Ohaus MB23 Manuel utilisateur

Taper
Manuel utilisateur
MB23 / MB25 Moisture Analyzer
Instruction Manual
Analizador de humedad MB23 / MB25
Manual de Instrucciones
Analyseur d’humidité MB23 / MB25
Guide de I’utilisateur
MB23 / MB25 Feuchtebestimmer
Bedienungsanleitung
MB23 / MB25 Analizzatore di umidità
Manuale d’instruzioni
MB23 / MB25 EN-1
1. INTRODUCTION
Safety Precautions
For safe and dependable operation of this moisture Analyzer, please comply with the following safety
precautions:
Use the Analyzer exclusively for determination of moisture in samples. Improper operation of the Analyzer
can endanger personnel and cause property damage.
If the Analyzer is used in a manner not specified in this manual, the protection provided may be impaired.
Verify that the input voltage printed on the voltage identification label information and the plug type matches
the local AC power supply.
The Analyzer has a 3-pin power cable that includes a ground connection. Intentionally disabling the
equipment grounding connection is prohibited.
Do not position the Analyzer so that it is difficult to disconnect the Power Plug from the local AC power supply.
Make sure that the power cord does not pose any obstacle or tripping hazard.
Do not operate the Analyzer in hazardous, wet or unstable environments.
Disconnect the Analyzer from the power supply when cleaning the Analyzer.
Ensure sufficient free space around the Analyzer as a safety zone. Allow at least 1 meter of free space above
the Analyzer.
The Analyzer must be operated only by trained personnel who are familiar with the properties of the samples
being tested and with the equipment operation.
Use appropriate personal safety equipment such as safety glasses, gloves, protective clothing and
respirators.
Do not make any modifications to the Analyzer.
Service should be performed only by authorized personnel.
The Moisture Analyzer works with heat!
Never place flammable materials on, below or next to the Analyzer.
Use caution when removing a test sample. The sample, the sample chamber, the heating element and the
surrounding areas may be very hot and can cause burns.
Some samples require special care!
Should there be any uncertainty regarding the safety of a substance, perform a careful risk analysis. In such
cases, never leave the Analyzer unattended.
Fire or explosion: Substances which contain solvents or release flammable or explosive vapors when
heated. With such samples, work at drying temperatures low enough to prevent the formation of flames or
an explosion.
Poisoning or burning: Substances which contain toxic or caustic components should only be dried in a fume
hood.
Corrosive: Substances which release corrosive vapors when heated should be tested in small amounts.
The user assumes responsibility for any damage caused by the use of these types of samples.
Application Disclaimer
Moisture determination applications must be optimized and validated by the user according to local regulations.
Application specific data provided by OHAUS is for reference purposes only. OHAUS waives all liability for
applications based on this data.
EN-2 MB23 / MB25
2. INSTALLATION
Package Contents
Moisture Analyzer
Draft Shield
In-Use Cover
Pan Support
Power Cable
50 Sample Pans
Glass Fiber Pad
Instruction Manual
Sample Pan Handler (MB25 only)
Selecting the Location
Operate the Analyzer on a firm, level surface.
Select a location that is safe and with adequate ventilation. Fire, corrosive or toxic fumes and other hazards associated with
the test samples will require specially prepared locations.
Ensure that the location has easy access to the local AC power supply.
Avoid locations with rapid temperature changes, excessive humidity, air currents, vibrations, electromagnetic fields, heat
or direct sunlight.
Installing Components
(1) Install and position Draft Shield
Connecting Power
Verify that the input voltage printed on the voltage identification label information and
the plug type matches the locations AC power supply.
Connect the supplied power cable to the power input receptacle at the rear of the Analyzer
and into a properly grounded power outlet.
Power: On (short press) / Off / Standby Mode (long press)
When powered on from standby mode, the Analyzer is ready for immediate use.
See also Section 1, Safety Precautions.
After connecting the Analyzer to the AC supply (standby mode), allow the Analyzer to warm up for at least
15 minutes for best results.
MB23 / MB25 EN-3
3. OPERATION
Display
Stable indicator
% Percent solids
%
Percent moisture
g grams
100° Temperature setting or current chamber temperature (Celsius)
10:00 Time setting (minutes : seconds)
Controls
Button:
Name:
Quick Functions:
Button:
Name:
Quick Functions:
Start / Stop
On (short press) / Off (long press)
Start/Stop (short press)
Back (short press)
Set
Increase value
(short or long press)
Tare
Tare (short press)
Enter / Accept value (short press)
Set
Decrease value
(short or long press)
Temp
Temperature Setting (short press)
Print / Cal
Print (short press)
Calibrate (long press)
Time
Time Setting (short press)
% g
Result Unit (short press)
There are also several modes of operation which affect how each button operates. Please refer to the following section.
Operating Modes
Standby Mode
When the Analyzer is connected to AC power and the display is off, the unit is in Standby mode.
Start (Short press) Turns on the display and enters Weigh mode.
Print (Long press) Displays the RS232 settings (see Section 6, RS232 Data Connection.). Thereafter,
Set (Short press) Switch the RS232 setting between two settings. (see Section 6, RS232 Data Settings.)
Set (Short press) Switch the RS232 setting between two settings. (see Section 6, RS232 Data Settings.)
Print (Short press) Re-enters Standby mode (off).
Weigh Mode
The display shows the weight of items placed on the pan support.
Tare (Short press) Sets the displayed weight value to zero
Start (Long press) Enters Standby mode (Off)
The following operations can be initiated:
Temp (Short press) Enters Edit mode for the temperature setting, the value will blink
Time (Short press) Enters Edit mode for the time setting, the value will blink
Start (Short press) Activates Run mode, initiates test with a sample greater than 0.5 g
Cal (Long press) Initiates weight or temperature Calibration (see Section 5, Maintenance)
EN-4 MB23 / MB25
Temperature / Time Edit Mode (Preparing a Test)
Temperature editing: (50° to 160°C in 5° steps)
Temp (Short press) Enters Edit mode for the temperature setting, the value will blink.
Set (Short or long press) Increases the value of the temperature setting.
Set (Short or long press) Decreases the value of the temperature setting.
Tare (Short press) Accepts the blinking value and returns to Weigh mode.
Start (Short press) Exits Edit mode without saving changes.
Note: After 5 seconds of no activity the blinking setting is automatically saved and the mode returns to Weigh.
Time editing: (1 to 60 minutes in 30 second steps, 61 to 99 minutes in 1 minute steps)
The time parameter can be set to timed duration or AUTO (the test ends when the Analyzer detects the weight loss has ended).
