Seca 777 Manuel utilisateur

Taper
Manuel utilisateur
seca 777
Instructions for use and warranty . . . . . . . . . . . . . . . . . . . . 3
Mode d'emploi et garantie . . . . . . . . . . . . . . . . . . . . . . . . . 32
Instrucciones de uso y garantía . . . . . . . . . . . . . . . . . . . . 61
US
FR
ES
Contents • 3
English
17-10-07-651-100a/07-2019S
English
CONTENTS
1. Description of device . . . . . . . . . . . . . . . . . . . . . . . . 4
1.1 Intended use . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.2 Description of function . . . . . . . . . . . . . . . . . . . . 4
Measuring station . . . . . . . . . . . . . . . . . . . . . . 4
USB interface . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.3 User qualification . . . . . . . . . . . . . . . . . . . . . . . . 4
Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2. Safety information . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2.1 Safety precautions in these instructions for
use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2.2 Basic safety precautions . . . . . . . . . . . . . . . . . . . 5
Handling the device . . . . . . . . . . . . . . . . . . . . . 5
Preventing electric shock . . . . . . . . . . . . . . . . . 6
Preventing injuries and infections . . . . . . . . . . . 6
Preventing damage to device . . . . . . . . . . . . . . 7
Handling measuring results . . . . . . . . . . . . . . . 7
Handling packaging material . . . . . . . . . . . . . . 8
Handling batteries and rechargeable
batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
3. Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
3.1 View of device . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
3.2 Display unit controls . . . . . . . . . . . . . . . . . . . . . 10
3.3 Symbols in the display . . . . . . . . . . . . . . . . . . . 11
3.4 Markings on the device and on the type
plate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
3.5 Markings on the packaging . . . . . . . . . . . . . . . . 12
4. Starting up device . . . . . . . . . . . . . . . . . . . . . . . . . . 13
4.1 Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . 13
4.2 Assembling the device . . . . . . . . . . . . . . . . . . . 13
Tools required . . . . . . . . . . . . . . . . . . . . . . . . 14
Fitting the measuring rod . . . . . . . . . . . . . . . . 14
Fitting the column . . . . . . . . . . . . . . . . . . . . . 15
Fitting the display head . . . . . . . . . . . . . . . . . 16
Switching the positions of the display unit
and the battery compartment . . . . . . . . . . . . 17
Setting up the device . . . . . . . . . . . . . . . . . . . 19
Disinfecting the device . . . . . . . . . . . . . . . . . . 19
4.3 Transporting the device . . . . . . . . . . . . . . . . . . 19
4.4 Establishing the power supply . . . . . . . . . . . . . . 20
Inserting batteries . . . . . . . . . . . . . . . . . . . . . 20
Connecting the plug-in power supply unit
(optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
5. Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
5.1 Switching the device on and off . . . . . . . . . . . . 21
Switching on the device . . . . . . . . . . . . . . . . . 21
Switching off the device . . . . . . . . . . . . . . . . . 21
5.2 Switching units of measurement (variant-
dependent) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
5.3 Measuring weight . . . . . . . . . . . . . . . . . . . . . . . 22
Weighing . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Taring off additional weight (Tare) . . . . . . . . . . 22
Displaying weight permanently (Hold) . . . . . . . 23
5.4 Measuring height . . . . . . . . . . . . . . . . . . . . . . . 23
Measuring heights < 1.22 m . . . . . . . . . . . . . 23
Measuring heights > 1.22 m . . . . . . . . . . . . . 24
5.5 Calculating and displaying body mass index
(BMI) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
6. Hygiene treatment . . . . . . . . . . . . . . . . . . . . . . . . . . 26
6.1 Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
6.2 Disinfecting . . . . . . . . . . . . . . . . . . . . . . . . . . . .26
6.3 Sterilizing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .27
7. Function check . . . . . . . . . . . . . . . . . . . . . . . . . . . . .27
8. Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
9. Servicing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
10. Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . .29
10.1 General technical data . . . . . . . . . . . . . . . . . . .29
10.2 Weighing data . . . . . . . . . . . . . . . . . . . . . . . . . 29
11. Optional accessories . . . . . . . . . . . . . . . . . . . . . . . 30
12. Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
12.1 Device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .30
12.2 Batteries and rechargeable batteries . . . . . . . . 30
13. Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
14. Declarations of conformity . . . . . . . . . . . . . . . . . . 31
14.1 For Europe . . . . . . . . . . . . . . . . . . . . . . . . . . .31
14.2 For USA and Canada . . . . . . . . . . . . . . . . . . . .31
4 •
17-10-07-651-100a/07-2019S
1. DESCRIPTION OF DEVICE
1.1 Intended use
The seca 777 measuring station is used in accordance with national regula-
tions primarily in hospitals, medical practices and in-patient care facilities.
