Megger ME6410920 Le manuel du propriétaire

Taper
Le manuel du propriétaire

Ce manuel convient également à

M
BM15/MJ15
Insulation Testers
USER GUIDE
GUIDE DE L’UTILISATEUR
GEBRAUCHSANLEITUNG
GUÍA DEL USUARIO
2
CONTENTS GUIDE DE L’UTILISATEUR - P.14 GEBRAUCHSANLEITUNG - S.26 GUÍA DEL USUARIO - P.38
Safety Warnings 3
General Description 4
Features and Controls 5
Working with Capacitive Loads 6
Operation
Preliminary Checks 7
Battery Replacement 7
Voltage Measurement 8
Insulation Testing 8
Choice of Test Leads 8
Using the Guard terminal 9
Specification 10
Accessories 12
Repair and Warranty 13
Before using the instrument, follow the separate instructions provided to fit
either the locking or non-locking test button.
If the locking button is fitted, extra care must be taken. See page 6.
Symbols used on the instruments
FRisk of electric shock.
GCaution, refer to accompanying documents
tEquipment protected throughout by Double Insulation
(Class II)
cEquipment complies with current EU Directives
3
GSAFETY WARNINGS
‘Safety Warnings‘ and Working with Capacitive Loads‘ must be read and understood before the instrument is used. Safety
precautions must be observed during use.
The circuit under test must be switched off, de-energised and isolated before any test connections are made.
Circuit connections must not be touched during a test.
On completion of a test, decaying voltage across the terminals is indicated on the display. Capacitive load circuits must be discharged
to below 60 V before disconnecting the test leads.
Capacitive load circuits should be shorted with a shorting link after discharge.
Remove the test leads from the instrument before opening the battery compartment.
Test leads and crocodile clips must be in good order, clean and with no broken or cracked insulation.
The instrument should not be used if any part of it is damaged.
U.K. Safety Authorities recommend the use of fused test leads when measuring voltage on high energy systems. See ‘Choice of Test
Leads‘.
NOTE
The instruments must only be used by suitably trained and competent persons.
NOTE
Users of this equipment and or their employers are reminded that Health and Safety Legislation requires them to carry out valid risk assessments of
all electrical work so as to identify potential sources of electrical danger and risk of electrical injury such as from inadvertent short circuits. Where the
assessments show that the risk is significant then the use of fused test leads constructed in accordance with the HSE guidance note GS38 ‘Electrical
Test Equipment for use by Electricians’ should be used. See ‘Choice of Test Leads’.
GENERAL DESCRIPTION
4
BM15, and MJ15 testers are completely self contained instruments designed
for high voltage insulation resistance testing in the maintenance and servicing
of rotating plant machinery, transformers, switchgear and industrial
installations.
Tests can be performed at 500 V, 1000V, 2500 V or 5000 V. Insulation measuring
range is 100 kΩto 20 GΩand Infinity. Automatic discharge for capacitive
circuits under test is provided and decaying voltage displayed.
A guard terminal can be used to minimise the effects of surface leakage when
carrying out insulation resistance tests.
Tests are initiated by pressing the BM15 ‘Test’ button (or by turning the MJ15
generator handle). Releasing the ‘Test’ button (or ceasing handle rotation)
causes the instrument to default to ac/dc voltmeter mode, with a discharge
resistor internally connected across the terminals. This feature will give
decaying voltage indication following the testing of equipment possessing
capacitance.
The moving coil meter operates a black pointer to display the readings on a
single logarithmic black scale on a white scale-plate for clarity of reading. The
movement is resiliently mounted for field use.
A slide - in ‘Pass Band‘ overlay can be inserted over the display. Appropriate
Pass/Fail bands can be marked on these windows with a permanent marker for
‘Go / No Go’ testing.
BM15 is powered by eight 1,5 V (IEC LR6) cells.