Selecting AUTO or timed duration:
Time (Short press) Enters Edit mode for the time setting, the value will blink.
Time (Short press) While the time setting is blinking, toggles between AUTO and a time value.
Tare Accepts the blinking setting.
If AUTO is selected Weigh mode is entered.
If time duration is selected:
Set (Short or long press) Increases the value of the time setting.
Set (Short or long press) Decreases the value of the time setting.
Tare (Short press) Accepts the blinking value and returns to Weigh mode.
Start (Short press) Exits Edit mode without saving changes.
Note: After 5 seconds of no activity the blinking setting is automatically saved and the mode returns to Weigh.
Run Mode (Performing a Test)
A moisture determination test is initiated. The results in progress are displayed.
Start (Short press) Activates Run mode, initiates test with a sample greater than 0.5 g.
%g Changes the unit of the displayed result: weight (grams) > % Moisture > % Solids.
Stop (Short press) Manually stops the test in progress.
Print Sends the current displayed value to the RS232.
Result Mode
At the end of the test (Run Mode), the display blinks the test result.
%g Changes the unit of the displayed result: weight (grams) > % Moisture > % Solids.
Tare Exits to Weigh mode.
Print Sends the current displayed value to the RS232.
How to Prepare a Test
Moisture determinations can be made very simply. The three steps are:
(1) Setting the drying temperature (see Section 3).
(2) Setting the drying time, and (see Section 3).
(3) Preparing the sample to be tested:
Place the Pan Handler with empty test pan on the Pan Support (see Section 2).
Press Tare to zero out the pan weight.
Remove the test pan and place the test sample onto the test pan. The sample must be greater than 0.5g.
Spread the sample evenly across the test pan.
Place the test pan with the sample on the Pan Support. The weight value of the sample will be displayed.
Section 4. Test Optimization gives hints on determining optimal drying temperatures and times, sample sizes,
and proper preparation of samples.
How to Perform a Test
(1) Close the Heater Cover.
(2) Press Start to begin the test (press Start again to stop the test in progress).
(3) When the test is over the display will blink showing the final result.
(4) To change the displayed units, press %g.
(5) To print the current displayed value, press Print.
(6) Press Tare to exit to Weigh mode.
MB23 / MB25 EN-5
Performing a Trial Test
Prior to actual testing, a test run can be performed with these suggested settings:
(1) Temperature = 120
(2) Time = AUTO
(3) Sample = 3g of water. Place a glass fiber pad (included with the Analyzer) on the test pan, place on Pan Support. Press
Tare to zero the pan weight. Add 3g of water to the fiber pad.
(4) Press Start to initiate the test. A perfect result on the trial test would be: 0g, 100% moisture or 0%solids.
Results may vary slightly due to weighing errors involved with a small sample, or other experimental errors.
See also Section 4. Test Optimization.
4. TEST OPTIMIZATION
Moisture is determined from the weight loss of a sample dried by heating.
The speed and quality of the measurement process will rely on the following parameters. Experimentation will also help ascertain
the optimal setup of these parameters:
Drying temperature
Drying time
Sample weight
Sample preparation
Type of sample
Drying Temperature
The drying temperature exerts a controlling influence on the drying time (e.g., a low temperature can prolong the drying time
unnecessarily).
Select a drying temperature that neither decomposes nor changes the chemical structure of the sample.
Some samples can give off different amounts of moisture at different drying temperatures. In these cases, deviations can
be compensated by changing the drying temperature.
Drying Time
This analyzer has three methods to establish the drying time.
Manual, where the user stops the test in progress by pressing the Stop button. The test must be longer than 30 seconds
to be a valid test.
Automatic, which ends the drying process when detecting less than 1mg loss in 60 seconds. To keep the drying time short,
select a small sample weight that still maintains the required measurement accuracy.
Timed duration, where the test ends when the pre-set drying time elapses.
Sample Weight
The weight of a sample influences the measurement time and repeatability of the results. With large amounts of samples, more
moisture must be vaporized and the process takes longer. Generally, the sample weight should be between 3g and 20g. 3g
samples give fast results with some sacrifice in accuracy. 20g samples generally give more consistent results but require a
longer test time.
Another way to determine the sample weight is to use the relation between sample weight and repeatability, as shown in the
following table. If requiring repeatability results better than ±0.3%, for example, the table indicates that a sample weight of
at least 2g will be needed.
Sample Weight
Repeatability
0.5g
1g
2g
5g
10g
±1.0%
±0.6%
±0.3%
±0.12%
±0.06%
Sample Preparation
Samples must always be uniform and representative of the total amount to obtain accurate and reproducible results. When
preparing samples, it is essential to ensure thin and uniform distribution of the sample on the test pan (i.e., avoid piling and
excessive amounts).
EN-6 MB23 / MB25
Types of Samples
Pasty, fat containing and melting substances
Use a glass fiber filter to increase the surface area of these types of samples (e.g., butter). The moisture in these substances
is more uniformly distributed through the filter. The increased surface area results in faster and more complete vaporization
of the moisture.
Liquid substances
Liquids (e.g., dispersions) tend to form drops on the test pan, which prevents rapid drying. Use of a glass fiber filter shortens
the drying time significantly as the filter distributes the liquid sample over a larger surface area.
Skin-forming and temperature sensitive substances
Formation of a film on the surface of these samples can prevent complete determination of moisture. Using a glass fiber filter
to cover the sample allow gentler and more beneficial heating, improving reproducitility.
Sugar-containing substances
Samples containing large amounts of sugar tend to caramelize. Ensure that a thin and uniform layer is applied and a moderate
temperature selected. The sample can also be covered with a glass fiber filter that improves reproducitility.
The following substances present risk of fire, explosion, damage or injury.
Should there be any uncertainty regarding the safety of a substance, always perform a careful risk analysis.
In such cases, never leave the Analyzer unattended.
Volatile substances
With volatile samples, rapid application of the sample on the test pan is advisable, to limit the moisture from escaping before
the initial weight is recorded. These substances also include samples treated with solvents, and substances which contain
solvents or release flammable or explosive vapors. Work at drying temperatures low enough to prevent the formation of flames
or an explosion. Always work with small samples (maximum 1g).
Poisonous and toxic substances
Substances which contain toxic or caustic components should only be dried in a fume hood.
Corrosive substances
Substances which release corrosive vapors when heated (e.g. acids) should be tested in small amounts. The vapors can
condense on the Analyzer parts causing corrosion.