The
seca 777 measuring station is for conventional determination of weight
and height and for determining the general state of nutrition; it supports the
attending physician in making a diagnosis or deciding on a course of treat-
ment.
To make an accurate diagnosis, however, the physician needs to commission
other specific examinations and take their results into account, in addition to
determining weight and height.
1.2 Description of function
Measuring station On the seca 777 measuring station, weight is recorded by four load cells.
Height is recorded with the integrated measuring rod.
Body mass index (BMI) can be calculated automatically from the measuring
results.
USB interface The USB interface allows the seca 777 measuring station to be connected to
a PC by means of a suitable USB cable (USB 2.0, mini USB type B, max.
cable length: 1.80 m). This makes it possible to process the weight values
displayed using the customer’s own software solution.
1.3 User qualification
Assembly Devices that are shipped partially assembled may only be mounted by suffi-
ciently qualified persons such as specialist dealers, hospital technicians or
seca Service technicians.
Operation The device may only be operated by healthcare professionals.
2. SAFETY INFORMATION
2.1 Safety precautions in these instructions for use
DANGER!
Used to identify an extremely hazardous situation. If you fail to take
note of this information, serious irreversible or fatal injuries will occur.
WARNING!
Used to identify an extremely hazardous situation. If you fail to take
note of this information, serious irreversible or fatal injuries may result.
CAUTION!
Used to identify a hazardous situation. If you fail to take note of this
information, minor to moderate injuries may result.
Safety information • 5
English
17-10-07-651-100a/07-2019S
NOTICE!
Used to identify possible incorrect usage of the device. If you fail to
take note of this information, you may damage the device, or the
measured results may be incorrect.
NOTE
Includes additional information about use of the device.
2.2 Basic safety precautions
Handling the device Please take note of the information in these instructions for use.
Keep the instructions for use in a safe place. The instructions for use are a
component of the device and must be available at all times.
DANGER!
Risk of explosion
Do not use the device in an environment in which one of the following
gases has accumulated:
oxygen
flammable anesthetics
other flammable substances/air mixtures
CAUTION!
Patient hazard, damage to device
Additional devices which are connected to electrical medical
devices must provide evidence of compliance with the relevant IEC
or ISO standards (e.g. IEC 60950 for data-processing devices).
Furthermore, all configurations must comply with the requirements
of standards for medical systems (see IEC 60601-1-1 or Section 16
of edition 3.1 of IEC 60601-1 respectively). Anyone connecting
additional devices to electrical medical devices is considered a
system configurer and is therefore responsible for ensuring that the
system complies with the requirements of standards for systems.
This also applies to additional devices recommended by seca. Your
attention is drawn to the fact that local laws take precedence over
the above-mentioned requirements of standards. In the event of any
queries, please contact your local specialist dealer or Technical
Service.
Have servicing carried out regularly as described in the relevant
section of this document.
Technical modifications may not be made to the device. The device
does not contain any parts for servicing by the user. Only have
servicing and repairs performed by an authorized seca Service
partner. You can find service partners in your area at
www.seca.com or by sending an e-mail to [email protected].
Only use original seca accessories and spare parts, otherwise seca
will not grant any warranty.
CAUTION!
Patient hazard, malfunction
Keep other electrical medical devices, e.g. high-frequency surgical
devices, a minimum distance of approx. 1 meter away to prevent
incorrect measurements or wireless transmission interference.
Keep HF devices such as cell phones a minimum distance of
approx. 1 meter away to prevent incorrect measurements or
wireless transmission interference.
The actual transmission output of HF equipment may require
minimum distances of more than 1 meter. Details can be found at
www.seca.com.