Normal power for the MJ15 is by low voltage, hand cranked, brushless a.c
generator. Fitting the battery container supplied, with eight 1,5 V IEC LR6 cells
gives the instrument a dual (independent) supply capability. Both instruments
are fitted with a battery check range, operated by pressing the ‘Test‘ push
button. Cranking the handle too slowly (<180 RPM) results in an unstable
pointer reading. Excess handle speed cannot harm the instrument as the
output voltage is electronically regulated.
The case is robust, yet light-weight, made from a flame retardant ABS, with a
polycarbonate display window. Mounted on top of the case is a 6 position,
rotary, range selection switch and a ‘Test‘ push button.
Three recessed sockets are provided, and marked ‘+’, ‘-’ and ‘G’. These have
safety covers which open when the plugs are inserted. When inserted into the
sockets, the shrouded test lead plugs lock into position. They are released by
twisting the plug a quarter turn and pulling out.
For this reason, only the test leads supplied or suitable Megger replacement
ones should be used.
Design safety features include:
External voltage, a.c. or d.c., displayed.
Load automatically discharged at the end of a test, and decaying
voltage displayed.
Test leads can lock into the case to prevent accidental disconnection.
5
FEATURES AND CONTROLS
Rotary Selector
Switch
Non Locking Button
(recommended
option)
Locking Button
(Extra care must
be taken if this
option is chosen.
See page 6)
Generator Handle (MJ15)
+ Terminal
Guard Terminal
- Terminal
TO INSERT TERMINAL PLUGS, PUSH IN - TO RELEASE, TWIST A QUARTER TURN AND PULL OUT
1. These instruments are designed to safely charge and discharge
capacitive loads up to 5 µF. To safeguard against malfunction however,
you are advised to take your own precautions when working with
capacitive loads. Remember that when charged, even low value
capacitors can be fatal on contact .
2. Extra care must be taken if the locking Test button is selected.
3. Circuit connections must not be touched when testing.
4. Care must be taken to prevent capacitive circuits becoming
disconnected during a test, leaving the circuit in a charged state.
5. On completion of a test, the instrument indicates circuit decaying
voltage. Do not disconnect test leads until the load capacitance has
discharged to below 60 V.
6. The voltmeter and automatic discharge feature of the instruments
should be regarded as additional safety features and not a substitute
for normal safe working practice.
7. If any part of the instrument is damaged, it should not be used, but
returned to the manufacturer or an approved service organization for
repair.
G
Circuit connections and the instrument terminals may become
hazardous when connected to Capacitive loads.
WORKING SAFELY WITH CAPACITIVE LOADS
6
Do not forget to release a locked Test button and discharge
any capacitance before touching the item under test or
disconnecting the test leads.
7
PRELIMINARY CHECKS
Preliminary Performance Checks
The instrument will operate in any position, but best results are achieved
when the instrument is face up, on a firm level surface. This is particularly true
for hand cranked units to obtain a crank speed of >180 rev/min.
1) If battery cells are fitted, switch to Battery Check range and press the
‘Test‘ push button. Confirm that the pointer settles within the battery
symbol portion of the scale.
2) With the test leads disconnected, set the rotary selector switch to the
appropriate insulation range, press and hold down the ‘Test‘ button,
(or turn the generator handle). The meter pointer should move up the
scale briefly and then return to the ‘‘ (infinity) position on the scale.
This establishes that there is no leakage through the instrument itself.
3) Check that the test leads and crocodile clips are in good order, clean
and with no broken or cracked insulation. Connect two of the test
leads to the ‘+’ and ‘-’ terminals and ensure that their clips are not
touching anything.
4) Press the ‘Test‘ button again (or turn the generator handle) and
observe the meter pointer. The pointer should rest over the ‘
position on the scale. If it does not, the test leads may be faulty and
should be inspected more closely for damage. Replace them if
necessary.
5) Connect the test lead clips together, press the ‘Test‘ button (or turn the
generator handle) and observe the meter pointer. The meter should
read ‘0Ω‘. If it indicates infinity or a high resistance value the leads
may be open circuit and should be inspected further. Replace them if
necessary. (Shorting the leads together and obtaining a ‘0 Ω‘ reading
also indicates that the instrument is working).
