Charging the Sensor
Connect the Go Direct Colorimeter to the included USB Charging Cable and
any USB device for two hours.
You can also charge up to eight Go Direct Go Direct Colorimeters using our
GoDirect Charge Station, sold separately (order code: GDX-CRG). An LED on
each Go Direct Colorimeter indicates charging status.
Charging
Orange LED next to the battery icon is solid
while the sensor is charging.
Fully charged
Green LED next to the battery icon is solid
when the sensor is fully charged.
Providing Power
Turning on the sensor
Press button once. Red LED indicator flashes
when unit is on.
Putting the sensor in sleep
mode
Press and hold button for more than three
seconds to put into sleep mode. Red LED
indicator stops flashing when sleeping.
Connecting the Sensor
See the following link for up-to-date connection information:
www.vernier.com/start/gdx-col
Connecting via Bluetooth
Ready to connect Red LED next to the Bluetooth icon flashes
when sensor is awake and ready to connect.
Connected Green LED next to the Bluetooth icon flashes
when sensor is connected via Bluetooth.
Go Direct
®
Colorimeter
(Order Code GDX-COL)
Go Direct Colorimeter is used to determine the
concentration of a solution by analyzing its color
intensity. The cuvette slot is designed to
accommodate most cuvettes with a 10 millimeter path
length. The Colorimeter measures the amount of light
transmitted through a sample at a user-selectable
wavelength. You may choose from four wavelengths: 430 nm, 470 nm, 565 nm,
and 635 nm.
Note: Vernier products are designed for educational use. Our products are not
designed nor are they recommended for any industrial, medical, or commercial
process such as life support, patient diagnosis, control of a manufacturing
process, or industrial testing of any kind.
What's Included
l Go Direct Colorimeter
l 15 polystyrene cuvettes and 15 plastic cuvette lids
l Micro USBcable
Compatible Software
See www.vernier.com/manuals/gdx-col for a list of software compatible with Go
Direct Colorimeter
Getting Started
Please see the following link for platform-specific connection information:
www.vernier.com/start/gdx-col
Bluetooth Connection USB Connection
1. Install Graphical Analysis 4 on
your computer, Chromebook™, or
mobile device. See
www.vernier.com/ga4 for software
availability.
2. Charge your sensor for at least
2hours before first use.
3. Turn on your sensor by pressing
the power button once. The
Bluetooth
®
LED will blink red.
4. Launch Graphical Analysis 4.
5. Click or tap Sensor Data
Collection.
6. Click or tap your Go Direct sensor
from the list of Discovered
Wireless Devices. Your sensor's ID
1. Install Graphical Analysis 4 on your
computer or Chromebook. If using
LabQuest 2, make sure LabQuest
App is up to date. See
www.vernier.com/ga4 for software
availability or
www.vernier.com/downloads to
update LabQuest App.
2. Connect the sensor to the USB port.
3. Launch Graphical Analysis 4 or turn
on LabQuest 2. You are now ready
to collect data.
2
Care and Maintenance
Battery Information
Go Direct Colorimeter contains a small lithium-ion battery. The system is
designed to consume very little power and not put heavy demands on the
battery. Although the battery is warranted for one year, the expected battery life
should be several years. Replacement batteries are available from Vernier (order
code: GDX-BAT-300).
Storage and Maintenance
To store Go Direct Colorimeter for extended periods of time, put the device in
sleep mode by holding the button down for at least three seconds. The red LED
will stop flashing to show that the unit is in sleep mode. Over several months,
the battery will discharge but will not be damaged. After such storage, charge
the device for a few hours, and the unit will be ready to go.
Exposing the battery to temperatures over 35°C (95°F) will reduce its lifespan. If
possible, store the device in an area that is not exposed to temperature extremes.
Water Resistance
Go Direct Colorimeter is not water resistant and should never be immersed in
water.
If water gets into the device, immediately power the unit down (press and hold
the power button for more than three seconds). Disconnect the sensor and
charging cable, and remove the battery. Allow the device to dry thoroughly
before attempting to use the device again. Do not attempt to dry using an
external heat source.
How the Sensor Works
Light from an LED light source passes through a cuvette containing a solution
sample. Some of the incoming light is absorbed by the solution. As a result,
light of a lower intensity strikes a photodiode.
Troubleshooting
Here are some tips for best data-collection practices:
l Allow the Colorimeter to warm up for about 5 minutes before calibrating.
l Fill a cuvette two-thirds to three-fourths full with liquid, including the
calibration blank, so that the light travels through the liquid reliably.
Connecting via USB
Connected and charging Orange LED next to the battery icon is solid
when the sensor is connected to Graphical Ana-
lysis via USB and the unit is charging. LED next
to Bluetooth icon is off.
Connected, fully charged Green LED next to the battery icon is solid
when the sensor is connected to Graphical Ana-
lysis via USB and fully charged. LED next to
Bluetooth icon is off.
Charging via USB,
connected via Bluetooth
Orange LED next to the battery icon is solid
when the sensor is charging. Green LED next to
the Bluetooth icon flashes.
