Patient Slings
Patient Slings 2 Part No 1023891-K
WARNING
DO NOT USE THIS PRODUCT OR ANY AVAILABLE OPTIONAL
EQUIPMENT WITHOUT FIRST COMPLETELY READING AND
UNDERSTANDING THESE INSTRUCTIONS AND ANY
ADDITIONAL INSTRUCTIONAL MATERIAL SUCH AS OWNER’S
MANUALS, SERVICE MANUALS OR INSTRUCTION SHEETS
SUPPLIED WITH THIS PRODUCT OR OPTIONAL EQUIPMENT. IF
YOU ARE UNABLE TO UNDERSTAND THE WARNINGS,
CAUTIONS OR INSTRUCTIONS, CONTACT A HEALTHCARE
PROFESSIONAL, DEALER OR TECHNICAL PERSONNEL BEFORE
ATTEMPTING TO USE THIS EQUIPMENT - OTHERWISE, INJURY
OR DAMAGE MAY OCCUR.
ACCESSORIES WARNING
Invacare products are specifically designed and manufactured for use in conjunction
with Invacare accessories. Accessories designed by other manufacturers have not
been tested by Invacare and are not recommended for use with Invacare products.
NOTE: Updated versions of this manual are available on www.invacare.com.
TABLE OF CONTENTS
Part No 1023891-K 3 Patient Slings
TABLE OF CONTENTS
SPECIAL NOTES ................................................................................ 4
Service Life...................................................................................................................................................4
SECTION 1—GENERAL GUIDELINES ................................................... 5
Using the Sling.............................................................................................................................................5
Care...............................................................................................................................................................6
SECTION 2—SLING STYLES ................................................................ 7
Standard Series Style Slings and Kits......................................................................................................7
One-Piece Style Slings ...............................................................................................................................7
SECTION 3—USING THE STANDARD STYLE SLING ............................. 8
Introduction.................................................................................................................................................8
Care...............................................................................................................................................................8
Removing/Installing the Metal Support Bars in the Standard Style Sling........................................8
Positioning the Standard Style Sling........................................................................................................9
Attaching the Chains or Straps to the Standard Style Sling............................................................10
Attaching the Standard Style Sling to the Patient Lift ......................................................................11
SECTION 4—USING FULL BODY, COMFORT AND HEAVY-DUTY
ONE-PIECE STYLE SLINGS ............................................................... 12
Positioning the Sling .................................................................................................................................12
Attaching the One-Piece Style Sling to the Patient Lift ...................................................................14
SECTION 5—USING DIVIDED LEG, UNIVERSAL HIGH, EASY-FIT,
TOILETING, DRESS TOILETING HIGH AND CRADLE ONE-PIECE
STYLE SLINGS ................................................................................. 15
Introduction...............................................................................................................................................15
Positioning the Sling .................................................................................................................................15
Attaching the One-Piece Style Sling to the Patient Lift ...................................................................17
Slings Using the Cradle Attachment................................................................................................17
SECTION 6—USING STANDING, STAND ASSIST, TRANSPORT AND
TRANSFER ONE-PIECE STYLE SLINGS .............................................. 18
Positioning the Sling Around the Patient ............................................................................................18
Standing Sling.........................................................................................................................................18
Transport Sling or Transfer Sling .....................................................................................................19
Attaching the Sling to the Stand Up Lift..............................................................................................20
Standing Sling.........................................................................................................................................20
Transport Sling or Transfer Sling .....................................................................................................22
LIMITED WARRANTY ..................................................................... 24
SPECIAL NOTES
Patient Slings 4 Part No 1023891-K
SPECIAL NOTES
Signal words are used in this manual and apply to hazards or unsafe practices which
could result in personal injury or property damage. Refer to the table below for
definitions of the signal words.
NOTICE
THE INFORMATION CONTAINED IN THIS DOCUMENT IS SUBJECT TO
CHANGE WITHOUT NOTICE.
Service Life
The expected service life is thirteen (13) months for this product, provided the product is
used in accordance with the intended use as set out in this document and all maintenance,
service requirements, and laundering instructions are followed. While the expected
service life might be exceeded if the product is carefully used and properly maintained,
it can also be reduced by excessive or incorrect usage.
