Minebea Intec Miras 2 IW Models Le manuel du propriétaire

Taper
Le manuel du propriétaire
98648-018-73
98648-018-73
Operating Instructions | Betriebsanleitung | Mode d’emploi |
Istruzioni per l’uso | Manual de instrucciones
Minebea Intec Miras 2
IW Models | IW Modelle | Modèles IW | Modelli IW | Modelos IW
2Miras 2
Miras 2
English page 3
In cases involving questions of
interpretation, the German-language
version shall prevail.
Deutsch Seite 35
Im Auslegungsfall ist die
deutsche Sprache maßgeblich.
Français page 67
En cas de questions concernant
l’interprétation, la version
en langue allemande fera autorité.
Italiano pagina 99
In caso di interpretazione dubbia,
fa testo la versione in lingua tedesca.
Español página 131
En caso de interpretación,
la versión en lengua alemana
será determinante.
2
Miras 2
Miras 2 3
Contents
Description Page
Safety Precautions 4
Getting Started 7
General View of the Equipment 8
Setting Up the Scale 9
Operating Design 12
Descriptions of the Keys 15
Configuration (Setup Menu) 16
Application Programs 18
Counting 18
Check Weighing 19
Toggling between Weight Units 20
Calibration/Span Adjustment 21
SBI Interface Protocol Descriptions 23
RS-232 Interface Diagram 26
Print Formats 27
Error Codes 31
Care & Maintenance 32
Specifications 33
Accessories (Options) 33
C Marking 34
33
Contents
Description Page
Safety Precautions . . . . . . . . . . . . . . . . . . . . . . . 4
Getting Started . . . . . . . . . . . . . . . . . . . . . . . . . 7
General View of the Equipment . . . . . . . . . . . . . 8
Setting Up the Scale . . . . . . . . . . . . . . . . . . . . . 9
Operating Design . . . . . . . . . . . . . . . . . . . . . . . . 12
Descriptions of the Keys. . . . . . . . . . . . . . . . . . . 15
Configuration (Setup Menu) . . . . . . . . . . . . . . . 16
Application Programs. . . . . . . . . . . . . . . . . . . . . 18
Counting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Check Weighing . . . . . . . . . . . . . . . . . . . . . . . . . 19
Toggling between Weight Units. . . . . . . . . . . . . . 20
Calibration/Span Adjustment . . . . . . . . . . . . . . . 21
SBI Interface Protocol Descriptions . . . . . . . . . . 23
RS-232 Interface Diagram. . . . . . . . . . . . . . . . . 26
Print Formats . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Error Codes . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Care & Maintenance . . . . . . . . . . . . . . . . . . . . . 32
Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Accessories (Options) . . . . . . . . . . . . . . . . . . . . . 33
CMarking . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
33
Contents
Description Page
Safety Precautions . . . . . . . . . . . . . . . . . . . . . . . 4
Getting Started . . . . . . . . . . . . . . . . . . . . . . . . . 7
General View of the Equipment . . . . . . . . . . . . . 8
Setting Up the Scale . . . . . . . . . . . . . . . . . . . . . 9
Operating Design . . . . . . . . . . . . . . . . . . . . . . . . 12
Descriptions of the Keys. . . . . . . . . . . . . . . . . . . 15
Configuration (Setup Menu) . . . . . . . . . . . . . . . 16
Application Programs. . . . . . . . . . . . . . . . . . . . . 18
Counting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Check Weighing . . . . . . . . . . . . . . . . . . . . . . . . . 19
Toggling between Weight Units. . . . . . . . . . . . . . 20
Calibration/Span Adjustment . . . . . . . . . . . . . . . 21
SBI Interface Protocol Descriptions . . . . . . . . . . 23
RS-232 Interface Diagram. . . . . . . . . . . . . . . . . 26
Print Formats . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Error Codes . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Care & Maintenance . . . . . . . . . . . . . . . . . . . . . 32
Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Accessories (Options) . . . . . . . . . . . . . . . . . . . . . 33
CMarking . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
33
Select Weight Units
Step Key (or instruction) Display
Select application setup (Fn/Up) > 2 sec
