Omron Healthcare Automatic Blood Pressure Monitor Manuel utilisateur

Taper
Manuel utilisateur
Compressor Nebulizer
MODEL NE-C28
INSTRUCTION MANUAL
ENGLISH
3
5
®
FRANÇAIS ESPAÑOL
PORTUGUÊS
Caution:
Federal (USA) la
w restricts this de
vice to sale by or on the
order of a ph
ysician or licensed healthcare practitioner
NE-C28- IM Final 6-12-06 6/12/06 3:25 PM Page 1
2
Before Using the Device
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
Safety Information
Operating the Device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
Risk of Electrical Shock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
Maintenance and Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
Know Your Unit
Main Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
Components and Replacement Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
Operating Instructions
Preparing the Nebulizer for Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
Attaching the Air Tube . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12
Using the Device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13
Care and Maintenance
Cleaning after Each Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16
Daily Disinfecting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17
Changing the Air Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .19
Caring for the Device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20
Troubleshooting
Troubleshooting Guide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22
Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .23
FCC Statement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .24
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25
The illustrations shown in this instruction manual are image drawings and may differ
slightly from the appearance of the actual product.
CONTENTS
NE-C28- IM Final 6-12-06 6/12/06 3:25 PM Page 1
3
Thank you for purchasing the OMRON
®
CompAir
®
NE-C28 Compressor Nebulizer.
INTRODUCTION
Fill in for future reference.
DATE PURCHASED: ________________
SERIAL NUMBER: _________________
Staple your purchase receipt here
This product was developed for the successful treatment of asthma, allergies and other
respiratory disorders. The compressor forces air to the nebulizer. When the air enters the
nebulizer, it converts the prescribed medication into an aerosol of microscopic droplets
that can easily be inhaled.
Your NE-C28 CompAir
®
comes with the following components:
Compressor (Main Unit)
Nebulizer Kit
Mouthpiece
Child Mask
Air Tube
Carrying Bag
Air Filters (Package of 5)
Air Filter Cover
Instruction Manual (English, Spanish, Portuguese, French)
User Registration Card
The following are optional accessories sold separately:
Adult Inhalation Mask
Nosepiece
SAVE THESE INSTRUCTIONS
Before Using the Unit
The CompAir
®
is a prescription medical de
vice. Operate this device only
as instructed by your physician and/or licensed healthcare practitioner.
NE-C28- IM Final 6-12-06 6/12/06 3:25 PM Page 2
4
SAFETY INFORMATION
To assure the correct use of the product basic safety measures should always be followed
including the warnings and cautions listed in this instruction manual.
OPERATING THE DEVICE
For type, dose, and regime of medication follow the instructions of your physician
or licensed healthcare practitioner.
Pentamidine is not an approved medication for use with this device.
Do not block the ventilation slots located on the sides of the device. Do not place
the device where the ventilation slots may be obstructed during operation. This
could result in the compressor overheating or malfunctioning.
Do not cover the compressor with a blanket, towel, or any other type of cover
during use. This could result in the compressor overheating or malfunctioning.
Do not use the device where the device may be exposed to flammable gas or vapors.
Do not use tap or mineral water in the nebulizer for nebulizing purposes.
Clean and disinfect the nebulizer kit, mouthpiece, and optional nosepiece or masks
before using them for the first time after purchase.
If the device has not been used for a long period of time clean and disinfect the
nebulizer kit, mouthpiece, and optional nosepiece or masks before using them.
Always dispose of any remaining medication in the medication cup after each use.
Use fresh medication each time you use the device.
Provide close supervision when this device is used by, on, or near infants, children
or compromised individuals.
Do not insert an
y object into the compressor
.
Make sure the air filter is clean. If the air filter has changed color or has not been
used for 60 days, replace the filter.
WARNING
Indicates a potentially hazardous situation which, if not avoided,
could result in death or serious injury.
CAUTION
Indicates a potentially hazardous situation which, if not avoided,
may result in minor or moderate injury to the user or patient or
damage to the equipment or other property.
SAFETY SYMBOLS USED IN THIS INSTRUCTION MANUAL
NE-C28- IM Final 6-12-06 6/12/06 3:25 PM Page 3
5
OPERATING THE DEVICE (continued)
Make sure the nebulizer kit is correctly assembled, the air filter is properly
installed, and the air tube is correctly connected to the compressor and the
nebulizer kit. Air may leak from the air tube during use if not securely connected.
Inspect the compressor (main unit) and the nebulizer parts each time before using
the device. Make sure no parts are damaged, the nozzle and air tube are not
blocked and the compressor operates normally.
