Amprobe IR-750 Infrared Thermometer Manuel utilisateur

Catégorie
Mesure
Taper
Manuel utilisateur

Ce manuel convient également à

IR-750
IR-750-EUR
50:1 Infrared
Thermometer
Users Manual
IR-750
IR-750-EUR
50:1 Infrared Thermometer with memory
Users Manual
4/2017, 4279850 D
©2017 Amprobe Test Tools.
All rights reserved. Printed in China
English
Limited Warranty and Limitation of Liability
Your Amprobe product will be free from defects in material and workmanship for one year from the date of
purchase, unless local laws require otherwise. This warranty does not cover fuses, disposable batteries or damage
from accident, neglect, misuse, alteration, contamination, or abnormal conditions of operation or handling.
Resellers are not authorized to extend any other warranty on Amprobe’s behalf. To obtain service during the
warranty period, return the product with proof of purchase to an authorized Amprobe Test Tools Service Center
or to an Amprobe dealer or distributor. See Repair Section for details. THIS WARRANTY IS YOUR ONLY REMEDY.
ALL OTHER WARRANTIES - WHETHER EXPRESS, IMPLIED OR STATUTORY - INCLUDING IMPLIED WARRANTIES OF
FITNESS FOR A PARTICULAR PURPOSE OR MERCHANTABILITY, ARE HEREBY DISCLAIMED. MANUFACTURER SHALL
NOT BE LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES, ARISING
FROM ANY CAUSE OR THEORY. Since some states or countries do not allow the exclusion or limitation of an
implied warranty or of incidental or consequential damages, this limitation of liability may not apply to you.
Repair
All test tools returned for warranty or non-warranty repair or for calibration should be accompanied by the
following: your name, company’s name, address, telephone number, and proof of purchase. Additionally, please
include a brief description of the problem or the service requested and include the test leads with the meter.
Non-warranty repair or replacement charges should be remitted in the form of a check, a money order, credit
card with expiration date, or a purchase order made payable to Amprobe.
In-Warranty Repairs and Replacement – All Countries
Please read the warranty statement and check your battery before requesting repair. During the warranty
period any defective test tool can be returned to your Amprobe distributor for an exchange for the same or
like product. Please check the “Where to Buy” section on www.amprobe.com for a list of distributors near you.
Additionally, in the United States and Canada In-Warranty repair and replacement units can also be sent to a
Amprobe® Test Tools Service Center.
Non-Warranty Repairs and Replacement – US and Canada
Non-warranty repairs in the United States and Canada should be sent to a Amprobe Service Center. Call
Amprobe or inquire at your point of purchase for current repair and replacement rates.
In USA In Canada
Amprobe Test Tools Amprobe Test Tools
Everett, WA 98203 Mississauga, ON L4Z 1X9
Tel: 877-AMPROBE (267-7623) Tel: 905-890-7600
Non-Warranty Repairs and Replacement – Europe
European non-warranty units can be replaced by your Amprobe distributor for a nominal charge. Please check
the “Where to Buy” section on www.amprobe.com for a list of distributors near you.
European Correspondence Address*
Amprobe® Test Tools Europe
Beha-Amprobe GmbH
In den Engematten 14
79286 Glottertal, Germany
Tel.: +49 (0) 7684 8009 - 0
www.amprobe.eu
*(Correspondence only – no repair or replacement available from this address. European customers please
contact your distributor.)
