Megger OTS60SX Le manuel du propriétaire

Taper
Le manuel du propriétaire
M
OTS60SX
Semi Automatic Oil Test Set
User Guide
Guide de l’utilisateur
Gebrauchsanleitung
Guía del usuario
Contents Guide de l’utilisateur - p14 Gebrauchsanleitung - s24 Guía del usuario - p34
Safety Warnings 3
General Description 4
Preparing the Oil Test Set 5
Operation 5
Oil Sampling 5
Preparing the Oil Test Vessel 5
Control Panel 6
Breakdown Testing 6
Withstand Testing 6
Warning messages 7
Testing to National Standards 7
Sampling and Testing Oil for Dielectric strength 8
Oil Testing Specifications 10
Test Set Specification 10
Accessories 12
Repair and Warranty 13
2.
3.
SAFETY WARNINGS
Safety warnings and precautions must be read and understood before the instrument is used.
They must be observed during use.
The oil test set must be properly earthed.
The test chamber must be kept scrupulously clean.
If the test chamber cover is cracked or damaged in any way the test set must not be used, but sent
for repair to an authorised agent.
The chamber door hinges are fitted with earth contact springs. These must not be damaged or
corroded.
The correct type of oil test vessel must be used, and correctly positioned on the support horns
before carrying out any testing.
Replacement fuses must be of the correct type and rating.
NOTE
THE TEST SET MUST ONLY BE USED BY SUITABLY TRAINED AND COMPETENT PERSONS.
Symbols used on the Test Set
Caution: Refer to accompanying notes.
Risk of electric shock.
Equipment complies with relevant EU Directives
Mains Input Panel
General Description
4.
The OTS60SX is a lightweight, semi-automatic, oil dielectric
strength test set. The instrument is suitable for protected field
use and can be powered from a range of mains supplies. The
maximum 60 kV output allows tests to be performed on oil from
a wide variety of electrical installations including transformers,
circuit breakers and other equipment. The operation of the test
set is extremely simple and the results are displayed on a bright
LED display. A selection of vessels enables the instrument to be
configured for a variety of tests.
The semi automatic operation allows all types of testing to be
performed. An automatic one minute timer operates when the
high voltage is paused for Withstand (proof) testing. Oil samples
can also be tested according to breakdown specification by using
a suitable sequence of tests.
The instrument is constructed in a strong, sheet steel case.
Handles are provided for ease of transportation. A pouch is
located on the side of the instrument to contain accessories such
as the power cord, additional electrodes and the spacing gauge.
An optional carrying case incorporates a shoulder strap.
A number of vessels are available suitable for testing to a wide
range of national specifications. Three types of electrode are
available; spherical (IEC type), mushroom (VDE/ASTM D1816)
and cylindrical (ASTM D877). The oil sample can be stirred by a
motor driven impeller available in selected vessels. The vessel is
located in the top of the instrument and covered by a transparent
polycarbonate door with a mesh screen so that the oil breakdown
can be viewed. A safety interlock ensures that the high voltage is
disconnected when the chamber door is opened.
Voltage
Selector
Mains
power
input
Fuseholders
Ensure that the instrument is properly earthed. The test set must
be connected to a socket (receptacle) with a protective earth
(ground) conductor.
Before connecting the instrument to the mains power supply,
ensure the voltage selector located on the mains input panel is
set to the required voltage. The correct fuse value should be
fitted for the supply voltage (See Specification). The fuse
holders are located next to the supply inlet.
Power cord
If the power cord plug is not suitable for your type of socket
outlets, do not use an adaptor. You should use a suitable
alternative power cord, or if necessary change the plug by cutting
the cord and fitting a suitable plug.
The colour code of the cord is:
Earth (Ground) Yellow / Green
Neutral Blue
Phase (Line) Brown
If using a fused plug, a 3 Amp fuse to BS 1362 should be fitted.
Note: A plug severed from the power cord should be destroyed,
as a plug with bare conductors is hazardous in a live socket
outlet.
Oil Sampling
The relevant test specification will give advice on the best
methods to use when sampling oil. Contamination of the oil may
lead to misleading results. Always run off a small amount of oil
before sampling and ensure that the flow of oil is steady.
