Omron NE-U22V Manuel utilisateur

Taper
Manuel utilisateur

Ce manuel convient également à

1
INSTRUCTION MANUAL
Thank you very much for purchasing
Nebulizer.
Read all instructions in the manual on the
care and use of this unit before operating.
Save these instructions for future reference.
VIBRATING MESH
NEBULIZER
Model NE-U22V
®
¤
This unit is a medical device. Operate this unit only as instructed by your physician
and/or Respiratory Therapist.
The illustrations shown in this Instruction Manual are image drawings.
Caution: Federal law restricts this device
to sale by or on the order of a physician.
Before Using the Unit
Safety Information . . . . . . . . . . . . . . . . . . . . . . 2
Features of the Product . . . . . . . . . . . . . . . . . . . 5
Components of the Product. . . . . . . . . . . . . . . . 6
Names and Functions of the Parts. . . . . . . . . . . 7
How to Assemble the Main Unit. . . . . . . . . . . . 8
Correct Use of the Unit
How to Use the Power Source . . . . . . . . . . . . 10
How to Fill the Medication Bottle . . . . . . . . . 12
How to Select the Nebulization Mode . . . . . . 14
How to Inhale . . . . . . . . . . . . . . . . . . . . . . . . . 15
How to Care for Unit After Use . . . . . . . . . . . 16
How to Disinfect the Unit
How to Disinfect. . . . . . . . . . . . . . . . . . . . . . . 18
How to Replace the Mesh Cap . . . . . . . . . . . . 19
Troubleshooting and Warranty
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . 20
Specifications /
Replacement Parts and Accessories . . . . . . . . 22
Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
¤
EnglishFrançais
Español
Safety Information
2
Safety Information
3
Before Using the Unit Safety Information
Safety InformationBefore Using the Unit
The warning signs and icons are shown below so that you use this product safely
and correctly to prevent the risk of damage to yourself and others.
Contents
Indicates matters in which the possibility of death or serious injury may
arise as a result of incorrect handling.
Examples of signs
For the type, dose, and regimen of the medication, be sure to follow the instructions of a doctor.
•Your symptom(s) may worsen.
When using the unit for the first time after purchasing it or after not using it for a long period of
time, be sure to wash and disinfect Medication Bottle and Mesh Cap. (Refer to Page 14-18)
Miscellaneous bacteria may be present and you could be infected.
Wash and disinfect Mesh Cap, Inhalation Mask, Mouthpiece, and Mask Adapter after each use.
(Refer to Page 18)
Be sure to dry the cleaned and disinfected parts promptly, and store them without
getting re-contamination.
Miscellaneous bacteria may be present and you could be infected.
If the unit is used by more than one person, be sure to use the disinfected Medication Bottle Unit,
Child’s Inhalation Mask, Mouthpiece, and Mouth Adapter for each user.
The user may be infected.
Keep the unit out of reach of infants or children.
They may swallow small pieces such as Mesh Cap. The unit may be damaged or may cause injury
to infants or children. In case a child swallows the Mesh Cap, immediately consult with a physician.
Do not plug in or pull out the AC Adapter with wet hands.
•You could suffer electric shock.
Do not wash the Main Unit and AC Adapter with water or hot water.
Avoid immersing them in water.
Can damage unit.
Warning signs
The icon indicates prohibitions (what you should not do).
Matters involving actual prohibitions are indicated by text or pictures
in or near .
The icon to the left refers to “do not to disassemble”.
The icon indicates prohibitions (including warning and danger).
Matters involving actual prohibitions are indicated by text or pictures
in or near .
The icon to the left refers to “caution for electric shock”.
Indicates matters in which bodily harm or material damage* may
arise as a result of incorrect handling.
*Material damage refers to a wide range of damage involving your house,
household goods, domestic animals, and pets.
The icon indicates something that is important (always follow).
Matters involving actual important actions are indicated by text
or pictures in or near .
The icon to the left refers to “important information”.
Do not disassemble, repair, or modify the unit. This may void the warranty.
If you are not going to use the unit for a long period of time, be sure to unplug the
power plug from the electric outlet.
Do not use disinfecting solution containing sodium hypochlorite.
Mesh cap will rust.
Do not spill medication on the Main Unit or the AC Adapter. If you spill
medication on the Main Unit or the AC Adapter, immediately wipe
it off with a soft cloth.
Do not drop or apply strong shock to the Main Unit, AC Adapter,
Medication Bottle, or the Mesh Cap.
Use only the Omron AC Adapter provided for this product.
Be sure to plug the AC Adapter into a 110 VAC outlet.
Rinse the disinfectant from the disinfected pieces with water thoroughly.
Before Using the Unit Features of the Product
Features of the Product
5
Safety Information
4
Safety InformationBefore Using the Unit
General Advice
(This is an image of the unit using an Inhalation Mask.)
Replace the worn batteries with new ones immediately.
Do not use different types of batteries together.
Do not use new and worn batteries together.
Do not use dry batteries and rechargeable batteries together.
Be careful not to insert batteries in the wrong way so that unit operates correctly.
You can use commercially available rechargeable NiMH batteries, but do not use
nickel cadmium batteries.
Charge the rechargeable batteries with a commercially available charger.
(The AC Adapter enclosed in the package cannot be used as a battery charger.)
Wash Medication Bottle, Mesh Cap, and Vibrator with clean tap water or commercially
available distilled water.
Wash Medication Bottle, Mesh Cap, and Vibrator with clean tap water or the commercially
available distilled water.
When rinsing Medication bottle under the tap, use gentle pressure, not strong running tap water.
Assemble the cleaned and disinfected parts after sufficiently drying them.
Do not use benzene, thinner, or gasoline to clean the unit.
Do not dry Main Unit, AC Adapter, Medication Bottle, and Mesh Cap in a microwave oven
or boil them to clean..
Do not store the unit under high temperature and high humidity, or under direct sunshine.
Do not leave or carry the Medication Bottle Unit filled with medication or water.
Remove batteries from the unit when you are not going to use the unit for a long period
(approximately three months or more).
To start inhalation, you must assemble the unit.