1
IR-750 / IR-750-EUR 50:1 Infrared Thermometer with memory
CONTENTS
SYMBOLS ...................................................................................................................................................................... 3
SAFETY INFORMATION ................................................................................................................................................ 3
UNPACKING AND INSPECTION .................................................................................................................................... 4
FEATURES ...................................................................................................................................................................... 4
HOW THE THERMOMETERS WORK ............................................................................................................................ 4
OPERATING THE THERMOMETER ................................................................................................................................ 4
Temperature Measurement .................................................................................................................................... 4
Locating a Hot or Cold Spot ................................................................................................................................... 8
Distance and Spot Size ............................................................................................................................................ 8
Field of View.............................................................................................................................................................8
Emissivity .................................................................................................................................................................. 9
Reminders ................................................................................................................................................................ 9
TYPICAL MEASUREMENTS .......................................................................................................................................... 10
OPERATING IR-750 / IR-750-EUR SOFTWARE ............................................................................................................. 13
DETAILED SPECIFICATIONS ......................................................................................................................................... 17
MAINTENANCE ............................................................................................................................................................. 18
TROUBLE SHOOTING .................................................................................................................................................... 19
BATTERY REPLACEMENT .............................................................................................................................................19
2
IR-750 / IR-750-EUR 50:1 Infrared Thermometer with memory
1
Laser aperture
2
Trigger
3
Battery cover
4
Display
5
USB connector
6
K-type thermocouple connector
7
Power adaptor connector (for optional accessory)
8
Laser “ON” symbol
9
Display backlight
10
Measurement lock (continuous measurement)
11
Temperature unit (Celsius / Fahrenheit)
12
8 seconds auto display hold
13
Making measurement (pulling the trigger)
14
Built in memory up to 99 points
15
Programmable high and low alarm
16
Low battery indicator
17
Primary display
18
Secondary display
19
Emissivity (adjustable from 0.10 to 1.00)
20
MAX, MIN, DIF, AVG temperature values
HAL, LAL temperature setting
DATA point selection
T-C K-type thermocouple measurement
1
5
6
7
4
89
2 3
17
12
1613
10
11
15
18
19
14
20
3
SYMBOLS
Caution! Refer to the explanation in this Manual.
Warning! Laser light. Do not stare into laser beam.
°C
Celsius.
°F
Fahrenheit.
Battery indication.
Complies with European Directives.
Do not dispose of this product as unsorted municipal waste.
Contact a qualified recycler.
SAFETY INFORMATION
The instrument complies with:
EN 61010-1 General Safety
EN 60825-1 Laser Safety
EN 61326-1 Electromagnetic Emissions and Susceptibility
Warning
• Do not stare into laser beam.
• Do not point laser directly at eye or indirectly off reflective surfaces.
• For use by competent persons only.
Replace the batteries as soon as the low-battery indicator appears.
• Do not use the thermometer if it operates abnormally.
Do not operate the thermometer around explosive gas, vapor, or dust.
• To avoid a burn hazard or fire, know that reflective objects may be much hotter than the indicated
temperature reading.
Do not leave the thermometer on or near objects of high temperature.
If the thermometer is used in a manner not specified by this manual, the protection provided by the
thermometer may be impaired or may result in hazardous laser radiation exposure.
Cautions
To avoid damaging the thermometer under measurement, protect them from the following:
EMF (electro-magnetic fields) from arc welders, induction heaters
• Static electricity
Thermal shock (caused by large or abrupt ambient temperature changes — allow 30 minutes for instrument to
stabilize before use)
Do not leave the thermometer on or near objects of high temperature
4
UNPACKING AND INSPECTION
Your shipping carton should include:
1 Thermometer (IR-750 / IR-750-EUR)
1 USB cable
1 K-type thermocouple probe
1 Carrying case
1 Hard carrying case
1 9V battery (installed)
1 Users manual
If any of the items are damaged or missing, return the complete package to the place of purchase for an
exchange.
FEATURES
The Amprobe IR-750 / IR-750-EUR, a precision performance 50:1 spot to distance ratio infrared thermometer,
offers unparalleled accuracy and response time with a temperature measurement range of -58°F to 2822°F
or -50°C to 1550°C. The IR-750 / IR-750-EUR is perfect for demanding quality and process control applications
with extremely high accuracy and distance to spot ratio. The IR-750 / IR-750-EUR also features 99 data points
recognition, thermocouple inputs, and USB download for advanced HVAC/R, electrical, industrial maintenance,
automotive as well as quality control and fire prevention applications.