SIGNAL WORD MEANING
DANGER
Danger indicates an imminently hazardous situation which, if not
avoided, will result in death or serious injury.
WARNING
Warning indicates a potentially hazardous situation which, if not
avoided, could result in death or serious injury.
CAUTION
Caution indicates a potentially hazardous situation which, if not
avoided, may result in property damage or minor injury or both.
SECTION 1—GENERAL GUIDELINES
Part No 1023891-K 5 Patient Slings
SECTION 1—GENERAL GUIDELINES
WARNING
SECTION 1 - GENERAL GUIDELINES contains important information for the safe
operation and use of this product.
WARNING
DO NOT move a person suspended in a sling any distance. The Invacare patient lift
or the sling are NOT transport devices. They are intended to transfer an individual
from one resting surface to another (such as a bed to a wheelchair). Otherwise,
injury or damage may occur.
Check all parts for shipping damage before using. In case of damage, DO NOT use the
equipment. Contact the dealer for further instructions.
DO NOT attempt any transfer without approval of the patient’s physician, nurse or
medical assistant. Thoroughly read the instructions in this owners manual, observe a
trained team of experts perform the lifting procedures and then perform the entire lift
procedure several times with proper supervision and a capable individual acting as a
patient.
Although Invacare recommends that two assistants be used for all lifting preparation,
transferring from and transferring to procedures, our equipment will permit proper
operation by one assistant. The use of one assistant is based on the evaluation of the health
care professional for each individual case. Bed rails may be used to minimize patient
movement during sling positioning.
ALWAYS keep hands and fingers clear of moving parts to avoid injury.
DO NOT exceed maximum weight limitation of the patient lift. These weights will vary
from 350 to 600 lbs. depending on the type of patient lift purchased. For maximum weight
limitation, refer to the Typical Product Parameters section in the patient lift manual or
contact customer service.
Invacare slings are made specifically for use with Invacare lifts. For the safety of the
patient, DO NOT intermix slings and lifts of different manufacturers.
Using the Sling
Use an Invacare approved sling that is recommended by the individual’s doctor, nurse or
medical assistant for the comfort and safety of the individual being lifted.
Standing Slings - Individuals that use the standing sling MUST be able to support the
majority of their own weight, otherwise injury can occur. Before lifting the patient, make
sure the bottom edge of the standing sling is positioned on the lower back of the patient
and the patient's arms are outside the standing sling. The belt MUST be snug, but
comfortable on the patient, otherwise the patient can slide out of the sling during transfer,
possibly causing injury.
SECTION 1—GENERAL GUIDELINES
Patient Slings 6 Part No 1023891-K
Transport/Transfer Slings - Before lifting the patient, make sure the bottom edge of the
transfer sling is at the base of the spine and the patient's arms are outside the transfer
sling. DO NOT raise the patient to a full standing position while using the transfer sling,
otherwise injury may occur.
The Jasmine Patient Lift can be used with the standard swivel bar or an optional cradle
attachment. The tilting cradle attachment enables a single caregiver to more easily
position a patient in an upright and seated position.
DO NOT use any kind of material (such as a plastic back incontinence pad or seating
cushion) between the patient and sling material that may cause the patient to slide out of
the sling during transferring.
After each laundering (in accordance with instructions on the sling), inspect sling(s) for
wear, tears, and loose stitching.
Bleached, torn, cut, frayed, or broken slings are unsafe and could result in injury. Discard
immediately.
DO NOT alter slings.
When connecting slings equipped with color coded straps to the patient lift, the shortest
of the straps MUST be at the back of the patient for support. Using the long section will
leave little or no support for the patient’s back. The loops of the sling are color coded and
can be used to place the patient in various positions. The colors make it easy to connect
both sides of the sling equally. Make sure that there is sufficient head support when lifting
a patient.
Care
NOTE: Laundering should always be done with dark colors. The sling should be washed regularly
in a water temperature not to exceed 180°F (82°C). DO NOT bleach. Air dry or dry at low
temperature. Inspect with each use. Refer to tagged washing instructions on the sling.
WARNING
After each laundering (in accordance with instructions on the sling), inspect sling(s)
for wear, tears, and loose stitching. Bleached, torn, cut, frayed, or broken slings are
unsafe and could result in injury. Discard immediately.
