-4-
IRREGULAR PULSE RHYTHM SYMBOL
Pulse rhythm can be disturbed from talking,
moving or arrhythmias. This product displays
[] when the difference of shortest and
longest time intervals between pulse waves is
over 25%, indicating irregular pulse rhythm.
MEMORY FUNCTION
The measured values are automatically saved for later review in either one of two memory
banks, [1] or [2]. These two memory banks can be used to save readings of two persons
separately or to save morning readings and evening readings separately.
ach bank can save up to 60 readings. When the number of stored readings reaches 60, the
oldest reading will be deleted to record new reading.
Saved readings are recalled with measurement date and time when clock function is
activated. *The clock needs to be activated before measurement for date and time to be
saved with the measured valued.
[Err] results are not saved.
CARE AND MAINTENANCE
Keep the product clean. Inspect its cleanliness after use. To clean, use only a soft dry cloth.
Do not use gasoline, paint thinner, or other strong solvents. Since the cuff may absorb
perspiration and other uids, inspect it for stain and discoloration after each use. When
cleaning the cuff, use a synthetic detergent and gently rub the surface. Air dry thoroughly.
ake sure uid never gets in AI HS. Do not machine wash or scrub it.
When storing the product, do not place heavy objects on it. Do not fold or bend the cuff
forcibly. To disconnect the cuff from the main unit, do not pull AIR PLUG but hold AIR
PLUG and unplug it. Do not coil AIR HOSE too tightly. In case the product is stored in the
environment with ambient temperature above 0 or below 10, please leave it for at least
2 hours before taking a measurement. Remove the batteries if the product is to be stored for
an extended period of time. Keep the batteries out of reach of children.
Stop using the product immediately and contact your dealer or the manufacturer in case any
regular pulse rhythm
An example of irregular pulse rhythm
compared to regular pulse rhythm
visible damage is found on the device.
We suggest that you have your blood pressure monitor checked every 2 years. This operation
may only be performed by the manufacturer or by rms authorized by the manufacturer.
PRODUCT SPECIFICATIONS
Model : DSK-1011
Operating principle : Oscillometric method
Indicator : 15 digits liquid crystal display
Pressure indicating range : 0 to 300 mmHg cuff pressure
Measuring range : 50 to 250 mmHg systolic, 0 to 180 mmHg diastolic,
0 to 160 bpm pulse rate
Accuracy* : 3 mmHg cuff pressure, 5 of reading pulse rate
Ination :Automatic ination
Exhaust : Automatic exhaust valve
Power supply : Four 1.5 volt 6 AA alkaline batteries or A adaptor
AD-W5 series
Electric rating : D6V/W, with four 6 batteries
Adaptor A100-20V, 50-60Hz, 0.12A,
Unit; DC6V, 500mA, with designated AC adaptor
Memory : 2 banks, each stores 60 readings
Operating condition : 10 to 0, 15 to 85 H noncondensing
Transportation/storage condition : -20 to 60, 10 to 95 H noncondensing
Cuff :
Coverage arm circumference; 22 to 32 cm, applied part; BF
Main unit : Weight Appro. 250 g without batteries
Size Appro. 115 115 65.9 mm W D H
Protection class IP : IP 20: Protected against solid foreign particles with a
diameter of more than 12.5 mm, no protection against
water.
Protection against electric shock : Internally powered equipment/Class II equipment, Type
BF applied part
Mode of operation : Continuous operation
lassication : Class II / Internally powered equipment