Using the Product
Connect the sensor following the steps in the Getting Started section of this user
manual. After connecting the sensor, the following steps must be completed
prior to starting data collection.
1. Press the < or > button on the Colorimeter to select the correct wavelength
for your experiment (430 nm, 470 nm, 565 nm, or 635 nm).
2. Allow the Colorimeter to warm up for about five minutes before calibrating.
3. Calibrate the Colorimeter.
a. Slide the lid of the Colorimeter open to reveal the cuvette slot.
b. Insert a cuvette, filled with distilled water or other solvent used to
prepare your solutions, for your calibration blank (100% transmittance or
0 absorbance).
Important: Line up one of the clear sides of the cuvette with the arrow
at the right side of the cuvette slot. Slide the Colorimeter lid closed.
c. Press the CAL button on the Colorimeter to begin the calibration
process. Release the CAL button when the red LED begins to flash.
d. When the red LED stops flashing, the calibration is complete. The
absorbance reading should be very close to 0.000 (100%T).
e. Remove the blank cuvette from the Colorimeter.
4. Continue with data collection.
Calibration
Calibration is required for use. See the Using the Product section for
instructions.
Specifications
Colorimeter range 0 to 3 (absorbance) (0 to 100% T)
Useful range 0.05 to 1.0 absorbance (90% to 10% T)
Wavelengths 430 nm, 470 nm, 565 nm, 635 nm
3
you help prevent potential negative consequences on human health or on the
environment. The recycling of materials will help to conserve natural resources.
For more detailed information about recycling this product, contact your local
city office or your disposal service.
Battery recycling information is available at www.call2recycle.org
Do not puncture or expose the battery to excessive heat or flame.
The symbol, shown here, indicates that this product must not be disposed of
in a standard waste container.
Federal Communication Commission Interference Statement
This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC rules.
These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment
generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause
harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If
this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and
on, the user is encouraged to try to correct the interference by one or more of the following measures:
Reorient or relocate the receiving antenna.
Increase the separation between the equipment and receiver.
Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
Consult the dealer or an experienced radio/TV technician for help.
FCC Caution
This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions:
(1) this device may not cause harmful interference and
(2) this device must accept any interference received, including interference that may cause undesired operation
RF Exposure Warning
The equipment complies with RF exposure limits set forth for an uncontrolled environment. The antenna(s) used for this transmitter must
not be co-located or operating in conjunction with any other antenna or transmitter. You are cautioned that changes or modifications not
expressly approved by the party responsible for compliance could void your authority to operate the equipment.
IC Statement
This device complies with Industry Canada license-exempt RSS standard(s). Operation is subject to the following two conditions:
(1) this device may not cause interference, and
(2) this device must accept any interference, including interference that may cause undesired operation of the device.
Industry Canada - Class B This digital apparatus does not exceed the Class B limits for radio noise emissions from digital apparatus
as set out in the interference-causing equipment standard entitled Digital Apparatus,” ICES-003 of Industry Canada. Operation is subject
to the following two conditions: (1) this device may not cause interference, and
(2) this device must accept any interference, including interference that may cause undesired operation of the device.
To reduce potential radio interference to other users, the antenna type and its gain should be so chosen that the equivalent isotropically
radiated power (e.i.r.p.) is not more than that permitted for successful communication.
RF exposure warning: The equipment complies with RF exposure limits set forth for an uncontrolled environment. The antenna(s) used
for this transmitter must not be co-located or operating in conjunction with any other antenna or transmitter.
Le présent appareil est conforme aux CNR d’Industrie Canada applicables aux appareils radio exempts de licence. L’exploitation est
autorisée aux deux conditions suivantes :
(1) lappareil ne doit pas produire de brouillage, et
(2) lappareil doit accepter tout interférence radilectrique, même si cela résulte à un brouillage susceptible d’en compromettre le
fonctionnement.
Cet appareil numérique respecte les limites de bruits radioélectriques applicables aux appareils numériques de Classe B prescrites dans
la norme sur le matériel interférant-brouilleur: Appareils Numériques,” NMB-003 édictée par industrie Canada. L’utilisation est soumise
aux deux conditions suivantes:
(1) cet appareil ne peut causer d’interférences, et
(2) cet appareil doit accepter toutes interférences, y comprises celles susceptibles de provoquer un disfonctionnement du dispositif.
Afin de réduire les interférences radio potentielles pour les autres utilisateurs, le type d’antenne et son gain doivent être choisie de telle
façon que léquivalent de puissance isotrope émis (e.i.r.p) n’est pas plus grand que celui permis pour une communication établie.
l After filling a cuvette with liquid, seal the cuvette with a cap to prevent
spills.
l Make sure to place a cuvette in the Colorimeter so the path of the light
source travels through the clear sides of the cuvette. An arrow to the right of
the cuvette slot shows the light path.
l For best results, use one cuvette to make all your measurements for a given
experiment.
l If you calibrate a Colorimeter and then change the wavelength, calibrate the
Colorimeter again to ensure the proper identification of the new wavelength.