Before starting to use the unit, please read the section on “How to
Assemble the Main Unit” and understand the points for assembling.
(Refer to Pages 8 and 9)
1
Pocket-sized unit can be easily carried anywhere you go
The unit is small enough to be placed in a carrying / storage case or purse, you can
easily carry it around wherever you go. The new titanium alloy mesh creates
a high efficiency aerosol.
2
Inhalation is possible at various angles
The new bottle structure enables inhalation at various angles with stable nebulization.
Therefore, the unit can be used even at bedside.
3
Simple care and easy use
The unit does not need any complex care. Just like the traditional compressor type
nebulizer, all you have to do is to remove the medication after inhalation and dry
the bottle immediately after washing it.
Before Using the Unit Names and Functions of the Parts
Names and Functions of the Parts
7
Components of the Product Please check the product.
6
Components of the ProductBefore Using the Unit
Main Unit Cover
Stores the main unit.
Main Unit Cover
Power Button
Turns Main Unit
on and off.
Mesh Cap
An alloy mesh is installed for
nebulization.
Handle mesh with care because
it is very delicate.
Medication Bottle Cap Locking Lever
Opens the Medication Bottle Cap for
washing.
Do not open Medication Bottle Cap
except for washing.
Medication Bottle
A container to store the medication
The following items are contained in the box.
If any item is missing, contact the store where you purchased the
product or the nearest OMRON dealer.
Main Unit Medication Bottle Mesh Cap
(with Mesh)
Instruction Manual
Instructional Video Storage Case
Power Plug
Main Unit Locking Board
AC Adapter as Standard Supply
Cord Band
AC Adapter Connection Stand
Battery Low Indicator
Orange light blinks when
the batteries are worn.
Electrode
Supplies the power from the
Main Unit to the Vibrator
of Medication Bottle.
Medication Port
Fill the medication by opening
the Mesh Cap.
Bottom of Main Unit
Power Indicator
A green lamp turns on when the
power is on.
Battery Cover
Disengaging Lever
AC adapter Jack on
Main Unit
Mask Adapter
®
Mouthpiece
Child’s Mask
(Optional) 9921
AC Adapter (Optional)
How to Assemble the Main Unit
9
Before Using the Unit How to Assemble the Main Unit
How to Assemble the Main Unit
8
How to Assemble the Main UnitBefore Using the Unit
This unit must be assembled before it is used.
–Parts may fit tightly since they are made to prevent the medication from leaking.
–You must wash and disinfect all removable parts before using the unit.
Assemble the unit after washing, disinfecting, and drying.
(For how to disinfect the unit, refer to Page 18)
2 Attach the Medication Bottle to the Main Unit.
3 Attach the Mask Adapter to the Main Unit.
If you are not going to use the unit soon after it is assembled,
disassemble the unit (as shown in the Figure to the right)
and store in the supplied Storage Case or store it in
clean environment.
(For the correct use of the unit, refer to Page 15)
Insert the
Mesh Cap
vertically.
Align both
electrodes with
each other.
Do not open the
bottle cap.
You have finished
installation.
You have finished
installation.
You have finished
assembly.
Click
You have finished
assembly.
Now you have completed assembling the Main Unit.
General Advice
Hold the unit securely with both hands.
Install the parts securely. (You may hear clicks depending on the parts
at the time of installation.)
1 Attach the Mesh Cap on the Medication Bottle.
Close it
securely.
You have finished
installation.
4 Attach the supplied Mouthpiece or Child’s Inhalation Mask to the Mask Adapter.
How to attach the Child’s Inhalation Mask How to attach the Mouthpiece
How to Use the Power Source
11
Correct Use of the Unit How to Use the Power Source
How to Use the Power Source
10
How to Use the Power Source
How to Use optional AC Adapter (optional)
Use the AC Adapter in a place where you have an electric outlet.
However, please note that this AC Adapter is not a battery charger.
1 Place the Main Unit on the AC Adapter Connection Stand as shown below.
How to remove the AC Adapter from the Main Unit
(1) Unplug the power plug from the electric outlet.
(2) Unlock the Main Unit by pushing in both ends
of the Main Unit Locking Board of the AC
Adapter Connection Stand.
(3) Remove the Main Unit.
(1) Unplug
(3) Remove
(2) Push both ends
Connection is completed.
How to Insert Batteries
Batteries are not supplied. Use two commercially available alkaline or NiMH AA batteries.
1 Remove the Battery Cover.
(1) Slightly rotate the Battery Cover Disengaging Lever
in the direction of an arrow.
(2) Remove the Battery Cover.
• The Battery Cover may not be removed easily when you try for the first time in order
to prevent water drop from getting inside the Unit.
2 Insert the batteries correctly by aligning the polarities
(+ and –) with the battery indication marks.
1
2
Bottom of the
Main Unit
If you store this unit with the batteries or rechargeable batteries in it, the batteries may leak and damage the unit.
If you are not going to use the unit for a long period of time (more than three months), remove the batteries.
•Do not use worn and new batteries together or batteries of different types (such as regular batteries and
rechargeable batteries) together.
•Follow the disposal method determined by the local community when it comes to disposal of used batteries.
Batteries
Main Unit Locking Board
Note: It will click and be locked to the Stand.
2 Plug in the power plug of the AC Adapter to the electric outlet.
When using NiMH batteries
–You may use NiMH batteries
(rechargeable batteries).
If you use a commercially available battery
charger suitable for the batteries you are
using, you can save money by recharging the
batteries again and again.
The unit can be used for approximately 8 days
(if it is used for 30 minutes a day when the
batteries are fully charged.)
The Battery Low Indicator (orange) flashes.
The rechargeable batteries have no residual
power. If the unit does not nebulize,
immediately recharge the batteries with
a commercially available charger.
The supplied AC Adapter does not function
as a battery charger.
Use a commercially available battery charger
to recharge the batteries.
Battery replacement and battery life
When using alkaline batteries
The unit can be used for
approximately 8 days (if it is
used for 30 minutes a day.)
• The Battery Low Indicator
(orange) flashes.
Batteries are low. Replace both batteries
with new ones.