Preparing the Oil Test Vessel
Optional vessels are available for testing to national
specifications (See table and Accessories). Some
specifications require stirring of the oil sample as well as different
test electrode shapes. Once the correct vessel has been chosen
for the required test the electrode gap should be set. This is
achieved by adjusting the threaded nut on the side of the vessel.
Small adjustment of the gap can be made by hand tightening the
nuts and rotating the spindle. A slot is provided on the end of the
spindle for this purpose.
Aset of individual 0,5mm feeler gauges is provided for accurate
measurement of the gap. Gauges can be combined to measure
from 1mm to 4mm. (see page 12)
Clean the vessel in accordance with the instructions given in the
relevant test specification. Allowing space for the lid to be fitted,
fill the vessel with the oil sample. The amount of oil required will
depend upon the type of lid used. Load the vessel in the test
chamber, connecting the motor power cable, if fitted, and
carefully shut the door so that the safety interlock switch
operates.
Preparing the Oil Test Set
5
Operation
Control Panel
The control panel contains the supply on/off switch, the high
voltage display, voltage ramp rate selection and the three control
keys. The operation of the instrument is very simple. When the
instrument is turned on, the on/off switch is illuminated, and the
software edition code will be flashed up on the display.
Breakdown Testing
The suitable ramp rate for the test to be performed should be
selected using the ‘s' key. The START key ‘ ‘ will then turn on
the high voltage. The red H.V. indicator will light to show the
output voltage is on. The voltage will then increase at the
selected ramp rate and the corresponding value will be shown on
the display. If oil breakdown occurs the instrument will detect this
and immediately shut off the high voltage. The breakdown
voltage value will be left on the display until the next test is
started. If no breakdown occurs, the high voltage will rise to
60 kV. At this point the test voltage is cut off and this maximum
value left on the display.
A test can be stopped at any time by pressing the STOP n
button. This will remove the high voltage from the oil sample and
leave the display at 0,0 kV.
Withstand Testing
AWithstand (proof) test can be carried out by pressing the
PAUSE ‘ ‘ key. This will maintain the high voltage at the current
value for one minute. After this time the high voltage will continue
to rise until one of the following:-
1) Oil breakdown occurs
2) The STOP n'key is pressed
6.
Mains
on/off
switch
HV on
indicator
Ramp rate
selection
Main
controls
LED
Display
s
Control Panel
7
3) The output voltage reaches the maximum value of the test set.
During withstand testing the ramp rate can be changed while the
pause is in progress. This can allow the withstand test value to be
accurately set by adjusting the ramp rate to the lowest value.
Warning Messages
Door Open
If the instrument overheats, the words ‘ ‘ and
‘ will be alternately displayed and the
operation disabled. This will reset automatically,
when cooled, after a few minutes.
Testing to National Standards
The can be used to test to any international standard.
Atable of the standards is given on page 10. To perform these
tests it is necessary to select the correct vessel so that the
specified electrodes and suitable stirring are used. Most standards
require repetitive breakdown tests to be carried out and the
average taken. Note that sometimes the first breakdown result is
omitted from the average calculation. If continuous stirring is not
required the stirrer motor supply can be disconnected when
appropriate by opening the chamber door.
100 ml Oil Test Vessel
The D877 Low Volume Test Vessel is an optional accessory for
use with the Megger® Oil Test Set.
The test vessel must be kept scrupulously clean. Before each
use, dismantle the test vessel, clean and dry, re-assemble, and
set the electrode gap to 2,54 mm (0.100 inch).
Dismantling the vessel
1. Loosen both white locking screws.
2. Hold each electrode in turn, and unscrew, and remove the
sliding arm from each electrode.
3. Remove both electrodes from the vessel.
4. Wash hands, and clean and dry the vessel as specified in
ASTM D.877.
Re-assembling the vessel
1. With clean, uncontaminated hands, place the cleaned
electrodes in the vessel.
2. Position each electrode in turn with the fingers, and screw
in a sliding arm through the vessel into the each electrode
in turn.
3. Centre the electrodes, and set the gap, using the
appropriate gauge.
4. Ensuring that the Viton ‘O’ rings are not fouled, hand tighten
both locking screws.
also supply a range of automatic oil test sets.
These have the parameters of a wide range of specifications
programmed into the instrument for unattended operation.