SECTION 2—SLING STYLES
Part No 1023891-K 7 Patient Slings
SECTION 2—SLING STYLES
Standard Series Style Slings and Kits
Invacare's standard series slings are designed specifically for use with the 9805/9805P
lifts. The standard style sling attaches to the lift with either adjustable straps or chains
(interchangeable).
One-Piece Style Slings
Invacare’s one-piece sling attaches to the patient lift with color coded sling loops that
provide an easy way to position the patient. There are no additional strap or chain kits
required for use with the one-piece sling styles.
STANDARD SERIES SLING DESCRIPTION CAN ONLY BE USED WITH...
Polyester Sling 9805/9805P Lifts
Polyester Sling With Commode Opening
Polyester Mesh Sling
Polyester Mesh Sling With Commode Opening
Strap Kit
Chain Kit
ONE-PIECE FABRIC SLING DESCRIPTION CAN ONLY BE USED WITH...
Full Body Sling with or without Commode Opening Jasmine, Reliant 450, Reliant 600 Lifts, I-Transia and
9805 Lifts (on the large hooks of the Swivel Bar)
Comfort Sling with or without Commode Opening
Divided Leg Sling
Universal High Sling
Easy Fit Sling
Toileting Sling
Dress High Toileting Sling
Heavy-Duty Sling with or without Commode Opening Reliant 450, Reliant 600, I-Transia and Jasmine Lifts
Cradle Sling Jasmine Lift equipped with optional Cradle Attachment
Standing Sling Reliant 350, Reliant 440, Get-U-Up™ and Roze Lifts
Stand Assist Sling
Transport Sling
Transfer Sling
SECTION 3—USING THE STANDARD STYLE SLING
Patient Slings 8 Part No 1023891-K
SECTION 3—USING THE STANDARD
STYLE SLING
Introduction
The hardware used with the standard sling includes four metal support bars. The support
bars insert into the pockets at the edge of the fabric. The pocket that holds the metal bar has
an opening in the center. The pocket is sewn closed on one end and has an open-end with a
strap on the other.
Care
Remove the four metal support bars before
laundering the standard style sling. The
metal support bars MUST be replaced
before the sling can be used for patient
lifting.
Removing/Installing the
Metal Support Bars in the
Standard Style Sling
NOTE: To remove the metal support bar,
reverse this procedure.
NOTE: For this procedure, refer to
FIGURE 3.1.
1. To insert the metal support bar, pull the
strap and insert the metal bar into the
opening in the sling pocket. Refer to
Detail “A”.
2. Push the metal support bar through
until the mounting hole clears the
center opening of the sling pocket.
Refer to Details “B” and “C”.
3. Once the opposite end of the bar is
inside the sewn pocket, push the strap
over the metal support bar. Refer to
Detail “D”.
NOTE: The hole in the metal bar MUST be in
the center of the pocket opening.
FIGURE 3.1 Removing/Installing the Metal
Support Bars in the Standard Style Sling
Metal Support Bar with Opening
for Chain or Strap
Standard Style
Sling
DETAIL “D”
Strap Over Metal
Support Bar
DETAIL “A”
Opening
DETAIL “B”
DETAIL “C”
SECTION 3—USING THE STANDARD STYLE SLING
Part No 1023891-K 9 Patient Slings
Positioning the Standard Style Sling
NOTE: For this procedure, refer to FIGURE 3.2.
NOTE: Use the following method to easily
move the patient and avoid strain to yourself.
1. With an attendant on each side of the
bed and up against the mattress, the
attendant on the right-hand side of the
bed will elevate the left knee and
position his/her left-hand on the
elevated knee and his/her right-hand
under the patient’s left shoulder, slowly
push on the knee and assist with a
slight lift of the shoulder and the
patient will easily roll onto their side.
Refer to Detail “A”.
2. Put the standard style sling onto the
mattress.
NOTE: The upper edge of the back section
should be positioned slightly above the armpit
level and the lower edge of the seat section
positioned a few inches above the back of the
patient’s knees Refer to Detail “A”.
3. Roll the patient onto his/her back.
4. Pull the sling across the mattress until
smooth.
NOTE: The patient should be approximately
centered on the sling section. Refer to Detail “B”.