Do not use the device if the air tube is bent.
Do not block the air filter cover.
Do not alter the baffle, the nozzle in the medication cup or any part of the
nebulizer kit.
Do not add more than 7ml of medication to the medication cup.
Do not operate the device at temperatures greater than +104˚F (+40˚C).
Do not tilt the nebulizer kit so the angle of the kit is greater than 45˚. Medication
may flow into the mouth.
Do not shake the nebulizer kit while using the device.
Do not subject the compressor or any of the components to strong shocks, such as
dropping on the floor.
This device is approved for human use only.
Do not disassemble or attempt to repair the device or components.
Operate the device only as intended. Do not use the device for any other purpose.
Dispose of the device, components and optional accessories according to
applicable local regulations. Unlawful disposal may cause environmental pollution.
Use only Omron authorized parts and accessories. Parts and accessories not
approved for use with the device may damage the unit.
Changes or modification not approved by Omron Healthcare will void the user warranty.
SAFETY INFORMATION
NE-C28- IM Final 6-12-06 6/12/06 3:25 PM Page 4
6
SAFETY INFORMATION
RISK OF ELECTRICAL SHOCK
Do not plug or unplug the power cord into the electrical outlet with wet hands.
Do not immerse the compressor (main unit) in water or other liquid.
Do not spill water or other liquids on the compressor and power cord.
The compressor and power cord are not waterproof. If liquid spills on these parts,
immediately unplug the power cord and wipe off the liquid with gauze or other
soft absorbent material.
Do not use or store the device in humid locations, such as a bathroom. Use the
device within the operating temperature and humidity.
Do not use a cellular phone near the device. It may result in an operational failure.
Do not overload power outlets. Plug the device into an outlet with the appropriate voltage.
Do not use extension cords. Plug the power cord directly into the electrical outlet.
Unplug the power cord from the electrical outlet after using the device.
Unplug the power cord from the electrical outlet before cleaning the device.
MAINTENANCE AND STORAGE
Keep the device out of the reach of unsupervised infants and children. The device
may contain small parts that can be swallowed.
Do not leave the cleaning solution in the nebulizer parts. Rinse the nebulizer parts
with clean hot tap water after disinfecting.
Wash the nebulizer parts after each use. Dry the parts immediately after washing.
Do not store the air tube with moisture or medication remaining in the air tube.
This could result in infection as a result of bacteria.
Store the device and the components in a clean, safe location.
Do not carry or leave the nebulizer with medication in the medication cup.
Do not place or attempt to dry the device, components, or any of the nebulizer
parts in a microwave oven.
Do not wrap the power cord around the compressor (main unit).
NE-C28- IM Final 6-12-06 6/12/06 3:25 PM Page 5
7
3
5
KNOW YOUR UNIT
Cap
Inhalation Air Inlet
Inhalation Top
Baf
f
le
Medication Cup
Nozzle
Air T
ube Connector
Main Unit:
Compressor
Components:
Nebulizer Kit
3
5
Power
Switch
Nebulizing Kit
Air Plug
Air Plug
Air Connect
or
Ventilation
Slots
Nebulizer Kit
Holder
Air Filter Cover
(*Air Filter Inside)
Front View
Back View
Power Cord
Ventilation
Slots
Power Plug
Mouthpiece
Child Mask
Air Tube
NE-C28- IM Final 6-12-06 6/12/06 3:25 PM Page 6
8
KNOW YOUR UNIT
Components:
Optional Accessories:
User
Registration
Card
Air Tube
with attached Air Plugs
Instruction Manual
Adult Mask
Nosepiece
Carrying Bag
Air Filters
Package of 5
Air Filter Cover
Accessories and Replacement Parts
Model No.
Adult Mask
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C920
Air Filters (Package of 5) . . . . . . . . . . . . . . . . . . . . . . . . . . .C28FL
Air Tube with attached Air Plugs . . . . . . . . . . . . . . . . . . . .C915
Carrying Bag . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C28BAG
Child Mask
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
C922
Nebulizer Kit Set . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C900
Includes Nebulizer Kit, Mouthpiece and Air Tube
Nosepiece . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C911
Mouthpiece . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C910
Compressor Nebulizer
MODEL NE-C28
3
5
®
Caution:
Federal (USA) law restricts this
NE-C28- IM Final 6-12-06 6/12/06 3:25 PM Page 7
9
PREPARING THE NEBULIZER FOR USE
WARNING
Clean and disinfect the nebulizer kit, mouthpiece, and optional nosepiece or masks
before using them for the first time after purchase.