Electrode
Viton
(Oil resistant)
‘O’ Ring
Locking
screw
Sliding arm
Fill level
The following notes should be read in conjunction with the relevant
oil testing specifications.
Causes of Bad Oil
Transformers and switchgear oil may be rendered unsuitable for
further use due to four main reasons:-
1. Low dielectric strength.
2. High acidity.
3. High sludge content.
4. Excessive free water content.
Low dielectric strength may result from many causes, the most
common of which is foreign particles or fibres and water in
combination. Individually their effect may be relatively small, but
together a contamination of only a few parts in a million can cause
considerable lowering of the breakdown voltage of the oil. High
acidity, sludge and free water should not be tolerated, but will not
necessarily reduce the dielectric strength below acceptable or
specified levels.
Appearance of the Sample
Only an experienced person can judge the condition of an
insulating oil from its appearance, but a general guide may be
obtained from the following observations:-
1. A cloudy appearance may indicate that sludge has
been formed.
2. A dark yellow colour could be a sign of overheating.
3. A blackish colour often results from an arc having taken
place with either carbonisation of the oil or of the
insulation within the equipment.
4. A green colour may be due to copper salts dissolving
in the oil.
Cleanliness of the Apparatus
The necessity for scrupulous cleanliness in the apparatus and
during the process of sampling, cannot be over emphasised. The
measurement of the dielectric strength of the sample is as
dependent on the cleanliness of the test cell and the sampling
apparatus as the condition of the oil itself.
After cleaning the apparatus, it should under no circumstances be
wiped, dried or even handled with a dusty or fibrous cloth as loose
dust or fibres are liable to cause contamination.
Precautions Necessary During Sampling
Since the oil taken from a drain cock will inevitably contain an
excess of settled out solid impurities, and will not necessarily be
representative of the bulk of the oil, it should be run to waste until
clear.
Always run off a quantity of oil into a glass bottle or a test tube for
an initial check on the appearance of the oil. Attention to this point
may prevent contamination of a clean test vessel.
If necessary, thoroughly clean and dry the sample point using a
suitable solvent. Any cloth used should be lint free. Open the
sample cock and drain to waste enough oil to ensure that the
sample cock is fully flushed and the sample is representative of
the bulk of oil to be tested. This quantity will depend upon the size
of the transformer or main container. About 2 litres is usually
suitable.
When sampling from a drum or supply container, the oil should
flow at a steady rate into the test vessel and after being swilled
around the sides it should be discarded. Without altering the rate
of flow of the oil the quantity required for the test should then be
Sampling and Testing Oil for Dielectric Strength
8
9
run off while taking extreme care to prevent the ingress of
atmospheric dust, cloth fibres or moisture.
Do not use a syphon. In the absence of a useable sample point,
use a ‘thief’.
Shield the sample from direct light until ready to be tested.
Turbulence and air bubbles should be avoided when pouring the
oil. Relevant national specifications should be observed.
Particular care to prevent contamination is necessary when a
test is to be made on a bulk delivery tanker, and on all occasions
when a high value of dielectric strength is anticipated. The
procedure suggested above should be rigorously applied.