NOTE: The sling is now ready to be attached to
the chain or strap. Refer to Attaching the Chains
or Straps to the Standard Style Sling on page 10.
FIGURE 3.2 Positioning the Standard Style
Sling
DETAIL “B”
DETAIL “A”
SECTION 3—USING THE STANDARD STYLE SLING
Patient Slings 10 Part No 1023891-K
Attaching the Chains or Straps to the Standard Style Sling
NOTE: For this procedure, refer to FIGURE 3.3.
NOTE: Each sling will work with either a pair of chain assemblies or a pair of strap assemblies.
CAUTION
When using either the chains or straps to connect the sling to the patient lift, the short-
est of the two sections of the chains or straps MUST be attached to the back section of
the sling. Using the long section will leave little or no support for the patient’s back.
1. With the patient laying on his/her back and positioned properly in the sling, lay the chains
or straps on the bed next to the sling. Refer to Detail “A”.
2. The short section of chain or strap will be placed parallel to the patient’s back. This will
position the longer section of chain or strap from the waist to the mid section of the thigh.
3. The “S” hooks on the ends of the chains or straps are inserted into the metal support bar
of the sling.
4. The “S” hook MUST be inserted through the hole in the support bar with the open end
facing away from the patient. Refer to Detail “C”.
5. Position the lift for use. Refer to Attaching the Standard Style Sling to the Patient Lift on
page 11.
FIGURE 3.3 Attaching the Chains or Straps to the Standard Style Sling
MODEL NUMBERS DESCRIPTIONS
9070 Strap Kit
9071 Chain Kit
S-Hook Open End Away
from Patient
S-Hook Open End
Away from Patient
O-Ring
Short Section of Chain
S-Hook Open End Away
from Patient
Short Section
Long Section
Long Section of Chain
D-Ring
Adjustable
DETAIL “C”
DETAIL “B”
DETAIL “A”
SECTION 3—USING THE STANDARD STYLE SLING
Part No 1023891-K 11 Patient Slings
Attaching the Standard Style Sling to the Patient Lift
NOTE: For this procedure, refer to FIGURE 3.4.
1. The D-ring (strap) or O-ring (chain) is the suggested location to attach the swivel bar,
however, by changing to different links or adjusting the strap, you can reposition the
patient into a reclining or more upright position.
WARNING
When changing to different lengths, DO NOT disconnect the O-rings of the chains
from the swivel bar. Position BOTH the O-rings and additional chain link onto the
hooks of the swivel bar.
NOTE: The straps can be simply adjusted to achieve the same effect.
2. With the sling positioned under the patient and the support bars in place, move the
lift into position to attach the sling hardware to the swivel bar.
3. Refer to Lifting the Patient in the Patient Lift Owners Manual.
FIGURE 3.4 Attaching the Standard Style Sling to the Patient Lift
Adjustable
D-Ring
S-Hook
Open-End
Away From
Patient
O-Ring
Long
Section
Short
Section
SECTION 4—USING FULL BODY, COMFORT AND HEAVY-DUTY ONE-PIECE STYLE
SLINGS
Patient Slings 12 Part No 1023891-K
SECTION 4—USING FULL BODY,
COMFORT AND HEAVY-DUTY
ONE-PIECE STYLE SLINGS
Positioning the Sling
NOTE: For this procedure, refer to FIGURE 4.1 and FIGURE 4.2 on page 13.
NOTE: Use the following method to easily move the patient and avoid strain to yourself.
1. Position the patient in the center of the bed and laying flat on his/her back.
2. Fold the sling in half (length-wise) and place the sling beside the patient. Refer to
Detail “A”.
NOTE: The closed-end or commode opening of the sling with positioning handle should be facing
the patient when folded.
NOTE: The top edge of the sling fabric should be slightly above the patient's head. The bottom edge
of the sling fabric should then be a few inches above the back of the patient’s knees.
3. With one assistant holding the patient, the second assistant pushes the folded sling
under the patient without rolling him/her over. Refer to Detail “B”.
FIGURE 4.1 Positioning the Sling
4. With an assistant on each side of the bed and up against the mattress, the assistant on
the left-hand side of the bed will position his/her right-hand on the elevated knee and
his/her left-hand under the patient’s right shoulder, slowly push on the knee and assist
with a slight lift of the shoulder. The patient will easily roll onto his/her side. Refer to
Detail “A”.