Select “Toggling Weight Units” (Fn/Up) repeatedly
Confirm selection (Print/Enter)
Select weight unit 1 (Fn/Up)
Confirm weight unit 1 (Print/Enter)
Select weight unit 2 (Fn/Up)
Confirm weight unit 2 (Print/Enter)
Toggling between Weight Units
Step Key (or instruction) Display
a. Select application program (Fn/Up) > 2 sec
b. Select “Toggling Weight Units” (Fn/Up) or (G/N/Down)
repeatedly
c. Confirm selection (Print/Enter)
d. Select weight unit 1 (Fn/Up) or (G/N/Down)
e. Confirm weight unit 1 (Print/Enter)
f. Select weight unit 2 (Fn/Up) or (G/N/Down)
g. Confirm weight unit2 (Print/Enter)
h. Place sample on the scale
i. Toggle weight unit (Fn/Up)
j. Unit toggling application: (Zero/Left) > 2 sec
Clear (unit changes to basic
unit set)
20
4Miras 2
Miras 2
Safety Precautions
4
Explanation of Symbols:
! Caution, risk of danger:
Read the accompanying text carefully.
The instrument is double-insulated.
Protective conductor terminal
Earth ground terminal
Safety tips, electrical protection class
! This instrument was built and tested in accordance with the safety regulations
for measuring and control instrumentation for protection class I (protective earth
connection) according to IEC 1010/ EN61010-1 or VDE 0411. The instrument was
delivered in safe condition. To maintain this condition and to ensure safe operation,
the operator must follow the instructions and warnings given in this documentation
Protective earth
! The connecting cable of the unit complies with the regulations in accordance with
VDE 0411 or EN61010. The mains plug must contain a protective earth conductor,
which must not be interrupted inside or outside this instrument (e.g. by using an
extension cable without protective earth). Before commissioning, acceptance of the
installation by a technically competent expert is required.
Measurement category
! This instrument is designed for measurement
category I, with a maximum of 8.5 V. To ensure safe operation do not use this
instrument for measurements within the measurement categories II, III or IV.
Miras 2
Miras 2 5
! Opening the instrument
CAUTION: DANGER TO LIFE!
Working on the instrument when it is
switched on can be dangerous to life.
Disconnect the instrument from the sup-
ply voltage! When removing covers of
parts by means of tools, live parts or ter-
minals may be exposed. Capacitors in the
unit may still be charged even after dis-
connecting the unit from all voltage
sources.
As the unit has no power switch, to com-
pletely remove AC power from the unit,
disconnect the AC power cable from the
mains socket.
Do not expose the scale unnecessarily to
aggressive chemical vapors or to extreme
temperatures,
moisture, shocks, or vibration.
Avoid exposing the scale to static elec-
tricity; be sure to connect the equipoten-
tial bonding conductor to the junction
box.
Repair and maintenance
Repairs are subject to checking and can
be carried out only at Minebea Intec. In
case of defect or functional trouble,
please contact your local Minebea Intec
dealer for repair. When returning the
instrument for repair, an exact and com-
plete fault description must be supplied.
Only a trained technician aware of the
hazards involved may carry out mainte-
nance work, during which the relevant
precautions must be taken.
Important note:
Make sure that the construction of the
instrument is not altered to the detri-
ment of safety. In particular, leakage
paths, air gaps (of live parts) and insulat-
ing layers must not be reduced.
The manufacturer cannot be held
responsible for personal injury or dam-
age caused by an instrument repaired
incorrectly by user or installer.
Only authorized service technicians who
have been trained by Minebea Intec and
who follow Minebea Intec’s standard
operating procedures for maintenance
and repair may open the scale. If you
see any indication that the scale cannot
be operated safely (for example, because
of equipment damage), turn off the
scale and lock it in a secure place so
that it cannot be used for the time
being.