10
Oil Testing Specifications
Continuous/None
Continuous/None
2.54mm
Specification
11
Output voltage 0 - 60 kV rms
Test Frequency 61,8 Hz
Transformer rating 500 VA
Output disconnection Within 1 ms of detection of breakdown
Power supply 115 V (99 - 132 V), 230 V (198 - 264 V) (switch selectable) 50/60 Hz, 80 VA
Fuses 500 mA (T) HBC to IEC127/5 for 220/240 V supply
1 A (T) HRC to IEC127/5 for 110/120 V supply
Mains power cord fused plug (when applicable): 3 Amp to BS1362
Operation Semi automatic with 0,5, 2 or 3 kV/s selectable rate of voltage rise
Pause function with automatic 1 minute timer for Withstand testing
Display 0,0 - 60,0 kV LED digital display
HV on LED
Ramp rate LED
Measurement Accuracy 2% ± 3 digits
Controls Start, Pause, Stop, Ramp Rate select and Power On/Off switch
Safety The instrument meets the requirements of IEC 61010-1
EMC In accordance with IEC 61326-1
Operational uncertainties Refer to www.megger.com
Temperature Range Operating: 0 to 40 ºC
Storage: -40 to +70 ºC
Humidity Range Operating: 80% RH at 40 ºC
Storage: 93% RH at 40 ºC, 95% RH at 25 ºC
Dimensions 336 mm (13,2 in) (H) x 400 mm (5,7 in) (L) x 235 mm (9,3 in) (D)
Weight 17,5 kg (38,5 lb)
Cleaning Wipe disconnected test set with a clean cloth dampened with soapy water or Isopropyl Alcohol (IPA)
12
Accessories
Supplied with Instrument Part Number
User Guide 6172-120
Spacing Gauge (0,5, 1,0, 1,5, 2,0, 2,5, 2,54 & 4mm) 6132-009
Accessory Pouch 6320-232
Fuses:
2x500 mA (T) for 230 V supply
(fitted to the instrument) 25950-014
2x1A (T) for 115 V supply 25950-004
Mains lead (fit your own plug) 25424-860
Mains lead with US plug 25970-002
Optional Accessories
Carrying Case 6420-106
Test Vessels
0,5 l for IEC156 (spherical electrodes) 6320-233
0,5 l for IEC156 with stirrer (spherical electrodes) 6320-236
0,5 l for VDE0370 (mushroom electrodes) 6320-234
0,5 l for ASTM D1816 with stirrer
(mushroom electrodes) 6320-237
0.15 litre vessel with cylindrical electrodes
for ASTMD 877 6111-356
Spare electrodes for 0.5 L vessels Part Number
Spherical (12,7 mm, 0,5 in. dia.) 6220-484
Mushroom (36 mm dia.) 6220-580
Cylindrical (25,4 mm, 1 in. dia.) 6220-483
Cylindrical with 0,5 mm edge radius 6220-538
Repair and Warranty
13
The instrument circuit contains static sensitive devices, and care
must be taken in handling the printed circuit board. If the
protection of an instrument has been impaired it should not be
used, and be sent for repair by suitably trained and qualified
personnel. The protection is likely to be impaired if, for example,
the instrument shows visible damage, fails to perform the
intended measurements, has been subjected to prolonged
storage under unfavourable conditions, or has been exposed to
severe transport stresses.
New Instruments are Guaranteed for 1 Year from the Date of
Purchase by the User.
Note: Any unauthorized prior repair or adjustment will
automatically invalidate the Warranty.
Instrument Repair and Spare Parts
For service requirements for Megger Instruments contact:
Megger LImited or Megger
Archcliffe Road Valley Forge Corporate Center
Dover 2621 Van Buren Avenue
Kent, CT17 9EN. Norristown, PA 19403
England. U.S.A.
Tel: +44 (0) 1304 502243 Tel: +1 (610) 676-8579
Fax: +44 (0) 1304 207342 Fax: +1 (610) 676-8625
or an approved repair company.
Approved Repair Companies
A number of independent instrument repair companies have been
approved for repair work on most Megger
instruments, using genuine Megger spare parts. Consult the
Appointed Distributor / Agent regarding spare parts, repair
facilities and advice on the best course of action to take.
Returning an Instrument for Repair
If returning an instrument to the manufacturer for repair, it should
be sent freight pre -paid to the appropriate address. A copy of the
Invoice and of the packing note should be sent
simultaneously by airmail to expedite clearance through
Customs. A repair estimate showing freight return and other
charges will be submitted to the sender, if required, before work
on the instrument commences.
Contenu User Guide - p2 Gebrauchsanleitung - s24 Guía del usuario - p34
Avertissments relatifs a la securité 15
Description générale 16
Préparation du boîtier d'essai d'huile 17
Fonctionnement 17
Echantillonnage d'huile 17
Préparation du récipient d'huile 17
Panneau de commande 18
Essai de claquage 18
Essai de résistance (Epreuve) 18
Témoins d’affichage 19
Essais selon les normes nationales 19
Echantillonnage et essai de résistance
diélectrique des huiles 20
Spécifications essai des huiles 22
Réparation et Garantie 23
14
15
GAVERTISSEMENTS RELATIFS A LA SECURITE
Les avertissements et précautions concernant la sécurité doivent être lus et compris avant de commencer
àutiliser l'instrument. Ils doivent être observés pendant l'emploi.