• 50:1 Distance to Spot ratio
• Temperature range of 0°F to -58°F to 2822°F or -50°C to 1550°C
• Precision accuracy and rapid response time
• Laser pointer, backlit dual LCD Display
• Auto display hold and MAX/MIN memory
• Adjustable emissivity for measuring a variety of materials
• 99 Memory locations and PC download cable included
HOW THE THERMOMETERS WORK
Infrared thermometers measure the surface temperature of an object. The thermometer’s optics sense emitted,
reflected, and transmitted energy, which is collected and focused onto a detector. The unit’s electronics translate
the signal into a temperature reading which the unit displays .
OPERATING THE THERMOMETER
Temperature Measurement
The thermometer will turn on when you pull the trigger and also features an auto-off function that
automatically powers down the thermometer after 8 seconds of inactivity.
To measure temperature, point the thermometer at an object and pull the trigger. You can use the laser pointer
to help aim the thermometer. Pull and hold the trigger when measuring the target surface. When release the
trigger, the display will hold the reading for 8 seconds. Be sure to consider distance-to-spot size ratio and field of
view. The laser is used for aiming only and is not related to temperature measurement.
5
Rotary Switch Positions
Button Description
YELLOW
button
Press YELLOW button to toggle between MAX, MIN, DIF, AVG, HAL, LAL, DATA and
T-C options. When the Thermometer goes into sleep mode, press MODE to turn the
thermometer ON again and it displays the last measurement result.
SET
Press to enter set-up mode stepping through Emissivity, Trigger Lock and Switching °C/°F
set-up.
Details refer to the below Emissivity, Trigger Lock and °C / °F set-up.
Press
to turn the display backlight ON or OFF.
Selectable two level of display backlight to adapt different lighting conditions.
Press to turn the laser light ON or OFF.
symbol on the display Indicates laser light is ON.
/
When the thermometer enters the setup up mode (SET), press or to select a set-up
option (Emissivity, Trigger lock, Switching °C / °F).
When the thermometer enters HAL, LAL and DATA mode, press
or to select a set-up
option (Emissivity, Trigger lock, Switching °C / °F).
Laser
Caution
To avoid injury, do not point the laser directly at eye or indirectly off reflective surfaces.
The thermometer is equipped with a laser used for aiming purposes only. The laser turns off when the trigger
is released.
To enable or disable the laser:
1. Press
button to enable or disable the laser. Symbol appears on the display when laser is enabled.
Emissivity set-up
1. Press SET button to select Emissivity set-up, icon is blinking on the display
2. Press
to increase the value by 0.01. Press and hold for quick setting. The maximum value is 1.00.
3. Press
to decrease the value by 0.01. Press and hold for quick setting. The minimum value is 0.01.
4. Press MODE button to complete the setting and exit Emissivity set-up, or press SET button to complete the setting
and continue setting for Trigger Lock.
Note: Default emissivity is 0.95.
6
Table of Surface Emissivity
Measure Surface Switch Setting
METALS
Aluminum
Oxidized 0.2 – 0.4
Alloy A3003
Oxidized 0.3
Roughened 0.1 – 0.3
Brass
Burnished 0.3
Oxidized 0.5
Cooper
Oxidized 0.4 – 0.8
Electrical Terminal Blocks 0.6
Haynes
Alloy 0.3 – 0.8
Inconel
Oxidized 0.7 – 0.95
Sandblasted 0.3 – 0.6
Electoropolished 0.15
Iron
Oxidized 0.5 – 0.9
Rusted 0.5 – 0.7
Iron Cast
Oxidized 0.6 – 0.95
Unoxidized 0.2
Molten 0.2 – 0.3
Iron Wrought
Dull 0.9
Lead
Rough 0.4
Oxidized 0.2 – 0.6
Molydbenum
Oxidized 0.2 – 0.6
Nickel
Oxidized 0.2 – 0.5
Platinum
Black 0.9
Steel
Cold-Rolled 0.7 – 0.9
Ground Sheet 0.4 – 0.6
Polished Sheet 0.1
Zinc
Oxidized 0.1
Measure Surface Switch Setting
NON-METALS
Asbestos 0.95
Asphalt 0.95
Basalt 0.7
Carbon
Unoxidized 0.8 – 0.9
Graphite 0.7 – 0.8
Carborundum 0.9
Ceramic 0.95
Clay 0.95
Concrete 0.95
Cloth 0.95
Glass
Plate 0.85
Gravel 0.95
Gypsum 0.8 – 0.95
Ice 0.98
Limestone 0.98
Paper (any colour) 0.95
Plastic
Opaque 0.95
Soil 0.9 – 0.98
Water 0.93
Wood, (natural) 0.9 – 0.95
7
Trigger Lock
The thermometer trigger can be locked on for continuous measurement.