5. MAINTENANCE
Weight Calibration
Weight calibration is rarely required. Moisture analyzers use relative weight values to determine the results, so a minor offset
from the absolute weight has little effect on accuracy. OHAUS moisture analyzers have rugged, high quality temperature
stabilized weighing modules that retain their calibration over long periods of time.
(1) Remove any load on the Pan Support including the sample pan.
(2) Press and hold Cal while in the Weigh mode. “CAL” will appear followed by “50.00g”.
(3) Place a 50g calibration mass on the Pan Support. The display will show “-----”.
(4) Remove the mass when “--0--” is shown on the display. The display will show -----”.
(5) The Analyzer will return to Weigh mode when calibration is complete.
Note: Pressing START will cancel the calibration without saving the changes.
Temperature Calibration
Temperature calibration is rarely required under normal use. If the heating elements become dirty, the usual setting may no
longer produce the same results. A temperature calibration can correct for these changes.
Use only the OHAUS Temperature Calibration Kit (accessory) to perform temperature calibration. Damage
to the Analyzer can occur if using other methods.
(1) Remove the Pan Support for the temperature calibration to start.
(2) Place the OHAUS Temperature Calibration Kit into the temperature chamber (see Kit Instructions). Close Cover.
(3) Press and hold Cal until “Cal” is displayed. When Cal is released TC100” will appear and the heater turns on.
(4) After 15 minutes, the unit will beep and blink 100 on the display.
(5) Read the thermometer on the Temperature Calibration Kit and press Set to change and match the displayed value
to the thermometer reading.
(6) Press Tare to enter the value. The display will show TC160 as the heater turns on again.
MB23 / MB25 EN-7
(7) After 15 minutes, the unit will beep and blink “160” on the display.
(8) Read the thermometer on the Temperature Calibration Kit and press Set to change and match the displayed value
to the thermometer reading.
(9) Press Tare to enter the value. The Analyzer returns to Weigh mode.
(10) Temperature calibration is now complete.
Note: If Tare is not pressed within 10 minutes the calibration will be aborted.
Cleaning
Disconnect the Analyzer from the power supply before cleaning the Analyzer.
Make sure that no liquid enters the interior of the Analyzer.
Make sure the Analyzer is cooled down before cleaning.
Clean the Analyzer at regular intervals.
Housing surfaces and the temperature sensor may be
cleaned with a lint-free cloth slightly dampened with water or
a mild cleaning agent.
Glass surfaces may be cleaned with a commercial glass
cleaner.
Do not use solvents, harsh chemicals, ammonia or abrasive
cleaning agents.
Troubleshooting
Symptom / Display
Possible Cause
Remedy
Cannot turn on
No power to analyzer
Verify connections, fuse and voltage
Weight value blinking
Sample weight less than 0.5g
Increase sample size
“Tare” shown
Pan weight needs set to zero
Press Tare
“Close Cover” shown
Cover needs to be closed before starting test
Close cover
Poor accuracy
Improper calibration
Unstable environment
Perform calibration
Move the analyzer to suitable location
Cannot calibrate
Unstable environment
Incorrect calibration masses
Move the analyzer to suitable location
Use correct calibration masses
Err 7.0
Time out
Err 8.1
Pan support has load during power on
Remove weight from pan support
Err 8.2
Pan support was removed prior to power on
Install pan support
Err 8.3
Weight on pan support exceeds capacity
Remove weight from the pan support
Err 8.4
Pan support was removed during weighing
Re-install pan support
Err 8.5
Analyzer temperature is above 50°C
Cool down the analyzer
Err 8.6
Analyzer temperature is higher than target setting
Cool down the analyzer
Err 9.5
Factory calibration data corrupted
Contact an Ohaus-authorized dealer
Err 10.3
Temperature sensor out of range high
Contact an Ohaus-authorized dealer
Err 10.4
Temperature sensor out of range low
Contact an Ohaus-authorized dealer
Err 10.5
Temperature sensor defective
Contact an Ohaus-authorized dealer
Err 10.6
Temperature sensor defective
Contact an Ohaus-authorized dealer
Err 10.7
Temperature sensor defective
Contact an Ohaus-authorized dealer
Err 11.0
AC power is unstable
Use a power line conditioner
Err 53
EEPROM checksum error
Contact an Ohaus-authorized dealer
Err 54
Calibration parameters loading failed
Contact an Ohaus-authorized dealer
Service Information
If the troubleshooting section does not resolve or describe your problem, contact your authorized OHAUS service agent. Please
visit our web site: www.ohaus.com to locate the OHAUS office nearest you.
Accessories
Description Part No.
Security Locking Cable 76288-01
Security Lock (Kensington
®
type) 470004-010
Temperature Calibration Kit 11113857
Description Part No.
STP103 Thermal Printer, US 80251992
STP103 Thermal Printer, EU 80251993
STP103 Thermal Printer, UK 80251994
Protective glass
Temperature sensor
EN-8 MB23 / MB25
Sample Pans (50/Box) 80850086
Pads, Glass Fiber (200/Box) 80850087
Reusable Sample Pans (3/pk) 80850088
Reusable Deep Sample Pans (3/pk) 80252479
Reusable Wire Cage 80252477
Sample Pan Handler 80252476
50g Calibration Weight 51054-16
CBM910 Impact Printer, JP 80252041
CBM910 Impact Printer, US 80252042
CBM910 Impact Printer, EU 80252043
Cable, STP103 80252581
Cable, CBM910 80252571
Data collection Software SW12
6. TECHNICAL DATA
Admissible Ambient Conditions
Location: Indoor use only
Temperature: 10°C to 4C
Relative humidity: 15 % to 80 % at 3C non-condensing
Warm up time: At least 15 minutes after connecting the Analyzer to the AC supply;
when switched from standby mode the Analyzer is ready for immediate use.
