61-337 Multimeter
Operation and Safety Manual
IDEAL
®
Test and Measurement
Introduction ...................................................................3
Contacting IDEAL INDUSTRIES, INC ......................................3
Safety Information ...........................................................4
Warnings ..............................................................................................4-5
Cautions ..................................................................................................5
Symbols ...............................................................................................6-7
Operation..................................................................8-25
Identification and description of operating controls and
functions .............................................................................................8-9
Operating Features ...........................................................................10-11
Using Test Leads ................................................................................... 12
Meter Operation ..............................................................................13-19
Non-Contact Voltage Testing ..........................................................13
Lead Warning..................................................................................14
Fuse Warning..................................................................................14
SEL Button ......................................................................................14
Measuring Voltage ..........................................................................15
Measuring Continuity ..................................................................... 16
Measuring Resistance .....................................................................16
Measuring Capacitance ..................................................................17
Measuring Diodes ..........................................................................17
Measuring Frequency ..................................................................... 18
Measuring Temperature ..................................................................18
Measuring Micro Amps .................................................................. 19
Measuring Milli Amps .................................................................... 19
Measuring Amps ............................................................................19
Functions Operation Table ................................................................20-21
Functions Indication Table ...............................................................22-23
Electrical Specifications ...................................................................24-25
Environmental Specifications ............................................ 26
Mechanical Specifications ............................................... 26
EMC / EMI ................................................................... 26
FCC ........................................................................ 27
Safety ........................................................................ 27
Maintenance and Service ............................................ 27-28
Table of Contents
2
Introduction
The IDEAL
®
61-337 Digital Multimeter is an auto ranging average responding
RMS meter that measures AC and DC current (amps) in series via test leads in
the designated terminals, measures voltage, frequency, resistance, continuity,
capacitance, diodes via test-leads in the designated terminal and measures
temperature via a K-Type thermocouple. It also detects the presence of voltage
between 40V to 600V AC via a non-contact sensor in the top center of the meter.
Contacting IDEAL INDUSTRIES, INC.
To contact IDEAL INDUSTRIES, INC., call one of the following telephone numbers:
IDEAL Industries USA Customer Service
• Phone Number: 800-435-0705
IDEAL Industries Canada Customer Service
• Phone Number: 905-683-3400
IDEAL Industries EMEA
• Phone Number: +44 (0)1925 444 446
• Email: eur[email protected]
IDEAL Industries Australia
• Phone Number: +61 3 9562 0175
Or visit the IDEAL Electrical Website at www.idealind.com
To register your product, find manuals, watch videos, simply scan this QR code.
3
Arc Flash and Shock Hazard, Proper PPE Required. Follow all safety procedures,
wear proper PPE in accordance to NFPA 70E. Read and fully understand the
instruction manual prior to using this product. Failure to comply can result in
serious injury or death.
Warning - Identifies conditions and actions that could result in
possible death or serious injury if the hazard is realized.
Caution - Identifies conditions and actions that could result in meter dam-
age, equipment under test damage or data loss if the hazard is realized.
WARNING
Safety Information
Arc Flash and Shock Hazard, Proper PPE Required. Follow all safety procedures,
wear proper PPE in accordance to NFPA 70E and follow the guidelines below and
the instructions in this manual when operating the meter. Failure to comply can
result in serious injury or death.
Choking Hazard, Small Parts. Keep Away from Children. Sharp Objects
Hazard, This is not a toy. It is not for use or play by children. Keep Away
from Children. Failure to do so can result in serious injury.
Only experienced or technically competent consumers should use this
equipment. When in doubt, call an experienced electrician to make any and
all necessary repairs or installations. At all times, perform any necessary
work on a de-energized circuit that has had its circuit breaker turned off and
has been locked out.
Use the Meter only as specified in this manual or protection provided by the
Meter can be compromised.
Before using or connecting the Meter, visually inspect it to ensure the cases
are not cracked and the back case is securely in place. Do not use if the
Meter appears damaged.
Before using the test leads, inspect carefully for damaged insulation,
exposed metal or cracked probes. Check test leads for continuity. Do not use
leads if they appear damaged.
Use only approved test leads. Do not use improvised connections that could
present a safety hazard.
When using the probes, keep fingers behind the guard ring on the probes.
Connect the common test lead before connecting the live test lead. When
disconnecting test leads, disconnect the live test lead first.
This Meter is intended for use by qualified electricians. Follow NFPA 70E
Standards for Electrical Safety in the Workplace when using this Meter.
Do not use without the batteries correctly in place and the battery door
closed and secured.
