Apex Digital XT Fit Manuel utilisateur

Taper
Manuel utilisateur
XT Fit
CPAP System
Instruction Manual
Model No.: 9S-005500 Please read the instruction manual before use.
User’ manual .................................................................. P. 1
Manual de Instrucciones
.................................................
P.20
Instructions d'utilisation ................................................ P. 40
1
English
English
IMPORTANT SAFEGUARDS PLEASE MAKE THIS
PRECAUTIONS LIST AVAILABLE TO ALL USERS and
PROVIDERS
READ ALL INSTRUCTIONS BEFORE USING THIS SYSTEM
WARNINGS:
1. THIS DEVICE IS NOT INTENDED FOR LIFE SUPPORT.
2. If oxygen is used with the device, the oxygen flow must be turned off when the device is not in use.
Explanation of the Warning: When the device is not in operation and the oxygen flow is left on,
oxygen delivered into the tubing may accumulate within the device’s enclosure and create a risk of
fire.
3. Oxygen supports combustion. Oxygen should not be used while smoking or in the presence of an
open flame.
4. Always ensured the device is generating airflow before the oxygen supply is turned. Always turn off
the oxygen supply before stopping the airflow from the device.
5. This device should not be used in the vicinity of a flammable anesthetic mixture in combination
with oxygen or air and nitrous oxide.
6. The airflow for breathing generated by this device may be as much as 7°C (12.6° F) higher than
the room temperature. This device should not be used if the room temperature is warmer than
35°C (95°F) to prevent the airflow temperature from exceeding 40° C (104° F) and causing
irritation to your airway.
7. If this device overheats, it will stop operating and show Error 002on the display. After cooling
down to proper temperature, the device can restart again.
8. This machine should be used only with masks (and connectors) recommended by the
manufacturer, or by your physician or respiratory therapist. A mask should not be used unless the
CPAP machine is turned on and operating properly. The vent holes associated with the mask
should never be blocked for proper exhaling purpose. If the vent hole is blocked, the CPAP
machine will stop and show message Error 002”, after cooling down, please re-connect the
power cord to reset the machine.
9. At low CPAP pressure, some exhaled gas may remain in the mask and be re-breathed.
CAUTION
1. Make sure the environment around the machine is dry and clean. Dust and foreign particles may
affect the treatment. Keep the air inlet on the back of the machine clear to prevent overheating and
damage of the device. Do not place the machine near a source of hot or cold air. Extreme cold or
hot environment may damage user's respiratory airway.
2. If there is a possibility of electro-magnetic interference with mobile phones, please increase the
distance between devices or turn off the mobile phone.
3. Do not connect the device to the personal computer for data downloading during the treatment.
This may cause the CPAP system failure.
4. To prevent from potential allergic reactions to mask used in the device, This machine shall be used
only with masks (and connectors) recommended by the manufacturer.
5. U.S. Federal law restricts this device to sale by or on the order of a licensed physician.
2
English
English
DANGER Reducing the risk of electrocution:
1. Always unplug this product immediately after using.
2. DO NOT use while bathing.
3. DO NOT place or store product where it can fall into or be pulled into a tub or sink.
4. DO NOT place or immerse the unit into water or other liquids.
5. DO NOT reach for the unit if it is immersed in water. Unplug the unit immediately.
WARNINGS - To reduce the risk of burns, electrocution, fire or injury:
1. This product should never be left unattended when plugged in.
2. Close supervision is necessary when this product is used by, on, or near children or invalids.
Choking resulting from a child swallowing a small part that has become detached from the device
or its accessories.
3. Use this product only for its intended purpose as described in this manual. DO NOT use auxiliary
attachments that are not approved by the manufacturer.
4. Never operate this product if it has a damaged cord or plug. If it is not working properly, or has
been dropped, damaged, or immersed in water, return the product to a service center for
examination and repair.
5. Keep the cord away from hot surfaces and objects.
6. Never block the air outlets on this product. DO NOT place the unit on a soft surface such as a bed
or couch, where the air outlets may become obstructed. Keep the air outlets free of lint, hair, and
other obstructive particles.
7. Never drop or insert any object into any opening on the unit or into the tubing.
8. When disposing the unit because of damage or failure, you must comply with the regulations of
your particular region or country for disposing such equipment.
9. For therapy pressure 4-18 cmH
2
O, the default pressure compensation is available for altitude from
sea level up to 8000 feet; for therapy pressure 18.5-20 cmH
2
O, the default pressure compensation
is available for altitude from sea level up to 5000 feet.