The Battery Low Indicator (orange) turns on.
Batteries are worn out. As the unit will not
nebulize, immediately replace both batteries
with new ones.
3 Put on the Battery Cover.
Put the Battery Cover on the Battery Case. Push in the Battery Cover
until it catches the claws and clicks.
Push both ends firmly.
You may find it difficult to push it through because the unit is made to prevent water
from getting inside the Unit.
Correct Use of the Unit
How to Fill the Medication Bottle
13
Correct Use of the Unit How to Fill the Medication Bottle
How to Fill the Medication Bottle
12
How to Fill the Medication Bottle Correct Use of the Unit
Have you removed Mask Adapter and the supplied Mouthpiece?
(For how to assemble the Main Unit, refer to Page 8)
1 Remove Medication Bottle
from the Main Unit.
5 Attach the Medication Bottle to the Main Unit.
If the Medication Bottle is not installed
correctly, the power will not turn on.
If the Electrodes on the Main Unit are
severely stained, the unit will not nebulize
normally. In that case, wipe off the stain.
(Refer to Page 17)
3 Fill Medication Bottle.
Be careful to prevent the Mesh Cap from closing
as shown in the Figure.
The maximum capacity of Medication Bottle is 7 ml.
4 Close the mesh cap.
Be sure to check that the Mesh Cap
is engaged. If it is not engaged,
medication will leak.
Click
Press gray tabs on
both side of Medication
Bottle at the same
time lift upwards.
Click
Align both
electrodes with
each other.
Check that the
Mesh Cap is closed.
You have
finished assembly.
You have
finished assembly.
Do not open the
bottle cap.
Lift up the mesh
cap while pressing
your thumb hard
on the brim.
2 Open the Mesh Cap.
Hold the medication bottle as if covering
it with your hand.
How to install the Child’s Inhalation Mask
7 Attach the Child’s Inhalation Mask or the Mouthpiece.
How to install the Mouthpiece
6 Attach the Mouthpiece, Mask Adapter.
You have
finished assembly.
How to Inhale
15
Correct Use of the Unit How to Inhale
How to Select the Nebulization Mode
14
How to Select the Nebulization ModeCorrect Use of the Unit
You can use this unit either in the continuous nebulization mode or in the manual mode.
The power indicator (green light) illuminates during nebulization.
Push the Button and
immediately release
your finger
(within 1 second).
1 Hold the Main unit with your
hand and start inhalation in
relaxed posture.
Slightly slant the Main Unit towards yourself.
• Immerse the Vibrator in the medication. If the
Vibrator is not immersed, the unit will not nebulize.
Medication
Caution
Do not drop or apply strong pressure to Main
Unit, Medication Bottle Unit, or Mesh Cap.
The unit may be damaged and cannot be used.
• If saliva or medication collects on the electrode
of Main Unit or on the Mesh, the unit stops
nebulization. In that case, turn off the power and
absorb the medication or saliva with tissue paper
or a soft cloth.
Keep on pressing
the Button
(for approximately
2 seconds or more).
Manual Nebulization Mode
When you keep pressing the Button (approximately
for 2 seconds or more), the unit nebulizes in Manual
Nebulization Mode. The unit nebulizes only while
you press the Button. You can inhale on demand.
Continuous Nebulization Mode
When you push the Button and immediately release
finger from the Button (within approximately
1 second), the unit nebulizes in Continuous
Nebulization Mode. When you push the Button
again, the unit stops nebulization.
2 When you finish inhalation,
push the Button to turn off the
power.
•The unit has a built-in timer to turn off the
power in approximately 30 minutes after
the power is turned on.
•When an AC Adapter is used, unplug the
power plug from the electric outlet.
(Refer to Page 11)
How to Care for Unit After Use
17
Correct Use of the Unit How to Care for Unit After Use
How to Care for Unit After Use
16
How to Care for Unit After UseCorrect Use of the Unut
Be sure to wash and store the unit after use.
If you do not wash the unit after inhalation, the medication will dry and adhere, and may not be
nebulized. In that case, disinfect Mesh Cap and wash off dried medication.
(Refer to Page 18)
5 After assembling the unit, store it in a clean environment.
Warning
Wash and disinfect Medication Bottle, Inhalation Mask,
Mouthpiece, and Mask Adapter after each use. (Refer to Page 16)
Be sure to dry the cleaned and disinfected parts promptly, and store them
without getting re-contamination.
Disengage the
Medication Bottle
Cap Locking
Lever and open
the Medication
Bottle Cap.
When the main unit is dirty
General Advice
Do not leave or carry the Medication Bottle filled with medication or water.
The medication may drip or spill and damage the Main Unit.
Do not store the unit under high temperature, high humidity,
or in direct sunlight.
Cord band
3 Wash the parts.
Wash the Inhalation Mask or Mouthpiece, Mask Adapter,
and Medication Bottle in the water.
General Advice
Do not wash Medication Bottle and Mesh Cap under strong
running tap water.
It may cause damage.
4 Dry parts thoroughly.
Immediately after washing the parts, wipe off excess water
with a dry cloth, and allow to air dry.
Remove the Mesh Cap from the Medication Bottle
and air dry. (Refer to Page 8)
Do not touch the Mesh with your hand.
Discard the
remaining
medication.
1 Remove the parts from the Main Unit.
Remove the Child’s Inhalation Mask or Mouthpiece, Mask Adapter, and Medication Bottle
from the Main Unit. (For how to assemble the Main Unit, refer to Page 8 to 9)
2 Discard the remaining medication.
Wipe the stain on the Main Unit with a soft cloth. If the Main Unit is heavily stained, wipe it with
a soft cloth moistened with water or detergent and squeezed well. Then wipe it with a dry soft cloth
.
General Advice
Do not clean the Main Unit with volatile chemical such as benzene or thinner.
• Do not immerse the Main Unit in Water.
How to Carry the Unit
•You can put Mouthpiece and Inhalation Mask in the storage case to carry them with the unit.
Put a Main Unit Cover over the Main Unit when carrying the unit.
Do not put the AC Adapter in the storage case along with the Main Unit.