Battery Replacement
The cells are housed in a battery compartment in the base of the instrument.
To change the cells, use a screwdriver to remove the battery cover securing
screws and lift off the battery compartment cover. Observing the correct
polarity as marked on the battery housing, install 8 replacement IEC LR6 (AA)
cells. Replace and secure the battery compartment cover on completion.
GWARNING
The battery contacts are not isolated from the test leads. Remove the test
leads from the instrument before opening the battery compartment.
OPERATION
8
Voltage measurement
When switched to the ‘Vposition, the instrument measures up to 600 V a.c.
to the specified accuracy of the instrument. DC voltage is also indicated on the
display, but not to the specified accuracy. When not testing (i.e. in standby
mode) and connected to a live circuit, the instruments default to voltmeter (0
to 600 Volts a.c. or d.c.) irrespective of the rotary switch position. Any voltage
present will immediately be shown. Thus indication is given that the item has
not been completely de-energized. The instrument also monitors circuit
discharge when the ‘Test‘ button is released following an insulation test on a
capacitive item, e.g. a long cable.
Insulation Testing
After connecting the test leads to the instrument and carrying out the
Preliminary Performance Checks:
1) Set the selector switch to the required test voltage. Connect the test
leads to the isolated circuit to be tested, as follows:-
(a) For insulation tests to earth (ground):- Connect either test
lead to earth (ground) or the frame of the equipment,and
the other lead to that part of the circuit to be tested.
b) For insulation tests between wires:- Connect a lead to the
core of each of the wires.
2) Press the ‘Test‘ button (or turn the generator handle).
3) The meter pointer will indicate the value of insulation resistance on
the Ωscale.
If a capacitive circuit is tested, the pointer will initially deflect towards 0 Ωand
then gradually rise to its final steady value as the capacitance is charged up to
the output voltage of the tester.
If several successive readings of (infinity) are obtained, connect the two
farthest ends of the test leads together and carry out a check on the leads. A
0Ωreading should result which double checks that the leads are not
disconnected or broken and therefore, the insulation resistance readings are
correct.
Capacitive circuits automatically discharge through the tester when the ‘Test‘
button is released. Decaying discharge voltage will be indicated on the voltage
scale. Wait a few moments for the voltage to decay to below 60 V before
disconnecting the test leads.
Choice of Test Leads
BM15/MJ15 are supplied with three unfused leads terminated in crocodile
clips. These are the best leads to use for insulation tests on non-live circuits.
The crocodile clips ensure that any capacitive load remains connected until it
is automatically discharged at the end of a test.
GS 38 (UK Safety Authority Guidance Note) advises the use of fused test prods
if the instrument is to be used for making voltage measurements on live, high
energy circuits. These leads are available as an optional extra. See
Accessories’.
Fused prods must not be used for insulation testing. If the fuse should
rupture, or the prods lose contact during a test, the system under test would
remain charged without any apparent evidence of danger!
Do not forget to release a locked Test button, and discharge
any capacitance before touching the item under test, or
disconnecting the test leads.
9
Using the Guard terminal (G)
For basic insulation tests and where there is little possibility of surface leakage
affecting the measurement, it is unnecessary to use the guard terminal (if the
insulator is clean and there are unlikely to be any adverse current paths).
However in cable testing, there may be surface leakage paths across the
insulation between the bare cable and the external sheathing due to the
presence of moisture or dirt. Where it is required to remove the effect of this
leakage, particularly at high testing voltages, a bare wire may be bound tightly
around the insulation and connected via the third test lead to the guard
terminal ‘G’.
The guard terminal is at the same potential as the negative terminal. Since the
leakage resistance is effectively in parallel with the resistance to be measured,
the use of the guard causes the current flowing through surface leakage to be
diverted from the measuring circuit. The instrument therefore reads the
leakage of the insulator, ignoring leakage across its surface.