Height above sea level: Up to 2000 m
Power Input: 100 VAC 120 VAC, 3 A, 50/60 Hz or 200 VAC -240 VAC, 3 A, 50/60 Hz
Voltage fluctuations: -15% +10%
Power load: MB25: 250 W; MB23:300 W (Maximum during drying process)
Power line fuse: 1 piece, 5 x 20 mm, 2.5 A 250 V
Protected against dust and water, Pollution degree: 2, Installation category: Class II
Specifications
MODEL
MB25
MB23
Capacity
110 g
Readability (% requires >10 g sample)
0.005 g, 0.05% (0-99 g)
0.01 g, 0.1% (100-110 g)
0.01 g, 0.1%
Temperature Settings
50°C to 160°C (5° increments)
Heat Source
Halogen
Infrared
Calibration - Weight
50 g
Calibration - Temperature
OHAUS Temperature Calibration Kit
Pan Size
90 mm diameter
Unit Dimensions (DxWxH)
11.0x6.5x5.0 in / 28.0x16.5x12.7 cm
Unit Weight
4.6 lb / 2.1 kg
Shipping Dimensions (LxWxH)
20.2x15.5x13.5 in / 51x40x35 cm
Shipping Weight
10 lb / 4.5 kg
Communication
RS232 Pin Connections
Female DB9 connector
Pin 2: Analyzer transmit line (TxD)
Pin 3: Analyzer receive line (RxD)
Pin 4 and 6: Internally connected
Pin 5: Ground signal (GND)
Pin 7: Clear to send (hardware handshake) (CTS)
Pin 8: Request to send (hardware handshake) (RTS)
RS232 Data Settings
Baud Rate (default ): 2400 Data Bits: 7 Parity: N Stop Bits: 2 Flow Control: Xon/Xoff
Baud Rate: 9600 Data Bits: 8 Parity: N Stop Bits: 1 Flow Control: Xon/Xoff
RS232 Commands
The RS232 Interface allows a computer to control the Analyzer, as well as to receive data such as displayed weight.
CTS
RTS
5
9
1
6
RxD
TxD
GND
MB23 / MB25 EN-9
Command
Function
ON
Turns Analyzer ON
OFF
Turns Analyzer OFF
T
Same as pressing Tare
U
Same as pressing %g
START
begins a test
STOP
ends a test
P
Same as pressing Print
xP
Interval Print x = Print Interval (1-3600 sec)
PSN
Print Serial Number
PV
Print software version
?
Print Header
H
Header on or off
RS
Print current RS232 settings
RS:2400,7,N,2,X
Change RS232 setting (The current RS232 setting can be displayed on the LCD by pressing Print
for 2 seconds while in standby mode.)
Baud:1200,2400,4800,9600,19200
Data Bits: 7 or 8
Parity: N = none, O = odd, E = even
Stop Bits: 1 or 2
Handshake: X = xon/xoff (software), R = RTS-CTS (hardware), N = none
The Analyzer will return “ES” for invalid commands.
All communication uses standard ASCII format.
Sent commands must terminate with a Line Feed or Carriage Return Line Feed (CRLF).
RS232 Output
^^^^^^^^^^^^^^^^^^^^^^^^
MOISTURE DETERMINATION
OHAUS MB2x SN #########
Switchoff Mode TIMED 10:00
Drying Temp 100C
Result Units %Moisture
00:10 0.0%MC
00:20 0.0%MC
Elapsed Time 00:02:21
Initial Weight 8.560 g
Final Weigh 8.555 g
Final Result 0.0%MC
Header on
Header on
Header on
Header on
Header on
If Print pressed or print interval set.
If Print pressed or print interval set.
Printed at successful end of test.
Printed at successful end of test.
Printed at successful end of test.
Printed at successful end of test.
^^^^^^^^^^^^^^^^^^^^^^^^
Compliance
Compliance to the following standards is indicated by the corresponding mark on the product.
Marking
Standard
This product conforms to the EMC directive 2004/108/EC and the Low Voltage Directive 2006/95/EC.
The complete Declaration of Conformity is available from Ohaus Corporation
AS/NZS4251.1 Emission, AS/NZS4252.1
CAN/CSA-C22.2 No. 61010-1-04; UL Std. No. 61010A-1
EN-10 MB23 / MB25
Disposal
In conformance with the European Directive 2002/96/EC on Waste Electrical and Electronic Equipment (WEEE) this
device may not be disposed of in domestic waste. This also applies to countries outside the EU, per their specific
requirements.
Please dispose of this product in accordance with local regulations at the collecting point specified for electrical and
electronic equipment.
If you have any questions, please contact the responsible authority or the distributor from which you purchased this
device.
Should this device be passed on to other parties (for private or professional use), the content of this regulation must
also be related.
Disposal instructions in Europe are available online at www.ohaus.com. Choose your, country then search for
“WEEE“.
Thank you for your contribution to environmental protection.
FCC Note
This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC Rules.
These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial
environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance
with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential
area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense.
Industry Canada Note
This Class A digital apparatus complies with Canadian ICES-003.
Cet appareil numérique de la classe A est conforme à la norme NMB-003 du Canada.
ISO 9001 Registration
In 1994, Ohaus Corporation, USA, was awarded a certificate of registration to ISO 9001 by Bureau Veritus Quality International (BVQI),
confirming that the Ohaus quality management system is compliant with the ISO 9001 standard’s requirements. On June 21, 2012,
Ohaus Corporation, USA, was re-registered to the ISO 9001:2008 standard.
LIMITED WARRANTY
OHAUS products are warranted against defects in materials and workmanship from the date of delivery through the duration of the
warranty period. During the warranty period OHAUS will repair, or, at its option, replace any component(s) that proves to be defective
at no charge, provided that the product is returned, freight prepaid, to OHAUS.
This warranty does not apply if the product has been damaged by accident or misuse, exposed to radioactive or corrosive materials,
has foreign material penetrating to the inside of the product, or as a result of service or modification by other than OHAUS. In lieu of
a properly returned warranty registration card, the warranty period shall begin on the date of shipment to the authorized dealer. No
other express or implied warranty is given by OHAUS Corporation. OHAUS Corporation shall not be liable for any consequential
damages.
As warranty legislation differs from state to state and country to country, please contact OHAUS or your local OHAUS dealer for further
details.
MB23 / MB25 ES-1
1. INTRODUCCN
Precauciones de seguridad
Para una operación segura y confiable de este analizador de humedad, implemente las siguientes medidas
de seguridad:
Use el analizador exclusivamente para determinar la humedad en muestras. La operación incorrecta de
analizador puede poner en riesgo al personal y causar daños físicos.
Si el analizador se usa en una forma no especificada en este manual, la protección que se proporciona
podría deteriorarse.
Verifique que el voltaje de alimentación impreso en la información de la etiqueta de voltaje y el tipo de enchufe
coincidan con la alimentación de corriente alterna local.
El analizador tiene un cable de tres patillas que incluyen una conexión a tierra. Está prohibido suprimir
intencionalmente la conexión a tierra.
No coloque el analizador en tal forma que sea difícil desconectar el enchufe de la toma de la corriente
eléctrica alterna.
Asegúrese de que el cordón eléctrico no represente un obstáculo o riesgo de tropezarse.