6 •
17-10-07-651-100a/07-2019S
Preventing electric shock WARNING!
Electric shock
Set up the device so that the power supply socket is easy to reach
and the device can be disconnected from the power supply quickly.
Ensure that your local power supply matches the information on the
power supply unit.
Do not touch the power supply unit with wet hands.
Do not use extension cables or power strips.
Make sure that cables are not pinched or damaged by sharp edges.
Make sure that cables do not come into contact with hot objects.
Do not operate the device at an altitude of more than 3000 m above
sea level.
Preventing injuries and infections WARNING!
Injury from falls
Ensure that the device is positioned firmly and level.
Route connecting cables (if present) in such a way that neither user
nor patient can trip over them.
The device is not designed as a standing aid. Assist people with
limited motor skills when they are getting up, e.g. from a wheelchair.
Make sure that the patient does not step onto and off the weighing
platform right at the edges.
Make sure that the patient steps onto and off the weighing platform
slowly and safely.
WARNING!
Danger of slipping
Ensure that the weighing platform is dry before the patient steps
onto it.
Ensure that the patients feet are dry before he or she steps onto the
weighing platform.
Make sure that the patient steps onto and off the weighing platform
slowly and safely.
WARNING!
Risk of infection
Before and after every measurement, wash your hands to reduce
the risk of cross-contamination and nosocomial infections.
Hygienically reprocess the device regularly as described in the
respective section in this document.
Make sure that the patient has no infectious diseases.
Make sure that the patient has no open wounds or infectious skin
alterations, which may come into contact with the device.
Safety information • 7
English
17-10-07-651-100a/07-2019S
Preventing damage to device NOTICE!
Damage to device
Ensure that no liquids enter the device. They can damage the
electronics.
Switch off the device before disconnecting the power supply unit
from the mains socket.
For devices with power supply operation: Disconnect the power
supply unit from the mains socket if you intend to not use the device
for a longer period of time. Only this way it can be ensured that the
device is de-energized.
For devices with battery or rechargeable battery operation: If you do
not use the device for an extended period of time, remove batteries
or rechargeable batteries. Only this way it can be ensured that the
device is de-energized.
Make sure not to drop the device.
Do not expose the device to any impacts or vibrations.
Perform function controls regularly as described in the relevant
section in this document. Do not operate the device if it is damaged
or not working properly.
Ensure that there is no heat source in the immediate vicinity. Do not
expose to direct sunlight. The excessive temperature could damage
the electronics.
Avoid rapid temperature fluctuations. When the device is
transported so that a temperature difference of more than 20 °C
occurs, it must stay turned off for at least 2 hours before it can be
turned on again. Otherwise, condensation water will form which can
damage the electronics.
Use the device only in the intended ambient conditions.
Store the device only in the intended storage conditions.
Do not use aggressive or abrasive cleaning agents.
Do not use organic solvents (e.g. white spirit or petroleum spirit).
Handling measuring results CAUTION!
Patient hazard
In order to avoid misinterpretations, measuring results for medical use
must be displayed and used in SI units (weight: kilogrammes, length:
metres) only. Some devices offer the ability to display measuring
results in other units. This is only an additional function.
Use the results exclusively in SI units.
The use of measuring results in non-SI units is the sole responsibility
of the user.
NOTICE!
Inconsistent measuring results
Before you electronically save measured values determined using
this device and use them further (e.g. in seca PC software or in an
EMR system), make sure that the measured values are plausible.
If measured values are transmitted to seca PC software or an EMR
system, make sure prior to further use that the measured values are
plausible and are assigned to the correct patient.
8 •
17-10-07-651-100a/07-2019S
Handling packaging material WARNING!
Risk of suffocation
Packaging material made of plastic foil (bags) is a choking hazard.
Keep packaging material out of reach of children.
In the event that the original packing material may not be available
anymore, only use plastic bags with security holes in order to reduce
the risk of suffocation. Use recyclable materials if possible.
NOTE
Keep the original packing material for future use (e.g. returning for ser-
vicing).
Handling batteries and
rechargeable batteries
WARNING!
Personal injury as a result of improper handling
Batteries and rechargeable batteries contain harmful substances
which may explode if not handled properly.
Do not try to recharge batteries.