If you use suction lifting equipment to
lift the load plate, always wear gloves,
hard-toed safety boots and protective
clothing. Warning: This procedure can
cause personal injury! Only reliable
personnel who are qualified to perform
such work are allowed to use suction
lifting equipment.
Always make sure the scale is discon-
nected from AC power before perform-
ing any installation,
cleaning, maintenance or repair work.
Check the pin assignment if you use
cables purchased from a different man-
ufacturer. Before connecting such a
cable to Minebea Intec equipment,
check the pin assignment on the corre-
sponding wiring diagram or chart and
disconnect any wires that are assigned
differently from those specified by
Minebea Intec.
The operator shall be solely responsible
for any damage or injuries that may
occur when using cables not supplied
by Minebea Intec.
5
6Miras 2
Miras 2
! Note:
Do not use this equipment in hazardous areas.
Make sure the voltage rating printed on the
power supply is identical to your local line
voltage.
Use only commercially available batteries
(rechargeable or dry-cell battery). If you use
a rechargeable battery, use only 12 V 2.3AH.
The scale is energized at all times unless you
disconnect the AC power and the battery.
Exposure to excessive electromagnetic interfer-
ence can cause the readout value to change.
Once the disturbance has ceased, the instrument
can be used again in accordance with its intend-
ed purpose.
6
Miras 2
Miras 2 7
Getting Started
Storage and Shipping Conditions
Do not expose the scale to shocks,
vibrations, moisture or extreme tem-
peratures.
Unpacking the Scale
After unpacking the scale, check it
immediately for any visible damage as
a result of rough handling during ship-
ment.
Note:
The display and control unit is attached
to the weighing platform via a cable.
If you see any sign of damage, proceed
as directed in the chapter entitled “Care
and Maintenance,” under the section on
“Safety Inspection.”
Save the box and all parts of the
packaging until you have successfully
installed your scale. Only the original
packaging provides the best protection
for shipment. Before packing your scale,
unplug all connected cables to prevent
damage.
Equipment Supplied
The equipment supplied includes the
components listed below:
Scale with attached display and control
unit
Load pan
Installation and operating instructions
Installation Instructions
The Minebea Intec Miras scales are
designed to provide reliable weighing
results under normal ambient condi-
tions. When choosing a location to set
up your scale, observe the following
so that you will be able to work with
added speed and accuracy:
Set up the scale on a stable, even sur-
face
Avoid placing the scale in close proxim-
ity to a heater or otherwise exposing the
scale to heat or direct sunlight
Protect the scale from drafts that come
from open windows or doors
Avoid exposing the scale to extreme
vibrations during weighing
Protect the scale from aggressive
chemical vapors
Do not expose the scale to extreme
moisture over long periods
Remove transport locking: see page 9
Conditioning the Scale
Moisture in the air can condense on
the surfaces of a cold scale whenever it
is brought into a substantially warmer
place. If you transfer the scale to a
warmer area, make sure to condition it
for about 2 hours at room temperature,
leaving it unplugged from AC power.
Afterwards, keep the scale continuously
connected to AC power.
77
8Miras 2
Miras 2
General View of the Equipment
1 Housing
2 Power cable
3 Display and control unit retainer
4 Display and control unit column
5 Column support foot
6 Level indicator (option)
7 Leveling feet
8 Stainless steel weighing pan
9 Power plug
10 Cable gland cover
(for optional RS-232 port outlet)
11 Load cell cable gland
12 Manufacturer’s ID tag
8
1
2
3
4
5
6
7
8
09
10
11
12
Miras 2
Miras 2 9
Setting Up the Scale
Connecting Electronic Peripheral Devices
§ Make absolutely sure to unplug the scale from AC
power or switch off (activate battery mode) before
you connect or disconnect a peripheral device (print-
er or PC) to or from the interface port.
Warm-up Time
To deliver exact results, the scale must warm up for
at least 30 minutes after
initial connection to AC power. Only after this time
will the scale have reached the required operating
temperature.
Transport Lock
Under the metal frame structure there are four
safety overload protection pins (Note: not on all
platform models).
Before using the scale, rotate the nut (upwards) until
tight.
