externas) se deben utilizar a una
distancia no menor a 30 cm (12 pulgadas) de alguna parte del concentrador de oxígeno, incluyendo los cables especicados por el fabricante. De lo contrario,
se podría ocasionar la degradación del rendimiento de este equipo.
A-1025
ES - 17
PRECAUCIÓN
Es muy importante seguir su prescripción de oxígeno. No aumente ni disminuya el ujo de oxígeno - consulte a su médico.
La temperatura de la supercie de los oricios de ventilación en la parte inferior de la unidad podría superar los 41 °C (105,8 °F) bajo ciertas condiciones.
Cuando se utiliza el dispositivo bajo condiciones de funcionamiento extremas, la temperatura cerca de la zona del respiradero de escape en la parte inferior
de la unidad puede alcanzar los 59 ˚C (138,2 ˚F). Mantenga el cuerpo a una distancia de 76,2 cm (30 pulg.) de esta área como mínimo.
No se recomienda utilizar productos químicos fuertes (incluido el alcohol). Si es necesario realizar una limpieza con agentes bactericidas, se deberá utilizar
un producto que no contenga alcohol para evitar daños involuntarios.
IMPORTANTE
Se recomienda que el proveedor de cuidados en el hogar bloquee la perilla de control de ujo para evitar ajustes involuntarios. Una conguración de ujo
diferente a la recomendada puede afectar la terapia del paciente.
No someta a servicio ni limpie el dispositivo mientras es utilizado por un paciente.
No utilice medidores de ujo de salida baja con este concentrador.
El dispositivo está clasicado como IP21, lo que signica que tiene protección para evitar que los dedos ingresen a las piezas peligrosas, y para evitar que
las gotas de agua caigan en forma vertical.
Este equipo no resulta adecuado para ser utilizado en presencia de una mezcla anestésica inamable con aire, con oxígeno o con óxido nitroso.
Este dispositivo contiene componentes eléctricos y/o electrónicos. Para desechar los componentes del dispositivo, se deben seguir los planes de reciclaje y
las ordenanzas locales reguladoras.
GUARDE ESTAS INSTRUCCIONES.
INTRODUCCIÓN
Esta guía de instrucciones le informará acerca del concentrador de oxígeno de Drive DeVilbiss. Asegúrese de leer y entender esta guía antes de usar su unidad. Se indican
las normas de seguridad importantes a través de esta guía; preste atención especial a toda la información de seguridad. Contacte a su proveedor de equipo médico del hogar
Drive DeVilbiss en caso de tener cualquier duda.
Uso previsto
El uso previsto del concentrador de oxígeno de 10 litros de Drive DeVilbiss es proporcionar una terapia suplementaria de oxígeno de bajo ujo para pacientes que padecen
EPOC, alguna enfermedad cardiovascular y trastornos pulmonares. El concentrador de oxígeno se utiliza en entornos de tipo hogareños, como hogares, residencias para
ancianos y centros de cuidados para pacientes, entre otros.
Indicaciones de Uso
El Concentrador de oxígeno Drive DeVilbiss se debe utilizar como un concentrador de oxígeno para proporcionar una terapia suplementaria de oxígeno de bajo ujo en el
hogar, residencia para ancianos y centros de cuidados para pacientes, entro otros.
Contraindicaciones
El dispositivo no está diseñado para usarse como dispositivo de auxilio o soporte vital.
Rendimiento esencial
El rendimiento esencial del Concentrador de oxígeno es brindar un ujo continuo de gas enriquecido con oxígeno. Las alarmas visuales y sonoras indican que el dispositivo
no cumple con la especicación o que se ha detectado una falla.
Vida útil
La vida útil prevista de la unidad 1025 es de 5 años de funcionamiento, cuando se la utiliza según todas las indicaciones del fabricante para el uso, mantenimiento,
almacenamiento, manipulación y funcionamiento general seguros. La vida útil prevista de la unidad, y en particular de los lechos de tamiz y el compresor, puede variar según
el ambiente operativo, el almacenamiento, la manipulación y la frecuencia e intensidad de uso.
Por qué su médico le prescribió un suplemento de oxígeno
Hoy en día mucha gente tiene padecimientos del corazón, pulmones y de otras enfermedades respiratorias. Una gran cantidad de esa gente puede resultar beneciada con la
terapia suplementaria de oxígeno. Su cuerpo requiere un suministro estable de oxígeno para funcionar apropiadamente. Su médico le prescribió un suplemento de oxígeno
debido a que no está recibiendo suciente oxígeno del aire del cuarto. El suplemento de oxígeno aumentará la cantidad de oxígeno que su cuerpo recibe.