No opere el analizador en ambientes peligrosos o inestables.
Desconecte el analizador de la fuente de energía antes de limpiarlo.
Asegúrese de que haya suficiente espacio libre alrededor del analizador como zona de seguridad. Deje
por lo menos 1 metro de espacio libre encima del analizador.
El analizador debe ser operado solamente por personal capacitado que esté familiarizado con las
propiedades de las muestras que se prueban y con el funcionamiento del equipo.
Use equipo de protección personal adecuado tal como lentes de seguridad, guantes, ropa protectora y
respiradores.
No haga ninguna modificación al analizador.
El servicio debe proporcionarse solamente por personal autorizado.
¡El analizador de humedad funciona con calor!
Nunca coloque materiales inflamables sobre, debajo ni junto al analizador.
Tenga cuidado cuando retire las muestras de prueba. La muestra, la cámara de muestras, el elemento de
calentamiento y las áreas cercanas pueden estar muy calientes y causar quemaduras.
¡Algunas muestras requieren atención especial!
En caso de duda en relación a la seguridad de una sustancia, lleve a cabo un análisis de riesgo cuidadoso.
En tales casos nunca deje el analizador solo.
Incendio o explosión: Las sustancias que contienen solventes o que liberan vapores inflamables o
explosivos cuando son calentadas. Con dichas muestras, trabaje a temperaturas de secado
suficientemente bajas para prevenir que se formen flamas o una explosión.
Envenenamiento o quemaduras: Las sustancias que contienen componentes tóxicos o cáusticos deben
secarse sólo en una capucha para humo.
Corrosivas: Las sustancias que liberan vapores corrosivos cuando se calientan deben probarse en
cantidades pequeñas.
El usuario asume la responsabilidad de cualquier daño causado por el uso de estos tipos de muestras.
Exención de responsabilidad de la aplicación
El usuario debe optimizar y validad las aplicaciones para determinar humedad de acuerdo con las regulaciones
locales. Los datos específicos de aplicación que proporciona OHAUS son lo para referencia. OHAUS renuncia
a cualquier responsabilidad para las aplicaciones que se basen en estos datos.
ES-2 MB23 / MB25
2. INSTALACIÓN
Contenido del paquete
Analizador de humedad
Pantalla contra corrientes de aire
Cubierta para uso
Soporte para bandeja
Cordón eléctrico
Bandeja para 50 muestras
Almohadilla de fibra de vidrio
Manual de instrucciones
Manipulador de bandeja para muestras (sólo MB25)
Selección del lugar de instalación
Opere el analizador sobre una superficie firme y nivelada.
Seleccione un lugar seguro y con ventilación adecuada. El fuego y los vapores corrosivos y xicos y otros riesgos asociados
con las muestras de prueba requerirán lugares preparados especialmente.
Asegúrese de que el lugar tenga fácil acceso a la corriente eléctrica alterna local.
Evite lugares donde ocurran cambios rápidos de temperatura, con humedad, corrientes de aire, vibraciones, campos
electromagnéticos, luz solar directa o calor excesivos.
Instalación de los componentes
(1) Instale y coloque la pantalla contra
corrientes de aire
(2) Instale el soporte de la bandeja y relo
hasta que encaje en su lugar
Soporte opcional para bandeja: Deslice
una bandeja para muestras vacía debajo
del manipulador de bandeja y colóquela
sobre el soporte de la bandeja.
Conexión de la energía
Verifique que el voltaje de alimentación impreso en la información de la etiqueta de
voltaje y el tipo de enchufe coincidan con la alimentación de corriente alterna local.
Conecte el cordón eléctrico en el receptáculo de entrada de energía en la parte posterior
del analizador y en una toma eléctrica con conexión a tierra adecuada.
Energía eléctrica: Encendido (presn corta) / Off / Modo En espera (presn larga)
Cuando se enciende estando en modo en espera, el analizador puede usarse de inmediato.
Vea también la Sección 1 Precauciones de seguridad.
Después de conectar el analizador en la corriente alterna (modo En espera), espere a que se caliente por
lo menos durante 15 minutos para obtener los mejores resultados.
MB23 / MB25 ES-3
3. OPERACIÓN
Pantalla
Indicador estable
% Porcentaje de sólidos
%
Porcentaje de humedad
g gramos
100° Ajuste de temperatura o temperatura actual de la cámara (°C)
10:00 Ajuste de la hora (minutos : segundos)
Controles
Botón:
Nombre:
Funciones rápidas:
Botón:
Nombre:
Funciones rápidas:
Iniciar/
detener
Encendido (presión corta) / Off
(presión larga)
Iniciar/Detener (presión corta)
Regresar (presión corta)
Ajustar
Aumentar el valor
(presión corta o larga)
Tara
Tarar (presión corta)
Ingresar / Aceptar valor (presión corta)
Ajustar
Disminuir el valor
(presión corta o larga)
Temp
Ajuste de temperatura (presión corta)
Imprimir /
Calibrar
Imprimir (presión corta)
Calibrar (presión larga)
Hora
Ajustar la hora (presión corta)
%g
Unidad de resultado (presión
corta)
También hay varios modos de operación que modifican el funcionamiento de cada botón. Vea la siguiente sección.
Modos de operación
Modo En espera
Cuando el analizador es conectado a la corriente alterna y la pantalla es apagada, la unidad está en modo En espera.
Iniciar (Presión corta) Enciende la pantalla e ingresa en el modo Pesar.
Imprimir (Presión larga) Muestra los ajustes del RS232 (vea la Sección 6, Comunicación.) Después de eso,
Ajustar (Presión corta) Cambiar la configuración predeterminada RS232 entre dos valores. (vea la Sección 6)
Ajustar (Presión corta) Cambiar la configuración predeterminada RS232 entre dos valores. (vea la Sección 6)
(Presión corta) Reingresa al modo En espera (apagado).
Modo Pesar
La pantalla muestra el peso de los artículos colocados en el soporte de la bandeja.
Tara (Presión corta) Establece el valor del peso mostrado a cero
Iniciar (Presión larga) Ingresa al modo En espera (apagado).
Se pueden iniciar las siguientes operaciones:
Temp (Presión corta) Ingresa en el modo Modificar para ajustar la temperatura; el valor destellará
Hora (Presión corta) Ingresa en el modo Modificar para ajustar la hora; el valor destellará
Iniciar (Presión corta) Activa el modo Ejecutar; inicia la prueba con una muestra mayor de 0.5 g
Cal (Presión larga) Inicia la calibración del peso o temperatura (vea la Sección 5. Mantenimiento)
ES-4 MB23 / MB25
Modo Modificar Temperatura / Hora (preparación de una prueba)
Modificación de temperatura: (50 a 160 °C en pasos de 5°)
Temp (Presión corta) Ingresa en el modo Modificar para ajustar la temperatura; el valor destellará.