WARNING
If the device has not been used for a long period of time clean and disinfect the
nebulizer kit, mouthpiece, and optional nosepiece or masks before using them.
For directions on cleaning and disinfecting refer to page 16 and page 17.
CAUTION
Make sure the air filter is clean. If the air filter has changed color or has not been
used for 60 days, replace the filter.
For directions on changing the air filter refer to page 19.
1. Turn the power switch off.
Make sure the power switch is in the O position.
2. Plug the power plug into the electrical outlet.
WARNING
Do not plug or unplug the power cord into the electrical
outlet with wet hands.
CAUTION
Do not overload power outlets. Plug the device into the
appropriate voltage outlet.
CAUTION
Do not use extension cords. Plug the power cord directly
into the electrical outlet.
3. Lift up the mouthpiece and the cap to remove them
from the nebulizer kit.
4. Rotate the inhalation top counterclockwise (A) and lift
(B) to remove the inhalation top from the medication cup.
NE-C28- IM Final 6-12-06 6/12/06 3:25 PM Page 8
PREPARING THE NEBULIZER FOR USE
5. Lift up the baffle and remove the baffle from the medication cup.
CAUTION
Do not alter the baffle, the nozzle in the medication cup or
any part of the nebulizer kit.
6. Add the correct amount of prescribed medication to the
medication cup.
WARNING
Pentamidine is not an approved medication for use with this device.
WARNING
For type, dose, and regime of medication follow the
instructions of your physician or licensed healthcare practitioner.
WARNING
Do not spill water or other liquids on the compressor and
power cord. The compressor and power cord are not
waterproof. If liquid spills on these parts, immediately
unplug the power cord and wipe off the liquid with gauze
or other soft absorbent material.
WARNING
Do not use tap or mineral water in the nebulizer for
nebulizing purposes.
CAUTION
Do not add more than 7ml of medication to the medication cup.
7. Place the baffle back into the medication cup. Make sure
the baffle is correctly positioned over the nozzle in the
medication cup.
8. Place the inhalation top on the medication cup o
ver the
baffle. The top of the baffle should be centered in the
inhalation top.
Align the protrusion on the inhalation top with the
indentation on the medication cup as illustrated (A).
Turn the inhalation top clockwise until securely closed (B).
max
5
3
10
NE-C28- IM Final 6-12-06 6/12/06 3:25 PM Page 9
PREPARING THE NEBULIZER FOR USE
9. Attach the desired inhalation accessory.
NOTE: Mouthpiece and Nosepiece
Place the cap on the medication cup when using the mouthpiece or
nosepiece. Using the cap with the mouthpiece or the nosepiece will
reduce the amount of medication that is discharged into the air.
USING THE MOUTHPIECE
Place the cap on the medication cup over the inhalation
air inlet (A).
Attach the mouthpiece (B).
USING THE NOSEPIECE
Place the cap on the medication cup over the inhalation
air inlet (A).
Attach the nosepiece (B).
NOTE: Child Mask and Adult Mask
Do not place the cap on the medication cup when using
the Child Mask or the Adult Mask.
USING THE CHILD MASK
Attach the mask to the inhalation top.
USING THE ADULT MASK
Attach the mask to the inhalation top.
A
B
11
NE-C28- IM Final 6-12-06 6/12/06 3:25 PM Page 10
12
ATTACHING THE AIR TUBE
1. Push the Air Plug on one end of the Air Tube onto the Air Connector
on the front side of the compressor (A).
2. Push the Air Plug on the other end of the Air Tube onto the Air Tube
Connector on the bottom of the Nebulizer Kit (B).
NOTE: Keep the nebulizer kit upright when attaching the air tube. Medication may
spill out of the medication cup.
CAUTION
Make sure the nebulizer kit is correctly assembled, the air filter is properly
installed, and the air tube is correctly connected to the compressor and the
nebulizer kit. Air may leak from the air tube during use if not securely
connected.
CA
UTION
Do not use the device if the air tube is bent.
3. Use the Nebulizer Kit Holder as a temporary
holder for the nebulizer kit.
NE-C28- IM Final 6-12-06 6/12/06 3:26 PM Page 11
13
USING THE DEVICE
WARNING
Do not use the device where the device may be exposed to flammable gas or vapors.
WARNING
Do not use or store the device in humid locations, such as a bathroom. Use the
device within the operating temperature and humidity.
CAUTION
Provide close supervision when this device is used by, on, or near infants,
children or compromised individuals.
CAUTION
Do not use a cellular phone near the device. It may result in an operational failure.