Do not use Meter if it operates incorrectly as protection may be
compromised. When in doubt, have the Meter serviced.
When servicing the Meter, use only specified replacement parts.
4
CAUTION
5
Arc Flash and Shock Hazard, Proper PPE Required. Follow all safety procedures,
wear proper PPE in accordance to NFPA 70E and follow the guidelines below and the
instructions in this manual when operating the meter. Failure to comply can result in
serious injury or death.
Have the Meter serviced only by qualified service personnel.
Do not use the Meter around explosive gas, dust, or vapor, or during
electrical storms, or in wet environments.
When measuring, keep fingers behind the Tactile Barrier. See “The Meter”
on pg. 8 and 9.
Do not apply more than the rated voltage, as marked on the Meter, between
the terminals or between any terminal and earth ground.
To avoid false readings that can lead to electrical shock and injury, replace
the batteries as soon as the low battery indicator ( ) appears.
Remove the test leads from the circuit prior to removing the battery door.
Voltages exceeding 30VAC or 60VDC pose a shock hazard so use caution.
Always ensure that test leads are secured so that they cannot be accidentally
snagged or tripped over.
Do not work alone so that assistance can be rendered in an emergency.
Use extreme caution when working around bare conductors or bus bars.
Contact with the conductor could result in electric shock.
Adhere to local and national safety codes. Individual protective equipment
must be used to prevent shock and arc blast injury where hazardous live
conductors are exposed.
Disconnect circuit power and discharge all high-voltage capacitors before
you measure resistance, continuity, or capacitance.
Never operate the Meter with the back cover removed or the case open.
Cancer and Reproductive Harm - www.P65Warnings.ca.gov
Meter damage, equipment under test damage or data loss can occur if the following
guidelines are not adhered to.
Use the proper terminals, function, and range for the measurement
application.
Clean the case and accessories with a damp cloth and mild detergents only.
Do not use abrasives or solvents. Make sure the meter is completely dry
before use.
CAUTION
WARNING
Symbols & Descriptions
SYMBOL DESCRIPTION
Arc Flash and Shock Hazard
Shock Hazard
Warning or Caution
Choking Hazard
AC (Alternating Current)
DC (Direct Current)
Low Battery Indicator
Earth Ground
10 A Maximum Current Specification
CAT III
IEC Measurement Category III
CAT III has protection against transients in equipment in fixed-
equipment installations such as distribution panels feeders,
and short branch circuits. Also included are lighting systems in
larger buildings.
NCV
Non-Contact Voltage Sensing
NCV
Non-Contact Voltage Sensing Point
A
Amperage AC and DC
Hz
Frequency measured via the test leads
V
Voltage AC or DC
Hz%
Frequency displayed as % measured with AC Voltage
6
SYMBOL DESCRIPTION
Ω
Ohms
Continuity
Capacitance
Diode
°F °C Temperature Degrees Fahrenheit / Celsius
LCD Liquid Crystal Display
Range Auto or Manual Range Selection
Delta (Relative)
Double Insulation
Do not dispose of this product as unsorted municipal waste.
It must be properly disposed of in accordance with local
regulations. Please see www.epa.gov or www.erecycle.org for
additional information.
Conforms to applicable North American Safety Standards
Conforms to applicable Australian Safety Standards
Conforms to European Directives
NOTE: The Measurement Category (CAT) and voltage rating of any combination
of test probe, test probe accessory, current clamp accessory, and the Meter is the
LOWEST rating of any individual component.
7
Identification and Description of Operating
Controls and Functions for the 61-337 Digital
Multimeter:
1. HV, & Continuity LED
2. LCD Display
3. Tactile Barrier
4. Backlight Button
5. Rubber Boot
6. Volts/Ohms Input Terminal
7. Common (COM) Input Terminal
8. Mili Amps, Micro Amps Input Terminal
9. Amps Input Terminal
10. Measuring Functions Dial
11. Function Select Button
12. Relative Button
13. NCV Sensing Point
14. Range
15. Hold
8
Operation
4
1
3
15
6
11
10
2
5
12
14
9
7
8
9
13
High Voltage Warning (HI-V)
The meter beeps once (for 1 second) and a red LED illuminates and remains on as
long as the voltage remains above 30V AC or DC, or when the meter’s voltage range
is exceeded.
NOTE: This feature does not work in the Ohm, capacitor, continuity or amperage
modes.
Data Hold Feature
Press the Hold button to toggle in and out of the data hold mode. “H” appears in the
upper left of the meter display when data hold is active. Use the data hold feature to
lock a measurement reading on the display. Press the Hold button again to unlock
the display and obtain a real-time reading.