OPERATION BLOCK DIAGRAM
to ‘-’ve
terminal
Leakage Path Tightly bound bare wire
to ‘+’ve
terminal
to ‘G’
terminal
10
Insulation
Range: 100 kΩto 20 GΩ(also 0 Ωand )
Test Voltages (d.c.): 500 V, 1000 V, 2500 V & 5000 V
Test Voltage Accuracy: ±5% of nominal test voltages on 20 MΩload
Test Voltage Stability: <±1% (180 r.p.m. to 240 r.p.m. MJ15)
Insulation Accuracy see page 11
Short Circuit Current: 1,5 mA ± 0,5 mA
Maximum capacitance of load: 5 µF
Interference Rejection: 1 mA rms at 50 to 60 Hz
Discharge Resistor: <500 kΩ
Voltage
Range: 0 to 600 V a.c. indication of d.c.
Accuracy: see page 11
General
Overload rating: 720 V a.c. or d.c.
Scale Length: 72 mm (96˚)
Power Supply: BM15 8 x LR6 (AA alkaline or rechargeable) cells
MJ15 Low voltage brushless Generator or 8 x LR6 (AA alkaline) cell battery
Battery life: Typically 2000 five second tests at 5 kV on 100 MΩload
Battery Indicator: Loaded battery test
SPECIFICATION
11
Safety: IEC 61010-1
Non replaceable Fuse: 1 Amp, 250 V, HBC type (F) (20 mm x 5 mm) to IEC 127/1
This fuse protects the instrument against any faults occurring when using rechargable batteries.
E.M.C: IEC 61326-1
Operational uncertainties: Refer to www.megger.com
Temperature Range:
Operating: 0˚C to 30˚C (32˚F to 86˚F) at full specification
Operating: -20˚C† to 50˚C (-4˚F to 122˚F) to temperature coefficient ±0,1% / ˚C
Storage: -25˚C to 65˚C (-13˚F to 149˚F)
Humidity: 90% RH maximum at 40˚C (104˚F)
Dimensions: 220 mm x 160 mm x 115 mm (8,9 in x 6,3 in x 4,5 in)
Weight: BM15 Approx 1.2 kg
MJ15 Approx 1,6 kg, or 1,8 kg with battery holder and cells.
Cleaning: Wipe the disconnected instrument with a clean cloth dampened with soapy water or Isopropyl Alcohol (IPA)
***Relates to transient overvoltage likely to be found in fixed installation wiring.
*Relates to transient overvoltage likely to be found in special equipment or parts of equipment, telecommunication, electronic etc.
† Note: MJ15 will operate as normal by generator at full temperature range
BM15 will require new healthy battery cells to operate as normal at -20˚C
ACCESSORIES
12
SUPPLIED WITH THE INSTRUMENT PART NUMBER
User Guide 6172-209
Battery of 8 x 1,5 V LR6 (AA Alkaline) cells
3 x 3 m HV leads 8101-181
Test Record Card (5 supplied) 6172-111
(U.S. 210949)
Carrying case with lead storage 6420-117
Slide in Pass Band overlay (5 supplied) 6121-401
AVAILABLE AS AN OPTIONAL EXTRA
5 kV Fused Prod Set 6320-240
5 kV Calibration Box - CB101 6311-077
Test Record Card (Pack of 20) 6111-216
PUBLICATIONS
A Stitch In Time’ AVTM21-P8B
’Lowdown on HV d.c. Testing’ AVTM22P-1
13
REPAIR AND WARRANTY
The instrument circuit contains static sensitive devices, and care must be taken
in handling the printed circuit board. If the protection of an instrument has
been impaired it should not be used, and be sent for repair by suitably trained
and qualified personnel. The protection is likely to be impaired if, for example,
the instrument shows visible damage, fails to perform the intended
measurements, has been subjected to prolonged storage under unfavourable
conditions, or has been exposed to severe transport stresses.
New Instruments are Guaranteed for 1 Year from the
Date of Purchase by the User.
Note: Any unauthorized prior repair or adjustment will automatically
invalidate the Warranty.