To lock the trigger:
1. Press SET button to select Trigger Lock set-up, icon
is blinking on the display
2. Press
or to select ON or OFF.
3. Press YELLOW button to complete the setting and exit Trigger Lock set-up, or press SET button to complete the
setting and continue setting for °C / °F.
°C / °F Set-up
1. Press SET button to select °C / °F set-up, icon °C or °F is blinking on the display
2. Press
or to select °C or °F.
3. Press YELLOW button to complete the setting and exit °C / °F set-up.
MAX, MIN, DIF, AVG
The thermometer can measure maximum (MAX), minimum (MIN), differential (DIF) or average (AVG)
temperatures each time a reading is taken. These values are available for infrared measurements.
1. Press YELLOW button until MIN or MAX or AVG or DIF appears on the display
2. Pull and hold the trigger while aiming at the target surface.
3. The value will show on secondary display.
High Alarm (HAL) and Low Alarm (LAL)
The thermometer has a programmable high and low alarm to designate high or low readings depending on
the thresholds entered. When the alarm level is reached, an alarm will sound. This feature is not available when
measuring by thermocouple.
To set either the high or low alarm:
1. Press YELLOW button to select HAL or LAL option.
2. Press
or to adjust the threshold value.
3. Press SET button to complete the setting. Symbol
or
will appear on the display.
To turn off the high or low alarm:
1. Press YELLOW button to select HAL or LAL option.
2. Press SET button. Disappearance of Symbol
or
on the display indicates high or low alarm is off.
DATA Memory
The thermometer has 99 points data storage and the stored data can be downloaded via IR-750 / IR-750-EUR
software. To save the data:
1. Press YELLOW button to select DATA option.
2. Press
or to select data record number
3. Pull and hold the trigger while aiming at the target surface, press DATA button to save the measurement value.
To erase the saved data, when the thermometer is in standby mode (screen is off), press and hold “DATA” and
” button while pulling the trigger until a beep occurs. The data records on screen displays “ – – – – “, all
saved data is cleared.
K-Type Thermocouple Measurement

To avoid electrical shock or personal injury, do not connect the thermocouple contact probe to live
electrical circuits.
The thermometer comes with a bead K-type thermocouple probe.
K-type thermocouple is connected to the thermometer via TC-K input connector located on right side of
the thermometer. The probe can be used simultaneously while the thermometer is taking non-contact
measurements.
1. Press YELLOW button to enter T-C mode. The display shows OL before the measurement is made.
2. Connect the K-type thermocouple to the thermometer via TC-K input connector located on right side of
the thermometer. Press the trigger to start measuring. Secondary display shows the measurement reading.
3. Release the trigger, the reading is kept on display (display hold: 8 seconds)
8
Locating a Hot or Cold Spot
To find a hot or cold spot, aim the thermometer outside the target area. Then, slowly scan across the area with
an up and down motion until you located the hot or cold spot.
Distance and Spot Size
As the distance (D) from the target being measured increases, the spot size (S) of the area measured by the
instrument becomes larger. The spot size indicates 90% encircled energy.
Field of View
Make sure that the target is larger than the spot size. The smaller the target, the closer you should be to it.
9
Emissivity
Emissivity describes the energy-emitting characteristics of materials. Most organic materials and painted or oxidized
surfaces have an emissivity of about 0.95.