The Main Unit may be scratched.
Bundle the power cord of the AC Adapter and fasten it to the Main Unit of the
AC Adapter with the cord band.
How to Replace the Mesh Cap
19
How to Disinfect the Unit How to Replace the Mesh Cap
How to Disinfect
18
How to Disinfect How to Disinfect the Unit
The Micro-Air
®
nebulizer can be cleaned and disinfected easily because the Medication
Bottle and the Mesh Cap can be removed from the Main Unit. The unit can
be disinfected with benzalkonium chloride.
1 Open the Mesh Cap.
2 Remove the Mesh Cap from the
Medication Bottle.
3 Install a new Mesh Cap.
You can usually use the Mesh Cap for approximately 1 year (according to the test results of OMRON).
However, this may vary depending on medication type and individual use and care. In some cases,
the unit may not nebulize or may nebulize weakly even after disinfection is done.
If the unit does not nebulize or nebulizes poorly even after cleaning the Mesh Cap,
purchase a new Mesh Cap from your nearest Omron dealer or where you purchased your unit from.
The part number is U22-4.
Lift up the Mesh Cap
by strongly applying
your thumb on the brim.
Insert a new
Mesh Cap
vertically.
Close it
securely.
1 Make 0.1% dilute solution using benzalkonium chloride, and soak the
Medication Bottle and Mesh Cap in the solution for 10 minutes.
2 Take the Medication Bottle and Mesh Cap out of solution
to rinse it with water.
3 Wipe off the remaining water for faster drying.
4 If using commercial cleaners, follow manufacturers instructions.
Caution
Do not use disinfection solution containing sodium hypochlorite.
The unit may not be able to be used after getting covered with rust.
5 The following parts can be disinfected:
Medication Bottle
Mesh Cap
Mask
Mouthpiece
Mouthpiece Adapter
Lift up the
Mesh Cap
vertically.
Troubleshooting
21
Troubleshooting and Warranty Troubleshooting
Troubleshooting
20
Troubleshooting Troubleshooting and Warranty
If a problem occurs while you are using the unit, please check the following information before calling.
The nebulization rate is extremely low, or the unit does not nebulize at all.
Where to Inspect How to correct
Is the Battery Low Indicator turned on?
Is the Mesh dirty?
Replace the batteries with new ones.
(Refer to Page 10)
Where to Inspect
Are the batteries worn?
Is the AC Adapter securely plugged into
the electric outlet or the connection port
of Main Unit?
Is the capacity of the rechargeable batteries
(NiMH) low?
Is the Medication Bottle installed to the Main
Unit correctly?
Replace the batteries with new ones.
(Refer to Page 10)
Insert it correctly. (Refer to Page 11)
Recharge the batteries with the commercially
available recharger or replace them
with new ones.
Install it correctly. (Refer to Page 13)
The Power Indicator is illuminated, but the unit does not nebulize.
Where to Inspect How to correct
Is the Medication Bottle filled with medication?
Is the Mesh broken, stained,
or extremely soiled?
Has any medication or water collected on the
electrode of the Main Unit?
Fill with medication. (Refer to Page 12)
Clean the Mesh Cap by disinfecting.
(Refer to Page 18)
If the unit does not nebulize even after
carrying out the above-mentioned procedure,
or the Mesh is broken, replace the Mesh Cap
with a new one.
(Refer to Page 19)
The Power Indicator does not turn on and the unit does not nebulize.
Clean the Mesh Cap by disinfecting to remove
any stains. (Refer to Page 18)
If the trouble is not corrected by the above
mentioned procedure, replace the Mesh cap
with a new one.
(Refer to Page 19)
Are the polarities of batteries or rechargeable
batteries placed in the wrong direction?
Insert batteries or rechargeable batteries in the
correct alignment.
(Refer to Page 10)
How to correct
Wipe off the stain. (Refer to Page 17)
Is the medication being supplied to the
Medication Bottle?
(Is the medication in contact with the Vibrator?)
Use the unit slightly slanting the Main Unit
(the side with the Button) towards yourself.
(Refer to Page 15)
Absorb the medication or water with
a soft cloth.
(Refer to Page 15)
Is the electrode of the Main Unit stained?
Warranty
23
Troubleshooting and Warranty Warranty
Specifications
22
Troubleshooting and Warranty
Specifications
Replacement Parts and Accessories
9921
Child’s Inhalation Mask, with rubber string
(Mask and string optional accessories)
U22-3
Medication Bottle
U22-5
AC Adapter (optional accessories)
U22-7
Storage case
U22-4
Mesh Cap
U22-2
Mask Adapter
U22-1
Mouthpiece
For the inquiry on where to purchase the replacement parts and accessories, contact the store where you purchased
the unit or the nearest OMRON dealer.
Your NE-U22V Omron Nebulizer is warranted to be free from defects in material and workmanship for
a period of 2 years from the date of purchase. This warranty extends only to the original retail
purchaser. This warranty does not cover failures resulting from misuse, abuse, alteration of the Omron
Nebulizer System, use of improper components or failure to comply with the operating instructions.
The 2 year warranty excludes mesh cap and accessories.
To make a warranty claim, return your Omron Nebulizer System, freight prepaid and insured
(or assume the risk of loss or damage during shipment) to Omron Healthcare, Inc., 300 Lak
eview
Parkway, Vernon Hills, IL 60061,
Attention: Service Department. All requests for warranty service
should include the model number of your Omron Nebulizer System, proof of purchase (including
date of purchase), your phone number and the nature or your service request. Omron will either repair
or at its option replace free of charge any parts necessary to correct defects in material or workmanship
covered by this limited warranty.
This limited warranty is in lieu of all other warranties, whether oral, written, express or implied.
There are no warranties of merchantability or fitness for a particular purpose, except for those
implied warranties arising by operation of law which are limited in duration to the 2 years term of this
warranty. Omron’s liability, whether based on contract, tort, warranty, strict liability, or any other
theory, shall not exceed the price of the individual Omron nebulizer system which is the basis of
the claim. In no event shall Omron be liable for any loss of profits, facilities, equipment or other
incidental, consequential or special damages.