CAUTION
Inspect the compressor (main unit) and the nebulizer parts each time before
using the device. Make sure no parts are damaged, the nozzle and air tube are
not blocked and the compressor operates normally.
1. Hold the nebulizer kit as illustrated on the right.
CAUTION
Do not tilt the nebulizer kit so the angle of the
kit is greater than 45˚. Medication may flow into
the mouth.
CAUTION
Do not shake the nebulizer kit while using the device.
2. Turn the power switch on.
The power switch is in the ( ) position.
NOTE: The compressor starts. Nebulization begins
and aerosol is generated.
Turn the power switch to the off (O) position to stop nebulization.
max
5
3
max
5
3
NE-C28- IM Final 6-12-06 6/12/06 3:26 PM Page 12
14
USING THE DEVICE
WARNING
Do not block the ventilation slots located on the sides of the device. Do not
place the device where the ventilation slots may be obstructed during
operation. This could result in the compressor overheating or malfunctioning.
WARNING
Do not cover the compressor with a blanket, towel, or any other type of cover
during use. This could result in the compressor overheating or malfunctioning.
CAUTION
Do not block the air filter cover.
3. Inhale medication as instructed by your physician or licensed
healthcare practitioner.
USING THE MOUTHPIECE
Insert the mouthpiece into your mouth. Inhale
medication breathing normally. Exhale normally
through the mouthpiece.
USING THE NOSEPIECE
Inhale the medication through the nosepiece as
illustrated. Exhale through the mouth.
NOTE: Child Mask and Adult Mask
If the amount of aerosol is too much, place the cap
on the medication cup over the inhalation air inlet.
USING THE CHILD MASK or THE ADULT MASK
Place the mask over the nose and mouth. Pull the elastic
strap over the head. Gently pull on the strap to securely
hold the mask o
ver the nose and mouth. Inhale the
medication. Exhale normally through the mask.
4. Complete the treatment.
W
ARNING
F
or type, dose, and regime of medication follow the instructions of your
physician or licensed healthcare practitioner.
m
a
x
5
3
m
a
x
5
3
ma
x
5
3
ma
x
5
3
max
5
3
max
5
3
NE-C28- IM Final 6-12-06 6/12/06 3:26 PM Page 13
15
USING THE DEVICE
5. Turn the power switch off.
The power switch is in the position.
The compressor turns off and nebulization stops.
6. Disconnect the Air Tube from the Nebulizer Kit. Hold the Air Plug and
gently pull down to disconnect the Air Plug from the Air Tube
Connector on the bottom of the nebulizer kit.
7. Check the Air Tube. No condensation or moisture should remain in the
Air Tube.
WARNING
Do not store the air tube with moisture or medication remaining in the air tube.
This could result in infection as a result of bacteria.
If any condensation or moisture remains in the air tube, remove the
moisture from the air tube. Follow the directions below:
1. Make sure the Air Tube is still connected to the Air connector
on the front side of the compressor.
2. Turn the power switch on. The power switch is in the ( )
position. The compressor will start and pump air through the air
tube to expel the moisture.
3. Turn the power switch off.
8. Disconnect the Air Tube from the compressor. Hold
the Air Plug and gently pull the Air Plug off the Air
Connector on the front side of the compressor.
9. Unplug the device from the electrical outlet.
CAUTION
Unplug the po
wer cord from the electrical outlet after
using the device.
NE-C28- IM Final 6-12-06 6/12/06 3:26 PM Page 14
16
CLEANING AFTER EACH USE
Following the cleaning instructions after each use will prevent any
remaining medication in the bottle from drying resulting in the device not
nebulizing effectively and will help prevent infections.
WARNING
Wash the nebulizer parts after each use. Dry the parts immediately after washing.
1. Remove the inhalation accessory (mouthpiece, mask, or nosepiece) from
the nebulizer kit.
2. Disconnect the air tubing from the nebulizer kit.
3. Gently twist the inhalation top counterclockwise and lift to separate the
nebulizer kit into two sections.
4. Remove the baffle.
5. Discard remaining medication in the medication cup.
WARNING
Always dispose of any remaining medication in the
medication cup after each use. Use fresh medication
each time you use the device.
6. Rinse all the parts of the nebulizer kit and the mouthpiece or nosepiece
with hot tap water. You may also wash these parts in a dishwasher.
Remove the parts after the wash and rinse cycle is completed.
DO NOT wash the mask in a dishwasher.
CAUTION
Place the nebulizer pieces in a basket recommended for dishwasher use. Place
the basket on the top shelf to avoid any damage to the nebulizer kit.