If possible, to compensate for inaccurate readings that may result from measuring shiny metal surfaces, cover
the surface to be measured with masking tape or flat black paint (<150°C / 302°F ) and use the high emissivity
setting. Allow time for the tape or paint to reach the same temperatures as the surface beneath it. Measure the
temperature of the tape or painted surface.
If you cannot use paint or use tape, then you could improve the accuracy of your measurements with the
emissivity selector. Even with emissivity selector, it can be difficult to get a completely accurate infrared
measurement of a target with a shiny or metallic surface.
The thermometer allows you to adjust the emissivity for the type of surface you’re trying to measure the
temperature of. Use a different setting for your application by referring to the Table of Surface Emissivity.
Reminders
1. Changes of surrounding ambient temperature can result in inaccurate reading, allow time for the
instrument to adopt the change of ambient temperature before use. Specified accuracy applies after 30
minutes when the instrument changes to a different temperature environment an inaccurate.
2. The instrument cannot measure through transparent surfaces such as glass. It will measure the surface
temperature of the glass instead.
10
3. See Table of Surface Emissivity for use in measuring shiny or polished metal surfaces (stainless steel,
aluminum, etc.).
4. Steam, dust, smoke, etc., can prevent accurate measurement by obstructing the instrument’s optics.
TYPICAL MEASUREMENTS
This section describes a variety of measurements often performed by technicians.
Reminder:
• User could select to turn on or off the backlight and laser whenever you are making readings with the
thermometer.
• Relatively high emissivity normally means emissivity setting of about 0.95.
• Relatively low emissivity normally means emissivity setting of about 0.30.
• When user cannot identify the emissivity of the object to be measured, user could cover the surface to be
measured (temperature >150°C) with black electric tape (emissivity of about 0.95). Allow time for the tape
to reach the same temperature as the object to be measured. Measure and record the temperature of the
tape.
Aim the thermometer to the object to be measured, adjust the emissivity setting to make it as the same temperature
as the tape. At this time, the thermometer emissivity setting is close to the emissivity of the object to be measured,
measurement can be taken.
Testing Contactors (Starters)
1. Press SET to select emissivity. Press / to select relatively low emissivity for bright contacts, or 0.7 mid
level for darkened contacts.
2. Press YELLOW button to select MAX.
3. Measure line and load side of one pole without releasing trigger.
4. A temperature difference between the line and load sides of a pole indicate increased resistance of one
point and a contactor may be failing.
Testing Enclosed Relays
1. Press SET and then press / to set emissivity to relatively low for un-insulated connectors or relatively
high for plastic encased relays or for Bakelite enclosed relays or insulated connectors.
2. Press YELLOW button to select MAX.
3. Start to scan.
4. Measure the relay casing, looking for hot spots.
5. Measure electrical connections on relay terminals looking for hot spots.
11
Testing Fuses and Buss Connections
1. Press SET and then press / to set emissivity to relatively high for paper covered fuse body or
insulated connections.
2. Press YELLOW button to select MAX.
3. Scan the paper covered length of fuse.
4. Without releasing the trigger, scan each fuse. Unequal temperatures between fuses may indicate voltage or
amperage imbalance.
5. Press SET and then press / to set emissivity to relatively low, for metal fuses and caps and insulated
buss connections.
6. Press MODE to select MAX.
7. Scan each end cap on each fuse.
Note: Unequal temperatures or a high temperature indicates loose or corroded connection through the fuse
buss spring clip
Testing Electrical Connections
1. Press SET and then press / to set emissivity to relatively low for
un-insulated connectors or buss connections or relatively high for insulated connections.
Note: Conductors are typically smaller than the thermometer’s spot size.
If the spot size is bigger than the connector, the temperature reading is the average within the spot.
2. Scan the conductor, moving toward direction of electrical connector (quick connector, wire nut, buss
connection, etc.).
Scanning Walls for Air Leaks or Insulation Deficiencies
1. Turn off heating, cooling, and blower.
2. Press SET to select emissivity. Press
/ to select emissivity relatively high for painted surfaces or
window surfaces.