El suplemento de oxígeno no crea adicción. Su médico le prescribió un ujo especíco de oxígeno para corregir síntomas como dolor de cabeza, somnolencia, confusión,
fatiga o aumento de irritabilidad. Si estos síntomas persisten después de iniciar el programa de suplemento de oxígeno, consulte a su médico.
La conguración del suministro de oxígeno se debe determinar de forma individual para cada paciente y debe incluir la conguración del equipo que se utilizará, incluidos los
accesorios.
La ubicación y colocación adecuada de las entradas de la cánula nasal en la nariz es fundamental para la administración de oxígeno en el sistema respiratorio del paciente.
Sus conguraciones de suministro del concentrador de oxígeno deben ser revaluadas periódicamente para garantizar la ecacia de la terapia.
Cómo funciona su Concentradors de Oxígeno Drive DeVilbiss
Los concentradores de oxígeno son la fuente de suplemento de oxígeno más conable, eciente y conveniente que se puede conseguir hoy en día. El concentrador se opera
eléctricamente. La unidad separa el oxígeno del aire del cuarto lo que le permite tener un suplemento de oxígeno de alta pureza proporcionado a Ud. a través del oricio de
oxígeno. Aunque el concentrador ltra el oxígeno del cuarto, no afectará la cantidad normal de oxígeno en su cuarto.
A-1025
ES - 18
Vista frontal (Figura A)
1. Instrucciones para el funcionamiento
2. Interruptor de encendido
| = ON
O = OFF
3. Perilla del medidor de ujo
4. Medidor de ujo
5. Cortacircuitos – reinicializa la unidad después de un apagado por exceso de
corriente eléctrica.
6. Salida de oxígeno el oxígeno se dispersa a través de este oricio.
7. Luz (verde) de oxígeno normal (vea la página 20).
8. Luz (amarilla) de oxígeno bajo (vea la página 20).
9. Luz roja de necesidad de servicio cuando se enciende contacte a su
proveedor DeVilbiss.
10. Medidor de horas
Vista posterior (Figura B)
11. Manija
12. Escape
PRECAUCIÓN
Cuando se utiliza el dispositivo bajo condiciones de
funcionamiento extremas, la temperatura cerca de la zona del
respiradero de escape en la parte inferior de la unidad puede
alcanzar los 59 ˚C (138,2 ˚F). Mantenga el cuerpo a una distancia
de 76,2 cm (30 pulg) de esta área como mínimo.
13. Cordón eléctrico y/o Conector de suministro IEC (siglas en inglés de
“Comisión Electrotécnica Internacional”).
14. Tira de cordón recto
15. Puerta del ltro con ventilación y compartimento para el ltro de partículas
gruesas opcional.
16. Puerto auxiliar de oxígeno: El concentrador tiene un puerto auxiliar de
oxígeno que puede utilizarse para llenar cilindros de oxígeno con un
dispositivo de llenado de cilindros aprobado por la FDA que está diseñado
para utilizar oxígeno proveniente de un concentrador para llenar cilindros. El
puerto se utiliza únicamente con dispositivos de llenado aprobados por la
FDA con especicaciones de entrada de oxígeno compatibles. Consulte la
guía de instrucciones del dispositivo de llenado de cilindros para conocer las
especicaciones de entrada/salida de oxígeno, al igual que las instrucciones
de conexión y operación.
ACCESORIOS
Carro para dispositivo trasiego ..................................................................................................................................................................................................DeVilbiss 525DD-650
Humidificador de burbuja de flujo elevado (de 6 a 15 LPM) ...................................................................................................................................... Salter Labs 7900 o equivalente
Humidificador de burbuja de flujo bajo (hasta 6 LPM) ............................................................................................................................................... Salter Labs 7600 o equivalente
Cánula nasal de flujo elevado ............................................................................................................................................................................... Salter Labs 1600HF o equivalente
Existe una amplia variedad de humidicadores, tubos de oxígeno y cánulas/mascarillas que pueden utilizarse con este aparato. Algunos humidicadores y accesorios pueden
impedir el buen desempeño del aparato. Puede utilizar una mascarilla o cualquier cánula nasal con suministro de ujo continuo y el tamaño del accesorio variará según lo que
le haya recomendado su proveedor de cuidados de la salud que también deberá aconsejarle sobre el uso, el mantenimiento y la limpieza adecuados.