Le boîtier d'essai d'huile doit être correctement mis à la masse.
La chambre d'essai doit être conservée scrupuleusement propre.
Le boîtier d'essai ne doit pas être utilisé si le couvercle de la chambre d'essai est fissuré ou endommagé
d'une manière quelconque, auquel cas il doit être renvoyé à un agent autorisé pour réparation.
Les charnières de la porte de la chambre sont équipées de ressorts de mise à la masse. Ces ressorts ne
doivent pas être endommagés ou rouillés.
Le type correct de récipient d'essai d'huile doit être utilisé, et positionné correctement sur les guignols de
support avant d'effectuer un essai quelconque.
Les fusibles de remplacement doivent être du type et de la puissance corrects.
REMARQUE
CONFIEZ LUTILISATION DE CET INSTRUMENT A UN PERSONNEL QUALIFIE ET COMPETENT.
Symboles utilisés sur cet instrument:
GAttention: Consultez les notes jointes.
FRisque de choc électrique
cCet équipement respecte les directives en viguer de l’UE
Le modèle OTS60SX est un boîtier d'essai de résistance
diélectrique des huiles, léger et semi-automatique. L'instrument
convient à un usage en service protégé et peut être alimenté à partir
de toute une gamme de tensions secteur. La puissance de sortie
maximale de 60 kV permet d'effectuer des mesures sur des huiles de
diverses installations électriques, y compris les transformateurs, les
disjoncteurs et autres équipements. L'exploitation du boîtier d'essai
est extrêmement simple, et les résultats sont affichés sur un
affichage clair à LED. Une sélection de récipients permet de
configurer l'instrument pour une variété d'essais.
Le fonctionnement semi-automatique permet la réalisation de tous
les types d'essais. Une minuterie automatique d'une minute se
déclenche lorsque la haute tension est mise en pause pour les essais
d'épreuve (résistance). Des échantillons d'huile peuvent aussi être
testés selon des spécifications de claquage en employant une
séquence appropriée d'essais.
L'instrument est monté dans un boîtier robuste en tôle d'acier, muni
de poignées pour faciliter le transport. Une pochette est prévue sur le
côté de l'instrument pour le rangement des accessoires tels que le
cordon d'alimentation, des électrodes supplémentaires et la jauge
d'espacement. Une valise de transport avec bandoulière est
disponible en option.
Des récipients convenant aux essais d'une gamme étendue de
spécifications nationales sont disponibles. Trois types d'électrodes
sont disponibles: sphérique (type IEC), à champignon (VDE/ASTM
D1816) et cylindrique (ASTM D877). Dans certains récipients,
l'échantillon d'huile peut être agité par une turbine motorisée. Le
récipient est situé en haut de l'instrument et est couvert par une porte
en polycarbonate transparent munie d'un grillage de manière à ce
que le claquage de l'huile puisse être constaté. Une sécurité assure
la coupure de la haute tension lorsque la porte de la chambre est
ouverte.
Description générale
16
Selecteur
de tension
Connecteur
de ligne
CEI
Fusibles
Panneau de alimentation secteur
Préparation du boîtier d'essai d'huile
17
Vérifier que l'instrument est correctement mis à la masse. Le
boîtier d'essai doit être branché dans une prise femelle ayant un
conducteur de mise à la masse (terre).
Avant de brancher l'instrument sur le secteur, vérifier que le
sélecteur de tension, situé sur le panneau d'entrée
d'alimentation, est bien réglé à la tension requise. En outre, un
fusible de valeur adaptée à la tension d'alimentation (voir
spécifications) doit être installé. Les porte-fusibles sont situés à
côté de l'entrée de l'alimentation.
Cordon électrique de recharge de batterie
Si la prise de ce cordon ne peut pas s'enficher dans l'une de
vos prises murales, vous ne devez pas utiliser d'adaptateur.
Vous devez employer un cordon électrique de type approprié
ou, le cas échéant, retirer la prise du cordon en tranchant ce
dernier avant d'y fixer une prise de type approprié.