3. Press YELLOW button to select MIN when opposite side of wall is at lower temperature and or select MAX
when opposite side of wall is at higher temperature.
4. Measure an interior partition wall surface temperature.
5. Do not release the trigger. Record this temperature as your baseline (or benchmark) for a “perfectly”
insulated wall.
6. Face the wall to be scanned. Stand 1.5m away to scan a 4cm spot on the wall.
7. Scan horizontal rows of wall from top to bottom, or horizontal rows of ceiling from wall to wall. Look for
greatest deviations from baseline temperature to identify problems. This completes the insulation test scan.
Turn on the blower (no heat, no cooling) and retest. If test results with the blower on are different than results
with the blower off, this may indicate air leaks in conditioned envelope walls. The air leaks are caused by duct
leaks that create a pressure differential across the conditioned space envelope.
Testing Bearings
Warning
To avoid injury when testing bearings:
1. Do not wear loose clothing, jewelry, or anything around neck when working around moving parts such as
motors, belts, blower, and fans.
2. Make sure an electrical disconnect is within reach and operating correctly and freely.
3. Do not work alone.
Note: It works best to compare two similar motors operating similar loads
12
1. Press SET and then press
/ to select relatively high emissivity.
2. Press SELECT to select MAX.
3. Enable motor and allow it to reach steady state operating temperatures.
4. Disable the motor if possible.
5. Measure the two motor bearing temperatures.
6. Compare the two motor bearing temperatures. Unequal temperatures or a high temperature can indicate
a lubrication or other bearing problem that is resulting from excess friction.
7. Repeat the sequence for the blower bearings.
Testing Belts and Sheaves
1. Press SET and then press / to select relatively high emissivity.
2. Press SELECT to select MAX.
3. Enable the motor and allow it to reach a steady state operating temperatures.
4. Aim the thermometer at the surface to be measured.
5. Start recording temperature.
6. Slowly move the thermometer up the belt toward second sheave.
• If belt is slipping, sheave temperature will be high from friction.
• If belt is slipping, belt temperature will remain high between sheaves.
• If belt is not slipping, belt temperature will reduce between sheaves.
• If inner surfaces of sheaves are not a true “V” shape, this indicates belt slippage and will continue to
operate at elevated temperatures until sheave is replaced.
• Sheaves must be properly aligned (include “pitch & yaw”) for belt and sheaves to operate at
appropriate temperatures. A straight edge or taut string, can be used to check alignments.
• Motor sheave should operate at a temperature consistent with blower sheaves.
• If motor sheave is at a higher temperature at motor shaft than at outer circumference, belt is probably
not slipping.
If outer circumference of sheave is at higher temperature than sheave at motor shaft, then belt is probably
slipping and sheaves may be misaligned.
Checking Hydronic Radiant Heat Applications
1. Press SET and then press / to select relatively high emissivity.
2. Press YELLOW button to select MAX.
3. To locate radiant heat tubes in floor, temporarily elevate the loop temperature to create hotter spots for
identifying tubing runs.
4. Before releasing trigger, press YELLOW button to toggle between MIN, MAX, DIF floor temperatures and
record the temperature for future comparison and trending under similar conditions.
Measuring Grill, Register, or Diffuser Discharge Temperature
1. Press SET and then press / to select relatively high emissivity.
2. Aim the thermometer at the discharge air grille, register, or diffuser.
3. Measure discharge temperature.
4. Release trigger to freeze the temperature reading for 8 seconds and record this temperature.
5. Grille, register, or diffuser temperature should be equivalent to discharge temperature at the air handler.
13
Checking for Blockage in Air-To-Air Evaporator or Condensers
1. Remove panels to gain access to coil return bends or hairpins.
2. Press SET and then press
/ to select relatively high emissivity for copper tube.
3. Start the refrigeration system.
4. Aim the thermometer at coil turn bends/hairpins.
5. Start recording temperature.
6. Take temperature of each return bend/hairpin.
All evaporator return bends/hairpins should be at or slightly above evaporator saturation temperature from the
pressure/temperature chart.