Instrument Repair and Spare Parts
For service requirements for Megger Instruments contact:
Megger Limited or Megger
Archcliffe Road Valley Forge Corporate Center
Dover 2621 Van Buren Avenue
Kent, CT17 9EN. Norristown, PA 19403
England. U.S.A.
Tel: +44 (0) 1304 502243 Tel: +1 (610) 676-8579
Fax: +44 (0) 1304 207342 Fax: +1 (610) 676-8625
ukenquires@megger.com usenquires@megger.com
or an approved repair company.
Approved Repair Companies
A number of independent instrument repair companies have been approved
for repair work on most Megger instruments, using genuine Megger spare
parts. Consult the Appointed Distributor / Agent regarding spare parts, repair
facilities and advice on the best course of action to take.
Returning an Instrument for Repair
If returning an instrument to the manufacturer for repair, it should be sent
freight pre -paid to the appropriate address. A copy of the Invoice and of the
packing note should be sent simultaneously by airmail to expedite clearance
through Customs. A repair estimate showing freight return and other charges
will be submitted to the sender, if required, before work on the instrument
commences.
14
GUIDE DE L’UTILISATEUR - TABLE DES MATIÈRES GEBRAUCHSANLEITUNG - S.26 GUÍA DEL USUARIO - P.38
Avertissements relatifs à la sécurité 15
Description Générale 16
Caractéristiques et Commandes 17
Opérations sur des charges capacitive 18
Exploitation
Vérifications Préliminaires 19
Remplacement des piles 19
Mesure des tensions 20
Essai d’isolement 20
Choix des câbles d’essai 20
Utilisation de la borne de protection 21
Spécifications 22
Accessoires 24
Réparation et Garantie 25
Avant d’utiliser l’instrument, suivre les instructions séparées pour monter soit
le bouton d’essai verrouillable soit le bouton d’essai non verrouillable. Le
bouton d’essai verrouillable exige des soins supplémentaires.
Symboles utilisés sur l'instrument
FRisque de choc électrique
GConsulter le Guide de l'utilisateur
tÉquipement entièrement protégé par un isolement
double (Classe II).
cÉquipement conforme aux Directives en vigueur de
l'UE.
15
GAVERTISSEMENTS RELATIFS À LA SÉCURITÉ
Les Avertissements relatifs à la sécurité et les instructions Opérations sur des charges capacitives doivent être lus et compris
avant d’utiliser l’instrument. Les précautions de sécurité doivent être observées pendant l’usage.
Le circuit à tester doit être coupé, désexcité et isolé avant d’effectuer des connexions d’essai quelconques.
Les connexions du circuit ne doivent pas être touchées pendant l’essai
A la fin de l’essai, la tension de déclin aux bornes est indiquée sur l’affichage. Les circuits à charge capacitive doivent être déchargés
en dessous de 60 V avant de déconnecter les câbles d’essai.
Les circuits à charge capacitive doit être court-circuités avec un cavalier de court-circuit après la décharge.
Enlever les câbles d’essai de l’instrument avant d’ouvrir le compartiment des piles.
Les câbles d’essai et les pinces crocodiles doivent être en bon état, propres et avec un isolement sans cassures ni fissures.
L’instrument ne doit pas être utilisé si l’un de ses constituants est endommagé.
Les Services de sécurité Britanniques recommandent l’utilisation de câbles d’essai équipés de fusibles pour lade la mesure de la
tension des systèmes à haute énergie. Voir “Choix des câbles d’essai”.
REMARQUE
Seule une personne düment formée et compétente doit se servir de cet instrument.
NOTE
ll est rappelé aux utilisateurs de cet instrument et/ou à leurs employeurs qu’ils doivent évaluer les risques de tous les travaux électriques de manière à
identifier les sources possibles de danger électrique et les risques de blessure tels que ceux des courts-circuits involontaires. Lorsque ces évaluations
indiquent des risques importants, des câbles d’essai équipés de fusibles tels que recommandés par les Services de sécurité Britanniques doivent être
employés. Des sondes équipées de fusibles ne doivent pas être utilisées pour les essais d’isolement. En cas de rupture du fusible ou si les sondes
perdent contact pendant l’essai, le circuit testé peut rester chargé sans aucun signe évident du danger !
