DataSheet
www.tektronix.com14
Trigger Modes
Mode Description
Edge
Positive or negative slope on any channel or front-panel
auxiliary input. Coupling includes DC, AC, HF reject, LF
reject, and noise reject
Glitch Trigger on or reject glitches of positive, negative, or
either polarity. Programmable glitch width is 4 ns
minimum to 8 s maximum
Runt
Trigger on a pulse that crosses one threshold but fails to
cross a second threshold before crossing the rst again
Width
Trigger on width of positive or negative pulse either
within or outside of selectable limits (4 ns to 8 s)
Timeout
Trigger on an event which remains high, low, or either, for
a specied time period (4 ns to 8 s)
Transition
Trigger on pulse edge rates that are faster or slower than
specied. Slope may be positive, negative, or either
Setup/Hold Trigger on violations of both setup time and hold time
between clock and data present on any two input
channels
Pattern
Trigger when any logical pattern of signals goes false
or stays true for specied period of time (4 ns to 1 s).
Pattern (AND, OR, NAND, NOR) specied for all analog
and digital input channels dened as High, Low, or Don’t
Care
State Any logical pattern of analog channels and digital
channels (MSO models) clocked by edge on another
channel. Trigger on rising or falling clock edge
Video
Trigger on all lines, specic line number, odd, even,
or all elds on NTSC, PAL, SECAM, and HDTV
480p/60, 576p/50, 875i/60, 720p/30, 720p/50, 720p/60,
1080/24sF, 1080i/50, 1080p/25, 1080i/60, 1080p/24,
1080p/25, 1080p/50, 1080p/60, Bi-level, Tri-level
Trigger Sequences
Main, Delayed by Time, Delayed by Events. All
sequences can include separate horizontal delay after
the trigger event to position the acquisition window in
time
A/B Sequence Event
Trigger Types
Edge
Trigger Delay by Time 4 ns to 8 s
Trigger Delay by Events 1 to 4,000,000 events
I
2
C (Optional) Trigger on Start, Repeated Start, Stop, Missing ACK,
Address (7 or 10 bit), Data, or Address and Data on I
2
C
buses up to 10 Mb/s
SPI (Optional) Trigger on SS or data on SPI buses up to 10 Mb/s
RS-232/422/485/UART
(Optional)
Trigger on Start Bit, End of Packet, Data, and Parity Error
up to 10 Mb/s
Mode Description
Low-speed: Trigger on Sync, Reset, Suspend, Resume,
End of Packet, Token (Address) Packet, Data Packet,
Handshake Packet, Special Packet, Error.
Token Packet Trigger – Any token type, SOF, OUT,
IN, SETUP; Address can be specied for Any, OUT,
IN, and SETUP token types. Address can be further
specied to trigger on ≤, <, =, >, ≥, != a particular
value, or inside or outside of a range. Frame number
can be specied for SOF token using Binary, Hex,
Unsigned Decimal, and Don’t Care digits.
Data Packet Trigger – Any data type, DATA0, DATA1;
Data can be further specied to trigger on ≤, <, =, >,
≥, != a particular data value, or inside or outside of a
range.
Handshake Packet Trigger – Any handshake type,
ACK, NAK, STALL.
Special Packet Trigger – Any special type, Reserved.
Error Trigger – PID Check, CRC5 or CRC16, Bit
Stufng.
Full-speed: Trigger on Sync, Reset, Suspend, Resume,
End of Packet, Token (Address) Packet, Data Packet,
Handshake Packet, Special Packet, Error.
Token Packet Trigger – Any token type, SOF, OUT,
IN, SETUP; Address can be specied for Any, OUT,
IN, and SETUP token types. Address can be further
specied to trigger on ≤, <, =, >, ≥, != a particular
value, or inside or outside of a range. Frame number
can be specied for SOF token using Binary, Hex,
Unsigned Decimal, and Don’t Care digits.
Data Packet Trigger – Any data type, DATA0, DATA1;
Data can be further specied to trigger on ≤, <, =, >,
≥, != a particular data value, or inside or outside of a
range.
Handshake Packet Trigger – Any handshake type,
ACK, NAK, STALL.
Special Packet Trigger – Any special type, PRE,
Reserved.
Error Trigger – PID Check, CRC5 or CRC16, Bit
Stufng.
High-speed: Trigger on Sync, Reset, Suspend, Resume,
End of Packet, Token (Address) Packet, Data Packet,
Handshake Packet, Special Packet, Error.
Token Packet Trigger – Any token type, SOF, OUT,
IN, SETUP; Address can be specied for Any, OUT,
IN, and SETUP token types. Address can be further
specied to trigger on ≤, <, =, >, ≥, != a particular
value, or inside or outside of a range. Frame number
can be specied for SOF token using Binary, Hex,
Unsigned Decimal, and Don’t Care digits.
Data Packet Trigger – Any data type, DATA0, DATA1,
DATA2, DATAM; Data can be further specied to
trigger on ≤, <, =, >, ≥, != a particular data value, or
inside our outside of a range.
Handshake Packet Trigger – Any handshake type,
ACK, NAK, STALL, NYET.
Special Packet Trigger – Any special type, ERR,
SPLIT, PING, Reserved. SPLIT packet components
that can be specied include:
Hub Address
Start/Complete – Don’t Care, Start (SSPLIT),
Complete (CSPLIT)
Port Address
Start and End bits – Don’t Care,
Control/Bulk/Interrupt (Full-speed Device,
Low-speed Device), Isochronous (Data is Middle,
Data is End, Data is Start, Data is All)
Endpoint Type – Don’t Care, Control, Isochronous,
Bulk, Interrupt
Error Trigger – PID Check, CRC5, CRC16, Any.
USB (Optional)
Note: High-speed support only available on 1 GHz and
2 GHz models.