7. Rinse the mask with hot tap water.
8. Hand dry or air dry in a clean
environment using a soft, clean,
lint-free cloth.
9. Assemble the nebulizer kit and store the
nebulizer kit in a dry, sealed bag.
CAUTION
The neb
ulizer kit should be replaced e
v
ery 6 months.
m
a
x
5
3
3
5
NE-C28- IM Final 6-12-06 6/12/06 3:26 PM Page 15
17
You can disinfect daily by soaking the parts in a vinegar solution or using
a commercially available medical disinfectant. If your physician or respiratory
therapist specifies a different cleaning procedure follow their instructions.
Disinfect the nebulizer kit and the mask, nosepiece or mouthpiece after the last
treatment of the day.
VINEGAR METHOD
1. Make a solution of 1 part white vinegar and 3 parts distilled water.
Make enough solution to submerge the parts.
2. Remove the inhalation accessory (mouthpiece, mask, or nosepiece)
from the nebulizer kit.
3. Disconnect the air tubing from the nebulizer kit.
4. Gently twist the inhalation top counterclockwise and lift to separate
the nebulizer kit into two sections.
5. Remove the baffle.
6. Wash all the parts of the nebulizer kit, the mouthpiece, nosepiece or
mask in warm water and a mild detergent. Rinse in hot tap water.
You may also wash the nebulizer kit, mouthpiece or nosepiece in a
dishwasher. DO NOT wash the mask in a dishwasher.
CAUTION
Place the nebulizer pieces in a basket recommended for dishwasher use.
Place the basket on the top shelf to avoid any damage to the nebulizer kit.
7. Submerge the parts in the white vinegar and distilled water solution
for 30 minutes.
8. Remove the parts and discard the solution. Rinse the parts with hot
tap water.
DAILY DISINFECTING
NE-C28- IM Final 6-12-06 6/12/06 3:26 PM Page 16
18
DAILY DISINFECTING
9. Hand dry or air dry in a clean environment using a soft, clean,
lint-free cloth.
10. Assemble the nebulizer kit and store the nebulizer kit in a dry, sealed bag.
WARNING
Store the device and the components in a clean, safe location.
MEDICAL DISINFECTANT METHOD
Consult your physician or respiratory therapist regarding the use of
medical disinfectant cleaners. Carefully follow the instructions supplied
by the manufacturer of the medical disinfectant.
NE-C28- IM Final 6-12-06 6/12/06 3:26 PM Page 17
19
Change the air filter every 60 days even if the air filter does not appear
dirty. If the air filter appears dirty, or if water or medication has spilled
on the air filter replace with a new air filter immediately.
1. Pull the air filter cover to remove from the
front side of the compressor.
2. Remove the dirty filter with a small pointed
object.
CAUTION
Do not attempt to wash or clean the air filter. Damp
air filters can cause blockages. Do not substitute
cotton or any other material for the air filter.
CAUTION
Wash the air filter cover regularly to prevent any
blockage in the cover. Do not boil. Make sure the
cover is dry before inserting the new air filter.
3. Insert a new air filter into the air filter cover.
CAUTION
Before inserting the new air filter make sure the air filter is clean and free of
dust. Do not operate the device without the air filter. Use only the Omron air
filter designed for this device.
4. Put the air filter cover back on the compressor.
CHANGING THE AIR FILTER
NE-C28- IM Final 6-12-06 6/12/06 3:26 PM Page 18
20
To keep your device in the best condition and protect the unit from damage follow
these directions:
CAUTION
Do not subject the compressor or any of the components to strong shocks, such as
dropping on the floor.
CAUTION
Use only Omron authorized parts and accessories.
CLEANING THE COMPRESSOR
Clean the casing of the main unit by using
a soft cloth moistened with water or a mild
detergent. Do not use abrasive cleaners.
Dry the casing immediately using a soft
clean cloth.
WARNING
Do not immerse the compressor (main unit) in water or other liquid.
CAUTION
Unplug the power cord from the electrical outlet before cleaning the device.
CAUTION
Do not insert any object into the compressor.
STORING THE DEVICE
1. Put the nebulizer kit and the inhalation accessory (mouthpiece,
nosepiece or mask) in a dry, sealed bag. Place the bag in the storage
pocket of the carrying bag.
2. Coil the air tubing and place in the storage pocket.
3. Place the compressor in the carrying bag.
CAUTION
Do not wrap the power cord around the compressor (main unit).
CARING FOR THE DEVICE
Mild Detergent
NE-C28- IM Final 6-12-06 6/12/06 3:26 PM Page 19
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112