ADVERTENCIA
Al usar el carro de trasvase con un dispositivo de trasvase, mantenga el sistema sobre una supercie plana en todo momento. Desmonte el sistema antes de
trasladarlo.
NOTA El dispositivo solo debe utilizarse con humidicadores de burbuja diseñados para usarse con ujos de hasta 10 litros por minuto y una presión de 20 psi.
NOTA La distancia máxima permitida entre el concentrador y el paciente es la establecida por un tubo de oxígeno de 15 metros (50 pies) resistente a los golpes, una
cánula de 2,1 metros (7 pies) y un humidicador de burbujas.
NOTA El profesional de la salud debe vericar la compatibilidad del concentrador de oxígeno con todas las piezas utilizadas para conectarlo al paciente antes del uso.
A-1025
PARTES IMPORTANTES DE SU CONCENTRADOR
Sírvase tomarse un momento para familiarizarse con su concentrador de oxígeno Drive DeVilbiss antes de usarlo.
1
2
3
4
5
6
A
11
12
13
12
14
11
15
B
16
10
7
9
8
ES - 19A-1025
PREPARACIÓN DE SU CONCENTRADOR DE OXÍGENO
1. Coloque su unidad cerca de un tomacorriente en el cuarto donde está la mayor parte del tiempo.
NOTANo lo conecte a un tomacorriente que se controla por medio de un interruptor de pared.
PELIGRO
El oxígeno se puede quemar rápidamente. No fume cuando el concentrador de oxígeno está funcionando o cuando está cerca de una persona que está bajo
terapia de oxígeno. Mantenga el concentrador de oxígeno y la cánula a al menos 2 m (6,5 pies) alejados del calor, de los objetos con chispas o de fuentes de
llama descubierta.
2. Coloque la unidad sobre una supercie plana a 16 cm (6 pulg.) de la pared como mínimo ya que las cortinas o cualquier otro objeto
podrían impedir el ujo de entrada y salida adecuado del concentrador de oxígeno. El concentrador de oxígeno debería estar ubicado
en un área bien ventilada para evitar contaminantes y emisiones de humo.
NOTA Para mover la unidad, tome con rmeza la manija situada en la parte superior de la unidad, haciéndola rodar y/o
levantándola sobre los obstáculos que se encuentren en el camino.
3. Antes de utilizar la unidad, compruebe siempre que los conductos de ventilación de la puerta del ltro (situados en la parte posterior
de la unidad) estén limpios. La limpieza adecuada se trata en la sección Cuidar su concentrador en la página 20.
4. Conecte los accesorios de oxígeno apropiados a la salida de oxígeno.
Conexión del tubo de oxígeno:
a. Enrosque el conector provisto de salida de oxígeno a la salida del oxígeno.
b. Fije directamente el conducto de oxígeno al conector Figura 1.
Conexióntubodeoxígenoconhumidicación:
Si su médico le ha prescrito un humidicador de oxígeno como parte de su terapia, siga estos pasos (Si utiliza un prellenado, reérase
al inciso b.):
a. Llene la botella del humidicador según las instrucciones del fabricante.
b. Enrosque la tuerca de mariposa ubicada en la parte superior de la botella del humidicador en la salida de oxígeno de tal
manera que quede ja Figura 2. Asegúrese que esté apretada.
c. Conecte el tubo de oxígeno directamente a la unión de la salida de la botella del humidicador Figura 3.
NOTA(Su médico le ha prescrito una cánula nasal o una mascarilla facial. En la mayoría de los casos, vienen conectados ya en el
tubo de oxígeno. En caso de que no lo estén, siga las instrucciones del.
NOTAEl profesional de la salud debe vericar la compatibilidad del concentrador de oxígeno con todas las piezas utilizadas para
conectarlo al paciente antes del uso.
5. Saque el cordón de suministro por completo de la tira de cordón recto. Asegúrese que el interruptor de encendido esté apagado “Off,”
y enchufe la clavija en el tomacorriente de pared. La unidad está doblemente aislada para protegerla contra choques eléctricos.
ADVERTENCIA
Asegúrese de que el cable de alimentación principal esté bien insertado en el conector del concentrador (230 voltios) y que el
tomacorriente del cable de alimentación esté completamente insertado en un tomacorriente de pared de CA que esté
funcionando correctamente. En caso de no cumplirse, puede ocurrir un riesgo relacionado con la seguridad eléctrica.