Place the load plate on the scale
9
10 Miras 2
Miras 2
Leveling the Scale
Remove the load plate
Using the 4 foot screws, level the scale so that the
air bubble is centered within the level indicator
Check to ensure that all leveling feet rest securely on
the work surface
10
Miras 2
Miras 2 11
Assembly Procedure
1. Affix the column bracket with 4 screws
2. Install the column support leveling foot
3. Push the load cell cable into the column
4. Insert the column into the bracket holder
5. Affix the column with 2 screws
6. Insert the display and control unit retainer into the
column
7. Tighten the screw on the display and control unit
bracket
Battery-Powered Operation
To activate battery operation:
Open the battery cover at the back of the Indicator.
Insert 1.5 V size D dry-cell batteries (not included)
into the battery compartment.
The battery symbol is displayed when the unit is
operating on battery power.
If the battery symbol flashes, the battery power is
weak.
Replace the batteries (if using dry-cell batteries).
11
4
5
7
6
1
2
3
12 Miras 2
Miras 2
Operating Design
The scales in the Miras series consist
of a weighing cell and a display and
control unit. In addition to the choice
of power supply or rechargeable main-
tenance-free lead acid battery or dry
battery.
The display and control unit is affixed
to the weighing platform. Operation of
the Miras scale is simple and uniform.
The Miras models include the features
like gross/net toggling, counting,
checkweighing & unit toggling.
Display
The display is divided into sections:
Busy Symbol, Plus/Minus Sign
If the symbol displayed here is a dia-
mond b, this indicates that the scale
is performing a function (busy symbol).
A plus or minus sign (+ or –) refers to
the value displayed.
Measured Value
Shows the measured value
Basic Unit and Additional Information
When the scale has stabilized, the
weight unit is displayed here (g, kg,..).
The arrows below the weight indications
shows the weighing mode (gross/net)
and the application mode (counting,
checkweighing, or unit toggling) ena-
bled, if any.
The ! symbol indicates the unit is
not in weighing mode (e.g.: calibration
mode, configuration mode, error mode)
12
Miras 2
Miras 2 13
Operating Limits
Never exceed the maximum capacity of the scale.
The maximum loading capacities of the scales in this series are
listed in the table below and depend on the position of the load
on platform:
Model Width Length Center* Side Corner
(mm) (mm) kg kg kg
IW2P1.–6ED… 300 4000 50 35 20
IW2P1.–15ED… 300 400 130 85 45
IW2P1.–30ED… 300 400 500 400 200
IW2P1.–60ED… 300 400 600 400 200
IW2P1.–60FE… 400 500 600 400 200
IW2P1.–150FE… 400 500 600 400 200
IW2P1U–150GF… 500 650 600 400 200
IW2P1.–300GF… 500 650 600 400 200
13
14 Miras 2
Miras 2
Scale connected to AC power:
On/Off Switch:
On: Scale is operational and display
shows function or mode that is in use
Off: Depends on menu setting (“4.
Standby option.”), scale is completely
off (no display and no internal circuitry)
or in standby (no display and internal
circuitry working).
Back Light:
Depends on menu setting (“2. Back-
light”). Backlight is on/off.
Auto Power Off:
Not active.
Scale running on battery power:
On/Off Switch:
On: Scale is operational and the display
shows the function or mode that is in
use
Off: Completely off (no display and no
internal circuitry working)
Back Light:
Depends on menu setting (“2. Back-
light”). Backlight is on/off. Backlight
goes off after 15 seconds if no opera-
tion is done (no key pressed/no weigh-
ing). Backlight comes back on whenever
a key (except on/off) is pressed or when
a load is placed on the scale.
Auto Power Off:
Depends on menu setting (“3. Auto
power off”); system power shuts down.
Time limit options for auto power in
setup menu are 5, 10 or 15 minutes,
or can be disabled.
14
Power Management Option
Miras 2
Miras 2 15
15
1) On/Off
“On/off” function will depend on
different power supply conditions.
Use the setup menu for backlight
control and for auto power off.