Some states do not allow limitations on how long an implied warranty lasts, or allow the exclusion
or limitations on how long an implied warranty lasts, or allow the exclusion or limitation of incidental,
consequential, or special damages, so the above limitations or exclusions may not apply to you.
This warranty gives you specific legal rights. You may also have other rights
which vary from state to state.
For Customer Service
Call Toll Free: 1-800-634-4350
Product name: OMRON Micro-Air
®
Vibrating Mesh Nebulizer
Model: NE-U22V
Power supply: 110 VAC and 50/60 Hz (With the use of an AC Adapter)
3 VDC (two AA alkaline batteries)
2.4 VDC (two AA NiMH batteries)
Power consumption: Approx. 1.5 W*
Vibration frequency: Approx. 180 kHz
Nebulization rate: 0.4 to 0.5 mL/min average*
Particle size: MMAD approx. 5µm*
(1 to 5 µm for 80% of the entire volume particle distribution)
(Measured by OMRON based on CEN/TC215)
Capacity of
Medication Bottle:
Maximum of 7 mL
External dimensions: 38 (W) x 104 (H) x 51 (D) mm
Weight: Approx. 97 g (not including batteries)
Battery life: Approx. 4 hours with continuous operation*
Approx. 8 days when used about 30 minutes a day
(with the use of alkaline batteries)
Operating temperature
and humidity:
10 to 40°C (No ice formation is allowed.) 30 to 85% RH
Accessories: Child’s Inhalation Mask, Mouthpiece, Mask Adapter,
Storage case, AC Adapter, Instruction Manual
* When physiological saline is used. These figures may vary depending on the type of medication.
Specifications and appearance may be subject to change without prior notice for improvement.
Specifications / Replacement Parts and Accessories
Distributed by:
OMRON HEALTHCARE, INC.
300 Lakeview Parkway
Vernon Hills, Illinois 60061
Toll Free Customer Service: 1-800-634-4350
www.omronhealthcare.com
Copyright © 2002 Omron Healthcare, Inc.
NEU22VINST Rev
A1
GUIDE D’UTILISATION
Merci d’avoir acheté le Nébuliseur
d’OMRON
®
. Avant d’utiliser
cet appareil pour la première fois, veuillez
lire attentivement ce guide d’utilisation.
Veuillez garder ce guide d’utilisation
à portée de la main pour référence future.
NÉBULISEUR À
FILTRE VIBRANT
Modèle NE-U22V
®
¤
Cet appareil est un dispositif médical. Utilisez cet appareil uniquement de la façon indiquée par
votre médecin et/ou votre inhalothérapeute.
Les illustrations présentées dans ce guide d’utilisation sont des dessins.
Mise en garde: La loi fédérale soumet cet appareil
à une restriction de vente par ou sur ordonnance
d’un médecin.
Avant d’utiliser l’appareil
Conseils de sécurité . . . . . . . . . . . . . . . . . . . . . 2
Caractéristiques du produit . . . . . . . . . . . . . . . . 5
Composants du produit . . . . . . . . . . . . . . . . . . . 6
Description des pièces . . . . . . . . . . . . . . . . . . . 7
Assemblage de l’appareil principal. . . . . . . . . . 8
Utilisation de l’appareil
Utilisation de la source d’énergie . . . . . . . . . . 10
Comment remplir bouteille de médicament . . 12
Sélection du mode de nébulisation . . . . . . . . . 14
Inhalation . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Entretien après utilisation . . . . . . . . . . . . . . . . 16
Désinfection de l’appareil
Désinfection . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Remplacement de la capsule-filtre . . . . . . . . . 19
Dépannage et garantie
Dépannage . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Spécifications, pièces de remplacement
et accessoires. . . . . . . . . . . . . . . . . . . . . . . . . . 22
Garantie . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
¤
Français
Conseils de sécurité
2
Conseils de sécurité
3
Avant d’utiliser l’appareil
Conseils de sécurité
Conseils de sécurité
Avant d’utiliser l’appareil
Les indicateurs d’avertissement et les exemples d’icônes sont présentés pour vous
aider à utiliser le produit correctement et de façon sécuritaire pour prévenir les
risques de vous blesser ou de blesser d’autres personnes.
Signification
Indique un point où la possibilité de décès ou de blessures sérieuses
peut se produire à la suite d’une manipulation incorrecte.
Exemples d’icônes
En ce qui concerne le type, la dose et le schéma posologique du médicament, assurez-vous de suivre
les directives d’un médecin.
•Vos symptômes peuvent s’aggraver.
Lorsque vous utilisez l’appareil pour la première fois après l’avoir acheté, ou après une longue période
sans l’utiliser, assurez-vous de laver et de désinfecter la coupelle à médicament et la capsule-filtre
(référez aux pages 14 à 18).
•Diverses bactéries peuvent être présentes et vous pourriez être contaminé.
Lavez et désinfectez la capsule-filtre, le masque d’inhalation, l’embout buccal et l’adaptateur
de masque après chaque usage (référez à la page 18).
Assurez-vous d’assécher rapidement les pièces nettoyées et désinfectées, et de les ranger
de façon qu’elles ne soient pas contaminées à nouveau.
•Diverses bactéries peuvent être présentes et vous pourriez être contaminé.
Si plus d’une personne utilisent l’appareil, assurez-vous d’utiliser la coupelle à médicament, le masque
d’inhalation pour enfant, l’embout buccal et l’adaptateur buccal désinfectés pour chaque utilisateur.
•L’utilisateur pourrait être contaminé.
Gardez l’appareil hors de portée des bébés et des enfants.
Ils peuvent avaler les petites pièces telle que la capsule-filtre. L’appareil peut être endommagé ou peut
occasionner des blessures aux bébés et aux enfants. Dans le cas où un enfant aurait avalé la capsule-filtre,
consultez immédiatement un médecin.
Ne branchez ou ne débranchez pas l’adaptateur CA si vous avez les mains mouillées.
- Vous pourriez subir un choc électrique.
Ne lavez pas l’appareil principal et l’adaptateur CA avec de l’eau ou de l’eau chaude.