Ce cordon respecte le code de couleurs suivant:
Terre (masse) jaune/vert
Neutre bleu
Sous tension marron
Si vous utilisez une prise protégée par un fusible, vérifiez qu'il
s'agit d'un fusible de 3 A conforme à la norme BS 1362.
Remarque: après avoir retiré du cordon la prise électrique en
la tranchant, mettez cette dernière au rebut car une prise
comportant des conducteurs dénudés présente un danger en
cas de branchement sur une prise murale sous tension.
Fonctionnement
Echantillonnage d'huile
La spécification d'essai appropriée fournira des conseils quant
aux meilleures méthodes à employer pour le prélèvement d'un
échantillon d'huile. La contamination de l'huile peut conduire à
des résultats erronés. Laisser toujours couler une petite quantité
d'huile avant de prélever l'échantillon et veiller à ce que
l'écoulement d'huile soit constant.
Préparation du récipient d'huile
Divers récipients optionnels sont disponibles pour les essais
selon les spécifications nationales (voir tableau et ‘Accessoires‘).
Certaines spécifications exigent l'agitation de l'huile ainsi que
différentes formes d'électrodes. Lorsque le récipient correct est
choisi pour l'essai requis, l'écartement des électrodes doit être
ajusté. Ce réglage est effectué à l'aide de l'écrou fileté sur le côté
du récipient. Un jeu de jauges d’espacement, de 0,5 mm
chacune, est fourni pour la verification de l’ecartement des
electrodes. Les jauges permettent de mesurer un ecartement
entre 1 mm et 4mm.
Nettoyer le récipient conformément aux instructions fournies
dans les spécifications d'essai concernées. Tout en prévoyant un
espace pour l'installation du couvercle, verser l'échantillon
d'huile dans le récipient. La quantité d'huile requise sera fonction
du type de couvercle utilisé. Placer le récipient dans la chambre
d'essai, en branchant, le cas échéant, le câble du moteur, et
fermer soigneusement la porte de manière à actionner
l'interrupteur de sécurité.
Panneau de commande
Le panneau de commande comporte l'interrupteur de mise
sous/hors tension, l'affichage haute tension, un sélecteur de taux
d'augmentation de tension et trois touches de commande.
L'exploitation de l'instrument est extrêmement simple. Lorsque
l'instrument est sous tension, l'interrupteur marche/arrêt s'allume
et le code de version du logiciel clignote sur l'affichage.
Essai de claquage
Le taux d'augmentation de tension approprié à l'essai à exécuter
doit être sélectionné en utilisant la touche 's'. La touche marche
' ' appliquera la haute tension. Le voyant rouge HT s'allumera
pour indiquer que la haute tension est appliquée. La tension
augmentera alors au taux sélectionné, et la valeur
correspondante sera affichée. L'instrument détecte le claquage
de l'huile et coupe immédiatement la haute tension. La valeur de
la tension de claquage reste affichée sur l'instrument jusqu'au
début de l'essai suivant. Si le claquage ne se produit pas, la
haute tension monte jusqu'à 60 kV. En ce point, la haute tension
est coupée et cette valeur maximum reste affichée.
Un essai peut être interrompu à tout moment en appuyant sur la
touche STOP n'. Cette touche coupe l'application de la haute
tension à l'échantillon d'huile, et l'affichage revient à 0,0 kV.
Essai de résistance (Epreuve)
Un essai de résistance (Epreuve) peut être effectué en appuyant
sur la touche de PAUSE ‘. Cette action maintient la haute
tension à la valeur courante pendant une minute, après quoi la
haute tension continuera à augmenter jusqu'à ce que :
1) l'huile claque
Fonctionnement (suite)
18
Interrupteur
Sous/Hors
tension
Voyant HT
en service
Sélection
de taux
d’augmentation
de tension
Commandes
principales
Affichage
LED
s
Panneau de commande
Essais selon les normes nationales
L'appareil peut être utilisé pour effectuer des essais
selon n'importe quelle norme internationale. Les normes sont
indiquées au tableau de la page 22. L'exécution de ces essais
exige la sélection du récipient correct de manière à ce que les
électrodes spécifiées et l'agitation appropriée puissent être
utilisées. La majorité des normes spécifient des essais de
claquage répétitifs, puis le calcul de la moyenne. Il convient de
noter que parfois le premier résultat de claquage est omis du
calcul. Si une agitation continue est requise, l'alimentation du
moteur de l'agitateur peut être déconnectée en ouvrant la porte
de la chambre.
fournit également toute une gamme de
boîtiers d'essai automatiques. Les paramètres d'une gamme
étendue de spécifications sont programmés dans l'appareil pour
assurer le fonctionnement automatique.