Do not expose (rechargeable) batteries to heat.
Do not burn (rechargeable) batteries.
If acid is leaking out, avoid contact with the skin, eyes and mucous
membranes. Rinse affected areas with plenty of clean water and
seek medical help at once.
NOTICE!
Damage to device and malfunctions with improper handling
Only use the type of (rechargeable) battery specified in this
document.
When replacing (rechargeable) batteries, always replace a complete
set at a time.
Do not short-circuit (rechargeable) batteries.
If you do not use the device for a long period of time, remove the
batteries (incl. rechargeable batteries). This prevents acid from leak-
ing into the device.
If acid leaked into the device, discontinue use. Have the device
checked by an authorised seca Service partner and repaired if
necessary.
Overview • 9
English
17-10-07-651-100a/07-2019S
3. OVERVIEW
3.1 View of device
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
Rear view
16
17
18
19
20
Item Device component Function
1
Display head
For supporting the display unit and housing the cables and batteries
2
Display unit
Serves as the central operating and display element
3
Column
For supporting the display head and the measuring rod
4
Column holder
For supporting the column
5
Power supply
connection socket
For connecting a seca plug-in power supply unit
6
Weighing platform
For weighing patients
7
Battery compartment
To take type AA batteries, 1.5 V
8
Measuring rod
For measuring height
9
Head slide
For sliding the measuring flap onto the lower telescopic element
10
Read-off edge 1
For reading off measuring results in measuring range 1
11
Read-off edge 2
For reading off measuring results in measuring range 2
12
Lower telescopic
element
For measuring height in measuring range 1: Up to 1.22 m
13
Upper telescopic
element
For measuring height in measuring range 2: From 1.22 m
14
Measuring flap
Serves as a head stop for measuring height
15
Latch
For releasing and latching the head slide
10 •
17-10-07-651-100a/07-2019S
3.2 Display unit controls
16
Cover panel
For covering the cables
17
Recessed handle
For transporting the measuring station
18
USB interface
For transmitting the weight values displayed to the customer’s own software
solution (if you have any questions, contact seca Service)
19
Battery cable socket
For connecting the battery cable
20
Connecting cable
socket
For connecting the connecting cable
Item Device component Function
1
2
3
4
Item Control Function
1
On/off key
- Switch the device on and off
2
BMI kg/lbs key
• During the measurement procedure:
- Activate BMI mode (press key briefly)
- Switch between the units kilograms and pounds (press and hold key) (function
depends on variant)
• In BMI mode:
- Confirm height value and display BMI (press key briefly)
- Exit BMI mode (press key briefly again)
3
Tare key
• During the measurement procedure:
- Activate/deactivate
Tare function
• In BMI mode:
- Reduce height value
4
Hold key
• During the measurement procedure:
- Activate/deactivate
Hold function
• In BMI mode:
- Increase height value
Overview • 11
English
17-10-07-651-100a/07-2019S
3.3 Symbols in the display
3.4 Markings on the device and on the type plate
I
D
EFGH
B
A
C
Item Symbol Meaning
A
Batteries are low
B
kg
Weight value in kilograms
C
lbs
Weight value in pounds
D
cm
Height in centimeters
E
BMI
Body mass index
F
NET
Additional weight tared off (Tare)
G
HOLD Weight value is displayed permanently (Hold)
H
Non-verifiable function active
I
Operation with power supply unit
Text/symbol Meaning
Name and address of manufacturer, date of manufacture
Model number
Serial number, consecutive
ProdID
Product identification number, consecutive
Approval Type
Type designation of design approval
Follow instructions for use
Medical electrical device, Type B
d
Value in mass units (non-verified models)
States the difference between two consecutive display values
12 •
17-10-07-651-100a/07-2019S
3.5 Markings on the packaging
Device complies with EU directives
FCC symbol (USA)
FCC ID
For USA: Device license number from the Federal Communications Commission (FCC)
IC
For Canada: Device license number from Industry Canada
Type plate on the power supply connection socket
xx V: Supply voltage required
max xx A: Maximum current consumption
: Note polarity of connector plug
: Operate device with direct current
Do not dispose of device with household waste
Text/symbol Meaning
0123
use compatible
seca adapter only
x-y V
max. xx A
Protect from moisture
Arrows indicate top of product
Transport and store in an upright position
Fragile
Do not throw or drop
Permitted min. and max. temperature for transport and
storage
Permitted min. and max. humidity for transport and
storage
Open packaging here
Packaging material can be disposed of through recycling
programs
Starting up device • 13
English
17-10-07-651-100a/07-2019S
4. STARTING UP DEVICE
4.1 Scope of delivery
4.2 Assembling the device
NOTE
• Use an assistant to perform the assembly.