In Main Mode
§ Off-Mode:
If standby active: Scale is in standby
mode, but there is no display visible
other than mode symbol (no backlight,
no display, no symbols)
If standby not active:
Scale is completely off
§ On-Mode:
Scale is in operation and the display
shows the function that is in use.
In this mode, auto power off function
is invalid.
In Battery Mode
§ Off-Mode: Scale is completely off.
§ On-Mode: Scale is in operation and the
display shows the function that is in use.
Backlight works based on setup menu
In this mode, auto power off function
is valid. After 5, 10 or 15 minutes, the
system shuts off when not in use in
accordance with the setup menu setting.
2) Zero/Arrow-Left (Zero/Left)
Weighing mode: Set scale to zero
Parameter mode: Shift position of digit
to the left
3) Tare/Arrow-Right (Tare/Right)
Weighing mode: Tare the scale
Parameter mode: Shift position of digit
to the right
4) Gross-Net/Arrow-Down (G/N/Down)
Weighing mode: Toggle weight unit
gross/net
Counting mode: Toggle weight unit
gross, net and pcs
Function mode: To scroll down menu
selection
Parameter mode: Reduce digit value by 1
5) Function/Arrow-Up (Fn/Up)
Weighing mode: Switch the scale from
normal weighing mode to application
setup mode
Function mode: Scroll up in menu
selection
Parameter mode:
Increase digit value by 1
6) Print/Enter (Print/Enter)
Weighing mode: Print key
Function mode: Scroll level up in menu
mode/confirm selected menu.
Descriptions of the Keys
16 Miras 2
Miras 2
16
Configuration (Setup Menu)
To configure the user interface of the scale to individual requirements
Step Key (or instruction) Display
1. Switch off the scale (On/Off)
2. Switch on the scale (On/Off)
3. While all the segments
are displayed: (Zero/Left) > 2 sec
Navigation in the Setup Menu
Key Function
(Print/Enter) Menu item: Confirm setting
(Fn/Up) Menu item: Scroll up
Digit input: Increase by 1
(G/N/Down) Menu item: Scroll down
Digit input: Reduce by 1
(Zero/Left) Digit input: Selection shift left
(Tare/Right) Digit input: Selection shift right
Miras 2
Miras 2 17
17
Setup Menu 1. Serial COM 1.1 Baud rate 1.1.1 600
1.1.2 1,200
1.1.3 2,400
1.1.4 4,800
1.1.5 9,600
1.1.6 19,200
1.2 Parity 1.2.1 7-bit-None
1.2.2 7-bit-Odd
1.2.3 7-bit-Even
1.2.4 8-bit-None
1.2.5 8-bit-Odd
1.2.6 8-bit-Even
1.3 Print option 1.3.1 Short
1.3.2 Long/block
1.3.3 Continuous
2. Backlight 2.1 On
2.2 Off
3. Auto Power Off 3.1 No options
3.2 5 Min.
3.3 10 Min.
3.4 15 Min.
4. Standby Options 4.1 On
4.2 Off
5. Adapter Filter 5.1 Very stable conditions
5.2 Stable conditions
5.3 Unstable conditions
5.4 Very unstable conditions
6. ADC Configuration 6.1 Decimal point 6.1.1 0
6.1.2 1
6.1.3 2
6.1.4 3
6.1.5 4
6.2 Basic unit 6.2.1 g
6.2.2 kg
6.2.3 lb For some countries,
6.2.4 oz these units
6.2.5 t are disabled
6.3 Step width 6.3.1 1
6.3.2 2
6.3.3 5
6.4 e<>d frame 6.4.1 On
6.4.2 Off
6.5 Adjust FSD (span) 6.5.1 (Enter FSD weight)
6.6 Calibration weight 6.6.1 (Enter CAL weight)
6.7 Linearisation weight 6.7.1 (Enter Lin weight)
6.8 Calibration unit 6.8.1 g
6.8.2 kg
6.8.3 lb
7. Reset Menu 7.1 Do not restore
7.2 Restore
• = Factory setting
18 Miras 2
Miras 2
18
Application Programs
Counting
Step Key (or instruction) Display
a. Select application program (Fn/Up) > 2 sec
b. Select “Counting” (Fn/Up) or (G/N/Down)
repeatedly
c. Press Print/Enter. (Print/Enter)
Display shows reference piece
selection menu.