Ajustar (Presión corta o larga) Aumenta el valor del ajuste de temperatura.
Ajustar (Presión corta o larga) Disminuye valor del ajuste de temperatura.
Tara (Presión corta) Acepta el valor que está destellando y regresa al modo Pesar.
Iniciar (Presión corta) Sale del modo Modificar sin guardar los cambios.
Nota: Después de 5 segundos de inactividad, el ajuste que destella se guarda automáticamente y el modo regresa a Pesar.
Modificación de la hora: (1 a 60 minutos en pasos de 30 segundos, 61 a 99 minutos en pasos de 1 minuto)
El parámetro de la hora puede fijarse en una duración de tiempo o en AUTOMÁTICO (la prueba finaliza cuando el analizador
detecta que ha finalizado la pérdida de peso). Selección de AUTOMÁTICO o duración de tiempo:
Hora (Presión corta) Ingresa en el modo Modificar para ajustar la hora; el valor destellará.
Hora (Presión corta) Mientras el ajuste de hora destella, alterna entre AUTMÁTICO y un valor de tiempo.
Tara Acepta el valor que es destellando.
Si se seleccionó AUTOTICO, ingresa en el modo Pesar.
Si se seleccionó duración de tiempo:
Ajustar (Presión corta o larga) Aumenta el valor del ajuste de hora.
Ajustar (Presión corta o larga) Disminuye el valor del ajuste de hora.
Tara (Presión corta) Acepta el valor que está destellando y regresa al modo Pesar.
Iniciar (Presión corta) Sale del modo Modificar sin guardar los cambios.
Nota: Después de 5 segundos de inactividad, el ajuste que destella se guarda automáticamente y el modo regresa a Pesar.
Modo Ejecutar (realización de una prueba)
Se inicia una prueba de determinación de humedad. Se muestran los resultados en curso.
Iniciar (Presión corta) Activa el modo Ejecutar; inicia la prueba con una muestra mayor de 0.5 g.
%g Cambia la unidad del resultado mostrado: peso (gramos) > % de humedad > % de sólidos.
Detener (Presión corta) Detiene manualmente la prueba en curso.
Imprimir Envía el valor actual mostrado al puerto RS232.
Modo Resultado
Al final de la prueba (modo Ejecutar), la pantalla destella el resultado de la prueba.
%g Cambia la unidad del resultado mostrado: peso (gramos) > % de humedad > % de sólidos.
Tara Sale del modo Pesar.
Imprimir Envía el valor actual mostrado al puerto RS232.
mo preparar una prueba
Las determinaciones de humedad pueden hacerse de manera muy simple. Los tres pasos son:
(1) Ajuste de la temperatura de secado (vea la Sección 3).
(2) Ajuste del tiempo de secado (vea la Sección 3).
(3) Preparación de la muestra a ser probada:
Coloque el manipulador de la bandeja con una bandeja de prueba vacía en el soporte de la bandeja (vea la Sección
2).
Presione Tara para poner en cero el peso de la bandeja.
Retire la bandeja de prueba y coloque la muestra de prueba en la bandeja de prueba. La muestra debe ser mayor de
0.5 g.
Esparza la muestra uniformemente en la bandeja de prueba.
Coloque la bandeja de prueba con la muestra sobre el soporte de la bandeja. Se mostrará el valor del peso de la muestra.
Sección 4. Optimización de la prueba da sugerencias para determinar temperaturas y tiempos óptimos de
secado, tamaños de muestras y preparación correcta de muestras.
mo realizar una prueba
(1) Cierre la tapa del calentador.
(2) Presione Iniciar para comenzar la prueba (presione Iniciar nuevamente para detener la prueba en curso).
(3) Cuando termine la prueba, la pantalla destellará para mostrar el resultado final.
(4) Para cambiar las unidades mostradas, presione %g.
(5) Para imprimir el valor actual mostrado, presione Imprimir.
(6) Presione Tara para salir del modo Pesar.
MB23 / MB25 ES-5
Realización de una prueba de ensayo
Antes de la prueba real, puede realizar una prueba con estos ajustes sugeridos:
(1) Temperatura = 120
(2) Hora = AUTOMÁTICO
(3) Muestra = 3 g de agua. Coloque una almohadilla de fibra de vidrio (incluida con el analizador) sobre la bandeja de prueba
y coloque ésta sobre el soporte. Presione Tara para poner en cero el peso de la bandeja. Agregue 3 g de agua a la
almohadilla de fibra.
(4) Presione iniciar para comenzar la prueba. Un resultado ideal de la prueba de ensayo sería: 0 g, 100% de humedad o
0% de sólidos.
Los resultados pueden variar ligeramente debido a errores de pesaje involucrados con una muestra pequeña
u otros errores experimentales. Vea también la Seccn 4. Optimización de la prueba.
4. OPTIMIZACIÓN DE LA PRUEBA
La humedad se determina mediante la pérdida de peso de una muestra secada mediante calentamiento.
La velocidad y calidad del proceso de medición dependen de los siguientes parámetros. La experimentación también ayudará
a determinar el ajuste óptimo de estos parámetros:
Temperatura de secado
Tiempo de secado
Peso de la muestra
Preparación de la muestra
Tipo de muestra
Temperatura de secado
La temperatura de secado ejerce una influencia controladora sobre el tiempo de secado (es decir, una temperatura baja
puede prolongar el tiempo de secado innecesariamente).
Seleccione una temperatura de secado que no descomponga ni cambie la estructura química de la muestra.
Algunas muestras pueden emitir diferentes cantidades de humedad a diferentes temperaturas de secado. En esos casos,
las desviaciones pueden compensarse al cambiar la temperatura de secado.
Tiempo de secado
El analizador tiene tres métodos para establecer el tiempo de secado.
Manual, en el que el usuario detiene la prueba en curso al presionar el bon Detener. La prueba debe durar s de 30
segundos para que sea válida.
Automático, el cual termina el proceso de secado cuando detecta una pérdida menor de 1 mg en 60 segundos. Para
mantener el tiempo de secado corto, seleccione un peso de muestra pequeño que n mantenga la precisión de medición
requerida.
Duración de tiempo, en el que la prueba termina cuando transcurre el tiempo de secado preestablecido.