REL Feature ()
A short press of the ‘’ button will activate or deactivate the REL function.
Range
The meter is auto ranging, however the user may select a specific range by pressing
the range button repeatedly.
Auto Power Off (APO) Feature Disable
The meter automatically powers itself down after about 30 minutes of no use. Press
any button, and the meter will wake up and enter the default function of that setting
before power down. To Disable APO, press and hold the SEL button while turning
the dial to any desired function. When APO is defeated, the “APO” will be removed
from the display. Turning the meter off and back on will restore the APO default.
10
Operating Features
11
Press the Backlight button on the meter to turn the backlight on and off. The white
backlight will remain lit for about 5 minutes before it automatically turns off to
conserve battery power. Or turn the backlight off by pressing the button again.
Backlight
Backlight is selectable
to be on in all
functions.
WARNING: Arc Flash and Shock Hazard, Proper PPE Required.
Follow all safety procedures, wear proper PPE in accordance to NFPA 70E and
assure that the Protective Caps are in place when operating a properly rated
electrical meter/tester using the TL-757 Test Leads in a CAT IV 600V or CAT III
1000V environment.
This meter is intended for use with the IDEAL TL-757 lead set (provided with this
product) or equivalent. The lead set must comply with requirements for Overvoltage
and Measurement Categories CAT IV 600V CAT III 1000V.
12
Using Test Leads
WARNING: Arc Flash and Shock Hazard, Proper PPE Required.
Follow all safety procedures, wear proper PPE in accordance to NFPA 70E and
follow the guidelines below and the instructions in this manual when operating
the meter with TL-757 Test Leads or equivalent. Test Leads must be rated for the
electrical environment the meter is being used in and have a voltage rating of at
least the voltage of the circuit to be measured. Failure to comply can result in
serious injury or death.
Choking Hazard, Small Parts. Keep Away from .....
Children. Sharp Objects Hazard, This is not a toy. It is not for use or play by
children. Failure to do so can result in serious injury or death.
Use only approved test leads. Do not use improvised connections that could
present a safety hazard.
Ensure that the test leads are inserted into the correct input jacks when
measuring AC or DC current.
Prior to using the test leads, inspect them carefully for damaged insulation,
exposed metal or bent probes. Check test leads for continuity. Do not use
leads if they appear damaged.
When using the probes, keep fingers behind the guard rings on the probes.
Connect the common test lead before connecting the live test lead. When
disconnecting test leads, disconnect the live test lead first.
Always ensure that test leads are secured so that they cannot be accidentally
snagged or tripped over.
Note: The 61-337 is only rated to 600V AC or DC MAX.
This meter is CAT III 600V ONLY
Protective Cap Guard Ring
CAT III 1000V, CAT IV 600V (with cap on)
WARNING: To prevent possible electrical shock or personal injury, the
protective caps must be in place when operating a properly rated electrical meter/
tester using the TL-757 Test Leads in a CAT IV 600V or CAT III 1000V environment.
Meter Operation
13
Non-Contact Voltage Sensing
First, rotate the function key to the NCV position. Place the sensing point marked
with NCV close to an AC outlet (or any AC conductor such as light switches or
power cords) and scan back and forth across the outlet. The meter beeps On/Off
continuously and the Red NCV LED above the display flashes if the sensing antenna
detects live voltage greater than 40V AC (50 -60 Hz). Voltages with frequencies
higher than 60Hz or electrostatic charges may also be detected by the NCV sensing
antenna. To differentiate between hot and neutral in an outlet, place the NCV tab
directly next to each slot in the outlet. The tone (buzzer) will sound over the slot
that is energized and not on the neutral slot. Either test lead can also be used to
differentiate between the hot and neutral. Plug the red or black test lead into the
V input jack on the meter. With the function switch in the NCV position, insert the
probe end of just one probe into the slots on the outlet. The meter will beep and the
Red LED will flash when a hot conductor is contacted.
NOTE: While the NCV is a helpful function, it is ALWAYS
RECOMMENDED that the operator verify that any electrical conductor is
completely de-energized and that no voltage is present by measuring
for voltage AND CONFIRMING THAT NO VOLTAGE IS PRESENT and that
all applicable PPE and lock out tag out procedures be followed before
attempting any work on ANY electrical distribution system.
Note: The 61-337 is only rated to 600V AC or DC MAX.
This meter is CAT III 600V ONLY
Protective Cap Guard Ring
CAT III 1000V, CAT IV 600V (with cap on)
WARNING: To prevent possible electrical shock or personal injury, the
protective caps must be in place when operating a properly rated electrical meter/
tester using the TL-757 Test Leads in a CAT IV 600V or CAT III 1000V environment.