Positioning Handle
Top Edge of Sling
Commode Opening
Sling Loop
Sling Loop
Bottom Edge of Sling
DETAIL “A”
DETAIL “B”
SECTION 4—USING FULL BODY, COMFORT AND HEAVY-DUTY ONE-PIECE STYLE
SLINGS
Part No 1023891-K 13 Patient Slings
5. With the patient on their side, push the fabrics of the seat and back gently under them.
Refer to Detail “B”.
FIGURE 4.2 Rolling Patient Onto His/Her Side
NOTE: The patient’s head should be positioned in the headrest just below the top edge for
maximum comfort and the lower edge of the seat section positioned a few inches above the back of
the patient’s knees.
6. Roll the patient on to his/her back.
NOTE: Assistants will reverse roles.
7. After the patient has been positioned once again on his/her back, you now need to roll
the patient to their right-side (facing the assistant on the left).
8. With an assistant on each side of the bed and up against the mattress, the assistant on
the right-hand side of the bed will elevate the left knee and position his/her left-hand
on the elevated knee and his/her right-hand under the patient’s left shoulder, slowly
push on the knee and assist with a slight lift of the shoulder and the patient will easily
roll onto their side. Refer to Detail “A”.
9. Pull the fabrics of the seat and back across the mattress until they are smooth.
10. Roll the patient onto their back and they should be approximately centered on the
sling.
11. Position the lift for use. Refer to Attaching the One-Piece Style Sling to the Patient Lift.
DETAIL “B”
DETAIL “A”
SECTION 4—USING FULL BODY, COMFORT AND HEAVY-DUTY ONE-PIECE STYLE
SLINGS
Patient Slings 14 Part No 1023891-K
Attaching the One-Piece Style Sling to the Patient Lift
NOTE: For this procedure, refer to FIGURE 4.3.
1. Place the straps of the sling over hooks of the swivel bar. Match the corresponding colors
on each side of the sling for an even lift of the patient.
CAUTION
When connecting sling to the patient lift, the shortest of the straps MUST be at the
back of patient for support. Using long section will leave little or no support for patient's
back. The loops of the sling are color coded and can be used to place patient in various
positions. The colors make it easy to connect both sides of the sling equally. Make sure
that there is sufficient head support when lifting a patient.
2. Refer to the Lifting the Patient in the Patient Lift Owners Manual.
NOTE: Invacare Lift Swivel Bars have three hookup points per side. The middle hookup is ONLY used
for slings that have three sets of straps per side or slings that use chains or adjustable straps.
FIGURE 4.3 Attaching the One-Piece Style Sling to the Patient Lift
Color
Coded
Straps
Patient in Reclined Position
(head and leg strap lengths equal)
Patient in Seated Position
(short strap length both sides of
head, long strap length at legs)
SECTION 5—USING DIVIDED LEG, UNIVERSAL HIGH, EASY-FIT, TOILETING, DRESS TOI-
LETING HIGH AND CRADLE ONE-PIECE STYLE SLINGS
Part No 1023891-K 15 Patient Slings
SECTION 5—USING DIVIDED LEG,
UNIVERSAL HIGH, EASY-FIT,
TOILETING, DRESS TOILETING
HIGH AND CRADLE ONE-PIECE
STYLE SLINGS
Introduction
Divided Leg, Universal High, Easy Fit, Toileting, Dress Toileting High and Cradle Sling
styles may be used for patients on which sling removal is more difficult. Support ranges
from upper back to beneath the thighs. These slings can be positioned with the patient
seated or lying flat.
FIGURE 5.1 Sling Styles
Positioning the Sling
NOTE: For this procedure, refer to
FIGURE 5.2 and FIGURE 5.3 on page 16.
NOTE: Use the following method to easily
move the patient and avoid strain to yourself.
WARNING
If the patient is in a wheelchair, secure
the wheel locks in place to prevent the
wheelchair from moving forward or
backwards.
1. Lock both rear wheel locks on the
wheelchair.