Peso de la muestra
El peso de la muestra influye en el tiempo de medición y repetibilidad de los resultados. Con mayores cantidades de muestras,
s humedad debe evaporarse y el proceso dura s tiempo. Generalmente el peso de la muestra debe ser entre 3 g y 20
g. Las muestras de 3 g dan resultados s pidos con un poco de sacrificio de la precisión. Las muestras de 20 g generalmente
dan resultados más consistentes pero necesitan tiempos de prueba más largos.
Otra forma de determinar el peso de la muestra es con el uso de la relacn entre el peso de la muestra y la repetibilidad, según
se muestra en la tabla siguiente. Si se requieren resultados de repetibilidad mejores de ±0.3%, por ejemplo, la tabla indica
que se necesitará un peso de muestra de por lo menos 2 g.
Peso de la muestra
Repetibilidad
0.5 g
1 g
2 g
5 g
10 g
±1.0%
±0.6%
±0.3%
±0.12%
±0.06%
Preparación de la muestra
Las muestras siempre deben ser uniformes y representativas de la cantidad total para obtener resultados precisos y
reproducibles. Cuando prepara muestras, es esencial asegurar la distribución delgada y uniforme de la muestra sobre la
bandeja (es decir, evitar apilar y cantidades excesivas).
ES-6 MB23 / MB25
Tipos de muestras
Sustancias pastosas, con contenido de grasa y derretibles
Use un filtro de fibra de vidrio para aumentar la superficie de estos tipos de muestras (por ejemplo, mantequilla). La humedad
de estas sustancias se distribuye de manera más uniforme a través del filtro. El área de la superficie aumentada resulta en
una vaporización más rápida y completa de la humedad.
Sustancias líquidas
Los líquidos (por ejemplo, las dispersiones) tienden a formar gotas en la bandeja, lo cual evita el secado pido. El uso de
un filtro de fibra de vidrio acorta el tiempo de secado significativamente ya que el filtro distribuye el quido sobre un área mayor.
Sustancias que forman una película y sensibles a la temperatura
La formación de una pecula sobre la superficie de estas muestras puede prevenir la determinación completa de la humedad.
El uso de un filtro de fibra de vidrio para cubrir la muestra permite un calentamiento más delicado y benéfico y mejora la
reproducibilidad.
Sustancias que contienen azúcar
Las muestras que contienen grandes cantidades de azúcar tienden a caramelizarse. Asegúrese de aplicar una capa delgada
y uniforme y que se seleccione una temperatura moderada. La muestra también puede cubrirse con un filtro de fibra de vidrio
que mejora la reproducibilidad.
Las siguientes sustancias presentan riesgo de incendio, explosión, daño y lesión.
En caso de duda en relación a la seguridad de una sustancia, siempre lleve a cabo un análisis de riesgo
cuidadoso. En tales casos nunca deje el analizador solo.
Sustancias volátiles
Con las muestras votiles, se aconseja la aplicacn pida de la muestra en la bandeja de la muestra para restringir el escape
de humedad antes de que se registre el peso inicial. Estas sustancias también incluyen muestras tratadas con solventes y
sustancias que contienen solventes o que emiten vapores inflamables o explosivos. Trabaje a temperaturas de secado
suficientemente bajas para prevenir que se formen flamas o una explosión. Siempre trabaje con muestras pequeñas (máximo
1 g).
Sustancias venenosas y tóxicas
Las sustancias que contienen componentes tóxicos o cáusticos deben secarse sólo en una capucha para humo.
Sustancias corrosivas
Las sustancias que liberan vapores corrosivos cuando se calientan (por ejemplo, los ácidos) deben probarse en cantidades
pequeñas. Los vapores pueden condensarse en las partes del analizador y causar corrosión.
5. MANTENIMIENTO
Calibración del peso
La calibración del peso se requiere esporádicamente. Los analizadores de humedad usan valores de peso relativos para
determinar los resultados, de manera que una desviacn menor del peso absoluto tiene poco efecto sobre la precisión. Los
analizadores de humedad de OHAUS poseen dulos de pesaje resistentes y de alta calidad estabilizados a alta temperatura
que retienen su calibracn durante largos periodos de tiempo.
(1) Retire cualquier carga que haya en el soporte de la bandeja incluyendo la bandeja.
(2) Presione y mantenga así Cal mientras está en el modo Pesar. Aparecerá la palabra “CAL” seguida de “50.00g”.
(3) Coloque un peso de calibración de 50 g sobre el soporte de la bandeja. La pantalla mostrará “-----”.
(4) Retire el peso cuando la pantalla muestre “--0--”. La pantalla mostrará “-----”.
(5) El analizador regresará al modo Pesar cuando la calibración haya finalizado.
Nota: Al presionar Iniciar se cancela la calibración sin guardar los cambios.
Calibración de la temperatura
La calibración de la temperatura se requiere esporádicamente durante el uso normal. Si los elementos de calentamiento se
ensucian, el ajuste usual puede dejar de producir los mismos resultados. La calibración de la temperatura puede corregir
estos cambios.
Use solamente el paquete de calibración de temperatura de OHAUS (accesorio) para calibrar la temperatura.
El analizador puede resultar dañado si se usan otros métodos.
MB23 / MB25 ES-7
(1) Retire el soporte de la bandeja para que inicie la calibración de la temperatura.
(2) Coloque el paquete de calibración de la temperatura de OHAUS en la mara de temperatura (lea las instrucciones del
paquete). Cierre la tapa.
(3) Presione y mantenga así Cal hasta que aparezca Cal” en la pantalla. Cuando suelte el botón Cal, aparecerá “TC100”
en la pantalla y el calentador se encenderá.
(4) Después de 15 minutos, la unidad emitirá una alarma “beep” y el número “100” destellará en la pantalla.
(5) Lea el termómetro del paquete de calibración de temperatura y presione Ajustar para cambiar y hacer coincidir el
valor con la lectura del termómetro.
(6) Presione Tara para ingresar el valor. La pantalla mostrará “TC160” mientras que el calentador se enciende nuevamente.
(7) Después de 15 minutos, la unidad emitirá una alarma “beep” y el número “160” destellará en la pantalla.
(8) Lea el termómetro del paquete de calibración de temperatura y presione Ajustar para cambiar y hacer coincidir el
valor con la lectura del termómetro.
(9) Presione Tara para ingresar el valor. El analizador regresa al modo Pesar.
(10) La calibración de la temperatura está finalizada.
Nota: Si no se presiona Tara dentro de 10 minutos, la calibración se cancelará.