14
SEL Button
Lead Warning
Fuse Warning
Lead Warning: When inserting the test
lead into any amperage input terminal, but
the rotary function switch IS NOT set to
the correct relative amperage position, the
meter will beep at 3Hz, LCD will display
LEAd“ and “ .”, flashes at 3Hz. If any
amperage function is selected first, but the
test lead IS NOT in the relative amperage
input jack,the meter will also beep at 3Hz
and the LCD will display “LEAd” and
.”, flashes at 3Hz.
Fuse Notification: The test leads
CANNOT be short circuited or used for
measuring while testing the fuse.
1. If the 400mA/600V fuse is blown,
setting the rotary switch to the “mA”
or “uA” setting, the LCD will display
FUSE” and “ .”, and the buzzer
will beep continuously at the same time.
2. If the 10A/600V fuse is blown, setting
the rotary switch to the “A” setting, the
LCD will display “FUSE” and “ .”,
and the buzzer will beep continuously at
the same time.
The SEL Button allows the user to toggle between multiple
functions when the function dial is used to select a specific
function. This can include but is not limited to toggling between
degrees F and C, AC and DC voltage or current, continuity,
resistance, capacitance and diode measurements.
Measuring DC ( ) Voltage
Measuring AC ( ) Voltage
15
SEL Button
16
Verifying Continuity ( )
Verify the circuit is de-energized.
The meter will sense the level of resistance and beep if the resistance is less than
10 Ωs to confirm that continuity is present.
The red LED will illuminate and the resistance value will be displayed.
The default setting is Continuity
• Verify the circuit is de-energized to obtain accurate measurements.
To exit Continuity and enter the Resistance mode, press the SEL button.
Measuring Resistance (Ohms / Ω)
17
Verifying Continuity ( )
• Verify the circuit is de-energized to obtain accurate measurements.
To exit Continuity and enter the Resistance mode, press the SEL button.
Measuring Resistance (Ohms / Ω)
Measuring Capacitance ( )
Measuring Diodes ( )
To exit Resistance and enter the Capacitance mode, press the SEL button.
To exit Capacitance and enter the Diode Test mode, press the SEL button.
Measuring Frequency ( Hz )
Measuring Temperature ( °F °C )
Hz using Voltage
For % Duty Cycle Press the
SEL button
Press SEL to toggle
between F and C
18
Measuring Frequency ( Hz )
Measuring Temperature ( °F °C )
Measuring Micro Amps (μA)
Measuring Milli Amps (mA)
Measuring Amps (A)
19
CAUTION
Meter damage. Measure current in
Micro Amps for a maximum duration
of 10 minutes. Failure to comply can
result in meter damage.
CAUTION
Meter damage. Measure current in
Milliamps for a maximum duration of
1 minute. Failure to comply can result
in meter damage..
CAUTION
Meter damage. Measure current in
Amps for a maximum duration of 10
seconds. Failure to comply can result
in meter damage.
Functions Operation Table
20
Button Response Default Function Operation
HOLD
HOLD: All Functions Normal Measurement
Short Press: Circularly enter or exit the data hold mode, LCD will display “ “ after enter
HOLD function.
SEL /
ACA / DCA DCA
1) Short press: Circularly select the relative measurement function.
2) When in HOLD, RANGE, MAX/MIN, REL function, Short press the SEL key to quit current
function and circularly change relative Select function.
ACV/DCV ACV
Hz/% Hz
/
Ω / /
Continuity
°F/°C °F
RANGE ACV, DCV, ACA, DCA, Ω
Auto range (LCD display
“AUTO”)
1) Short press the RANGE key to enter manual range mode (LCD will not display “AUTO”), and
the current range, press again to enter cyclical selection range. Long press: exits manual range
and enters Auto range. Switching measurement setting or restarting the unit returns the unit
to the default setting.
2) In HOLD, MAX/MIN, REL function, Short press the RANGE key to exit the current function
and enter the relative range function.
: ACV, DCV, ACA, DCA,
CAP, Ω, °F/°C
Normal Measurement
1) Short press: Circularly enter or exit REL function (when entering REL function, LCD will
display “ ” symbol )
2) Entering the REL function, exits auto range and enters into the current range.
3) Entering the REL measurement mode does not change the actual measurable range of the
current range
4) When HOLD or MAX/MIN is in use, REL function is invalid.
Non-Contact Voltage Indication
Displays “EF” – Electromagnetic Field
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90

Ideal 61-337 Manuel utilisateur

Catégorie
Mesure, test
Taper
Manuel utilisateur

dans d''autres langues