NOTA (Sólo en los unidades de 120 Volts) La clavija del concentrador tiene una espiga más ancha que la otra. Para reducir el riesgo
de choque eléctrico, esta clavija se debe conectar a un tomacorriente de pared solamente de una manera. No intente modicar esta característica de seguridad.
NOTA Para vericar que el concentrador de oxígeno y los accesorios funcionen adecuadamente; 1. Verique el ujo de salida al colocar el extremo de la cánula nasal
debajo de la supercie de una taza de agua medio llena y observe las burbujas. 2. Compruebe si existen pérdidas en el sistema doblando las cánulas nasales y apretándolas
bien para detener el ujo de oxígeno. Observe el medidor de ujo para vericar que el indicador del medidor baje a cero. Si el indicador no baja a cero, revise todas las
conexiones para detectar posibles fugas. Las piezas que debe revisar para detectar fugas son: las conexiones de los tubos, la botella del humidicador y otros accesorios
como cortafuegos. Repita estos pasos hasta que el indicador del medidor baje a cero. Si encuentra algún problema, póngase en contacto inmediatamente con su distribuidor
o proveedor de servicio.
ADVERTENCIA
El uso inapropiado del cordón de suministro y de las clavijas puede causar quemaduras, incendios y riesgos de choque eléctrico. No use la unidad si el cordón de
suministro está deteriorado.
USO DE SU CONCENTRADOR
PELIGRO
El oxígeno se quema rápidamente. No fume cuando esté usando su concentrador de oxígeno o cuando esté cerca de una
persona que esté bajo terapia de oxígeno. Mantenga el concentrador de oxígeno y la cánula a al menos 2 m (6,5 pies)
alejados del calor, de los objetos con chispas o de fuentes de llama descubierta.
Los concentradores de oxígeno Drive DeVilbiss están equipados con una salida de mitigación de fuego que evita la
propagación del fuego hacia el interior de la unidad.
1. Pulse el interruptor para colocarlo en posición de encendido (“On”). Cuando la unidad está encendida, las tres luces (Servicio
requerido, Nivel bajo de oxígeno y Nivel normal de oxígeno [Service Required, Low Oxygen and Normal Oxygen]) del panel frontal se
iluminarán brevemente y se emitirá una breve alarma auditiva, para conrmar que las luces LED y la señal auditiva funcionan adecuadamente. La unidad operará en
modo de “arranque”, y la luz de Nivel bajo de oxígeno (Low Oxygen) permanecerá encendida hasta que se alcance un nivel normal de oxígeno. En ese momento,
permanecerá encendida la luz de Nivel normal de oxígeno (Normal Oxygen). El “arranque” puede tardar hasta 15 minutos.
NOTA Para una óptima vida útil de servicio, Drive DeVilbiss recomienda que el concentrador de oxígeno DeVilbiss se utilice durante al menos 30 minutos una vez
que se encienda. Los períodos de funcionamiento más breves, el funcionamiento bajo extremas temperaturas/condiciones de humedad o en presencia de agentes
contaminantes, o bajo condiciones de manipulación y almacenamiento fuera de las especicadas pueden afectar el funcionamiento conable a largo plazo del producto.
NO FUMAR
3
1
2
ES - 20
PELIGRO
El oxígeno facilita que se produzca y se extienda un incendio. No deje la cánula nasal ni la mascarilla sobre acolchados ni cojines de silla si el concentrador
de oxígeno está encendido pero no está en uso, ya que el oxígeno podría provocar el incendio de los materiales. Apague el concentrador de oxígeno cuando
no lo esté utilizando para evitar la sobreoxigenación.
NOTA Si la señal auditiva suena pero la unidad no está funcionando, la unidad no recibe alimentación. Consulte la Tabla de detección y solución de problemas
menores en la página 21 y contacte a su proveedor Drive DeVilbiss de ser necesario.
NOTA La unidad no funciona correctamente, si se percibe un sonido vibratorio de baja frecuencia. Remítase a la tabla sobre Detección y Solución de Problemas
Menores de la página 21, y póngase en contacto con su proveedor de Drive DeVilbiss si fuera necesario.
2. Revise la medidor de ujo para cerciorarse que la bola del medidor de ujo está centrada en la línea junto al número de su ujo prescrito.
PRECAUCION
Es muy importante seguir su prescripción de oxígeno. No aumente ni disminuya el ujo de oxígeno - consulte a su médico.