• We recommend placing the components on the floor and only set-
ting the device in an upright position once you have fitted the mea-
suring rod and the column.
Item Component Pcs.
a
Weighing platform with column holder
1
b
Display head with display unit and connecting cable
1
c
Measuring rod
1
d
Column (with four pre-fitted screws)
1
e
Spacer sleeves
2
f
Slot-head screws
2
g
Cross-head screws
2
h
Lock washers
4
i
T20 hex key
1
j
Batteries (Type AA, 1.5 V)
6
k
Instructions for use
1
Instructions for Use
seca
Measuring Device
a
b
c
d
e
f
g
hi
j
k
14 •
17-10-07-651-100a/07-2019S
Tools required In addition to the scope of delivery, you require the following tools:
Fitting the measuring rod Proceed as follows to fit the measuring rod to the column.
1. Lie the column down on the floor.
2. Place the spacer sleeves on the bores in the column.
3. Extend the upper telescopic element far enough to expose the lower
assembly bore in the lower telescopic element.
4. Position the measuring rod on the spacer sleeves as shown in the
diagram.
5. Place a slot-head screw in the lower assembly bore and the spacer sleeve
underneath it.
6. Tighten the slot-head screw.
7. Extend the upper telescopic element far enough to expose the upper
assembly bore in the lower telescopic element.
Tool Size
Slot-head screwdriver 1x5.5 mm
Cross-head screwdriver PH 2
Starting up device • 15
English
17-10-07-651-100a/07-2019S
8. Place a slot-head screw in the upper assembly bore and the spacer
sleeve underneath it.
9. Tighten the slot-head screw.
10. Push the upper telescopic element into its lowest position.
Fitting the column Proceed as follows to fit the column to the weighing platform.
1. Lie the column on the floor and position the weighing platform as shown
in the diagram.
2. Hold the weighing platform steady.
3. Push the column and measuring rod into the openings in the column
holder of the weighing platform.
The screws on the column protrude from the underside of the weighing
platform.
4. Place one lock washer in each of the bores for the screws.
16 •
17-10-07-651-100a/07-2019S
5. Tighten the screws using the T20 hex key supplied (recommended tight-
ening torque: 3 Nm).
6. Set the measuring station in an upright position.
Fitting the display head Proceed as follows to fit the display head to the column.
1. Extend the upper telescopic element of the measuring rod approx.
10 cm.
2. Guide the connecting cable of the display head through the center
channel of the column as far as the lower part of the device.
3. Put the display head on the column.
4. Push the upper telescopic element into its lowest position.
5. Place one cross-head screw in each of the two side bores of the column.
6. Tighten the cross-head screws.
7. Place the measuring station on the floor with the display unit facing
upward.
Starting up device • 17
English
17-10-07-651-100a/07-2019S
NOTICE!
Damage due to assembly faults
If the cables are bent at too sharp an angle, they may be damaged
and thus cause a functional failure.
Route all cables to prevent sharp bends and kinked connectors.
8. Connect the connecting cable:
a) Draw the connecting cable out of the column
b) Route the connecting cable past the cable clip and insert in the
socket on the weighing platform
c) Fix the connecting cable in position with the cable clip
9. Set the measuring station in an upright position.
Switching the positions of the
display unit and the battery
compartment
At the factory, the display unit is fitted to the right-hand operating side of the
display head. The battery compartment is on the left-hand operating side.
Proceed as follows to switch the positions of the display unit and the battery
compartment.