Limpieza
Desconecte el analizador de la fuente de energía antes de limpiarlo.
Asegúrese de que no entre líquido al interior del analizador.
Asegúrese de que el analizador se enfríe antes de limpiarlo.
Limpie el analizador a intervalos regulares.
Las superficies del gabinete y el sensor de temperatura
pueden limpiarse con una tela sin pelusa ligeramente
empapada con agua o con un agente suave de limpieza.
Las superficies de cristal pueden limpiarse con un
limpiador comercial para cristal.
No use solventes, químicos fuertes, amoníaco ni agentes
de limpieza abrasivos.
Identificación y solución de problemas
ntoma o texto
Causa posible
Solución
ntoma / Display
Posible Causa
Solución
No se pone en marcha
No llega corriente al analizador
Verifique las conexiones el fusibles y la tensn
El indicador de peso
parpadea
Peso de la muestra inferior a 0,5 g
Aumentar el tamaño de la muestra
Se indica “Tara”
El platillo de peso necesita ponerse a cero
Presione Tara
Se indica “Cierre tapa”
La tapa se tienen que cerrar antes de iniciar le test
Cierre la tapa
Precisión insuficiente
Calibración incorrecta
Entorno inestable
Realice calibración
Mueva el analizador a un lugar más adecuado
No se puede calibra
Entorno inestable
Masas de calibración incorrectas
Mueva el analizador a un lugar más adecuado
Utilice masas de calibración adecuadas
Err 7.0
Time out
Err 8.1
El soporte del platillo ha sido cargado durante la puesta en
marcha
Saque el peso del soporte del platillo
Err 8.2
El soporte del platillo se ha quitado antes de la puesta en
marcha
Coloque el soporte del platillo
Err 8.3
El peso en el soporte del platillo excede la capacidad
Quite el peso del soporte del platillo
Err 8.4
El soporte del platillo ha sido quitado durante la pesada
Re-instale el soporte del platillo
Err 8.5
Analizador de temperatura es superior a 50 ° C
Enfríe el analizador
Err 8.6
Analizador de temperatura es más alta que la fijacn de
objetivos
Enfríe el analizador
Err 9.5
Datos de calibración de fábrica dañados
ngase en contacto con un distribuidor
autorizado de Ohaus
Err 10.3
Sensor de temperatura fuera de rango - alto
ngase en contacto con un distribuidor
autorizado de Ohaus
Cristal protector
Sensor de temperatura
ES-8 MB23 / MB25
Err 10.4
Sensor de temperatura fuera de rango - bajo
ngase en contacto con un distribuidor
autorizado de Ohaus
Err 10.5
Sensor de temperatura defectuoso
ngase en contacto con un distribuidor
autorizado de Ohaus
Err 10.6
Sensor de temperatura defectuoso
ngase en contacto con un distribuidor
autorizado de Ohaus
Err 10.7
Sensor de temperatura defectuoso
ngase en contacto con un distribuidor
autorizado de Ohaus
Err 11.0
Corriente AC inestable
Utilice un regulador de nea eléctrica
Err 53
EEPROM checksum error
ngase en contacto con un distribuidor
autorizado de Ohaus
Err 54
Los parámetros de calibración de carga no
ngase en contacto con un distribuidor
autorizado de Ohaus
Información de servicio
Si la sección de identificación y solución de problemas no resuelve o describe su problema, comuníquese con su agente
autorizado de servicio de OHAUS. Visite nuestra página web: www.ohaus.com para encontrar la sucursal de OHAUS más
cercana a usted.
Accesorios
Descripción Parte No.
Cable de bloqueo de seguridad 76288-01
Candado de seguridad (tipo Kensington®) 470004-010
Paquete de calibración de la temperatura 11113857
Bandejas para muestras (50/caja) 80850086
Almohadillas, fibra de vidrio (200/caja) 80850087
Bandejas para muestras reutilizables (3) 80850088
Bandejas profundas para muestras reutilizables(3) 80252479
Jaula de alambre reutilizable 80252477
Manipulador de bandeja para muestras 80252476
Peso de calibración de 50 g 51054-16
Descripción Parte No.
Impresora térmica STP103, EE.UU. 80251992
Impresora térmica STP103, U.E. 80251993
Impresora térmica STP103, R.U. 80251994
Impresora Impact CBM910, JP 80252041
Impresora Impact CBM910, EE.UU. 80252042
Impresora Impact CBM910, U.E. 80252043
Cable, STP103 80252581
Cable, CBM910 80252571
Software de recopilación de datos SW12
6. DATOS TÉCNICOS
Condiciones ambientales admisibles
Ubicación: Para uso exclusivo en interiores
Temperatura ambiente: 10 a 40 °C
Humedad relativa: 15 a 80 % a 30 °C, no condensante
Tiempo de calentamiento: Por lo menos 15 minutos después de conectar el analizador a la fuente de corriente alterna;
cuando cambia del modo En espera el analizador puede usarse de inmediato.
Altura sobre el nivel del mar: Hasta 2000 m
Entrada: 100 VCA 120 VCA, 3 A, 50/60 Hz o 200 VCA 240 VCA, 3 A, 50/60 Hz
Fluctuaciones de voltaje: -15% +10%
Carga eléctrica: MB25: 250 W; MB23:300 W (máxima durante el proceso de secado)
Fusible para corriente eléctrica: 1 pieza, 5 x 20 mm, 2.5 A 250 V
Protección: Protegida contra polvo y agua, grado de contaminación: 2, categoría de instalación: Clase II
Especificaciones
MODELO
MB25
MB23
Capacidad
110 g
Capacidad de lectura (% requiere
muestra >10 g)
0.005 g, 0.05% (0-99 g)
0.01 g, 0.1% (100-110 g)
0.01 g, 0.1%
Ajustes de temperatura
50 a 160 °C (incrementosde 5 °C )
Fuente de calor
Halógena
Infrarroja
Peso de calibración
50 g
Calibración de la temperatura
Paquete de calibración de la temperatura de OHAUS
Tamaño de la bandeja
90 mm de diámetro
Dimensiones de la unidad (PxAxH)
11.0x6.5x5.0 pulgadas / 28.0x16.5x12.7 cm
Peso de la unidad
4.6 lb / 2.1 kg
Dimensiones de transporte (LxAxH)
20.2x15.5x13.5 pulgadas / 51x40x35 cm
Peso de transporte
10 lb / 4.5 kg
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60

Ohaus MB23 Manuel utilisateur

Taper
Manuel utilisateur