• All condenser return bend/hairpins should be at or slightly less than condenser saturation temperature.
• If a group of return bends/hairpins do not conform to expected temperatures, that indicates a blocked
or restricted distributor or distributor tube.
OPERATING IR-750 / IR-750-EUR SOFTWARE
System Prerequisites
Recommended configuration
Processor Pentium® 4 or higher with support of all customary operating systems
Operating system Microsoft Windows® 2000 / XP / Vista / Win7
Memory 512 MB RAM or higher
Monitor VGA (1024 x 768)
Drive CD-ROM
Input device Mouse or compatible
Interface USB for data transmission
Install IR-750 / IR-750-EUR Software
• Insert the provided CD disk into CD-ROM drive.
• Double-click “IR-750_VX.XX.exe” (X can be 0 to 9 or blank,
indicating the software version)
• The pop-up windows will guide you through the program setup process.
Connecting USB cable to IR-750 / IR-750-EUR Thermometer
14
Main Screen
1
command enables you to create New, Save file and set the numbers of measurement (1000, 5000,
10000, 50000, 100000 or limitless).
2
command enables you to view the software operation instructions.
3
Blinking indicates IR-750 / IR-750-EUR is connected to the software
4
Main display screen of measurement
5
Secondary display screen for measurement data
6
Measurement data in graph
Using Program
button enables you to connect to IR-750 / IR-750-EUR thermometer and start measurement. “USB
Connect” blinking indicates that the IR-750 / IR-750-EUR Thermometer is successfully connected to the
PC and that the thermometer will automatically start taking measurements. The primary display screen shows
current measurement temperatures and the secondary display screen shows all measurement data.
Download Data enables you to download temperature measurements saved in IR-750 / IR-750-EUR Thermometer.
Tick “Download Data” and then click “USB Connect”; the saved data will automatically upload to your PC and
display on secondary display.
Exit and close the program
1
Print data
2
Save data (*.txt, *.xls, *.xml)
3
Open files (*.txt, *.xls, *.xml)
4
Deleting current data and create a new data.
1
2
3
4
5
6
1
2
3
4
15
Repeat: Tick to display every measurement with the interval you set. Un-tick “Repeat”, the secondary display will
only show measurement results that are different from the last measured temperature. The graph display will
show complete measurement results in the graphics despite whether “Repeat” is ticked or un-ticked.
Sampling Interval allows you to set the sampling intervals of measurement ( 1 to 9999 seconds).
Note: When measurement data reaching to 10000 sets (measurement number), a warning message will pop up
and stop measurements. Save or clear the measurement data before continuing next measurements.
1
Last page
5
First page
2
Next page
6
Save graph in current page as *.bmp
3
Current page
7
Set numbers of measurement data on graph display
4
Previous page
Test Report
Test data can be saved as *.txt, *.xls, *.xml by clicking button. When opening data report, make sure
selecting font format as Amprobe-A1 or Amprobe-A2 in “Unit” column in order to display correct temperature
unit °C or °F.
1
2
3
4
5
6
7
16
Zoom In Selected Graph
To enlarge the particular part of the graph, use the mouse pointer to select area from upper-left point to lower-right
point.
You can use button to save the selected enlarged graph.
To resume to its original full view, use the mouse pointer to select any area on the graph from lower-right point
to upper-left point.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114
  • Page 115 115
  • Page 116 116
  • Page 117 117
  • Page 118 118
  • Page 119 119
  • Page 120 120
  • Page 121 121
  • Page 122 122
  • Page 123 123
  • Page 124 124
  • Page 125 125
  • Page 126 126
  • Page 127 127
  • Page 128 128
  • Page 129 129
  • Page 130 130
  • Page 131 131
  • Page 132 132
  • Page 133 133
  • Page 134 134

Amprobe IR-750 Infrared Thermometer Manuel utilisateur

Catégorie
Mesure
Taper
Manuel utilisateur
Ce manuel convient également à