NOTA Su proveedor Drive DeVilbiss puede haber prejado el medidor de ujo de tal forma que no se pueda cambiar.
NOTA Si la perilla del medidor de ujo se gira en sentido de las manecillas del reloj, el ujo disminuirá (y eventual-mente cerrará el suministro de oxígeno). Si la
perilla se gira en sentido contrario de las manecillas del reloj, el ujo aumentará.
NOTAPara prescripciones de 10 LPM, asegúrese de que la bola esté centrada en la línea del litro 10 y de que no llegue a tocar la línea roja. Si se establece el
ujo por encima de 10, podría disminuir el nivel de pureza del oxígeno.
NOTA La alarma de ujo bajo se podrá activar si la bola del medidor de ujo está ajustada a menos de 2 lpm. La unidad
continuará funcionando; sin embargo, la luz de aviso de Mantenimiento requerido se encenderá y sonará la alarma. Ajuste el medidor
de ujo de acuerdo con el ujo prescrito.
IMPORTANTE
No utilice medidores de ujo de salida baja con este concentrador.
3. Su concentrador Drive DeVilbiss está listo para ser usado. Coloque la cánula correctamente con las cánulas nasales hacia arriba y
luego inserte las cánulas en la nariz. Envuelva el tubo de la cánula sobre los oídos y colóquelo frente al cuerpo (Figura 4). Espere 15
minutos para que el concentrador de oxígeno alcance el funcionamiento previsto.
Funcionamiento de DeVilbiss OSD
®
El OSD (siglas en inglés de “Aparato Detector de Oxígeno”) es un aparato dentro de su concentrador que monitorea el oxígeno producido en su unidad.
Las luces OSD en la parte superior del panel se denen de la siguiente manera:
Luz verde de oxígeno normal – nivel de oxígeno aceptable.
Luz amarilla de bajo oxígeno – por debajo del nivel de oxígeno aceptable.
Si la pureza del oxígeno cae por debajo del nivel aceptable: La luz verde de Oxígeno Normal [Normal Oxygen] se apaga, la luz amarilla de Oxígeno Bajo [Low Oxygen] se
ilumina y se emite una alarma auditiva intermitente.
Consulte la sección de Solución de problemas menores de esta guía en la página 21 y cambie al sistema de oxígeno de reserva. No intente realizar ningún otro tipo de
mantenimiento. Póngase en contacto inmediatamente con su proveedor de Drive DeVilbiss.
SISTEMA DE OXÍGENO DE RESERVA
Como precaución, su proveedor Drive DeVilbiss le debe proporcionar un sistema de oxígeno de reserva. En caso de falta de suministro de corriente o que la unidad tuviera
alguna falla, el sistema de alerta del paciente sonará para avisarle que utilice el sistema de oxígeno de reserva (si está provisto) y contacte a su proveedor Drive DeVilbiss.
CUIDADO DE SU DEVILBISS CONCENTRADORS DE OXÍGENO
Drive DeVilbiss recomienda usar solamente piezas y ltros originales DeVilbiss para garantizar un funcionamiento conable del producto.
ADVERTENCIA
No use lubricantes, aceites ni grasas.
Antes de realizar los procedimientos de limpieza apague la unidad “Off.”
Cánula/Mascarilla, Tubo y Botella del humidificador
Limpie y reemplace la cánula/mascarilla, tubo y botella del humidicador de acuerdo con las instrucciones del fabricante.
Puerta del filtro con conductos de ventilación
Inspeccione los conductos de ventilación periódicamente, y límpielos con un paño seco según sea necesario para eliminar el polvo.
Cubierta Exterior
Limpie el gabinete exterior del concentrador semanalmente con un paño húmedo y séquelo; los conductos de ventilación también pueden limpiarse con un paño húmedo.
Limpieza
Intervalo de limpieza
recomendado
Cantidad de ciclos de
limpieza* Método de limpieza compatible
Gabinete exterior 7 días 260 Utilice solamente un paño humedecido con agua
Conductos de ventilación de la
puerta del ltro
7 días 260 Limpie con un paño seco, o un paño humedecido con
agua para quitar el polvo.
Conector de salida de oxígeno 7 días 104 Detergente suave (2 cucharaditas) y agua tibia (2 tazas)
*cantidad de ciclos de limpieza determinados por el intervalo de limpieza recomendado y la vida útil esperada.
A-1025
4
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40

Drive DeVilbiss 1025DS Manuel utilisateur

Taper
Manuel utilisateur
Ce manuel convient également à