DESCRIPTION GÉNÉRALE
16
Les testeurs BM15 et MJ15 sont des instruments autonomes conçus pour la
mesure des résistances d’isolement sous haute tension pour l’entretien et le
dépannage des machines rotatives, des transformateurs, des appareillages de
commutation et des installations industrielles.
Les tensions d’essai peuvent être sélectionnées à 500 V, 1000 V, 2500 V ou 5000
V. La plage des mesures d’isolement est de 100 kΩà 20 GΩet l’infini. La
décharge automatique des circuits capacitifs est prévue, et la tension de déclin
est affichée. La borne de protection peut être utilisée pour minimiser les effets
de fuites de surface lors des essais de résistance d’isolement.
Les essais sont initiés en appuyant sur le bouton “Test” du BM15 (ou en
tournant la manivelle du générateur du MJ15). Le relâchement du bouton
“Test” (ou l’arrêt de la rotation de la manivelle) remet l’instrument dans le
mode voltmètre ca/cc par défaut, avec une résistance de décharge connectée
intérieurement aux bornes. Cette fonction donnera une indication de la
tension de déclin à la suite d’essai d’équipements ayant une capacité.
L’appareil de mesure à bobine mobile actionne une aiguille noire pour afficher
clairement les résultats sur une échelle logarithmique noire simple sur fond
blanc. Le mouvement est monté de façon élastique pour utilisation sur le
terrain.
Un gabarit transparent de bande passante peut être glissé sur l’affichage. Des
repères Bon/Mauvais appropriés peuvent être marqués sur ces fenêtres avec
un marqueur permanent.
L’instrument BM15 est alimenté par huit piles de 1,5V (IEC LR6).
L’alimentation normale du MJ15 est assurée par une génératrice c.a. basse
tension, sans balais, tournée à la main. L’installation du boîtier de piles et de
huit piles de 1,5 V (IEC LR6) donne à l’instrument une alimentation double
(indépendante). Une rotation trop lente de la manivelle (< 180-t/mn) produit
des lectures instables. Une rotation excessive ne peut pas endommager
l’instrument car la tension de sortie est régulée électroniquement.
Le boîtier robuste, mais léger, est réalisé en ABS ignifuge, avec une fenêtre
d’affichage en polycarbonate. Le dessus du boîtier comporte un commutateur
rotatif de sélection de gamme à six positions, et un bouton poussoir “Test”.
Trois douilles encastrées sont prévues, marquées “+”, “-” et “G”. Elles sont
dotées de couvercles de sécurité qui s’ouvrent lorsque les prises sont insérées.
Les prises gainées des câbles d’essai se verrouillent dans les douilles. Elles sont
libérées en tournant la prise d’un quart de tour et en la tirant.
Pour cette raison, seuls les câbles d’essai fournis, ou les remplacements
Megger convenables, doivent être utilisés.
Les caractéristiques intégrales de sécurité comprennent:
Affichage de tension extérieure c.a. ou c.c.
Décharge automatique de la charge en fin d’essai, avec affichage de la
tension de déclin.
Les douilles d’essai verrouillent les câbles d’essai dans le boîtier et
empêchent toute déconnexion accidentelle.
17
CARACTÉRISTIQUES ET COMMANDES
Commutateur de
sélection rotatif
Bouton non
verrouillable (option
recommandée)
Bouton verrouillable
(prendre des soins
particuliers lorsque
cette option est
choisie)
Manivelle de génératrice (MJ15)
Borne +
Borne de protection
Borne -
POUR INSÉRER LA PRISE, POUSSER - POUR LA LIBÉRER, TOURNER D’UN QUART DE TOUR ET TIRER.