FIGURE 5.2 Wheel Lock
Toileting
Easy-Fit
Divided Leg
Cradle
Universal High
Dress Toileting High
Wheel Lock
SECTION 5—USING DIVIDED LEG, UNIVERSAL HIGH, EASY-FIT, TOILETING, DRESS TOI-
LETING HIGH AND CRADLE ONE-PIECE STYLE SLINGS
Patient Slings 16 Part No 1023891-K
2. With the patient sitting in a wheelchair
one assistant in front of the wheelchair
and the other assistant in the back of
the wheelchair, lean the patient
forward. Refer to Detail “A”.
NOTE: The front assistant will be supporting
the weight of the patient.
3. Place the sling behind the patient (with
"grab handle" on the outside) and bring
the flaps out alongside the patient’s
legs. Refer to Detail “B”.
NOTE: Toileting slings have a waist buckle
which helps hold the sling in place during
positioning.
NOTE: For the Divided Leg, Universal High,
Easy-Fit, Dress Toileting High and Cradle
sling: the back of the sling should be parallel to
the patient’s upper arms and be positioned
between the top of the patient’s head and the
base of the spine.
NOTE: For the Toileting Sling: the back of the
sling should be parallel to the patient’s upper
arms and be positioned between the top of the
patient’s midback and the base of the spine.
4. With the back of the sling positioned
properly, push the bottom of the back
section to the seat.
5. Lean the patient back into the
wheelchair with the assistant in the rear
supporting his/her weight.
6. Lift the patient’s legs (one at a time) and
reach under the patient’s leg to pull the
front of the sling until it is behind the
patient’s knees (about 3 inches). The
back of the sling should remain in
position. Refer to Details “C” to “F”.
NOTE: The straps may be around, crossed
between or underneath the patient's legs.
Refer to Detail “G”.
FIGURE 5.3 Positioning the Sling
DETAIL “A”
DETAIL “B”
DETAIL “C”
DETAIL “D”
DETAIL “E”
DETAIL “F”
DETAIL “G”
Underneath
Between
Around
SECTION 5—USING DIVIDED LEG, UNIVERSAL HIGH, EASY-FIT, TOILETING, DRESS TOI-
LETING HIGH AND CRADLE ONE-PIECE STYLE SLINGS
Part No 1023891-K 17 Patient Slings
Attaching the One-Piece Style Sling to the Patient Lift
NOTE: For this procedure, refer to FIGURE 5.4.
1. Place the straps of the sling over hooks of the swivel bar. Match the corresponding colors
on each side of the sling for an even lift of the patient.
CAUTION
When connecting sling to the patient lift, the shortest of the straps MUST be at the
back of patient for support. Using long section will leave little or no support for patient's
back. The loops of the sling are color coded and can be used to place patient in various
positions. The colors make it easy to connect both sides of the sling equally. Make sure
that there is sufficient head support when lifting a patient.
2. Refer to the Lifting the Patient in the Patient Lift Owners Manual.
NOTE: Invacare Lift Swivel Bars have three hookup points per side. The middle hookup is ONLY used
for slings that have three sets of straps per side or slings that use chains or adjustable straps.
FIGURE 5.4 Attaching Slings to Lift - Slings Using Color Coded Straps
Slings Using the Cradle Attachment
The cradle sling has four straps that attach over pegs on the cradle attachment.
FIGURE 5.5 Attaching Slings to Lift - Slings Using the Cradle Attachment
Color Coded
Straps
Patient in Reclined Position
(head and leg strap lengths equal)
Cradle Attachment
Sling Attachment Point
Two Sling Attachment Points on Handle (One Attachment Point Not Shown)
Handle
Sling Attachment Point
SECTION 6—USING STANDING, STAND ASSIST, TRANSPORT AND TRANSFER ONE-
PIECE STYLE SLINGS
Patient Slings 18 Part No 1023891-K
SECTION 6—USING STANDING,
STAND ASSIST, TRANSPORT AND
TRANSFER ONE-PIECE STYLE
SLINGS
Positioning the Sling Around the Patient
Standing Sling
WARNING
Individuals that use the Standing Sling MUST be able to support the majority of
their own weight, otherwise injury can occur.
DO NOT use the standing sling in combination with the stand up lift as a transport
device. It is intended to transfer an individual from one resting surface to another
(such as a bed to a wheelchair). Moving a person using the standing sling in
combination with the stand up lift over any distance is not recommended.