19
2) le bouton STOP n‘ (Arrêt) est pressé
3) la tension de sortie atteint la valeur maximum du boîtier
d'essai.
Pendant l'essai de résistance, le taux d'augmentation peut être
modifié pendant la pause. Ceci permet de régler avec précision
la valeur d'essai de résistance en ajustant le taux
d'augmentation à sa valeur la plus basse.
Témoins d’affichage
Porte ouverte
En cas de surchauffe de l'instrument, les mots
'' (trop) et ' ' (chaud) sont affichés alter-
nativement et l'opération est abandonnée.
L'instrument se remet à zéro automatiquement
une fois qu'il s'est refroidi après quelques minutes.
Echantillonnage et essai de résistance diélectrique des huiles
20
Ces notes doivent être lues conjointement aux spécifications
d’essai d’huile appropriées.
Causes des mauvaises huiles
L’huile de transformateur et d’appareillage de commutation peut
être devenir inutilisable à cause de quatre raisons :
1. Faible résistance diélectrique
2. Forte acidité
3. Haute teneur en boues
4. Teneur excessive en eau libre
La faible résistance diélectrique a de nombreuses causes, la plus
courante étant la présence de particules étrangères ou d’une
combinaison de fibres et d’eau. Individuellement, leurs effets
peuvent être relativement faibles, mais ensemble une
contamination de quelques parties par million peut causer une
réduction considérable de la tension de claquage de l’huile. Une
forte acidité, les boues et l’eau libre ne doivent pas être tolérées,
mais ne réduiront pas la résistance diélectrique en dessous des
niveaux acceptables ou spécifiés.
Aspect de l’échantillon
Seule une personne compétente peut juger l’état d’une huile
d’isolement à partir de son aspect, mais une indication générale
peut être déduite des observations suivantes :
1. Un aspect nuageux peut indiquer la formation de boues.
2. Une couleur jaune sombre peut indiquer un échauffement
excessif.
3. Une couleur noirâtre est souvent le résultat d’un arc avec soit
carbonisation de l’huile ou de l’isolement à l’intérieur du
matériel.
4. Une teinte verdâtre peut être causée par la dissolution
desels de cuivre dans l’huile.
Propreté des appareils
La nécessité d’une propreté scrupuleuse des appareils et pendant
l’échantillonnage ne peut pas être suffisamment soulignée. La
mesure de la résistance diélectrique de l’échantillon dépend tout
aussi bien de la propreté de la cellule d’essai que de l’état même
de l’huile.
Une fois nettoyé, l’appareil ne doit jamais être essuyé, séché ou
même manipulé avec un chiffon poussiéreux ou fibreux étant
donné que les poussières ou les fibres libres peuvent causer une
contamination.
Précautions à prendre pendant l’échantillonnage
Etant donné que l’huile prélevée à partir d’un robinet de
prélèvement contiendra inévitablement une forte teneur
d’impuretés solides décantées, et ne sera pas nécessairement
représentative de la masse de l’huile, il convient de laisser couler
l’huile dans un bac de déchet jusqu’à ce qu’elle soit claire.
Laisser toujours couler une certaine quantité d’huile dans une
bouteille en verre ou dans une éprouvette pour effectuer une
inspection initiale de l’aspect de l’huile. Cette observation peut
éviter la contamination d’un récipient d’essai propre.
Si nécessaire, nettoyer et sécher le point d’échantillonnage en
utilisant un solvant approprié. Tout chiffon utilisé doit être du type
non pelucheux. Ouvrir le robinet de prélèvement et laisser couler
l’huile pour s’assurer que le robinet est totalement rincé et que
l’échantillon est représentatif de la masse d’huile à tester. Cette
quantité est fonction de la taille du transformateur ou du récipient
principal. En règle générale, 2 litres suffisent.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44

Megger OTS60SX Le manuel du propriétaire

Taper
Le manuel du propriétaire