OPÉRATIONS SUR DES CHARGES CAPACITIVE
18
1. Ces instruments sont conçus pour charger et décharger en toute
sécurité des charges capacitives allant jusqu’à 5 µF. Cependant, pour
vous protéger contre tout mauvais fonctionnement, nous vous
conseillons de prendre vos propres précautions lorsque vous travaillez
sur des circuits capacitifs. N’oubliez jamais qu’une fois chargés, mêmes
les condensateurs de faible valeur peuvent être mortels au contact.
2. Prendre des soins particuliers lorsque le bouton Test” verrouillable
est choisi.
3. Les connexions du circuit ne doivent pas être touchées pendant les
essais.
4. Veiller à ce que les circuits capacitifs ne se déconnectent pas en cours
d’essai, laissant le circuit dans un état chargé.
5. A la fin de l’essai, l’instrument indique la tension de déclin. Ne jamais
déconnecter les câbles d’essai tant que la capacitance de charge n’est
pas déchargée en dessous de 60 V.
6. Le voltmètre et la fonction de décharge automatique de l’instrument
doivent être considérés comme des fonctions de sécurité
supplémentaires et non pas comme remplaçant les pratiques normales
de travail sûr.
7. L’instrument ne doit pas être utilisé si l’un de ses constituants est
endommagé. Il doit être renvoyé au fabricant ou à un organisme
d’entretien approuvé pour réparation.
G
Les connexions du circuit et les bornes del’instruments
peuvent devenir dangereuses lorsque connectées à des
charges capacitives.
Ne pas oublier de libérer le bouton “Test” verrouillé,et de
décharger toute capacitance avant de toucher au composant
testé, ou de débrancher les câbles d’essai.
19
EXPLOITATION
Vérifications préliminaires de performance
L’instrument fonctionne dans n’importe quelle position, mais les meilleurs
résultats sont obtenus lorsque l’instrument est face vers le haut, sur une
surface horizontale ferme. Ceci est plus particulièrement vrai pour les
instruments à manivelle qui nécessitent une vitesse de rotation constante et
supérieure à 180 t/mn.
1. Lorsque des piles sont montées, sélectionner la gamme Contrôle Pile et
appuyer sur le bouton “Test”. Vérifier que l’aiguille se stabilise dans la
partie de l’échelle portant le symbole des piles.
2. Les câbles d’essai étant déconnectés, placer le commutateur rotatif sur
la gamme d’isolement requise, appuyer sur le bouton “Test” et le
maintenir appuyé, (ou tourner la manivelle de la génératrice). L’aiguille
de l’instrument devrait monter brièvement sur l’échelle puis revenir à
la position de l’échelle. Ceci vérifie qu’il n’y a pas de fuite dans
l’instrument lui-même.
3. Vérifier que les câbles d’essai et les pinces crocodiles sont en bon état,
propres et avec un isolement sans cassures ni fissures. Connecter deux
des câbles d’essai aux bornes “-” et “+”, et vérifier que leurs pinces ne
touchent rien.
4. Appuyer de nouveau sur le bouton “Test” (ou tourner la manivelle de
la génératrice) et observer l’aiguille. Elle doit reposer sur le symbole
de l’échelle. Sinon, les câbles d’essai peuvent être défectueux et
devront être soigneusement inspectés pour tout signe
d’endommagement. Les remplacer si nécessaire.
5. Connecter les pinces des câbles d’essai ensemble et appuyer sur le
bouton Test” (ou tourner la manivelle de la génératrice) et observer
l’aiguille. L’instrument doit indiquer “0Ω“. S’il indique l’infini ou une
résistance élevée, les câbles peuvent être en circuit ouvert et devront
être inspectés. Les remplacer si nécessaire. (Le court-circuitage des
câbles et l’obtention d’une lecture de “0Ω“ indiquent aussi que
l’instrument fonctionne correctement).
Remplacement des piles
Pour changer les piles, utiliser un tournevis pour déposer les vis de fixation du
couvercle et déposer le couvercle du compartiment des piles.
En observant la polarité correcte marquée dans le logement des piles, installer
huit piles de rechange IEC LR6 (AA). Remonter et fixer le couvercle du
compartiment des piles.