10. No modification of this equipment is allowed.
11. Do not modify this equipment without authorization of the manufacturer.
12. If this equipment is modified, appropriate inspection and testing must be conducted to ensure
continued safe use of the equipment.
13. Do not leave long lengths of air tubing around the top of your bed. It could twist around your head
or neck while you are sleeping and lead to strangulation.
3
English
English
1. Introduction
This manual is provided to explain initial set up and operating instructions. This manual should
be stored in a safe location and be readily accessible for future reference.
1.1 General Information
Obstructive Sleep Apnea (OSA) is a condition that is an intermittent and reoccurring obstruction of
the upper respiratory tract, causing a complete (apnea) or partial (hypopnea) blockage of
respiratory airflow during sleep. The syndrome varies in degree, depending on the position of the
tongue and soft palate muscle when in a relaxed state.
The most common treatment for OSA is Continuous Positive Airway Pressure (CPAP). CPAP
devices can deliver constant air pressure into your upper respiratory tract via a nasal mask. This
constant air pressure can keep your airway open during sleep, preventing occurrences of OSA.
This unit is a CPAP device regulated by an internal micro-processor. It features a menu-driven
LCD display, universal power supply, and Ramp Time Adjustment. The Ramp Time Adjustment
feature allows the unit to begin at low pressures, allowing you to comfortably fall asleep. The unit
will then gradually “ramp up” the pressure over a configurable period of time until the appropriate
therapy pressure is reached. This unit also has a Compliance Meter that records patient usage
data that can be reviewed by a physician at a later time.
The system has been tested and successfully approved for the following standard:
EN 60601-1
EN 60601-1-2
EN 61000-3-2 Class A
EN 61000-3-3
For US and CANADA only
E348970
53DG
Medical Equipment- CPAP
WITH RESPECT TO ELECTRICAL SHOCK,
FIRE AND MECHANICAL HAZARDS ONLY
IN ACCORDANCE WITH ANSI/AAMI
ES60601-1 (2005,3rd ed.) and CAN / CSA C22.2
No. 60601-1 (2008)
4
English
English
EMC Warning Statement
This equipment has been tested and found to comply with the limits for medical devices for EN
60601-1-2:2007. These limits are designed to provide reasonable protection against harmful
interference in a typical medical installation. This unit generates, uses and can radiate radio
frequency energy and, if not installed and used in accordance with the instructions, may cause
harmful interference to other devices in the vicinity. However, there is no guarantee that
interference will not occur in a particular installation. If this equipment does cause harmful
interference to other devices, which can be determined by turning the equipment off and on, the
user is encouraged to try to correct the interference by one or more of the following measures:
Reorient or relocate the unit.
Increase the distance between the conflicting equipment.
Connect the unit into an outlet that is on a different circuit than the other device(s).
Consult the manufacturer or field service technician for help.
This system has been tested to comply with the following voluntary standards: FDA
1.2 Intended Use
This device is intended to provide continuous positive airway pressure (CPAP) for the treatment of
adult Obstructive Sleep Apnea (OSA).
Cautions: Some patients might have pre-existing contraindications for CPAP therapy, or
might experience some potential side effects of using CPAP device, please consult
your physician if you have any questions concerning your therapy.
5
English
English
2. Product Description
Components including:
1. Main CPAP device
2. Detachable power cord
3. User manual
4. Flexible air tubing with 1.8 m length
5. Carrying bag
Menu Button
Start/Standby Button
LCD Display
Up Button
Down Button
Air Outlet
Air Filter
6
English
English
3. Installation
3.1 Unpacking
To secure its contents inside, the CPAP device and accessories are bundled in a paper-packaged
box. Unpack this box by removing the CPAP and its accessory and checking for any damage,
which may have occurred during shipping. If there are damages, please contact your dealer
immediately.
3.2 Setting Up
1) Connect the power cord to CPAP device and
plug into main electrical outlet.
Once the power cord is plugged into the
electrical outlet, the device is in ready to operate
position (“STANDBY sign appears in LCD
display)
NOTE: The plug is also served to
disconnect the device.
2) Connect one end of the air tubing firmly onto the
air outlet of the CPAP.
3) Connect the other end of the air tubing to the
mask system. Putting on the mask and
headgear according to the mask instruction
manual.
7
English
English
4. Operation
NOTE: Always read the operating instruction before use.