DO NOT use any kind of material (such as a plastic back incontinence pad or
seating cushion) between the patient and sling material that may cause the patient
to slide out of the sling during transferring.
Before lifting the patient, make sure the bottom edge of the standing sling is posi-
tioned on the lower back of the patient and the patient's arms are outside the stand-
ing sling.
NOTE: For this procedure, refer to FIGURE 6.1 on page 19.
1. Position the standing sling around the patient's torso just below the shoulder blades
and under the arms. Refer to Detail “A”.
NOTE: Make sure the bottom edge of the sling is on the lower back and the patient's arms are
outside the standing sling.
NOTE: The belt should be towards the patient with the labels at the bottom of the standing sling.
2. Secure the standing sling to the patient by buckling the belt. Refer to Detail “B”.
WARNING
The belt MUST be snug, but comfortable on the patient, otherwise the patient can
slide out of the sling during transfer, possibly causing injury.
3. Adjust the belt for a snug, but comfortable fit. Refer to Detail “C”.
4. Attach the standing sling to the stand up lift. Refer to Attaching the Sling to the Stand
Up Lift on page 20.
SECTION 6—USING STANDING, STAND ASSIST, TRANSPORT AND TRANSFER ONE-
PIECE STYLE SLINGS
Part No 1023891-K 19 Patient Slings
FIGURE 6.1 Positioning the Slings Around the Patient - Standing and Stand Assist Slings
Transport Sling or Transfer Sling
WARNING
DO NOT raise the patient to a full standing position while using the transport or
transfer sling, otherwise injury can occur.
DO NOT use any kind of material (such as a plastic back incontinence pad or seating
cushion) between the patient and sling material that may cause the patient to slide out
of the sling during transferring.
Before lifting the patient, make sure the bottom edge of the transport or transfer sling
is at the base of the spine and the patient's arms are outside the transport or transfer
sling.
NOTE: For this procedure, refer to FIGURE 6.2 on page 20.
1. Position the transport/transfer sling behind the patient. Make sure the bottom edge of the
sling is at the base of the spine. Refer to Detail “A”.
NOTE: The labels on the transport/transfer sling should be up and facing away from the patient.
2. Lift up one of the patient's legs and position the thigh support under the leg. Refer to
Detail “B”.
3. Repeat STEP 2 for the opposite thigh support.
4. Pull the thigh support of the sling under each thigh of the patient until the thigh supports
reach completely around the thighs of the patient.
NOTE: Make sure the sling fabric is smooth underneath the thighs of the patient.
5. Attach the transport/transfer sling to the stand up lift. Refer to Attaching the Sling to the
Stand Up Lift on page 20.
SECTION 6—USING STANDING, STAND ASSIST, TRANSPORT AND TRANSFER ONE-
PIECE STYLE SLINGS
Patient Slings 20 Part No 1023891-K
FIGURE 6.2 Positioning the Slings Around the Patient - Transport or Transfer Sling
Attaching the Sling to the Stand Up Lift
Standing Sling
NOTE: For this procedure, refer to FIGURE 6.3.
1. Determine the best attachment point for the sling on each lift arm. Use the following
chart as a guideline to determine attachment point. Refer to Detail “A”.
NOTE: The following chart is for reference ONLY.
WARNING
ALWAYS use the color coded strap on the standing sling closest to the patient
while still maintaining patient stability and comfort.
Make sure the color coded strap sits flush against the attachment knob in Detail
“B”, otherwise injury or damage can result.
NOTE: The color coded straps can be used to place the patient in various positions. The colors make
it easy to connect both sides of the sling equally.
2. From the outside of the stand up lift, attach the sling to the stand up lift at the
attachment point determined in STEP 1. Refer to Detail “A”.
ATTACHMENT POINT PATIENT HEIGHT RANGE
Short under 5 ft. 6 in.
Medium 5 ft. 6 in. to 5 ft. 10 in.
Tall over 5 ft. 10 in.
DETAIL “A”
DETAIL “B”
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76

Invacare ONE-PIECE FABRIC Series Owner's Operator And Maintenance Manual

Taper
Owner's Operator And Maintenance Manual
Ce manuel convient également à