GATTENTION
Les contacts des piles ne sont pas isolés des câbles d’essai. Retirer les câbles
d’essai de l’instrument avant d’ouvrir le compartiment des piles. Les piles
sont logées dans un compartiment dans la base de l’instrument.
EXPLOITATION
20
Mesure des tensions
Lorsqu’il est à la position “V, l’instrument peut mesurer des tensions allant
jusqu’à 600 V c.a.avec la précision spécifiée pour l’instrument. Les tensions c.c.
sont également indiquées sur l’affichage, mais pas avec la précision spécifiée.
Lorsqu’il n’effectue pas d’essai (c’est-à-dire en mode d’attente) et s’il est
connecté à un circuit sous tension, l’instrument passe automatiquement au
mode de voltmètre (0 à 600 V c.a. ou c.c.) quelle que soit la position du
commutateur rotatif. Toute tension présente sera immédiatement affichée.
Ceci indique donc que le circuit testé n’est pas complètement désexcité.
L’instrument contrôle également la décharge du circuit lorsque le bouton
“Test” est relâché à la fin d’un essai d’isolement sur un article capacitif, par
exemple un câble long.
Essai d’isolement
Une fois que les câbles d’essai sont connectés à l’instrument et que les
vérifications préliminaires sont terminées:
1) Placer le commutateur de sélection sur la tension requise. Connecter
les câbles d’essai au circuit isolé à tester, comme suit-:
(a) Pour les essais d’isolement à la terre (masse) :- Connecter
un des câbles d’essai à la terre (masse) ou au châssis de
l’équipement, et l’autre câble à la partie du circuit à tester.
(b) Pour les essais d’isolement entre conducteurs : -
Connecter un câble au fil de chaque conducteur.
2) Appuyer sur le bouton “Test” (ou tourner la manivelle de la
génératrice).
3) L’aiguille de l’instrument indiquera la valeur de la résistance
d’isolement sur l’échelle Ω.
Dans le cas d’essai d’un circuit capacitif, l’aiguille déviera initialement vers 0Ω
puis augmentera progressivement vers sa valeur stable finale au fur et à
mesure de la charge de la capacitance jusqu’à la tension de sortie du testeur.
Si plusieurs lectures successives de (infini) sont obtenues, connecter les
deux extrémités les plus éloignées des câbles d’essai et effectuer un contrôle
des câbles. Une lecture de 0Ωdoit être obtenue, ce qui confirme que les
câbles ne sont pas déconnectés ou cassés, et que par conséquent les lectures
d’isolement sont correctes.
Les circuits capacitifs se déchargent automatiquement par l’intermédiaire du
testeur lorsque le bouton “Test” est relâché. La tension de déclin de la
décharge est affichée sur l’échelle de tension. Attendre quelques instants pour
que la tension tombe en dessous de 60 volts avant de déconnecter les câbles
d’essai.
Choix des câbles d’essai
Les instruments BM15 et MJ15 sont fournis avec trois câbles d’essai sans
fusible, équipés de pinces crocodile. Ce sont les meilleurs câbles d’essai à
utiliser pour les essais d’isolement sur circuits hors tension. La pince crocodile
garantit que toute charge capacitive reste connectée jusqu’à ce qu’elle soit
automatiquement déchargée à la fin de l’essai.
Les Services de sécurité britanniques recommandent l’utilisation de câbles
d’essai équipés de fusibles pour la mesure de la tension des systèmes sous
tension à haute énergie. Ces câbles sont disponibles en option contre
supplément. Voir Accessoires”.
Les sondes équipées de fusible ne doivent pas être utilisées pour les essais
d’isolement. En cas de rupture du fusible ou si les sondes perdent contact
pendant l’essai, le circuit testé peut rester chargé sans aucun signe évident
du danger!
Ne pas oublier de libérer un bouton “Test” verrouillé, et de
décharger la capacitance avant de toucher au circuit testé, ou
de déconnecter les câbles d’essai.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48

Megger ME6410920 Le manuel du propriétaire

Taper
Le manuel du propriétaire
Ce manuel convient également à