4.1 Control Panel Description
Main control panel and button descriptions:
START/STANDBY
To start the treatment, simply press the "START/STANDBY" button. To stop the
treatment, press the "START/STANDBY" button again. The display will switch
between [STANDBY] and Therapy Pressure [ XX.X cmH
2
O] in cmH
2
O unit.
MENU
When the unit is in “Standby” mode, you can press the “MENU” button to enter the
configuration menus. The patient can adjust the “Ramp Up Time” (Ramp), Ramp
Starting Pressure (Ramp P) and Altitude Compensation (Alt). The three other
options, Therapy Pressure (P), Compliance Meter (CM), and Total Operating meter
(TM) are to be set only by a physician or other care provider. To confirm a change to a
setting, you must press the “MENU” button again. Refer to section 4.2 (Menu Option
Descriptions) for a detailed explanation.
UP
Press the “UP button to the increasing values when adjusting configurations.
DOWN
Press the "DOWN" button to decrease values when adjusting configurations.
8
English
English
4.2 Getting Started First time use.
For first-time use, follow the steps below:
1. Make sure that your physician or care provider has initialized and configured the appropriate
treatment values into the unit.
2. Make sure that your package contains a face or nasal mask with headgear. If not, your
physician or care provider can recommend the type of mask and headgear you must obtain
3. Attach tubing and mask (see Setup section).
4. Attach power cord to the unit and an electrical outlet (see Setup section). Unit will power-on
automatically.
5. Use the “MENU” button to set the Current Time, Ramp Up Time, Ramp Starting Pressure,
and Altitude Compensation (see “MENU Button Section for description of these options).
Note that these values may have been already set by your physician or care provider.
6. Press the “START/STANDBY” button Airflow to the mask will begin.
7. You can now place the Headgear and Mask over your head and face and begin treatment.
4.3 Menu Button - Option Descriptions
The following items describe the options presented to you after you press the “MENU” button on
the control panel.
(1) Ramp Time (Ramp XX MIN)
The Ramp Time feature on this unit allows the patient to fall sleep at a lower, more
comfortable pressure, after which the unit will gradually ramp up to the appropriate therapy
pressure. The first option of the “MENU” button is the Ramp Time, which is the time (in
minutes) it will take for the unit to “ramp” up to the appropriate therapy pressure. Press the
“UP” or "DOWN" buttons to set the preferred ramp time and press “MENU for confirmation.
There are ten timing options; 0, 5, 10, 15, 20, 25, 30, 35, 40 and 45 minutes.
(2) Ramp Starting Pressure (Ramp P XX.X)
Press "MENU" button to select [Ramp P XX.X] menu, press "UP" or "DOWN" button to
set the preferred ramp starting pressure and press "MENU" for confirmation. The ramp
starting pressure can be changed from 3 cmH
2
O to “Therapy Pressure 1” cmH
2
O. For
example, if your therapy pressure is 10 cmH
2
O, the maximum ramp starting pressure you can
select is 9 cmH
2
O.
(3) Therapy Pressure (P XX.XcmH
2
O)
The third option of the “MENU” button is the Therapy pressure displayed in units of cmH
2
O.
The Therapy pressure option should only be set by the equipment provider, a respiratory
therapist or a physician.
NOTE: The Therapy Pressure value itself should only be prescribed by a
physician.
9
English
English
(4) Altitude Compensation (Alt XXX)
The fourth option of the “MENU” button is the Altitude Compensation option, used to adjust
pressure in the mask based on the altitude relative to sea level. Press the “UP” or "DOWN"
buttons to select from eight options (1 to 8) that correspond to a range of altitudes listed in the
table below. The level should be set relative to your elevation above sea level. Once the
preferred option has been selected, press the “MENU” button to confirm.
NOTE:
Users can operate XT Fit at wide range of altitude within 0~8000 ft (limit to 5000 ft if
the pressure is set at 18.5-20 cmH
2
O). The altitude function provides a method of
accuracy pressure output at high altitude. Users must follow the below table to select
correct altitude setting depending on your elevation above sea level, the device will
automatically regulate airflow output to achieve targeted pressure at high altitude.
Otherwise, wrong altitude setting will cause inaccuracy pressure output.
There are eight altitude compensation levels. The level should be set depending on
your elevation above sea level. Below table is the comparison of altitude setting and
altitude.