Évitez de les plonger dans l’eau.
•Vous pourriez endommager l’appareil.
Indicateur de
mise en garde
L’icône indique une interdiction (vous ne pouvez pas faire).
Les sujets impliquant de réelles interdictions sont indiquées par
un texte ou des illustrations dans ou près de .
L’icône de gauche réfère à « l’interdiction de désassembler ».
L’icône indique une interdiction (incluant mise en garde et
avertissement). Les sujets impliquant de réelles interdictions sont
indiquées par un texte ou des illustrations dans ou près de .
L’icône de gauche réfère à « avertissement de choc électrique ».
Indique un point où des lésions corporelles ou des dommages matériels*
peuvent se produire à la suite d’une manipulation incorrecte.
*Dommages matériels réfère à une vaste gamme de dommages impliquant votre
maison, vos biens, les animaux domestiques et les animaux de compagnie.
L’icône indique un point obligatoire (vous devez toujours
respecter). Les sujets impliquant de réels points obligatoires sont
indiqués par un texte ou des illustrations dans ou près de .
L’icône de gauche réfère à une « renseignement important ».
Ne pas désassembler, réparer ou modifier cet appareil. Cela pourrait
annuler la garantie.
Si vous n’utilisez pas l’appareil durant une longue période, assurez-vous
de le débrancher.
N’utilisez pas de solution de décontamination contenant de l’hypochlorite de sodium.
La capsule-filtre rouillerait.
Ne pas renverser de médicament sur l’appareil principal ou sur l’adaptateur c.a.
Si un médicament est renversé sur l’appareil principal ou sur l’adaptateur c.a.,
l’essuyer immédiatement avec un chiffon doux.
N’échappez pas ou ne soumettez pas l’appareil principal, l’adaptateur CA,
la coupelle à médicament ou la capsule-filtre à un choc violent.
Utiliser seulement l’adaptateur Omron c.a. fourni avec ce produit.
S’assurer que l’adaptateur c.a. est branché dans une prise de courant de 110 V c.a.
Après avoir désinfecté les pièces, rincez à fond le désinfectant avec de l’eau.
MISE EN
GARDE
ATTENTION
MISE EN GARDE
ATTENTION
MISE EN GARDE
Avant d’utiliser l’appareil
Caractéristiques du produit
Caractéristiques du produit
5
Conseils de sécurité
4
Conseils de sécurité
Avant d’utiliser l’appareil
Conseils généraux
(Ceci est une représentation de l’appareil doté du masque d’inhalation)
Remplacez immédiatement les piles à plat par des neuves.
N’utilisez pas différents types de piles ensemble.
N’utilisez pas des piles neuves et usagées ensemble.
N’utilisez pas des piles sèches et des piles rechargeables ensemble.
Pour que l’appareil fonctionne adéquatement, assurez-vous d’insérer les piles correctement.
Vous pouvez utiliser les batteries rechargeables disponibles dans le commerce de NiMH,
mais n'utilisez pas des batteries de cadmium de nickel.
Rechargez les piles rechargeables à l’aide d’un chargeur disponible en magasin.
(L’adaptateur CA inclus dans l’emballage ne peut être utilisé pour charger une pile.)
Lavez bien la capsule-filtre et le dispositif de vibration après chaque usage pour qu’aucun
résidu de médicament n’y soit laissé (référez à la page 16).
Lavez la coupelle à médicament, la capsule-filtre et le dispositif de vibration avec de l’eau
courante propre ou de l’eau distillée disponible en magasin.
Lorsque vous rincez la coupelle à médicament sous l’eau du robinet, n’utilisez pas l’eau
provenant d’un robinet ouvert à pleine capacité; utilisez plutôt un léger filet d’eau.
Rassemblez les pièces nettoyées et désinfectées après qu’elles aient suffisamment séché.
N’utilisez pas de benzène, de diluant ou d’essence pour nettoyer l’appareil.
Ne faites pas sécher l’appareil principal, l’adaptateur CA, la coupelle de médicament
et la capsule-filtre au micro-ondes; ne les nettoyez pas par ébullition.
Ne rangez pas l’appareil dans un endroit où il subira des niveaux de température ou d’humidité
élevés, ou dans un ensoleillement direct.
Ne laissez pas de médicament ou d’eau dans la coupelle à médicament, et ne la transportez
pas alors qu’elle en contient.
Si vous n’utilisez pas l’appareil durant une longue période (trois mois ou plus),
retirez les piles de l’appareil.
Avant de commencer l’inhalation, vous devez
assembler l’appareil.
Avant d’utiliser l’appareil, veuillez lire la section concernant
l’ « Assemblage de l’appareil principal »; assurez-vous de
comprendre les étapes d’assemblage (référez-vous aux pages 8 et 9).
1
L’appareil de format réduit peut facilement être transpor
partout où vous allez
L’appareil est assez petit pour être placé dans un étui de rangement ou un sac à main;
vous pouvez facilement le transporter partout où vous allez. Le nouveau filtre en alliage
de titane crée un aérosol à rendement élevé.
2
Inhalation possible à divers angles
La nouvelle structure de la bouteille permet l’inhalation à divers angles, avec une
nébulisation stable. Ainsi, l’appareil peut même être utilisé au lit.
3
Entretien et utilisation faciles
L’appareil ne requiert aucun soin compliqué. Tout comme le nébuliseur traditionnel
à compresseur, tout ce que vous avez à faire est d’enlever le médicament après inhalation
et d’assécher la coupelle immédiatement après l’avoir lavée.
Avant d’utiliser l’appareil
Description des pièces
Description des pièces
7
Composants du produit
(Veuillez vérifier sur le produit)
6
Composants du produit
Avant d’utiliser l’appareil
Couvercle de
l’appareil principal
Protège l’appareil principal
Couvercle de
l’appareil principal
Bouton de Puissance
Unité principale de tours
marche-arrêt.
Capsule-filtre
Un filtre en alliage est installé
pour la nébulisation.
- Manipulez le filtre avec soin car
il est très délicat.