For best results, a sample's absorbance or transmittance values should fall within
these ranges:
percent transmittance:10–90%
absorbance: 0.05–1.0
Beer’s law experiment results begin to lose their linearity at absorbance values
above 1.0 (percent transmittance values less than 10%). If you have a solution
that transmits such a low level of light, consider diluting the solution so that it
falls within this range.
Repair Information
If you have followed the troubleshooting steps and are still having trouble with
your Go Direct Colorimeter, contact Vernier Technical Support at
support@vernier.com or call 888-837-6437. Support specialists will work with
you to determine if the unit needs to be sent in for repair. At that time, a Return
Merchandise Authorization (RMA) number will be issued and instructions will
be communicated on how to return the unit for repair.
Accessories/Replacements
Item Order Code
Micro USB Cable
CB-USB-MICRO
USB-C to Micro USB cable
CB-USB-C-MICRO
Go Direct 300 mAh Replacement Battery
GDX-BAT-300
Plastic Cuvettes (Visible Range)
CUV
Warranty
Vernier warrants this product to be free from defects in materials and
workmanship for a period of five years from the date of shipment to the
customer. This warranty does not cover damage to the product caused by abuse
or improper use. This warranty covers educational institutions only.
Disposal
When disposing of this electronic product, do not treat it as household waste. Its
disposal is subject to regulations that vary by country and region. This item
should be given to an applicable collection point for the recycling of electrical
and electronic equipment. By ensuring that this product is disposed of correctly,
4
Vernier Software & Technology
13979 SW Millikan Way Beaverton, OR 97005-2886
Toll Free (888) 837-6437 (503) 277-2299 Fax (503) 277-2440
info@vernier.com www.vernier.com
Rev. 04/02/18
Go Direct, Graphical Analysis, LabQuest, and other marks shown are our trademarks or registered trademarks in
the United States. All other marks not owned by us that appear herein are the property of their respective
owners, who may or may not be affiliated with, connected to, or sponsored by us.
The Bluetooth
®
word mark and logos are registered trademarks owned by the Bluetooth SIG, Inc. and any use of
such marks by Vernier Software & Technology is under license. Other trademarks and trade names are those of
their respective owners.
Avertissement d’exposition RF: L’équipement est conforme aux limites d’exposition aux RF établies pour un environnement non
supervisé. L’antenne (s) utilisée pour ce transmetteur ne doit pas être jumelés ou fonctionner en conjonction avec toute autre antenne ou
transmetteur.
Note: This product is a sensitive measurement device. For best results, use the cables that were provided. Keep the device away from
electromagnetic noise sources, such as microwaves, monitors, electric motors, and appliances.
IC Statement
This device complies with Industry Canada license-exempt RSS standard(s). Operation is subject to the following two conditions:
(1) this device may not cause interference, and
(2) this device must accept any interference, including interference that may cause undesired operation of the device.
Industry Canada - Class B This digital apparatus does not exceed the Class B limits for radio noise emissions from digital apparatus
as set out in the interference-causing equipment standard entitled Digital Apparatus,” ICES-003 of Industry Canada. Operation is subject
to the following two conditions: (1) this device may not cause interference, and
(2) this device must accept any interference, including interference that may cause undesired operation of the device.
To reduce potential radio interference to other users, the antenna type and its gain should be so chosen that the equivalent isotropically
radiated power (e.i.r.p.) is not more than that permitted for successful communication.
RF exposure warning: The equipment complies with RF exposure limits set forth for an uncontrolled environment. The antenna(s) used
for this transmitter must not be co-located or operating in conjunction with any other antenna or transmitter.
Le présent appareil est conforme aux CNR d’Industrie Canada applicables aux appareils radio exempts de licence. L’exploitation est
autorisée aux deux conditions suivantes :
(1) lappareil ne doit pas produire de brouillage, et
(2) lappareil doit accepter tout interférence radilectrique, même si cela résulte à un brouillage susceptible d’en compromettre le
fonctionnement.
Cet appareil numérique respecte les limites de bruits radioélectriques applicables aux appareils numériques de Classe B prescrites dans
la norme sur le matériel interférant-brouilleur: Appareils Numériques,” NMB-003 édictée par industrie Canada. L’utilisation est soumise
aux deux conditions suivantes:
(1) cet appareil ne peut causer d’interférences, et
(2) cet appareil doit accepter toutes interférences, y comprises celles susceptibles de provoquer un disfonctionnement du dispositif.
Afin de réduire les interférences radio potentielles pour les autres utilisateurs, le type d’antenne et son gain doivent être choisie de telle
façon que léquivalent de puissance isotrope émis (e.i.r.p) n’est pas plus grand que celui permis pour une communication établie.
Avertissement d’exposition RF: L’équipement est conforme aux limites d’exposition aux RF établies pour un environnement non
supervisé. L’antenne (s) utilisée pour ce transmetteur ne doit pas être jumelés ou fonctionner en conjonction avec toute autre antenne ou
transmetteur.
Note: This product is a sensitive measurement device. For best results, use the cables that were provided. Keep the device away from
electromagnetic noise sources, such as microwaves, monitors, electric motors, and appliances.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4

Vernier GDX-COL Manuel utilisateur

Taper
Manuel utilisateur

dans d''autres langues