Altitude Setting
Altitude (Imperial)
Altitude (Metric)
1
0 ~ 1000 ft
0 ~ 304 m
2
1001 ~ 2000 ft
305 ~ 609 m
3
2001 ~ 3000 ft
610 m ~914m
4
3001 ~ 4000 ft
915 m ~1219m
5
4001 ~ 5000 ft
1220 m ~1524m
6
5001 ~ 6000 ft
1525 m ~1829m
7
6001 ~ 7000 ft
1830 m ~2134m
8
7001 ~ 8000 ft
2135 m ~2438m
(5) Total Meter (TM XXXX.X)
The fifth “MENU” item is the Total Meter which displays the total number of hours that the
device has been “On” and delivering airflow. In other words, it is the time the
“START/STANDBY” button was pressed to begin airflow to the time the
“START/STANDBY” button was pressed again to stop airflow. This meter should only be
re-set by the equipment provider, a respiratory therapist or by a physician.
(6) Compliance Meter (CM XXXX.X)
Press "MENU" button to select [CM XXXX.X hr] menu, the compliance meter records the
total therapy hours for the device. The compliance meter should be re-set only by the provider,
a respiratory therapist or by a physician.
(7) Session Counter (CM XXX 4hr)
Press "MENU" button to select [CM XXX 4hr] menu, the session counter displays the
number of sessions greater than four hours. Your home care provider may ask for this
information. The session counter will be erased once the compliance meter has been reset.
10
English
English
(8) Check Value (c XXX)
Press "MENU" button to select [c XXX] menu, which displays a check value. Your home
care provider may ask for this information.
(9) Turn off the Device
Remove the power cord from the electrical outlet, and disconnect power cord from the
power socket on the back of device.
4.4 Flowchart of Menu settings
Enter the user's menu mode by pressing the “MENU” button.
In each setting, when the preferred value has been selected, press "MENU" for confirmation and
press "MENU" again to enter next selection.
NOTE: For physicians, please refer to a separated “Physician’s Additional Instruction”
page.
NOTE:
Once the setting is confirmed, press “MENU button. Otherwise, the device will
automatically go back to standby without saving the modification if no action is
taken in 20 seconds.
Ramp
(Ramp Time)
P
(Therapy Pressure)
Alt.
(Altitude Compensation)
TM
(Total Meter)
< 0, 5, 10, 15, ……………., 45> min
< 4, 4.5, 5, 5.5, ……………., 20> cmH2O
< 1, 2, 3, 4, 5, 6, 7, 8> if P set at 4-18 cmH2O
< XXXX.X > hr
< 1, 2, 3, 4, 5> if P set at 18.5-20 cmH2O
Ramp P
(Ramp Starting Pressure)
< 3, 3.5, 4, 4.5, ……………., 19> cmH2O
CM
(Compliance Meter)
CM 4 hr
(Session Counter)
c
(Check Value)
< XXXX.X > hr
< XXX >
< XXX >
11
English
English
5. Adding a Humidifier
XT Fit CPAP device can be used with XT Heated Humidifier
(9S-006500) which is available from the home care provider.
The heated humidifier may reduce nasal dryness and irritation
by providing adequate moisture and heat to the airflow. Please
refer to the XT heated humidifier (9S-006500) instruction
manual for complete setup information.
NOTE: When XT Fit CPAP device is used with the heated humidifier, its power supply is
from the power socket outlet of the heated humidifier. DO NOT connect the power cord to
CPAP device and plug into main electrical outlet.
6. Cleaning & Maintenance
6.1 Cleaning the Main Unit
The main unit should be cleaned regularly (at least every 30 days). To clean the unit, wipe it down
with a damp cloth using a mild detergent. When selecting a cleaning agent, take precautions to
choose one that is not chemically abrasive to the surface of the plastic case. All parts should be
air-dried thoroughly before use. During storage, the unit should be kept in a dust-free location.
WARNING:
Don’t try to open this device. Repairs and internal servicing should only
performed by an authorized service agent. Don’t drop any subject into the air
tube or air outlet.
6.2 Tubing and Mask
The tubing and mask should be checked and cleaned daily. Please refer to the cleaning instruction
packaged with the accessories.
1. Disconnect the air tubing from the air outlet of the device.
2. Remove the air tubing and headgear straps from the nasal mask.
3. Use a mild detergent and prepare the detergent according to manufacturer's
recommendations. Tubing and mask can be washed and rinsed in warm water.