Levier de blocage du bouchon de la
coupelle à médicament
Ouvre la capsule de médicament pour le lavage.
- N'ouvrez pas la capsule de médicament
excepté le lavage.
Coupelle à médicament
Remplissez médicament en ouvrant
le dessus de la capsule.
Les items suivants sont contenus dans la boîte.
S’il manque un de ces items, veuillez communiquer immédiatement
avec le magasin où vous avez acheté le produit, ou le marchand
OMRON le plus près.
Appareil
principal
Coupelle
à médicament
Capsule-filtre
(avec filtre)
Étui de rangement
Fiche
d’alimentation
Verrouillage pour appareil principal
Adaptateur CA pour approvisionnement standard
Bande pour cordon
Base de connexion pour adaptateur CA
Indicateur de
décharge de pile
Un indicateur orange
clignote lorsque les piles
sont à plat.
Électrode
Alimente la puissance de l’appareil
principal au dispositif de vibration
de la coupelle à médicament.
Orifice d’injection du médicament
Introduisez le médicament en ouvrant
la capsule-filtre.
Dessous de l’appareil principal
Voyant de mise en circuit
Un voyant vert s’allume lorsque
l’appareil est en marche.
Levier de blocage
du couvercle à pile
Prise pour adaptateur CA
sous l’appareil principal
Adaptateur
de masque
®
Embout buccal
Masque pour enfant
(en option) 9921
Adaptateur CA (en option)
Guide d’utilisation
Vidéo d’instruction
Assemblage de l’appareil principal
9
Avant d’utiliser l’appareil Assemblage de l’appareil principal
Assemblage de l’appareil principal
8
Assemblage de l’appareil principal
Avant d’utiliser l’appareil
Cet appareil doit être assemblé avant d’être utilisé.
- Les pièces peuvent s’emboîter de façon assez serré puisqu’elles sont conçues pour
empêcher le médicament de couler.
- Vous devez laver et désinfecter toutes les pièces amovibles avant d’utiliser l’appareil.
Assemblez l’appareil après en avoir lavé, désinfecté et asséché les composants.
(Pour connaître la façon de procéder pour désinfecter l’appareil, référez-vous à la page 18)
2 Fixez la coupelle à médicament à l’appareil principal
3 Fixez l’adaptateur de masque à l’appareil principal
Si vous n’utilisez pas l’appareil après l’avoir
assemblé, désassemblez-le (tel qu’indiqué ci-contre),
et rangez-le dans l’étui de rangement fourni ou dans
un environnement propre. (Pour en savoir davantage
sur l’utilisation adéquate de l’appareil, référez-vous
à la page 15)
Insérez
verticalement
la capsule-filtre
Alignez les
électrodes
l’une sur
l’autre
N’ouvrez pas
le bouchon de
la coupelle
à médicament
Vous avez terminé
l’assemblage
Vous avez terminé
l’assemblage
Vous avez terminé
l’assemblage
Vous avez terminé
l’assemblage
Vous avez maintenant complété l’assemblage de l’appareil principal.
Conseils généraux
•Tenez bien l’appareil avec vos deux mains.
Assemblez solidement les pièces (selon les pièces, il se peut que vous entendiez un petit
bruit sec au moment de l’assemblage).
1 Fixez la capsule-filtre sur la coupelle à médicament
Vous avez terminé
l’assemblage
4 Fixez l’embout buccal fourni ou le masque d’inhalation pour enfant
à l’adaptateur de masque.
Fixation du masque d’inhalation pour enfant Fixation de l’embout buccal
Fermez-la bien
Utilisation de la source d’énergie
11
Utilisation de l’appareil Utilisation de la source d’énergie
Utilisation de la source d’énergie
10
Utilisation de la source d’énergie
Utilisation facultative de l’adaptateur CA (en option)
Utilisez l’adaptateur CA aux endroit où vous disposez d’une prise électrique.
Notez cependant que cet adaptateur CA n’est pas un chargeur de piles.
1 Placez l’appareil principal sur la base de connexion pour adaptateur CA,
tel qu’indiqué ci-après.
Enlèvement de l’adaptateur CA de l’appareil principal
(1) Débranchez le cordon d’alimentation de la prise électrique.
(2) verrouillez l’appareil principal en poussant
sur les deux extrémités du dispositif de
verrouillage de la base de connexion pour
adaptateur CA.
(3) Enlevez l’appareil principal.
(1) Débranchez
(3) Enlevez
(2) Poussez sur les deux extrémités
La connexion est complète.
Installation des piles
Les piles ne sont pas fournies. Utilisez deux piles alcalines ou au NiMH
AA disponibles en magasin.
1 Enlevez le couvercle de pile.
(1) Tournez légèrement le levier de blocage du couvercle
de pile dans la direction de la flèche.
(2) Enlevez le couvercle de pile.
• La première fois, il se peut que le couvercle de pile soit difficile à enlever puisqu’il
est serré afin d’empêcher l’eau de pénétrer à l’intérieur de l’appareil.
2 Insérez correctement les piles en alignant les polarités
(+ et -) sur les marques d’indication.
1
2
Dessous de
l’appareil
principal
Si vous rangez l’appareil alors que les piles ou les piles rechargeables sont en place, elles peuvent couler
et endommager l’appareil.
Si vous n’utilisez pas l’appareil durant une longue période (plus de trois mois), enlevez les piles.
N’utilisez pas des piles neuves et usées, ou des piles de différents types (telles que piles régulières et piles
rechargeables), ensemble.
Respectez les méthodes d’élimination déterminées par votre communauté locale lorsque vous devez disposer
des piles usagées.
Piles
Verrouillage pour appareil principal
Notez: cependant que cet adaptateur CA
n’est pas un chargeur de piles.
2 Branchez le cordon d’alimentation de l’adaptateur CA dans la prise électrique.
Piles au NiMH
–Vous pouvez utiliser des piles au NiMH
(piles rechargeables).
Si vous utilisez un chargeur de piles
convenant aux piles que vous utilisez
(disponible en magasin), vous épargnerez de
l’argent en rechargeant régulièrement les piles.