Counting is blinking.
d. Automatically the display
will show the reference pieces. (Tare/Right)
Place container and press Tare
if required (display “tared”).
e. Select reference sample quantity: (Fn/Up) or (G/N/Down)
5, 10, 20, 50 or 100, and put repeatedly
the reference on the scale.
(quantity flashes)
f. Press Print/Enter . Place reference sample on scale
Now “O” flashes. Automatic optimization is
active. (Print/Enter)
g. Place more reference samples
on scale (The number is
x+2 ~ 2x times the reference
pieces (x) selected in step e.)
When the sample is still, the
reference sample weight is calculated.
The flashing “O” goes out.
h. Place uncounted parts/samples
on the scale
i. Toggle between reference piece weight, (G/N/Down)
total weight, and total pieces.
j. Unload the scale
k. Counting application: Clear (Zero/Left) > 2 sec
Miras 2
Miras 2 19
19
Check Weighing
Step Key (or instruction) Display
a. Select application program (Fn/Up) > 2 sec
b. Select “Checkweighing” (Fn/Up) or (G/N/Down)
repeatedly
c. Press Print/Enter to confirm. (Print/Enter)
“LO” flashes.
d. Set lower limit (Fn/Up) or (G/N/Down) and/or
(Zero/Left) or (Tare/Right)
e. Press Print/Enter to confirm (Print/Enter)
and save. “HI” flashes
f. Set upper limit, (G/N/Down) and/or
and then press Print/Enter. (Zero/Left) or (Tare/Right)
g. Press Print/Enter to confirm (Print/Enter)
and save.
h. Place the sample on the scale.
Display shows the difference
and HI/LO symbol
i. Unload the scale
j. Checkweighing application: Clear (Zero/Left) > 2 sec
20 Miras 2
Miras 2
Toggling between Weight Units
Step Key (or instruction) Display
a. Select application program (Fn/Up) > 2 sec
b. Select “Toggling Weight Units” (Fn/Up)
repeatedly
c. Confirm selection (Print/Enter)
d. Select weight unit 1 (Fn/Up)
e. Confirm weight unit 1 (Print/Enter)
f. Select weight unit 2 (Fn/Up)
g. Confirm weight unit2 (Print/Enter)
h. Place sample on the scale
i. Toggle weight unit (Fn/Up)
j. Unit toggling application: (Zero/Left) > 2 sec
Clear (unit changes to basic
unit set)
20
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114
  • Page 115 115
  • Page 116 116
  • Page 117 117
  • Page 118 118
  • Page 119 119
  • Page 120 120
  • Page 121 121
  • Page 122 122
  • Page 123 123
  • Page 124 124
  • Page 125 125
  • Page 126 126
  • Page 127 127
  • Page 128 128
  • Page 129 129
  • Page 130 130
  • Page 131 131
  • Page 132 132
  • Page 133 133
  • Page 134 134
  • Page 135 135
  • Page 136 136
  • Page 137 137
  • Page 138 138
  • Page 139 139
  • Page 140 140
  • Page 141 141
  • Page 142 142
  • Page 143 143
  • Page 144 144
  • Page 145 145
  • Page 146 146
  • Page 147 147
  • Page 148 148
  • Page 149 149
  • Page 150 150
  • Page 151 151
  • Page 152 152
  • Page 153 153
  • Page 154 154
  • Page 155 155
  • Page 156 156
  • Page 157 157
  • Page 158 158
  • Page 159 159
  • Page 160 160
  • Page 161 161
  • Page 162 162
  • Page 163 163
  • Page 164 164

Minebea Intec Miras 2 IW Models Le manuel du propriétaire

Taper
Le manuel du propriétaire