4. Rinse thoroughly and make sure they are completely dry before next use.
5. All items of the mask and air tubing are subject to normal wear and tear and may eventually
be replaced. Replace the mask and the air tubing if they are damaged.
CAUTION:
Do not use blench, chlorine-, alcohol-, or aromatic-based (including all
scented oils), moisturizing or antibacterial soaps to clean the cushion,
mask, air tubing. These solutions may cause hardening and reduce the life
of the product.
12
English
English
CAUTION:
Do not wash or dry the mask or air tubing at a temperature above 70°C
(160°F)
WARNING:
Do not use any cleaner containing fragrance or conditioners as they will
leave a residue.
WARNING:
The mask must not be re-used by another person. This is to avoid the risk
of cross-infection.
6.3 Air Filter Maintenance
For an optimum operation of the device, the air filter can be cleaned by the user at least once
every two weeks or more often if this device is operated in a dusty environment. It is recommended
for the filter to be replaced with a new one after 6 months of use of the device; however, depending
on the air environment, the replacement time may vary.
CAUTION
:
Dirty air filter may cause high operating temperatures that affect device
performance. Ensure the air
filter is cleaned and fitted at all times.
1. Remove the dirty filter from the enclosure on the rear of the
device.
2. Wash the filter in warm water with a mild detergent, and rinse
with water. Allow the filter to air dry completely before
reinstalling. DO NOT use a filter that is not completely dry. If the
filter is torn, replace it.
3. Reinstall the filter.
NOTE: Please follow national requirements to dispose the unit properly.
13
English
English
7. Troubleshooting
The table below lists troubleshooting solutions for the problems that may happen. If the problem
persists, contact your equipment provider service agent.
Problem
Possible Causes
Solutions
No display
1. The power cord is not
connected to the power
socket.
2. LCD failure or controlled
PCB failure.
1. Ensure the power cord is
connected.
2. Contact your equipment
provider for repair.
Display code
incorrect
LCD failure or controlled PCB
failure.
Contact your equipment provider
for repair.
Illuminant under
LCD is not on
LED failure
Contact your equipment provider
for repair.
Buttons disable
Button failure
Contact your equipment provider
for repair.
Air delivered is slow
1. During ramp time.
2. Filter is too dirty.
3. Flow generator failure.
1. Check the ramp time setting
2. Change or clean the filter
regularly.
3. Contact your equipment
provider for repair.
Error / Warning Messages show in LCD
Message type
Definition
Message in LCD
Error:
Primary function can’t
execute.
Error for abnormal system settings
Error 001
Error for flow generator failure
Error 002
Warning:
Out of system memory
Warn 001
1. System memory is nearly full
2. Memory resetting problem
Warn 002
NOTE: When the warning message appears, contact your physician or equipment
provider to download the memory data and reset the meter.
14
English
English
8. Technical Specifications
Item
Specifications
Mode of Operation
Continuous
Pressure Range
4 20 cmH
2
O (adjustable in 0.5 cmH
2
O increment)
Ramp Time
0 45 minutes (adjustable in 5-minute increment)
Ramp Starting Pressure
3 to therapy pressure
Altitude Compensation
(Manual setting)
Level 1~8 for 0 ~ 8000 ft when the pressure set at 4-18 cmH
2
O.
Level 1~5 for 0 ~ 5000 ft when the pressure set at 18.5-20 cmH
2
O
Dimensions (W x D x H)
14.5 x 13.0 x 10.0 cm or 5.7” x 5.1” x 3.9”
Weight
Approximately 800 g or 1.76 lb
Sound Level
28 dBA at 10 cmH
2
O, 1 meter distance
Power Input
AC100-240V, 50/60 Hz, 0.5-0.3A
Environment
Temperature
Operating: +5° C to +35° C (+41° F to +95° F)
Storage/Transport: -15º C to +50º C (+5° F to +122° F)
Humidity
Operating: 15%RH to 95%RH non-condensing
Storage/Transport: 10%RH to 90%RH non-condensing
Air Tubing
Flexible plastic, 1.8m (approx.)
Classification:
Class II
Type BF, Applied Parts Nasal Mask.
Not suitable for use in the presence of a flammable anesthetic
mixture.
IP21: Protected against solid foreign objects of 12.5 mm and
greater; Protected against vertically falling water drops.
Continuous operation.Continuous operation.
Note: The manufacturer reserves the right to modify the specification without notice.
15
English
English
Symbols
BF symbol, which indicated this product, is according to the degree of protecting
against electric shock for type BF equipment.