L’appareil peut être utilisé durant environ 8 jours
(à raison de 30 minutes par jour lorsque les piles
sont complètement chargées).
•L’indicateur de décharge de pile (orange) clignote.
Les piles rechargeables n’ont pas de puissance
résiduelle. Si l'unité pas nebulize, rechargez
immédiatement les batteries avec un chargeur
disponible dans le commerce.
•L’adaptateur CA fourni ne peut servir à charger
les piles.Utilisez un chargeur de piles disponible
en magasin pour recharger les piles.
Remplacement et durée de vie des piles
Piles alcalines
L’appareil peut être utilisé durant
environ 8 jours (à raison de 30
minutes par jour)
• L’indicateur de décharge
de pile (orange) clignote.
Les piles sont presque déchargées.
Remplacez les deux piles par des neuves.
•L’indicateur de décharge de pile (orange) s’allume.
Les piles sont à plat. L’appareil ne
produira aucune nébulisation.
Remplacez immédiatement les
deuxpiles par des neuves.
3 Replacez le couvercle de pile.
Placez le couvercle de pile sur le compartiment à piles. Poussez le
couvercle jusqu’à ce qu’il s’adapte aux pattes et produise un petit bruit sec.
Poussez fermement sur les deux extrémités.
Il se peut que vous trouviez difficile de le pousser jusqu’au bout puisque l’appareil a été
conçu pour empêcher l’eau d’atteindre l’intérieur.
Utilisation de l’appareil
Comment remplir bouteille de médicament
13
Utilisation de l’appareil Comment remplir bouteille de médicament
Comment remplir bouteille de
médicament
12
Comment remplir bouteille de médicamentUtilisation de l’appareil
Avez-vous enlevé l’adaptateur de masque et l’embout buccal fourni?
(Pour les directives sur l'assemblage de l'appareil principal, référez-vous à la page 8)
1 Enlevez la coupelle à médicament
de l’appareil principal.
5 Fixez la coupelle à médicament à l’appareil principal.
Si la coupelle à médicament n’est pas correctement installée,
l’appareil ne se mettra pas en marche.
Si les électrodes de l’appareil principal
sont très sales, l’appareil ne pourra
produire une nébulisation normale.
Dans ce cas, nettoyez les saletés.
(Référez-vous à la page 17)
3 Remplissez la coupelle à médicament
Tel qu’indiqué à la Figure ci-contre, empêchez
la capsule-filtre de se refermer lorsque
vous remplissez la coupelle à médicament
La capacité maximale de la coupelle
à médicament est de 7 ml.
4 Fermez la capsule-filtre.
Assurez-vous de vérifier que la capsule-filtre
soit bien fermée. Dans le cas contraire,
le médicament s’écoulera.
Appuyez sur les leviers
de blocage gris des deux
côtés de la coupelle
à médicament tout
en la soulevant
Alignez les
électrodes l’une
sur l’autre.
Vérifiez que la
capsule-filtre soit
bien fermée.
Vous avez terminé
l’assemblage
Vous avez terminé
l’assemblage
N’ouvrez pas le bouchon
de la coupelle à médicament
Soulevez la
capsule-filtre tout
en appuyant
fermement sur
l’arête du rebord
avec votre pouce
2 Ouvrez la capsule-filtre Retenez
la coupelle à médicament en
la couvrant avec votre main
.
Installation du masque d’inhalation pour enfant
7 Fixez la masque d’inhalation pour enfant ou l’embout buccal.
Installation de l’embout buccal
6 Fixez l’embout buccal et l’adaptateur de masque
Vous avez terminé
l’assemblage
Inhalation
15
Utilisation de l’appareil Inhalation
Sélection du mode de nébulisation
14
Sélection du mode de nébulisation Utilisation de l’appareil
Vous pouvez utilisez l’appareil en mode de nébulisation continue ou en mode manuel.
Le voyant de mise en circuit (vert) s’allume durant la nébulisation.
Appuyez sur la touche et
dégagement immédiatement
votre doigt (dans la seconde)
1 Tenez l’appareil principal d’une
main et commencez l’inhalation
dans une position détendue.
Inclinez légèrement l’appareil principal vers vous.
vous (la touche vous faisant face).
• Immergez le dispositif de vibration dans le médicament.
Si le dispositif de vibration n’est pas immergé, l’appareil
ne produit pas de nébulisation.
Médicament
Avertissement
N’échappez pas ou ne soumettez pas l’appareil
principal, la coupelle à médicament ou la capsule-filtre
à un choc violent.
•L’appareil peut être endommagé et ne peut être utilisé.
Essuyez le médicament
ou la salive.
• Si de la salive ou du médicament se dépose sur
l’électrode de l’appareil principal ou sur le filtre,
l’appareil arrête la nébulisation. Le cas échéant,
éteindre l’appareil et absorber le médicament ou la
salive avec un mouchoir en papier ou un chiffon doux.
Continuez d’appuyer sur
la touche (durant environ
2 secondes ou plus)
Nébulisation en mode manuel
Lorsque vous continuez d’appuyer sur la touche
(durant 2 secondes ou plus), l’appareil produit une
nébulisation en mode manuel. L’appareil produit une
nébulisation seulement pendant que vous appuyez sur
la touche. Vous pouvez inhaler à votre convenance.
Nébulisation en mode continu
Lorsque vous appuyez sur la touche et enlevez
immédiatement votre doigt de la touche (après
environ 1 seconde), l’appareil produit une
nébulisation en mode continu.
Lorsque vous appuyez de nouveau sur la touche,
l’appareil arrête la nébulisation.
2 Lorsque vous avez terminé
l’inhalation, appuyez sur la
touche et arrêtez l’appareil.
•L’appareil est doté d’une minuterie intégrée qui
arrête l’appareil environ 30 minutes après qu’il
ait été mis en circuit.
•Lorsqu’on utilise l’adaptateur CA, débrancher
le cordon d’alimentation de la prise électrique.
(Référez-vous à la page 11)
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37

Omron NE-U22V Manuel utilisateur

Taper
Manuel utilisateur
Ce manuel convient également à

dans d''autres langues