1. Remove the cover panel:
a) Unscrew two cross-head screws
b) Remove the cover panel
2. Remove the battery compartment lid:
a) Hold the battery compartment lid steady
b) Remove the cross-head screw behind the battery compartment lid
c) Remove the battery compartment lid
Right-hand
operating side
Left-hand
operating side
18 •
17-10-07-651-100a/07-2019S
3. Remove the display unit:
a) Undo the cross-head screw behind the display unit
b) Remove cable tie on the cable hook, if present
c) Draw cables out of the cable hook
d) Lift off display unit
4. Switch the positions of the display unit and the battery holder:
a) Put the battery holder in its new position
b) Carefully put the display unit in its new position
5. Route the cables so that they do not dangle out of the display head.
NOTE
When fitting on the right-hand operating side: You can gather all the
cables into the cable hook and secure them with a cable tie if neces-
sary.
6. Fit the display unit:
a) Hold the display unit steady
b) Tighten the cross-head screw behind the display unit
7. Fit the battery compartment lid:
a) Insert the battery compartment lid and hold it steady
b) Tighten the cross-head screw behind the battery compartment lid
Starting up device • 19
English
17-10-07-651-100a/07-2019S
8. Fit the cover panel:
a) Insert cover panel
b) Place and tighten two cross-head screws
Setting up the device 1. Place the device on a firm, level surface.
CAUTION!
Incorrect measurement as a result of incorrect setup location and
incorrect ambient conditions
Soft floors such as wooden hallways, for example, give under the
patient's weight and falsify the measuring result. Incorrect ambient
conditions lead to measuring faults.
Select a setup location with a level, stable floor in order to achieve
precise measuring results.
Use the device only in the intended ambient conditions
( “Technical data”, page 29).
2. Remove the protective film from the display.
Disinfecting the device Disinfect the following components before initial commissioning
( “Disinfecting”, page 26):
Column and recessed handle
Display head and display unit
Measuring rod, head slide with measuring flap
Weighing platform
4.3 Transporting the device
The device is equipped with two casters to facilitate transport over short
distances.
1. If necessary, remove the plug-in power supply unit from the power socket
and from the device.
2. Fold down the measuring flap and push the upper telescopic element of
the measuring rod into its lowest position.
3. Take hold of the recessed handle of the device.
4. Tilt the device until it can be moved freely on the casters.
5. Move the device slowly and carefully over door sills.
CAUTION!
Incorrect measurement as a result of incorrect setup location and
incorrect ambient conditions
Soft floors such as wooden hallways, for example, give under the
patient's weight and falsify the measuring result. Incorrect ambient
conditions lead to measurement errors.
Select a setup location with a level, stable floor in order to achieve
precise measuring results.
Use the device only in the intended ambient conditions
( “Technical data”, page 29).
6. Transport the device to its new setup location.
7. Set the device in an upright position.
20 •
17-10-07-651-100a/07-2019S
8. If necessary, plug the plug-in power supply unit into the device and into
the power socket ( “Connecting the plug-in power supply unit
(optional)”, page 20).
4.4 Establishing the power supply
Inserting batteries You need six AA-type batteries, 1.5 V. To establish the power supply,
proceed as follows:
1. Open the battery compartment:
a) Hold the battery compartment lid steady
b) Remove the cross-head screw behind the battery compartment lid
c) Remove the battery compartment lid
2. Put the new batteries into the battery holder in the battery compartment.
NOTE
• If replacing batteries: Dispose of the old batteries ( “Batteries and
rechargeable batteries”, page 30).
• Ensure the correct polarity of the batteries (markings on the battery
holder). If
bAtt appears in the display, you have inserted one of the
batteries the wrong way round or the batteries have discharged. If
batteries are inserted the wrong way round, they must be removed
again immediately.
3. Close the battery compartment:
a) Insert the battery compartment lid
b) Tighten the cross-head screw behind the battery compartment lid
Connecting the plug-in power
supply unit (optional)
The seca plug-in power supply unit is available as an optional accessory.
WARNING!
Personal injury or damage to the device as a result of incorrect power
supply units
Conventional power supply units may deliver a higher voltage than is
indicated on them. The measuring station may overheat, catch fire,
melt or short-circuit.
Use only original seca plug-in power supply units with controlled
12 V output voltage.
The connection for the plug-in power supply unit is located on the side of
column holder. To establish the power supply, proceed as follows:
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90

Seca 777 Manuel utilisateur

Taper
Manuel utilisateur

dans d''autres langues