Attention, should read the instructions.
Refer to instruction manual
Attention, should read the instructions.
Class II
Protected against solid foreign objects of 12.5 mm and greater; Protected against
vertically falling water drops.
Disposal of Electrical & Electronic Equipment (WEEE):
This product should be handed over to an applicable collection point for the recycling
of electrical and electronic equipment. For more detailed information about the
recycling of this product, please contact your local city office, household waste
disposal service or the retail store where you purchased this product.
Authorized representative in the European community
Manufacturer
9. NOTE, CAUTION, AND WARNING STATEMENTS
NOTE:
Indicate information that you should pay special attention to.
CAUTION:
Indicate correct operating or maintenance procedures in order to prevent damage to
or destruction of the equipment or other property.
WARNING:
Calls attention to a potential danger that requires correct procedures or practices in
order to prevent personal injury.
16
English
English
10. Servicing
The XT series systems are intended to provide safe and reliable operation according to the
instructions provided by Apex Medical. Apex Medical recommends that the XT system be
inspected and serviced by authorized technicians if there is any sign of wear or concern with
device function. Otherwise, service and inspection of the devices generally should not be required
during the five year design life of the device.
11. Limited Warranty
Apex Medical Corp. (hereafter ‘Apex’) gives the customer a limited manufacturer warranty on new
original Apex products and any replacement part fitted by Apex in accordance with the warranty
conditions applicable to the product in question and in accordance with the warranty periods from
date of purchase as listed below. This warranty does not cover damage caused by accident,
misuse, abuse, alteration and other defects not related to material or workmanship. To exercise
your rights under this warranty, please contact your local, authorized Apex dealer.
NOTE: Warranty is only valid in the country of purchase.
Product
Warranty period
CPAP devices
2 years
Water chamber
6 months
Mask systems
6 months
Disposable products
None
17
English
English
Appexdix A: EMC Information
Guidance and Manufacturer’s Declaration- Electromagnetic Emissions:
This device is intended for use in the electromagnetic environment specified below. The user of this device should
make sure it is used in such an environment.
Emissions Test
Compliance
Electromagnetic Environment-Guidance
RF emissions
CISPR 11
Group1
The device uses RF energy only for its internal function. Therefore, its RF
emissions are very low and are not likely to cause any interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class B
The device is suitable for use in all establishments, including domestic
establishments and those directly connected to the public low-voltage power
supply network.
Harmonic emissions
IEC61000-3-2
Class A
Voltage fluctuations /
Flicker emissions
IEC61000-3-3
Complies
Guidance and Manufacturer’s Declaration- Electromagnetic Immunity:
This device is intended for use in the electromagnetic environment specified below. The user of this
device should make sure it is used in such an environment.
Immunity Test
IEC60601 test level
Compliance
Electromagnetic
Environment-Guidance
Electrostatic
Discharge(ESD)
IEC61000-4-2
±6kV contact
±8kV air
±6kV contact
±8kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30 %.
Electrical fast transient/
burst
IEC61000-4-4
±2kV for power
supply line
±1kV for input/out
line
±2kV for power
supply line
±1kV for input/out
line
Mains power quality should be that
of atypical commercial or hospital
environment
Surge
IEC61000-4-5
±1kV for differential
mode
±2kV for common
mode
±1kV for differential
mode
Mains power quality should be that
of atypical commercial or hospital
environment.
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC61000-4-11
<5 % UT(>95 % dip
in UT)for 0,5 cycle
40 % UT(60 % dip
in UT)for 5 cycles
70 % UT(30 % dip
in UT)for 25 cycles
<5 % UT(>95 % dip
in UT)for 5 sec
<5 % UT(>95 % dip
in UT) for 0,5 cycle
40 % UT(60 % dip
in UT) for 5 cycles
70 % UT(30 % dip
in UT) for 25 cycles
<5 % UT(>95 % dip
in UT) for 5 sec
Mains power quality should be that
of atypical commercial or hospital
environment. If the user of this
device requires continued
operation during power mains
interruptions, it is recommended
that the device be powered from an
uninterruptible power supply or a
battery.
Power frequency
(50/60Hz) magnetic field
IEC61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristic of
atypical location in a typical
commercial or hospital
environment.
NOTE: U
T
is the a.c. mains voltage prior to the application of the test level
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64

Apex Digital XT Fit Manuel utilisateur

Taper
Manuel utilisateur